Best in class MACSQuant® Instrument quick guide How to create an instrument setting Setup of detectors 1.Perform PMT calibration (refer to the MACSQuant Instrument quick guide ‘PMT calibration’ for details). 2.In a new workspace, open a template and acquire some unstained cells 3.In the ‘Channels’ tab in the sidebar adjust voltage of scatter and fluorescent detectors to set cell populations on the desired position in the plot. Select a trigger channel and set an appropriate value. Note: Details on manual detector adjustment can be reviewed in the MACSQuantify™ Software guide Setup of automated compensation Figure 1 : Rack configuration for CompensationMultiColor 1.Prepare single-stained controls representing all fluorochromes to be used in the experimental staining panel. Ensure that there is a comparable positive and negative population for setting compensation. 2.Select the ‘Chill5’ rack from drop-down list in the ‘Experiment’ tab. Note: It is recommended to have one tube that represents a true blank. 3.Click on the ‘Rack’ button in the toolbar. 4.In the ‘Racks’ dialog box, the Chill 5 Rack will be displayed. Select the appropriate number of sample positions to match the number of samples that will be used for compensation. 5.While all wells are selected (orange), click on the ‘Group’ button at the bottom of the window. Each of the selected rack positions should now be labeled with the number ‘1’. Figure 2: Settings for CompensationMultiColor and fluorochrome selection 8.To set the fluorochrome for each rack position, highlight the respective circle in the rack window. Only one circle must be selected (orange highlight) at a time. In the ‘Sample ID’ field a drop-down list will be available from which the representative fluorochrome can be chosen. 6.Choose ‘Express’ from the ‘Settings’ tab within the ‘Experiment’ tab. 9.Each fluorochrome that appears in the drop-down list represents a specific detection channel. For example, PerCP-Vio770™ represents the channel B3, therefore using another B3-compatible fluorochrome, e.g, PE-Cy™5, PerCP-Vio700™ should be selected. 7.Select ‘Setup’ from the ‘Type’ drop-down list and ‘CompensationMultiColor’ from the ‘Mode’ drop-down list. 10.The position ‘Blank’ is selected for a sample of unstained cells. This is the reference for negative cells, which should always be included. Note: Sample should be arranged in columns, not rows. 11.The position ‘PI’ is selected, if propidium iodide (PI) will be used for dead cell exclusion in later experiments. It is used for compensating all fluorescence channels against the channel B3. This ensures that there is no fluorescence spillover into the B3 channel. PI itself will not be compensated. Note: For compensation use unstained cells and do not add PI. 12.Place samples in proper rack positions and start the acquisition. Then follow onscreen instructions. Manual compensation using compensation matrix 1.Choose an analysis template (e.g. four dot plots) by clicking on the ‘New analysis window’ button in the toolbar. 2.In the left dot plot, set up to visualize FSC on the x-axis and SSC on the y-axis. 3.Change one dot plot into a histogram and display the respective channel of the stained particles (e.g. for FITC use B1). 4.In the ‘Experiment’ tab, program all necessary parameters, such as Sample ID and Uptake volume. 5.Open the compensation matrix by clicking the ‘Instrument settings’ button in the toolbar or under ‘Edit > Instrument settings’ and selecting the ‘Compensation’ tab. 6.Click the ‘Matrix’ checkbox. 7.Place the first single-stained tube (e.g. FITC-stained cells) into the single tube holder. 8.Start acquisition. 9.When events start to appear on the plots pause the measurement by right-clicking the stop button in the instrument status bar and choosing ‘Pause’. 10.Draw a region around the population of interest (e.g. lymphocytes) within the FSC versus SSC dot plot. This will be P1. 11.Display only events in P1 in the FITC histogram plot, by selecting ‘P1’ from the drop-down menu of the plot header. 12.Draw two interval regions such that the positive and negative populations are in two separate regions (e.g. P3 for FITC positive and P2 for FITC negative populations). Figure 3: Setup of histograms and regions 13.Display the medians for all spillover channels used in the experiment by clicking on the button, selecting the spillover channels (e.g. B2, B3, etc.) and ‘median’ from the ‘Feature functions’ tab. 14.Resume measurement by clicking on the ‘pause’ button in the instrument status bar. 15.To add compensation to the combination of channels in the plot, choose the appropriate cell in the matrix. The columns represent the fluorochrome being measured, the rows represent the detection channels, where the spillover fluorescence should be corrected. For example, to compensate a FITC-stained sample against the PE channel, go to the cell ‘FITC/B2’ and adjust the value to achieve equal median fluorescence intensities for the positive and negative populations. 16.Increase the value if the median of positive population is higher than the median of negative population. Decrease the value if the median of positive population is lower than the median of negative population. 17.Adjust values for other spillover channels, if necessary. 18.Once compensation is adjusted for this fluorochrome, repeat compensation for all additional fluorochromes. 19.When finished, save as an instrument settings file. Miltenyi Biotec provides products and services worldwide. Visit www.miltenyibiotec.com/local to find your nearest Miltenyi Biotec contact. MACS, the MACS logo, MACSQuant, MACSQuantify, and Vio770 are registered trademarks or trademarks of Miltenyi Biotec GmbH. Cy is a trademark of GE Healthcare companies. Unless otherwise specifically indicated, Miltenyi Biotec products and services are for research use only and not for therapeutic or diagnostic use. Copyright © 2014 Miltenyi Biotec GmbH. All rights reserved. 130-106-092 Are you in need of additional assistance? Visit www.miltenyibiotec.com/locations to find your nearest Miltenyi Biotec contact.