Research Support Dual-Source CT: What is it and How Do I Test it? • National Institute of Health – EB04898, CA75333, AR27065, HR46158 Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN • Flight Attendant Medical Research Institute • Siemens Medical Solutions • Bayer Healthcare • RTI Electronics Tube A: 50 cm FOV Tube A: 50 cm FOV Tube B: 26 cm FOV Tube B: 26 cm FOV Each tube/detector pair has: 80 kW generator 32 x 0.6 mm detector array Double z sampling with Z-FFS No gantry tilt I. Single Tube Operation - Performs Same as a Sensation 64 Temporal Resolution = Rotation Time 330 ms = 2 2 = 165 ms 83 ms vs. 165 ms (same raw data, same phase) II. Dual Tube Operation - x 2 Improved Temporal Resolution Single tube recon Dual tube recon Temporal Resolution = Rotation Time 330 ms = 4 4 = 83 ms RCA RCA LAD Mean HR=90 bpm LAD Dose 82 ms • • 125 ms 165 ms McCollough et al., Radiology 243(3):775-784 (2007) 70 MDCT DSCT 0.2 MDCT pulsed DSCT 310 ms 0.2 DSCT 210 ms 0.2 DSCT 110 ms 0.2 60 Continuous irradiation (spiral acquisition) with two x-ray tubes would double the radiation dose Four dose reduction strategies: – – – – Cardiac beam shaping filter (bowtie filter) 3-D adaptive noise filtration Variable pitch values (based on heart rate) ECG-pulsing with arrhythmia detection and adjustable temporal windows III. Dual Source (Obese) Mode Use both tubes at same kV to double the available power to 160 kW 50 40 30 0.2 0.265 0.36 0.46 20 10 No ECG pulsing Pulsing No ECG pulsing Pulsing No ECG pulsing Pulsing No ECG pulsing Pulsing 0 <55 55-70 70-90 Heart Rate (bpm) >90 •Reconstructed images are the SUMMATION of Tube A and B •Avoids need to trade off exam speed (pitch) or narrow slices (collimation) for photon flux •Allows much higher mAs for – Obese patients – 80 kVp imaging IV. Material Specific Imaging Use each tube at a different kV to exploit the kV-dependent nature of CT # 80kV Bone 670 HU Iodine 296 HU Applications of Dual-Energy CT • Iodine imaging – Automated bone removal in CT angiography – Plaque removal – Blood pool imaging (Perfused blood volume) • Soft tissue imaging – Enhanced visualization of tendons & ligaments – Virtual non-contrast (Iodine removal) Bone 450 HU Iodine 144 HU 140kV • Dose reduction • Material characterization TR Johnson et al. Eur Radiology 17(6):1510-1517 (2007) Applications of Dual-Energy CT Direct subtraction of bone in complicated anatomical regions • Iodine imaging – Automated bone removal in CT angiography – Plaque removal – Blood pool imaging (Perfused blood volume) • Soft tissue imaging – Enhanced visualization of tendons & ligaments – Virtual non-contrast (Iodine removal) • Dose reduction • Material characterization Courtesy of University Hospital of Munich - Grosshadern / Munich, Germany Applications of Dual-Energy CT • Iodine imaging Contrast enhanced Nephrographic phase – Automated bone removal in CT angiography – Plaque removal – Blood pool imaging (Perfused blood volume) • Soft tissue imaging – Enhanced visualization of tendons & ligaments – Virtual non-contrast (Iodine removal) DE image from same scan Virtual non-contrast • Dose reduction • Material characterization O. Dzyubak, Tuesday 2:42 pm, Rm L100J Implications for Performance Testing Image Quality • Single source mode: No change from any MDCT • Dual source cardiac mode: optional – Use demo ECG mode – Coronary CTA: 120 kVp, 240 mAs/rot, 0.33 sec, 32x0.6, pitch auto selected, 3 mm, B26f, retrospectively-gated spiral mode – ACR phantom: CT number accuracy – Large (> 30 cm diameter) water or uniform acrylic phantom • Uniformity and noise in axial and coronal images – mAs/rot represents mAs sum from both tubes (not effective mAs = mAs/pitch) Implications for Performance Testing Image Quality • Dual energy mode: Suggested – Mixed image (0.3 • 80 kV + 0.7 • 140 kV) – DE abdomen: 140/80 kVp, 95/400 eff. mAs, 0.5 s, 14x1.2 mm, pitch < 0.7, 5 mm, D30f, spiral – ACR CT phantom: CT number accuracy • Fat (polyethylene) and bone (Teflon) expected to have different CT numbers relative to 120 kVp • Check accuracy of water and air CT numbers – Large (> 30 cm) water or uniform acrylic phantom • Uniformity and noise in axial and coronal images Implications for Performance Testing Image Quality • Dual source non-cardiac (obese) mode: Optional – Average image (0.5 • Tube A + 0.5 Tube B) – Obese XXL: 120 kVp, 300 eff. mAs, 0.5 s, 24x1.2 mm, pitch < 1.0, 5 mm, B20f, spiral – ACR phantom: CT number accuracy – Large (> 30 cm) water or uniform acrylic phantom • Uniformity and noise in axial and coronal images Implications for Performance Testing Dose • Must test both tubes, recommend to do separately • Tube A: No change from any MDCT – CTDIw in head and body phantoms for routine head and body – CTDIair for all collimations at routine kVp – CTDIair for all kVp settings at routine collimation – IEC 61223-2-6-Ed2 (Constancy Testing) • Tube B: Fewer collimations available • HVL for both tubes at acceptance testing (required after any tube chance in some states) • Change of one tube does NOT affect other tube • Doses are additive Body Dose Measurement Cardiac Dose Measurement (Body CTDI phantom, Body filter) (Body CTDI phantom, Head filter) 80kVp 100kVp 120kVp QC(A,B) AT (A) AT (A) QC(A,B) QC(A,B) QC(A) QC(A) QC(A,B) AT (B) AT (B) CTDIw 32x0.6(UHR) AT (A,B) UHR =Ultra high resolution focalspot. 140kVp 2x1 1x5 1x10 14x1.2 24x1.2 32x0.6 QC(A) AT (B) AT = acceptance test only. QC = AT + scheduled QC or after tube change Head Dose Measurement (Head CTDI phantom, Head filter) 80kVp 120kVp QC(A,B) AT(A) AT(A) QC(A,B) QC(A,B) QC(A) QC(A) QC(A,B) AT(B) AT(B) CTDIw 32x0.6(UHR) AT(A,B) UHR =Ultra highresolutionfocalspot. 2x1 1x5 1x10 14x1.2 24x1.2 32x0.6 100kVp 140kVp QC(A) AT(B) AT = acceptance test only. QC = AT + scheduled QC or after tube change 80kVp 120kVp QC(A,B) AT(A) AT(A) QC(A,B) QC(A) QC(A) QC(A,B) AT(B) AT(B) CTDIw 32x0.6(UHR) AT(A,B) UHR =Ultra highresolutionfocalspot. 2x1 1x5 1x10 24x1.2 32x0.6 100kVp 140kVp QC(A) AT(B) AT = acceptance test only. QC = AT + scheduled QC or after tube change Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Dose Measurement – Service mode Mayo CT Clinic Innovation Center and Dept. of Radiology L Yu, MR Bruesewitz, JM Kofler, AN Primak, AP Dzyubak, X Liu JG Fletcher, TJ Vrtiska, EE Williamson, JF Breen, DM Hough Additional Reference: Flohr, TG et al., European Radiology 16:256-268 (2006)