NUMERICAL SIMULATION OF HEAT TRANSFER IN BLOOD VESSELS SALAHUDDIN AZRAIE A thesis submitted in fulfillment of the requirements for the award of the degree of Master of Engineering (Mechanical) Faculty of Mechanical Engineering Universiti Teknologi Malaysia MARCH 2010 iii ABSTRACT There is inadequate work done on tapered and stenois blood vessel under thermotheraphy, experimental studies being impossible to be conducted since the system is unlike mechanical system which could be modeled in the lab. In this study, the blood flow and heat transfer in an artery under various heating conditions have been numerically simulated. Three different cases of blood vessels are considered: straight vessel, converging tapered vessel and stenois vessel. Constant heat flux was applied at several intensities in each case. It was shown that the flow simulated at the vena contracta is analogous to the flow in a converging-diverging nozzle. However, the dimensionless center velocity at outlet region was found to be higher than that at the inlet region which could be due to the slow recovery of pressure after the stenois region. The local Nusselt number fluctuated at both the downstream and upstream which could have been caused by recirculation as the flow entered and exit the stenois region. Results in general have shown that blood flow in biological systems may well behave like the mechanical systems. iv ABSTRAK Kajian terhadap pembuluh darah meruncing dan pembuluh ‘stenois’ dalam keadaan terapi panas adalah kurang. Ujikaji adalah mustahil dilakukan kerana tidak seperti sistem mekanikal, model sistem untuk ujikaji tidak boleh dilakukan di dalam makmal. Dalam kajian ini, pengaliran darah dan pemindahan haba di dalam saluran arteri di bawah pelbagai keadaan pemanasan telah disimulasikan secara berangka. Tiga kes pembuluh darah telah dilihat: pembuluh lurus, pembuluh meruncing dan pembuluh ‘stenois’. Pelbagai kekuatan fluks haba kenaan telah dikaji dalam setiap kes. Hasil simulasi menunjukkan bahawa aliran pada vena kontrakta adalah seperti aliran di dalam muncung tumpu capah. Halaju pusat tak berdimensi pada kawasan alur keluar didapati lebih tinggi daripada yang terdapat pada kawasan alur masuk. Ini mungkin disebabkan oleh pemulihan perlahan tekanan selepas melewati kawasan stenois. Angka Nusselt tempatan turun-naik di hulu dan hilir kawasan stenois yang mungkin disebabkan oleh edaran semula apabila aliran memasuki dan keluar dari kawasan tersebut. Keputusan secara umumnya menunjukkan bahawa aliran darah dalam sistem biologi berkelakuan sama seperti aliran sistem mekanikal. v TABLE OF CONTENTS CHAPTER TITLE TABLE OF CONTENTS PAGE v LIST OF TABLES viii LIST OF FIGURES ix LIST OF SYMBOLS xii LIST OF APPENDICES xiv CHAPTER 1 INTRODUCTION 1.1 Introduction 1 1.2 Literature Review 3 1.3 Objectives 9 1.4 Scopes of Study 10 vi CHAPTER 2 GOVERNING EQUATIONS 2.1 Continuity and Momentum Equations 11 2.2 Energy Equations 13 2.3 Related Equations 14 2.4 Assumptions 16 CHAPTER 3 METHODOLOGY 3.1 Computational Fluid Dynamics 18 3.2 Pre-processing: GAMBIT 19 3.3 Solver: FLUENT 19 3.4 Post-processing: FLUENT 21 3.5 Pre-processing: Domain of Computational 22 3.6 Solver: Parameters 26 3.7 Solver: Method of Solution 28 3.8 Expected Outcome 29 CHAPTER 4 MODEL VALIDATION 4.1 Model Validation Methodology 30 4.2 Grid Dependency Test 32 4.3 Model Validation with Numerical Results Comparison 34 vii CHAPTER 5 RESULTS AND DISCUSSIONS 5.1 Simulation of the Normal Blood Flow in an Artery (Straight Tube) 5.2 Simulations of Abnormal Blood Flow in an Artery under Heating Conditions (Straight Tube) 5.3 37 39 Simulations of Blood Flow in an Artery (Tapered and Stenois Tubes) 5.3.1 Simulations of Blood Flow in an Artery (Tapered Tube) 5.2.2 47 47 Simulations of Blood Flow in an Artery (Stenois Tube) 56 CHAPTER 6 CONCLUSION 6.1 Conclusion 68 REFERENCES 70 APPENDICES A-M 73 viii LIST OF TABLES TABLE NO TITLE PAGE 3.6.1 Blood and water properties 26 3.6.2 Variation for normal blood flow in an artery 26 3.6.3 Variation for abnormal blood flow in an artery under heating condition 3.6.4 4.1.1 27 Variation for abnormal blood flow under tapered and stenois channel 28 Summary of meshing size for grid dependency test 31 ix LIST OF FIGURES FIGURE NO TITLE PAGE 1.2.1 Blood flow categories 8 1.2.2 History of blood flow studies in arteries 8 1.2.3 History of blood flow studies in veins and capillaries 9 3.3.1 Segregation solution method 3.3.2 Coupled solution method 21 3.5.1 Normal blood flow in an artery 22 3.5.2 Abnormal blood flow in an artery under heating condition 23 3.5.3 Normal blood flow in a tapered artery 3.5.4 Abnormal blood flow in a tapered artery under heating 20 23 condition 24 3.5.5 Normal blood flow in a stenois artery 25 3.5.6 Abnormal blood flow in a stenois artery under heating condition 4.2.1 25 Percentage difference changing from one mesh size to another 33 4.2.2 Number of iteration versus number of cells 33 4.3.1 Local Nusselt number for simultaneously developing region in a circular duct 4.3.2 35 Comparison of local Nusselt number with correlation results 36 x 5.1.1 Axial velocity of blood flow (q = 0W/m2, Um = 0.05m/s) at various radii 5.1.2 Skin friction coefficient of blood flow (q = 0w/m2, Um = 0.05m/s) at various radii 5.2.1 41 Axial velocity of blood flow (r = 5mm, Um = 0.05m/s) under various heating conditions 5.2.5 40 Local Nusselt number of blood flow (q = 4W/m2, Um = 0.05m/s) at various radii 5.2.4 39 Skin friction coefficient of blood flow (q = 4W/m2, Um = 0.05m/s) at various radii 5.2.3 38 Axial velocity of blood flow (q = 4W/m2, Um = 0.0 m/s) at various radii 5.2.2 38 42 Skin friction coefficient of blood flow (r = 5mm, Um = 0.05m/s) under various heating conditions 43 5.2.6 Local Nusselt number of blood flow (r = 5mm, Um = 0.05m/s) under various heating conditions 44 5.2.7 Plot of correlation of local Nusselt number for blood and water flow 5.2.8 45 Plot of simulation and correlation of local Nusselt number for blood flow 45 5.2.9 Zoom- in of Figure 5.2.6 46 5.3.1.1 Axial velocity of blood flow (q = 0W/m2, Um = 0.05m/s) under various taper angles 5.3.1.2 Skin friction coefficient of blood flow (q = 0W/m2, Um = 0.05m/s) under various taper angles 5.3.1.3 49 Skin friction coefficient of blood flow (q = 4W/m2, Um = 0.05m/s) under various taper angles 5.3.1.5 48 Axial velocity of blood flows (q = 4W/m2, Um = 0.05m/s) under various taper angles 5.3.1.4 47 50 Local Nusselt numbers of blood flow (q = 4W/m2, Um = 0.05m/s) under various taper angles 51 xi 5.3.16 Axial velocity of blood flow (Um = 0.05m/s, θ = 0.50) under various heating conditions 5.3.1.7 52 Skin friction coefficient of blood flow (Um = 0.05m/s, θ = 0.50) under various heating conditions 53 5.3.1.8 Local Nusselt number of blood flow (Um = 0.05m/s, θ = 0.50) under various heating conditions 54 5.3.2.1 Axial velocity of blood flow (q = 0W/m2, Um = 0.0 m/s) with various stenois heights 5.3.2.2 Center axial velocity of blood flow (q = 0W/m2, Um = 0.05m/s) with various stenois heights 5.3.2.3 62 Axial velocity of blood flow (a / r = 0.2, Um = 0.05m/s) under various heating conditions 5.3.2.9 61 Local Nusselt number of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights 5.3.2.8 60 Skin friction coefficient of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights 5.3.2.7 59 Center axial velocity of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights 5.3.2.6 58 Axial velocity of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights 5.3.2.5 57 Skin friction coefficient of blood flow (q = 0W/m2, Um = 0.05m/s) with various stenois heights 5.3.2.4 56 63 Center axial velocity of blood flow (a / r = 0.2, Um = 0.05m/s) under various heating conditions 64 5.3.2.10 Skin friction coefficient of blood flow (a / r = 0.2, Um = 0.05m/s) under various heating conditions 65 5.3.2.11 Local Nusselt number of blood flow (a / r = 0.8, Um = 0.05m/s) under various heating conditions 66 xii LIST OF SYMBOLS cp Specific heat d Diameter F External body forces or any user-defined sources h Enthalpy (ideal gases / incompressible flows) h Heat transfer coefficient I Unit tensor J j Diffusion flux of species j k eff Effective conductivity p Static pressure q Constant heat flux r Radial coordinate / radius Sh Heat of chemical reaction and any other user-defined heat sources Sm Mass added to the continuous phase from the dispersed second phase or any user-defined sources t Time Tm Fluid bulk temperature Tw Wall temperature Uc Axial velocity at center of tube Um Fluid average inlet velocity x Axial coordinate ρ Fluid density xiii ρg Gravitational body force τ Tensor stress µ Molecular viscosity / fluid dynamics viscosity ν Fluid velocity νr Radial velocity νx Axial velocity ∇ν T Effect of volume dilation Non-dimensional parameter Nu Nusselt number Pr Prandlt number Re Reynolds number xiv LIST OF APPENDICES APPENDIX A B TITLE Blood properties set up in Fluent Axial velocity profile for 0.5 tapered tube 77 Axial velocity profile for stenois height of 1.5mm (q = 0W/m2, Um = 0.05m/s) J 76 Axial velocity profile for stenois height of 1mm (q = 0W/m2, Um = 0.05m/s) I 76 Temperature contour for 0.50 tapered tube (q = 6W/m2, Um = 0.05m/s) H 75 Temperature contour for 0.50 tapered tube (q = 5W/m2, Um = 0.05m/s) G 75 Temperature contour for 0.50 tapered tube (q = 4W/m2, Um = 0.05m/s) F 74 Axial velocity profile for 1.50 tapered tube (q = 4W/m2, Um = 0.05m/s) E 74 Axial velocity profile for 1.00 tapered tube (q = 4W/m2, Um = 0.05m/s) D 73 0 (q = 4W/m2, Um = 0.05m/s) C PAGE 77 Axial velocity profile for stenois height of 2mm (q = 0W/m2, Um = 0.05m/s) 78 xv K Temperature contour for stenois height of 1mm (q = 4W/m2, Um = 0.05m/s) L Temperature contour for stenois height of 1mm (q = 5W/m2, Um = 0.05m/s) M 78 79 Temperature contour for stenois height of 1mm (q = 6W/m2, Um = 0.05m/s) 79 xvi CHAPTER I INTRODUCTION 1.1 Introduction The understanding of the function of the circulation system under both normal and abnormal conditions, designing cardiovascular artificial devices as well as diagnosing and treating disease is not complete without the full understanding of blood flow. The first attempt to formulate the governing equation of motion of blood in an artery was made by Euler in 1775. Before the age of computer, the analytical solutions were done with much considerable simplifications. Following the introduction of computers, blood flow equation can be treated with fewer simplification and enable the theoretical results to be close to experimental ones (Hunter, 1972). Since the earlier attempt by Euler until recently development, many models of blood and blood flow have been introduced by researchers. Earlier researchers tend to assume blood as a Newtonian fluid which is flowing in a rigid tube wall in order to make the calculation easier. Later on, non-Newtonian fluid assumption was applied to study blood flow in tube with less than 1mm diameter. Lattice xvii Boltzmann approach was introduced around year 2009 to cater to non-Newtonian complexity of blood flow (Ashrafizaadeh and Bakshaei, 2009). Currently, the study of blood flow in a straight tube involves complex simulation by introducing bifurcation and stenosis effects. Heat transfer and magnetic field effect on blood flow have also been introduced to bring more understanding about blood rheology. Higher heart failure mortality in most developed countries (Espino et al, 2006) makes the study of blood flow significantly important. Certain diseases lead to heart failure due to the changing of blood flow pattern in blood vessels. From a microscopic point of view, interaction between blood particles (red blood cells, white blood cells and platelets) and bifurcation walls can lead to changing in velocity patterns especially when there are irregularities in the said regions such as fat or stenosis. Blood contains many enzymes and hormones, and transports oxygen and carbon dioxide from the lung to the cells of the tissues. Blood circulation also plays an important role in thermoregulation and mass transport. The quantitative prediction becomes more significant in human thermal comfort, drug delivery and noninvasive measurements research. Earlier heat transfer analysis introduced noninvasive measurement of glucose in blood by means of measuring heat generation from glucose itself. Another significant importance of heat transfer analysis in blood vessels to surrounding walls or tissues is to understand more on drug penetration in tumor affected tissues. Thermal therapy is one of the widely used methods in treating tumor. Various method of heating is introduced in medical field to enhance necrosis rate of tumor such as ultrasound, laser irradiation, and laser ablation. Hyperthermia therapy method is used to enhance the effect of radiotherapy and chemotherapy (Yuan, 2009). These methods require no surgery to perform. Other method that requires small surgery is radiofrequency tumor ablation (Santos et al, 2008). All abovementioned methods are aim to kill tumor affected tissue. However, heating does change behavior of blood flow and heating effectiveness relies on blood xviii vessels geometry. Therefore, further understanding of heating effects on blood flow in various blood vessels geometry is indeed important to improve tumor treatment under thermal therapy. 