Slide 1 Slide 2 Introduction Digital Mammography Detector QC z What is FFDM QC and why is it Eric A. Berns, Ph.D. z What to know before you start important? z Overview and compare QC tests eric.berns@gmail.com University of Colorado at Denver and HSC Denver, CO Slide 3 z Key take home points Slide 4 Introduction Introduction 5,911 SFM Units Certified Statistics – Past Year July 1, 2007 July 1, 2008 Difference Total Certified Mammo Facilities 8,837 8,714 -123 Total Accredited Mammo Units 13,450 12,875 -575 Certified Facilities with FFDM units 3,366 4,659 +1,293 Accredited FFDM units 5,119 6,964 +1,845 Slide 5 z MQSA – Mammography Quality Standards Act z ACR – American College of Radiology Slide 6 ScreenScreen-Film FFDM z z In FFDM, the manufacturer designs and mandates their own QC program and action limits In FFDM, you must the manufacturers’ manufacturers’ QC program 1 Slide 7 Slide 8 z AGFA IMPAX MA3000 – – – – – – – z Barco Clean monitor surface Measure display white & black Measure quality level Measure uniformity Calibrate displays View SMPTE pattern Viewing conditions z – – – – I-Guard Check Quality Level Display White Calibration Settings Check z GE Seno Advantage – – – – z Geometric distortion Reflection Luminance Response Luminance Uniformity Resolution Noise Overall Evaluation Slide 9 z Planar – – – – – Monitor Cleaning Viewing Conditions Check Monitor Calibration Check Image Quality – SMPTE Visual Inspection for Display Defects Slide 10 Quality Control: Why? Reduce exposure to patients and personnel z Pattern Check Luminance Check Grayscale Check Uniformity Check Siemens MammoReport Plus – – – – – – – z Eizo – – – – Viewing conditions check and setting Monitor calibration Image quality – SMPTE pattern Analysis of screen uniformity Consistent image quality z Detect and correct for potential problems, before they impact image quality What is Quality Control? z Determination of what is “Normal” Normal” z Detection of what is “Abnormal” Abnormal” z Understanding of how to return to “Normal” Normal” from “Abnormal” Abnormal” z In particular, in FFDM, how do you know what you are seeing is what it is supposed to be? Slide 11 Slide 12 Before You Begin QC Before You Begin z Golden Rules for FFDM QC – Must use manufacturer’ manufacturer’s QC procedures • Mandate action limits – Manufacturers’ Manufacturers’ QC may refer to Monitor & Printer Manufacturers’ Manufacturers’ QC z Obtain proper training & – HandsHands-on training on actual unit: • Mechanics – Multimodality Workstations have own separate QC • Software – Printers may have their own QC • Artifacts – Most failures result in stopping clinical imaging until failure can be corrected CE credits (8 hours) • Learn vendor specific tests and tricks 2 Slide 13 Slide 14 Before You Begin z Must have proper continuing experience: Before You Begin z Note: – For new unit: unit: Must use most current – 2 facilities version – 6 units – For renewal unit: unit: Can use older – Within last 2 years version (version used when tested previously) Slide 15 Slide 16 Before You Begin z Before You Begin ACR Accreditation - www.acr.org – Physics forms • http://acr.org/accreditation/mammography/mammo_qc_forms.aspx – – – – – GE Senographe 2000D, DS, and Essential Fischer Senoscan Lorad