Quality Control Siemens MAMMOMAT NovationDR Donald Jacobson, Ph.D. Asst. Adjunct Professor, Radiology Medical College of Wisconsin Milwaukee, WI drj@mcw,edu 414-805-3163 Siemens Opdose W or Mo target Mo or Rh (tilted) filter Effective X-ray Energy Target Mo Mo Rh Mo Rh W Filter Mo Rh Rh Al Al Rh kVp 22 14.1 24 14.6 15.4 14.7 26 14.9 15.8 15.7 17.6 17.8 28 15.4 16.2 16.2 17.8 18.5 18.4 30 15.7 16.5 16.6 18 32 16.9 18.9 18.6 18.1 19.1 on-off on-off Flex paddles are optional, additional to standard paddles Cone or tapered paddle for MLO view 20 x 22 cm FOV The measurement devices listed below must always display a valid calibration stamp to ensure compliance with the described measurement and calibration intervals. • • • • • • • • • • • • • • • • • • • • • • Mammography line pair phantom (2-10 lp/mm) 3 x 20 mm plexi (PMMA) phantom (10x10 cm) Test phantom (RMI model 156 or Nuclear Associates model 18-220) One collimator mounted plexi (4 cm thick) absorber Three plexi (PMMA) absorbers, each with a thickness of 20 mm Lint free non-woven cotton or gauze (100% cotton) Water or lukewarm diluted aqueous solution of household dishwashing liquid Luminance/Illuminance meter Densitometer A non-invasive digital kV meter Dose meter calibrated at the mammography X-ray beam energies Film or CR cassette 24x30 cm or larger Magnification table Compression plate (set for a 1.5 magnification) Compression plate simulator* 2 mm steel plate (FD object table size) 2 mm thick steel bars (approx. 3 x 10 cm) Six sheets of 99.9% pure aluminum Four coins Four paper clips Timer Ruler with a mm scale SMPTE pattern, AAPM TG18 test images * Provided with every system. Testing device allows exposures to be performed without compression plate attached. Detach compression plate and insert metal plug compression plate simulator. 4 cm acrylic attenuator Compression plate simulator Grid Quality Control Manual MAMMOMAT NovationDR Version 03, dated 10/05 or Version 04 dated 01/06 or Version 05, dated 04/07 Version 4 adds SNR & CNR for T weekly Compression force semi-annually Repeat rate quarterly 3. Start up and Login Procedure CAUTION! If the MAMMOMAT is started from being completely shut down (for details about complete shut down, see Instructions for Use for MAMMOMAT NovationDR) the detector should be powered on at least 1 hour before intended use. Otherwise, detector image quality can be affected. The start-up time can be shortened to approximately 5 minutes if the MAMMOMAT is started from the standby mode (for details about the standby mode, see Instructions for Use for MAMMOMAT NovationDR). 1. Press the enable button ( ) on the control panel to enable the MAMMOMAT. Allow a warm up time of approximately 5 minutes. The internal monitoring system automatically performs a functional check of the MAMMOMAT. A “DR” is displayed on the density display on the control panel to indicate that the MAMMOMAT NovationDR system is operational. 2. Switch ON the acquisition workstation if it is off. Otherwise proceed to step # 3. NOTE! A message about warming-up of the detector might be displayed. If the detector has been ON for more than 60 minutes, dismiss. Otherwise, wait for the detector to warm up. Every other week, a message about calibration of the detector is displayed. If the detector has been calibrated within the intervals described in the chapter: “Calibrating the detector” in the MAMMOMAT NovationDR Instructions for Use, dismiss. Otherwise, perform a calibration procedure according to the same chapter. 