OVERVIEW QC for FFDM: What You Must Do and What Really Matters Eric A. Berns, Ph.D. Northwestern University Medical School Lynn Sage Comprehensive Breast Center Chicago, IL Important PrePre-Survey Events • Obtain proper training & CE credits (8 hours) – HandsHands-on training on actual unit: • Mechanics • Software • Artifacts • Learn vendor specific tests and tricks • Important prepre-survey events • Manufacturer’ Manufacturer’s tests and equipment • Performing the survey • Summary of important points and what really matters Important PrePre-Survey Events • ACR Accreditation - www.acr.org – GE Senographe 2000D – Fischer Senoscan – Lorad Selenia • FDA Accreditation - www.fda.gov/cdrh/mammography/ www.fda.gov/cdrh/mammography/ – GE Senographe DS – Siemens, Sectra, Sectra, Fuji, etc., when approved 1 Important PrePre-Survey Events Important PrePre-Survey Events •ACR Accreditation – Equipment Evaluation Forms • ACR Accreditation – Before clinical use MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital – GE Medical Systems Site • Medical Physicist equipment evaluation and indicate it passes Report Date Survey Date Model GE Medical Systems X-Ray Unit Manufacturer Date of Installation Medical Physicist Fischer X-Ray Unit Manufacturer Date of Installation Medical Physicist Medical Physicist's QC Tests ? 2277390-100 Rev 1, 2000 ? 2371472-100 Rev 0, 2003 • New unit application PASS/FAIL/NA Flat Field (as described in radiologic technologist tests) Image Quality (Phantom) (as described in radiologic technologist tests) CNR (required) Change in CNR =0.2 (NA for Equipment Evaluations) Fibers Specks Masses Phantom IQ Test on AWS Phantom IQ Test on RWS – Left* (* NA for units with Seno Phantom IQ Test on RWS – Right* Advantage workstations) Phantom IQ Test on Printer 3. MTF Measurement (as described in radiologic technologist tests) 4. AOP Mode and Signal-to-Noise (SNR) (as described in radiologic technologist tests) 5. Collimation Assessment Deviation between X-ray field and light field is less than 2% of SID X-ray field does not extend beyond any side of the IR by more than 2% of SID Chest wall edge of compression paddle doesn't extend beyond IR by more than 1% of SID 6. Evaluation of Focal Spot Performance Measured performance within acceptable limits for large focal spot Measured performance within acceptable limits for small focal spot 7. Breast Entrance Exposure, Average Glandular Dose and Reproducibility Average glandular dose for average breast is below 3 mGy (300 mrad) Average glandular dose to a 4.2-cm-thick breast on your unit is mrad Exposure reproducibility (CV) for R and mAs must be less than 0.05 8. Artifact Evaluation and Flat Field Uniformity 9. Viewing Condition Check and Setting 10. Monitor Calibration 11. Image Quality – SMPTE Pattern 12. Analysis of Review Work Station (RWS) or Seno Advantage Screen Uniformity • However, no reimbursement without FDA receiving ACR app. • Approximately 3 days for accreditation approval from ACR (required for Equipment Evaluations and as necessary) 13. kVp Accuracy and Reproducibility Measured average kVp within ±5% of indicated kVp kVp coefficient of variation =0.02 14. Beam Quality Assessment (Half-Value Layer Measurement) 15. Radiation Output Radiation output is =800 mR/s 16. Mammographic Unit Assembly Evaluation Important PrePre-Survey Events (not required for Equipment Evaluations) Frequency (Fischer, continued) (Lorad, continued) Evaluation of Site's Technologist QC Program Evaluation of Site's Technologist QC Program PASS/FAIL MQSA Regs 1. Monitor cleaning Daily Daily 2. Viewing conditions for Review Work Station (RWS) or Seno Advantage Daily, if applicable 3. Darkroom cleanliness 4. Processor QC Daily, if applicable 5. Mobile unit quality control After every move, if applicable 6. Flat field Weekly 7. Image quality (phantom) Weekly Weekly 8. Viewbox and viewing conditions Monthly 9. MTF measurement 10. AOP mode and signal-to-noise (SNR) Monthly 11. Visual checklist Monthly 12. Monitor calibration check Monthly Quarterly 13. Repeat analysis - performed, records maintained, radiologist reviewed Quarterly, if applicable 14. Analysis of fixer retention 15. Compression force (pressure) Semi-annually 16. Darkroom fog Semi-annually, if applicable 17. Laser film printer QC (according to printer manufacturer recommendations) Medical Physicist's Recommendations for Quality Improvement (not required for Equipment Evaluations ) (not required for Equipment Evaluations) PASS/FAIL Frequency 1. 