Selenia Siemens Novation Fuji FCRm • Note: Note: if using unit for both screenscreen-film & CR, you must accredit for both z FDA Accreditation – www.fda.gov/cdrh/mammography/ www.fda.gov/cdrh/mammography/ – Other vendors when approved Slide 17 Slide 18 Before You Begin Before You Begin 3 Slide 19 Slide 20 Before You Begin Before You Begin Slide 21 Slide 22 Collimation z Limiting Spatial Resolution ScreenScreen-Film Mammo z ScreenScreen-Film – Assure XX-ray field aligns with the light field – LineLine-pair test tool – Collimator allows full coverage of receptor • Focal spot – Chest wall edge of compression paddle aligns with chest wall of receptor z Mammo z FFDM FFDM – LineLine-pair test tool – Same • Focal spot – Multiple target/filters and/or multiple paddle positions • Detector & imaging chain – Chest Wall Missed Tissue – Compression Plate Overlap on the Chest Wall Side Slide 23 Slide 24 Spatial Resolution Fuji FCRm GE 2000D, DS, Essential Fischer Senoscan AEC Performance Lorad Selenia Siemens Novation z ScreenScreen-Film Mammo – Measure optical densities • Different thicknesses using clinical modes Same as ScreenScreen-Film (Focal Spot Eval) Eval) Measure using LP phantom on top of acrylic MTF X • Different density settings ((-2, -1, 0, +1, +2, etc.) z FFDM X X X – Measure resultant techniques X X – Measure signal, exposure, & SNR values 4 Slide 25 Slide 26 AEC GE 2000D Fuji FCRm GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation z AOP Mode and SNR Check – Variable thicknesses of acrylic – Std, Auto – Evaluate: Same as ScreenScreen-Film Density Control Function X ACR Reproducibility X Image Mode Tracking X CNR X SNR • Correct techniques? • Adequate SNR? X X X X X Acrylic Thickness 2.5 cm TargetFilter Mo-Mo Selected kVp 27 kVp Selected mAs 20-60 X X X 4.0 cm Mo-Rh 28 kVp 35-90 6.0 cm Rh-Rh 32 kVp 35-90 4.0 cm Acrylic Each “raw” raw” image must have a measured SNR of at least 50 Slide 27 6.0 cm Acrylic Slide 28 GE DS & Essential z Lorad Selenia DS specific QC tests z – AOP Mode and SNR Check Automatic Exposure Control – AUTOAUTO-FILTER – 2, 4, 6, 8 cm • Use builtbuilt-in software for image acquisition – 4 cm ((-5 thru +5, or -3 to +4) • AOP: STD/Auto – Use ROI to measure mean pixel value inside virtual AEC detector Exposure Parameters For AOP STD mode oly Acrylic Thickness (mm) Track/Filter mAs kV 25 Mo/Mo 2020-60 26 50 Rh/Rh 4040-90 29 60 Rh/Rh 4545-95 31 z Slide 29 Performance Criteria – Pixel value should not vary by more than 10% of mean – Exposure compensation steps shall SNR must be > 50 meet requirements in pixel value table Slide 30 Siemens Novation z 2.5 cm Acrylic AEC Image Stability and Reproducibility and SNR – ACR Phantom Siemens Novation DR z AEC Thickness Tracking – 2, 4, and 6 cm – “H” mode – Mo/Mo, 28 kV, “H”, sensor 1 – Max deviation = (Max difference / mean value ) * 100 – Record mAs, mAs, SNR, Mean, Entrance Exposure (5 times) z Action Limits: – Coef. Coef. of Var for mAS and X < 5% – SNR and Mean shall not vary by + 15% of Mean 5 Slide 31 Slide 32 Fuji FCRm Fuji FCRm z z Density Control Function Reproducibility & Image Mode Tracking – 4 cm acrylic with clinical technique – Position ion chamber in beam – 4 cm acrylic, ACR phantom technique – Record mAs and X – Repeat 3 times – Repeat