FLAT PANEL FLUOROSCOPY ACCEPTANCE TESTING AND QUALITY CONTROL by EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY & NEW YORK –PRESBYTERIAN HOSPITAL NEW YORK, NEW YORK Dr. Ed Nickoloff, Columbia University ACCEPTANCE TESTING GOALS • • • • • • • • PRIOR TO 1st CLINICAL USAGE VERIFY PURCHASE ORDER CHECK ARCHITECTURAL DESIGN & FLOW EVALUATE RADIATION SHIELDING, SCATTER LEVELS & RADIATION PROTECTION MOST COMPREHENSIVE TESTING MAKE REPAIRS & ADJUSTMENTS UNDERSTAND ABC OPERATION ESTABLISH BASELINE VALUES FOR QC Dr. Ed Nickoloff, Columbia University ACCEPTANCE TESTING GOALS • COMPARISON OF MEASURED IMAGE QUALITY TO MANUFACTURER’S SPEC’S • ITEMS NOT ROUTINELY TESTED – – – – MECHANICAL SYSTEMS LASER CAMERAS ARCHIVEAL STORAGE DEVICES ELECTRICAL SAFETY • REGULATORY COMPLIANCE • RECOMMENDATIONS TO STAFF • RADIATION DOSE CHARTS Dr. Ed Nickoloff, Columbia University ROUTINE QC TESTING • LIMITED NUMBER SPOT CHECKS • INTENDED TO IDENTIFY DEGRADATIONS OR CHANGES • ENSURE PATIENT & STAFF RADIATION SAFETY • FOLLOW-UP TO REPAIRS & UPGRADES • NEEDED FOR REGULATORY COMPLIANCE • SOME TEST SIMILAR TO ACCEPTANCE – FEWER PARAMETERS EVALUATED – COMPARED TO BASELINE VALUES Dr. Ed Nickoloff, Columbia University 1 THE FIELD – of – VIEW (FoV) GAME? GE FoV (cm) SIEMENS FoV (cm) 25 20 20 17 16 15 12 GE measures horizontal & vertical. Siemens measures diagonal. Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University NON-UNIFORMITIES • IMAGE INTENSIFIERS + TV ARTIFACTS & IMAGE UNIFORMITY – VIGNETTING, DISTORTION, PHOSPHOR BURNS, DIRT, SATURATION, GRID PROBLEMS • FLAT PANEL DETECTORS – BAD DETECTORS, NON-UNIFORM GAIN, CALIBRATION ERRORS, DIRT, SATURATION, GRID PROBLEMS, PHYSICAL DAMAGE, SOFTWARE PROBLEMS, IMAGE PERSISTENCE Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University 2 SYSTEM SPATIAL RESOLUTION • DEPENDS UPON GEOMETRICAL MAG ( M ), FOCAL SPOT ( f )& INHERENT IMAGE SPATIAL RESOLUTION ASSESSMENTS RECEPTOR RESOLUTION ( ir ) • LESS THAN THE LOWEST OF: – Mx ir – M / [ (M - 1) x f ] ir DOMINATES • AT HIGH “M”, FOCAL SPOT BLUR DOMINATES • AT LOW “M”, Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University SUMMARY OF BAR PATTERN PLACEMENT CAVEATS OF SPATIAL RESOLUTION MEASUREMENTS • PARTIAL PIXEL SHIFT EITHER VERTICAL OR HORIZONTAL CAN ALTER MEASURED RESOLUTION – RESOLUTION BEST = 1 / [ 2 x PIXEL SIZE] – OR, ZERO FOR 50% PIXEL SHIFT – OR, ANYWHERE IN BETWEEN • 45 DEGREE ALIGNMENT OF BAR PATTERN INCREASES SPACING BY 