AAPM 2004, Pittsburgh, PA TG18 Display Evaluation Demonstration Workshop Director: Ehsan Samei, PhD, DABR Duke University samei@duke.edu Workshop format 1:30 2:10 2:25 Lecture Moderated review of workstations Group/individual hands-on rotations 3:20 BREAK 4:00 4:30 4:45 Lecture (repeated) Moderated review of workstations Group/individual hands-on rotations Workshop participants Ehsan Samei Hans Roehrig Duke University Medical Center University of Arizona, Tucson Manufacturers: 1. BarcoView, LLC 2. Data Ray Corp. 3. Image Smiths, Inc. 4. Image Systems Corp. 5. National Display Systems 6. Planar Systems, Inc. 7. Totoku Electric Co., Ltd. Rationale Electronic displays, a key component of digital imaging Image Processing Imaging Device 1 1 Rationale Soft-copy display quality is susceptible to variations in hardware, calibration, and to degradations over time Poor display quality leads to reduced overall effectiveness of diagnostic imaging, misdiagnosis (lawsuits), quality mismatch leading to inconsistent clinical decisions Monitoring and maintaining display quality is essential for quality patient care TG18 Document In final editing phase Reviews – current display metrology standards – current display technologies Describes AAPM TG18 AAPM task group aimed to define standard and objective methods for Assessment of display performance for medical imaging systems Version 9, Oct. 2002 Version 10, Aug. 2004 IEC Display WG 36 TG18 => IEC Formed in 2003 to develop an international standard – Prerequisites and tools for display assessment Public draft soliciting comments – Assessment methods expected before the end of 2004 Defines – Acceptance testing and quality control criteria 2 2 Current Display Evaluation Standards Basics Luminance Diverse – Not standardized – Visible light emitting from a surface into a unit solid Do not cover all aspects of display performance angle per second Often do not indicate acceptance criteria Either too subjective or too complicated – SI unit: lumens/steradian/m2 , cd/m2 or nit Not yet fully endorsed by the medical community – Conventional unit: foot-lambert = 3.43 cd/m2 Basics Illuminance – Quantity of visible light striking a surface per second – SI unit: lumens/m2 or lux 5 Initial steps 1. 2. 3. 4. 5. Display class The purpose of testing Availability of test patterns Availability of required instrumentation Prerequisite steps – 1 lux striking a perfect white diffuser surface => 1/ nit 3 3 2. Purpose of Testing 1. Display Class Primary (i.e., diagnostic) – Used for primary interpretation Secondary (i.e., clinical) – Used for secondary view of medical images – Acquisition QC displays 1. Acceptance testing (by medical physicist) – Includes all key quantitative display characteristics (~1 hr) 2. Annual compliance evaluation (by medical physicist) – Redoing of almost all acceptance tests (~30-45 min) 3. Monthly/quarterly evaluation (by QC technician) – Mix of quantitative and visual tests (~ 15-20 minutes) – Results logged for consistency checks 4. Daily evaluation (by user) – Checks with a login test pattern (<1 min) – Clear instructions about who to call 3. Test Patterns 4. Instrumentation Luminance meter Illuminance meter Magnifier Mask Light source Other devices – High-grade CCD camera 4 4 5. Prerequisites Display placement Display placement Absence of strong magnetic fields Clean face-plate Ambient lighting setup Display calibration to GSDF Ambient light flowchart Set L’max Desired LR’ Measured Rd No specular reflection Reflection, diffuse Illuminator L’min = L’max/LR’ Lamb Light sources Telescopic Luminance-meter ~5 L’min Display E Lamb /Rd Rd= Lr / I 5 5 Reflection criteria Testing methods 100 Rd = 0.001 /sr Maximum allowable illuminance (lux) Lmin = 1.0 cd/m2 Rd = 0.002 /sr Rd = 0.004 /sr Rd = 0.008 /sr Rd = 0.016 /sr 20 Rd = 0.032 /sr 10 Rd = 0.064 /sr 1 0.001 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 9. Other tests 0.004 0.01 0.1 Specular reflection coefficient (Rs) Testing methods 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity Geometry GD = max(di-dmin)/dmin GD < 2% 9. Other tests TG18-QC 6 6 Geometry Geometry Image clips: – Whole pattern