at -2, -1, 0, +1, +2 etc. density – Repeat in each mode (small, large, mag, mag, no grid) – Action Limits: Limits: – Record mAs • Coefficient of variation for Exposure or mAs must not exceed 0.05 – mAs change should be 5 to 15% per step • No significant difference in exposure between small and large bucky when using similar grids Slide 33 Slide 34 Uniformity of Screen Speed & Detector Performance Fuji FCRm z z CNR Per Object Thickness – Measure if OD’ OD’s are consistent – Check for artifacts – CNR using clinical technique for 2 cm z – Repeat for 4 and 6 cm FFDM – FlatFlat-field uniformity – Action Limits: Limits: – Detector calibration • CNR of 2 cm relative to 4 cm must be > 100% – Pixel correction test • CNR of 6 cm relative to 4 cm must be > 75% Slide 35 – CR Reader Scanner Performance Slide 36 Detector Performance GE Fuji FCRm 2000D, DS, Essential Fischer Senoscan Image Quality Lorad Selenia Siemens Novation Same as ScreenScreen-Film (Screen Uniformity) X Detector Calibration X X CR Reader Scanner Performance X Dynamic Range X Primary Erasure X InterInter-Plate Consistency X Geometric Distortion z ScreenScreen-Film Mammo – ACR Phantom Scores FlatFlat-Field Uniformity X X – Optical Density & Contrast z FFDM – ACR Phantom Scores • Pass/fail requirements differ by vendor – SignalSignal-toto-Noise Ratio (SNR) X Pixel Correction Ghosting ScreenScreen-Film Mammo X X – ContrastContrast-toto-Noise Ratio (CNR) X 6 Slide 37 Slide 38 Image Quality Fuji FCRm Image Quality GE 2000D, DS, Essential Fischer Senoscan X X X X Lorad Selenia ACR Phantom Imaging Siemens Novation GE & Fischer & Fuji Lorad & Siemens Fibers 4 5 Specks 3 4 Masses 3 4 Same as ScreenScreen-Film Manual Techniques Clinical Technique X CNR X X Partial X X Slide 39 Slide 40 GE 2000D ACR Phantom Imaging z GE 2000D z – Manual technique (Mo/Mo, 26 ContrastContrast-toto-Noise Test (CNR) – To examine consistency of CNR ratio measured over time kVp, 125 mAS) mAS) – Use the raw image – Score the processed image – + 20% of baseline – Acquisition workstation Background ROI – Each monitor of the RWS Mass ROI – Laser imager CNR = (Mean (Meanbackground - Meanmass)/SDbackground Slide 41 Slide 42 GE DS z DS specific QC tests – Phantom Image Quality • Manual technique: Rh/Rh, Rh/Rh, 29 kVp, 56 mAs – MTF and CNR Measurement • Use IQST test tool • Use builtbuilt-in software for image acquisition • Manual technique: Rh/Rh 30 kVp, 56 mAs Lorad Selenia z Phantom Image Quality – Select clinical exposure mode (i.e. AUTOAUTO-FILTER) – Print film – Measure background OD and density difference – Plot on tech worksheets • Background must be > 1.20 OD + 0.20 • DD must be > 0.40 + 0.05 – Score on each Soft Copy Workstation • 5 fibers • Results are automatically displayed (pass/fail) • 4 speck groups • Same action limits as 2000D • 4 masses 7 Slide 43 Slide 44 Siemens Novation DR z Lorad & Siemens Phantom Image Quality – Position phantom 1 cm over chest wall edge z SNR and CNR Measurements – Select: 28 kV, AECAEC-Auto, Mo/Mo – SNR at least > 40 – Score on RWS or Film – CNR should stay within ±15% of baseline – Fiber > 5 • Obtained during acceptance testing – Specks > 4 – Masses > 4 Slide 45 Slide 46 Fuji FCRm z ContrastContrast-toto-Noise Test (CNR) Dose z ScreenScreen-Film – To examine