An Improved Microprobe Using Direct Undulator Radiation M. R. Weiss1 , V. Wüstenhagen2 , C. Heske1 , R. Fink1 , E. Umbach1,2 1 Universität Würzburg, Experimentelle Physik 2, Am Hubland, D-97074 Würzburg, Germany 2 Universität Stuttgart, 4. Physikalisches Institut, Pfaffenwaldring 57, D-70550 Stuttgart, Germany Abstract. The concept of a microspectrocope using direct undulator radiation is described. This instrument utilizes the quasi monochromatic beam from the first undulator at BESSY without monochromatization. Thus X-ray induced Auger spectroscopy as well as photoemission using the 2 eV wide first harmonic of the undulator are possible. Spatial resolution is achieved by an adjustable aperture in front of the sample yielding a diffraction limited resolution of 3–4 µm. An improved concept using an ellipsoidal mirror is described resulting in a spatial resolution in the submicrometer range. As an example a microanalysis of Cu(InGa)Se2 , a material used for thin film solar cells, is briefly discussed. 1 Introduction Microscopes which provide spectroscopic information are presently under development because of their benefits in materials research. Especially photoelectron spectroscopy has been implemented since one can analyze not only the composition but also the chemical state of a sample surface. The spatial resolution is obtained either by imaging the emitted photoelectrons or by scanning focused photons over the surface. Both methods, scanning or imaging, have specific advantages and disadvantages. The imaging method is very well suited for surveys and spatially evolving patterns since it gives an image of the surface for a certain electron energy in short time. Experimental setups based on this concept have reached the highest spatial resolution today [1, 2, 3, 4]. This is mostly due to the fact that in the past many efforts were concentrated on the imaging technique in electron microscopy. The scanning method, on the contrary, allows to direct all photons onto the spot in question, thus providing faster spectroscopic analysis of this spot with better signal-to-noise ratio. It is, in principle, as fast as imaging provided that the time for the scanning motion is short. In the scanning mode one can also use additional analysis techniques, for example mass spectroscopy of photo-desorbed atoms [5]. X-Ray Microscopy and Spectromicroscopy Eds.: J. Thieme, G. Schmahl, D. Rudolph, E. Umbach c Springer-Verlag Berlin Heidelberg 1998 III - 16 M. R. Weiss et al. The focusing of a UV- or X-ray spot can be achieved by Fresnel micro zoneplates [6], by normal incidence X-ray mirrors, e.g. in Schwarzschild geometry with multilayer coatings [7, 8], or by grazing incidence, (often) non-spherical X-ray mirrors [9, 10]. The photon sources for these microscopes are dipole magnets or insertion devices at synchrotron radiation facilities. Present laboratory sources do not provide the flux and brightness necessary for such low numerical aperture optics. For high spectral resolution X-ray monochromators are used which reduce the flux considerably. To overcome the shortcomings of other concepts, namely low flux, no adjustable photon energy and no availability of other methods of analysis, we have developed a microspectroscope of the scanning type which is described in the following. For very high flux and hence short measurement times we use direct undulator light. This source is quasi-monochromatic by interference effects of its periodic magnet structure. Since the linewidth is rather broad (here: ∆E ≈ 2eV ), this microspectroscope is best suited for X-ray induced Auger electron spectroscopy (XAES) for which monochromaticity is not required. Still, X-ray photoemission (XPS) analysis is possible, with reduced energy resolution. The specific advantages of this concept which we named Photon Induced Scanning Auger Microscope (PISAM) are discussed in more detail in section 2. For first tests, focusing was achieved by a simple aperture which demonstrated the possibilities of the concept and yielded first results. This apparatus and one example of an application are also described in section 