Hardware Pace using Slope Detection

advertisement
Tony Calabria
TI Precision Designs: Verified Design
Hardware Pace using Slope Detection
TI Precision Designs
Circuit Description
TI Precision Designs are analog solutions created by
TI’s analog experts. Verified Designs offer the theory,
component selection, simulation, complete PCB
schematic & layout, bill of materials, and measured
performance of useful circuits. Circuit modifications
that help to meet alternate design goals are also
discussed.
This hardware pace detection circuit is designed to
monitor for occurrences of a pace maker signal in an
electrocardiogram (ECG) application by providing an
alert to a GPIO. The circuit combines three individual
circuits: a differentiator circuit used for slope
detection, a window comparator to monitor for an
event and SR latch to indicate an event has occurred.
A pacemaker signal at the input will latch a digital I/O
pin high to indicate the signal is present in the
waveform.
Design Resources
Design Archive
TINA-TITM
ADS1298
OPA348
TLV3401
SN74LVC2G00DCTR
Ask The Analog Experts
WEBENCH® Design Center
TI Precision Designs Library
All Design files
SPICE Simulator
Product Folder
Product Folder
Product Folder
Product Folder
Vcc
Vcc
R7
R5
V1
C2
R1
+
R6
R2
C1
+
Alert Out
R11
ECG + Pace
+
–
Vbias
Vcc
R3
+
V2
R4
Reset
An IMPORTANT NOTICE at the end of this TI reference design addresses authorized use, intellectual property matters and
other important disclaimers and information.
TINA-TI is a trademark of Texas Instruments
WEBENCH is a registered trademark of Texas Instruments
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
1
www.ti.com
1
Design Summary
The design requirements are as follows:

Supply Voltage: 5V dc

Minimum pace signal width: 100µs

Minimum pace signal amplitude: 2mV

Monitor for slope of pacemaker signal while ignoring ECG signal

Set GPIO to indicate pacemaker signal is present
The design functionality is displayed below showing a pulse when the slope of the pace maker signal is
present. The Differentiator Out signal triggers a falling edge on the window comparator thereby setting the
SR latch.
ECG + Pace Input
Differentiator Out
Window
Comparator Out
SR Latch Out
PACE: 20mV Amplitude
100us period
Figure 1: Hardware Pace Results
2
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
2
Theory of Operation
To understand the design approach discussed in this TI Design note, basic pacemaker functionality must
first be discussed. The primary role of the pacemaker is to apply a small surge of voltage to the heart, prior
to the QRS pulse, to assist with blood flow through the body. The signal can take on a variety of form
factors which can be broken down to a specific amplitude and signal width. Figure 2 is used to model a
typical QRS pulse with an example of a Ventricle pacemaker signal shown prior.
QRS
Complex
PR
Interval
R
ST
Segment
T
P
Pace
Q
U
S
QT
Interval
Figure 2: ECG Signal with Pace
The height and width of the pacemaker signal can range anywhere from 2mV to over 500mV in amplitude
and 0.1ms to 500ms in width while possibly sharing similar signal characteristics of the ECG pulse. The
shape of the signal will differ depending on how the pacemaker is positioned to assist the heart and the
type of pacemaker installed. A common characteristic of pacing pulses is to have a very steep rising edge
transition which can be used to distinguish from the QRS complex. Using the higher frequency
components associated with this steep rising edge can determine if a pacemaker is present. Taking a Fast
Fourier Transfer (FFT) of an ECG signal with a pacemaker, the characteristics show that the ECG
components lie within the 150Hz and lower range while the pace signal resides in the higher bandwidths,
at times 1kHz and greater.
Determining if the pace signal is present in an ECG system can be done in a variety of ways. Wide
bandwidth ADCs are commonly used to digitize the ECG signal, where post processing can determine if a
pace signal appears before the QT interval shown in Figure 2. There are also hardware methods which
can be used focused around analog design to determine if the pacemaker signal is present.
This TI Design is a hardware pace detection method designed around using slope detection. A complete
schematic for this design is shown in Figure 3. The circuit is designed to be specifically used with the
ADS1298 which includes an internal differential to single ended converter for the ECG lead. Once routed
from the ADC, a differentiator circuit is used for slope detection of the pacemaker signal. A reference
voltage, Vbias, is used to add DC bias to the window comparator circuit to allow the circuit to be powered off
of a single supply. Resistors set the threshold values for the window comparator prior to the latch stage to
indicate an event. The design can be segmented into three stages: Slope Detection, Alert Stage, and
Latch to GPIO.