consistency of CNR ratio measured over time Mammo – Dose for single CC view of ACR phantom shall – Use 4 cm acrylic & 0.2 mm Al not exceed 3.0 mGy per exposure per FDA – Manual technique (Mo/Mo, 26 kVp, 125 mAs) mAs) – Calculate CNR using software z FFDM – + 20% of baseline Slide 47 Slide 48 Film Processing z – Same Film Processor QC ScreenScreen-Film Mammo Fuji FCRm – Measure optical densities • Density difference, midmid-density, base+fog • Measures consistency Follow Printer Manufacturers QC If Not, Use Theirs Follow FFDMs’ FFDMs’ QC z GE 2000D, DS, Essential Fischer Senoscan X X X X X X Lorad Selenia Siemens Novation X X FFDM – Manufacturers’ Manufacturers’ recommendations – Some refer to printer manufacturers’ manufacturers’ recommendations – Typically identical to ACR SFM Manual 8 Slide 49 Slide 50 Film Processing Film Processing z MidMid-Density z Density Difference Wet Processing Laser Printer Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Mid-Density Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Mid-Density Wet Processing 2.30 2.20 Laser Printer Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Density Difference Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Density Difference 2.60 1.70 2.50 1.60 2.40 1.50 2.30 1.40 2.20 HD Step=13 2.00 LD Step=10 MD Step=12 1.90 2.10 2.00 1.90 1.30 1.80 1.20 1.70 1.10 1.80 1.70 1.60 2.10 1.50 2.00 1.40 Mid-Density Upper Control Limit = 1.48 1.90 Mid-Density Daily Value 1.00 Mid-Density Upper Control Limit = 2.06 Mid-Density Daily Value Operating Level = 1.33 Mid-Density Daily Value Density Difference Upper Control Limit = 1.77 1.30 Density Difference Upper Control Limit = 2.25 Mid-Density Lower Control Limit = 1.18 0.90 Mid-Density Daily Value Operating Level = 1.91 1.50 1.80 Density Difference Daily Value Density Difference Daily Value 1.20 Density Difference Daily Value Operating Level = 1.62 1.70 Mid-Density Lower Control Limit = 1.76 Density Difference Daily Value Operating Level = 2.1 0.80 1.60 1.40 1.50 1-M ar0 2-M 6 ar0 3-M 6 ar0 6-M 6 ar0 7-M 6 ar0 8-M 6 ar0 9-M 6 ar06 10 -M ar06 13 -M ar06 14 -M arse 06 rv ic e 3/1 15 4 -M ar06 16 -M ar06 17 -M ar06 20 -M ar06 21 -M ar06 22 -M ar06 23 -M ar06 24 -M ar06 27 -M ar06 28 -M ar06 29 -M ar06 30 -M ar06 31 -M ar06 r2 ,2 cto 00 be 6 r3 ,2 cto 00 be 6 r4 ,2 cto 00 be 6 r5 ,2 cto 00 be 6 r6 O ,2 cto 00 be 6 r9 O cto ,2 00 be 6 r1 O 0, cto 20 be 06 r1 O 1, cto 20 be 06 r1 O 2, cto 20 be 06 r1 O 3, cto 20 be 06 r1 O 6, cto 20 be 06 r1 O 7, cto 20 be 06 r1 O 8, cto 20 be 06 r1 O 9, cto 20 be 06 r2 O 0, cto 20 be 06 r2 O 3, cto 20 be 06 r2 O 4, cto 20 be 06 r2 O 5, cto 20 be 06 r2 O 6, cto 20 be 06 r2 O 7 cto ,2 00 be 6 r3 0, 20 06 O O O O O cto be 1-M a r0 2-M 6 a r0 3-M 6 a r0 6-M 6 a r0 7-M 6 a r0 8-M 6 a r0 9-M 6 a r06 10 -M ar06 13 -M ar06 14 -M arse 06 rv ic e 3/1 15 4 -M ar06 16 -M ar06 17 -M ar06 20 -M ar06 21 -M ar06 22 -M ar06 23 -M ar06 24 -M ar06 27 -M ar06 28 -M ar06 29 -M ar06 30 -M ar06 31 -M ar06 Density Difference Lower Control Limit = 1.47 1.00 0.70 1.30 1.10 Density Difference Lower Control Limit = 1.95 Date Date Date O cto be r2 O ,2 cto 00 be 6 r3 O ,2 cto 00 be 6 r4 O ,2 cto 00 be 6 r 5 O ,2 cto 00 be 6 r6 O ,2 cto 00 be 6 r O 9, cto 20 be 06 r 10 O cto ,2 00 be 6 r1 O 1, cto 20 be 