2. For improved spatial resolution a focusing optics based on two grazing incidence mirrors was designed (see Fig. 1). Electron-analyzer Storage Ring eSample Wiggler/Undulator Condenser Mirror Field Aperture Focussing Mirror Fig. 1. Scheme of the new PISAM II concept. The condenser mirror increases the flux at the field aperture and the ellipsoidal focusing mirror provides the high spatial resolution. An Improved Microprobe Using Direct Undulator Radiation III - 17 Although the highest possible spatial resolution has not yet been demonstrated for a grazing incidence ellipsoidal mirror it was chosen because it allows to focus a broad range of photon energies. Thus the proper conditions can be selected by the experimentalist, e.g. to gain higher count rates or additional information by resonant spectroscopy. The design considerations and first results of the improved setup can be found in section 3. 2 2.1 The First Apparatus: PISAM-1 Description of the Setup and Its Performance The first experiment using this concept was conducted with a set of apertures which reduced the direct undulator beam down to a width of approximately 4 µm. The sample was mounted on a scanning stage motorized by steppermotors (1.25 µm per step). The excited photoelectrons were detected by a VG CLAM2 spherical sector analyzer. A preparation chamber with standard equipment (sputter gun, mass spectrometer, etc.) and a sample transfer system completed the apparatus [11, 12, 13]. We performed several experiments to characterize the potential of this simple setup [14, 15, 16]. Measurements of microstructured samples demonstrated a photon energy-dependent spatial resolution of 15 µm at 45 eV and 3-4 µm at 280 eV. The accessible photon energies ranged from 40 eV to 550 eV covering the absorption edges of sulfur, carbon, nitrogen, and oxygen. The surprisingly good resolution regarding the simple setup and the diffraction limit is partly due to the proximity of the sample to the aperture (12 mm). The resulting diffraction is of the Fresnel type which leads to a smaller spot than expected by Fraunhofer diffraction calculations [13, 14]. One might object that the spatial resolution of conventional electron-induced Auger microscopes (SAM) is well beyond this resolution (typically 50-200 nm, even 5 nm reported [17]), but the comparison of spatial resolutions leaves out other important aspects. SAM resolutions of this kind can only be reached by highly accelerated electrons (10 keV up to several 100 keV). Unfortunately, these electrons contribute to a large background of inelastically scattered electrons which reduce the signal-to-background ratio (S/B) considerably. For good spectroscopic analysis longer sampling times are hence required. Additionally, sensitive adsorbates and samples cannot withstand the required high fluxes and energies in SAM instruments. Photons as excitation source have the advantage that much less damaging secondary electrons are produced, and hence that less background and a better S/B ratio result. This can even be improved by tuning the photon energy to a resonance or threshold which also leads to additional spectroscopic information [14]. Furthermore, we performed experiments to utilize the microanalytical potential of this setup (for an example see the next section). Additionally, we tested microfabricational applications, for example writing of microwires either by depositing molybdenum from an organometallic precursor [15] or by photolytic polymerization of condensed monothiophene[14, 16]. III - 18 2.2 M. R. Weiss et al. Analysis of Cu(In,Ga)Se2 Thin Films for Solar Cells Thin CuInGaSe2 (CIGS) films are very promising absorber materials for low cost and high efficiency solar cells. A typical cell consists of a soda lime glass substrate, a 1µm thick Mo back-electrode, a 1-2µm CIGS absorber, a CdS buffer layer, and a ZnO top-electrode. The efficiency of such a solar cell is found to correlate with the XPS-derived sodium content segregating from the glass substrate through the Mo back-electrode into the CIGS absorber film [18]. A laser-cutting technique is applied on the Mo