MARP FFDM in the Field: Physicist's Role in the

advertisement
MARP, Inc.
FFDM in the Field: Physicist's Role in the
QC of Third-Party Review Work Stations
July 2009
Carl R. Keener, Ph.D., DABMP, DABR
keener@MARPinc.com
MARP
Medical & Radiation Physics, Inc.
Physicist's Role in the QC of Third-Party Review
Work Stations
MARP, Inc.
‹
consulting physics group with variety of
clients
‹ FFDM:
GE, Fischer, Hologic, Siemens
‹ RWS:
GE, Hologic, Siemens
Barco, Planar
Sectra, Carestream, Agfa, McKesson
‹ how
do we keep all of them in compliance?
‹ review of QC manuals, TG18
‹ research of RWS vendors at RSNA
‹ incorporation of RWS requirements into
database
FFDM review workstations
MARP, Inc.
‹
‹
‹
‹
‹
‹
‹
MQSA /FDA guidance
TG18 guidance
FFDM manufacturer guidance
monitor manufacturer guidance
PACS vendor guidance
medical physicist survey & documentation
summary
caveat
MARP, Inc.
This lecture involves MQSA.
Use with care.
MQSA Guidance
MARP, Inc.
‹
21 CFR 900.12(e)(6)
Quality control tests—other modalities. For
systems with image receptor modalities other
than screen-film, the quality assurance
program shall be substantially the same as
the quality assurance program recommended
by the image receptor manufacturer, except
that the maximum allowable dose shall not
exceed the maximum allowable dose for
screen-film systems in paragraph (e)(5)(vi) of
this section.
FDA Guidance
MARP, Inc.
‹
allowed digital mammography in 2000
‹ FFDM
manufacturer supplied QC manual to FDA
‹ initially
allowed final interpretation from laser film
only
‹ later
allowed soft copy mammography
interpretation
‹
requires MP to follow QC program in FFDM
manufacturer’s QC manual
‹ varies
with manual version
• & alternative standards
FFDM vendors
MARP, Inc.
‹
‹
GE
‹
Senographe 2000D
‹
Senographe DS
‹
Essential
Fischer
‹
‹
Hologic
‹
‹
Senoscan
Selenia
Siemens
‹
Novation
‹
Fuji
‹
FCRm
third party vendors
MARP, Inc.
‹
used for all studies
‹
‹
not just mammography
PACS (with manuals)
‹
Carestream
• Totoku +
‹
Agfa
• Barco
‹
5 MP monitors
‹
NDS (Planar)
‹
EIZO
‹
WIDE
‹
Totoku
Fimi
Sectra
• Barco, EIZO, Planar, Siemens
Barco
‹
‹
‹
‹
Richard Electronics
• Image Systems (WIDE)
‹
Quest
• Alphaview (Totoku)
‹
PACS (no manuals)
‹
Amicas
• Barco, Planar
‹
McKesson
• Barco
MARP, Inc.
scenario 1
MARP, Inc.
‹
soft copy workstation provided by FFDM
manufacturer
scenario 2
MARP, Inc.
‹
units at different sites
share soft copy
workstation by FFDM
manufacturer
scenario 3
MARP, Inc.
‹
FFDM units of different manufacturers share
a single soft copy workstation
scenario 4
MARP, Inc.
‹
single FFDM unit uses multiple soft copy
workstations by various manufacturers
FFDM
RWS x 5
How do FFDM manufactures address 3rd party
monitors?
MARP, Inc.
‹
some defer to monitor manufacturer:
‹ Hologic
• “If any workstation other than the (Hologic) SecurViewDX is
used, the recommended QC of that workstation should be
used”
• weekly phantom must be scored on diagnostic workstation
‹ Fuji
• “Follow monitor manufacturer’s QC program for test
frequency, procedure, performance and corrective action”
• For manufacturer’s w/out QC program, set up program
using
• TG18 QC or SMPTE RP133
• luminance response using gray scale images and
photometer
• weekly phantom scored on soft copy workstation
How do FFDM manufactures address 3rd party
monitors?
MARP, Inc.