06 r1 O 2, cto 20 be 06 r1 O 3, cto 20 be 06 r1 O 6, cto 20 be 06 r1 O 7, cto 20 be 06 r1 O 8, cto 20 be 06 r 19 O ,2 cto 00 be 6 r2 O 0, cto 20 be 06 r 23 O ,2 cto 00 be 6 r2 O 4, cto 20 be 06 r 25 O cto ,2 00 be 6 r 26 O cto ,2 00 be 6 r2 O 7, cto 20 be 06 r 30 ,2 00 6 1.60 Slide 51 Date Slide 52 Artifacts Film Processing z Dmax z ScreenScreen-Film Mammo Laser Printer Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Dmax – Evaluate cassettes 4.00 3.90 3.80 z FFDM 3.70 3.60 3.50 3.40 3.30 3.20 – Evaluate detector 3.10 3.00 Dmax Daily Value 2.90 Dmax - Operating Level = 3.44 2.80 Dmax Action Limit = 3.19 2.70 2.60 O cto O cto be r2 ,2 00 be 6 r3 O ,2 cto 00 be 6 r4 O ,2 cto 00 be 6 r5 O ,2 cto 00 be 6 r6 O ,2 cto 00 be 6 r9 O cto ,2 00 be 6 r1 O 0, cto 20 be 06 r1 O 1, cto 20 be 06 r1 O 2, cto 20 be 06 r1 O 3, cto 20 be 06 r1 O 6, cto 20 be 06 r1 O 7, cto 2 0 be 06 r1 O 8, cto 20 be 06 r1 O 9, cto 20 be 06 r2 O 0, cto 2 00 be 6 r2 O 3, cto 20 be 06 r 24 O cto ,2 00 be 6 r2 O 5, cto 2 00 be 6 r2 O 6, cto 20 be 06 r2 O 7, cto 20 be 06 r3 O 0, cto 20 be 06 r3 1, 20 06 – CR: Imaging Plates Date Slide 53 Artifacts Artifacts Fuji FCRm GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation X X X X X Same as ScreenScreen-Film Expose Using Acrylic 9 Artifacts Artifacts ¾ Description: Description: ¾ Description: Description: Hologic Good Flat GE Good FlatFlat-field Field Image Artifacts Artifacts ¾ Description: Description: Good ¾ Description: Description: Acceptable image ACR Phantom image ¾ Raw image ¾ WW = 200 Artifacts ¾ Artifacts Artifact evaluation – Contact Mode GE 2000D ¾ Artifact evaluation - windowing Window width = 30 Mo/Mo Mo/Rh Mo/Rh Window width = 200 Window width = 1000 Rh/Rh 10 Artifacts Artifacts ¾ Description: Description: NonNon- ¾ Description: Description: Low uniform background ¾ Possible Cause: Cause: Calibration file ¾ Solution: Solution: Recalibrate contrast overall ¾ Possible Cause: Cause: Window width too wide – (~750) ¾ Solution: Re-window Solution: Re- Artifacts Artifacts ¾ Description: Description: Poor ¾ Description: Description: Vertical contrast, fibers and masses failing ¾ Possible Cause: Cause: Ultra low dose – Mo/Mo, 29 kVp, 25 mAs, mAs, 0.58 mGy ¾ Solution: Solution: Increase techniques Artifacts and horizontal lines ¾ Possible Cause: Cause: Gridlines, grid artifacts in in cal file ¾ Solution: Solution: Check grid mechanism, recalibrate Artifacts ¾ Description: Description: White vertical bands ¾ Possible Cause: Cause: Calibration file ¾ Solution: Solution: Recalibrate detector 11 Artifacts Artifacts ¾ Description: Description: Magnification image, ¾ Description: Description: White several small flecks, pixels grayscale gradation ¾ Possible Cause: Cause: ¾ Possible Cause: Cause: Detector going bad Object calibrated into Cal file, grayscale ¾ Solution: Solution: New gradation ~ normal Detector ¾ Solution: Solution: Clean detector and tube head, recalibrate detector Artifacts ¾ 2x Artifacts ¾ 4x Artifacts ¾ 6x Artifacts ¾ 8x 12 Artifacts Artifacts ¾ Description: Description: ¾ Description: Description: Image Gridlines, black flecks ¾ Possible Cause: Cause: Grid failure, calibration file ¾ Solution: Solution: Grid repair, clean the imaging chain, and recalibrate Artifacts processing