Recent Developments in Medical Imaging Characteristics and Performance Evaluation of Digital Image Displays Hans Roehrig University of Arizona The Totally Digital and Film-less Radiology Department is rapidly replacing the conventional film-screen based radiology department. Digital imaging sensors and digital displays are used instead of the traditional sensor and display, namely the film-screen combination and the associated film-light box. CRTs and LCDs are the most common digital displays In digital systems the functions of detection and display are independent. As a result, the detector and the display can be optimized separately by image processing. Medical Imaging is a visual science. It is the task of the radiologist to diagnose and quantify disease processes in patients with the aid of mainly anatomical images that are created by a variety of physical and physiological processes. Some of these processes are: •In Projection Radiography: The absorption of x-rays according to the attenuation coefficients µ/ρ and thickness t with different energy in the diseased organs, tissues and bones of the patients. I = I0 × e− ( µ / ρ )× ρ × t Radiologist “reading” radiographs. Notice the environment, the ambient light is moderate in order to enhance the perception 1 Computed Tomography: Imaging of slices. Acquisition using projections and reconstruction by (usually filtered) back-projection. •In Diagnostic Ultrasound: The reflection and transmission of the mechanical energy of inaudible sound waves according to differences of the acoustic impedance at both sides of tissue interfaces, using pulse-echo techniques. Snell’s law of Refraction •In Magnetic Resonance Imaging: The magnetic resonance of mainly Hydrogen nuclei (Protons), generated using radio-frequency pulses at different pulse repetition times TR and different pulse echo times TE. T1 Weighting T2 Weighting The correct diagnosis depends critically on several physical and psycho-physical factors, all of which are related to the quality of the displayed image and the observer’s perception capabilities: M or Mz represents longitudinal magnetization Mxy represents transverse magnetization 2 In particular: •The accuracy and precision with which the clinical images represent the patient’s anatomy and pathology. •The specifics of digital image processing and image manipulation, i.e., contrast enhancement and spatial resolution compensation. •The accuracy and precision with which the human observer, the radiologist, can perceive the patient’s pathology. •A particular influence on the perception by the radiologist has the widely applied DICOM “Greyscale Standard Display Function” (GSDF). This display function perceptually linearizes the display. •Perceptual linearization means equal numbers of digital steps in the image at the display input lead to equal number of perceptual luminance steps at the display output It is clear that amongst all components of the imaging chain, the display has a prominent importance: It is the component, where the bits meet the brain. That’s where the psycho-physical characteristics of the human observer come in. The display certainly must meet the requirements for spatial resolution, display contrast and SNR (see Rose-Model below). Digital X-Detector (DR) Quantum Efficiency ηx Integration Time tint,detector Image on the display Image in the radiologist’s brain X-ray photon fluence rate Φx Background area ab Object area ao Display signal-to-noise in image detector, on the display and hopefully in the brain: (S/N)Display = (Φ Φ x ao tint, detector ηx)1/2 (C/(2-C)1/2) Display of large amounts of data on film, CT slices of one patient study on the traditional alternator. Above: Characteristic curves (H&D) of film and film-screen combinations : DLog E response Below: MTFs of film and film-screen combinations; superb resolution The problem of film: Film serves as detector and as display. Image processing is very hard to do with data displayed on film. One needs a film digitizer at least. 