‹
some do not address it:
‹ GE
• 2000D requires weekly phantom IQ on review workstation
‹ Siemens
• phantom evaluated on RWS if it fails on AWS
‹ Fischer
• requires weekly IQ on RWS
How does FDA address 3rd party RWS?
MARP, Inc.
‹
Printer and/or Monitors without QC Manuals
‹
‹
‹
‹
Under the Mammography Quality Standards Act (MQSA) final
regulation 900.12(e)(6), facilities using a mammographic modality other
than screen-film must follow a quality control program that is
“substantially the same as the quality assurance program
recommended by the image receptor manufacturer…”.
While all full field digital mammography (FFDM) manufacturers have
quality control (QC) manuals, in some cases, the QC manual instructs
the facility to test monitors and printers according to the component’s
QC manual.
In these cases, it is the responsibility of each facility to ensure that it
obtains and follows the component’s QC manual for its monitors and
printers.
FDA Inspection News Update February 4, 2008
how does FDA address other RWS?
MARP, Inc.
‹
Facilities Using the Same Printer or Monitor with FFDM Units
from Different Manufacturers
‹
For facilities that are using FFDM units from different
manufacturers, each with its own QC requirements for printers
and monitors:
‹
Each FFDM manufacturer QC manual requires that the same or
equivalent test be done, but at different time frequencies
• facilities need to perform the test at the more stringent frequency.
‹
Each FFDM manufacturer QC manual requires that different but
equivalent tests be done.
• facilities may perform only one of the tests at the more stringent frequency.
• medical physicist should provide a written statement for the facility’s quality
control records, indicating that in his or her opinion, the two tests are equivalent.
‹
Each FFDM manufacturer QC manual requires that different tests
(not equivalent) be done.
• facilities need to perform each test at the frequency required in the respective
FFDM manufacturer QC manual.
‹
FDA Inspection News Update February 4, 2008
How does FDA address laser printers?
MARP, Inc.
‹
To avoid unnecessary follow-ups by MQSA inspectors, we
urge all medical physicists to review the FFDM
manufacturer’s QC manual and, where necessary, the laser
printer QC or operator’s manual to determine the
appropriate testing.
‹
They then need to clearly document the testing of the laser
printer in their reports.
‹
FDA Inspection News Update February 4, 2008
TG18 tests
MARP, Inc.
‹
‹
‹
‹
monitor placement
ambient light
maximum & minimum
luminance
gray scale calibration
‹
‹
‹
‹
‹
geometric distortions
reflections
luminance response
quantitative evaluations
visual evaluations
TG18 test patterns
MARP, Inc.
TG18-QC
MARP, Inc.
‹
visual & quantitative
‹
resolution
‹
contrast
‹
luminance
‹
distortion
‹
artifacts
TG18-CX
MARP, Inc.
‹
visual
‹
resolution
TG18-LN
MARP, Inc.
‹
quantitative
‹
luminance
‹
DICOM grayscale calibration
TG18-UN & UNL
MARP, Inc.
‹
quantitative and visual
‹
luminance
‹
color uniformity
200 × ( Lmax − Lmin )
≤ 30%
( Lmax + Lmin )
• TG18-UN80
• bright
• TG18-UNL80
• bright, lined
• TG18-UN10
• dark
• TG18-UNL10
• dark, lined
TG18-CT
MARP, Inc.
‹
visual
‹
luminance response
TG18-AFC
MARP, Inc.
‹
visual
‹
noise
TG18-MP
MARP, Inc.
‹
visual
‹
luminance response
• bit-depth resolution
TG18-MM
MARP, Inc.
‹
anatomical
‹ reference
anatomical mammogram pattern
FFDM manufacturer guidance
MARP, Inc.
‹ GE
‹ Siemens
‹ Fischer
‹ Fuji
‹ Hologic
FFDM manufacturer – GE
MARP, Inc.
‹
Senographe 2000D
‹
RWS tasks in FFDM
manual
• 2277390-100 Rev 1, 2000
• 2371472-100 Rev 0, 2003 (postOnyx)
‹ CRT
workstation
• Siemens
‹ photometer
• Dome
• SMfit ACT
‹ SMfit
‹ no
ACT software
TG18 patterns
‹
technologist QC
‹
monitor cleaning
• daily
‹
RWS viewing conditions
• daily
‹
monitor calibration check
• SMPTE pattern
• monthly
• fix immediately if fails
FFDM manufacturer – GE
MARP, Inc.