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
3
www.ti.com
Vcc
Vcc
R7
R5
V1
C2
R1
+
R6
R2
C1
+
Alert Out
R11
ECG + Pace
+
–
Vcc
R3
Vbias
+
V2
R4
Reset
Figure 3: Complete Circuit Schematic
2.1
Slope Detection
The first stage monitors for the leading edge of the pace signal while attenuating the QRS complex. A
differentiator circuit is selected to monitor for the steep leading edge of the pacemaker signal. This circuit
takes advantage of the characteristic of current across a capacitor and uses the op amp feedback resistor
to develop a proportional voltage. The output from the circuit will reflect a signal relative to the slope, dV/dt,
of the input, shown in Figure 4.
R1
Vout
C1
+
Vin
+
–
Vout   R1C1
dV
 Vbias
dt
Vbias
Figure 4: Typical Differentiator Circuit
The resistor and capacitor used in this circuit form a high pass filter which is set to monitor for the higher
frequency components from the pacemaker signal. Thus, the output will be proportional to the slope
amplitude and duration while ignoring the QT interval and being offset by the Vbias voltage.
4
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
2.1.1
Circuit Compensation
The passive components will set the poles and zeroes for the transfer function and must be selected to
ensure stability. A second capacitor, C2, may be placed in parallel with the resistor in the feedback loop to
help stabilize the circuit, along with a second resistor, R 2, in the input path. The procedure for selecting the
stability network is beyond the scope of this design note; please see reference [1] for a detailed
explanation of stability.
C2
R1
R2
Vout
C1
+
Vin
+
–
R11
Vbias
Figure 5: Differentiator Circuit with Compensation
In order to ensure 40dB gain before compensating for stability, the first order high pass filter for the circuit
must begin at around 1Hz. C1 is chosen to be 1µF to set the HPF cut off of the differentiator circuit to
attenuate the unwanted QRS complex. Setting R1 to 392kΩ and C2 to 10nF help maximize the gain while
keeping the circuit stable. R2 is short circuited to 0Ω but included for design flexibility. R11 is in place as a
1MΩ load resistor for the output. The Vbias voltage is set to 2.5V, the midpoint of the power supplies, to
allow the differentiator output maximum swing in either the positive or negative direction allowing design
flexibility if the pace maker pulse is of opposite polarity. The Vbias voltage is generated by dividing the 5V
supply using two 2MΩ resistors to help conserve power. As mentioned in the modifications section later, a
second order high pass filter would help improve gain before compensating for stability.
2.2
Alert Stage
The output of the differentiator circuit, discussed above, will produce a pulse when a pace event occurs
which must then be recognized to register an alert. By design, this pulse from the output of the
differentiator can range anywhere from a few hundred mVs to a couple volts depending on the pacemaker
signal characteristics. Capturing the signal is done using a window comparator circuit designed to output a
logic low signal when the comparator input pulse exceeds either a high or low threshold voltage. The open
drain output window comparator circuit used in the design is shown in Figure 6.
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
5
www.ti.com
Vcc
Vcc
 R6 
V1  Vcc 

 R5  R6 
 R4 
V 2  Vcc 

 R3  R 4 
R7
R5
V1
+
R6
Vin
*From Differentiator Vout
Comparator Vout
Vcc
R3
+
V2
R4
Figure 6: Window Comparator Circuit
The threshold limits, listed as V1 and V2 in Figure 6, are set as a fraction of the power supply determined
by the R3, R4, R5, R6 resistor values. When the output from the differentiator circuit, listed as V in, exceeds
the set limits V1 or V2, the comparator output will trigger by pulling the comparator Vout line low.