3 Schematic of CR, Computed Radiography, the first practical digital imaging techniques for Chest Radiography Electronic Display Devices CRT LCD In digital systems the functions of detection and display are independent. As a result, the detector and the display can be optimized separately by image processing. In The most common displays for filmless radiology Single LCD Pixel: Plane of transmitted light CRT Display of chest images on two CRTs (above right) and on two LCDs (below right). Both CRTs and both LCDs are standing side by side, and each one having the capability to display images with a pixel matrix of 2048 x 2560 pixels. LCD Plane of back light 4 Psychophysical Evaluation of Displays Human observers are the detectors of objects and backgrounds. • Experimental Setup: Human observer views objects displayed on a background and determines presence or absence of object. The human observer serves as the detector. Clinical Object Simulated Object Signal = ∆L = LB – LO SNR = Noise = σ LB , LO ,Tot = (1 / 2) × σ Background LB , B Background LB , B Object LO, O Object LO, O Schematic illustrating the simple Rose Model of Vision. It relates the Probability of Detection to signal-to-noise ratio and contrast threshold. Presence of spatial noise on the display may reduce the signal-to-noise ratio and consequently lead to the wrong diagnosis. The Rose Model OBJECT AREA Detection of simulated objects on simulated backgrounds, all of which can be described analytically: Psychometric curves, Threshold Contrasts, Detection Probability (PD), Just Noticeable Differences (JND), Rose-Model, signal detection theory. BACKGROUND AREA 2 LB +σ 2 LO ∆L σ O ,B , L Contrast = ∆L/Lmax Threshold − Contrast = CT = Background LB Object LO ∆L JND = LB LB Detection of a uniform aperiodic object on a uniform aperiodic background Psychometric Curve (2 – C) k2 ao = ---------- * ---------ΦO η tint C2 aO = Area of object to be detected ΦO = Photon fluence rate incident on object Observer η = Efficiency of photon utilization C tint k RANGE OF VIEWING ANGLES: 0.1 Degrees to 0.53 Degrees Observer (detector or human observer) = Contrast of object relative to background, defined as (Φ O - ΦB)/ΦO = Integration time or exposure time = Signal-to-noise ratio SNR = ∆L σ O,B , L JND 5 There are different detection strategies and configurations with different objects, all of which result in different threshold values and different JNDs Uniform aperiodic object (square or disc) on uniform background Uniform periodic object (sine-wave or square-wave patterns) on uniform background Non-uniform aperiodic object (Gaussian intensity distribution) on uniform background Series of uniform aperiodic objects (disks of different diameter and different intensity) on uniform background DICOM Calibration, Barten Model ∆L JND Threshold − Contrast = CT = = LB LB Perceptual Linearization: Equal steps in driving levels DDL result in equal steps in perception or JNDs Threshold Contrasts and JNDs, the Basis for the DICOM Standard Sinewave Pattern Frequency: 4 lp/deg Background, Luminance Lb Object, Luminance Lo > Lb Typical Viewing Conditions: Viewing Distance dv = 0.5 m Viewing Angle αv = 2 deg Contrast C = (|Lb - Lo|) / Lb At Threshold (50 % Probability of Detection): Lb - Lo = JND Lb, n+1 = Lb, n + JNDn Standard Display Function: Mathematical Interpolation of 1023 Luminance Levels from L = 0.050 (cd/m2) L = 4000 (cd/m2) Log10 L I = a + c • ( Ln I) + e • ( Ln I )2 + g • ( Ln I )3 + j • ( Ln I )4 2 3 4 5 1 + b • ( Ln I) + d • ( Ln I ) + f • ( Ln I ) + h • ( Ln I ) + k • ( Ln I ) I = index (1 to 1023) of the Luminance Levels LI of the JNDs a = -1.30119 b = -0.025840 C = 0.080243 d = - 0.10320 f = 0.028746 g = - 0.025468 h = - 0.0031979 j = 0.0013635 JND 1 2 3 4 . . 20 21 22 to L (cd/m2) 0.0500 0.0546 0.0594 0.0644 JNDI (cd/m2) 0.0046 0.0048 0.0050 0.0052 0.1750 0.1839 0.1930 0.0089 0.0091 0.0094 JND 993 994 995 996 . . 