1.2 Literature Review J.P. Hunter is probably the first to formulate and solved numerically the governing equations of blood flow in a nonlinear viscoelastic artery in 1972 in his Masters’ thesis (Hunter, 1972). He assumed that the blood was Newtonian, homogenous and incompressible. In his model, blood was assumed to flow in a cylindrical tube under non-turbulent flow condition with an elastic artery wall. The artery wall showed both elastic and collagen fiber behavior. These two behaviors lead to a non-linear time dependent stress-strain relationship. The artery wall became stiffer at higher transmural (through the wall of an organ) pressure due to higher Young’s modulus value. In vito testing was used to obtain experimental data to get the relationship between transmural pressure and arterial radius of the wall. According to Hunter, an empirical approach was better than synthetic approach using finite elasticity theory since the synthetic approach implied the artery wall material is homogenous, isotropic and incompressible. Furthermore, Hunter also considered blood flow through bifurcation in order to fully describe his heart-aorta model. The computational blood flow through bifurcation was solved using FORTRAN software by adopting finite difference approach. Heat transfer in blood flow was mentioned in the last chapter. The general one dimensional heat equation was used to calculate the transfer of heat along the artery assuming no heat loss to the surroundings. Hunter did not consider any heat transfer between bloods to the artery wall. xix Khaled et al (Khaled et al, 2003) reviewed transport theories in porous media and its application in biological tissues. Mass diffusion and different convective flow models are used to study the transport theories. Development of bio-heat equation was analyzed. The most appropriate bio-heat model for heat transfer in biological tissues was found in the theory of porous media. Tsubota et al (Tsubota et al, 2006) introduced a new approach to describe the behavior of blood flow in 2005 called particle method. Blood was assumed to be incompressible and Newtonian fluid and modeled as an assembly of discrete particles. Blood was also assumed to be homogenous fluid from macroscopic point of view. Tsubota et al argued that continuum mechanics approach such as finite difference method (FDM), finite volume method (FVM), and finite element method (FEM) can be used to analyze blood flow numerically. One of obvious limitation for this approach is the simulation is its two-dimensional analysis instead of threedimensional. However, his simulation results are yet to be verified by experimental results. Furthermore, there is no mention of heat transfer analysis in blood vessel using this particle method. Blood flow and heat transfer in human finger was studied by He et al in 2008 (He et al, 2008). In his blood flow modeling, he simulated the blood flow in large vessels with the diameter larger than or equal to 1mm. One-dimensional flow model was used for this case. Darcy model in porous medium approach was assumed in his blood flow modeling for micro-vessels case. Finite element model was used to model a human finger. Magnetic resonance imaging (MRI) method was used to obtain a geometric model of a human finger and different sequences of images were taken to simulate flow of blood and heat transfer. From his images data, the inflow velocity in the artery was found to be 19cm/s and in the vein3.5cm/s. Reynolds number was 50 for this case when the arterial velocity was the reference velocity. Ying He concluded that porous media transport model provided good result to heat transfer analysis in living tissues (in this case, human finger) using unified energy equation. xx Numerical modeling of blood flow in a variable shape artery wall was performed by Mukhopadhyay and Layek (Mukhopadhyay and Layek, 2008). In their study, blood was assumed to be a Newtonian fluid and the flow was simulated in the rigid artery wall under steady state situations. Three shapes of the stenois (abnormal narrowing in a blood vessel) were used in this modeling and the shape was symmetric as well as asymmetric about the middle cross sectional area. Mukhopadhyay et al agreed in their research that blood could be treated as a Newtonian fluid since the viscosity coefficient of blood approaches a constant value under relatively large shear rates. Furthermore, the diameter of the rigid tube used in the modeling was equivalently the same as the diameter of a large vessel in which Newtonian fluid condition was satisfied. Hemodynamic factors in the rigid tube like wall shear stress, pressure and velocity were analyzed. No heat transfer analysis was performed under this rigid artery wall assumption with variable shape of stenois. Thermal effect on tumor was studied by Horng et al (Horng et al, 2007) using pulsatile blood flow. Cyclic effect of heart beat was implemented by using a sinusoidal pulsatile velocity profile. Horng et al judged from their literature that blood flow in all arteries undergo pulsatile flow due to the heart pump nature. The energy transport equation with Pennes’s bioheat equation was used to solve the heat transfer between blood vessels and surrounding tumor tissues. The blood flow was modeled with assumptions of incompressible, laminar and Newtonian fluid. Blood vessel was modeled as a rigid artery wall in perfect cylindrical shape. Other blood properties such as thermal conductivity; specific heat and perfusion mass flow rate can be found in their literature. They concluded in their study by suggesting that steady uniform velocity profile and pulsatile velocity profile for blood flow exhibited minor differences on thermal analysis. Heat transfer analysis to the surrounding tissues was done in their study with an assumption of inelastic artery wall. The study of dynamics response of heat and mass transfer in blood flow through bifurcation arteries was done by Chakravarty and Sen (Chakravarty and xxi Sen, 2005). A mathematical model of the response was developed under the assumption of Newtonian fluid. Navier-Stokes equation was used to describe nonlinear unsteady flow phenomenon and heat conduction and convectiondiffusion equations were used to performed thermal analysis. In this study, the artery wall was assumed to be elastic on local fluid mechanics but not on the stresses and strains in the wall itself. The wall motion effect on velocity profile of blood flow was found to be insignificant. Temperature profiles at apex of arterial bifurcation also gave insignificant difference when considering wall motion compared to the rigid wall. In his concluding remark, there was significant reduction in heat flux magnitude when the arterial wall was considered as rigid. Hyperthermia therapy simulation in biological tissue was done by Yuan (Yuan, 2008). An equivalent heat transfer coefficient in a porous model was done by numerical analysis. In this study, all arterial blood vessels were assumed to be straight and treated as porous medium, constant heat transfer coefficient and blood velocity were used, and thermal properties for tissues and blood were treated as isotropic. Ultrasound was used as a source of energy input for hyperthermia treatment. The whole domain of calculation was considered as an assembly of repeated hexagon units. Tri-Diagonal Matrix Algorithm was employed to solve two-equation porous model for blood and tissue and Pennes’s bio-heat transfer equation to predict the temperature distribution. Yuan concluded in his summary that a decrease in vessel diameter increases the equivalent heat transfer coefficient. The equivalent heat transfer coefficient also is not dominated by blood velocity and heating conditions. Three equivalent heat transfer coefficient were calculated for three different vessel diameters. Yuan did not consider in his study to for expanding or narrowing vessel diameter or non-symmetric blood vessel. Cooling effect of thermally significant blood vessels in perfused tumor tissue during thermal therapy was done by Shih et al (Shih et al, 2006). A single blood vessel with rigid wall inside and throughout the perfused tumor tissue was used as computational domain. The velocity profile in the blood vessel was assumed to be uniform. The temperature of blood flow and tissue were governed by xxii the energy transport equation. During thermal simulation, ultrasound was used as source of heat power and the heat absorption rate of blood is assumed to be onetenth of tissue. Heating power density and heating durations were varied while the absorbed total energy density was kept constant during the simulation. Shih et al concluded that short duration and high intensity heating method can eliminate the cooling the cooling effect of the blood vessels 0.2mm in diameter. Cooling effect was stronger for blood vessel larger than 0.2mm diameter and led to incomplete necrosis even though double heating power or density was used. Shih also did not consider in his study to for expanding or narrowing vessel diameter or nonsymmetric blood vessel. Radiofrequency electric current is used in radiofrequency tumor ablation procedure. An active electrode is inserted by mean of surgery and radiofrequency energy is applied to heat the surrounding tissue. Santos et al (Santos et al, 2008) studied the effect of variable heat transfer coefficient on tissue temperature next to a larger vessel during radiofrequency tumor ablation. In his journal, a large artery 10mm diameter was used and an electrode was put at two different distances away from the artery. The blood was assumed as a Newtonian fluid and the flow was assumed to be laminar. The vessel of artery was assumed to be straight with constant blood velocity. Both constant and variable of the convective heat transfer coefficient were simulated. The convective heat transfer coefficient was increased initially and rapidly up to 6000W/m2K followed by a sharp decrease to 554W/m2K for 1mm distance configuration. The convective heat transfer coefficient was increased initially and rapidly up to 7000W/m2K followed by a sharp decrease to 336W/m2K for 5mm distance configuration. Santos et al concluded that assumption of constant convective heat transfer coefficient sufficient enough to produce precise result for thermal procedure more than 2 minutes. The preceding paragraphs described some of the past studies related to blood flow. Figure 1.2.1 shows blood flow major categories. xxiii Figure 1.2.1: Blood flow categories Figure 1.2.2 shows summarized most of the work associated with blood flow completed. Figure 1.2.2: History of blood flow studies in arteries xxiv Figure 1.2.3: History of blood flow studies in veins and capillaries Figure 1.2.2 and Figure 1.2.3 show that most of heat transfer studies were done with straight vessel assumption. He et al (He et al, 2008) did performed heat transfer study by considering bifurcation vessel but He assumed that both parent and daughter vessels were straight vessel. Apparently, there is inadequate work done on tapered and stenois blood vessel under heating condition. 