‹
physicist tests
• 2277390-100 Rev 1, 2000
• 2371472-100 Rev 0, 2003
‹
monitor calibration check
‹
Lmax & Lmin
• L0 < 1.0 cd/m²
• L255 = 240 ± 10 cd/m²
viewing conditions
• ambient < 50 lux & stable ± 5 lux
• room drawing
• fix immediately if fails
‹
‹
‹
gray scale vs baselines
•
•
•
•
•
•
SMPTE pattern
• smallest vertical bars do not need
to be resolved
• fix immediately if fails
L0
L10
L60
L120
L180
L255
baseline ± 0.5 cd/m²
baseline ± 0.5 cd/m²
baseline ± 1.0 cd/m²
baseline ± 2.0 cd/m²
baseline ± 7 cd/m²
baseline ± 10 cd/m²
• background = L180
• fix within 30 days
‹
uniformity
• visual check at L255
FFDM manufacturer – GE
MARP, Inc.
‹
SenoAdvantage Workstation
‹
separate RWS QC manual
‹
physicist tests
‹
• ambient < 20 lux & stable ± 5 lux
• room drawing
• 2391082-4-1EN Rev 1 2007
‹
usually LCD monitors
• Siemens monitors
• SMfit ACT software
• Barco monitors
• MediCal X software
• TG18 test patterns
‹
‹
SMPTE pattern
• smallest vertical bars do not need
to be resolved
‹
monitor calibration check
•
•
•
•
•
•
•
technologist QC
• monitor cleaning
• daily
• RWS viewing conditions
• daily
• monitor calibration check
• SMPTE pattern
• monthly
• fix immediately if fails
viewing conditions
‹
L10
baseline ± 0.5 cd/m²
L60
baseline ± 1.0 cd/m²
L120
baseline ± 2.0 cd/m²
L180
baseline ± 7 cd/m²
L255
baseline ± 10 cd/m²
background = L20%
fix within 30 days
uniformity
• visual check at L255
FFDM manufacturer – Hologic
MARP, Inc.
‹
‹
SecureView Dx
‹
Barco Coronis monitor
‹
MediCal Pro software
Technologist QC
‹
weekly
‹
open MediCal Pro
software & run tests
which are due
•
•
•
•
white level
quality level performance
other tests for CRTs only
preprogrammed limits
‹
Physicist tests
‹
viewing conditions
• <50 lux
• (<20 recommended)
‹
MediCal Pro software
‹
white level
• 500 cd/m² ± 6%
• 600 cd/m² ± 6%
‹
quality level
performance
• < 10%
‹
CRTs only
• black level
• uniformity < 15%
FFDM manufacturer – Siemens
MARP, Inc.
‹
‹
syngo MammoReport
VB22
technologist QC
‹
‹
‹
•
•
•
•
•
•
•
daily monitor cleaning
physicist tests
‹
monitor location
‹
specular reflections
‹
TG18-QC
‹
ambient light
TG18-LN12-01
TG18-LN12-18
Lmin < 1.0 cd/m²
Lmax > 250 cd/m²
Lmax/Lmin > 250
ΔLmax < 10%
physicist photometer
uniformity
• TG18-UNL10
• TG18-UNL80
• including CX scores
‹
luminance settings
‹
( Lmax − Lmin )
≤ 30%
Lcenter
resolution
• TG18-QC (Cx scores)
• < 10 cd/m² (recommended)
‹
noise
• TG18-AFC
monitor manufacturer guidance
MARP, Inc.
‹ Barco
‹ WIDE
‹ NDS
‹ Totoku
‹ EIZO
(Planar)
‹ Fimi
available tests
MARP, Inc.
‹ room
layout
• ambient light
‹ gray
scale luminance
‹ TG18QC
or SMPTE
‹ uniformity
‹ luminance
response
‹ phantom
‹ technologist
QC tests
Barco
MARP, Inc.