1021 1022 1023 e = 0.13647 k = 0.00012993 L (cd/m2) 3283.6680 3304.9660 3326.4010 3347.9740 JNDI (cd/m2) 21.42 21.55 21.7 21.83 3934.7960 3960.2810 3985.9310 25.4850 25.6500 25.8150 In General: JNDn+1 > JNDn 6 One Single Display Function is better as long as it can be repeated and serve as a standard Additional advantage: Perceptual Linearization: The display is matched to the contrast sensitivity of the human eye! LUMINANCE [CD/M**2] 1000.0 Equal steps in driving levels DDL result in equal steps in Perception or JNDs 100.0 1000 10.0 600 Display Function Before Calibration Luminance(Cd/m2) 0.1 0 256 512 JND INDEX 768 1024 JND Distance 100 1.0 400 10 Display Function After Calibration 200 1 DICOM Standard Display Function. The function is defined for the luminance range from 0.05 to 4000 cd/m2 . The just-noticeable difference applies to 2-degree targets with sinusoidal modulation of 4 c/deg calculated with the Barten model of human contrast sensitivity 0 590 JNDs Lmax,2 Lmax,1 80:1 200 100.0 200:1 Lmin,2 Monitor 1 1000.0 10.0 Lmin,1 610 JNDs 1.0 Right: Multitude of display functions of CRT for different contrast and brightness setting. Select one as standard. 0.1 0 256 512 JND INDEX 768 1024 Luminance [cd/m^2] LUMINANCE [CD/M**2] 100 Digital Driving Levels(DDLs) Left: Characteristic curve of Film-Screen Combination: Only one fixed curve; no contrast change possible 1000.0 Monitor 2 0 0.1 Gamma 1.97 100.0 Brightness 189 100 Gamma 1.93 10.0 Contrast 255 200 Gamma 2.77 Gamma 2.97 1.0 0.1 1 10 100 Command Level [ADU] 1000 7 What can go wrong in the Reading Room? What can go wrong with a display ? The luminance of the digital displays can change, the spatial resolution can change, the display can be set-up incorrectly…. a myriad of problems. This needs to be evaluated in the Reading Room. There is a need for Image Quality Control in the Reading Room The image quality of the digital displays also needs to be tested and certified when the displays are acquired from the display companies. This testing is called “Acceptance Testing”. Acceptance Testing is done in the Laboratory. Original Image Good MTF 1.00 0.80 MTF 0.60 0.40 0.20 0.00 0.00 0.50 1.00 1.50 2.00 Spatial Frequency (lp/mm) 8 Good Display Function Bad MTF 1.00 Luminance 0.80 MTF 0.60 0.40 0.20 Command Level 0.00 0.00 0.50 1.00 1.50 2.00 Spatial Frequency (lp/mm) Bad Display Function Noise free Luminance Command Level 9 Construction and Operation of CRTs and LCDs Noise present Schematic of a Cathode Ray Tube Display, an analog Display Cathode Modulator (G1) D First Anode (G2) D f b B t Notice the “Triode-like” control grid G1 and the brightness and contrast controls Triode configuration in CRT electron gun Serial input of information requires very high bandwidth 10 1000.0 ε Generalized current vs. voltage equation for triode Ic = Cathode Current K = a constant of proportionality, Vd = Drive voltage on the modulator (G1) Vc = the cutoff voltage and ψ, ε are exponents that must satisfy the equation ψ+ε=3/2. ψ is commonly called gamma γ Typical values for γ are between 1.5 and 3.5 Gamma 1.97 100.0 Brightness 189 100 Gamma 1.93 Contrast 255 200 10.0 Gamma 2.77 Characteristic curves of a CRT monitor for contrast settings 200 through 255 at brightness settings100 (dashed curves) and 189 (solid curves). Addressable matrix: 2048 x 2560 pixels Gamma 2.97 1.0 1000 0.1 1 10 100 Command Level [ADU] Characteristic curves of a CRT monitor for brightness settings 200 through 255 at contrast settings 150 (dashed curves) and 222 (solid curves). Addressable matrix: 2048 x 2560 pixel 1000 Luminance [cd/m^2] I c = KVd Vc Luminance [cd/m^2] ψ Gamma 1.66 Gamma 1.94 100 Brightness 255 200 10 Contrast 150 Gamma 1.88 Gamma 2.32 1 1 Video Signal and Scanning Contrast 222 10 100 Command Level (ADU) 1000 Ideal line raster with no overlap and no residual raster modulation, and real CRT line raster with overlap of lines and finite raster modulation. 