Acknowledgements QA for the TomoTherapy Hi ™ Hi--Art Art™ Katja Langen, PhD Sanford Meeks, PhD Disclosures MDACCO has received grant funding from TomoTherapy, Inc. Robert J. Staton, PhD Medical Physicist Topics The TomoTherapy units at MD Anderson Orlando • Review of the TomoTherapy system Tomo1 (SN 02) Tomo2 (SN 50) • Commissioning and Twinning • QA Program • Routine • After Component Replacement • Patient DQA • TQA TomoTherapy TomoTherapy • Helical delivery of IMRT 6 MV Accelerator Primary Collimator • Same hardware is used for MVCT Binary MLC 85cm (64 leaves – 6.25mm ) • IGRT for every patient 85 cm Gantry Aperture 40 cm CT FOV Approximately 50cm TomoTherapy TomoTherapy Fan beam, continuously rotates Patient Divides beam into 64 beamlets each beamlet can be on/off for a percentage of each projection Target OAR TomoTherapy TomoTherapy Fan beam, Fan beam, continuously rotates Patient continuously rotates Target OAR TomoTherapy TomoTherapy Fan beam, Fan beam, continuously rotates continuously rotates Fan Beam lateral Helical Tomotherapy longitudinal 400 mm lateral 10, 25, or 50 mm longitudinal Delivery - 51 projections per rotation: 360º/51=7º -Temporal modulation leaf opening time during projection -Typical gantry rotation: 15-30 seconds -Typical treatment time: Prostate: 5 Min H&N: 7 Min from: W. Kalender, Computed Tomography: Fundamentals, System Technology, Image Quality, Applications Delivery Instructions (Sinograms) Sinogram Sinogram 1 gantry rotation Sinogram Megavoltage CT images 6 MV Accelerator (tuned to 3.5 MV for MVCT) Primary Collimator ( 5 mm Slice Width) Binary MLC 85cm (all leafs open during MVCT) 85 cm Gantry Aperture 40 cm CT FOV Approximately 50cm Commissioning IGRT using megavoltage CT images Prostate Head + Neck • Done in conjunction with TomoTherapy physicist • Machines are assembled and tested in the factory and then partially disassembled for shipment • Verify that the beam properties matches the machine data taken in the factory and the gold standard • TomoTherapy TPS uses gold standard data for all dose calculations Commissioning Tests • • Hardware Alignment • Source position • Jaws • MLC • Lasers Beam • Output Stability • Beam Profiles - Transverse, Beam Width, PDD • Static Output Calibration • MVCT image quality • IMRT phantom tests Water Tank Standard Imaging – 2D Scanning Tank Scanning Data Scanning Data PDD 25 mm Field Width Cone 25 mm Field Width 15mm Calculated 50mm Calculated 100 100mm Calculated 100 Calculated 150mm Calculated Measured 90 90 200mm Calculated Install measured 15mm Install Measured 80 80 70 70 50mm Install Measured 100mm Install Measured Dose (Normalized) Dose (Normalized) 150mm Install Measured 60 50 40 200mm Install Measured 60 50 40 30 30 20 20 10 10 0 0 0 5 10 15 20 25 -300 -200 -100 0 Distance (cm) 100 200 300 Distance (mm) Scanning Data Commissioning - IMRT Tests Penumbra 25 mm Field Width • On-axis and off-axis target for each beam 15mm Calculated 50mm Calculated 100 100mm Calculated 90 150mm Calculated 80 Calculated & Delivered Dose: 25mm Beam, On-Axis Tumor 200mm Calculated Calculated & Delivered Dose: 25mm Beam, Off-Axis Tumor Calculated Calculated 15mm Install Measured 70 2.0 Measured 2.0 Measured Dose (Normalized) 50mm Install Measured 60 1.5 1.5 Dose (Gy) 150mm Install Measured Dose (Gy) 100mm Install Measured 50 1.0 1.0 200mm Install Measured 40 0.5 0.5 30 -20 10 0 -25 -20 -15 -10 -5 0.0 0.0 20 0 Distance (mm) 5 10 15 20 25 -15 -10 -5 0 Distance (cm) 5 10 15 20 -20 -15 -10 -5 0 Distance (cm) 5 10 15 20 Twinning • For multiple machines, you can match the beam characteristics After Commissioning • • Recommissioning • Facilitates the transfer of patients from one machine to the other in case of down time • End to End phantom tests • H&N phantom • Pelvis phantom • Cheese phantom • Film & Ion Chamber DQAs for phantom plans and test cases • Independent calibration check - RPC TLDs • Setup your QA program • • QA Program Daily QA Reference data for monthly and part replacement QA • TMR/PDD in solid water • TomoDose Profiles Daily Output