1.41 x’s – PARTIAL PIXEL SHIFT STILL A PROBLEM Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University 3 SUMMARY OF NOISE EFFECTS M TF VALUE MTF GRAPH & SPATIAL RESOLUTION LIMIT 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 • AS IMAGE NOISE INCREASES THE LIMITING HIGH CONSTRAST SPATIAL RESOLUTION DECREASES NO NOISE 5 % NOISE MAGE NOISE I 0 0.5 1 1.5 2 2.5 3 SPATIAL FREQUENCY (LP/mm) – QUANTUM MOTTLE IS A KEY FACTOR IN IMAGE NOISE – RECORD MODE (CINE) WILL HAVE LESS QUANTUM MOTTLE THAN FLUORO MODE WHICH MIGHT IMPROVE RESOLUTION – ABC DOSE MODES WILL INFLUENCE RESULTS Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University 80 kVp X-RAY SPECTRA WITH DIFFERENT COPPER FILTRATION EFFECTS OF FILTRATION AND kVp UPON SPATIAL RESOLUTION 7.00E+04 #X-RAYS / keV 6.00E+04 5.00E+04 4.00E+04 3.00E+04 2.00E+04 1.00E+04 0.00E+00 0 10 20 30 40 50 60 70 80 X-RAY ENERGY (keV) 0 mm Cu 0.3 mm Cu 0.6 mm Cu Dr. Ed Nickoloff, Columbia University 0.9 mm Cu • SOME NEW FLUORO UNITS USE CONSIDEABLE FILTRATION IN X-RAY BEAM • HIGH kVp AND FILTRATION REDUCE INHERENT CONTRAST OF BAR PATTERN • 0.10 mm LEAD BAR PATTERN OR THICKER WOULD BE BEST FOR SPATIAL RESOLUTION • SCATTERED RADIATION IS MORE PENETRATING & WILL AFFECT SHIELDING OR ROOM & LEADED APPAREL Dr. Ed Nickoloff, Columbia University 4 X-RAY BEAM QUALITY (HVL) HVL MINIMUM REGULATORY HVLmin > 3.50 (kVp/100) + 0.08 in mm aluminum • kVp & FILTER MAY BE AUTOMATIC • HVL DEPENDS UPON kVp, CONTOUR & XRAY FILTER • FILTER MAY BE DIFFERENT IN FLUORO & CINE` • TYP. HVL=3.5 - 9.0 mm @ 80 kVp & CAN BE HIGHER IEC STANDARD (REGULATORY FOR IR??) HVL MAXIMUM EMPIRICAL HVL max < 4.0 to 9.0 mm Al @ 80 kVp MODERN ANGIO ROOMS HAVE VARIABLE FILTRATION WHICH CAN BE UP TO 0.9 mm Cu Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University COMPARISON OF HVL vs. ADDED FILTRATION MEASURED HVL ( mm Al EQ. ) 9 8 SPATIAL RESOLUTION AT INPUT SURFACE OF IMAGE RECEPTOR 7 6 5 4 3 @ 80 kVp 2 1 0 0 0.2 0.4 0.6 0.8 1 ADDED FILTRATION ( mm COPPER ) SIEMENS Dr. Ed Nickoloff, Columbia University GE Dr. Ed Nickoloff, Columbia University 5 SUMMARY ABOUT INHERENT SPATIAL RESOLUTION SPATIAL RESOLUTION (LP/mm) COMPARISON OF INHERENT DETECTOR SPATIAL RESOLUTION 3.8 3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 2 10 12 14 16 18 20 22 24 FIELD-of-VIEW (LP/mm) Axiom Artis I.I. Axiom Artis Flat Panel Innova Flat Panel Dr. Ed Nickoloff, Columbia University • SPATIAL RESOLUTION OF IMAGE INTENSIFIER + T.V. IMPROVES WITH SMALL FoV’S • FLAT PANEL RESOLUTION DOES NOT INCREASE WITH FoV CHANGES • SPATIAL RESOLUTION OF FLAT PANEL SYSTEMS IS LESS THAN IMAGE INTENSIFIER PLUS T.V. SYSTEMS IN SMALL FoV’S • SPATIAL RESOLUTION OF FLAT PANEL SYSTEMS IS BETTER THAN IMAGE INTENSIFIER + T.V. IN LARGE FoV’S Dr. Ed Nickoloff, Columbia University SUMMARY ABOUT INHERENT SPATIAL RESOLUTION • FLAT PANEL RESOLUTION IS APPROX. – 1 / [ 2 x PIXEL SIZE (mm)] • GE PIXEL = 200 T & SIEMENS PIXEL = 185 T • ESTIMATE SPATIAL RESOLUTION FLAT PANELS – 2.5 LP/mm & 2.7 LP/mm – CORRESPONDS TO MEASUREMENTS SPATIAL RESOLUTION AT DISTANCES AWAY FOR INPUT SURFACE OF IMAGE RECEPTOR • ABC MODES, RADIATION DOSE, SOFTWARE AND ANGULATION HAVE AN EFFECT Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University 6 COMPARISON OF SPATIAL RESOLUTION IN MEDIUM FoV ( I.I.= 17cm, SFPD=20cm & GEFPD=17cm) COMPARISON OF SPATIAL RESOLUTION FOR SMALL Fov (I.I. = 13 cm, SFPD = 16 cm & GEFPD = 13 cm) 4 SPATIAL RESOLUTION (LP/mm) SPATIAL RESOLUTION (LP/mm) 3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 3.8 3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 2 2 0 5 10 15 20 25 30 35 40 45 0 5 DISTANCE FROM ENTRANCE (cm) Axiom Artis I.I. Calc Axiom Artis Flat Panel Calc Dr. Ed Nickoloff, Columbia University Innova Flat Panel Calc Axiom Artis I.I. Calc I 10 15 20 25 30 DISTANCE FORM ENTRANCE (cm) Axiom Artis Flat Panel Calc S 35 40 45 Innova Flat Panel Calc G Dr. Ed Nickoloff, Columbia University EFFECTS OF DISTANCE ON SPATIAL RESOLUTION • DISPLACING THE PATTERN AWAY FROM ENTRANCE SURFACE IMPROVES RESOLUTION UP TO 20 – 25 cm – SID=100 cm USED MAG FACTOR ~ 1.30 – FLUORO FOCAL SPOTS ~ 0.7 mm effective FLUORO vs. CINE (RECORD MODE) SPATIAL RESOLUTION • LONGER DISTANCE RESULT IN FOCAL SPOT BLUR DEGRADING RESOLUTION • FLAT PANEL SYSTEMS BETTER AT LARGE FoV’S & BOTH SYSTEMS TESTED SIMILAR • IMAGE INTENSIFIER SYSTEM BETTER AT ALL SMALL FoV’S Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University 7 SUMMARY OF FLUORO vs CINE SPATIAL RESOLUTION RESOLUTION COMPARISON OF FLUORO vs. RECORD MODE FOR SIEMENS AXIOM ARTIS FLAT PANEL 4 MEASURED SPATIAL RESOLUTION (LP?mm) 3.8 3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 2 0 5 10 15 20 25 30 35 40 DISTANCE FORM INPUT SURFACE (cm) FLUORO 16 cm Fov RECORD 16 cm FoV Dr. Ed Nickoloff, Columbia University SUMMARY OF FLUORO vs. CINE SPATIAL RESOLUTION • FOR IMAGE INTENSIFIER UNITS WITH DIGITIZED TV – FLUORO MAY BE 1023 