around wax insert ¾ Possible Cause: Cause: Image processing algorithm ¾ Solution: Solution: None Artifacts ¾ Description: Description: Washed out Image ¾ Cause: Cause: Phantom not at chest wall edge ¾ Solution: Re-position the phantom Solution: Re- ¾ Description: Description: Linear and rectangular banding ¾ Possible Cause: Cause: Calibration file ¾ Solution: ReSolution: Recalibrate Artifacts ¾ Description: Description: Shadows just outside of breast skin line ¾ Possible Cause: Cause: Saturated detector due to overexposure from dense breast ¾ Solution: Solution: Siemens recommends to increase kVp to reduce exposure time Artifacts ¾ Description: Description: CR Noisy Background, dark speck ¾ Possible Cause: Cause: Background somewhat normal ¾ Solution: Solution: Check screen and system to identify black speck 13 Artifacts ¾ Description: Description: Horizontal, vertical, and curved lines multiple density differences – printed from “flatflat-field” field” menu on Lorad Selenia ¾ Possible Cause: Cause: Dirty spinner assembly in optics casting shadows ¾ Solution: Solution: Clean spinner assembly Printer Artifacts Readout Line Artifact Collimator needs adjustment at chest wall edge Artifacts ¾ Description: Description: Artifacts Processing Steps for Digital Images Horizontal and curving lines Image Detection Image Correction Raw Image Processing Processed Image Display ¾ Possible Cause: Cause: Readout error, ghosting Calibration File ¾ Solution: Solution: Recalibrate, let detector sit idle, replace detector 14 Slide 86 Key Take Home Points Summary on Artifacts •Most artifacts due to calibration file z Obtain relevant handshands-on training •Window/level adjustments can appear as AEC and/or exposure problems z Must perform manufacturer specific QC tests •Printers can cause fine, linear streaking artifacts - rare •Look at technique factors, breast thickness, and breast density for clues •Objects on bucky, mag stand, and up in tube head often make their way into the image – dust on accessories z Artifacts – most problems can be seen here z ReRe-booting and/or rere-calibrating fixes most problems z Laser Printer – Dmax at least 3.5 OD – MidMid-density about 1.5 OD •Detectors can, and do, fail Slide 87 Slide 88 SAMs Questions 20% 20% 20% 20% 20% In digital mammography, who mandates the pass/fail criteria for site QC? 1. The American College of Radiology 2. The FDA 3. The FFDM unit manufacturer 4. NEMA 5. MQSA 10 Slide 89 Slide 90 Answer: 3 - The FFDM unit manufacturer Answer 3: The FFDM unit manufacturer References Explanation z “the quality assurance program shall be substantially the same as the quality assurance program recommended by the image receptor manufacturer, except that the minimum allowable dose for screenscreenfilm systems in this section” section” z MQSA Regulations 900.12(e)(6) http://www.fda.gov/CDRH/MAMMOGRA PHY/frmamcom2.html#s90012 15 Slide 91 Slide 92 How do you accredit a Fuji CR Mammography system which uses both CR and screenscreen-film on the same x-ray system? 25% 1. As 1 mammography unit? 25% 2. As 2 mammography units? 