3. Login to the acquisition workstation’s operating system as apps-level user. 4. Sign in to the DirectRay FFDM Acquisition Station software, see chapter “Signing in to the DirectRay FFDM Acquisition Station software” in the MAMMOMAT NovationDR Instructions for Use. User name: mammo Password: mammo Patient menu task cards Test 4.1 Detector calibration 4.2 Artifact detection (field flatness) Test 4.1 Detector calibration 4.2 Artifact detection (field flatness) Test 4.3 Acquisition workstation monitor check SMPTE test pattern 5%, 95% squares high contrast lines visual evaluation Test 4.4: Mechanical Inspection and Follow up 18 safety and operational checks Test 4.5: Collimator Assessment (from the film) (from the digital image) Light field illuminance > 160 lux Test 5.2: Chest Wall Missed Tissue (Optional) (Not optional) Measure diameter and imaged portion of coin this is called “α" Test 4.6: Compression Plate Position Make an image of each paddle to make sure it is not visible at the chest wall. < 6.5 mm this is called “α" Test 4.7: Spatial Resolution AEC, on 4 cm plexiglas > 7 lp/mm Test 4.8: Phantom Image Quality RMI 156 phantom Place flush 5 4 4 Test 4.9: Detector Uniformity < 10% deviation from mean of means. Test 4.10 Radiation Output Rate > 800 mR/sec Test 4.11: Tube Voltage Measurement & Reproducibility Use non-invasive kVp meter Place on the 2 mm steel plate accy < 5%, CV < 2% Test 4.12: Beam Quality (HVL) Use spot compression plate (not mag. paddle) to hold Al sheets (99.9% pure) all anode-filter combinations > 0.28 mm Al (at 28 kVp) Test 4.13: Mean glandular dose < 3 mGy required < 2 mGy desired (for both H and D modes) Phantom should be centered L-R Opdose AEC mode Use Program 2 for phantom imaging (W/Rh 27-28 kVp). H and D modes. PMB 45 (2000) 3225-3240 Patient Radiation Dose Average glandular dose Dg (mrad) = exp x HVL x 1/T x 2.1 exp in mR HVL in mm Al Thickness (T) in cm 100 mrad = 1 mgrey (mG) Jacobson, DR. Radiographic Exposure Calculator and Mammographic Dose Calculator. Radiology, 1992; 182:578-580 Test 4.14: SNR and AEC reproducibility Place flush, expose 5 times AUTO mode Sensor posn. 2 SNR = meanb – 50 SDb mAs exp. <pix> SNR Test 4.15: SNR and CNR Measurements SNR > 40 SNR, CNR dev within + 15% of baseline value SNR = meanb – 50 SDb CNR = meanb – meanm SDb Test 4.16: AEC Thickness Tracking Test 2,4,6 cm (24,26,28 kVp) SNR > 40 Deviation < 15% from mean of the means Place flush Test 4.17: Printer check Follow Siemens or camera manufacturers instructions Print SMPTE pattern, measure 11 optical densities “If your values do not meet the values specified above, then please refer to the printer manufacturer’s quality control procedure. You can create your own baseline with the measured values that result from the initial install of the printer that was installed according to the manufacturer’s QC procedure.” syngo MammoReportPlus syngo MammoReportPlus User name: mammo Password: mammo SCR Administration Equipment needed: • • • • • Ruler Luminance meter Illuminance meter Magnification lens Test images (installed on Mammotest Plus) • Multi: TG18-QC • glare: TG18-GQ, -GQB, -GQN • lumin: TG18-LN12-01, -LN12-18, -UNL10, -UNL80 • noise: TG18-AFC Mammo Report QC Tests 1. Reflection 2. Overall evaluation 3. Geometric distortion (CRT only 4. Luminance response 5. Luminance uniformity 6. Resolution 7. Noise Acceptance and annual testing – MP – tests 1-7 Daily – RT/Rad. – tests 1-2 Technologist performs daily (pass/fail) Test 3: Geometric distortion Illuminance refers to light falling on the screen. Lamb is ambient diffuse illuminance Loff is reflected Lamb with monitor o Luminance refers to light emitted from the screen. Must supply own light meter. Test 4: Luminance response Ambient light should be < 10 lux. Lmin should be < 1.0 cd/m2 (TG18-LN-12-01) Lmax should be > 250 cd/m2 (TG18-LN-12-18) LR = Lmax / Lmin > 250 (Lmax, left –LLmax.right ) / Lmax. left. < 10% Loff should be < 0.6 Lmin Test 5: Luminance Uniformity Maximum deviation: (Lmax – Lmin) / Lcenter should be < 30%. TG18-UNL10 TG18-UNL80 Test 6: Resolution Evaluate TG18-QC. CX pattern. x x Test 7: Noise Evaluate TG18-AFC. Visualization of all targets in 3 quadrants of the pattern AFC (alternative forced choice)