‹
‹
Coronis 5MP monitor
QC “manual”
‹
‹
Version 7.0 29 August 2008
software:
‹
MediCal Pro
‹
MediCalX (more recent)
‹
most TG18 patterns
‹
built-in photometer at front
Barco
MARP, Inc.
‹
technologist QC:
‹
display
• weekly
• if MediCal QA Web is
available
‹
compliance
• monthly
• weekly if MediCal QA Web
not available
• compares calibration to
calibration curve
‹
autocalibration
• annually
• calibrates gray scale
‹
visual
• TG18 QC
• annually
‹
physicist tests
‹
display cleanliness
‹
viewing conditions
• < 20 lux
‹
TG18QC
•
•
•
•
ramps
gray-scale patches
5% & 95% squares
resolution
NDS (Planar)
MARP, Inc.
‹
‹
DOME C5i & DOME E5 monitors
QC manual
‹
‹
MQSA QC Manual 020-0517-00 Rev:B
software:
‹
Cxtra
• RightLight Conformance
‹
external photometer
‹
most TG18 patterns
NDS (Planar)
MARP, Inc.
‹
technologist QC:
‹
monitor cleaning
‹
physicist tests
‹
• soft cloth or “mild
cleanser”
‹
viewing conditions
• drawing of room layout
• < 20 lux
‹
• daily
‹
‹
monitor calibration
check
• SMPTE pattern
• monthly
monitor calibration
• “privileged user”
• displays LUM, white level,
graph
Cxtra operation
verification
• daily
• observe Windows taskbar
for Cxtra icon
viewing conditions
‹
SMPTE pattern
‹
visual inspection for
defects
EIZO
MARP, Inc.
‹
‹
RadiForce GS 520 monitor
QC “manual”
‹
‹
RadiForce QC management Revision 1.0 2007
software:
‹
RadiForce
• guides user & writes report
‹
external sensor
‹
most TG18 patterns
EIZO
MARP, Inc.
‹
user QC
‹
pattern check & constancy
‹
physicist tests
‹
• daily
‹
•
•
•
•
technologist QC:
‹
constancy
• monthly
‹
‹
‹
luminance
• monthly
Lmax/Lmin > 250
Lmax > 170 cd/m²
ΔLmax < 10%
Lamb < (Lmin-Lamb)/4
gray scale
• target error rate < 10% GSDF
‹
gray scale
pattern check
•
•
•
•
•
•
• monthly
‹
luminance check
uniformity
• monthly
‹
TG18-QC
TG18-CX
TG18-AD
TG18-UN80
TG18-AFC
TG18-CT
uniformity
• bright & dark screens
• < 30%
WIDE
MARP, Inc.
‹
‹
IF2105MP monitor
QC manual
‹
‹
MQSA QC Manual OMI-1034 Rev 1.3
software:
‹
LumiCal
• built-in internal photometer
‹
CFS
• built-in internal photometer
• or external phototimer
‹
some TG18 patterns?
WIDE
MARP, Inc.
‹
technologist QC:
‹
screen cleanliness
• daily
• soft cloth or special
cleaning fluid
‹
viewing conditions
‹
physicist tests
‹
• drawing of room layout
• < 20 lux
‹
• daily
‹
display luminance & gray
scale calibration
• black level (± 20% of target)
• white level (± 10% of target)
• JND root mean error < 20%
image quality check
• TG18-QC pattern
• weekly
viewing conditions
‹
image quality
• TG18-QC
Totoku
MARP, Inc.
‹
‹
ME511L & ME551i2 monitors
QC manual
‹ MQSA
Quality Control Manual Version 1.0
September 2007
‹
software:
‹ Medivisor
‹ external
photometer
• DTP94
‹ some
TG18 patterns?
Totoku
MARP, Inc.
‹
technologist QC:
‹
cleaning
‹
physicist tests
‹
• drawing of room layout
• < 20 lux
• daily
• cleaning agent & cloth
‹
viewing conditions
‹
• daily
‹
viewing conditions
display calibration
•
•
•
•
•
image quality
• SMPTE pattern
• daily
maximum (400 cd/m²)
minimum (0.0 cd/m²)
ambient (option)
record gamma
(displays calibration graph)
‹
SMPTE pattern
‹
visual uniformity
• white pattern or TG18-UN80
PACS vendor guidance
MARP, Inc.
