Collaborators/Support C-arm Flat Panel CT: Image Quality Considerations Rebecca Fahrig Department of Radiology, Stanford University Introduction o C-arm CT for visualization in 3D of vasculature and other high-contrast structures has become commonplace o The transition from XRIIs to digital flat panels opened the doors to the possibility of low-contrast and high resolution 3D CT imaging in the interventional suite o The Physics Gang:: ‘Bob’ R.L.Dixon, ‘Tom’ J. Payne, ‘Rick’ R.L. Morin o A. Ganguly, N. Strobel and T. Moore o L. Hoffman, N. Kothary, D. Sze, M. Marks, H. Do o Technical support : A. White, N.R. Bennett, J. Kneebone, M. Lozada-Parks, W. Baumgardner o Siemens Medical Solutions o Varian GTC o NIH R01 EB003524 o Lucas Foundation C-arm System :: CT System half scan, area detector, ‘thin’ CsI converter vs. full scan, narrow detector, ‘thick’ Gd2O2S 1 Creating 3D Images in the Interventional Lab 1) Rotational Angiography Run 4) In-room Display Topics of Discussion o Dose measurement o Signal and Noise o Scatter 3) Reconstruction and Visualization 2) Image transfer Dose Measurement o small 0.6cc Ion Chamber, measuring dose max. o CTDI phantom (16cm diameter, 15cm long) o Dose measured at center and eight peripheral positions for : (30x40) cm detector format based on 543 views o Beam Size (iso-center): Width: 26.67cm 16 cm Height: 20.00cm 32 cm Dose Measurement o small 0.6cc Ion Chamber, measuring dose max. o averaging with a 10-cm pencil chamber doesn’t make sense since the slab thickness is variable o CTDI phantom (16cm diameter, 15cm long) o Dose measured at center and eight peripheral positions for : (30x40) cm detector format based on 543 views 16 cm o Beam Size (iso-center): Width: 26.67cm 32 cm Height: 20.00cm 2 Dose Measurements: 81kVp Dose Measurements: 81kVp D (0) = (1 / 3) D0 + ( 2 / 3) D p 2/3 mean peripheral + 1/3 central source side 2/3 mean peripheral + 1/3 central detector side Peak Dose (mGy) 86 Center Dose (mGy) 34 “CTDIw” (mGy) 1167 Detector dose (uGy/view) 0.46 81 608 0.44 63 28 37 109 310 0.70 66 31 40 125 260 0.92 76 38 46 kVp Total mAs 70 48 Variation in CTDIw in spite of AEC. The EU guidelines for routine head CT scans specify a CTDIw of 60mGy. Dose in Abdomen Phantom Dose Slab width @ mA ms mGy det/iso 22.2/17 cm 348 4.9 34 18.4/14 cm 357 4.9 32 14/11 cm 367 4.9 27 8.4/6.5 cm 383 4.9 21 28cm 26cm 16cm Measured Doses : ‘Medium-High’ dose requested 37cm Circumference = 104cm system uses automatic exposure control : mA ramped up to max at thickest part of phantom 3 Which statement is true about dose in C-arm CT... 20% 1. 2. 20% 20% 3. 20% 4. 20% 5. Answer Collimation does not affect the measured dose CTDIw is defined as ½ the dose measured at the center of the phantom + ½ the average dose measured at the periphery of the phantom The peripheral dose is higher on the x-ray tube side of the C-arm sweep The average dose measured in the periphery of a 16-cm phantom is lower than the dose measured at the center of the phantom CTDI100 represents the angular average dose at z=0 for a scan length of L=100 cm 1. Collimation does not affect the measured dose FALSE 2. CTDIw is defined as ½ the dose measured at the center of the phantom + ½ the average dose measured at the periphery of the phantom FALSE : 1/3 center + 2/3 periphery 3. The peripheral dose is higher higher on the xray tube side of the C-arm sweep TRUE The average dose measured in the periphery of a 16-cm phantom is lower than the dose measured at the center of the phantom FALSE CTDI100 represents the angular average dose at z=0 for a scan length of L=100 cm FALSE : gotcha question … L=10 cm 4. 5. 10 Visibility vs. kVp Visibility of Low-Contrast Objects? 120 Nominal Contrast U) (3H 15mm 9mm 0% 6mm Visibility [%] 100 0.5% (5HU) % 0 .3 1. 0 (1 80 60 40 Average (70kVp) Average (81kVp) 20 Average (109kVp) ) HU o Catphan Module CTP515 used as image quality phantom (20cm housing) o Acquired 543 views over 20sec at various dose and kVp settings, Zoom 0 o Reconstructed soft tissue segment (smooth kernel, 10mm slice width) o Analyzed visibility of (outer) 5HU insets Detail Diameters [mm] (2, …, 9, 15) Average (125kVp) 0 0 2 4 6 8 10 12 14 16 Normalized Diameter [mm] Scoring Question: What size “5HU” objects can you see? 4 Nominal vs. Measured Contrast Values for a 0.5% contrast object 70 kVp 0.49 81 kVp 0.47 109 kVp 0.43 o CsI vs. Gd2O2S of thickness 1.4 mm With16cmwater o Measured contrast depends on kVp o Contrast decreases as kVp increases 1.1 Absorbedenergyfra kVp Measured Contrast Detector Efficiency? 