Cardiac Rhythm Management // Tachyarrhythmia Therapy // Protego Steroid Eluting ICD Leads Protego Steroid Eluting ICD Leads Technical Manual CAUTION Federal (U.S.A.) law restricts this device to sale by, or on the order of, a physician. © 2014 BIOTRONIK, Inc., all rights reserved. Table of Contents Protego Steroid Eluting ICD Leads Technical Manual Contents 1. General................................................................................................................................................ 1 1.1 Description..................................................................................................................................... 1 1.2 Indications for Use......................................................................................................................... 1 1.3 Contraindications........................................................................................................................... 1 1.4 Protego Lead Names..................................................................................................................... 2 1.5 Warnings........................................................................................................................................ 2 1.6 Precautions.................................................................................................................................... 3 1.6.1 Sterilization, Storage, and Handling����������������������������������������������������������������������������������������� 3 1.6.2 Implantation and Evaluation����������������������������������������������������������������������������������������������������� 3 1.6.3 Pulse Generator Explant and Disposal������������������������������������������������������������������������������������� 4 1.6.4 Hospital and Medical Hazards�������������������������������������������������������������������������������������������������� 5 1.7 Adverse Events.............................................................................................................................. 6 1.7.1 Potential Adverse Events���������������������������������������������������������������������������������������������������������� 6 1.7.2 Observed Adverse Events�������������������������������������������������������������������������������������������������������� 6 1.8 Clinical Studies............................................................................................................................... 7 1.8.1 Protego ICD Leads.................................................................................................................... 7 1.8.2 Kentrox SL‑S Steroid and Kentrox RV‑S Steroid���������������������������������������������������������������������� 7 2. Sterilization and Storage..................................................................................................................11 3. Implant Procedure............................................................................................................................ 13 3.1 Implant Preparation...................................................................................................................... 13 3.2 Product Inspection....................................................................................................................... 13 3.3 Opening the Sterile Container����������������������������������������������������������������������������������������������������� 14 3.4 Protego T and Protego TD Lead Insertion���������������������������������������������������������������������������������� 14 3.5 Active Fixation Lead Insertion����������������������������������������������������������������������������������������������������� 15 3.5.1 Checking the Function of the Fixation Helix before Implantation 16 3.5.2 Accessing the Vein and Inserting the Lead����������������������������������������������������������������������������� 16 3.6 Intraoperative Measurements����������������������������������������������������������������������������������������������������� 19 3.7 PA10 DF4 Adapter....................................................................................................................... 20 3.7.1 Technical description............................................................................................................... 20 3.8 Anchoring the Lead...................................................................................................................... 21 3.9 Connecting the Lead to the Device��������������������������������������������������������������������������������������������� 22 3.10 Final Implant Steps and Defibrillation Testing��������������������������������������������������������������������������� 24 3.11 Testing the Complete Device System���������������������������������������������������������������������������������������� 24 4. Follow-Up Procedures...................................................................................................................... 27 4.1 ICD Patient Follow-Up.................................................................................................................. 27 4.2 Explantation................................................................................................................................. 27 5. Package Contents............................................................................................................................ 29 6. Technical Data (Protego Passive Fixation Leads)������������������������������������������������������������������������ 31 7. Technical Data (Protego Active Fixation Leads)��������������������������������������������������������������������������� 35 8. Disclaimer......................................................................................................................................... 38 PAGE i Table of Contents Protego Steroid Eluting ICD Leads Technical Manual PAGE ii Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual 1. General 1.1 Description The Protego ICD lead is a steroid-eluting lead for use with any ICD with a DF4 connector port according to ISO 27186. The Protego S and Protego T Leads have two pace/sense electrodes (distal tip and ventricular ring electrode) and one defibrillation electrode (ventricular shock coil). The Protego SD and Protego TD Leads have two pace/sense electrodes (distal tip and ventricular ring electrode) and two defibrillation electrodes (ventricular and superior vena cava shock coils These leads, in conjunction with a compatible ICD, perform the following functions: • sense electrical signals from cardiac tissue and conduct those signals to the ICD; • conduct bradycardia and antitachycardia pacing pulses emitted from the ICD to cardiac tissue; • conduct cardioversion/defibrillation shocks of both high and low energies from the ICD to cardiac tissue. The Protego Leads are intended for placement in the right ventricle. The tip and ring electrodes form the most distal portion of the lead and provide dedicated bipolar sensing and pacing. All Protego Leads have one shock electrode that is positioned in the right ventricle (RV). The Protego SD and Protego TD dual-coil ICD leads have an additional shock electrode for placement in the superior vena cava (SVC). All Protego leads feature Silglide® surface treatment, designed to reduce the force required to maneuver the lead during the implant procedure. The Protego S and Protego SD