visible w/o clips Clocking distortion: – All edge line patterns w/o distortion Local distortion: – 2 directional measurement per quadrant – Max deviation, max(di-dmin)/dmin < 2% (5% secondary) Global distortion: – Directional measures added together – Max deviation <2% (5% secondary) TG18-GD / IEC-GD Testing methods 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 9. Other tests Visual assessment Specular: – High-contrast white objects should not be visible Diffuse: – Equivalent visibility threshold of lowcontrast targets in a dark field with and without ambient lighting TG18-AD 7 7 Reflection, specular Light sources Reflection criteria The maximum ambient lighting should be such that – specular reflections of white objects (with 90% reflection) at low luminance levels is < contrast threshold E < ( /0.9) x (Ct Lmin / Rs) Display Telescopic Luminance-meter – contrast reduction due to diffuse reflection at low luminance levels < 20% E < 0.25 Lmin / Rd Rs= Lr / Lo Testing methods 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 9. Other tests Lamb Measured OR Lamb = ERd Display off 8 8 Luminance response Luminance response L’max = Lmax+Lamb > 170 cd/m2 for primary displays > 100 cd/m2 for secondary displays ~ 5% of desired value L’min = Lmin+Lamb > 5Lamb ~ 10% of desired value LR’ = L’max/L’min > 250 for primary displays > 100 for secondary displays TG18-LN Luminance response Luminance response 1000 Luminance, cd/m 2 – Lmin and Lmax mapped to corresponding GSDF JND index (J) 100 280 cd/m 10 1.5 cd/m2 2 J min = J ( L min ) J max = J ( L max ) J = J max J min Jp = P J – JNDs ascertained corresponding to luminance values by linearly mapping the JND and pixel value (P) spaces 1 0 200 400 JND indice (DICOM 3.14) 600 J i = J min + P i J max P 9 9 Luminance response Luminance response – Measured and GSDF luminance ascertained as a function of JND – Maximum deviation between measured and GSDF JND per pixel value calculated = incremental contrast per JND L di @ J i – measured and GSDF incremental contrast per JND ascertained = i d i = 2 (L i (L i + L i (L d i 2 (L di + L di Li 1 )( J i 1 Ji L di 1 )( J ) i 1 ) Ji 1 ) 1 @ 0 .5 ( J i + J i ) @ 0 .5 ( J i + J i 1 = Max ) 1 ) Luminance results JNDmax for L’max GSDF calibrated JNDmin for L’min Uncalibrated 0.100 dL/L for a JND 0.100 JND < 10% deviation from GSDF’s for primary < 20% deviation from GSDF’s for secondary d i i 0.2 for secondary Luminance response L’/L’avg per ( 0.1 for primary ) TG18-LN 313 cd/m2 dL/L for a JND L i @ Ji 0.010 0.5 cd/m2 0.001 237 cd/m2 0.010 1 cd/m2 0.001 0 100 200 300 400 JND indice 500 600 700 0 100 200 300 400 500 600 700 JND indice 10 10 versus # of steps Luminance results 0.4 0.100 n 0.35 Barten model + tolerance < 2.0 < 1.0 dL/L per JND Jp < 3.0 0.010 Complianace with GSDF contrast Measured 0.3 0.25 0.2 0.15 n 0.1 0.05 0 0.001 0 200 400 600 1 Visual assessment of luminance response 10 100 1000 Number of levels (n) JND index (Dicom 3.14) Testing methods 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 9. Other tests TG18-CT pattern: Equivalent perception of lowcontrast objects TG18-MP pattern: Assessment of bit-depth resolution via ramps 11 11 Viewing angle problem Luminance uniformity TG18-ULN10 Max L/Lavg @ 10% < 30% TG18-ULN80 Max L/Lavg @ 80% < 30% Angle of regard problem Lmax and Lmin vs angle 300 6 Phi=0 Phi=0 Phi=45 Phi=45 Phi=90 250 Minimum luminance (cd/m^2) Maximum luminance (cd/m^2) 200 150 100 50 Viewing distance: 30 cm Display size: 30x40 cm Max off-axis viewing angle: 35 degrees Phi=90 5 Phi=135 Phi=135 4 3 2 1 0 -80 -60 -40 -20 0 20 Viewing angle (degrees) 40 60 80 0 -80 -60 -40 -20 0 20 40 60 80 Viewing angle (degrees) 12 12 Angular response LR vs angle 300 Phi=0 Phi=45 Phi=90 250 Luminance ratio Phi=135 200 180 150 145 Viewing distance: Display size: Max off-axis viewing angle: 30 cm 30x40 cm 35 degrees 95 100 60 50 0 -80 -60 -40 -20 0 20 40 60 80 Viewing angle (degrees) LR vs angle Angular response, hor. 0.100 800 Phi=0 Phi=45 700 Phi=90 Dicom 3.14 Phi=135 0 -20 500 400 LR=250 300 dL/L per JND Luminance ratio 600 -40 -60 0.010 20 40 60 200 + tolerance - tolerance 100 0 -80 -60 -40 -20 