1 0.9 0.8 0.7 Flat Panel Detector 0.6 CTDetector 0.5 125 kVp 60 0.36 MTF: 100 µm steel wire o Take advantage of the high resolution of the flat panel : ~150 µm native at panel 70 80 90 100 kVp 110 120 130 140 High Resolution Imaging In-vivo Stent in the Superficial Femoral Artery displayed at constant win/lev 1000/900 150 µm no binning 300 µm 2x2 binning 600 µm 4x4 binning 1.0 Sharp 0.8 Normal MTF Smooth 0.6 0.4 0.2 0.0 0.0 0.5 1.0 1.5 2.0 Spatial Frequency (cycles/mm) 2.5 var{µ ( x, y )} ∝ 1 ∆R 3 S 5 Visibility vs. Dose CNR ∝ Which statement is true about signal and noise in C-arm CT? 1 Dose Visibility Chart (81kVp, 543view s) 20% 100 80 20% Visibility (%) 70 60 20% 50 4. CTDIw = 19.93mGy CTDIw = 26.31mGy 20 20% CTDIw = 37.01mGy CTDIw = 54.12mGy 10 0 0 3 6 9 12 15 Detail Diameter [mm] Answer 2. 3. 4. 5. 3. 40 30 1. 1. 2. 90 Inherent object contrast increases with increasing kVp TRUE In C-arm CT head scans, better detectability occurs for kVps above 100 because the detector has been optimized for visualization of iodinefilled vessels FALSE : detector optimized for 70-80 kVp The variance in a projection image is proportional to the slice width FALSE : inversely proportional to slice width The variance in a reconstructed C-arm CT slice is inversely proportional to the square of the inslice reconstructed pixel size FALSE : inversely proportional to cube of pixel size The contrast to noise ratio is proportional to 1/dose FALSE : inversely proportional to 1/sqrt(dose) 20% 5. Inherent object contrast increases with increasing kVp In C-arm CT head scans, better detectability occurs for kVps above 100 because the detector has been optimized for visualization of small iodine-filled vessels The variance in a projection image is proportional to the slice width The variance in a reconstructed C-arm CT slice is inversely proportional to the square of the inslice reconstructed pixel size The contrast to noise ratio is proportional to 1/dose 10 Intracranial Imaging C-arm CT Clinical CT In vivo pig model, Autologous blood, NO iodine contrast ARTIFACT: Beam hardening Scatter Conebeam M. Marks, H. Do et al. 6 Body Applications : TACE Scatter is Significant o depends on such geometric factors as ARTIFACTS: Scatter Truncation • • • • air gap between object and detector focal-spot-to-object distance focal-spot-to-detector distance weakly dependent on spectrum energy (kVp) over the range of diagnostic energies of interest o The magnitude of the scatter-to-primary ratio (SPR) depends on irradiated volume L. Hoffman et al. SPR is increasing as Cone Angle (Imaged Volume) Increases A Rough Rule-of-Thumb (no collimation!) o Cone angles on clinical scanners are getting larger • 64-slice conventional scanners have a cone angle of ~ 4°° • 256-slice scanners have a cone angle of ~13°° o CT imaging using large-area flat panels • C-arm CT in interventional suites have a cone angle of ~ 12°° • On-board kV CT on radiation therapy machines have a cone angle of ~ 15°° RESULT: Scatter-to-primary ratios are on the rise… modified from Fox et al. Proceedings of SPIE Vol. 4320 (2001) 7 For a 16-cm diameter head-sized object 64-slice scanner For a 16-cm diameter head-sized object Conebeam C-arm CT 30-cm z-extent @ detector For a 16-cm diameter head-sized object 256-slice scanner For a 16-cm diameter head-sized object Conebeam Rad Therapy 41-cm z-extent @ detector 8 For a 32-cm diameter body-sized object 64 slice Rad Therapy 256-slice C-arm CT Image Quality vs. SPR A Simple Phantom Study o scatter field is shown as a lowfrequency background signal that does not vary much from view to view o the primary varies significantly depending on the angle o signal (with scatter) in view 1 behind the two rods is significantly over-estimated compared to signal (with scatter) behind a single rod in view 