Constancy Routine: Daily QA Output measurements: Daily, Patient DQA, Monthly, Annual • Static procedure, gantry at 0°, 5x40 cm field • Readings taken at a depth of 5 cm and 10 cm After Component Replacement: Magnetron, MLC, Target Daily Output Summary Daily Output Constancy Hi*Art II, S/N 002 850 +5% 840 830 +3% +2% 810 800 790 -2% -3% 780 770 760 - Very temperature sensitive -if machine is not at operating temp. (38-40 ºC) output measurement can be unreliable -5% 17-Feb-08 7-Sep-07 28-Nov-07 17-Jun-07 4-Jan-07 27-Mar-07 25-Jul-06 15-Oct-06 4-May-06 11-Feb-06 1-Sep-05 22-Nov-05 11-Jun-05 21-Mar-05 9-Oct-04 30-Dec-04 19-Jul-04 6-Feb-04 28-Apr-04 17-Nov-03 6-Jun-03 27-Aug-03 16-Mar-03 750 Date Daily Energy Constancy Daily IMRT Procedure HI*ART II S/N 002, ENERGY CHECK 0.4350 0.4300 +3 % +2 % 0.4250 0.4200 1) Scan 0.4150 0.4100 DATE 14-Nov-07 28-Apr-07 10-Oct-06 24-Mar-06 5-Sep-05 17-Feb-05 1-Aug-04 14-Jan-04 28-Jun-03 0.4000 1J 0.4050 08 -2 % -3 % 10-Dec-02 DEPTH DOSE AT 15 cm Output (cGy) (60sec Readings) 820 - Typically within 2 % of nominal output - If outside 3 %, adjust PFN (output) at end of day - Observed long term drifts - About 2 weeks after magnetron replacement - End of target life 2 ) Register 4) Treat -compare to known offsets - two point dose measurements 3) Align -test couch and laser movements DQA Setup Patient Plan DQA - Deliver plan into phantom - Measure dose point with Ion chamber and dose distribution with EDR film Ion chamber high dose, low gradient region DQA Setup DQA – Point Dose Measurement TomoTherapy Point Dose Verification Temperature 24.0 Pressure 759.0 CTP 1.008 Electrometer Calculate dose distribution in the phantom C mm Hg Inovision 35040 Chamber Phantom 0 Exradin A1SL, SN 30562 Solid H20 ("Cheese" Phantom) System Calibration Factor (C21) 0.6660 [Gy/nC] Ion Chamber Position In Phantom 1 BELOW Film Plane Electrometer Reading for Phantom Irra 2.081 Doseplan 1.381 Dosemeas 1.397 [nC] [Gy] Ratio (Dm eas /Dplan) 1.012 Pass 1.1780 % Difference [Gy] DQA Film Analysis DQA film analysis Expected isodose lines vs measurement Film DQA Film Analysis DQA – Point Dose Measurement Film and dose profiles 5% 3% 06-07 DQA Results 10 8 6 % Difference 4 2 0 -2 -4 -6 -8 -10 11/14/2007 8/6/2007 4/28/2007 1/18/2007 10/10/2006 7/2/2006 3/24/2006 12/14/2005 Date DQA with Array devices Available array devices DQA with Array devices Advantages: Sun Nuclear – MapCHECK Time savings IBA Dosimetry – Matrixx No film processing issues PTW – 2D-ARRAY seven29 Real time comparison Disadvantages: Resolution Tomo DQA with MapCHECK Tomo DQA with MapCHECK Head and Neck - Initial tests have shown good results for TomoTherapy DQA -Gamma test results - Absolute dose (3%, 3mm) - HN: 98.5% passing - Prostate: 99.6% passing Tomo DQA with MapCHECK Tomo DQA with MapCHECK Prostate Import into TomoTherapy Monthly QA Monthly QA -Output measurement with farmer chamber -Output measurement with farmer chamber -Laser alignments, x-z localization procedure -Laser alignments, x-z localization procedure - Cone profile, lateral profile (TomoDose - diode array) - Cone profile, lateral profile (TomoDose - diode array) - Beam profile, longitudinal profile - Beam profile, longitudinal profile - MVCT dose and HU - MVCT dose and HU -Table position accuracy (ruler vs. readout) -Table position accuracy (ruler vs. readout) Fan Beam TomoDOSE lateral Check profile monthly longitudinal 400 mm lateral 10, 25, or 50 mm longitudinal TomoDOSE lateral beam profile Accuracy of TomoDOSE 1.2 1 Relative Reading 0.8 0.6 0.4 0.2 0 -30 -20 -10 0 Off axis distance (mm) 10 20 30 Accuracy of TomoDOSE Monthly QA: Cone Profile Monthly QA: Cone Profile Monthly QA -Output measurement with farmer chamber -Laser alignments, x-z localization procedure - Cone profile, lateral profile (TomoDose - diode array) - Beam profile, longitudinal profile - MVCT dose and HU -Table position accuracy (ruler vs. readout) Longitudinal Profile Monthly QA -Output measurement with farmer chamber - Static Gantry at 0° - Continuous table movement - Measure current -Laser