‹ Carestream
‹ Richard
‹ Agfa
‹ Quest
‹ Sectra
Electronics
Carestream
MARP, Inc.
‹
Quality Control Manual
‹
‹
‹
‹
KODAK CARESTREAM Mammography
Workstation Pub. #9F6423 May 17, 2007
Totoku monitors (relabeled)
TG18 patterns
Totoku technologist QC: ‹ Carestream technologist QC:
‹
cleaning
‹
• daily
• daily
‹
viewing conditions
‹
image quality
• SMPTE pattern
• daily
viewing conditions
• daily
• daily
‹
clean monitor
‹
image quality
• SMPTE pattern
• weekly
Carestream
MARP, Inc.
‹
physicist tests:
‹
viewing conditions
‹
• draw room layout
• < 10 lux
‹
200 × ( Lmax − Lmin )
≤ 30%
( Lmax + Lmin )
‹
‹
chromaticity
• TG18-UNL80
• nonuniformity between
displays in same position
<10%
photometer (yours?)
TG18-LN12-01
TG18-LN12-18
Lmax/Lmin >250
image quality
• TG18-QC
‹
pixel defects
• black & white
• score pixels & subpixels
luminance response
• TG18-CT
• TG18-MP
‹
•
•
•
•
luminance uniformity
• TG18-UNL10
• visual & qualitative
min & max luminance
‹
phantom
• meets FFDM specs
Agfa
MARP, Inc.
‹
Quality Control Manual
‹
‹
‹
‹
IMPAX 3000 Version 2.7
September 2005
Barco monitors
(relabeled)
MediCal Pro software
Barco technologist QC:
‹
autocalibration
‹
Agfa (LCD) technologist
QC:
‹
• daily
‹
display
‹
compliance
‹
visual
• annually
• TG18 QC
measure display white
• weekly
‹
calibration setting check
• weekly
‹
• monthly (weekly)
‹
SMPTE
• daily
• weekly (or NA)
‹
viewing conditions
• daily
• annually
‹
clean monitor
measure quality level
• monthly
‹
automatic I-guard check
• annually
Afga
MARP, Inc.
‹
Barco physicist tests
‹
Agfa physicist tests:
‹
display cleanliness
‹
clean monitor
‹
viewing conditions
‹
viewing conditions
• < 20 lux
‹
• post room conditions
• < 20 lux
TG18QC
•
•
•
•
ramps
gray-scale patches
5% & 95% squares
resolution
‹
measure display white
• 600 cd/m² ± 6
‹
black level
• 1.0 cd/m²
‹
calibration settings
check
‹
measure quality level
• <10%
‹
calibrate displays
‹
SMPTE pattern
‹
auto I-guard
Sectra
MARP, Inc.
‹
Quality Control Procedures
Sectra PACS for Breast
Imaging
‹
‹
‹
‹
Version 11.1 P3 September
2008
various monitors:
‹
Planar DOME E5i
‹
Eizo RadiForce GS520
‹
Barco Coronis 5MP
‹
Siemens SMD 21500
Sectra technologist QC:
daily:
‹
monitor cleaning
‹
viewing conditions
‹
display check
•
‹
TG18-QC
printer check
‹
‹
Sectra technologist QC:
monthly:
‹
contrast visibility
•
‹
gray scale display function
•
‹
visual (TG18-QC) & photometer
display artifacts
•
‹
photometer
Lmin, Lmax
Lmax/Lmin > 250
black level
•
‹
photometer & calibration program
monitor luminance range
•
•
•
‹
TG18-QC
stuck pixels
printer test
•
print TG18-QC
Sectra
MARP, Inc.
‹
Sectra physicist tests:
‹ acceptance
testing
• 4-page chart signed by physicist
• includes CAD
‹ annual
testing
‹ viewing conditions
• room drawing
• Lamb < 0.25Lmin
‹ luminance
•
•
•
•
•
& grayscale
Lmin
Lmax
Lmax/Lmin > 250
Lmax & Lmin match for all monitors
GSDF passes
‹ phantom
medical physicist surveys
MARP, Inc.
medical physicist surveys
MARP, Inc.