Selection of component values is not as straightforward as it is for the differentiator circuit. The boundary
limits for the window comparator are dependent on the amplitude of the differentiator amplifier output
under the worst case condition, 2mV amplitude and 100µs width pacemaker pulse. Through simulation
(shown later) we can estimate a value for the differentiator output amplitude to set the threshold limits of
the window comparator, and then make modifications in the verification process. The values chosen are
setting V1 to 2.77V and V2 to 2.48V, requiring R3 set to 10.2kΩ, R4 set to 10kΩ, R5 set to 8.06kΩ, and R6
set to 10kΩ. R7 is a weak pull up resistor for the open drain output and will be set to 10kΩ. Vcc is the 5V
supply used to power the system.
2.3
Latch to GPIO
The width of the comparator output pulse is determined by the time that Vin (from Figure 6) exceeds the
window comparator boundaries. This duration can be as small as a few milliseconds and may be missed if
not latched. Placing a SR-latch circuit after the window comparator will latch the signal in a steady state to
be read back by a microcontroller or DSP until a reset is performed. The SR latch will perform two
functions: latching the window comparator out signal, and serving as an inverter to create an active high
alert out signal. The design of the SR-latch using two NAND gates is shown in Figure 7.
6
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
Comparator Vout
Alert Out
Vcc
R8
Reset
S1
Figure 7: SR Latch
The alert out signal can be routed to a GPIO and the state monitored for a pace event. The R 8 resistor and
S1 switch are used to reset the latch. R8 is a weak pull up resistor and will be set to 10kΩ, and S1 is a
single pull, single throw (SPST) push button switch.
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
7
www.ti.com
3
Component Selection
3.1
Differentiator Circuit Amplifier Selection
The slope detection circuit has a few requirements on the amplifier selection. The op amp must have
adequate bandwidth to respond to the pace signal change on the input before compensation. Earlier it was
specified that the minimum pace signal we would like to detect is 100µs in width, therefore, an amplifier of,
at least, a decade greater bandwidth is preferred. Depending on the duration of the pace pulse slope, the
output may swing close to the power rails, requiring a rail to rail out op amp. The OPA348 was chosen
because it has 1MHz bandwidth, is capable of reaching within 25mV of the rail, and maintains adequate
bandwidth, while staying cost effective in the TI value line of devices. A wider bandwidth op amp can be
selected to provide more flexibility on the high pass filter design prior to compensation for stability.
3.2
Differentiator Vbias Voltage Selection
The role of the Vbias voltage is to set the dc operating point to the window comparator input when a pace
pulse is not detected. This becomes the idle voltage state of the window comparator input used to set the
high and low threshold values of the individual comparators. By setting Vbias to 2.5V, the midpoint of the
system power supplies, the swing of individual comparators is maximized in either the positive or negative
direction. Adjusting the Vbias value would require that the window comparator threshold voltages are
properly set.
3.3
Comparator Device Selection
The window comparator circuit design requires two comparators sharing a common output that each has
the ability to control. Open drain output comparators allow for either device to pull the line low when the
window comparator input voltage exceeds set voltage values. The TLV3401 was chosen for its open drain
output and small package option. The TLV3402 is a secondary option as it includes two devices per
package.
3.4
NAND Gates for SR Latch Selection
Two NAND gates are required for the SR latch design with adequate propagation delay. The
SN74LVC2G00 was chosen for its two devices per package and tpd max of under 5ns. A simple push
button switch will work for the reset.
3.5
Passive Component Selection
The differentiator circuit design and window comparator threshold design are the two paths for which the
passive component selection values are critical for this design. To meet the design goal of the
differentiator circuit responding to the pace signal, the resistor tolerance values of the passive components
setting the high pass filter were chosen to be 1%.
Resistors R3, R4, R5, and R6 in Figure 6 set the threshold values for the window comparator. These
threshold voltage values may be set very close to the dc operating point set in the differentiator stage, in
order to meet the specification of responding to a 2mV amplitude, 100µs period pace signal. Therefore,
resistor tolerances of 0.1% were chosen to set the voltages as accurate as possible. When this was not
possible due to reasonable cost or availability, the tolerance of the resistor was chosen to be 1%. The
TM
values, themselves, were obtained experimentally during TINA-TI simulation and then modified based
on lab testing.
Other passive components in this design may be selected for 1% or greater as they will not directly affect
the transfer function of this design.
8
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
4
Simulation
TM
The TINA-TI schematic shown in Figure 8 includes the circuit values obtained in the design process. The
first OPA348 in a unity gain buffer configuration is used to model the internal buffer of the ADS1298 prior
to the V_PACE_OUT pin.