electrode for microstructuring and connecting the single absorber cells in a module. Because this laser-cut drastically influences crystal growth and segregation chemical microanalysis of the CIGS film on top of this microstructure and its correlation with the electrical performance of a solar cell in operation is of particular interest. An microscopic image of the vicinity of a laser-cut in the Mo back-electrode (width ≈30 µm) written prior to CIGS deposition and formation is shown in Fig. 2 a). The film has a changed morphology near the cut indicated by differ- a) b) Counts 190 m In 4d Na 2p VB(Cu) 400 m 30 35 40 45 50 55 60 65 70 kin. energy [eV] Fig. 2. (a) Optical microscopic image of the vicinity of a laser-cut. Brighter areas show locations of different morphology. The analyzed position, line, and area are indicated. (b) Valence spectrum using the PISAM microprobe set at the indicated small area on the sample. The photon energy of the first harmonic was 70 eV. ent grey scales. Measurements of the optical beam induced current (OBIC) [19] yielded areas of increased efficiency at the position of the brighter strips. The microspectroscopical analysis of a spot (Fig. 2 b) displays the occurrence of sodium, indium, and copper in the surface region indicated by the photoemission structures from Na 2p, In 4d and valence band, the latter being predominantly due to copper, for hν=70 eV. With the electron analyzer set at the corresponding energies we scanned across the laser-cut as indicated by the white box in the left frame of Fig. 2. An Improved Microprobe Using Direct Undulator Radiation III - 19 intensity Na 2p In 4d height [ m] VB(CU) 2 1 profile 0 -1 0 100 200 300 400 500 position [ m] Fig. 3. Line scans across the laser-cut with the analyzer set at the Na 2p, In 4d, and valence (i.e. Cu VB) energies indicating the relative concentration of the elements. The narrow rectangular box in the middle (position: 200-230 µm) indicates the actual size of the cut, the grey filled areas show regions of high OBIC signal. The resulting line scans are displayed in Fig 3. A close examination of these and similar data (not shown) leads to the following conclusion. At the lasercut all intensities appear to drop strongly. This is probably an artefact of the measurement which is due to due to charging effects since the absorber material has no back-electrode there. An area of ≈100 µm width left and right from the cut shows reduced indium and enhanced copper concentration indicating a new Cu-rich phase. In the areas of increased OBIC signal (grey boxes in Fig. 3), the sodium concentration shows a strong variation. This variation cannot be explained by topographical effects alone, such as height variation of the film (as indicated by the height profile in Fig. 3, bottom). We rather believe that sodium influences the size of the crystallites [18], which are larger in this area than elsewhere. Sodium is also expected to segregate predominantly to the grain boundaries [18]. The distribution of elements displayed in Fig. 3 can be found all along the cut as seen in area scans. Fig. 4 shows 2D grey-scale intensity images for sodium and indium displaying the 2D-distribution of these elements along the laser-cut. Obviously, the patterning step of the Mo-back contact leads to compositional variations on a significantly larger length scale than that of the actual laser-cut. III - 20 M. R. Weiss et al. a) [ m] 150 100 50 0 0 100 200 300 400 300 400 [ m] b) [ m] 150 100 50 0 0 100 200 [ m] Fig. 4. 2D-intensity maps for sodium (a) and indium (b) in the vicinity of a laser-cut in the Mo back-electrode. The scanned area is about 400 × 190µm2 . An improvement of the complete solar cell therefore must be combined with improved patterning methods. Moreover, new structuring techniques or substrate modifications might be necessary in order to optimize the spatial distribution of the sodium content in the absorber material. 