1.3 Objective The objective of this study is to numerically simulate the blood flow and heat transfer in an artery under heating conditions. xxv 1.4 Scope of Study The scope of this study will involve: 1 – Simulation of the normal blood flow in an artery. 2 – Repeating (1) under abnormal condition of heating (i.e. chemotherapy, thermotherapy) at various degrees. 3 – Repeating (1) for a tapered channel flow and stenois (abnormal narrowing) channel flow. CHAPTER I INTRODUCTION 1.1 Introduction The understanding of the function of the circulation system under both normal and abnormal conditions, designing cardiovascular artificial devices as well as diagnosing and treating disease is not complete without the full understanding of blood flow. The first attempt to formulate the governing equation of motion of blood in an artery was made by Euler in 1775. Before the age of computer, the analytical solutions were done with much considerable simplifications. Following the introduction of computers, blood flow equation can be treated with fewer simplification and enable the theoretical results to be close to experimental ones (Hunter, 1972). Since the earlier attempt by Euler until recently development, many models of blood and blood flow have been introduced by researchers. Earlier researchers tend to assume blood as a Newtonian fluid which is flowing in a rigid tube wall in order to make the calculation easier. Later on, non-Newtonian fluid assumption was applied to study blood flow in tube with less than 1mm diameter. Lattice 2 Boltzmann approach was introduced around year 2009 to cater to non-Newtonian complexity of blood flow (Ashrafizaadeh and Bakshaei, 2009). Currently, the study of blood flow in a straight tube involves complex simulation by introducing bifurcation and stenosis effects. Heat transfer and magnetic field effect on blood flow have also been introduced to bring more understanding about blood rheology. Higher heart failure mortality in most developed countries (Espino et al, 2006) makes the study of blood flow significantly important. Certain diseases lead to heart failure due to the changing of blood flow pattern in blood vessels. From a microscopic point of view, interaction between blood particles (red blood cells, white blood cells and platelets) and bifurcation walls can lead to changing in velocity patterns especially when there are irregularities in the said regions such as fat or stenosis. Blood contains many enzymes and hormones, and transports oxygen and carbon dioxide from the lung to the cells of the tissues. Blood circulation also plays an important role in thermoregulation and mass transport. The quantitative prediction becomes more significant in human thermal comfort, drug delivery and noninvasive measurements research. Earlier heat transfer analysis introduced noninvasive measurement of glucose in blood by means of measuring heat generation from glucose itself. Another significant importance of heat transfer analysis in blood vessels to surrounding walls or tissues is to understand more on drug penetration in tumor affected tissues. Thermal therapy is one of the widely used methods in treating tumor. Various method of heating is introduced in medical field to enhance necrosis rate of tumor such as ultrasound, laser irradiation, and laser ablation. Hyperthermia therapy method is used to enhance the effect of radiotherapy and chemotherapy (Yuan, 2009). These methods require no surgery to perform. Other method that requires small surgery is radiofrequency tumor ablation (Santos et al, 2008). All abovementioned methods are aim to kill tumor affected tissue. However, heating does change behavior of blood flow and heating effectiveness relies on blood 3 vessels geometry. Therefore, further understanding of heating effects on blood flow in various blood vessels geometry is indeed important to improve tumor treatment under thermal therapy. 1.2 Literature Review J.P. Hunter is probably the first to formulate and solved numerically the governing equations of blood flow in a nonlinear viscoelastic artery in 1972 in his Masters’ thesis (Hunter, 1972). He assumed that the blood was Newtonian, homogenous and incompressible. In his model, blood was assumed to flow in a cylindrical tube under non-turbulent flow condition with an elastic artery wall. The artery wall showed both elastic and collagen fiber behavior. These two behaviors lead to a non-linear time dependent stress-strain relationship. The artery wall became stiffer at higher transmural (through the wall of an organ) pressure due to higher Young’s modulus value. In vito testing was used to obtain experimental data to get the relationship between transmural pressure and arterial radius of the wall. According to Hunter, an empirical approach was better than synthetic approach using finite elasticity theory since the synthetic approach implied the artery wall material is homogenous, isotropic and incompressible. Furthermore, Hunter also considered blood flow through bifurcation in order to fully describe his heart-aorta model. The computational blood flow through bifurcation was solved using FORTRAN software by adopting finite difference approach. Heat transfer in blood flow was mentioned in the last chapter. The general one dimensional heat equation was used to calculate the transfer of heat along the artery assuming no heat loss to the surroundings. Hunter did not consider any heat transfer between bloods to the artery wall. 4 Khaled et al (Khaled et al, 2003) reviewed transport theories in porous media and its application in biological tissues. Mass diffusion and different convective flow models are used to study the transport theories. Development of bio-heat equation was analyzed. The most appropriate bio-heat model for heat transfer in biological tissues was found in the theory of porous media. Tsubota et al (Tsubota et al, 2006) introduced a new approach to describe the behavior of blood flow in 2005 called particle method. Blood was assumed to be incompressible and Newtonian fluid and modeled as an assembly of discrete particles. Blood was also assumed to be homogenous fluid from macroscopic point of view. Tsubota et al argued that continuum mechanics approach such as finite difference method (FDM), finite volume method (FVM), and finite element method (FEM) can be used to analyze blood flow numerically. One of obvious limitation for this approach is the simulation is its two-dimensional analysis instead of threedimensional. However, his simulation results are yet to be verified by experimental results. Furthermore, there is no mention of heat transfer analysis in blood vessel using this particle method. Blood flow and heat transfer in human finger was studied by He et al in 2008 (He et al, 2008). In his blood flow modeling, he simulated the blood flow in large vessels with the diameter larger than or equal to 1mm. One-dimensional flow model was used for this case. Darcy model in porous medium approach was assumed in his blood flow modeling for micro-vessels case. Finite element model was used to model a human finger. Magnetic resonance imaging (MRI) method was used to obtain a geometric model of a human finger and different sequences of images were taken to simulate flow of blood and heat transfer. From his images data, the inflow velocity in the artery was found to be 19cm/s and in the vein3.5cm/s. Reynolds number was 50 for this case when the arterial velocity was the reference velocity. Ying He concluded that porous media transport model provided good result to heat transfer analysis in living tissues (in this case, human finger) using unified energy equation. 5 Numerical modeling of blood flow in a variable shape artery wall was performed by Mukhopadhyay and Layek (Mukhopadhyay and Layek, 2008). In their study, blood was assumed to be a Newtonian fluid and the flow was simulated in the rigid artery wall under steady state situations. Three shapes of the stenois (abnormal narrowing in a blood vessel) were used in this modeling and the shape was symmetric as well as asymmetric about the middle cross sectional area. Mukhopadhyay et al agreed in their research that blood could be treated as a Newtonian fluid since the viscosity coefficient of blood approaches a constant value under relatively large shear rates. Furthermore, the diameter of the rigid tube used in the modeling was equivalently the same as the diameter of a large vessel in which Newtonian fluid condition was satisfied. Hemodynamic factors in the rigid tube like wall shear stress, pressure and velocity were analyzed. No heat transfer analysis was performed under this rigid artery wall assumption with variable shape of stenois. Thermal effect on tumor was studied by Horng et al (Horng et al, 2007) using pulsatile blood flow. Cyclic effect of heart beat was implemented by using a sinusoidal pulsatile velocity profile. Horng et al judged from their literature that blood flow in all arteries undergo pulsatile flow due to the heart pump nature. The energy transport equation with Pennes’s bioheat equation was used to solve the heat transfer between blood vessels and surrounding tumor tissues. The blood flow was modeled with assumptions of incompressible, laminar and Newtonian fluid. Blood vessel was modeled as a rigid artery wall in perfect cylindrical shape. Other blood properties such as thermal conductivity; specific heat and perfusion mass flow rate can be found in their literature. They concluded in their study by suggesting that steady uniform velocity profile and pulsatile velocity profile for blood flow exhibited minor differences on thermal analysis. Heat transfer analysis to the surrounding tissues was done in their study with an assumption of inelastic artery wall. The study of dynamics response of heat and mass transfer in blood flow through bifurcation arteries was done by Chakravarty and Sen (Chakravarty and 6 Sen, 2005). A mathematical model of the response was developed under the assumption of Newtonian fluid. Navier-Stokes equation was used to describe nonlinear unsteady flow phenomenon and heat conduction and convectiondiffusion equations were used to performed thermal analysis. In this study, the artery wall was assumed to be elastic on local fluid mechanics but not on the stresses and strains in the wall itself. The wall motion effect on velocity profile of blood flow was found to be insignificant. Temperature profiles at apex of arterial bifurcation also gave insignificant difference when considering wall motion compared to the rigid wall. In his concluding remark, there was significant reduction in heat flux magnitude when the arterial wall was considered as rigid. Hyperthermia therapy simulation in biological tissue was done by Yuan (Yuan, 2008). An equivalent heat transfer coefficient in a porous model was done by numerical analysis. In this study, all arterial blood vessels were assumed to be straight and treated as porous medium, constant heat transfer coefficient and blood velocity were used, and thermal properties for tissues and blood were treated as isotropic. Ultrasound was used as a source of energy input for hyperthermia treatment. The whole domain of calculation was considered as an assembly of repeated hexagon units. Tri-Diagonal Matrix Algorithm was employed to solve two-equation porous model for blood and tissue and Pennes’s bio-heat transfer equation to predict the temperature distribution. Yuan concluded in his summary that a