A thorough characterization of the frequency response of various in vivo blood pressure measurement technologies Kroehle, J.P. Jr.; Grenwis, J.; Sarazan, R.D Data Sciences International, (DSI) St. Paul, MN Introduction Figure 2 is an actual rat left ventricular pressure signal that has a high heart rate and a high dP/dtMAX. This time point is where the highest frequency component would exist in an LVP signal. From the FFT plot (Figure 3) it can be seen that the highest frequency content measured is below 100 H Hz, as measured. d Frequency Response 10 0.1 1 10 The question posed at the beginning of the study was: Do DSI devices made specifically for the measurement of left ventricular pressure have the correct characteristics to accurately measure dP/dtMAX and other endpoints from the species they are intended? The answer is, yes. As seen in Figure 10 and Figure 11 below, the Millar sensor and the HD-S21 implantable transmitter correspond overlay the rat LVP signal directly at low and high dP/dtMAX values. For the F th Statham St th PX23L and d th the Ut Utah hM Medical di l sensor, we concluded l d d that these devices are appropriate for measuring systemic arterial pressures but they would not be recommended for the measurement of LVP due to their lower frequency response. 5 0 Geddes Handbook of Blood Pressure Measurement (1991) offers a recommendation that can be used as a starting point for accurately measuring blood pressure. It states that “for an accurate representation of the blood pressure the frequency response needs to be six times higher than the highest heart rate in seconds”. If a dog has a 300 BPM heart rate (5 beats-per-second) then the frequency response should be 30 Hz (5 seconds * 6). This rule of thumb works reasonably well for a systemic blood pressure signal since it more closely represents a sine wave but this is not a good rule for the measurement of LVP due to the more complex morphology of the signal. 100 HD‐S21 (8cm) Utah Medical disposable, PE50 25 cm ‐5 Statham, PE50 25cm It is important that the user select the correct device for the goals of the experiment. Information provided in this report should facilitate such a well informed decision. D70 ‐10 ‐15 ‐20 Frequency (Hz) 200 Figure 4: Bode plot of different pressure sensors Figure 3: FFT of the rat LVP Control Figure 5 contains Fi t i simulated i l t d sine i waves att different diff t rates t collected ll t d by b four f diff different t pressure sensor systems. t A can be As b seen, there th i very little is littl change h i signal in i l amplitude lit d as frequency f changes. h This is true for the external fluid filled type catheter systems as well as the transducer tipped and fully implantable sensors. LVP (mmH Hg) Figure 2: LVP signal of a rat 150 100 50 0 30 DSI HD-S21 Small Animal Telemetry Device with High Frequency Pressure Sensor dP/dt (mmHg/sec) The preferred method to determine what is needed to measure LVP is to record a LVP signal with a high heart rate with a known drug that causes high contractility. The LVP signal needs to be recorded on a high fidelity system and then perform a Fast Fourier Transformation (FFT) on the recorded signal. The FFT results allow the user to determine the highest frequency component of the LVP signal and from that, it can be determined what the frequency response and sample rate of the system needs to be. Conclusions/Perspectives In the Bode plot below (Figure 4), a comparison of sine wave signals measured by the different pressure sensing systems is plotted across a range of frequencies. The Y axis represents signal amplitude and the X axis represents frequency in Hz. Traditionally, the frequency response of a device is defined as the frequency above which the amplitude of the signal is decreased by greater than 3 decibels (db), which is known as the 3 dB point. The data shown in this plot confirm that the D70-PCTP large animal transmitter has >70 Hz frequency response and the HD-S21 small animal transmitter has >100 Hz frequency response. Both devices are more than adequate to measure dP/dtMAX in the appropriate animal model. Amplitude (dB) A question exists; what is needed to accurately measure blood pressure or left ventricle pressure (LVP) to obtain accurate end points such as dP/dtMAX values in animal models? Th answer depends The d d on severall variables i bl such h as ffrequency response off th the pressure sensor, frequency response of the associated electronics and sample rate of the digital acquisition system. All of these variables must be looked at as a system to determine the best method for measuring the pressure signal of interest. Determining Frequency Response of Pressure Sensors To accurately measure end points such as dP/dt, the frequency content of the pressure signal needs to be known. For this a rat left ventricular pressure signal was measured using a system with a bandwidth of 1000 Hz to make sure that the highest frequency component could be measured. The collected pressure signal was then analyzed using an FFT (Fast Fourier Transformation), which is the common method to determine the highest frequency component in a signal. 20 HD-S21 Transmitter mmHg Since Hale first measured blood pressure by using a water manometer in 1733, there have been many improvements in the technology used in laboratory animal models. Throughout history, devices have evolved from mercury manometers to fluid filled catheter systems using a strain gauge device, followed by implanted solid state transducers connected via a tether system. Today there are systems that allow a scientist to measure blood pressure in freely roaming animals by using micro-machined sensors and telemetry. 