VCC
ECG+PACE
R6 10k
+
ECGp
+
V_PACE_OUT
To SR Latch
-
U3 TLV3401
-
OPA348
VPDetect
U1 OPA348
+
VCC
VCC
+
VCC
VCC
Vref
V1 2.5
+
U2 TLV3401
+
-
R8 10k
Vref
VCC
R7 10.2k
+
R2 0
+
+
C1 1u
-
+
R1 1M
Vpace Pos
R4 392k
VECG_block
VCC
R5 8.06k
C2 10n
R16 100k
Buffer internal to ADS1298
V2 2.5
TM
Figure 8: TINA-TI
4.1
Schematic
Time Domain Analysis
The transient analysis is first performed to get an understanding of the events at each stage of the circuit.
ECGp and Vpace Pos are arbitrary waveforms used to model a typical ECG pulse with a 2mV, 100µs pace
signal beforehand, similar to what is shown in Figure 2. VPDetect is the output from the differentiator which
is used to get an understanding of where to set the voltage threshold levels for the window comparator.
V_PACE_OUT is the output from the window comparator which will get pulled low when VPDetect appears
outside of the set boundaries. The time domain response is shown in Figure 9.
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
9
www.ti.com
TM
Figure 9: TINA-TI
- Transient Response
The simulation results in Figure 9 are used to make any adjustments to the window comparator levels if a
latch does not take place. VPDetect reaches a value of 2.45V when the target of a 2mV, 100µs pace
signal is applied to the input. Giving a little room for error, a level of 2.48V is used for the lower limit.
10
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
4.2
Step Response
The time domain step response of the differentiator stage can be seen in Figure 10 to verify stability.
T
779.23m
Differentiator Out
1.99m
10.00m
Vin
0.00
0.00
100.00m
200.00m
300.00m
Time (s)
TM
Figure 10: TINA-TI
- Step Response
At the moment of the step input, there is a large spike on the Differentiator Out (VPDetect) relative to the
slope before stabilizing to a steady DC value. This is the expected behavior of the differentiator circuit
without ringing or oscillations verifying that we have a stable design.
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
11
www.ti.com
5
PCB Design
The PCB schematic and bill of materials can be found in Appendix A.
5.1
PCB Layout
The layout of the ECG hardware pace board was designed to pair with the ADS1298ECG-FE board
requiring minor modifications to the connecting headers. General layout practices based on the general
PCB Layout Guidelines document found on the e2e Wiki were followed. The PCB layout for this design is
shown in Figure 11.
Figure 11: PCB Layout
12
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
6
Verification & Measured Performance
6.1
Time Domain Analysis
This circuit is required to pass functionality tests comparing the real-world circuit performance to
TM
expectations from the TINA-TI simulation and the requirements laid out in the Design Summary. Test
points were included in the board design at the VPDetect, V_PACE_OUT, and the SR Latch points out to
monitor signal progress through the circuit. Figure 12 displays the transient analysis results when a 20mV
amplitude, 100µs period pacemaker pulse is applied prior to the QRS interval. The Fluke Medsim 300B
was used to simulate a pacemaker pulse and ECG signal.
ECG + Pace Input
Differentiator Out
Window
Comparator Out
SR Latch Out
PACE: 20mV Amplitude
100us period
Figure 12: Time Domain Results (20mV amplitude)
A 20mV pace signal amplitude was first selected to show a visual demonstration of how a pace signal
(shown on channel 1) will cause the SR latch to change to a high state (shown on channel 4).
Once functionality has been proven in Figure 12, a second test is required to demonstrate circuit response
to a 2mV amplitude and 100µs period pulse width pace signal. Figure 13 displays the transient analysis for
these conditions which can be compared to the Figure 9 simulation results.
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
13
www.ti.com
ECG + Pace Input
Differentiator Out
Window
Comparator Out
SR Latch Out
PACE: 2mV Amplitude
100us period
Figure 13: Time Domain Results (2mV Amplitude)
The 2mV amplitude pacemaker signal does successfully trigger the window comparator and SR latch
indicating an event. Figure 12 and Figure 13 also show how the amplitude of the differentiator out circuit is
dependent on the duration of the input pacemaker slope. This concludes testing verification of the
hardware pace detection circuit.