‹
initial evaluations
‹ newly
‹
installed RWS
annual surveys
‹ at
least every 14 months
‹ MQSA
does not require that all mammography tests
be performed at same time
‹
evaluations
‹ RWS
‹ one
‹
moved
or both monitors replaced
if requested by FFDM service engineer following
service or upgrade
survey scenarios
MARP, Inc.
‹
‹
‹
‹
perform tests required by FFDM QC manual
perform tests required by monitor or PACS
QC manual
perform tests as deemed appropriate by
medical physicist
perform all available tests
‹ including:
• TG18
• FFDM manufacturer
• monitor / PACS manufacturer
‹ for
each 5 MP monitor used for final soft-copy
interpretation
survey caveats
MARP, Inc.
‹
privileges
‹ most
required QC applications do not require
privileges to run
‹ many
consoles require special privileges to access
the menu which start the QC applications
‹ communicate
‹
with IT
photometers
‹ calibrations
‹ gray
vary
scale specs based on a specific photometer
medical physics survey
MARP, Inc.
‹
common tests
‹
room layout
• drawing
• ambient light
‹
image quality
• TG18-QC or SMPTE
‹
luminance values
• Lmax
• Lmin
• Lmax/Lmin > 250
‹
visual uniformity
‹
MAP phantom
‹
technologist QC
‹
other tests
‹
gray scale calibration
‹
uniformity
• <30%
‹
other available TG18
tests
room layout
MARP, Inc.
TG18 QC
MARP, Inc.
SMPTE pattern
MARP, Inc.
TG18 QC Cx
MARP, Inc.
TG18 CT, MP, AFC
MARP, Inc.
luminance calibration
MARP, Inc.
‹
internal photometers
‹ behind screen
‹ in
front of screen
luminance calibration
MARP, Inc.
‹
external photometers
‹ connected
to RWS
• supplied with RWS
• referred to in QC manual
and application
• read by calibration program
• values shown on monitor
‹ not
connected to RWS
• values on its on display
• physicist or service
engineer
luminance calibration
MARP, Inc.
gray scale calibration - GE
MARP, Inc.
uniformity - visual
MARP, Inc.
bright screen (dark pixels)
L255
TG18-UN80
TG18-UNL80
dark screen (white pixels)
L0
TG18-UN10
TG18-UNL10
uniformity – measured on bright monitor
MARP, Inc.
bright screen
L255
TG18-UN80
TG18-UNL80
uniformity – measured on dark monitor
MARP, Inc.
dark screen
L0
TG18-UN10
TG18-UNL10
MAP phantom
MARP, Inc.
technologist QC
MARP, Inc.
technologist QC
MARP, Inc.
‹
do the test intervals match the QC manual?
‹ automated
test frequency may not be programmed
correctly
‹ IT
can change automated test frequency
review workstation report
MARP, Inc.
‹
part of FFDM
survey report
‹
separate from FFDM
survey report
‹ may
be surveyed at
different time
• allowable under MQSA
‹
allows for:
‹ survey
of monitor only
• evaluation due to
replacement or relocation
‹ multiple
monitor surveys
for single FFDM
‹ monitor
survey for FFDM
in another part of
country
monitor identification & recommendations
MARP, Inc.
1
room layout
MARP, Inc.
2
phantom
MARP, Inc.
3
TG18 QC or SMPTE
MARP, Inc.
3
gray-scale calibration
MARP, Inc.
3
visual uniformity
MARP, Inc.
4
measured uniformity
MARP, Inc.
4
QC program
MARP, Inc.
5
summary
MARP, Inc.
‹
‹
‹
FFDM sites may use 5 MP monitors from
manufacturers other than the FFDM unit
manufacturer
many different QC manuals may describe a
FFDM unit / softcopy review workstation
a comprehensive set of monitor tests may
cover most or all of the MQSA required tests
MARP, Inc.
FFDM in the Field: Physicist's Role in the
QC of Third-Party Review Work Stations
July 2009
Carl R. Keener, Ph.D., DABMP, DABR