leads feature an electrically active extendable/ retractable fixation helix for use in lead placement. The helix is extended and retracted by rotating the connector pin with a fixation tool. Both the fixation helix and ring electrode are comprised of a platinum/iridium alloy base with fractal iridium. The Protego T and Protego TD leads feature a passive fixation tip coated with fractal iridium. The fractal surface structure on the electrode provides a larger effective tissue interface that is a major factor in determining a lead’s sensing characteristics. All Protego ICD leads include a steroid eluting collar at the distal end (tip electrode), which elutes dexamethasone acetate (DXA) to the surrounding tissue after implantation. Release of the steroid is intended to decrease the inflammatory response at the contact site between the lead tip and the endocardium, thereby decreasing the elevated pacing thresholds of the endocardial lead that often occur after lead implantation. Each Protego lead has a quadripolar DF4 connector according to ISO 27186:2010-03, which connects to the shock coils as well as to the tip and ring electrodes for bipolar sensing and pacing. 1.2 Indications for Use The Protego 8F steroid-eluting, bipolar, IS-1 transvenous lead system is intended for use in the right ventricle of patients for whom implantable cardioverter defibrillators are indicated. 1.3 Contraindications Do not use the Protego Lead System in patients with severe tricuspid valve disease or patients who have a mechanical tricuspid valve implanted. PAGE 1 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual Active Fixation The Protego steroid-eluting leads with active fixation are additionally contraindicated for patients who cannot tolerate a single systemic dose of up to 1.3 mg of dexamethasone acetate (DXA). Passive Fixation The Protego steroid-eluting leads with passive fixation are additionally contraindicated for patients who cannot tolerate a single systemic dose of up to 1.0 mg of dexamethasone acetate (DXA). 1.4 Protego Lead Names The Protego lead names include both the lead type and the overall length of the lead in centimeters. The letter "T" designates passive fixation (tined) and "S" designates active fixation (helix). The suffix "D" denotes leads with dual shock coils. The number following the letter designation indicates the overall length of the lead in centimeters (60, 65 or 75 cm). For dual coil leads, the number after the slash denotes the distance in centimeters between the SVC shock coil and the lead tip. Thus, the name "Protego TD 65/16" refers to a Protego passive fixation, dual coil lead, with a total length of 65 cm and a distance of 16 cm between the SVC shock coil and the lead tip. 1.5 Warnings Adapter Inspection—Always visually check the adapter prior to usage. Do not use wet or damaged adapters. Wet or damaged adapters may result in malfunction of the adapter and pose a risk to the patient. Advancing the Lead—Use of excessive force with a preformed stylet can result in a significant level of pressure being placed on the ventricular myocardium via the lead tip. Defibrillation Threshold—Be aware that changes in the patient’s condition, drug regimen, and other factors may affect the defibrillation threshold (DFT) which may result in nonconversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post‑operatively. Electric Shock—Do not touch the adapter during defibrillation. Touching the adapter during defibrillation may endanger people involved in the procedure. Fixation Tool Rotation—Only rotate the fixation tool as many times as necessary for complete helix extension. Always verify the position of the fixation helix using fluoroscopy. The myocardium can be damaged if the fixation helix is over-rotated. Helix Extension—Do not operate the helix mechanism if it has become sticky due to coagulated blood or bodily fluid or if it has been substantially overwound during retraction or extension. The lead may be too damaged for use. Leakage Currents—Only connect implanted leads to battery-operated measurement and pacing devices or to devices that are classified as type CF (Cardiac Floating) applied parts complying with IEC 60601. Always follow the instructions in the respective technical manuals. All other line-powered devices connected to the patient must be properly grounded. Leakage currents can trigger ventricular fibrillation. MRI (Magnetic Resonance Imaging)—Do not expose a patient to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient. Resuscitation Availability—Do not perform induction testing unless an alternate source of patient defibrillation such as an external defibrillator is readily available. PAGE 2 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual 1.6 Precautions Always observe the technical manuals and accompanying documents for devices combined with this lead (ICD, additional leads) and for devices and implant accessories used during implantation. 1.6.1 Sterilization, Storage, and Handling Lead Packaging—Do not use the lead if the lead’s packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the lead to BIOTRONIK. Resterilization—These leads are intended for single use. Do not resterilize and re-implant explanted leads. Storage (temperature)—Storage at temperatures up to 25° C (77° F); excursions permitted from 5° to 55° C (41° to 131° F). Exposure to temperatures outside this range may result in lead malfunction. Use Before Date—Do not implant the lead after the USE BEFORE DATE. 1.6.2 Implantation and Evaluation Adapter Modification­—Never modify the adapter. Modifications to the adapter may result in loss of function and thereby pose a risk to the patient. Allergic Reaction—Prevent the adapter from contacting the skin or wounds, since in extremely rare cases the adapter material may trigger allergic reactions. Anchor Site—Consider the following when determining where to anchor the lead at the entry site: • Contraction of the heart and other movement of the patient should not put tension on the fixated lead. • The tricuspid valve's function must not be hindered by the lead. Anchoring Sleeve—Always use an anchoring sleeve (lead fixation sleeve) when implanting a lead. Use of the anchoring sleeve, which is provided with the lead, will lessen the possibility of lead dislodgment and protect the lead body from damage by a securing ligature. Blind Plugs—Always close unused ports with blind plugs. Open IS-1 or DF-1 ports can cause undesired current flows to the body. Bodily fluid can penetrate into the device and damage it. Blood Ingress—Coagulated blood can affect the maneuverability of the stylet in the lead and inhibit or block the screw mechanism. • Ensure that no blood reaches the interior of the lead on, or with, the stylet • As far as possible, prevent blood from entering the lead from other pathways Capping Leads—If a lead is abandoned rather than removed, it must be capped to ensure that it is not a pathway for currents to or from the heart. Cross-Threading—To prevent cross-threading the set screw, do not back the set screw completely out of the threaded hole. Leave the screwdriver in the slot of the set screw while the lead is inserted. DFT Testing—For leads with a DF4 connector, the intraoperative DFT test cannot be performed using patient cables. For this purpose, the lead needs to be connected directly to an ICD (see "Testing the complete device system"). Gripping Leads—Do not grip the lead with surgical instruments or use excessive force