14
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
7
Modifications
The components selected for this design were based on the design goals and outlined at the beginning of
the design process. Those goals are focused on monitoring for a 2mV positive polarity pacemaker pulse
that can be as small as 100µs in signal width. In reality, a pacemaker can be positioned in the body a
variety of ways, where the monitored pacemaker signal could either read back with positive or negative
polarity. Following the design method described above, R5 and R6 can be selected for the circuit to
respond to a negative polarity pace signal in addition to the positive pulse.
The first order high pass filter and slope detection circuit is designed to demonstrate how manipulation of
frequency components can alert a user when a pace maker signal is present prior to the ECG signal. All
tests in this document were performed with a Fluke patient simulator, not taking into account any motion
artifacts which would be present in a commercial medical system. This circuit is not designed to be used
directly with a patient without proper IEC certification.
When designed to be used in a patient monitoring system, additional design effort is required to help filter
motion artifacts which will cause false positive triggers. The poles and zeroes on the differentiator may
need to be manipulated to further attenuate the lower frequency components avoiding unwanted triggers
on the window comparator. Additional gain may be required on the differentiator circuit to allow a wider
range on the window comparator threshold limits to help with motion artifacts. Adding gain to the
differentiator circuit will allow the threshold limits to be pushed further away from the idling voltage, set by
Vbias. Designing a second order high pass filter may be required to help attenuate the lower frequency
components while providing enough gain at the higher frequencies to help with motion artifacts. A wider
bandwidth op amp in the differentiator stage can also help improve the gain before stability compensation.
Moving forward, software detection methods are becoming a more accurate way to monitor for a pace
maker. Using a wide bandwidth SAR ADC, the ECG and pace signal can be digitized and then processed
to remove unwanted signal components. The Q-T interval can be removed, along with electromyography
(EMG) signals from muscle movement and any other motion artifacts leaving solely the pace maker signal
components. Further algorithms can be put in place to not only alert the user if a pacemaker is present, but
also provide details about the pace device.
8
About the Author
Tony Calabria is a Systems Engineer and Product Definer in the Delta Sigma ADC group at Texas
Instruments where he supports medium to wide bandwidth Delta Sigma ADCs. Tony received his BSEE
from the University of Arizona.
9
References
1.
Green, T., Operational Amplifier Stability, Parts 1 – 11, November 2008, http://www.engenius.net/site/zones/acquisitionZONE/technical_notes/acqt_050712
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
15
www.ti.com
Appendix A.
A.1 Electrical Schematic
Figure A-1: Electrical Schematic
16
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
SLAU511-June 2013-Revised June 2013
www.ti.com
A.2 Bill of Materials
Figure A-2: Bill of Materials
SLAU511-June 2013-Revised June 2013
Hardware Pace using Slope Detection
Copyright © 2013, Texas Instruments Incorporated
17
IMPORTANT NOTICE FOR TI REFERENCE DESIGNS
Texas Instruments Incorporated ("TI") reference designs are solely intended to assist designers (“Buyers”) who are developing systems that
incorporate TI semiconductor products (also referred to herein as “components”). Buyer understands and agrees that Buyer remains
responsible for using its independent analysis, evaluation and judgment in designing Buyer’s systems and products.
TI reference designs have been created using standard laboratory conditions and engineering practices. TI has not conducted any
testing other than that specifically described in the published documentation for a particular reference design. TI may make
corrections, enhancements, improvements and other changes to its reference designs.
Buyers are authorized to use TI reference designs with the TI component(s) identified in each particular reference design and to modify the
reference design in the development of their end products. HOWEVER, NO OTHER LICENSE, EXPRESS OR IMPLIED, BY ESTOPPEL
OR OTHERWISE TO ANY OTHER TI INTELLECTUAL PROPERTY RIGHT, AND NO LICENSE TO ANY THIRD PARTY TECHNOLOGY
OR INTELLECTUAL PROPERTY RIGHT, IS GRANTED HEREIN, including but not limited to any patent right, copyright, mask work right,
or other intellectual property right relating to any combination, machine, or process in which TI components or services are used.
Information published by TI regarding third-party products or services does not constitute a license to use such products or services, or a
warranty or endorsement thereof. Use of such information may require a license from a third party under the patents or other intellectual
property of the third party, or a license from TI under the patents or other intellectual property of TI.
