Laser Tracker Test Facility at SLAC Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration Laser Tracker Investigation ADM Tests Conclusion February, 2008 IWAA08 1 G. Gassner, R. Ruland gassner@slac.stanford.edu Introduction Particle Physics experiments at SLAC require high accuracy positioning, e. g. 100 µm over a distance of 140 m or 25 µm in a 7 x 10 x 3 meter room Realization -> Laser Tracker Measurements Rotary calibration table (Kugler GmbH) providing an accuracy of better than 0.2 arcsec Interferometer bench February, 2008 IWAA08 2 G. Gassner, R. Ruland gassner@slac.stanford.edu Rotary Table Renishaw Signum RESM angle encoder system 1 arcsec graduation accuracy 0.01 arcsec resolution February, 2008 IWAA08 3 G. Gassner, R. Ruland gassner@slac.stanford.edu Calibration of the rotary table (1) Rosette technique calibrate precision polygon prisms sum of the angles measured must result in 360 degrees Moeller-Wedel electronic autocollimator (0.1arcsec accuracy, 0.05arcsec reproducibility) February, 2008 IWAA08 4 G. Gassner, R. Ruland gassner@slac.stanford.edu Calibration of the rotary table (2) Derived Technique set the fixture with the mirrors in a way that mirror ‘a’ is in line with autocollimator ‘a’ rotate the table together with the fixture until mirror ‘b’ is in line with autocollimator ‘b’ February, 2008 IWAA08 5 G. Gassner, R. Ruland gassner@slac.stanford.edu Calibration of the rotary table (3) Absolute value of an angle between two positions of the rotary table can be determined with ±0.2 arcsec std. February, 2008 IWAA08 6 G. Gassner, R. Ruland gassner@slac.stanford.edu Laser Tracker Horizontal Angle Measurement System Investigation Test Setup mirror as a target for laser tracker parallelism of the two rotation axes axial displacement of the rotation axes Mirror Faro Laser Tracker Rotary table February, 2008 IWAA08 7 G. Gassner, R. Ruland gassner@slac.stanford.edu Comparison SMR vs. SMM Test of Mirror (SMM) and Retroreflector (SMR) Only small deviations can be found February, 2008 IWAA08 8 G. Gassner, R. Ruland gassner@slac.stanford.edu Calibration Results Two calibration runs with Faro Laser Tracker Xi February, 2008 IWAA08 Two calibration runs with Faro Laser Tracker Xi V2 9 G. Gassner, R. Ruland gassner@slac.stanford.edu ADM Calibration on Interferometer Bench Calibration of the scale factor of the ADM in the field. Tests of the general performance and tests to detect malfunctions in the laboratory. Carriage • Horizontal Bench Laser Tracker February, 2008 IWAA08 Dd Interferomete 10 G. Gassner, R. Ruland gassner@slac.stanford.edu Calibration Results Calibration run with Faro Laser Tracker Xi February, 2008 IWAA08 Calibration runs with Faro Laser Tracker Xi V2 11 G. Gassner, R. Ruland gassner@slac.stanford.edu Conclusion Rotary table - absolute angle accuracy < 0.2 arcsec Rotary table to test any horizontal angle measurement instrument Faro laser trackers angle measurement system deviations stable and correctible With corrections, angle accuracies of 1 arcsec are possible. Test of the ADM showed its usability as a replacement for the interferometer mode February, 2008 IWAA08 12 G. Gassner, R. Ruland gassner@slac.stanford.edu End of Presentation February, 2008 IWAA08 13 G. Gassner, R. Ruland gassner@slac.stanford.edu