3 Improved Design: PISAM-2 After successful testing of the performance of PISAM-1 we thought of ways to improve the spatial resolution. An obvious solution is to image the aperture demagnified onto the sample. Several alternatives are available to focus photons in the XUV region from 10 eV to 600 eV, as mentioned above. Microzoneplates have reached spatial resolutions down to 100 nm [6]. Unfortunately, their efficiency is very low, and they focus only for a fixed preselected photon energy. Bragg-Fresnel lenses provide higher efficiency but are still in the exploration phase and hence still limited in their spatial resolution. Normal incidence mirrors are coated with a multilayer to improve their reflectivity at a certain photon energy, e.g. 100 eV. These yielded the highest spatial resolution achieved at present with mirrors: 100 nm [7]. This is mostly due to the fact that normal incidence mirrors are rotational symmetric and spherical (or only little aspheric), and hence can be manufactured with very high precision. We decided to use grazing incidence optics. Although still difficult to manufacture they are the only focusing elements with energy-independent focus- An Improved Microprobe Using Direct Undulator Radiation III - 21 ing. Therefore, one needs not to sacrifice the tunability of the photon energy which can be an important experimental parameter in photoemission and allows to perform X-ray absorption experiments. Thus, we implemented an ellipsoidal mirror for demagnification and also added a condenser mirror between undulator and aperture to even increase the photon flux (see Fig. 1). To find optimized parameters for the system we calculated the expected focus for each possible configuration, characterized by demagnification factor and grazing angle of incidence and limited by the manufacturing errors: tangential error and microroughness [20, 21]. Based on this consideration an optimized mirror was selected with a grazing angle of 6◦ and a demagnification factor of 20. Larger angles would lead to lower transmission at high photon energies, smaller angles to increased curvature and therefore to even more problems to obtain the required surface quality. Higher demagnification also leads to unsurmountable problems in producing the mirror. The described evaluation process resulted in a design with the following properties: – focus width: 0.2 − 0.3 µm – focus height: 0.9 − 1.1 µm – photon flux: >2 · 1012 phot/s 1%BW in the smallest possible focus (0.2 × 0.9 µm2 ). From the ray tracings a focal intensity-distribution as shown in Fig. 5 was derived. Also, the sensitivity to maladjustments was determined by ray tracing. It 1 0.8 0.6 0.4 0.2 0 2 y [ m] intensity, norm. grey scale image -2 -1 x[ 1 0 -1 -2 0 m] 1 2 -2 -1 0 1 ] y[ m 2 -2 -1 0 1 2 x [ m] Fig. 5. Calculated intensity distribution of the PISAM-2 in the image plane including all broadening effects for a photon energy of 600 eV . resulted in stringent requirements for the mirror mount and fine adjustment: the mirror must be positioned with an accuracy of 1 µm and oriented within 1µrad for optimum conditions. III - 22 M. R. Weiss et al. Manipulator Electron Analyzer Ellipsoidal Mirror Piezo Stage Sample Holder Mirror Holder Fig. 6. Cut open view of the central experimental PISAM-2 chamber showing details of the sample stage and mirror holder. The recently completed PISAM II apparatus (see Fig. 6) appears to meet these requirements and is briefly described in the following. To reach the goal of high accuracy and stability, a complete reconstruction of the original PISAM-1 apparatus was neccessary: The sample positioning was divided in coarse movements with stepper motors and fine movements by piezo-stacks in flexure hinge frames (see Fig. 6), thus giving a full scan range of 25 mm in 50 nm steps. The ellipsoidal mirror is held by a mirror mount with 5 degrees of freedom, each of which acts directly on the mirror, thus increasing the stability. The whole setup is mounted on a marble plate to reduce vibration. With a prototype ellipsoidal mirror of reduced surface quality first tests of the new apparatus were recently performed on an integrated circuit. We scanned across wire-edges on a chip. High count rates indicate the conductor material as opposed to the adjacent insulator (SiO2 ). From the steep edges of the linescan a spatial resolution of better than 1µm in both directions can be derived (Fig. 7). This result was unexpected because of the large tangential error of this prototype mirror. It is probably due to the fact that only a very small part of the mirror was used which has a much higher surface quality than the rest of the mirror. 