AbstractID: 7756 Title: Loss of Beam Flatness Due to the Double Penumbra Effect in Closely Collimated Electron Applicators and the Necessity of Selecting Oversized Cones Electron beams from Siemens accelerators exhibit some advantageous features, including, in many applicator configurations, reduced output dependency on Cerrobend cutout geometry. Certain Siemens electron applicators, however, show an undesired broadening in the electron beam penumbra due to the close alignment of the mid and lower level trimmers. On Siemens MD model accelerators, for example, light field observations show that the two trimmers are divergently aligned within 2 mm. A result of this alignment is that the penumbra from the mid-level trimmer, which is located about 25 cm above the final trimmer, is essentially superimposed on the final trimmer’s penumbra and causes significant broadening. This effect is particularly problematic for the 5-cm diameter cone, which is often used for facial lesions where smaller source-to-skin distances are critical. Here the double-penumbra effect, along with the narrow field, significantly reduces the flatness of the dose profiles. In this work we include comparisons of field isodose profiles generated with the open 5-cm cone with those of an "identical" field, a 5-cm Cerrobend cutout mounted in a 10x10 cm cone. For 6 and 9 MeV beams, the open 5-cm cone shows isodose widths that are about 1 cm narrower than those of the 5-cm cutout for 80 to 95% isodoses. Hence, for the same target, an additional 5 mm must be added to the margins when the 5-cm open cone is used. Therefore, whenever patient setup constraints allow, the use of a cutout in a larger, oversized, cone is often preferable on Siemens MD accelerators.