10 0 1 Hz (60 beats/min) 2 Hz (120 beats/min) 5 Hz (300 beats/min) 10 Hz (600 beats/min) 20 Hz (1200 beats/min) -10 -20 -30 30 Materials and Methods DSI D70-PCTP Large Animal Telemetry Device with High Frequency Pressure Sensor 10000 8000 6000 4000 2000 0 -2000 -4000 -6000 -8000 Time_sec vs LVP_Millar Time_sec vs LVP_DSI-offest 0.0 In order to determine what is needed to accurately measure blood pressure, left ventricle pressure and endpoints such dP/dtMAX, we used a pressure generator that simulates the same conditions as the heart to test pressure sensing systems systems. The pressure generator can output different waveforms at different amplitudes and frequencies. mmHg 20 D70-PCTP Transmitter 0 0.5 1 Hz (60 beats/min) 2 Hz (120 beats/min) 5 Hz (300 beats/min) 10 Hz (600 beats/min) Figure 10: Millar and HD-S21 signal comparison at low contractility -30 Statham Pressure Transducer with 25 cm Fluid Filled PE Catheter 30 200 10 0 1 Hz (60 beats/min) 2 Hz (120 beats/min) 5 Hz (300 beats/min) 10 Hz (600 beats/min) LVP (mmHg) mmHg 20 A Millar catheter (SPC-330A) was used as the reference since it is well established in animal research and there are many publications on the performance of this sensor. Other devices tested included a fluid filled Statham (P23XL) pressure transducer with a 25cm PE50 catheter, a fluid filled disposal Utah Medical (Deltran IV) pressure transducer with a 25cm PE50 catheter, a DSI HD-S21 small animal transmitter and a DSI D70-PCTP large animal transmitter. 20 Hz (1200 beats/min) -10 -20 -30 Millar Catheter-Tipped Pressure Transducer dP/dt (mmHg/sec) mmHg 20 Millar catheter 10 0 Dobutamine (60 μg/kg/hr) 150 100 50 0 30 1 Hz (60 beats/min) 2 Hz (120 beats/min) 5 Hz (300 beats/min) 10 Hz (600 beats/min) 20 Hz (1200 beats/min) -10 -20 -30 Figure 5: Amplitude of different pressure sensors at different rates 10000 8000 6000 4000 2000 0 -2000 -4000 -6000 -8000 Time_sec vs LVP_Millar Time_sec vs LVP_DSI-offest 0.0 Canine dP/dtMAX Comparison p Figure 7 contains the dP/dtMAX values plotted over time during the escalating doses of dobutamine. From the anesthetized rat study, Figure 8 contains the dP/dtMAX values from LVP for both a Millar pressure sensor and an HD-S21 transmitter. dP/dtMAX was increased to approximately 14,000 mmHg/Sec by administering increasing doses of dobutamine. The Millar compared with the HDS21 and shows a very good correlation between the two signals with an R2 value of 0.9463. Figure 9 contains the dP/dtMAX values plotted over time during the escalating doses of dobutamine. Figure 1: DSI HD-S21 small animal transmitter and D70-PCTP large animal transmitter 5sec Averages dP/dtMAX 9000 5sec Averages dP/dtMAX 5sec Averages dP/dtMAX 18000 16000 8000 16000 8000 14000 7000 14000 7000 12000 6000 mmHg/S 5000 4000 5000 4000 I1+dP/dt D70 I6+dP/dt Millar 12000 10000 +dP/dt 6000 Millar y = 0.9725x + 227.03 R² = 0.9463 5sec Averages +dP/dtMAX 18000 9000 HD‐S21 All devices were simultaneously connected to the pressure generator and different signals were applied. Signals were measured with a Ponemah acquisition system sampling at 2000 Hz with no filters. y = 1.0374x ‐ 418.21 R² = 0.9893 8000 10000 8000 Series1 Millar Series2 HD‐S21 3000 3000 6000 6000 2000 2000 4000 1000 2000 0 2000 0 1000 2000 3000 4000 5000 6000 7000 8000 9000 0 0 D70 0 2000 4000 6000 8000 10000 12000 14000 16000 0 Millar Figure 6: Millar compared to D70-PCTP dP/dtMAX values Figure 7: Millar compared to D70-PCTP dP/dtMAX values over time Figure 8: Millar compared HD-S21 dP/dtMAX values 1.0 1.5 Time (seconds) Figure 11: Millar and HD-S21 signal comparison at high contractility References • R.D. Sarazan. Validation of a Digital Data Acquisition and Analysis System in a Cardiovascular in a Toxicology Laboratory. Toxicology Methods Vol.4 No. 4 pp 274-291, 1994 • Main B.W., C.A. Strnat, R. D. Sarazan. Addition of left ventricular pressure telemetry in dogs and monkeys is a powerful tool in safety pharmacology. Safety Pharmacology Society, Annual Meeting. Noordwijk, The Netherlands. September 2003 • Jason Segreti, Andrew Lisowski, James Polakowski, Eric Blomme, Bryan Cox, Andrew King. Simultaneous measurement of arterial and left ventricular pressure in conscious rats by telemetry. Safety Pharmacology Society, Annual Meeting poster 101, September 2010. • Angela Jenkins, Dusty Sarazan. Validation of a Digital Data Acquisition and Analysis System. Safety Pharmacology Society, Annual Meeting, September 2006. 4000 1000 0.5 Millar DSI Rat dP/dtMAX Comparison p From the anesthetized canine study, Figure 6 contains the dP/dtMAX values from LVP for both a Millar pressure sensor and a D70-PCTP transmitter. dP/dtMAX was increased to approximately 8000 mmHg/Sec by administering increasing doses of dobutamine. The Millar compared to the D70-PCTP and shows a very good correlation between the two signals with an R2 value of 0.9893. In each study a Millar catheter tipped transducer was used to compare and correlate data to the transmitters. 1.5 20 Hz (1200 beats/min) -20 Statham After the frequency response of the devices was determined, two animal studies were performed. First an anesthetized rat with an HD-S21 implanted was administered doses of dobutamine to create high values of dP/dtMAX. A second study was done with an anesthetized canine with a D70-PCTP implanted that was administered doses of dobutamine to create high values of dP/dtMAX. 1.0 Time (seconds) Millar DSI 10 -10 Figure 9: Millar compared to HD-S21 dP/dtMAX values over time • Geddes, L.A. “Handbook of Blood Pressure Measurement” Clifton: The Humana Press, 1991.