2. 3. 4. 5. 6. 7. 8. 9. Laser Imager Quality Test (if applicable) Image Display Monitor Test (Review Station) Phantom Image Acquisition Test Phantom Image Quality Test Detector Calibration and Flat Field Test System Resolution (Detector Alignment/Scan Speed Uniformity) System Operation Reject/Repeat Analysis Compression Force Test MQSA Regs Daily Daily Weekly Weekly Weekly Monthly Monthly Quarterly Semi-annually Medical Physicist's Recommendations for Quality Improvement 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. Darkroom Cleanliness (if applicable) Processor Quality Control (if applicable) Laser Printer Quality Control Viewboxes and Viewing Conditions Softcopy Workstation QC Artifact Evaluation Signal-To-Noise and Contrast-To-Noise Measurements Phantom Image Detector Flat-Field Calibration Compression Thickness Indicator Visual Checklist Analysis of Fixer Retention in Film (if applicable) Repeat Analysis Darkroom Fog (if applicable) Compression mR/s Fischer QC manual version at facility (check one) : ? P-55943-OM Issue 1, Rev. 000 (2002) ? P-55943-OM Issue 1, Rev. 3 (July 2003) ? P-55943-OM Issue 1, Rev. 1 (April 2003) ? P-55943-OM Issue 1, Rev. 7 (March 2004) PASS/FAIL/NA X-Ray Field Size Alignment and Chest Wall Missed Tissue Checks X-ray field size aligns with field area indicated on breast support within =2% SID Chest wall edge of the digital image to the ruler reference mark of breast support is =8.5 mm Compression Paddle Alignment Compression paddle edges not visible within field of view Chest wall edge of compression paddle extends beyond image by =1% SID 3. kVp Accuracy Test Measured average kVp within ±5% of indicated kVp kVp coefficient of variation =0.02 4. Linearity, Reproducibility and Accuracy Linearity <0.08 Reproducibility <0.035 for each technique 5. Half-Value Layer and Output HVL =0.33 mm Al at 30 kVp, 100 mA 6. Dosimetry – Average Glandular Dose and Output Average glandular dose for average breast is below 3 mGy (300 mrad) mrad Average glandular dose to a 4.2-cm-thick breast on your unit is 7. Phantom Image Acquisition Test No obvious artifacts (ADU values required for both QC and Equipment Evaluations; P/F results are NA for Equipment Evaluations) within +50/-0 ADU counts of baseline Background StDev within ±100 ADU counts of baseline Background mean within ±300 ADU counts of baseline ADU level difference 8. Image Quality Fibers Specks Masses Largest 4 fibers, 3 speck groups and 3 masses Phantom IQ Test on Review Work Station 9. System Resolution/Scan Speed Uniformity Standard imaging mode @ 7 lp/mm: =5 transitions, =10% modulation High resolution imaging mode @ 11.1 lp/mm: =5 transitions, =5% modulation 10. Flat Field Test Flat field test Deviations between corner ROIs and center within ±20% 11. Geometric Distortion and Resolution Uniformity 12. Automatic Decompression Control 13. System Artifacts 14. Image Display Monitor(s) Check (review daily log) 15. Image Viewing Room Illuminance Test (=50 lux) 1. 2. Lorad X-Ray Unit Manufacturer Date of Installation Medical Physicist Report Date Survey Date Model Room ID Signature Selenia Medical Physicist's QC Tests ? June 11, 2002 ? 9-500-0285, Rev. 004 (2004) Lorad QC manual version at facility (check one) : ? 9-500-0285, Rev. 003 (2003) ? 