LINE & CINE 512 LINES – FLUORO RESOLUTION IS MUCH BETTER THAN CINE • FOR IMAGE INTENSIFIER UNITS WITH CINE FILM – CINE IS BETTER THAN FLUORO AT I.I. BECAUSE IT IS NOT LIMITED BY TV LINES & BANDPASS – AT DISTANCES, FOCAL SPOT BLUR DOMINATES – CINE FILM HAS BETTER RESOLUTION THAN RECORD MODE ON FLAT PANEL SYSTEMS Dr. Ed Nickoloff, Columbia University • FLAT PANEL UNITS AT INPUT SURFACE – RECORD MODE (CINE) USE MORE RADIATION – RECORD MODE HAS LESS Q.M. – RECORD MORE IS A LITTLE BETTER • FLAT PANEL UNITS DISPLACED FROM INPUT SURFACE – RECORD MODE HAS LARGER FOCAL SPOT & MORE FOCAL SPOT BLUR – FOR FLAT PANEL UNITS FLUORO RESOLUTION BETTER AT LARGE DISTANCES – GE FLAT PANEL PROBABLY USED SMALL FOCAL SPOT IN RECORD Dr. Ed Nickoloff, Columbia University TEMPORAL RESOLUTION & PERSISTANCE • 6 WIRES 0.13 - 0.56 mm DIAMETER AT 30 RPM • FLUORO & CINE`/ DSA • STATIONARY vs. ROTATING • WATER PHANTOM SIMULATES PATIENT ATTENUATION • 2 - 3 SPOKES VISIBLE MOVING Dr. Ed Nickoloff, Columbia University 8 TYPICAL PATIENT RADIATION DOSES MEASUREMENTS OF PATIENT ENTRANCE RADIATION DOSES I.I. FIXED 30 cm ATTENUATION MATERIAL T RADIATION DETECTOR TABLE SID = 100 cm Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University CINE DOSE RATES FOR DIFFERENT FoV's & THICKNESS--- SIEMENS AXION ARTIS dBC SUMMARY OF PATIENT ENTRANCE RADIATION EXPOSURE RATES 100 NORMAL, 15 FPS CINE • EACH VENDORS HAS MULTIPLE RADIATION DOSE PROGRAMS!!! • GE HAS: EXPO SU RE(R/M IN ) 10 1 0 5 10 X-RAY TUBE 15 20 25 – IQ+, IQ STAND., SMART IQ, RDL+, RDL STAND. – RANGE OF Max / Min = 3 or 4 : 1 – “LOW” is about 50% of “NORMAL” • SIEMENS HAS: – MINUS = -50%; NORMAL; PLUS = +50% 0.1 ACRYLIC THICKNESS (cm) 25 cm FoV 20 cm FoV Dr. Ed Nickoloff, Columbia University 16 cm FoV • FOR BOTH FLAT PANELS, EACH CINE FRAME = 4 TO 10 FLUORO FRAMES Dr. Ed Nickoloff, Columbia University 9 SUMMARY OF PATIENT ENTRANCE RADIATION EXPOSURE RATES • USEFUL TO PLOT “LN( EXPOSURE RATE)” VERSES THICKNESS • HVL FOR TISSUE 3.7 TO 4.4 cm COMPARING 17 cm FoV, IN THE 15 – 25 cm THICKNESS ACRYLIC: – SIEMENS FLAT PANEL = 2x’s I.I. WITH DIGITIZED TV FOR RECORD – GE FLAT PANEL (“RDL+,NORMAL”) ABOUT SAME EXPOSURE RATES AS SIEMENS FLAT PANEL IN FLUORO & LESS IN CINE • VARIOUS MODES ALLOW A LARGE RANGE Dr. Ed Nickoloff, Columbia University ABC CONTROLS Dr. Ed Nickoloff, Columbia University I.I. INPUT EXPOSURE RATES (IIIER) FLUORO ABC CONTROL OF FLAT PANEL SYSTEMS TUBE POTENTIAL (kVp) 110 100 90 80 70 