decrease in vessel diameter increases the equivalent heat transfer coefficient. The equivalent heat transfer coefficient also is not dominated by blood velocity and heating conditions. Three equivalent heat transfer coefficient were calculated for three different vessel diameters. Yuan did not consider in his study to for expanding or narrowing vessel diameter or non-symmetric blood vessel. Cooling effect of thermally significant blood vessels in perfused tumor tissue during thermal therapy was done by Shih et al (Shih et al, 2006). A single blood vessel with rigid wall inside and throughout the perfused tumor tissue was used as computational domain. The velocity profile in the blood vessel was assumed to be uniform. The temperature of blood flow and tissue were governed by 7 the energy transport equation. During thermal simulation, ultrasound was used as source of heat power and the heat absorption rate of blood is assumed to be onetenth of tissue. Heating power density and heating durations were varied while the absorbed total energy density was kept constant during the simulation. Shih et al concluded that short duration and high intensity heating method can eliminate the cooling the cooling effect of the blood vessels 0.2mm in diameter. Cooling effect was stronger for blood vessel larger than 0.2mm diameter and led to incomplete necrosis even though double heating power or density was used. Shih also did not consider in his study to for expanding or narrowing vessel diameter or nonsymmetric blood vessel. Radiofrequency electric current is used in radiofrequency tumor ablation procedure. An active electrode is inserted by mean of surgery and radiofrequency energy is applied to heat the surrounding tissue. Santos et al (Santos et al, 2008) studied the effect of variable heat transfer coefficient on tissue temperature next to a larger vessel during radiofrequency tumor ablation. In his journal, a large artery 10mm diameter was used and an electrode was put at two different distances away from the artery. The blood was assumed as a Newtonian fluid and the flow was assumed to be laminar. The vessel of artery was assumed to be straight with constant blood velocity. Both constant and variable of the convective heat transfer coefficient were simulated. The convective heat transfer coefficient was increased initially and rapidly up to 6000W/m2K followed by a sharp decrease to 554W/m2K for 1mm distance configuration. The convective heat transfer coefficient was increased initially and rapidly up to 7000W/m2K followed by a sharp decrease to 336W/m2K for 5mm distance configuration. Santos et al concluded that assumption of constant convective heat transfer coefficient sufficient enough to produce precise result for thermal procedure more than 2 minutes. The preceding paragraphs described some of the past studies related to blood flow. Figure 1.2.1 shows blood flow major categories. 8 Figure 1.2.1: Blood flow categories Figure 1.2.2 shows summarized most of the work associated with blood flow completed. Figure 1.2.2: History of blood flow studies in arteries 9 Figure 1.2.3: History of blood flow studies in veins and capillaries Figure 1.2.2 and Figure 1.2.3 show that most of heat transfer studies were done with straight vessel assumption. He et al (He et al, 2008) did performed heat transfer study by considering bifurcation vessel but He assumed that both parent and daughter vessels were straight vessel. Apparently, there is inadequate work done on tapered and stenois blood vessel under heating condition. 1.3 Objective The objective of this study is to numerically simulate the blood flow and heat transfer in an artery under heating conditions. 10 1.4 Scope of Study The scope of this study will involve: 1 – Simulation of the normal blood flow in an artery. 2 – Repeating (1) under abnormal condition of heating (i.e. chemotherapy, thermotherapy) at various degrees. 3 – Repeating (1) for a tapered channel flow and stenois (abnormal narrowing) channel flow. 11 CHAPTER II GOVERNING EQUATIONS 2.1 Continuity and Momentum Equations Conservation of mass or continuity equation in FLUENT is written as follows: ( ) ∂ρ + ∇ • ρν = S m ∂t (2.1) ρ is fluid density, t is time, ν is fluid velocity, S m is the mass added to the continuous phase from the dispersed second phase or any user-defined sources. For 2D axis-symmetric geometries flow condition, the Eq. (2.1) can be written as: ρν r ∂ρ ∂ ∂ + ( ρ ν x ) + (ρ ν r ) + = Sm ∂t ∂x ∂r r (2.2) 12 x is axial coordinate, r is radial coordinate, ν x is axial velocity, ν r is radial velocity. Conservation of momentum equation in FLUENT is written as follows: ( ) () ( ) ∂ ρν + ∇ • ρν ν = − ∇ p + ∇ • τ + ρ g + F ∂t (2.3) p is static pressure, τ is tensor stress, ρ g is gravitational body force, F is external body forces or any user-defined sources. The tensor stress τ is given by: T 2   τ = µ  ∇ν + ∇ν  − ∇ •ν I  ï£°ï£ ï£¸ 3  (2.4) µ is the molecular viscosity, I is the unit tensor, T ∇ν is the effect of volume dilation. For 2D axis-symmetric geometries flow condition, the axial momentum equation can be written as: ∂ (ρ ν x ) + 1 ∂ (r ρ ν x ν x ) + 1 ∂ (r ρ ν r ν x ) = − ∂p + ∂t r ∂x r ∂r ∂x 1 ∂   ∂ν x 2  1 ∂   ∂ν x ∂ν r  + − ∇ •ν  +  + Fx r µ  2 r µ  r ∂x  ï£ ∂x 3 ∂x   r ∂r  ï£ ∂r ( ) (2.5) For 2D axis-symmetric geometries flow condition, the radial momentum equation can be written as: 13  ∂ν  ∂ (ρ ν r ) + 1 ∂ (r ρ ν x ν r ) + 1 ∂ (r ρ ν r ν r ) = − ∂p + 1 ∂ r µ  2 ∂ν r + x  ∂t r ∂x r ∂r ∂r r ∂x  ï£ ∂x ∂r  1 ∂   ∂ν r 2 + r µ 2 − ∇ •ν r ∂r  ï£ ∂r 3 ν ν 2µ  ∇ •ν + ρ x + Fr  − 2 µ 2r + 3 r r r  ( ) 2 ( ) (2.6) ∇ •ν is defined as: ∇ •ν = ∂ν x ∂ν r ν r + + . ∂x ∂r r 2.2 Energy Equation (2.7) The energy equation in FLUENT is solved in the following form:   ∂ (ρ E ) + ∇ • ν (ρ E + p ) = ∇ •  k eff ∇ T − ∑ h j J j + τ eff •ν  + S h ∂t j ï£ ï£¸ ( ) ( ) (2.8) k eff is the effective conductivity, J j is the diffusion flux of species j , k eff ∇ T is energy transfer due to conduction, ∑h j J j is energy transfer due to species diffusion, j ∑ (τ eff ) •ν is energy transfer due to viscous dissipation, j S h is the heat of chemical reaction and any other user-defined heat sources, E is defined as: E=h− p ν2 + , ρ 2 (2.9) h is enthalpy (ideal gases) and is defined as: h = ∑Y j h j , j (2.10) 14 or h is enthalpy (incompressible flows) and is defined as: h = ∑Y j h j + j p ρ . (2.11) In Eq. (2.10) and Eq. (2.11), Y j is the mass fraction of species j and h j is defined as follows: T h j = ∫ c p , j dT (2.12) Tref Tref is at 298.15K. 2.3 Related Equations  r Axial velocity, u = U m 2 1 −   ï£ r0 where u is axial velocity U m is fluid average inlet velocity r is radial position r0 is tube radius    2    (2.13) 15 Skin friction coefficient, C f = 64 Re 4 (2.14) where Re is Reynolds number Re = ρUm d µ (2.15) ρ is fluid density µ is fluid dynamics viscosity d is tube diameter Axial temperature gradient, 2q dT = dx ρ U c c p r (2.16) where c p is fluid specific heat q is constant heat flux Temperature difference between wall and fluid bulk, 2 Tw − Tm = 11 U c r ρ c p dT 96 k dx (2.17) where Tw is wall temperature Tm is fluid bulk temperature hd k (2.18) Hydrodynamic entrance length, Lh = 0.056 Re d (2.19) Thermal entrance length, Lt = 0.05 Re Pr d (2.20) Nusselt number, Nu = where h is heat transfer coefficient 16 where Pr is Prandlt number Pr = cp µ 2.4: k (2.21) Assumptions Blood is assumed to be homogenous, incompressible, and a Newtonian fluid (Hunter, 1972, He et al, 2005, Tsubota et al, 2006, Horng et al, 2007, Jafari et al, 2008, Mukhopadhyay and Layek, 2008, Santos et al, 2008, and Yuan, 2008). The blood can be assumed as a Newtonian fluid since temperature in normal human body is almost constant all the time at 37 degree Celsius. According to Fahraeus and Lindqvist, blood flow through tubes more than approximately 1mm diameter has a constant apparent blood viscosity (Waite and Fine, 2007 and Chandran et al, 2007). Blood flow in artery is assumed to be uniform velocity profile instead of pulsative velocity profile in a perfect cylindrical channel. Even though the actual flow in artery is in pulsative velocity profile, thermal analysis result shows no significant difference if uniform velocity profile is used (Horng et al, 2007). Arterial wall is assumed to be rigid channel instead of elastic channel. Elastic arterial wall was used by Hunter (Hunter, 1972) and Chakravarty and Sen (Chakravarty and Sen, 2005) while rigid arterial wall was used by Horng et al (Horng et al, 2007), Mukhopadhyay and Layek (Mukhopadhyay and Layek, 2008), Santos et al (Santos et al, 2008), and Yuan (Yuan, 2008). Study by Chakravarty and Sen (Chakravarty, 2005) showed that both velocity profile and thermal analysis in elastic arterial wall provided insignificant difference compare to the rigid one. 17 Figure 1.2.2 and Figure 1.2.3 in the previous chapter show that there are lacks of study for blood vessel with a tapered and stenois channel under various heating conditions. Most of the heat transfer studies were done by assuming the blood vessel is straight. Therefore, simulation of the normal blood flow in an artery with various shapes of channels under various heating conditions will performed in this study. 18 CHAPTER III METHODOLOGY 3.1 Computational Fluid Dynamics There are many commercially available softwares to perform computational fluid dynamics (CFD) in general. FLUENT, POLYFLOW and CFX are among the major ones used in flow modeling. In this study, FLUENT version 6.2.26 and GAMBIT 2.3.16 (GAMBIT) from ANSYS Incorporation (ANSYS) are used. There are three elements in the CFD package in ANSYS, which are: 1. Pre-processing 2. Solver 3. Post-processing 19 3.2 Pre-processing: GAMBIT Pre-processing is an input process for any flow modeling simulation in ANSYS software. In this stage, geometry of desired part or system region is defined. The region of interest is called computational domain. Once the computational domain is defined, the domain is subdivided into predefine number of smaller region. Fluid properties also will be defined in this stage. Boundary layers will be defined on the computational domain such as wall, center line, inlet and outlet. Overall accuracy of flow simulation is heavily relies in this stage. In general, the fineness of the grid, the results will be more accurate. GAMBIT is relatively old pre-processing software for ANSYS package. ANSYS is no longer support GAMBIT once it is replaced by ANSYS Design Modeler and ANSYS Meshing in ANSYS version 12 under ANSYS. Now, GAMBIT is no longer considered the latest state-of-the-art pre-processing software for engineering analysis, but it still suitable to run simulation for this study. 3.3 Solver: FLUENT Flow calculations are performed by solver in CFD package. The governing integrals equations for the conservation of mass and momentum, and energy equation are solved using two numerical methods. FLUENT 6.2 version provides two methods: 1- Segregated solver 2- Coupled solver 20 Control-volume-based technique is used in both solvers. In general, both solvers consist of: 1- The computational domain is divided by computational grid into discrete control volumes. 2- Algebraic equations for the discrete dependent variables are constructed by integrating the governing equations on the individual control volumes. 3- Values of the dependent variables are updated by linearization of the discrete equations and solution of the resultant linear equation system. The segregation solver solves the governing equations of continuity, momentum, and energy sequentially. The overview of the segregated solution method is shown in Figure 3.3.1. Update properties. Solve momentum equations. Solve continuity equation. Update pressure, face mass flow rate. Solve energy, species, turbulence, and other scalar equations. Converged? Stop Figure 3.3.1: Segregation solution method The coupled solver solves the governing equations of continuity, momentum, and energy simultaneously. The overview of the coupled solution method is shown in Figure 3.3.2. 