or surgical instruments to insert a stylet into a lead. Helix Mechanical Function Test—Do not implant the lead if this function test is not successful. In this case, test a replacement lead in the same way, and if the test is successful, implant the replacement lead. PAGE 3 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual Lead Accessories—Please take the following precautions into account to prevent damage to the lead: • Only use the lead with a stylet inserted, even if you only want to check the helix mechanism. • The stylet must not be kinked or overbent. • Only use the provided fixation tool, which is clamped to the connector pin, to extend and retract the fixation helix. Do not use any other tools or accessories. Lead Handling—Make sure that the lead is neither knotted, twisted, nor bent at a sharp angle. This could cause damage to the conductors or result in abrasion of the lead’s insulation. Lead Positioning—If the ICD is implanted underneath the pectoral muscle, ensure that no parts of the lead lie between the ribs and clavicle or between the housing of the ICD and the ribs/clavicle. Chafing and pressure on the lead between the housing of the ICD and the ribs/clavicle could damage the lead’s insulation and thus cause premature failure. o prevent mechanical overstress from causing the failure of pacing/sensing functions, make sure that T the lead does not become pinched between the clavicle and the first rib after implantation. Liquid Immersion—Do not immerse leads in mineral oil, silicone oil, or any other liquid. Over-Tightening—Do not over-tighten the set screw(s). Use only a torque wrench which automatically prevents over-tightening. PA10 DF4 Adapter—Only connect a DF4 connector with an intraoperative test system using a BIOTRONIK PA10 DF4 adapter. Do not contact DF4 connectors directly with alligator clips, or the connector may be damaged. Pacing Interruption—During intracardiac measurements, pacing will be temporarily interrupted. Repositioning or Explanting—Before repositioning or explanting the lead, use fluoroscopy to ensure that the fixation helix has been fully retracted. Sealing System—Failure to properly insert the torque wrench into the perforation at an angle perpendicular to the connector receptacle may result in damage to the sealing system and its selfsealing properties. Set screw(s)—Failure to back off the set screw(s) may result in damage to the lead(s), and/or difficulty connecting the lead(s). Shock Impedance—Never implant a device with a lead system that has a measured shock impedance that is less than what is recommended in the appropriate ICD technical manual. Damage to the device may result. If the shock impedance is less than the recommended value, reposition the lead system to allow a greater distance between the electrodes. Short Circuit—Ensure that none of the lead electrodes are in contact (a short circuit) during delivery of shock therapy as this may cause current to bypass the heart or cause damage to the ICD system. Suitable Stylets—Use only the stylet supplied with the respective lead (based on length and diameter). Never use extremely curved or bent stylets. The use of unsuitable stylets or improper handling of the stylet can result in damage to the lead. Tricuspid Valve Bioprosthesis—Use ventricular transvenous leads with caution in patients with a tricuspid valvular bioprosthesis. 1.6.3 Pulse Generator Explant and Disposal Device Incineration—Never incinerate an ICD due to the potential for explosion. An ICD must be explanted prior to cremation. PAGE 4 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual Explanted Devices—Return all explanted devices to BIOTRONIK. Unwanted Shocks—Prior to explanting the ICD, program the detection status of the device to OFF to prevent unwanted shocks. 1.6.4 Hospital and Medical Hazards Diathermy—Diathermy therapy is not recommended for ICD patients due to possible heating effects at the implant site. If diathermy therapy must be used, it should not be applied in the immediate vicinity of the pulse generator or lead system. Following the procedure, proper ICD function should be checked and monitored. Electrocautery—Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or may cause device malfunction or damage. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pulse generator and leads as possible. Excessive Pressure—Excessive pressure and hyperbaric oxygen therapy should be avoided, as it may cause damage to the implant. External Defibrillation—External defibrillation may result in permanent myocardial damage at the electrode-tissue interface as well as temporarily or permanently elevated pacing thresholds. When possible, observe the following precautions: • Position the adhesive electrodes or defibrillation paddles of the external defibrillator anterior‑posterior or along a line perpendicular to the axis formed by the implanted ICD system and the heart. • Set the energy to a level not higher than is required to achieve defibrillation. • Place the paddles as far as possible away from the implanted device and lead system. • After delivery of an external defibrillation shock, interrogate the ICD to confirm device status and proper function. Lithotripsy—Lithotripsy treatment should be avoided since electrical and/or mechanical interference with the ICD is possible. If this procedure must be used, the greatest possible distance from the point of electrical and mechanical strain should be chosen in order to minimize a potential interference with the implant. Medical Procedures—For any medical procedures that may affect the device (e.g., therapeutic ultrasound, external defibrillation, electrophysiological ablation, HF surgery, lithotripsy), perform a complete follow-up after the procedure. MRI (Magnetic Resonance Imaging)—Do not expose a patient to MRI device scanning. Strong magnetic fields may damage the device and cause injury to the patient. Radio Frequency Ablation—Prior to performing an ablation procedure, deactivate the ICD. Avoid applying ablation energy near the implanted lead system whenever possible. The ICD system should be checked for proper operation after the procedure. Therapeutic Ultrasound—Therapeutic ultrasound is not recommended due to the possible heating effects of the device at the implant site. If therapeutic ultrasound must be considered, it should not be applied in the immediate vicinity of the implant. Transcutaneous Electrical Nerve Stimulation (TENS)—Transcutaneous Electrical Nerve Stimulation must be avoided, as it may lead to unintended heart stimulation. PAGE 5 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual 1.7 Adverse Events 1.7.1 Potential Adverse Events Adverse events (in alphabetical order) associated with ICD systems include, but are not limited to: • • • • • • • • • • • • • • • • • Acceleration of arrhythmias Air embolism Arrhythmias Bleeding Cardiac tamponade Chronic nerve damage Device migration Elevated pacing thresholds Erosion Fluid accumulation Foreign body rejection phenomena Formation of hematomas, cysts or fibrotic tissue Heart valve damage Inappropriate shocks Infection Keloid formation Lead abrasion and discontinuity • • • • • • • • • • • • • • Lead migration/dislodgment Lead fracture/insulation damage Muscle or nerve stimulation Myocardial damage Myopotential sensing Pacemaker mediated tachycardia Pneumothorax Potential mortality due to inability to defibrillate or pace Shunting current or insulating myocardium during defibrillation with internal or external paddles Thromboembolism Undersensing of intrinsic signals Valvular damage Venous occlusion Venous or cardiac perforation Patients susceptible to frequent shocks despite antiarrhythmic medical management may develop psychological intolerance to an ICD system that may include the following: • Dependency • Depression • Fear of premature battery depletion • Fear of shocking while awake • Fear that shocking ability may be lost • Imagined shocking (phantom shock) 1.7.2 Observed Adverse Events Since the Protego ICD leads are based on BIOTRONIK’s legally marketed Kentrox SL Steroid ICD lead, the observed adverse events for this predecessor have been included in this section of the manual. 