3. You cannot use CR and film on the same unit. CR is exempt from accreditation 25% 25% 4. Answer: 2 – As 2 mammography units References z http://www.acr.org/accreditation/mammography/ mammo_faq/mammo_faq_mamac.aspx#8.0.1 10 Slide 93 Slide 94 What are the minimum passing ACR phantom scores for the Siemens Novation DR for weekly QC? Answer: 2 - As 2 mammography units Explanation As of November 15, 2006 facilities using both screenscreen-film and CR on the same mammography units must accredit these 2 systems as 2 separate units. 5 Fibers, 4 Speck Groups, 3 Masses 4 Fibers, 3 Speck Groups, 3 Masses 3. 5 Fibers, 4 Speck Groups, 4 Masses 4. 4 Fibers, 4 Speck Groups, 3 Masses 5. 4 Fibers, 4 Speck Groups, 4 Masses 20% 1. 20% 2. 20% 20% 20% 10 Slide 95 Slide 96 Answer 3: 5 Fibers, 4 Speck Groups, 4 Masses Answer 3: 5 Fibers, 4 Speck Groups, 4 Masses References Explanation z The Siemens Novation DR QC manual states that the ACR Phantom minimum passing scores are as follows: Siemens. Siemens Quality Control Manual Version 05, Erlangan, Erlangan, Germany. 2007 z z z 5 Fibers 4 Speck Groups 4 Masses 16 Slide 97 Slide 98 Answer 4: 3.00 mGy/exposure mGy/exposure For FFDM, the exposure for a single CC view of the ACR Phantom shall not exceed? References z 0.75 mGy/exposure mGy/exposure 1.25 mGy/exposure mGy/exposure 20% 3. 2.00 mGy/exposure mGy/exposure 20% 4. 3.00 mGy/exposure mGy/exposure 20% 5. 4.00 mGy/exposure mGy/exposure 20% 1. 20% 2. 900.12(e)(5)(vi): Dosimetry. Dosimetry. The average glandular dose delivered during a single craniocaudal view of an FDAFDAaccepted phantom simulating a standard breast shall not exceed 3.0 milligray (mGy) mGy) (0.3 rad) rad) per exposure. The dose shall be determined with technique factors and conditions used clinically for a standard breast. 10 Slide 99 Slide 100 Answer 4: 3.00 mGy/exposure mGy/exposure To meet the FDA requirement for continuing experience, how many mammography facilities and mammography unit surveys must be performed within the previous 24 months? Explanation The FDA requires: z 3.00 mGy/exposure mGy/exposure 20% 1. 20% 2. 20% 3. 20% 4. 20% 5. 6 Facilities and 2 mammography units 4 Facilities and 12 mammography units 2 Facilities and 6 mammography units 1 Facilities and 6 mammography units 2 Facilities and 12 mammography units 10 Slide 101 Slide 102 Answer 3: 2 Facilities and 6 Mammography Units References z z http://www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/ START.HTM 900.12(a)(3)(iii)(B): Continuing experience. Following the second anniversary date of the end of the calendar quarter in which the requirements of paragraphs (a)(3)(i) and (a)(3)(ii) of this section were completed or of April 28, 1999, whichever is later, the medical physicist shall have surveyed at least two mammography facilities and a total of at least six mammography units during the 24 months immediately preceding the date of the facility’ facility’s annual MQSA inspection or the last day of the calendar quarter preceding the inspection or any date in between the two. The facility shall choose one of these dates to determine the 2424-month period. No more than one survey of a specific facility within a 1010-month period or a specific unit within a period of 60 days can be counted towards this requirement. Answer 3: 2 Facilities and 6 Mammography Units Explanation The FDA requires: z z z 2 Facilities 6 Mammography Units Within past 24 months 17 Slide 103 Thank You 18