4 Conclusion With this project we have further developed the idea of spatially resolved spectroscopy. The result is an apparatus that has reached submicron resolution utilizing the direct undulator radiation. The microspectroscope can be operated in An Improved Microprobe Using Direct Undulator Radiation 3000 2500 2500 2000 2000 intensity intensity III - 23 1500 1500 1000 1000 500 500 0 7550 7575 7600 7625 x-position [ m] 7650 0 7575 7600 7625 7650 7675 y-position [ m] Fig. 7. Line-scans across a wire of an integrated circuit. The steep edges indicate a resolution of better than 1µm in both directions. the Auger or photoemission mode with variable photon energies and provides the highest photon density ever achieved in the sampling spot of such an instrument. Thus PISAM-2 is an interesting microprobe that may be useful for several applications one of which, a structured film of a Cu(In,Ga)Se2 solar cell, has been sketched as example. Acknowledgements We like to thank Dr. W. Riedl, Siemens Corp., for the solarcell samples. We also like to thank the BESSY-crew and especially Prof. W. Peatman for support regarding all stages of development. Financial support by the BMBF (projects 05 644WWA and 05 5WWAXB) is gratefully acknowledged. References 1. E. Bauer, Rep. Prog. Phys. 57, 895 (1994). 2. M. Keenlyside and P. Pianetta, J. Electron Spectrosc. Relat. Phenom. 66, 189 (1993). 3. B. Tonner, G. Harp, G. Koranda, and J. Zhang, Rev. Sci. Instrum. 63, 564 (1992). 4. R. Fink, M.R. Weiss, E. Umbach, D. Preikszas, H. Rose, R. Spehr, P. Hartel, W. Engel, R. Degenhardt, R. Wichtendahl, H. Kuhlenbeck, W. Erlebach, K. Ihmann, R. Schlögl, H.-J. Freund, A. M. Bradshaw, G. Lilienkamp, T. Schmidt, E. Bauer, and G. Benner (to be published in this issue). 5. J. Voss, M. Fornefett, J. Friedrichs, C. Kunz, M. Ptrorius, A. Ranck, K. Berens v. Rautenfeld, M. Schroeder, H. Sievers, and V. Wedemeier (to be published in this issue). III - 24 M. R. Weiss et al. 6. H. Ade, J. Kirz, S. Hulbert, E. Johnson, E. Anderson, and D. Kern, Appl. Phys. Lett. 56, 1841 (1990). 7. C. Capasso, W. Ng, A. Ray-Chaudhuri, S. Liang, R. Cole, Z. Guo, J. Wallace, F. Cerrina, J. Underwood, R. Perera, J. Kortright, G. De Stasio, and G. Margaritondo, Surf. Sci. 287/288, 1046 (1992). 8. K. Ninomya and M. Hasegawa, Surf. Sci. 287/288, 1051 (1993). 9. J. Voss, H. Dadras, C. Kunz, A. Moewes, G. Roy, H. Sievers, I. Storjohann, and H. Wongel, J. X-Ray Sci. Technol. 3, 85 (1992). 10. R. Nyholm, M. Eriksson, K. Hansen, O.-P. Sairanen, S. Werin, A. Flodström, C. Törnevik, T. Meinander, and M. Sarakontu, Rev. Sci. Intr. 60, 2168 (1989). 11. V. Wüstenhagen, J. Taborski, E. Umbach, I. Storjohann, and J. Voß, HASYLAB Technical Report, 555 (1989). 12. V. Wüstenhagen, M. Schneider, J. Taborski, W. Weiss, and E. Umbach, Vacuum 41, 1577 (1990). 13. V. Wüstenhagen, Photoemissions- und photoinduzierte Auger-Mikroskopie unter Verwendung von Undulatorstrahlung, Dissertation, Universität Stuttgart (1992). 14. M. Weiss, V. Wüstenhagen, R. Fink, and E. Umbach, J. Electr. Spectrosc. Relat. Phenom. 84, 9 (1997). 15. P. Väterlein, V. Wüstenhagen, and E. Umbach, Appl. Phys. Lett. 66, 2200 (1994). 16. P. Väterlein, M. Weiß V. Wüstenhagen, and E. Umbach, Appl. Surf. Sci. 70/71, 278 (1993). 17. G. G. Hembree, J. S. Drucker, F. C. H. Luo, M. Krishnamurthy, and J. A. Venables, Appl. Phys. Lett. 58, 1890 (1991). 18. C. Heske, R. Fink, and E. Umbach, Appl. Phys. Lett. 68, 3431 (1996). 19. W. Riedl, private communication. 20. M. Weiß, Hochortsauflösende Elektronenspektroskopie mit Undulatorstrahlung, Diplomarbeit, Universität Stuttgart (1993). 21. M. Weiss, V. Wüstenhagen, and E. Umbach, in: Proc. SPIE , vol. 2279, 2 (1994). This article was processed using the LATEX macro package with LLNCS style