1.7.2.1 Kentrox The Kentrox RV‑S Steroid and Kentrox SL‑S Steroid ICD leads clinical evaluation included a total of 64 patients implanted with 43 Kentrox RV‑S Steroid leads and 21 Kentrox SL‑S Steroid leads. All adverse events are classified into two types: observations and complications. Observations are defined as clinical events that do not require additional invasive intervention to resolve. Complications are defined as clinical events that require additional invasive intervention to resolve. NOTE: The Kentrox RV‑S Steroid and Kentrox SL‑S Steroid ICD leads are earlier generation of BIOTRONIK devices. The Protego family is based upon the Kentrox family and other BIOTRONIK ICD Leads (i.e., Linox and Kainox families of ICD leads). PAGE 6 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual Table 1 provides a summary of the observed complications with the steroid version of the Kentrox ICD leads. Of the 64 patients, 35 were followed for at least 3 months and only 1 (1.6%) patient underwent a lead revision due to elevated pacing thresholds/loss of ventricular capture in this steroid lead registry. Table 1: Kentrox RV‑S Steroid and Kentrox SL‑S Steroid Lead Related Complication Rate Complication Number of Patients with Complications Percentage of Patients with Complications Number of Complications Steroid (Total Number of Patients = 64, Followed for 3 months = 35) Elevated Ventricular Pacing Threshold 1 1.6% 1 Lead Repositioning 1 1.6% 1 Total Steroid 2 3.2% 2 1.8 Clinical Studies 1.8.1 Protego ICD Leads Since the Protego ICD leads are based on BIOTRONIK’s legally marketed Linoxsmart, Linox and Kentrox SL Steroid ICD leads, engineering tests and an animal study were performed in lieu of human clinical data. 1.8.2 Kentrox SL‑S Steroid and Kentrox RV‑S Steroid BIOTRONIK conducted a prospective registry outside the United States (OUS) of the Kentrox RV‑S Steroid and Kentrox SL‑S Steroid ICD leads. Because no reasonable non-steroid control data from leads with the exact physical specifications was available, no comparison analysis was performed. However, data from this registry is presented in the following sections simply to support the efficacy of the Kentrox RV‑S and Kentrox SL‑S Steroid ICD leads including pacing threshold measurements that are within the normal range for steroid leads. 1.8.2.1 Patients Studied The Kentrox RV‑S and Kentrox SL‑S Steroid ICD leads clinical evaluation included a total of 64 patients implanted with 43 Kentrox RV‑S Steroid leads and 21 Kentrox SL‑S Steroid leads. The study population had a mean age of 65 years (range: 54 to 76 years) and included 59 males (92%) and 5 females (8%). Patients presented with ventricular fibrillation (52%–VF) and ventricular tachycardia (48%–VT). One month follow-up data for 49 patients and three month follow-up data for 35 patients was received and reviewed for this summary. 1.8.2.2 Methods Investigators were required to use the implanted ICD to obtain ventricular lead measurements including intrinsic sensing amplitudes, pacing thresholds and lead impedance values at the implant, predischarge, one-month, three-month, and subsequent routine follow-ups. PAGE 7 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual 1.8.2.3 Results Table 2, Table 3 and Table 4 provide summaries of measured ventricular pacing thresholds, R-waves, and ventricular pacing impedance measurements, respectively. Additionally, Table 2 depicts the ventricular pacing thresholds over time. Both the Kentrox RV‑S Steroid and the Kentrox SL‑S Steroid ICD leads were pooled for presentation of this data because the leads are identical except for the second high energy shock coil embedded in the Kentrox SL‑S Steroid ICD lead. Sixty-one tests with two successful 20 J shocks or less demonstrated the ICD lead system’s ability to convert VF and provide a minimum 10 J safety margin with BIOTRONIK ICDs. Table 2: Ventricular Pacing Threshold Pacing Threshold (Volts @ 0.5ms) Results Implant Number of Tests 64 Mean ± SE 0.5 ± 0.1 Range 0.3 to 1.2 Pre-discharge Follow-up Number of Tests 59 Mean ± SE 0.8 ± 0.1 Range 0.3 to 4.6 One-Month Follow-up Number of Tests 49 Mean ± SE 1.1 ± 0.2 Range 0.3 to 5.2 Three-Month Follow-up Number of Tests PAGE 8 35 Mean ± SE 0.9 ± 0.2 Range 0.3 to 4.6 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual Figure 1: Ventricular Pacing Threshold Table 3: R-Wave Amplitude R-Wave (mV) Results Implant Number of Tests 63 Mean ± SE 12.0 ± 0.4 Range 5.2 to 17.6 Pre-discharge Follow-up Number of Tests 58 Mean ± SE 11.5 ± 0.4 Range 4.1 to 17.5 One-Month Follow-up Number of Tests 49 Mean ± SE 11.7 ± 0.5 Range 6.3 to 17.3 Three-Month Follow-up Number of Tests 35 Mean ± SE 11.7 ± 0.5 Range 6.1 to 17.1 PAGE 9 Chapter 1 General Protego Steroid Eluting ICD Leads Technical Manual Table 4: Ventricular Pacing Impedance Pacing Impedance (Ohms) Results Implant Number of Tests Mean ± SE Range 42 573 ± 19 400 to 1126 Pre-discharge Follow-up Number of Tests Mean ± SE Range 40 468 ± 12 339 to 871 One-Month Follow-up Number of Tests Mean ± SE Range 34 507 ± 13 377 to 829 Three-Month Follow-up Number of Tests Mean ± SE Range 24 525 ± 23 354 to 989 1.8.2.4 Discussion and Conclusion There were a total of 64 patients implanted with 21 Kentrox SL‑S Steroid leads and 43 Kentrox RV‑S Steroid leads. • Two of the 64 patients (3.125%) experienced lead related complications; patient #17 experienced an increase in pacing threshold and had an additional pace/sense lead added; patient #18 experienced a lead repositioning, an expected risk occurrence when using ICD leads, which was successfully revised. • Sixty-one tests with two successful 20 J shocks or less demonstrated the ICD lead system’s ability to convert VF and provide a minimum 10 J safety margin with BIOTRONIK ICDs. • All pacing thresholds, R-wave amplitudes and pacing impedances are within clinically-acceptable normal ranges for single-coil and dual-coil, dedicated bipolar, ICD leads. The data received and analyzed demonstrate that the Kentrox SL‑S Steroid and Kentrox RV‑S Steroid ICD leads are safe and effective for the implanted patients. PAGE 10 Chapter 2 Sterilization and Storage Protego Steroid Eluting ICD Leads Technical Manual 2. Sterilization and Storage This lead is shipped in packaging equipped with a quality control seal and product information label. The label contains the model specifications, technical data, serial number, expiration date, sterilization, and storage information. The lead and its accessories have been sealed in a container and gas sterilized with ethylene oxide. PAGE 11 Chapter 2 Sterilization and Storage Protego Steroid Eluting ICD Leads Technical Manual PAGE 12 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual 3. Implant Procedure Always observe the technical manuals and accompanying documents for devices combined with this lead (ICD, additional leads) and for devices and implant accessories used during implantation. 