11 Rise- and Falltime of the Amplifier and the Effect on Bandwidth Pixel on White Level Black Level Pixel off Pixel on 90 % 10 % tpix trise fall tpix >> tfall + trise Assuming tfall = trise Bandwidth f = 1/ (4 trise ) For 2048 x 2560 pixel matrix, 71 Hz refresh rate, blanking time 26 % of frame time: Ideally: Typically: tpix = 2 E-9 sec trise = 1.0 E-10 sec: Df = 2.5 GHz trise = 6.25 E-10 sec: Df = 400 MHz Resolution-Addressability Ratio: RAR Scanned Raster, (horizontal and vertical) defining nominal pixel size ah,nx av,n = Anominal Beam Spot Shape, defining actual pixel size aFWHM in terms of FWHM FWHM 50% Amplitude CRT-Vertical Raster LCD –Pixel Structure The CRT is to a certain degree an analog-discrete system, analog and continuous in the horizontal direction, discrete in the vertical direction, while the LCD is discrete in both directions, a totally digital system RAR = ((1 / 2) × FWHM ) 2 × π AFWHM = Ano min al ah , n × av , n 12 AMPLITUDE [RELATIVE UNITS] 100 SPOT FIXED 80 60 SPOT BROADENED DUE TO MOTION 40 SPOT BROADENED DUE TO LIMITED ELECTRONICS BANDWIDTH 20 0 2 4 6 8 TIME [NSEC] 10 12 Schematic illustrating the width of a single pixel as given by a Gaussian spot, which moves during the “video-on time” for a single pixel (approx 2 nsec) and which is convolved with the time response function of the electronics (rise time and fall time about 1.4 nsec each. Attenuation of ambient light by adding a neutral-density filter (“panel”) to the faceplate and by making the faceplate itself a neutral-density glass. 33 % of the phosphor light is transmitted, but only 9.3 % of the incident ambient light is returned 13 Spatial Noise of CRTs: Phosphor Granularity CCD Camera images of two types of CRT phosphor layers taken with high optical magnification P45 Binned 8 x 8 CCD Pixels P104 P45 P104 P 45 P 104 In this case the CRT line structure was removed by defocussing CRT electron optics, which is why the SNR is high. Schematic of LCD Display Single LCD Pixel: Plane of transmitted light Plane of back light Single pixel consisting of 3 sub-pixels 14 15 Intrinsic display function of LCD and the use of reference Potentials to achieve a gamma of about 2.2 Display Functions of an LCD (LCD2) and their differential before and after calibration at a precision of 12.58 bits. 1000 20 1.20 Schematic Illustrating Cause of Non-Monotonic Display Functions of LCD Display. Before Calibration 16 Intrinsic display function of liquid crystal with steep slope, not useful for display of gray scale 0.40 Intrinsic display function of liquid crystal, made useful by applying 8 reference voltages to achieve gradually changing grayscale. Frequently the effect of the reference voltages is a nonsmooth display function with edges and kinks. L/ DDL (Cd/m2) 0.80 Luminance(Cd/m2) Relative Transmission of LCD 100 12 After Calibration 10 8 Derivative before calibration 1 Discontinuity due to one of the 8 reference potentials 4 Derivative after calibration 0.00 0.1 0.00 0.04 0.08 1000 600 Display Function Before Calibration JND Distance 100 400 Display Function After Calibration 200 Digital Driving Levels(DDLs) 300 590 JNDs Lmax,2 1000.0 LUMINANCE [CD/M**2] Equal steps in driving levels DDL result in equal steps in Perception or JNDs Luminance(Cd/m2) 100 0 Additional advantage: Perceptual Linearization: The display is matched to the contrast sensitivity of the human eye! 10 0 0.12 Relative Driving Level Monitor 2 Lmax,1 80:1 100.0 200:1 Lmin,2 Monitor 1 10.0 Lmin,1 610 JNDs 1.0 200 1 0.1 0 0.1 0 100 200 Digital Driving Levels(DDLs) 0 256 512 JND INDEX 768 1024 16 Example of Spatial Noise for LCDs: Temporal Noise: LCDs exhibit also temporal noise Spatial Noise are the Local and Stationary Luminance Variations from Pixel to Pixel and from Sub-Pixel to Sub-Pixel In general, subtraction of two CCD images generated shortly one after another in time and at the same exposure time results in the disappearance of the spatial features. The RMS in the subtracted images is substantially smaller than that in the original images. In addition, subtraction of two subtracted images leads to an increase in the RMS by a factor of square-root of about 2, indicating that the noise remaining in the subtracted images is practically temporal noise which increases when the respective images are subtracted. Temporal