c TECH LETTER #3 MEASUREMENT OF LINE AND LOAD REGULATION OF DC POWER SUPPLIES HARRISON LABORATORIES DIVISION OF HEWLETT-PACKARD COMPANY (F 100 Locust Avenue Berkeley H e i g h t s , New J e r s e y 0 7 9 2 2 c ---.___-I - -- WWW.HPARCHIVE.COM I TECH LETTER # 3 December 16, 1964 MEASUREMENT OF LINE AND LOAD REGULATICNOF DC POWER SUPPLIES ? I NTRODTJCTIO N T h e measurement of the regulation characteristics of a highly regulated DC power supply involves several subtle procedures and precautions which, i f disregarded, may completely invalidate the measurement r e s u l t s . T h i s tech l e t t e r will present two suggested measurement setups--one for measuring the s t a t i c regulation c h a r a c t e r i s t i c s of constant voltage power s u p p l i e s , the other for the measruement of constant current supplies. THEORY Load regulation for a constant voltage supply is defined a s t h e change of output voltage AVoutwith a change in load c u r r e n t A Iload. Load regulation for a c o n s t a n t current supply is defined a s the change i n output c u r r e n t A I o u t with a change in load voltage A V l o a d . Line regulation is defined a s the change in output voltageAVout ( A Iout for c o n s t a n t current supplies) for a change in l i n e v o l t a g e A V l i n e . Regulation c a n b e expressed either a s a n absolute quantity or as a percentage. Vout or A Iout The DC output r e s i s t a n c e RO c a n e a s i l y be found by using t h e equations: for constant voltage and nvOUt %= 9 AIload PROCEDURE I. Requlation Measurement of Constant Voltage Power Supplies. Figure i shows a s e t u p s u i t a b l e for measuring both load and l i n e regulation of a constant voltage power supply. The following precautions and procedures should b e observed. (1) I t is e s s e n t i a l t h a t the l e a d s connecting t h e load d e v i c e to the power supply and the monitoring device to the power supply be completely s e p a r a t e . This cannot b e overemphasized. Connecting the monitoring d e v i c e to points AA' (see Figure 1) i n s t e a d of the power supply output terminals will r e s u l t i n a measurement not of the regulation characteristics of t h e power supply, but of the power supply plus the r e s i s t a n c e of the l e a d s between the output terminals and points A and A ' . Even connecting the load path by means of c l i p l e a d s to the power supply and then connecting t h e monitoring path by means of c l i p l e a d s t o the load c l i p l e a d s will r e s u l t in a measurement error. Remember that the power supply being measured may have a n output impeda n c e of 1 milliohm or less, and contact r e s i s t a n c e between c l i p l e a d s and power supply terminals may be considerably greater than the specified output impedance of the power supply itself. 3 WWW.HPARCHIVE.COM _q_- c Because this is undoubtedly the m o s t common error which occurs in the measurement of static regulation of a power supply, Figures 2 and 3 have been included to s h o w the proper method of connecting the monitoring and load l e a d s t o the power supply terminals. If the measurement is being made a t the front terminals (Figure 2 ) , the load should be plugged into t h e front of the terminal (at B) while t h e monitoring d e v i c e is connected by means of a l e a d which is inserted through the hole in the neck of t h e binding p o s t s (at A ) . In t h i s way the only r e s i s t a n c e which is common to both the l o a d a n d monitoring paths is located from t h e neck of the binding post back. When the s u p p l y is monitored a t the rear terminal barrier s t r i p (Figure 3 ) , the load device should b e connected t o t h e plus and minus output terminals while the monitoring device should be connected directly to the plus and minus s e n s i n g terminals. S i n c e there is only a s m a l l current flowing i n the l e a d s connecting the sensing terminals to the output terminals, the monitoring device connected to the s e n s i n g terminals obtains a valid indication of output performance. In fact, the monitoring device sees the s a m e performance a s t h e feedback amplifier within the power supply which i s s e n s i n g and correcting t h e output. If a n ammeter is not included on the front panel of the power supply b e i n g (2) measured, then one should b e inserted i n s e r i e s with t h e load r e s i s t a n c e so t h a t t h e full load current which is being imposed on the power supply c a n be verified to b e t h e same as t h e full load current rating of t h e supply. c The load r e s i s t o r should be chosen t o draw the full rated current of t h e (3) supply a t the voltage of measurement. S i n c e the performance of the supply with regard to both load regulation and l i n e regulation will be t o a large degree independent of t h e output voltage s e t t i n g , the output voltage may be adjusted to a v a l u e which, together with the load r e s i s t o r s readily available, will yield the rated v a l u e of load current. The current l i m i t setting of t h e power supply should be set t o its m a x i (4) mum v a l u e , well above the v a l u e of current actually being drawn from t h e supply. Failure to observe t h i s precaution may r e s u l t i n the output voltage dropping a t full load due to current limiting action. The monitoring device to b e used c a n b e a differential voltmeter, or (5) s e n s i t i v e digital voltmeter, o r ' e v e n a battery or variable power supply in s e r i e s with a s e n s i t i v e multimeter or DC oscilloscope. The meter to b e u s e d should b e plugged into a n A C power l i n e which is reasonably isolated from the AC power line feeding t h e power supply, s i n c e large current changes in the power supply may r e s u l t i n l i n e volta g e changes which would c a u s e the reading of the meter to change independent of a n y output change of the power supply. The Variac to be used to v