21 Update properties. Solve continuity, momentum, energy, and species equations simultaneously. Solve turbulence, and other scalar equations. Converged? Stop Figure 3.3.2: Coupled solution method 3.4 Post-processing: FLUENT Post-processing is the last element in CFD package. All interpretation of numerical results as well as CFD image of simulation is produced in this stage. Beside as a solver, FLUENT also can be used during post-processing stage. 22 3.5 Pre-processing: Domain of Computational Figure 3.5.1 shows domain of computational for simulation of normal blood flow in artery. Figure 3.5.1: Normal blood flow in an artery Figure 3.5.2 shows domain of computational for simulation of abnormal blood flow in artery under heating condition. 23 Figure 3.5.2: Abnormal blood flow in an artery under heating condition Figure 3.5.3 shows domain of computational for simulation of normal blood flow in a tapered artery. Figure 3.5.3: Normal blood flow in a tapered artery 24 Figure 3.5.4 shows domain of computational for simulation of abnormal blood flow in a tapered artery under heating condition. Figure 3.5.4: Abnormal blood flow in a tapered artery under heating condition Figure 3.5.5 shows domain of computational for simulation of normal blood flow in a stenois artery. 25 Figure 3.5.5: Normal blood flow in a stenois artery Figure 3.5.6 shows domain of computational for simulation of abnormal blood flow in a stenois artery under heating condition. Figure 3.5.6: Abnormal blood flow in a stenois artery under heating condition 26 3.6 Solver: Parameters The following data for blood (Horng et al, 2007, and Shih et al, 2006) and water properties in Table 3.6.1 will be use during simulation: Table 3.6.1: Blood and water properties Properties Density, ρ Dynamics viscosity, µ Specific heat, c p Thermal conductivity, k Unit kg m3 kg ms Ws kg K Water at 310K 993 Blood at 310K 1060 6.95 ×10 −4 3.00 ×10 −3 4174 3770 W mK 0.628 0.500 4.6 22.6 Prandlt number, Pr The following variations will be use during simulation of normal blood flow in artery: Table 3.6.2: Variation for normal blood flow in an artery Case #1 Constant length Diameter #1 Diameter #2 Diameter #3 Vel. pattern #1 Vel. pattern #1 Vel. pattern #1 27 The following variations will be use during simulation of abnormal blood flow in artery under heating condition: Table 3.6.3: Variation for abnormal blood flow in an artery under heating condition Case #1 Constant heating source Case #2 Constant diameter Diameter #1 Diameter #2 Diameter #3 Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 Heat flow pattern Heat flow pattern Heat flow pattern #1 #2 #3 Heating source #1 Heating source #2 Heating source #3 Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 Heat flow pattern Heat flow pattern Heat flow pattern #1 #2 #3 The following variation will be use during simulation of abnormal blood flow under tapered (Case #1-3) and stenois (Case #4-6) channel: 28 Table 3.6.4: Variation for abnormal blood flow under tapered and stenois channel Case #1 Tapered angle #1 Tapered angle #2 Tapered angle #3 No heating Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 Case #2 Tapered angle #1 Tapered angle #2 Tapered angle #3 Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 Heat flow pattern Heat flow pattern Heat flow pattern #1 #2 #3 Case #3 Heating source #1 Heating source #2 Heating source #3 Constant tapered Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 angle Heat flow pattern Heat flow pattern Heat flow pattern #1 #2 #3 Case #4 Stenois height #1 Stenois height #2 Stenois height #3 No heating Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 Case #5 Stenois height #1 Stenois height #2 Stenois height #3 Constant heating Vel. pattern #1 Vel. pattern #2 Vel. pattern #3 source Heat flow pattern Heat flow pattern Heat flow pattern #1 #2 #3 Case #6 Heating source #1 Heating source #2 Heating source #3 Constant stenois Vel. pattern #1 Ve. pattern #2 Vel. pattern #3 height Heat flow pattern Heat flow pattern Heat flow pattern #1 #2 #3 Constant heating source 3.7 Solver: Method of Solution The segregated solver will be used in this study. In section 2.3, blood is assumed to be incompressible fluid. Therefore, the segregate solution method is sufficient to run this simulation. Under-relaxation factor for energy equation is set 29 to 1.0 since there are no temperature-dependent properties or buoyancy forces present. 3.8 Expected Outcome 1) Velocity and heat flow patterns when the vessels are heated at various degrees. 2) Comparative analysis between artery flow and mechanical systems under similar conditions to help us to understand future trends since mechanical systems tend to be more predictable. 30 CHAPTER IV MODEL VALIDATION 4.1 Model Validation Methodology Three models are constructed in order to evaluate all three cases as mentioned in Chapter 1. Model 1, Model 2 and Model 3 are constructed for a flow in straight tube geometry, a flow in tapered tube geometry, and a flow in stenois tube geometry respectively. Water is used as working fluid instead of blood for validation purposes. Grid dependency test is performed on Model 1 by doing the simulations under various meshing sizes. Each simulation results are then compared to correlation results. Grid dependency test is performed for Model 1 for two reasons. First, to ensure that selected meshing size is sufficient for the simulations. Second, to ensure the simulation results are converged and with agreement with theoretical results. 31 Four sizes of meshing are selected for this test. The following table shows summary of meshing size for grid dependency test. Table 4.1.1: Summary of meshing size for grid dependency test Mesh Type Element in X- Element in Y- Number of direction direction Elements Mesh A 50 10 500 Mesh B 100 20 2000 Mesh C 150 30 4500 Mesh D 200 40 8000 Local Nusselt numbers are calculated from simulation results for each mesh size then compared to correlation results. The following correlation equations are suggested Kakac and Yener (1995) for simultaneously developing region in a circular duct. Constant temperature difference (Langhaar velocity profile) Nu = 4.36 +  Re Pr   0.104   x  ï£ d   Re Pr   1 + 0.016   x  ï£ d  0.8 (4.1) Constant heat input (Parabolic velocity profile) Nu = 4.36 +  Re Pr   0.023   x  ï£ d   Re Pr   1 + 0.0012   x  ï£ d  0.8 (4.2) 32 Constant heat input (Langhaar velocity profile) Nu = 4.36 +  Re Pr   0.036   x  ï£ d   Re Pr   1 + 0.011  x  ï£ d  0.8 (4.3) Each meshing sizes are then compared with theoretical results for validation. The smallest size of meshing which is in agreement with theoretical results within 5% of error is the chosen as the optimum meshing size. The smallest size is selected because it consumes less calculation time compare to bigger one and still produces reasonable results. Once Model 1 is validated, it can be used for blood flow simulations. The blood flow simulations are performed for all cases as mentioned in Table 3.6.2 to Table 3.6.4 and results are shown in the following sections. 4.2 Grid Dependency Test The simulations for Model 1 are performed with fluid average inlet velocity is taken at 0.05m/sand constant wall heat flux is taken at q = 4W/m2. The radius is taken at r = 5mm. The following figures show results of the simulations for grid dependency test. 33 Grid Dependency Test 8 Grid Dependency Test 7 6 Difference (%) 5 4 3 2 1 0 1000 cells to 2000 cells 2000 cells to 4500 cells 4500 celss to 8000 cells -1 Meshing Improvement Figure 4.2.1: Percentage difference changing from one mesh size to another Iterations Convergence vs Number if Cell 800 700 Number of Iteration 600 500 400 300 200 100 0 0 1000 2000 3000 4000 5000 6000 7000 8000 Number of Cell Figure 4.2.2: Number of iteration versus number of cells Figure 4.2.1 shows that the local Nusselt number varies around 7.3% as meshing size is changed from 1000 cells to 2000 cells. As the meshing size is refined further from 2000 cells to 4500 cells, results virtually do not vary at all. 34 Same result shown as the meshing size is refined further more from 4500 cells to 8000 cells. Therefore, further refinement of meshing size give no better convergence results. Figure 4.2.2 shows number of iteration required to get convergence result versus number of cell. Number of iteration required increases as number of cells increase. From Figure 4.2.1 and 4.2.2, Mesh B is selected since it has the optimum size compare to other meshing size. 4.3 Model Validation with Numerical Results Comparison The following figure shows correlation plots of local Nusselt number variation for simultaneously developing region in a circular duct. Flow conditions are same as in grid dependency test and water properties are taken at 310K as shown in Table 3.6.1. 35 Variation of Local Nusselt Number 10 Langhaar Vel. Profile (Cons. Temp. Diff.) Parabolic Vel. Profile (Cons. Heat Input) 9 Langhaar Vel. Profile (Cons. Heat Input) Nux 8 7 6 5 4 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 4.3.1: Local Nusselt number for simultaneously developing region in a circular duct The following figure shows the simulation result with a parabolic velocity profile and a Langhaar velocity profile. Local Nusselt number for the simulation converge at dimensionless distance higher than 0.2. Percentage of error around 3% compares to parabolic velocity profile correlation and 1% compares to Langhaar velocity profile correlation. Higher percentage of error occurs at dimensionless distance less 0.2. Since present study only concentrates on fully developed region, higher percentage of error at the said region is ignored in this study. 36 Comparison of Local Nusselt Number 20 Simulation Parabolic Vel. Profile Langhaar Vel. Profile 18 16 Nux 14 12 10 8 6 4 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 4.3.2: Comparison of local Nusselt number with correlation results 37 CHAPTER V RESULTS AND DISCUSSIONS 5.1 Simulation of the Normal Blood Flow in an Artery (Straight Tube) The axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and zero wall heat flux boundary condition at various radii is shown in Figure 5.1.1. Maximum axial velocities occur at center of the tube and gradually become zero at the wall. At the same radius and average inlet velocity, the axial velocity of blood is quantitatively the same as that of water. This is because the velocity profile is only influenced by the average inlet velocity and tube diameter as stated in Equation 2.13. 38 Axial Velocities Pattern 7 Water (r = 5mm) Blood (r = 5mm) 6 Blood (r = 6mm) R adial Position (mm) Blood (r = 7mm) 5 4 3 2 1 0 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 U / Um Figure 5.1.1: Axial velocity of blood flow (q = 0W/m2, Um = 0.05m/s) at various radii Skin Friction Coefficient 0.10 Cf 0.08 0.06 Water (r = 5mm) Blood (r = 5mm) Blood (r = 6mm) 0.04 Blood (r = 7mm) 0.02 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 5.1.2: Skin friction coefficient of blood flow (q = 0w/m2, Um = 0.05m/s) at various radii 39 Figure 5.1.2 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and zero heat flux at various radii. The skin friction coefficient increases as the tube radius decreases. The skin friction coefficient remain constant at dimensionless length of tube greater than 0.1. This is because the skin friction coefficient does not depend on length but on radius. Water has lower skin friction coefficient compared to blood under the same conditions. This is due to the ratio of density and dynamics viscosity for water which is almost four times lower than blood in general. 