noise represents a small fraction of an LCD’s total noise (see next slide). CRT-Vertical Raster LCD –Pixel Structure The CRT is to a certain degree an analog-discrete system, analog and continuous in the horizontal direction, discrete in the vertical direction, while the LCD is discrete in both directions, a totally digital system 17 (a) (a) (b) (b) Figure 13 Display contrast expressed as dL/L as a function of the viewing direction for a medical imaging CRT (a) and for an AMLCD [(b) is in the horizontal direction and © is in the diagonal direction] Figure 4.12 Display luminance curves as a function of the viewing direction for a medical imaging CRT (a), and for an AMLCD [(b) is in the horizontal direction, and (c) is in the diagonal direction] (c) (c) Physical Evaluation of Displays Use of CCD Camera for evaluation of an LCD 18 Quantitative evaluation of Veiling Glare Comparision of Veiling Glare in CRTs and LCDs 0.0500 Measurement of Veiling Glare 0.0400 (Image of a Circle) White Surround Veiling Glare CRT Black Center 0.0300 0.0200 Disk Diameter D VG = 0.0100 L(Disk, Center) - L(Monitor, Dark) L(White Surround) LCD 0.0000 0 50 100 150 200 250 300 350 400 Diameter of the disk (mm) 19 Uniformity of Totoku LCD 1 at 38.78 cd/m2 (View 1) Maximum Luminance Deviation: 27.79% 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Use of CCD Camera for evaluation of an LCD Measurement of Luminance at 40 positions throughout the display Optical magnification: Select 8 x 8 CCD pixels to represent 1 CRT pixel in order to achieve “Oversampling” CRT, 1 Pixel, width wCRT CCD, 8 Pixels, total width (8 * wCCD) Squarewave Pattern for MTF Single Line for MTF Examples of actual CCD Images Optical Magnification: mo = (8 * wCCD) /wCRT Result of “Oversampling”: Oversampling will permit to consider the actually discrete LCD and CRT (at least in the vertical direction) a continuous system, so we can do FFTs as in continuous systems. “Pixel Emitted Intensity” Horizontal pixel strings of different length: 1 pixel, 2 pixels 3 pixels, 4 pixels …. 10 pixel 20 Evaluation of Spatial Resolution: The Modulation Transfer Function (MTF) Spatial Resolution: MTF: Involve 1-D or 2-D Fourier Transform of image with stimulus The MTF is defined as the Fourier Transform of the imaging system’s Point Spread Function (PSF) or Impulse Response, normalized to unity at zero spatial frequency: MTF = |P(f)| / P(0) Requirement: The imaging system must be a linear system Noise: Noise Power Spectrum (NPS): Involve 2_D Fourier Transform of uniform image Unfortunately, a CRT is not a linear system: The luminance as a function of the command level ADU is given by L = a (ADU) with between 1.5 and 3.5 However, with a small signal ∆ADU as input superimposed on a mean value ADUmean,, a non-linear system can be considered a linear system: ADU = ADUmean + ∆ ADU So we display a small signal ∆ ADU on a uniform background ADUmean, The general input-output relation for a linear shift invariant system may be written as a convolution or folding integral: wout(u) = p(u - u') win(u') du' The general shorthand notation is wout(u) = p(u) * win(u) If the delta function is the input (“impulse”), win(u) = δ(u), the output is then woutδ(u) = p(u-u')δ(u')du' = p(u) = impulse response Fourier transform: 1{w in(u)*p(u)} Examples of input stimuli for the measurement of MTF: Squarewaves (superpositions of sinewaves), lines (rectangles) and white noise Vertical SquareWave ADUmean ADUmean + ∆ADU = Win(f) P(f) In Fourier Space: convolution turns in to multiplication: Wout(f) = Win(f) P(f) Vertical Line ADUmean ADUmean + ∆ADU Also: p(u), which was considered the system Point Spread Function, turns in Fourier Space into MTF = |P(f)| / P(0) If the input cannot be a perfect “impulse”, i.e. a delta function, corrections for the actual input have to be made P(f) = Wout(f) / Win(f) White Noise ADUmean Standard Deviation ∆(ADU) 21 1.20 Comparison of the MTFs of an aperture modulated monochrome LCD (Nyquist frequency: 1.46 lp/mm) and a 5M-pixel CRT with a P104 phosphor (Nyquist frequency: 3.47 lp/mm) Horizontal MTF obtained from the Line Spread Function (Totoku LCD, 8 bits mode, Nyq. Freq: 2.42 lp/mm) 1.2 32.75 cd/m2 0.8 181.38 cd/m2 LCD: Horizontal MTF 0.80 6.21 cd/m2 MTF LCD: Vertical MTF Pixel Size 0.206 mm 0.4 Vertical MTF obtained from the Line Spread Function (SIEMENS 5M LCD, Nyq. Freq: 3.03 lp/mm) 1.2 MTF 43.82 cd/m2 CRT: Vertical MTF 0 0.8 0 0.40 0.5 1 1.5 2 2.5 Spatial Freq (lp/mm) Pixel Size 0.165 mm 10.26 cd/m2 MTF CRT: Horizontal MTF 203.02 cd/m2 0.4 0.00 0.00 0.40 0.80 1.20 0 Relative Spatial Frequency (Nyquist) 0 1 2 3 4 Spatial Frequency (lp/mm) 10 CCD-Camera Image of Uniform LCD display. Notice the LCD structure, leading to the spikes in the NPS NPS of a P104 CRT 1 The pixel dimension of the CRT is 2048 * 2560. It is measured from a 127 ADU uniform image. The mean value of the CCD image has been normalized to 1. The Mag. is 8. 0.1 NPS (m m 2 ) 0.01 2D NPS of Totoku LCD at 32.75 cd/m2 (8 bits mode) 0.001 0.0001 Y direction Spatial Noise of CRTs: Phosphor Granularity X direction 1E-005 CCD Camera images of two types of CRT phosphor 1E-006 layers taken with high optical magnification Nyquist Frequency (N f): 1E-007 3.4722 lp/mm 4*Nf : 13.8889 lp/mm 1E-008 0 10 20 Frequency (lp/mm) 30 Spatial Noise for CRT 2-D Noise Power Spectrum P 45 P 104 22 Pixel Structure of LCDs used for Frame-Rate Modulation or for Sub-Pixel Modulation or for both. Two Pixels are shown. Each Pixel has 3 Sub-Pixels Pixel 1 with 3 sub-pixels Pixel 2 with 3 sub-pixels Comparison of Horizontal and Vertical NPS ( DDL 55) 1 0.1 0.01 10 NPS of a LCD The pixel dimension is 2048 * 1536. It is measured from a 137 ADU uniform image.The mean value of the CCD image has been normalized to 1. The Mag. is 15. 1 0.001 0.1 X direction LCD spatial noise alone 0.01 NPS (mm 2 ) LCD Structure Noise and LCD spatial noise 0.0001 Y direction 1E-005 0.001 0.0001 1E-006 1E-005 Nyquist Frequency (Nf): 1E-006 1E-007 2.4369 lp/mm 0 1E-007 0 10 The pixel dimension of the CRT is 2048 * 2560. It is measured from a 127 ADU uniform image. The mean value of the CCD image has been normalized to 1. The Mag. is 8. 0.1 10 Y direction NPS of a LCD 0.1 1E-006 X direction Nyquist Frequency (Nf): 3.4722 lp/mm 0.01 4*Nf : 13.8889 lp/mm 1E-008 0 10 20 Frequency (lp/mm) NPS (mm2) 1E-007 20 o o Display a uniform field on display Take an image with the CCD camera o Perform 2-D Fourier Transform (FT) o Integrate NPS and reduce to 1 display pixel (“down-sampling”) The pixel dimension is 2048 * 1536. It is measured from a 137 ADU uniform image.The mean value of the CCD image has been normalized to 1. The Mag. is 15. 1 X direction 1E-005 16 Procedure for finding the SNR per display pixel of a display from the response to a uniform input to the display under test: 0.001 0.0001 12 40 LCD Noise seemes to be similar or even larger than CRT noise 0.01 NPS (m m 2 ) 20 30 Frequency (lp/mm) Comparison of LCD noise with CRT Noise: NPS of a P104 CRT 1 8 Structure permitting Aperture Modulation with almost 1800 levels 1E-008 10 4 4*Nf : 9.7276 lp/mm Y direction 0.001 0.0001 30 1E-005 Nyquist Frequency (Nf): 1E-006 2.4369 lp/mm 1E-007 Example of a uniform display field 4*Nf : 9.7276 lp/mm 1E-008 0 10 20 30 Frequency (lp/mm) 40 23 Illustration of derivation of the 5 x 5 pixel image of a 5 x 5 pixel section of an LCD from a 15 x 15 CCD pixel CCD image. “Convolution and Down-Sampling”. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.00E+0 Normalized Spatial Noise Power Spectra for Sub-Pixel Modulated LCD 1.00E-1 NPS in Horizontal Direction NPS in Vertical Direction 1.00E-2 NPS (mm^2) . . . . . . . . . . . . . . . . . . Integration of the Noise-Power Spectrum to find the Variance σ2 1.00E-3 1.00E-4 CCD image of 5 x 5 pixel section of an LCD, taken at a 3:1 pixel-ratio (3 x 3 CCD pixels per LCD pixel). The CCD image is a 15 x 15 array of CCD pixels Ideal 5 x 5 pixel image of a 5 x 5 pixel section of an LCD. For simplicity we assume an LCD pixel is a square 1.00E-5 LCD-Nyquist Frequency 2.42 lp/mm 1.00E-6 0.00 4.00 8.00 12.00 16.00 20.00 Frequency (lp/mm) σ2 LCD, pix = NPSonedf ≈ f ≠0 NPSupsamp*sin c2df f ≠0 Plotting the Resulting 9 SNRs What tests should be done in the Reading Room and in the Laboratory? Ask the AAPM Task Group 18! 