alignments, x-z localization procedure - Cone profile, lateral profile (TomoDose - diode array) - Beam profile, longitudinal profile FWHM - MVCT dose and HU -Table position accuracy (ruler vs. readout) Monthly MVCT QA Test Consistency 1.824 0.48 1.56 0.33 HU Noise Uniformity Spatial resolution Dose - takes 1 MVCT scan Weekly QA (by TomoTherapy Service) - Output - Cone beam profile - Check error Log files - Check TCS - Check Data Server - Check various mechanical parameters Weekly QA (by TomoTherapy Service) Rotational Variation (by TomoTherapy Service) 6 MV Accelerator - Output - Cone beam profile Rotational Variation - Check error Log files - Check TCS - Check Data Server - Check various mechanical parameters Primary Collimator 85cm Binary MLC 40 FOV CT cm - Treatment beam - All MLC leafs open - Rotate gantry - Collect detector data - Compare detector data to Gold Standard Approximately 50cm Rotational Variation Annual QA -Beam data (PDD, long. and lateral profiles) -Mechanical alignment tests (Jaw, MLC alignment) Follow TomoTherapy’s recommended test procedures Phantom end-to-end test Annual QA, Beam Data -Collect beam data for all (three) commissioned beam widths (1.0, 2.5, 5.0 cm) -Cheese Phantom based end-to-end tests - RPC TLD RPC TLD Measurements (Static Field) 20 s beam-on time with 5x40 field gives ~270 cGy to TLDs 1 cm 2.5 cm 5 cm Central Target + 0.5 % + 2.8 % + 2.1 % Peripheral Target + 1.3 % + 1.0 % + 0.8 % Annual QA Alignment Tests -Source should be aligned with center of jaw Measure output at different jaw positions Max output indicates jaw center Annual QA Alignment Tests Annual QA: Y-Divergence/Jaw Twist -Test Y-divergence and jaw twist Film below iso, irradiate from 0º and 180º Aligned Jaw Iso Film Twisted Jaw Divergence center off from rotation plane Annual QA Alignment Tests Time Requirement -Source should be aligned with MLC - Daily: Including warm up time: 1 hour Run odd leaf only and even leaf only profile Compare to all leaf open profile -> T-G profile Look for symmetry in TG-Profile - DQA: 20-30 minutes per case - Monthly: 2-3 hours - Annual: 1-2 days QA Following Major Component Replacement QA Following Major Component Replacement Magnetron Typically every 3-4 months MLC Typically lasts about 18 Months Tomo service engineer -Adjusts beam profile and output using a rotational variation procedure Tomo service engineer -Checks alignment with T+G test -Measures leaf latency update leaf latency file for future plans Physicist: (2 h) -Check output and energy -Check beam profile using TomoDose array -Run 1-2 DQA procedures (IC+Film) QA Following Major Component Replacement Target Typically lasts 8-10 Months Tomo service engineer -Checks Linac alignment -Adjusts beam profile and output using a rotational variation procedure Physicist: (1-2 days) - Beam profiles (TomoDose) Do annual QA Physicist: (3-4 h) -Check multiple current patient DQAs. 6-8 DQA Target Wear Warning Signs Beam profile change: • Service engineers checks beam profiles weekly • Physics QA includes monthly TomoDOSE test Target Wear Warning Signs Variations in lateral beam profile TomoDOSE Energy Variation Component replacement Target : HI*ART II S/N 002, ENERGY CHECK Target changes 0.4350 • Can typically be predicted +3 % +2 % 0.4250 0.4200 • Scheduled over weekend Magnetron : 0.4150 • Fails suddenly 0.4100 14-Nov-07 28-Apr-07 10-Oct-06 24-Mar-06 5-Sep-05 17-Feb-05 1-Aug-04 14-Jan-04 28-Jun-03 0.4000 1J 0.4050 08 -2 % -3 % 10-Dec-02 DEPTH DOSE AT 15 cm 0.4300 DATE On-board detector • 1 day (or one night job) TomoTherapy TQA • TQA software being released by TomoTherapy • Uses detector data to measure beam properties and delivery systems • Conclusion Routine Machine QA: • Comparable to other machines in terms of time • Output • Beam Profile • Energy Major component replacement: • Lasers • More frequent than ‘regular’ Linac • Couch synchrony • MLC Allows access to detector data for QA for development of personalized QA tests Thank You • Some replacement can be predicted • TQA may help with prediction of these failures