9-500-0285, Rev. 002 (2003) PASS/FAIL 1. Mammographic Unit Assembly Evaluation Autodecompression can be overidden to maintain compression (& status maintained) Manual emergency compression release can be activated in the event of power failure Collimation Assessment Deviation between X-ray field and light field is less than 2% of SID X-ray field does not extend beyond any side of the IR by more than 2% of SID Chest wall edge of compression paddle doesn't extend beyond IR by more than 1% of SID Artifact Evaluation Artifacts were not apparent or not significant 4. kVp Accuracy and Reproducibility Measured average kVp within ±5% of indicated kVp kVp coefficient of variation =0.02 5. Beam Quality Assessment - HVL Measurement Half-value layer is within acceptable lower and upper limits at all kVp values tested 6. Evaluation of System Resolution Measured performance within acceptable limits 7. Breast Entrance Exposure and Average Glandular Dose Average glandular dose for average breast is below 3 mGy (300 mrad) Average glandular dose to a 4.2-cm-thick breast on your unit is mrad 8. Radiation Output Rate Radiation output rate is greater than 800 mR/sec mR/sec 9. Phantom Image Quality Evaluation 5 largest fibers, 4 largest speck groups and 4 largest masses are visible Fibers Background Optical Density Phantom image quality scores: Specks Disk Optical Density Disk Contrast Masses Hard copy background density must be =1.20 (with operating level =1.40) Hard copy density difference over acrylic disk must be =0.35 (with operating level =0.40) 10. Signal-To-Noise Ratio and Contrast-To-Noise Ratio Measurement Signal-To-Noise Ratio should be equal or greater to 40 SNR Contrast-To-Noise Ratio should not vary by more than ±15% CNR 11. Viewbox Luminance and Room Illuminance Mammographic viewbox is capable of a luminance of at least 3000 cd/sq m (nit) Room illuminance (viewbox surface as seen by observer) is 50 lux or less Room illuminance (monitor surface) is 20 lux or less 12. Softcopy Workstation QC White level performance Black level performance Quality level performance Uniformity performance 2. 3. • FDA Forms MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY (GE Medical Systems, continued) Evaluation of Site's Technologist QC Program Site SenoScan Important PrePre-Survey Events •ACR Accreditation – Equipment Evaluation Forms MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY ? 2277390-100 Rev 3, 2001 ? 2277390-100 Rev 3 + Addendum 2354312-100, 2003 ? Mobile 2371698-100 Rev 0, 2003 ? Seno Advantage 2391082-100 Rev 1, 2003 1. 2. – Not required to wait for ACR response Report Date Survey Date Model Room ID Signature Medical Physicist's QC Tests GE QC manual version(s) at facility (check all that apply): MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital – Lorad MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital - Fischer Site Senographe 2000D Room ID Signature Frequency Daily Daily Weekly* Weekly Weekly Weekly Weekly Weekly Bi-weekly Bi-weekly Monthly Quarterly Quarterly Semi-annually Semi-annually PASS/FAIL MQSA Regs * Dry laser printer (daily if wet processor used) Medical Physicist's Recommendations for Quality Improvement – MQSA Certification Extension for GE Senographe DS • Site must already be screenscreen-film certified by FDA • Three things to submit to FDA: – Application of qualifications – Medical Physicist Survey – Printed ACR Phantom • Must have approval letter from FDA before clinical imaging can be performed 2 Important PrePre-Survey Events • Contact the site – things to confirm: – Site is aware of ACR or FDA application process – FFDM unit is operable – Review workstation is operable • Images can be transmitted – Laser printer works, can print mammo images, and hooked up to all RWS’ RWS’s – Discuss QC issues • Many QC failures result in stopping clinical imaging – ACR Phantom – do they have one onon-site? Important PrePre-Survey Events • Gather forms – Copy of ACR or FDA forms – Physics test forms • Ensure you have tests that are required by manufacturer • Gather test tools – Check required tests in manufacturer’ manufacturer’s manual – Artifact test tool – 1 or 2 inches of acrylic Important PrePre-Survey Events • Contact Manufacturer’ Manufacturer’s Service Engineers – Complete rere-calibration – Can