5.2 Simulation of Abnormal Blood Flow in an Artery under Heating Condition (Straight Tube) Axial Velocities Pattern 7 Water (r = 5mm) Blood (r = 5mm) 6 Blood (r = 6mm) R adial Position (mm) Blood (r = 7mm) 5 4 3 2 1 0 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 U / Um Figure 5.2.1: Axial velocity of blood flow (q = 4W/m2, Um = 0.0 m/s) at various radii 40 The axial velocities of blood flow at constant average inlet velocity at 0.05m/s and constant heat flux of 4W/m2 boundary condition at various radii is shown in Figure 5.2.1. The profiles are similar to the case without flux. Heating of the fluid at 4W/m2 did not seem to affect the velocity profiles probably because it is too small. This observation is expected because the velocity profiles are, again, influenced by tube diameters and average inlet velocity only. A property that might have been affecting the velocity is the viscosity which with the heat flux applied did not affect the velocity profile here. Skin Friction Coefficient 0.10 Cf 0.08 0.06 Water (r = 5mm) Blood (r = 5mm) Blood (r = 6mm) 0.04 Blood (r = 7mm) 0.02 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 5.2.2: Skin friction coefficient of blood flow (q = 4W/m2, Um = 0.05m/s) at various radii Figure 5.2.2 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 under various radii. The skin friction coefficient values are comparable with Figure 5.1.2 which is simulated under zero heat flux. With or without heat being applied, the skin friction coefficient values do not change since it does not depend on the heat input. This is also because the velocity profile remains the same as shown in Figure 5.2.1. 41 Variation of Local Nusselt Number 16 Water (r = 5mm) Blood (r = 5mm) Blood (r = 6mm) Blood (r = 7mm) Nux 12 8 4 0.2 0.4 0.6 0.8 1.0 x/L Figure 5.2.3: Local Nusselt number of blood flow (q = 4W/m2, Um = 0.05m/s) at various radii Figure 5.2.3 shows the local Nusselt number of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 at various radii. Higher local Nusselt number occurs at dimensionless distance less than 0.6. This is because the flow is still in the thermally developing region. The local Nusselt number begins to converge after a dimensionless distance higher than 0.6. The local Nusselt number for blood in bigger tubes is higher compared to the smaller tubes. Bigger tubes have larger surface area which leads to higher heat absorption from externally applied heat flux. The local Nusselt number for water is clearly lower compared to blood under the same conditions at dimensionless distance less than 0.5. This is because water has a higher thermal conductivity (kwater = 0.628W/m0C) compared to blood (kblood = 0.500W/m0C) and Nusselt number is decreasing with increasing of thermal conductivity. Average Nusselt number of blood flow is found to be 5.14, 5.79 and 6.83 for radius of 5, 6 and 7mm respectively. Average Nusselt number of water flow is found to be 4.96 for 5mm tube. 42 Axial Velocities Pattern 5 Water (q = 4w/m^2) Blood (q = 4w/m^2) Blood (q = 5w/m^2) R adial P osition (mm) 4 Blood (q = 6w/m^2) 3 2 1 0 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 U / Um Figure 5.2.4: Axial velocity of blood flow (r = 5mm, Um = 0.05m/s) under various heating conditions Figure 5.2.4 shows the axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and tube radius of 5mm under various heating conditions. Maximum axial velocity occurs at the center of the tube and gradually becomes zero at the wall. Each heating condition yields the same velocity profiles as expected. Heating of the fluid at these ranges of heat flux did not seem to affect the velocity profiles probably because it is too small. The results here are identical to those in Figures 5.2.1 and 5.1.1. 43 Skin Friction Coefficient 0.10 Cf 0.08 0.06 Water (q = 4w/m^2) Blood (q = 4w/m^2) Blood (q = 5w/m^2) 0.04 Blood (q = 6w/m^2) 0.02 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 5.2.5: Skin friction coefficient of blood flow (r = 5mm, Um = 0.05m/s) under various heating conditions Figure 5.2.5 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and tube radius of 5mm under various heating conditions. As the applied heat flux increases, the skin friction remains unchanged. This is because the skin friction coefficient does not depend on heat input (see Equation #2.14). 44 Variation of Local Nusselt Number 16 Water (q = 4 w/m^2) Blood (q = 4 w/m^2) Blood (q = 5 w/m^2) Blood (q = 6 w/m^2) Nux 12 8 4 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 5.2.6: Local Nusselt number of blood flow (r = 5mm, Um = 0.05m/s) under various heating conditions Figure 5.2.6 shows the local Nusselt number of blood flow simulations at constant average inlet velocity of 0.05m/s and tube radius of 5mm under various heating conditions. The local Nusselt number is higher in thermally developing region which is at dimensionless distance less than 0.4. The local Nusselt number converges to a constant value for dimensionless distance higher than 0.4. Blood under various heating conditions show almost identical local Nusselt number. Water produces lower local Nusselt number compared to blood under the same conditions as clearly shown at dimensionless distance less than 0.5. Comparison of the results for this study and those using the correlation used by Langhaar is shown in Figure 5.2.8. Using his correlation showed slight differences between water and blood as seen in Figure 5.2.7. Simulation results showed that large disagreement occurs at the inlet developing region. It maybe possible to modify Langhaar’s correlation to be used for blood but further investigation will be required and this is beyond the scope of this study. 45 Variation of Local Nusselt Number 10 Parabolic Vel. Pro. (Cons. Heat Input Blood) Langhaar Vel. Pro (Cons. Heat Input Blood) 9 Par. Vel. Pro. (Cons. Heat Input Water) Langhar Vel. Pro. (Cons. Heat Input Water) Nux 8 7 6 5 4 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 5.2.7: Plot of correlation of local Nusselt number for blood and water flow Comparison of Local Nusselt Number 14 Simulation Blood Parabolic Vel. Profile Blood Langhaar Vel. Profile Blood 13 12 11 Nux 10 9 8 7 6 5 4 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 x/L Figure 5.2.8: Plot of simulation and correlation of local Nusselt number for blood flow 46 Variation of Local Nusselt Number 5.0 Water (q = 4 w/m^2) Blood (q = 4 w/m^2) Blood (q = 5 w/m^2) 4.9 Blood (q = 6 w/m^2) Nux 4.8 4.7 4.6 4.5 0.90 0.95 x/L Figure 5.2.9: Zoom- in of Figure 5.2.6 Figure 5.2.9 shows the zoom in region of Figure 5.2.6. Local Nusselt number converges to a constant value as expected. Blood under higher applied heat flux produces slightly higher local Nusselt number. This is because local Nusselt number is directly proportional to the amount of applied heat. Water produces slightly higher local Nusselt number compared to blood under the same imposed conditions at fully developed region. Possible explanation for this situation is due to water has lower skin friction coefficient compared to blood as shown in Figure 5.2.5. Average Nusselt number is found to be 5.14 for each case of applied heat flux. Figures 5.1.1, 5.2.1 and 5.2.4 have shown that blood profiles follow the general characteristic of flow in a circular straight tube. Furthermore, axial velocity profiles of blood do not change with the increasing amount heat applied in the simulation. 47 Figures 5.2.3 and 5.2.9 have shown that the tube radius does affect the local Nusselt numbers when the amount of heat flux applied was varied. Bigger radius of tube absorbs more heat compared to smaller one. 5.3 Simulation of Blood Flow in an Artery (Tapered and Stenois Tubes) 5.3.1 Simulation of Blood Flow in an Artery (Tapered Tube) Axial Velocities Pattern Out: 0.0 degree D im en sio n less R ad ial P o sitio n at T ap ered Ou tlet Out: 0.5 degree Out: 1.0 degree Out: 1.5 degree 0 0.5 1 1.5 U / Um 2 2.5 3 3.5 Figure 5.3.1.1: Outlet axial velocity of blood flow (q = 0W/m2, Um = 0.05m/s) under various taper angles The axial velocity (at tapered outlet) of blood flow simulations under various converging taper angles is shown in Figure 5.3.1.1. The simulations were completed at constant average inlet velocity of 0.05m/s and zero heat flux. 48 Maximum axial velocity occurs at the tube center and gradually becomes zero at the wall for the no slip condition. As the tapered angle increases, maximum axial velocity at the center of the tube also increases. It is also observed that velocity profile for bigger tapered angle assumes flat shape instead of parabolic shape. This observation can be explained using conservation of mass principle. To conserve mass, the inflow rate must equal the outflow rate. Since increasing of the tapered angle reduces the cross sectional area, thus average velocity increases. This result agrees qualitatively well with Chakravarty and Mandal (2000). It is noted that Chakravarty and Mandal (2000) used different geometrical dimensions as compared to present study. They found that the velocity profile increased with the increase of the tapered angle. Skin Friction Coefficient 0.30 Taper 0.0 degree Taper 0.5 degree 0.25 Taper 1.0 degree Taper 1.5 degree Cf 0.20 0.15 0.10 0.05 0.0 0.2 0.4 0.6 0.8 1.0 x/L Figure 5.3.1.2: Skin friction coefficient of blood flow (q = 0W/m2, Um = 0.05m/s) under various taper angles Figure 5.3.1.2 shows the skin friction coefficients of blood flow simulations at constant average inlet velocity of 0.05m/s and zero heat flux under various converging tapered angles. It is observed that the skin friction coefficient values are the same value at the tapered inlet. This is because the tube diameters are the same 49 at the tapered inlet region. The skin friction coefficient for zero tapered angle remains constant for the entire dimensionless length of tapered region. On the other hand, the skin friction coefficient increases as the dimensionless distance from tapered region inlet increases. Increments of skin friction coefficient at a particular angle seem to occur at a constant rate. Higher converging taper angle has a higher rate of increment. For the 0.50 angle, the gradient is constant at 0.04. For the 1.00 angle, the gradient is constant at 0.11. For the 1.50 angle, however, the gradient is 0.17 until x/L = 0.5 and then increases further with a gradient of 0.25. This observation is expected because the skin friction is inversely proportional to the Reynolds number as stated in Equation #2.14. Reynolds number is directly proportional to diameter. As the tapered angle increases, the diameter decreases and lead to decreasing in the Reynolds number. Axial Velocities Pattern Out: 0.0 degree D im en sio n less R ad ial P o sitio n at T ap ered Ou tlet Out: 0.5 degree Out: 1.0 degree Out: 1.5 degree 0 0.5 1 1.5 U / Um 2 2.5 3 3.5 Figure 5.3.1.3: Axial velocity of blood flows (q = 4W/m2, Um = 0.05m/s) under various taper angles Figure 5.3.1.3 shows the axial velocity (at tapered outlet) of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 under various converging taper angles. As the tapered angle increases, 50 maximum axial velocity at the center of the tube also increases. Figures 5.3.1.3 and 5.3.1.1 are identical since the axial velocity patterns remain the same with the amount of heat flux applied. This is because the velocity profile is influenced by the tube diameter and average inlet velocity only. Refer to Appendix B to D for axial velocity profile plots from Fluent. Skin Friction Coefficient 0.30 Taper 0.0 degree Taper 0.5 degree 0.25 Taper 1.0 degree Taper 1.5 degree Cf 0.20 0.15 0.10 0.05 0.0 0.2 0.4 0.6 0.8 1.0 x/L Figure 5.3.1.4: Skin friction coefficient of blood flow (q = 4W/m2, Um = 0.05m/s) under various taper angles Figure 5.3.1.4 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 under various converging tapered angles. The skin friction coefficient values are the same as that under no heat flux condition. This observation is expected since the average inlet velocity does not change with or without heat flux under present study. Skin friction coefficient also is not affected by the heat flux as stated in Equation #2.14. 