3.1 Implant Preparation Prior to beginning an implant of the Protego Lead System, ensure that all of the necessary equipment is available. The implant procedure requires the ICD, the lead system, a programmer, other external testing equipment, and the appropriate cabling and accessories. Additional sterile equipment and devices should be available in the event of accidental contamination or damage. CAUTION Capping Leads—If a lead is abandoned rather than removed, it must be capped to ensure that it is not a pathway for currents to or from the heart. Gripping Leads—Do not grip the lead with surgical instruments or use excessive force or surgical instruments to insert a stylet into a lead. Liquid Immersion—Do not immerse leads in mineral oil, silicone oil, or any other liquid. Short Circuit—Ensure that none of the lead electrodes are in contact (a short circuit) during delivery of shock therapy as this may cause current to bypass the heart or cause damage to the ICD system. Tricuspid Valve Bioprosthesis—Use ventricular transvenous leads with caution in patients with a tricuspid valvular bioprosthesis. 3.2 Product Inspection The lead and its accessories are packaged in a double blister and sterilized with ethylene oxide. To assure sterility, inspect the packaging and check for integrity prior to opening. Do not use products if the lead or packaging appears damaged. Should a breach of sterility be suspected, return the lead to BIOTRONIK. CAUTION Lead Packaging—Do not use the lead if the lead’s packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the lead to BIOTRONIK. Storage (temperature)—Storage at temperatures up to 25° C (77° F); excursions permitted from 5° to 55° C (41° to 131° F). Exposure to temperatures outside this range may result in lead malfunction. Use Before Date—Do not implant the lead after the USE BEFORE DATE. Should a replacement lead be required, contact your local BIOTRONIK representative. PAGE 13 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual 3.3 Opening the Sterile Container 1. In a non-sterile area, open the outer blister by peeling the paper seal in the direction of the arrow. 2. In the sterile area, remove the inner, sterile blister by using the gripping the tab. Open the inner blister by peeling the paper seal in the direction of the arrow. 3.4 Protego T and Protego TD Lead Insertion The following procedure is recommended for implanting the Protego T and Protego TD leads. 1. Prior to lead placement, inspect the lead to ensure that the fixation sleeve is positioned close to the junction near the connector portion of the lead. 2. Venous access may be obtained through a standard cut-down or introducer technique. If the introducer method is used, select an appropriate introducer for the lead. A vein lifter is supplied inside the sterile packaging for use, if needed. If using the cephalic cutdown procedure for lead implant, carefully insert the pointed end of the disposable vein lifter into the lumen after opening the vein, and then lift the vein to permit easy lead insertion (see the following figure). PAGE 14 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual NOTE: The recommended introducer size for the Protego T and Protego TD leads is 8 French. 3. The lead should be inserted into the vein and advanced into the ventricle under fluoroscopic guidance. Four stylets are included in the sterile package. Additional stylets are available in separate packaging, if required. Do not shape the stylet while it is inserted in the lead body. When changing stylets, care should be taken to keep the stylet free of blood to help ensure easy insertion and removal. 4. Advance the lead into the right ventricle. The tip should be placed in the apex of the right ventricle. Ensure the lead tip is stable within the ventricle and is in contact with the endocardium. Once adequate lead placement is achieved, the stylet should be carefully withdrawn under fluoroscopy so as not to dislodge the lead. 3.5 Active Fixation Lead Insertion CAUTION Suitable Stylets—Use only the stylet supplied with the respective lead (based on length and diameter). Never use extremely curved or bent stylets. The use of unsuitable stylets or improper handling of the stylet can result in damage to the lead. CAUTION Helix Mechanical Function Test—Do not implant the lead if this function test is not successful. In this case, test a replacement lead in the same way, and if the test is successful, implant the replacement lead. PAGE 15 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual 3.5.1 Checking the Function of the Fixation Helix before Implantation Use the included fixation tool to extend the fixation helix (turn right, clockwise) and retract it (turn left, counterclockwise). When fully extended, the fixation helix protrudes a maximum of 1.8 mm from the lead body. It typically takes 6 to 10 complete rotations to completely extend the fixation helix, however it can take up to 20 rotations. Prior to implantation, test the operation of the fixation helix with inserted stylet. CAUTION Lead Accessories—Please take the following precautions into account to prevent damage to the lead: • Only use the lead with a stylet inserted, even if you only want to check the helix mechanism. • The stylet must not be kinked or overbent. • Only use the provided fixation tool, which is clamped to the connector pin, to extend and retract the fixation helix. Do not use any other tools or accessories. 1. Remove the stylet guide from the lead connector. This leaves it on the part of the stylet that is protruding from the lead. The stylet remains entirely in the lead. 2. Clamp one of the included fixation tools to the connector pin of the lead connector. 3. Turn the fixation tool in clockwise direction until the fixation helix is fully extended. 4. Turn the fixation tool in counterclockwise direction until the fixation helix is fully retracted. 5. Remove the fixation tool from the connector pin and place the stylet guide back on the lead connector. 6. Do not implant the lead if it fails the function test. Instead, use a replacement lead that has passed the same test. 3.5.2 Accessing the Vein and Inserting the Lead 1. 2. 3. 4. Ensure that the fixation helix is completely retracted. Position the premounted lead fixation sleeve close to the lead connector. Ensure that a straight stylet is completely inserted into the lead. If the subclavian stick approach is used to gain venous access, select an appropriate introducer for the lead. CAUTION Suitable Stylets—Use only a suitable stylet for the respective lead (based on length and diameter). Never use extremely curved or bent stylets The use of unsuitable stylets or improper handling of the stylet can result in damage to the lead PAGE 16 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual CAUTION Helix Mechanical Function Test—Do not implant the lead if this function test is not successful. In this case, test a replacement lead in the same way, and if the test is successful, implant the replacement lead. NOTE: The recommended introducer size for the Protego S and Protego SD leads is 8 French. 