they be present? • If not, how can they be contacted? – Is the system working properly? – Can the laser printer service engineer be present? Performing the Survey • Must perform manufacturer’ manufacturer’s tests • Turn off auto push and/or auto print – Remember to turn them back on • Order of tests is important • Use “Raw” Raw” of “Processed” Processed” images for testing – Lead sheet 3 Performing the Survey GE 2000D http://www.gemedicalsystems.com/services/repl_parts/documentation.html http://www.gemedicalsystems.com/services/repl_parts/documentation.html Raw Image Processed Image GE 2000D - Performing the Survey • Manufacturer’ Manufacturer’s tests – Mammography unit evaluation * – Flat field uniformity – Artifact evaluation * – AOP Mode and SNR Check – ACR Phantom and Contrastto-Noise Ratio (CNR) Check Contrast-to– MTF measurement – Collimation Assessment * – Evaluation of Focal Spot * GE Review Workstation *30 Days to Repair 4 GE 2000D - Performing the Survey • Manufacturer’ Manufacturer’s tests – Breast entrance exposure, average glandular dose, and reproducibility GE 2000D - Performing the Survey • FlatFlat-field uniformity – Test to ensure detector performance acceptable • Measures detector uniformity (signal & noise) • Measures bad pixels – Beam quality (HVL) * – kVp accuracy and reproducibility * – System automatically calculates pass/fail – Radiation output * – Viewing conditions check and setting – Monitor calibration * – Image quality – SMPTE pattern QAP – Analysis of RWS screen uniformity * *30 Days to Repair GE 2000D - Performing the Survey • Artifact evaluation – 1 inch acrylic phantom – Use clinical techniques – Image at each target/filter No grid or compression paddle GE 2000D - Performing the Survey • AOP Mode and SNR Check – Variable thicknesses of acrylic - 2.5, 4, 6 cm – Std, Auto – Evaluate: • Correct techniques? • Adequate SNR? Acrylic Thickness 2.5 cm TargetFilter Mo-Mo Selected kVp 27 kVp Selected mAs 20-60 – Review images at window width ~ 400 to 450 4.0 cm Mo-Rh 28 kVp 35-90 – Review artifact images on RWS, AWS, and printed film 6.0 cm Rh-Rh 32 kVp 35-90 – Image at each magnification mode 2.5 cm Acrylic 4.0 cm Acrylic Each “raw” raw” image must have a measured SNR of at least 50 6.0 cm Acrylic 5 GE 2000D - Performing the Survey GE 2000D - Performing the Survey • ContrastContrast-toto-Noise Test (CNR) • ACR Phantom Imaging – To examine consistency of CNR ratio – Manual technique & 3 auto measured over time – Manual technique modes – Use the raw image – Score the processed image – A control level is established – Acquisition workstation over 5 days – + 20% of baseline – Each monitor of the RWS Background ROI Mass ROI CNR = (Mean (Meanbackground - Meanmass)/SDbackground – Laser imager GE 2000D - Performing the Survey • MTF Measurement GE 2000D - Performing the Survey • Collimation – Mo/Mo – Rh/Rh MTF (%) = (Std. Dev.) x 222 / (Mean Dark ROI – Mean Light ROI) MTF (%) @ 2 lp/mm lp/mm > 58% MTF (%) @ 4 lp/mm lp/mm > 25% 6 GE 2000D - Performing the Survey • Focal Spot Evaluation GE 2000D - Performing the Survey • Beam Quality (HVL) •Remove Compression Paddle •Mo/Mo •Rh/Rh •Large & Small Spots (1.5 mag) mag) Note the lead sheet GE 2000D - Performing the Survey • Average Glandular Dose GE 2000D - Performing the Survey • kVp Accuracy – Measure using AOP system, or – Measure entrance exposure and calculate AGD 7 GE 2000D - Performing the Survey • Monitor Quality Control – Image Quality - SMPTE Pattern – Luminance Levels – RWS Screen Uniformity – Viewing Conditions Check High Contrast Line Pair Patterns 1 pixel/line Vert Horiz RWS SMPTE Patterns: 5% & 95% contrast boxes 10% contrast boxes High contrast contrast lineline-pair resolution Monitor Calibration Measure luminance at: 10 60 2 pixel/line Vert Horiz 120 180 255 4 pixels/line Vert Horiz Compare to baseline – must be within action limits 8 Analysis of Screen Uniformity Viewing Conditions Check 14. Viewing Conditions Map •Set monitor luminance to completely white (255) via menu • To ensure