51 Variation of Local Nusselt Number 5.0 N ux 4.5 4.0 Taper 0.0 degree 3.5 Taper 0.5 degree Taper 1.0 degree Taper 1.5 degree 3.0 0.0 0.2 0.4 0.6 0.8 1.0 x/L Figure 5.3.1.5: Local Nusselt numbers of blood flow (q = 4W/m2, Um = 0.05m/s) under various taper angles Figure 5.3.1.5 shows the local Nusselt number of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 under various converging taper angles. It is observed that the local Nusselt numbers are at the same at the tapered inlet region. The local Nusselt numbers for zero tapered angle remains constant. As the tapered angle increases, the local Nusselt number decreases at almost a constant rate. Decrement trends of the local Nusselt number can be explained from a geometrical point of view. As the tapered angle increases, the surface area decreases. Therefore, less heat is able to be absorbed by blood as it passes through and it leads to a lower local Nusselt number. Average Nusselt number is found to be 5.14, 4.60, 4.37 and 4.04 for 0.0, 0.5, 1.0 and 1.5 taper angles respectively. 52 Axial Velocities Pattern 5 R ad ial P o sitio n at T ap ered O u tlet (m m ) q = 4w/m^2 q = 5w/m^2 q = 6w/m^2 4 3 2 1 0 0 0.5 1 U / Um 1.5 2 2.5 Figure 5.3.1.6: Axial velocity of blood flow (Um = 0.05m/s, θ = 0.50) under various heating conditions Figure 5.3.1.6 shows the axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and constant converging tapered angle of 0.50 under various heating conditions. Maximum axial velocity occurs at the center of the tube and gradually becomes zero at the wall. Each heating condition yields the same profile as expected. This is because the velocity profile does not change with the changing of amount of applied heat unless the viscosity is greatly affected. The heat flux applied in this study seemed insignificant enough to influence the velocity profile. 53 Skin Friction Coefficient 0.20 q = 4w/m^2 q = 5w/m^2 q = 6w/m^2 Cf 0.15 0.10 0.05 0.0 0.2 0.4 0.6 0.8 1.0 x/L Figure 5.3.1.7: Skin friction coefficient of blood flow (Um = 0.05m/s, θ = 0.50) under various heating conditions Figure 5.3.1.7 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and constant converging tapered angle of 0.50 under various heating conditions. Each heating condition produces identical skin friction coefficient value. The skin friction coefficient increases as the dimensionless distance from tapered inlet region increases. Increment of the skin friction coefficient occurs at a same constant rate. The amount of heat flux applied in this study is quite insignificant to influence the velocity profile change as mentioned earlier, the skin friction coefficient value remains the same under various heating conditions. 54 Variation of Local Nusselt Number 4.8 q = 4 w/m^2 q = 5 w/m^2 q = 6 w/m^2 Nux 4.6 4.4 4.2 4 0.0 0.2 0.4 0.6 0.8 1.0 x/L Figure 5.3.1.8: Local Nusselt number of blood flow (Um = 0.05m/s, θ = 0.50) under various heating conditions Figure 5.3.1.8 shows the local Nusselt number of blood flow simulations at constant average inlet velocity of 0.05m/s and constant converging tapered angle of 0.50 under various heating conditions. The local Nusselt number shows downward trend at almost a constant rate. The highest local Nusselt number occurs at the inlet of the tapered region while the lowest local Nusselt number occurs at the outlet tapered region. Note that the local Nusselt number is higher for higher amount of applied heat flux. This observation is expected since local Nusselt number is directly proportional to heat transfer coefficient and heat transfer coefficient is directly proportional to heat flux. Therefore, as amount of applied heat flux increases, heat transfer coefficient increases, thus local Nusselt number increases as well. It is also observed that local Nusselt number dipped close to the outlet region. Increasing the amount of heat flux seems to move this sudden decrease region further downstream. However, further studies into lengthening the tapered region may show if this trend is repeated with higher heat flux. Average Nusselt number is found to be 4.64, 4.65 and 4.66 for 4, 5 and 6W/m2 heat flux respectively. Refer to Appendix E to G for temperature contour plots from Fluent. 55 Figures 5.3.1.1, 5.3.1.1 and 5.3.1.6 showed that blood follows the general characteristic of flow in a converging tapered tube. Axial velocity profiles of blood do not change regardless of the heat being applied. Profiles are identical for both straight and converging tapered tube. Figures 5.3.1.2, 5.3.1.4 and 5.3.1.7 showed that skin friction coefficient does not vary with applied heat but vary with tapered angle. This is because tapered region has different cross sectional area which leads to a changing radius and consequently to a changing area. If tapered angle was zero, it is completely analogous to constant radius tube cases as shown in Figures 5.1.2, 5.2.2 and 5.2.5. Figures 5.3.1.5 and 5.3.1.8 showed that applied heat flux has a less significant effect on local Nusselt number compared to that from the variation of tapered angle. Rate of changes of local Nusslet number are different for each case in Figure 5.3.1.5. This is because as dimensionless distance changes, radius also changes due to the tapered angle. Different tapered angle produces different rate of radius reductions. Therefore, local Nusslet number decreases at a difference rate. On the other hand, in Figure 5.3.1.8, the rate of changes of local Nusselt number remains the same under various heating conditions in this study. This is because the amount of heat flux being applied is constant as dimensionless distance moves from inlet to outlet. 56 5.3.2 Simulation of Blood Flow in an Artery (Stenois Tube) Axial Velocities Pattern 5 Cen: a / r = 0.2 Cen: a / r = 0.3 Radial Position (mm) 4 Cen: a / r = 0.4 Out: a / r = 0.2 Out: a / r = 0.3 3 Out: a / r = 0.4 2 1 0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 U / Um Figure 5.3.2.1: Axial velocity of blood flow (q = 0W/m2, Um = 0.0 m/s) with various stenois heights The axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and zero heat flux under various stenois heights is shown in Figure 5.3.2.1. It is observed that the maximum axial velocity occurs at the center of the tube. The axial velocity gradually reduces to zero at the tube wall. Velocity shapes gradually changed from a parabolic profile to a flat profile at the center of the stenois. Note that all stenois cases have identical velocity profiles at the inlet region of the stenois. This observation can be explained using conservation of mass principle. To conserve mass, the inflow rate must equal the outflow rate. Since increases of stenois height reduces the cross sectional area, thus average velocity increases at that region. This leads to a pressure decrease, according to the Bernoulli equation. As the stenois height reduces, the cross sectional area increases, thus average velocity decreases. In this region, the pressure increases. This behavior is analogous to flow in converging-diverging nozzle. This observation 57 agrees well with Chakravarty and Mandal (2000). They also found that as the stenois height increases, the velocity increases. It is also found that the velocity profile at the center of the stenois is comparable to the converging tapered case as shown in Figure 5.3.1.1. Center Axial Velocities Pattern 4.0 3.5 Uc / Um 3.0 2.5 a / r = 0.2 a / r = 0.3 2.0 a / r = 0.4 1.5 0 0.2 0.4 x/L 0.6 0.8 1 Figure 5.3.2.2: Center axial velocity of blood flow (q = 0W/m2, Um = 0.05m/s) with various stenois heights Figure 5.3.2.2 shows the center axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and zero heat flux under various stenois heights. It is observed that the maximum dimensionless center axial velocity occurs at the center of the stenois. As the stenois height increases, maximum dimensionless center axial velocity increases. This observation is expected since the flow simulated here at the vena contracta is analogous to the flow in a converging-diverging nozzle. It is also observed that the dimensionless center axial velocity at outlet region is higher compared to the inlet region. This is because the pressure is not fully recovered after the stenois due to possible of backflow. This observation agrees well with Mukhopadhyay and Layek (2008), their axial velocity profile being similar to that in Figure 5.3.2.2. 58 Skin Friction Coefficient a / r = 0.2 0.8 a / r = 0.3 a / r = 0.4 Cf 0.6 0.4 0.2 0.0 0 0.2 0.4 0.6 0.8 1 x/L Figure 5.3.2.3: Skin friction coefficient of blood flow (q = 0W/m2, Um = 0.05m/s) with various stenois heights Figure 5.3.2.3 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and zero heat flux under various stenois heights. It is observed that the skin friction coefficient is higher for higher stenois height at dimensionless distance less than 0.6. For dimensionless distance more than 0.6, the highest stenois height does not produce a higher skin friction coefficient. This is due to the skin friction coefficient decreasing with increasing product of average velocity and tube diameter. Changes of skin friction coefficient are due to the stenois increasing as the stenois height increases. The skin friction coefficient rate of change at both inlet and outlet region of stenois are different since the rate of change of axial velocities are different as well, as shown in Figure 5.3.2.2. This is because the skin friction affects the Reynolds number and the Reynolds number influences the fluid velocity. Refer to Appendix H to J for axial velocity profile plots from Fluent. 59 Axial Velocities Pattern 5 Cen: a / r = 0.2 Cen: a / r = 0.3 Radial Position (mm) 4 Cen: a / r = 0.4 Out: a / r = 0.2 Out: a / r = 0.3 3 Out: a / r = 0.4 2 1 0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 U / Um Figure 5.3.2.4: Axial velocity of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights Figure 5.3.2.4 shows the axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4w/m2 under various stenois heights. It is observed that the maximum axial velocity occurs at the center of the tube. The axial velocity gradually reduces to zero at the tube wall. Velocity shapes gradually changed from a parabolic profile to a flat profile at the center of the stenois. All stenois cases have identical velocity profiles at the inlet region of stenois. It is noted that the velocity profiles are identical to Figure 5.3.2.1 for the case of zero heating. This is because only the tube diameter and the velocity affect the velocity profile. 60 Center Axial Velocities Pattern 4.0 3.5 Uc / Um 3.0 2.5 a / r = 0.2 a / r = 0.3 2.0 a / r = 0.4 1.5 0 0.2 0.4 x/L 0.6 0.8 1 Figure 5.3.2.5: Center axial velocity of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights Figure 5.3.2.5 shows the center axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 under various stenois heights. The center axial velocity values are similar to that in Figure 5.3.2.2 which is simulated under zero heat flux condition. 61 Skin Friction Coefficient a / r = 0.2 0.8 a / r = 0.3 a / r = 0.4 Cf 0.6 0.4 0.2 0.0 0 0.2 0.4 0.6 0.8 1 x/L Figure 5.3.2.6: Skin friction coefficient of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights Figure 5.3.2.6 shows the skin friction coefficients of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4w/m2 under various stenois heights. The skin friction coefficient values are similar to those in Figure 5.3.2.3 which is simulated under zero heat flux condition. 