24 American Association of Physicists in Medicine (AAPM) Task Group 18 Assessment of Display Performance for Medical Imaging Systems Types of TG 18 tests: • There are visual tests, often called qualitative or basic tests, where a human observer views the display and makes a decision on the presence or absence of a test object. AAPM On-Line Report No. 03, 2005 Can be downloaded from website: www.duke.edu/~samei • There are quantitative or advanced tests, where an instrument like a photometer or a CCD camera is used to make a measurement and provide quantitative data. 25 and therefore 26 27 28 29 We developed a CCD Camera-lens combination. It can be used hand-held in the Reading Room It can be used for both types of TG18 tests. The Optical Layout for the It can be used for both CRTs and LCDs. 04 Case of an LCD Focus Adjustment Image Quality Control of Display Systems in the PACS Environment. Hans Roehrig1, Jerry Gaskill2, Jiahua Fan1 1University 2Image Total camera length < 110mm Working F/#: 9.46 The Optical Layout for the Case of a CRT of Arizona, Tucson AZ Smith’s Inc., Germantown, MD This work was supported by a grant from NIH and by the EIZO NANAO Corporation. CCD Camera Focus Adjustment Display Glass Thickness CCD Camera Pixel matrix: 640 x 480 8 Bits Contrast Total camera length < 90mm Working F/#: 9.46 30 The Software: VeriLUM from Image-Smiths VeriLUM provides • Tracking • Calibration • MTF measurement (click MTF button) • SNR measurement (click SNR button) • Results Verifying Focusing of the CCD Camera Note the status bar at the bottom of the image. It displays the gain, max pixel value and the standard deviation of the image that help in focusing the camera. The Gain is automatically adjusted so that the image is not saturated. Check of Focusing and Alignment of the Camera and Determining the Magnification Ratio MR VeriLUM displays this test pattern to check for alignment, to verify focusing of the camera and to determine the pixel ratio MR, (the ratio of the number of CCD pixels per display pixel) while the camera is held firmly against the display or held on a firm stand on a table. The MTF of the display under test is found from the line response Here we input all the parameters required for the calculation of an MTF: • Horizontal or Vertical Line • Line and Background Values Gain Max Pixel Value Focus Index 31 The input stimulus for the MTF from the line response: A line displayed on the display, and the line image taken with the CCD The Profile of the Line Image: The LSF 60 Pixel Intensity 40 20 0 Single line test pattern the stimulus or input to the display system CCD camera image of the single line from the display which is the output of the display system i.e. lineresponse of interest Plotting the Resulting MTF -20 0 100 200 300 400 500 Pixel Position This line profile from the subtracted image is subjected to a Fourier Transform to get the MTF. Procedure for finding the SNR per display pixel of a display from the response to several (perhaps even 9) uniform inputs to the display under test: o o Display a uniform field on display Take an image with the CCD camera o Perform 2-D Fourier Transform (FT) o Integrate NPS and reduce to 1 display pixel Example of one of several perhaps even 9 uniform fields 32 Illustration of derivation of the 5 x 5 pixel image of a 5 x 5 pixel section of an LCD from a 15 x 15 CCD pixel CCD image. “Convolution and Down-Sampling”. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Plotting the SNRs for the case of 9 different luminance values . . . . . . . . . . . . . . . . . . CCD image of 5 x 5 pixel section of an LCD, taken at a 3:1 pixel-ratio (3 x 3 CCD pixels per LCD pixel). The CCD image is a 15 x 15 array of CCD pixels Ideal 5 x 5 pixel image of a 5 x 5 pixel section of an LCD. For simplicity we assume an LCD pixel is a square 33