optimal viewing conditions angles to search for defects in glass or other nonnon-uniformities • Configure room for optimal viewing conditions surface of monitors - < 20 lux Monitor position: Room Lights: Desk Lights: Other Viewboxes: Other Alternators: Doors: Other: Left Monitor Right Monitor Ambient room illuminance (lux): Room Layout: • Record data on form • Post in reading room • Update if there is a room reconfiguration Fischer Date: Unit: Senographe 2000D ACR Site # or MAP#: Room Description: • Measure ambient room light at •Look at monitor at various Site: Room #: Action Limit: The ambient light levels must not exceed 50 lux. The measured value must stable over one minute, with a tolerance of + 5 lux over the measuring time with regards to the nominal lighting values. It the system fails the test, the problem must be fixed before further reviews are performed. Fischer Senoscan Some slides courtesy of Fischer Imaging and Idris Elbakri, Elbakri, Ph.D. 9 Fischer Senoscan Fischer Senoscan - Performing the Survey • Manufacturer’ Manufacturer’s tests – X-ray field size and Chest wall missed tissue – Compression paddle alignment – kVp accuracy – Linearity, reproducibility, and accuracy – Beam Quality (HVL) – Dosimetry – average glandular dose – Phantom image acquisition Fischer Senoscan - Performing the Survey • Manufacturer’ Manufacturer’s tests – Image quality – System resolution/scan speed uniformity – Flat field test – Geometric distortion and resolution uniformity – Automatic decompression control Fischer Senoscan - Performing the Survey • Flat Field and Artifact Evaluation – – – – – – – Use 4 cm acrylic Exposure resulting in 1000 ADU’ ADU’s Display “raw” raw” image No Artifacts at WW > 800 Compute deviation between corner and center means Automatic in new software release Must be within + 20% of center ROI – System artifacts – Image display monitor(s) check – Tech Review – Image viewing room illuminance 10 Fischer Senoscan - Performing the Survey • Collimation Fischer Senoscan - Performing the Survey • Compression Paddle Alignment Fischer Senoscan - Performing the Survey • Collimation Fischer Senoscan - Performing the Survey • Collimation – Error between fieldfield-size markers and image receptor must be less than 2% of SID – Missed chest wall tissue less than 8.5 mm – Compression paddle: distance between image receptor at chest wall and inside of edge of paddle must be < 8.5 mm 11 Fischer Senoscan - Performing the Survey • kVp Accuracy Fischer Senoscan - Performing the Survey • Phantom Image Acquisition Test and Image Quality – Select Techniques to give 1000 ADU – Performed invasively through special BNC connectors – Compare background mean, StdDev, StdDev, and ADU level difference to baseline values – NonNon-invasive method under evaluation – Score : 4 fibers, 3 speck groups, and 3 masses Fischer Senoscan - Performing the Survey • Mean Glandular Dose – 30 kVp – mA between 150 to 190 to result in 1000 ADU – mRad/R mRad/R conversion table provided in manual Fischer Senoscan - Performing the Survey • HalfHalf-Value Layer – Ion chamber affixed 5 cm to the right of center and close to chest chest wall – Position paddle 4.2 cm above chest wall – kVp = 30 mA = 100 – Criterion: HVL > 0.33 mm – Watch collimation! 12 Fischer Senoscan - Performing the Survey Fischer Senoscan - Performing the Survey • Exposure Linearity and Reproducibility • System Resolution/Scan Speed Uniformity – Procedure: – Purpose: Ensure correct detector/beam alignment and constant • Center ion chamber on breast support near chest scanning speed wall with paddle in the beam • Measure exposure at specified techniques – Equipment: line pair phantom • Calculate: – Perform imaging with scan lines parallel and perpendicular to • Output = Average exposure reading/mA reading/mA • Linearity = (A(A-B)/(A+B) < 0.08 • Reproducibility = StdDev/Avg < 0.035 for each technique scan direction – Normal and high resolution modes Fischer Senoscan - Performing the Survey Normal Mode: 7 lp/mm: lp/mm: > 