5. If using the cephalic cutdown approach, open the vein, carefully insert the tip of the provided vein lifter into the lumen of the vein, carefully lift the vein and insert the lead into the opening. See Figure 2. Figure 2: Vein Lifter NOTE: Suitable spare stylets are included in sterile packaging with the lead. They can also be ordered individually as accessories. CAUTION Lead Positioning—Make sure the lead does not become pinched between the clavicle and the first rib after implantation. Securing the Lead Tip Remove the stylet guide from the lead connector. It is now on the end of the stylet that is protruding from the lead. The stylet remains entirely in the lead. Clamp one of the included fixation tools onto the contact pin of the lead connector. Secure the lead tip in the myocardium by keeping the lead in position while rotating the fixation tool clockwise until the helix is fully extended. PAGE 17 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual 6. Advance the lead into the right ventricle under fluoroscopic guidance. WARNING Advancing the Lead—Use of excessive force with a preformed stylet can result in a significant level of pressure being placed on the ventricular myocardium via the lead tip. 7. Use fluoroscopy to verify the position of the fixation helix. Both the helix and the tip are visible under fluoroscopy as noted in the X-rays below. Figure 3: Helix Retracted Figure 4: Helix Extended CAUTION Repositioning or Explanting—Before repositioning or explanting the lead, use fluoroscopy to ensure that the fixation helix has been fully retracted. 8. If, due to repeated extension and retraction of the fixation helix (from repositioning of the lead tip), the helix mechanism becomes difficult to handle or if it sticks, the lead should be removed and replaced with a new one by following these measures: • No longer use the helix mechanism. • Rotate the entire lead with inserted stylet counterclockwise in order to unhelix the lead from the myocardium without using the helix mechanism. WARNING Fixation Tool Rotation—Only rotate the fixation tool as many times as necessary for complete helix extension. Always verify the position of the fixation helix using fluoroscopy. The myocardium can be damaged if the fixation helix is over-rotated. PAGE 18 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual WARNING Helix Extension—Do not operate the helix mechanism if it has become sticky due to coagulated blood or bodily fluid or if it has been substantially overwound during retraction or extension.The lead may be too damaged for use. 3.6 Intraoperative Measurements Standard baseline lead measurements should be performed before connecting the lead to a BIOTRONIK ICD. Medical judgment should be used in cases where optimal lead signals cannot be reliably obtained. If tunneling is required, baseline R-wave and capture threshold measurements are recommended after the tunneling procedure to confirm lead performance and ensure system integrity. Refer to the appropriate BIOTRONIK ICD manual for the configuration of the ports prior to connecting the leads. NOTE: A suitable patient cable has to be used for temporary connection of the lead to an intraoperative test system. The stylet must be removed prior to the measurement. CAUTION PA10 DF4 Adapter—Only connect a DF4 connector with an intraoperative test system using a BIOTRONIK PA10 DF4 adapter. Do not contact DF4 connectors directly with alligator clips, or the connector may be damaged. WARNING Leakage Currents—Only connect implanted leads to battery-operated measurement and pacing devices or to devices that are classified as type CF (Cardiac Floating) applied parts complying with IEC 60601. Always follow the instructions in the respective technical manuals. All other line-powered devices connected to the patient must be properly grounded. Leakage currents can trigger ventricular fibrillation. CAUTION Pacing Interruption—During intracardiac measurements, pacing will be temporarily interrupted. PAGE 19 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual CAUTION DFT Testing—For leads with a DF4 connector, the intraoperative DFT test cannot be performed using patient cables. For this purpose, the lead needs to be connected directly to an ICD (see "Testing the complete device system"). 3.7 PA10 DF4 Adapter 3.7.1 Technical description 1 2 3 1. Drawer for the connector 2. Slide switch 3. Contact surfaces for alligator clips WARNING Electric Shock—Do not touch the adapter during defibrillation Touching the adapter during defibrillation may endanger people involved in the procedure. WARNING Adapter Inspection—Always visually check the adapter prior to usage. Do not use wet or damaged adapters. Wet or damaged adapters may result in malfunction of the adapter and pose a risk to the patient. CAUTION Allergic Reaction—Prevent the adapter from contacting the skin or wounds, since in extremely rare cases the adapter material may trigger allergic reactions. PAGE 20 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual CAUTION Adapter Modification—Never modify the adapter. Modifications to the adapter may result in loss of function and thereby pose a risk to the patient. Instructions for Use 1. Hold the adapter between thumb and index finger. Press the slide switch down slightly and push it forward at the same time. The drawer for the connector will open. 2. Place the DF4 connector of the lead into the drawer for the connector from above. Make sure that the connector is inserted until it stops. 3. Close the drawer for the connector until it snaps in place with an audible click. 4. Verify that the lead is properly seated by ensuring that the blue tip of the lead is visible. 5. Connect the alligator clips to the contact surfaces: • Contact surface + (for 1st ring) • Contact surface - (for connector pin of the DF4 connector) 6. Hold the alligator clips by the protective sleeves, otherwise there is a direct electrical connection to the myocardium. 7. Perform all required measurements. To reposition a lead with active fixation, you need to remove the patient adapter but not the alligator clips. 8. Remove the patient adapter by opening the drawer for the connector. Press the slide switch down slightly and forward at the same time. 9. Pull the DF4 connector upwards to remove it from the drawer for the connector. 3.8 Anchoring the Lead Fixating the lead at the entry site in the vein or in the muscle minimizes the risk of dislodgment. The lead comes with a lead fixation sleeve which has ligature grooves and ligature tabs. The lead fixation sleeve enables secure and smooth fixation of the lead at its entry site and decreases the risk of damaging the insulation or coil during fixation. PAGE 21 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual CAUTION Anchor Site—Consider the following when determining where to anchor the lead at the entry site: • Contraction of the heart and other movement of the patient should not put tension on the fixated lead. • The tricuspid valve's function must not be hindered by the lead CAUTION Anchoring Sleeve—Always use an anchoring sleeve (lead fixation sleeve) when implanting a lead. Use of the anchoring sleeve, which is provided with the lead, will lessen the possibility of lead dislodgment and protect the lead body from damage by a securing ligature. 3.9 Connecting the Lead to the Device In BIOTRONIK ICDs with a DF4 connection, the set screw of the DF4 port is accessible from the labeled side of the housing. WARNING Blind Plugs—Always close unused ports with blind plugs. Open IS-1 or DF-1 ports can cause undesired current flows to the body. Bodily fluid can penetrate into the device and damage it. PAGE 22 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual NOTE: Shock coil(s) and electrodes for ventricular sensing and pacing are commonly connected via the DF4 connection. 