a r y the input AC l i n e voltage should h a v e a n adequate current rating: a l i n e regulator should not be u s e d anywhere i n the l i n e feeding the power supply. J'ailure t o observe t h e s e precautions may r e s u l t i n waveform distortion of the input l i n e voltage t o the power supply with resulting degradation of the regulating performance or, in the case of SCR regulated or preregulated s u p p l i e s , serious malfunctioning of the firing circuit. (6) c - WWW.HPARCHIVE.COM (7) The AC Voltmeter should be known to have a n a c c u r a c y over t h e s p a n from 105-125 volts of 1% or better. Wherever possible t h i s voltmeter should be connected directly to the power supply input terminals rather than to t h e output terminals of the Variac--particularly in t h e case of high current power s u p p l i e s , s i n c e the input voltage a t the power supply terminals may differ significantly from the output voltage of the Variac due to the IR drop in the power l i n e s connecting t h e Variac to t h e power supply. ?. (8) Measurement of l i n e regulation should be accomplished by changing the input l i n e voltage from 105-125 and 125 -105 volts a t the desired output voltage and current combination. (9) Load regulation should b e measured by throwing t h e s w i t c h S-1 and noting the change in t h e output voltage while the supply is being energized with a n y l i n e voltage within its input rating. Notice should b e taken of t h e direction in which the output voltage changes. An output voltage drop concurrent with a n i n c r e a s e i n load current is indicative of a positive output r e s i s t a n c e a t DC, whereas a n output voltage rise concurrent with a n i n c r e a s e in load current is indicative of a negative output r e s i s t a n c e . Measurement of Load and Line Regulation of C o n s t a n t Current Supplies. Figure 4 shows a s u i t a b l e s e t u p for measurement of a constant current supply. Comments Nos. 6 , 7 , and 8 under Measurement of a Constant Voltage Regulation apply to the measurement of constant current regulation. In addition, t h e following precautions should b e followed. 11. (1) The series monitoring r e s i s t o r should have a reasonably s m a l l v a l u e s o that t h e voltage drop a c r o s s it is only a small fraction of t h e total output voltage of t h e power supply. I t should be a wirewound r e s i s t o r using 20 ppm/oC wire (or better) and should b e operated at a power level of less than 1/10 of its rating. This last precaution is n e c e s s a r y so that the surface temperature of the power r e s i s t o r w i l l be only slightly higher than the ambient temperature. Failure to observe t h i s precaution will r e s u l t in a surface temperature "bobble" of the power r e s i s t o r with a corresponding change in the value of RI. This c a u s e s short-term variations i n the measured voltage drop a c r o s s t h i s r e s i s t o r which a r e not the r e s u l t of output current c h a n g e s of the power supply. ") (2) The load r e s i s t a n c e should be c h o s e n so t h a t the voltage drop a c r o s s it a t the current level of measurement is equal to the voltage rating of t h e power supply minus the drop a c r o s s the sensing resistor. (3) Switch S-2 should be placed i n parallel with the load so t h a t the curr e n t regulation of t h e power supply will be effectively measured between a very low voltage (equal to the voltage drop a c r o s s RI) and t h e maximum rated voltage of the supply I I (3) WWW.HPARCHIVE.COM (4) The voltage a d j u s t control should be s e t well above the maximum rated voltage of t h e supply so that voltage limiting will not occur under t h e condition of maximum load r e s i s t a n c e . If t h e voltage control is s e t a t too low a value, voltage limiting may occur with the r e s u l t that the output current will drop under the condition when s w i t c h S-2 is opened. (5) T h e monitoring d e v i c e should be a differential voltmeter, digital voltmeter, DS o s c i l l o s c o p e or other voltage monitoring device capable of measuring accurately the low voltage drop a c r o s s the c h o s e n value of RI. I t is imperative t h a t no voltmeter be connected a c r o s s the terminals BB' (6) of Figure 4. Such a meter will have a constant resistance, and with the change in voltage which will occur a c r o s s the power supply output terminals a s a result of load r e s i s t a n c e change, t h e current drawn by the monitoring device connected between BB' will vary. Even i f the true output current of the power supply were t o remain constant under t h i s condition, t h e small current change through terminals BB' would be compens a t e d by a n equal current c h a n g e through the monitoring resistor RI, with the result that t h e voltage drop a c r o s s t h i s r e s i s t o r would change even though the total power supply current had not changed a t all. It is desirable to u s e a voltmeter i n shunt with the load r e s i s t o r or u s e the front panel voltmeter of t h e power supply itself a s a n indication of t h e output voltage. Special circuitry has been included in Harrison Laboratories c o n s t a n t current s u p p l i e s so t h a t the presence of the front panel voltmeter does not degrade the c o n s t a n t current performance. k I /7 Load regulation should b e measured by throwing the switch S-2 and (7) noting t h e change i n the output voltage as measured a c r o s s RI while the supply is operated with a n y l i n e voltage within its rating. Then: AI Av = RI is used t o calculate the load regulation. n "I WWW.HPARCHIVE.COM 7 c tu c I I WWW.HPARCHIVE.COM - i LOAD LEAD MONITOR HERE FIG 2 l v Proper Method of Connecting Monitoring and Load Lead Connections to Front Panel Terminals. I F-IG 3 Proper Method of Connecting Monitoring and Load Lead Connections to Barrier Strip Terminals. - WWW.HPARCHIVE.COM c c WWW.HPARCHIVE.COM I I _ _ _ _ _ I