10% modulation Fischer Senoscan - Performing the Survey •Geometric Distortion –Use a 40 mesh Cu screen HighHigh-Res Mode –Acquire image at low technique 11.1 lp/mm: lp/mm: > 5% modulation –Visually inspect for distortions or blurring 13 Fischer Senoscan - Performing the Survey • Viewing Room Luminance – Need a photometer Lorad Selenia – Monitors off – Measure luminance at the monitor screen and 50 cm away – Illuminance must be < 50 lux Lorad Selenia Some slides courtesy of Lorad and Nikolaos A. Gkanatsios, Gkanatsios, Ph.D. Lorad Selenia 14 Lorad Selenia - Performing the Survey Lorad Selenia - Performing the Survey • Manufacturer’ Manufacturer’s tests • Manufacturer’ Manufacturer’s tests – Unit assembly evaluation * – kVp accuracy and reproducibility * – Artifact evaluation * – Beam quality— quality— HVL * – Phantom image quality – Breast Entrance exposure and average glandular dose – Evaluation of system resolution – Radiation output rate * – SignalSignal-toto-Noise and ContrastContrast-toto-Noise Measurements – Viewbox luminance and room illuminance * – Collimation assessment * – Softcopy Workstation QC *30 Days to Repair *30 Days to Repair Lorad Selenia - Performing the Survey • Collimation Assessment – Use coin techniques as described in ACR Manual • Test 24x29 cm detector mode • Test 18x24 cm detector mode – X-Ray field to light field coincidence – X-Ray field to image receptor alignment Lorad Selenia - Performing the Survey • Artifact Evaluation – 4 cm acrylic block – Mo/Mo – Mo/Rh Mo/Rh – Large & Small Spot – Evaluate for artifacts at WW ~ 250 – Print films – check printer – Compression Paddle to Image Receptor Alignment 15 Lorad Selenia - Performing the Survey • kVp – – Described in the 1999 ACR QC Manual • HVL – – Described in the 1999 ACR QC Manual • Phantom Image Quality Lorad Selenia - Performing the Survey • Breast Entrance Exposure and Average Glandular Dose – Cover detector for protection – lead sheet – Technique set to clinically image average breast • 28 kVp, kVp, 65 mAs, mAs, Mo/Mo – Calculate dose - < 3.0 mGy – 28 kVp, kVp, 65 mAs, mAs, Mo/Mo – Until AEC system implemented – Print film – measure OD and Contrast – Score on each SCW (Soft Copy Workstation) • 5 fibers • 4 speck groups • 4 masses Lorad Selenia - Performing the Survey • Radiation Output Rate – Cover detector for protection – lead sheet – Technique set to clinically image average breast • 28 kVp, kVp, 320 mAs, mAs, Mo/Mo – Output Exposure - > 800 mR/sec mR/sec Lorad Selenia - Performing the Survey • Evaluation of System Resolution – 5-15 lp/mm lp/mm Test Pattern – 4-cm Attenuation Block – Pattern at 4545-Degree Angle to the Detector • The system limiting spatial resolution must be > 7 lp/mm lp/mm 16 Lorad Selenia - Performing the Survey • SNR and CNR Measurements Lorad Selenia - Performing the Survey • Viewbox Luminance and Room Illuminance – Desribed in the 1999 ACR QC Manual – SNR at least equal or greater than 40 • SNR = (Mean (MeanBkgd-DCoffset)/SDBkgd DCoffset = 50 – Establish CNR during acceptance testing • SoftCopy Workstation QC – Use supplied photometer and run monitor QC software • White level = 300 cd/m2 for Barco MG521 & 400 cd/m2 for Barco MG521M – Warning level = + 3% Tolerance level = + 6% Recalibrate Tolerance level = 1.0% Recalibrate • Black level = 0 cd/m2 • CNR = (Mean (MeanBkgd-MeanDisk)/SDBkgd – Warning level = 0.5% • Quality Level Performance – checks full monitor calibration automatically – CNR should stay within ±15% of measurement obtained during – Warning level = 5% Tolerance level = 10% Recalibrate • Uniformity Performance – minimize nonnon-uniformities away from center of display acceptance testing GE Seno DS – Warning level = 10% Tolerance level = 15% Recalibrate GE Seno DS Some slides courtesy of GE and Vince Polkus and Marcia Hill 17 GE Senographe DS Seno Advantage RWS GE Senographe DS Quality Assurance Plan • • • • • • Automated Evaluative Procedure Patented Technology Tracks IQ Over Time • Spatial Resolution • Small Signal Contrast • Dynamic Range • Resolution Uniformity • Distortion • Other Run by Technologist Pass/Fail Result In-Site Interactive – Remote Corrections – Automatic Service Dispatch GE Image Quality Signature Test (IQST) Phantom QAP Not a calibration but a process that maximizes Senographe digital image quality consistency Fuji CR GE Senographe DS QA Phantom Overview Uniform area for noise power spectrum measurement Step wedge for contrast measurement Mesh for resolution uniformity measurement Edge object for MTF measurement CR Reader Fuji FCR 5000 MA CR QC Workstation CR Cassettes Rulers for measuring distance of detector from chest wall edge of Bucky Tabs to position IQST with respect to chest wall edge of image receptor. Rails to position the IQST leftleft-toto-right 24 x 30 cm 18 x 24 cm 18 Fuji CR Fuji FCR 5000 MA Performing the Survey • Manufacturer’ Manufacturer’s tests – Fuji – Physics tests unique to Fuji CR FFDM • CR reader sensitivity (“ (“S” number) • CR reader shading correction • Imaging plate fogging test • Verification of AEC with CR cassettes on each unit that is tested Siemens Mammomat NovationDR Siemens - Performing the Survey • Manufacturer’ Manufacturer’s tests – AWS Monitor and Viewing Conditions – Chest Wall and Missed Tissue – Collimator – Compression Plate Position – Spatial Resolution – Phantom Image Quality – Mechanical Inspection – Detector Uniformity – Radiation Safety – (Optional in U.S) 19 Siemens - Performing the Survey Siemens - Performing the Survey • Manufacturer’ Manufacturer’s tests – kVp Accuracy and Reproducibility – HVL – Mean Glandular Dose • Manufacturer’ Manufacturer’s tests – Film Printer – AEC Stability, Reproducibility & SNR – Viewing Conditions – Ghost Image – Illuminance – SNR and CNR – AEC Thickness Tracking – Monitor Constancy – Pixel Correction – Detector Calibration Sectra Artifacts 20 Performing the Survey Performing the Survey • Artifact evaluation - windowing • Artifact evaluation – Contact Mode GE Mo/Mo Performing the Survey Mo/Rh Rh/Rh Performing the Survey • Artifact evaluation – Mag mode 1.5 Mo/Mo Mo/Rh • Artifact evaluation – Mag mode 1.8 Rh/Rh Mo/Mo Mo/Rh Rh/Rh 21 Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation 22 Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation 23 Performing the Survey • Artifact evaluation Performing the Survey • Artifact evaluation 24 Performing the Survey • Artifact evaluation Artifact at edge of image receptor Cause: hole in the graphite cover of the digital detector 25 There Are Currently Seven FDAFDA-Approved Laser Imagers for Digital Mammography OD Requirements for HiHi-Resolution Laser Imagers • Agfa LR5200 Laser Imager (Wet Chemistry) • Agfa DS4500M Dmax > 3.5 OD • Kodak 8600 Laser Imager • Kodak 8610 Laser Imager • Kodak 8900M MidMid-density > 1.5 OD • Fuji Drypix 7000 • Fuji Drypix FMFM-DP L Laser Processor QC Kodak daily sensitometry Base + Fog • Does the background match? Density Difference – OD closest to 2.20 minus OD closest but not less than 0.45 MidMid-density – step closest to but not less than 1.20 Action Limits: MD & DD + 0.15 OD Dmax RWS Clinical Image Check B+Fog = 0.03 • Is the background dark enough? • Does the dense tissue area match? • Is the dense tissue light enough? • Is the contrast adequate? Dmax + 0.25 26 RWS Clinical Image Check RWS Clinical Image Check RWS Clinical Image Check RWS Clinical Image Check 27 Key TakeTake-home Points • Obtain proper handshands-on training Key TakeTake-home Points • Review workstation monitors – look at the clinical images! – Do they match? • ACR & FDA applications and forms • Turn off and on auto print and/or auto push • Artifacts – most problems can be seen on this test • Lead sheet protecting detector for Focal Spot, HVL & kVp – Appropriate dark and light levels • Do all work on correct images – raw vs. processed • Can be archiving physics test images on one workstation while working on another • Laser Printer – Dmax at least 3.5 OD • Take your time and use your professional judgement – MidMid-density about 1.5 OD Thank You 28