1. Remove the stylet and the stylet guide. 2. Using the screwdriver (provided with the active device), pierce the center of the silicone plug vertically, and insert the tip of the screwdriver into the respective set screw. 3. Rotate the set screw counterclockwise with the screwdriver until the connector port of the active device is completely clear. 4. Push the lead connector into the port without bending the conductor or rotating the lead connector until the blue color mark of the DF4 connector becomes visible behind the set screw block. 5. If the lead connector cannot be easily pushed into the connector port, sterile water may be used for lubrication, any other agents are prohibited, however. 6. If the lead connector cannot be inserted completely, the set screw may be protruding into the connector port of the set screw block. Carefully loosen the set screw without completely unscrewing it, so that it does not become tilted upon retightening. 7. Turn the set screw clockwise until the torque control starts (you will hear a clicking sound). • Carefully withdraw the screwdriver without retracting the set screw. • The other contacts of the DF4 port are designed as spring contacts. 8. When the screwdriver is withdrawn, the silicone plug automatically seals the lead connector port safely. . CAUTION Set screw(s)—Failure to back off the set screw(s) may result in damage to the lead(s), and/or difficulty connecting the lead(s). Cross-Threading—To prevent cross-threading the set screw, do not back the Set screw completely out of the threaded hole. Leave the screwdriver in the slot of the set screw while the lead is inserted. Over-Tightening—Do not over-tighten the set screw(s). Use only a torque wrench which automatically prevents over-tightening. Sealing System—Failure to properly insert the torque wrench into the perforation at an angle perpendicular to the connector receptacle may result in damage to the sealing system and its self-sealing properties. PAGE 23 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual 3.10 Final Implant Steps and Defibrillation Testing The ICD and/or lead system may be placed in the pocket at this time. The following steps will help ensure appropriate chronic Protego Lead System function: BIOTRONIK ICD leads are made of highly flexible materials. Depending on where the lead is implanted and the patient’s anatomy, the lead may be longer than required. If this is the case, we recommend placing the excess lead length around the ICD in loose loops (see figure below). CAUTION Lead Handling—Make sure that the lead is neither knotted, twisted, nor bent at a sharp angle. This could cause damage to the conductors or result in abrasion of the lead’s insulation. Lead Positioning—If the ICD is implanted underneath the pectoral muscle, ensure that no parts of the lead lie between the housing of the ICD and the ribs. Chafing and pressure on the lead between the housing of the ICD and the ribs could damage the lead’s insulation and thus cause premature failure. To prevent mechanical overstress from causing the failure of pacing/ sensing functions, make sure that the lead does not become pinched between the clavicle and the first rib after implantation. • Place the device into the pocket with the printed side facing anteriorly. 3.11 Testing the Complete Device System WARNING Resuscitation Availability—Do not perform induction testing unless an alternate source of patient defibrillation such as an external defibrillator is readily available. A final function test of the device system, including defibrillation, must be performed when the ICD has been implanted and the lead has been connected. To do so, communication with the test or programming devices is to be established via the programming head. An external defibrillator must be readily available at all times because the patient is temporarily in a lifethreatening state during this test. PAGE 24 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual CAUTION Shock Impedance—Never implant a device with a lead system that has a measured shock impedance that is less than what is recommended in the appropriate ICD technical manual. Damage to the device may result. If the shock impedance is less than the recommended value, reposition the lead system to allow a greater distance between the electrodes. WARNING Defibrillation Threshold—Be aware that changes in the patient’s condition, drug regimen, and other factors may affect the defibrillation threshold (DFT) which may result in nonconversion of the arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post‑operatively. • Close the device pocket using standard surgical technique. Ensure the ICD detection status has been deactivated prior to using electrocautery. • If a Protego lead is being implanted without an ICD at this time, the lead connector must be capped with an appropriate lead cap and carefully laid in the subcutaneous pocket. Ensure that the lead system is carefully coiled and not twisted. PAGE 25 Chapter 3 Implant Procedure Protego Steroid Eluting ICD Leads Technical Manual PAGE 26 Chapter 4 Follow-Up Procedures Protego Steroid Eluting ICD Leads Technical Manual 4. Follow-Up Procedures 4.1 ICD Patient Follow-Up Follow the instructions described in the appropriate BIOTRONIK ICD technical manual. 4.2 Explantation To explant the Protego S or Protego SD lead, the fixation helix must be fully retracted. An explanted lead may not be reused. Please complete the appropriate explant form and return the form to BIOTRONIK with the explanted lead system. Explanted devices should be sent to BIOTRONIK for analysis and/or disposal. Contact BIOTRONIK if you are in need of assistance with returning any explanted devices. The explanted devices should be cleaned with a sodium hypochlorite solution of at least 1% chlorine rinsed with water prior to shipping. CAUTION Resterilization—These devices are intended for single use. Do not resterilize and re-implant explanted leads. PAGE 27 Chapter 4 Follow-Up Procedures Protego Steroid Eluting ICD Leads Technical Manual PAGE 28 Chapter 5 Package Contents Protgeo Steroid Eluting ICD Leads Technical Manual 5. Package Contents Protego S Lead The Protego S lead and contents of the inner blister package are sterile. Each Protego S Lead system package contains: • 1 Protego S 60, Protego S 65 or Protego S 75 lead with one silicone anchoring sleeve attached and a pre-inserted stylet (0.36 mm) • 1 stylet (0.36 mm) • 2 stylets (0.38 mm) • 1 stylet guide • 2 fixation tools • 1 vein lifter Protego T Lead The Protego T lead and contents of the inner blister package are sterile. Each Protego T Lead system package contains: • 1 Protego T 65 lead with one silicone anchoring sleeve attached and a pre-inserted stylet (0.36 mm) • 1 stylet (0.36 mm) • 2 stylets (0.40 mm) • 1 stylet guide • 1 vein lifter Protego SD Lead The Protego SD lead and contents of the inner blister package are sterile. Each Protego SD Lead system package contains: • 1 Protego SD 60/xx, Protego SD 65/xx or Protego SD 75/xx lead with one silicone anchoring sleeve attached and a pre-inserted stylet (0.36 mm) • 1 stylet (0.36 mm) • 2 stylets (0.38 mm) • 1 stylet guide • 2 fixation tools • 1 vein lifter Protego TD Lead The Protego TD lead and contents of the inner blister package are sterile. Each Protego TD Lead system package contains: • 1 Protego TD 65/xx or Protego TD 