62 Variations of Local Nusselt Numbers a / r = 0.2 5.0 a / r = 0.3 a / r = 0.4 4.5 Nux 4.0 3.5 3.0 2.5 2.0 0 0.2 0.4 0.6 0.8 1 x/L Figure 5.3.2.7: Local Nusselt number of blood flow (q = 4W/m2, Um = 0.05m/s) with various stenois heights Figure 5.3.2.7 shows the local Nusselt number of blood flow simulations at constant average inlet velocity of 0.05m/s and constant heat flux of 4W/m2 under various stenois heights. The local Nusselt number decreases as the stenois height increases. The local Nusselt number is at minimum at the center of the stenois height. Higher stenois height has lower local Nusselt number in general. This is because higher stenois height has a smaller radius and the smaller radius leads to a smaller surface area. Therefore, less heat is able to be absorbed and it leads to a lower local Nusselt number. It is also observed that local Nusselt number at the inlet is slightly higher compared to the outlet. This could be due to the cooling effect since the outlet region of stenois height has higher axial velocity compared to the inlet region (Figure 5.3.2.5). For a dimensionless distance less than 0.5, the local Nusselt number is comparable to the tapered case as shown in Figure 5.3.1.5. Average Nusselt number is found to be 4.12, 3.80 and 3.49 for stenois height ratio of 0.2, 0.3 and 0.4 respectively. 63 Axial Velocities Pattern 5 In: q = 4w/m^2 In: q = 5w/m^2 In: q = 6w/m^2 Radial Position (mm) 4 Cen: q = 4w/m^2 Cen: q = 5w/m^2 Cen: q = 6w/m^2 Out: q = 4w/m^2 3 Out: q = 5w/m^2 Out: q = 6w/m^2 2 1 0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 U / Um Figure 5.3.2.8: Axial velocity of blood flow (a / r = 0.2, Um = 0.05m/s) under various heating conditions Figure 5.3.2.8 shows the axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and constant stenois height of 1mm under various heating conditions. It is observed that maximum axial velocity occurs at the center of the tube. The axial velocity gradually reduces to zero at the tube wall. Velocity shapes gradually changed from a parabolic profile to a flat profile at the center of the stenois. It is noted that all heating conditions cases have identical velocity profiles as in the tapered case discussed previously. 64 Center Axial Velocities Pattern 2.75 q = 4 w/m^2 q = 5 w/m^2 q = 6 w/m^2 Uc / Um 2.50 2.25 2.00 1.75 0 0.2 0.4 x/L 0.6 0.8 1 Figure 5.3.2.9: Center axial velocity of blood flow (a / r = 0.2, Um = 0.05m/s) under various heating conditions Figure 5.3.2.9 shows the center axial velocity of blood flow simulations at constant average inlet velocity of 0.05m/s and constant stenois height of 1mm under various heating conditions. It is observed that maximum dimensionless center axial velocity occurs at the center of the stenois. It is noted that all heating conditions cases have identical velocity profiles. 65 Skin Friction Coefficient q = 4w/m^2 0.25 q = 5w/m^2 q = 6w/m^2 0.20 Cf 0.15 0.10 0.05 0.00 0 0.2 0.4 0.6 0.8 1 x/L Figure 5.3.2.10: Skin friction coefficient of blood flow (a / r = 0.2, Um = 0.05m/s) under various heating conditions Figure 5.3.2.10 shows the skin friction coefficient of blood flow simulations at constant average inlet velocity of 0.05m/s and constant stenois height of 1mm under various heating conditions. It is observed that maximum skin friction coefficient occurs at the center of the stenois. It is noted that all heating conditions cases have identical skin friction coefficient values. The skin friction coefficient values remain the same regardless of the amount of heat flux applied. 66 Variation of Local Nusselt Numbers 4.8 q = 4 w/m^2 4.6 q = 5 w/m^2 q = 6 w/m^2 4.4 Nux 4.2 4.0 3.8 3.6 3.4 0 0.2 0.4 0.6 0.8 1 x/L Figure 5.3.2.11: Local Nusselt number of blood flow (a / r = 0.8, Um = 0.05m/s) under various heating conditions Figure 5.3.2.11 shows the local Nusselt number of blood flow simulations at constant average inlet velocity of 0.05m/s and constant stenois height of 1mm under various heating conditions. It is observed that minimum local Nusselt number occurs at the center of the stenois height. It is noted that the local Nusselt number variations with the variation in heat flux are small at both the inlet and outlet of the stenois regions. It is observed that the local Nusselt number fluctuated at the downstream of stenois as well as upstream. This was observed in the tapered case (Figure 5.3.1.8). The phenomenon could be due to the recirculation caused by the sudden decrease in flow region at the speed studied. The increase in fluctuations after the stenois could be contributed by the slow decrease in velocity after the stenois (Figure 5.3.2.9). Average Nusselt number is found to be 4.12, 4.15 and 4.17 for 4, 5 and 6W/m2 heat flux respectively. Refer to Appendix K to M for temperature contour plots from Fluent. Figures 5.3.2.1, 5.3.2.4 and 5.3.2.8 showed that blood follows general characteristic of flow in stenois tapered tube. Axial velocity profiles of blood do 67 not change with the amount of heat being applied under the stenois condition. The observations are identical for straight tube, converging tapered tube and stenois tube. Figures 5.3.2.2, 5.3.2.5 and 5.3.2.9 showed that maximum center axial velocities occur at the center of stenois. Under the same stenois condition, center axial velocities do not change with the amount of heat being applied. Figures 5.3.2.3, 5.3.2.6 and 5.3.2.10 showed that skin friction coefficient does not vary with applied heat but vary with stenois height. If stenois height was zero, it is completely analogous to the constant radius tube cases as shown in Figures 5.1.2, 5.2.2 and 5.2.5. 68 CHAPTER VI CONCLUSION 6.1 Conclusion Simulation of blood flow in a straight tube under various radius and heating conditions were completed. Both blood and water flows exhibit identical axial velocity patterns for flow in a straight circular tube. The velocity, skin friction coefficient and local Nusselt number were found to be unaffected by the heating conditions, the latter changed insignificantly with variations in tube radius. Blood flow simulations in converging tapered tube were performed under various tapered angles and heating conditions. Axial velocities under zero heat flux agreed well with past results of Chakravarty and Mandal (2000). Axial velocity patterns did not change under various heating conditions but did so under various tapered angles. Skin friction coefficients did not vary under various heating conditions but did under various tapered angles. Local Nusselt numbers decreased as dimensionless distance from tapered region increased. Blood flow simulations in stenois tube were performed under various stenois heights and heating conditions. Center axial velocity pattern under zero heat 69 flux agreed well with past work of Mukhopadhyay and Layek (2008). Axial velocity patterns did not change under various heating conditions but did under various stenois heights. Skin friction coefficients did not vary under various heating conditions but did under various stenois heights. Local Nusselt numbers varied with stenois heights with small variations under various heating conditions. The results in general have shown that blood flow in biological systems may well behave like the mechanical systems. Since experimental studies on blood flow in biological system are near impossible to be conducted under desirable conditions, their behavior can possibly be predicted with the current findings and future extensions of the current work. It has been demonstrated that heat transfer in blood flow within tapered cross-section is affected by various heating conditions. This could be further investigated in terms of the various parts of the biological system if and when responding to thermal treatment. 70 REFERENCES Ashrafizaadeh, M, and Bakshaei, H, A Comparison of Non-Newtonian Models for Lattice Boltzmann Blood Flow Simulations, Computer and Mathematics with Application (2009) Bijan, A, Convective Heat Transfer, Wiley Interscience Publication (1984) Chandran, K.B., Yoganathan, A.P., and Rittgers, S.E., Biofluid Mechanics: The Human Circulation, Taylor and Francis (2007) Chakravarty, S. and Mandal, P.K, Two Dimensional Blood Flow through Tapered Arteries under Stenotic Conditions, International Journal of Non-linear Mechanics (2000) Chakravarty, S, and Sen, S, Dynamics Response of Heat and Mass Transfer in Blood Flow through Stenosed Bifurcation Arteries, Korea-Australia Rheology Journal, Vol 17, No. 2 (June 2005) Espino, D.M., Watkins, M.A., Shepherd, D.E., Hukins, D.W., and Buchan, K.G., Simulation of Blood Flow through Mitral Valve of the Heart: A Fluid Structure Interaction Model, COMSOL Users Conference (2006) Birmingham Fluent Incorporation, Fluent 6.2 User’s Guide, Fluent Incorporation (2005) Goshdastidar, P.S, Heat Transfer, Oxford University Press (2004) Jafari, A., Mousavi, S.M., and Kolari, P, Numerical Simulation of Blood Flow: Part 1: In Micro-vessels Bifurcations, Communications in Nonlinear Science and Numerical Simulation 12 (2008) 1615-1626 Jafari, A, Zamankhan, P, Mousavi, S.M., Kolari, P, Numerical Investigation of Blood Flow. Part II: In Capillaries, Communication in non-Linear Science and Numerical Simulations 14 (2009) 1396-1402 71 He, Y, Liu, H, Himeno, R, Sunaga, J, Kakusho, N, and Yokota, H, Finite Element Analysis of Blood Flow and Heat Transfer in an Image-based Human Finger, Computer in Biology and Medicine 38 (2008) 555-562. Horng, T.L, Lin, W.L, Liauh, C.T, Shih, T.C, Effects of Pulsative Blood Flow in Large Vessels on Thermal Dose Distribution during Thermal Theraphy, Medical Physic 34 (4) (April 2007) Hunter, P.J, Numerical Simulation of Arterial Blood Flow, University of Auckland, Masters Thesis (1972) Kakac, S and Yener, Y, Convetive Heat Transfer, CRC Press (1995) Khaled, A.R.A, and Vafai, K, The Role of Porous Media in Modeling Flow and Heat Transfer in Biological Tissues, International Journal of Heat and Mass Transfer 46 (2003) Mukhopadhay, S and Layek, G.C, Numerical Modeling of a Stenosed Artery using Mathematical Modeling of Variable Shape, Applications and Applied Mathematics: An International Journal Vol. 3, No. 2 (December 2008) pp. 308-328 Olufsen M.S, Peskin, C.S, Kim, W.Y, Pedersen, E.M, Nadim, A, and Larsen, J, Numerical Simulation and Experimental Validation of Blood Flow in Arteries with Structured-Tree Outflow Conditions, Annals of Biomedical Engineering, Vol. 28, pp. 1281-1299 (2000) Qioa, A, and Liu, Y, Medical Application Oriented Blood Flow Simulation, Clinical Biomechanics 23 (2008) S130-S136 Santos, I.D, Haemmerich, D, Pinheiro, C.D.S, Rocha, A.F.D, Effect of Variable Heat Transfer Coefficient on Tissue Temperature Next to a Large Vessel during Radiofrequency Tumor Ablation, Biomedical Engineering Online, (2008) Shih, T.C, Liu, H.L, Horng, T.L, Cooling Effect of Thermally Significant Blood Vessels in Perfused Tumor Tissue during Thermal Therapy, International Communications in Heat and Mass Transfer 33 (2006) 135-141 Tsubota, K.I, Wada, S, Kamada, H, Kitagawa, Y, Lima, R, and Yamaguchi, T, A Particle Method for Blood Flow Simulation – Application to Flowing Red 72 Blood Cells and Platelets, Journal of Earth Simulator, Volume 6, (March 2006) 2-7 Waite, L and Fine, J Applied Bio-fluid Mechanics, McGraw Hill (2007) Yuan, P, Numerical Analysis of an Equivalent Heat Transfer Coefficient in a Porous Model for Simulating a Biological Tissue in a Hyperthermia Therapy, International Journal of Heat and Mass Transfer 52 (2008) 17341740 73 APPENDICES Appendix-A: Blood properties set up in Fluent 74 Appendix-B: Axial velocity profile for 0.50 tapered tube (q = 4W/m2, Um = 0.05m/s) Appendix-C: Axial velocity profile for 1.00 tapered tube (q = 4W/m2, Um = 0.05m/s) 75 Appendix-D: Axial velocity profile for 1.50 tapered tube (q = 4W/m2, Um = 0.05m/s) Appendix-E: Temperature contour for 0.50 tapered tube (q = 4W/m2, Um = 0.05m/s) 76 Appendix-F: Temperature contour for 0.50 tapered tube (q = 5W/m2, Um = 0.05m/s) Appendix-G: Temperature contour for 0.50 tapered tube (q = 6W/m2, Um = 0.05m/s) 77 Appendix-H: Axial velocity profile for stenois height of 1mm (q = 0W/m2, Um = 0.05m/s) Appendix-I: Axial velocity profile for stenois height of 1.5mm (q = 0W/m2, Um = 0.05m/s) 78 Appendix-J: Axial velocity profile for stenois height of 2mm (q = 0W/m2, Um = 0.05 m/s) Appendix-K: Temperature contour for stenois height of 1mm (q = 4W/m2, Um = 0.05 m/s) 79 Appendix-L: Temperature contour for stenois height of 1mm (q = 5W/m2, Um = 0.05 m/s) Appendix-M: Temperature contour for stenois height of 1mm (q = 6W/m2, Um = 0.05 m/s)