60 50 0 5 10 15 20 25 THICKNESS ACRYLIC (cm) GE INNOVA SIEMENS 0.3 mm Cu SIEMENS 0.9 mm Cu SIEMENS 0.2 mmCu Dr. Ed Nickoloff, Columbia University SIEMENS 0.6 mm Cu 30 • IIIER IS HOW X-RAY SERVICE ADJUSTS RADIATION LEVELS • GRID REMOVED • FLUORO < 100 T R / SEC* • CINE` < 20 T R / FRAME* • DSA < 1000 T R / FRAME* • * 9 INCH (23 cm) FoV I.I. DETECTOR NO GRID ATTENUATION X-RAY TUBE Dr. Ed Nickoloff, Columbia University 10 SUMMARY OF INPUT RADIATION TO DETECTORS • ALL SYSTEM MEET < 100 µR/ sec FLUORO & < 20 µR / frame IN RECORD FOR 23 cm FoV • GE HAS ABOUT SAME RAD LEVELS IN 15 pps & 30 pps FLUORO • IN LARGE FoV’s LEVELS ABOUT SAME FOR ALL TYPES OF FLUORO & RECORD • THERE ARE DIFFERENCE BETWEEN SETTINGS AND VENDORS Dr. Ed Nickoloff, Columbia University LOW CONTRAST IMAGING • STANDARD PENETRAMETER WITH LOW CONTRAST HOLES TOO EASY • kVp & FILTER DEPENDENT • C-D PHANTOMS USEFUL FOR RELATIVE MEASUREMENTS Dr. Ed Nickoloff, Columbia University CONTRAST RATIO (RC) • MEASURES DEGRADATION DUE TO LIGHT & PHOTON SCATTER • CAN DEGRADE VISIBILITY OF SMALL VESSELS • USE LIGHT ON MONITOR • RC > 60 : 1 Dr. Ed Nickoloff, Columbia University CONTRAST OF HOLES C0 = EXP [ - µ ( T) ] VISIBILITY OF HOLES C0 * MTF ( ) = K / SNR where (1 / 2D) < < (1 / D) Dr. Ed Nickoloff, Columbia University 11 DATA FROM “C-D” PHANTOM GE FPD FLUOR. 55%@ 20cm 60%@ 17 cm 57%@ 15 cm 62%@ 12 cm SIEMENS FPD CINE FLUOR. 55% @ 62% @ 20 cm 25 cm 60% @ 62% @ 17 cm 20 cm 63% @ 62% @ 15 cm 16 cm 73% @ 12 cm Dr. Ed Nickoloff, Columbia CINE 71% @ 25 cm 71.5% @ 20 cm 75% @ 16 cm MONITOR CRITERIA • MONITORS MUST HAVE A BALANCED GRAY SCALE • 5% CONTRAST MUST BE VISIBLE • MINIMUM BRIGHTNESS (BLACK LEVEL) <1 TO 2 nit (Cd / m2 ) • MAXIMUM BRIGHTNESS >400 TO 500 nit • NO DEFECT & DISTORTIONS • TINT OR COLOR DIFFERENCES University Dr. Ed Nickoloff, Columbia University OTHER PHYSICS TESTS OTHER PHYSICS TEST • X-RAY EQUIPMENT TEST – kVp, WAVEFORM, HVL, mA LINEARITY, FOCAL SPOT SIZE, ETC. • REGULATORY TESTS – MAX FLUORO EXPOSURE RATE, COLLIMATION, SAFETY CHECKS, ETC. • MECHANICAL CHECKS – ISOCENTER, CENTRAL BEAM, MOTION, ALIGNMENT, COLLISION SENSOR, ETC. • “DAP” METER ACCURACY • ARCHIVING & PACS TRANSMISSION • ELECTRICAL SAFETY – GROUNDING, POLARITY, EMERGENCY POWER, CUT-OFF SWITCHES, ETC. • GRID EVALUATION • SID TRACKING • RADIATION PROTECTION SURVEY Dr. Ed Nickoloff, Columbia University Dr. Ed Nickoloff, Columbia University 12 THE END! Dr. Ed Nickoloff, Columbia University 13