75/xx lead with one silicone anchoring sleeve attached and a pre-inserted stylet (0.36 mm) • 1 stylet (0.36 mm) • 2 stylets (0.40 mm) • 1 stylet guide • 1 vein lifter PAGE 29 Chapter 5 Package Contents Protego Steroid Eluting ICD Leads Technical Manual PAGE 30 Chapter 6 Technical Data (Protego Passive Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual 6. Technical Data (Protego Passive Fixation Leads) DF4-LLHH (RV) 1.8 mm², Ir 24.5 mm², Ir 410 mm² 290 mm² 0.75 mg DXA 9 mm 15 mm 160/180 mm Table 5: General Specifications Protego TD Protego T 65, 75 65 Quadripolar Tripolar Connectors 1 x DF4 1 x DF4 Connector Pin Material MP35N MP35N Pt/Ir Pt/Ir Passive fixation, with four silicone tines Passive fixation, with four silicone tines 2.6 mm (7.8 F) 2.6 mm (7.8 F) 8F 8F Leads, Length (cm) Polarity Connector Ring Contact Material Fixation Type Lead Diameter Diameter of Recommended Introducer Table 6: Tip Electrode Specifications - Linoxsmart and Protego Leads Protego TD Protego T Passive fixation with 4 silicone tines Passive fixation with 4 silicone tines 1.8 mm² 1.8 mm² 90% Pt / 10% Ir 90% Pt / 10% Ir Iridium, fractal Iridium, fractal Coil Coil Conductor Material MP35N MP35N Outer Diameter of Inner Conductor Coil 0.70 mm 0.70 mm 4 4 ≤ 85 Ω (65 cm length) ≤ 100 Ω (75 cm length) ≤ 85 Ω (65 cm length) Fixation Design Electrically Active Surface Area Base Material Surface Material, Structure Conductor Style Number of Filaments Conduction Resistance PAGE 31 Chapter 6 Technical Data (Protego Passive Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual Table 7: Steroid Collar Active Pharmaceutical Agent (API) DXA Content Steroid Carrier Protego TD Protego T DXA DXA 0.75 mg 0.75 mg Silastic 7-6860 Silastic® 7-6860 ® Table 8: Ring Electrode Protego TD Protego T Surface Area 24.5 mm Base Material 80% Pt / 20% Ir 80% Pt / 20% Ir Iridium, fractal Iridium, fractal Tip to Ring Distance 9 mm 9 mm Conductor Style Cable Cable Conduction Material MP35N MP35N Number of Filaments 7*7 (cable) 7*7 (cable) ≤ 45 Ω (65 cm length) ≤ 50 Ω (75 cm length) ≤ 45 Ω (65 cm length) Surface Material, Structure Conduction Resistance 2 24.5 mm2 Table 9: RV Coil Protego TD Protego T 2.9 cm2 2.9 cm2 2.6 mm (7.8F) 2.6 mm (7.8F) 5.0 cm 5.0 cm 90% Pt / 10% Ir with Tantalum core 90% Pt / 10% Ir with Tantalum core Distance From Lead Tip 15 mm 15 mm Conductor Style Cable Cable MP35N/DFT (41% Ag) MP35N/DFT (41% Ag) 7*7 7*7 ≤ 1.6 Ω (65 cm length) ≤ 1.9 Ω (75 cm length) ≤ 1.6 Ω (65 cm length) Geometric Surface Area Diameter Length Base Material Conduction Material Number Of Filaments Conduction Resistance * MP35N®, MP and Multiphase are registered trademarks of the Standard Pressed Steel Co. PAGE 32 Chapter 6 Technical Data (Protego Passive Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual Table 10: SVC Coil Protego TD Geometric Surface Diameter Length Base Material Distance From Lead Tip Conductor Style Conduction Material Number Of Filaments Conduction Resistance Protego T 4.1 cm 4.1 cm2 2.6 mm (7.8F) 2.6 mm (7.8F) 7.0 cm 7.0 cm 90% Pt / 10% Ir with Tantalum core 90% Pt / 10% Ir with Tantalum core 16, 18 cm 16, 18 cm Cable Cable MP35N/DFT (41% Ag) MP35N/DFT (41% Ag) 7*7 7*7 ≤ 1.2 Ω (65 cm length) ≤ 1.4 Ω (75 cm length) ≤ 1.2 Ω (65 cm length) ≤ 1.4 Ω (75 cm length) 2 PAGE 33 Chapter 6 Technical Data (Protego Passive Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual PAGE 34 Chapter 7 Technical Data (Protego Active Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual 7. Technical Data (Protego Active Fixation Leads) DF4-LLHH (RV) 4.5 mm², Ir 24.5 mm², Ir 290 mm² 410 mm² 1 mg DXA 1.8 mm 11 mm 17 mm 160 /180 mm Table 11: General Specifications Protego SD Protego S 61, 65, 75 61, 65, 75 Quadripolar Tripolar Connectors 1 x DF4 1 x DF4 Connector Pin Material MP35N MP35N Pt/Ir Pt/Ir Active Fixation Helix Active Fixation Helix 2.6 mm (7.8F) 2.6 mm (7.8F) 8 F 8 F Leads, Length (cm) Polarity Connector Ring Contact Material Fixation Type Lead Diameter Diameter of Recommended Introducer Table 12: Steroid Collar Specifications - Linoxsmart and Protego Leads Parameter Active Pharmaceutical Agent (API) DXA content Steroid Carrier Protego SD Protego S DXA DXA 1.0 mg 1.0 mg Silastic 7-6860 Silastic® 7-6860 ® PAGE 35 Chapter 7 Technical Data (Protego Active Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual Table 13: Tip Electrode Parameter Protego SD Protego S Electrically active, extendable / retractable helix Electrically active, extendable / retractable helix Maximum Number of Turns for Helix Extension 20 20 Maximum Length of Extended Helix 1.8 mm 1.8 mm Fixation Helix: Wire Diameter 0.25 mm 0.25 mm Electrically Active Surface Area 4.5 mm2 4.5 mm2 70% Pt / 30% Ir 70% Pt / 30% Ir Iridium, fractal Iridium, fractal Coil Coil Conductor Material MP35N MP35N Outer Diameter of Inner Conductor Coil 0.70 mm 0.70 mm 4 4 ≤ 50 Ω (61, 65 cm length) ≤ 60 Ω (75 cm length) ≤ 50 Ω (61, 65 cm length) ≤ 60 Ω (75 cm length) Fixation Design Base Material Surface Material, Structure Conductor Style Number Of Filaments Conduction Resistance Table 14: Ring Electrode Parameter Protego SD Protego S Surface Area 24.5 mm² 24.5 mm² Base Material 80% Pt / 20% Ir 80% Pt / 20% Ir Iridium, fractal Iridium, fractal Tip to Ring Distance 11 mm 11 mm Conductor Style Cable Cable MP35N MP35N 7*7(cable) 7*7(cable) ≤ 45 Ω (61, 65 cm length) ≤ 50 Ω (75 cm length) ≤ 45 Ω (61, 65 cm length) ≤ 50 Ω (75 cm length) Surface Material, Structure Conductor Material Number of Filaments Conduction Resistance * MP35N®, MP and Multiphase are registered trademarks of the Standard Pressed Steel Co. PAGE 36 Chapter 7 Technical Data (Protego Active Fixation Leads) Protego Steroid Eluting ICD Leads Technical Manual Table 15: RV Coil Parameter Protego SD Protego S 2.9 cm² 2.9 cm² 2.6 mm (7.8 Fr) 2.6 mm (7.8 Fr) 5.0 cm 5.0 cm 90% Pt / 10% Ir with Tantalum core 90% Pt / 10% Ir with Tantalum core Distance From Lead Tip 17 mm 17 mm Conductor Style Cable Cable MP35N/DFT (41% Ag) MP35N/DFT (41% Ag) 7*7 7*7 ≤ 1.6 Ω (61, 65 cm length) ≤ 1.9 Ω (75 cm length) ≤ 1.6 Ω (61, 65 cm length) ≤ 1.9 Ω (75 cm length) Geometric Surface Area Diameter Length Base Material Conductor Material Number of Filaments Conduction Resistance Table 16: SVC Coil Specifications - Linoxsmart and Protego Leads Parameter Geometric Surface Area Diameter Length Base Material Distance From Lead Tip Conductor Style Conductor Material Number of Filaments Conduction Resistance Protego SD Protego S 4.1 cm² N/A 2.6 mm (7.8 Fr) N/A 7.0 cm N/A 90% Pt / 10% Ir with Tantalum core N/A 16, 18 cm N/A Cable N/A MP35N/DFT (41% Ag) N/A 7*7 N/A ≤ 1.2 Ω (61, 65 cm length) ≤ 1.4 Ω (75 cm length) N/A PAGE 37 Chapter 8 Disclaimer Protego Steroid Eluting ICD Leads Technical Manual 8. Disclaimer BIOTRONIK leads, lead extensions, adapters and accessories used in connection with these devices (referred to as: leads and accessories) have been qualified, manufactured and tested in accordance with well proven and accepted standards and procedures. The physician should be aware, however, that leads and accessories may be easily damaged by improper handling or use. Except as set forth in BIOTRONIK’s Lead Limited Warranty, BIOTRONIK makes no express or implied warranties for its leads and accessories. PAGE 38 Protego Steroid Eluting ICD Leads Technical Manual M4158-A 05/14 ©2014 BIOTRONIK, Inc. All rights reserved. MN025 5/20/14 BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035-5369 (800) 547-0394 (24-hour) (800) 291-0470 (fax) www.biotronik.com