0 20 View ing angle (degre es) 40 60 80 0.001 0 200 400 600 800 JND index (Dicom 3.14) 13 13 GSDF calib. angular tolerance Lmin Lmax LR (0.3) (250) mLCD 3mpx 0.87 715.1 820 ±23 ±25 Max. Max.dev. dev.btw. btw.measured measuredand andGSDF GSDFinc. inc.dL/L dL/L 1.2 1.2 1.0 1.0 Testing methods Phi=0 Phi=0 Phi=45 Phi=45 Phi=90 Phi=90 Phi=135 Phi=135 0.8 0.8 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 0.6 0.6 0.4 0.4 =0.3 0.2 0.2 0.0 0.0 -80 -80 9. Other tests -60 -60 -40 -40 -20 -20 0 0 20 20 40 40 60 60 80 80 wing a ngle (degrees) (degrees) Vie V iew ing angle (assumed perfect on-axis calibration) TG18-RV50 Resolution evaluation 14 14 TG18-QC Display MTF FWHM = 0.20 mm FWHM = 0.27 mm MTF(fn) > 35% for primary displays 1 0.8 > 25% for secondary displays 4500 0.6 LSF MTF 3500 2500 0.4 1500 0.2 500 -500 -0.8 0 -0.4 0 0.4 0.8 Distance (m m ) TG18-CX 0 1 2 3 4 5 6 7 8 Spatial Frequency (cycles/mm) CX resolution criteria CX score < 4 for primary displays (RAR < 1.15) CX score < 6 for secondary displays (RAR < 1.47) 15 15 Testing methods Noise TG18-AFC 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity CD = 0.012 mm CD = 0.027 mm CD = 0.049 mm CD = 0.110 mm 9. Other tests Noise Testing methods Noise in radiographic imaging 250,000 quanta/mR-mm2 X 1 mR/exposure X 0.22 mm2/pixel X 0.25 detected quanta/incident quanta = 3,500 detected quanta/pixel ~ 2% relative noise Assuming display noise < 1/3 quantum noise ~ 0.7% w/ threshold SNR ~ 3 CD threshold ~ 0.023 mm (Rose model) w/ 0.018 C/PV (8 bit) and w/ 0.15 mm/display pixel (5 Mpx) square target pixels 1 2 3 4 5 PV contrast 8 4 3 2 2 CD value (mm) 0.024 0.024 0.027 0.024 0.030 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 9. Other tests 16 16 Veiling glare GR = LB L LN LN Veiling glare, visual TG18-GQ TG18-GQN Cone Telescopic Luminance-meter Display TG18-GV TG18-GVN TG18-GQB Glare criteria Visual test 3rd target (96/4096 PV contrast) visible in both patterns for primary displays 5rd target (160/4096 PV contrast) visible in both patterns secondary displays Testing methods 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity 9. Other tests 17 17 Chromaticity Testing methods u' = 4x/(-2x +12y+3) v' = 9y/(-2x+12y+3) CU = u’2 + v’2 1. Geometry 5. Resolution 2. Reflection 6. Noise 3. Luminance 7. Veiling glare 4. Lum. dependencies 8. Chromaticity CU < 0.01 9. Other tests TG18-ULN80 Overall display QC Display artifacts More visually-based and oriented toward QC “Clinical” test patterns – TG18-CH – TG18-KN – TG18-MM1 – TG18-MM2 Comprehensive test pattern: TG18-QC TG18-QC Line-pair patterns CX targets CX scoring reference 16 luminance steps Low-contrast targets 5 and 95% patches Continuous Ramps “Quality Control” targets Contrast bars Cross-talk target 18 18 TG18-CH Clinical patterns TG18-MM1 1. 2. 3. 4. 5. 6. 7. 8. 9. 1. 2. 3. 4. 5. 6. Evaluation elements Degree of difficulty for exam Overall Contrast Overall Sharpness Symmetrical reproduction of the thorax Medial border of the scapulae Reproduction of rib cage above the diaphragm 7. Sharp reproduction of the vascular patterns, the trachea and proximal bronchi, the borders of the heart and the aorta, and the diaphragm 8. Visibility of the retrocardiac lung and the mediastinum 9. Visibility of the spine through the heart shadow 10. Visibility of small details in the whole lung 11. Visibility of linear and reticular details out to the lung periphery TG18-MM and TG18-KN Evaluation elements Degree of difficulty for exam Overall contrast Overall exposure/penetration Overall sharpness (no blur) Overall positioning - nipple in profile Adequate compression Sharp appearance of Cooper's ligaments Structure of the clip and the presence of the gap at its apex Appearance and visibility of subtle microcalcifications Visibility of structures at the margins of the breast 19 19 TG18 info Workshop participants Ehsan Samei Hans Roehrig deckard.duhs.duke.edu/tg18 Duke University Medical Center University of Arizona, Tucson Manufacturers: 1. BarcoView, LLC 2. Data Ray Corp. 3. Image Smiths, Inc. 4. Image Systems Corp. 5. National Display Systems 6. Planar Systems, Inc. 7. Totoku Electric Co., Ltd. 20 20