2 A Simple Phantom Study As SPR increases: o overall cupping increases, and therefore mean offset increases o variation due to streaking reaches a max of ~ 200 HU 9 How can image quality be maintained? o Scatter decreases Contrast-to-Noise ratio CNRs CNR0*(1-SPR/2) (an approximation in the ‘small scatter, small contrast’ regime) o To recover CNR, the dose can be increased, but CNR ∝ 1/sqrt(dose) so the dose penalty is high • e.g. for an SPR of 1, need 4x dose to get the same CNR as a scatter-free image Scatter Correction o removes mean scatter signal from the recorded signal, thereby reducing artifact, but does not remove the noise o can lead to increased artifact if the estimate of SPR is incorrect Pencil beam is attenuated and produces a scatter “impulse response” at detector. Transmitted Primary Scatter PointSpread Function Detector 1. L.A. Love and R.A. Kruger, Med Phys 1987 14(2) 1978-85. 2. B. Ohnesorge, T Flohr, K Klingenbeck-Regn, Eur Radiol. 1999; 9(3) 563-9 3. J. Maltz et. al. Med Phys. 2006 33(6), 2280. 4. Star-Lack et. al. AAPM07, EPI2k8 Grids and/or Collimators o Anti-scatter grid • removes scatter from the beam, but also removes primary • depending on the efficiency of the grid, the SPR and the resolution, it may or may not be dose efficient o Siewerdsen et al. (Med Phys 31, 2006) have shown that grids used with fluoroscopic systems (e.g. Ptrans = 75%, Strans = 25%) are not efficient unless SPR is high (>1) , and resolution is low (vox size > 1 mm) o Endo et al. (Med Phys 33, 2006) showed that a 30:1 Mo collimator was dose efficient for a 256-slice CT scanner Direct Measurement : Modulator Technique primary o “code” the primary without affecting the scatter high-frequency semi-transparent grid o Decoding (primary modulator) provides scatter-free estimate of the primary o no additional dose required same window for both = [-50, 100] HU scatter uncorrected corrected 10 Impact of Scatter on Image Quality o Images of a 16-cm ‘medium contrast’ insert in the abdomen phantom o Objects are -20, -25, -30 and -45 HU relative to background o 32, 16, 8, 4 and 2 mm diameter Which statement is true about scatter in C-arm CT? 20% 1. 20% 2. 20% 3. 4. 20% 20% 5. Scatter-to-primary ratio (SPR) increases with decreasing cone angle SPR decreases approximately linearly (at small volumes) as a function of increasing irradiated volume Scatter decreases contrast-to-noise ratio by (1-SPR/2) For a SPR of 1, an increase in dose of a factor of 2 is required to get the same contrast-to-noise ratio as a scatter free image For best image quality in C-arm CT, the collimators should always be as wide open as possible 10 Answer 1. 2. Scatter-to-primary ratio increases with decreasing cone angle FALSE : decreases SPR decreases approximately linearly (at small volumes) as a function of increasing irradiated volume FALSE : increases 3. Scatter decreases contrast-to-noise ratio by (1SPR/2) TRUE 4. For a SPR of 1, an increase in dose of a factor of 2 is required to get the same contrast-to-noise ratio as a scatter free image FALSE : need an increase in dose of a factor of 4 For best image quality in C-arm CT, the collimators should always be as wide open as possible FALSE 5. Conclusions o For C-arm CT with current detector technologies, operate at ‘low’ kVp o Present reconstructed data using the largest pixel size appropriate for the detection task o expose no more than you need to see since SPR increases as the irradiated volume increases o Consider grid/collimator properties to determine if contrast improvement outweighs noise increase for the particular task 11 References o Fahrig R, Dixon R, Payne T, Morin RL, Ganguly A, and Strobel S. “Dose and Image Quality for a Cone-Beam Carm CT System.” Medical Physics 33(12), 4541-4550, 2006. o Riederer, SJ, Pelc, NJ, and Chesler DA, „The Noise Power Spectrum in Computed X-ray Tomography“Phys Med Biol 23 (3), 446-454 (1978) o Faulkner, K and Moores, BM. “Noise and contrast detection in computed tomography images” Phys Med Biol 29 (4), 329-339 (1984). o Endo M, Mori S, Tsunoo T, Miyazaki H. “Magnitude and effects of x-ray scatter in a 256-slice CT scanner” Med Phys 33(9), 3359-3368 (2006). 12