TI REFERENCE DESIGNS ARE PROVIDED "AS IS". TI MAKES NO WARRANTIES OR REPRESENTATIONS WITH REGARD TO THE
REFERENCE DESIGNS OR USE OF THE REFERENCE DESIGNS, EXPRESS, IMPLIED OR STATUTORY, INCLUDING ACCURACY OR
COMPLETENESS. TI DISCLAIMS ANY WARRANTY OF TITLE AND ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS
FOR A PARTICULAR PURPOSE, QUIET ENJOYMENT, QUIET POSSESSION, AND NON-INFRINGEMENT OF ANY THIRD PARTY
INTELLECTUAL PROPERTY RIGHTS WITH REGARD TO TI REFERENCE DESIGNS OR USE THEREOF. TI SHALL NOT BE LIABLE
FOR AND SHALL NOT DEFEND OR INDEMNIFY BUYERS AGAINST ANY THIRD PARTY INFRINGEMENT CLAIM THAT RELATES TO
OR IS BASED ON A COMBINATION OF COMPONENTS PROVIDED IN A TI REFERENCE DESIGN. IN NO EVENT SHALL TI BE
LIABLE FOR ANY ACTUAL, SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT DAMAGES, HOWEVER CAUSED, ON ANY
THEORY OF LIABILITY AND WHETHER OR NOT TI HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES, ARISING IN
ANY WAY OUT OF TI REFERENCE DESIGNS OR BUYER’S USE OF TI REFERENCE DESIGNS.
TI reserves the right to make corrections, enhancements, improvements and other changes to its semiconductor products and services per
JESD46, latest issue, and to discontinue any product or service per JESD48, latest issue. Buyers should obtain the latest relevant
information before placing orders and should verify that such information is current and complete. All semiconductor products are sold
subject to TI’s terms and conditions of sale supplied at the time of order acknowledgment.
TI warrants performance of its components to the specifications applicable at the time of sale, in accordance with the warranty in TI’s terms
and conditions of sale of semiconductor products. Testing and other quality control techniques for TI components are used to the extent TI
deems necessary to support this warranty. Except where mandated by applicable law, testing of all parameters of each component is not
necessarily performed.
TI assumes no liability for applications assistance or the design of Buyers’ products. Buyers are responsible for their products and
applications using TI components. To minimize the risks associated with Buyers’ products and applications, Buyers should provide
adequate design and operating safeguards.
Reproduction of significant portions of TI information in TI data books, data sheets or reference designs is permissible only if reproduction is
without alteration and is accompanied by all associated warranties, conditions, limitations, and notices. TI is not responsible or liable for
such altered documentation. Information of third parties may be subject to additional restrictions.
Buyer acknowledges and agrees that it is solely responsible for compliance with all legal, regulatory and safety-related requirements
concerning its products, and any use of TI components in its applications, notwithstanding any applications-related information or support
that may be provided by TI. Buyer represents and agrees that it has all the necessary expertise to create and implement safeguards that
anticipate dangerous failures, monitor failures and their consequences, lessen the likelihood of dangerous failures and take appropriate
remedial actions. Buyer will fully indemnify TI and its representatives against any damages arising out of the use of any TI components in
Buyer’s safety-critical applications.
In some cases, TI components may be promoted specifically to facilitate safety-related applications. With such components, TI’s goal is to
help enable customers to design and create their own end-product solutions that meet applicable functional safety standards and
requirements. Nonetheless, such components are subject to these terms.
No TI components are authorized for use in FDA Class III (or similar life-critical medical equipment) unless authorized officers of the parties
have executed an agreement specifically governing such use.
Only those TI components that TI has specifically designated as military grade or “enhanced plastic” are designed and intended for use in
military/aerospace applications or environments. Buyer acknowledges and agrees that any military or aerospace use of TI components that
have not been so designated is solely at Buyer's risk, and Buyer is solely responsible for compliance with all legal and regulatory
requirements in connection with such use.
TI has specifically designated certain components as meeting ISO/TS16949 requirements, mainly for automotive use. In any case of use of
non-designated products, TI will not be responsible for any failure to meet ISO/TS16949.
Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265
Copyright © 2013, Texas Instruments Incorporated
Download