Slide 1 Slide 2 Introduction Digital Mammography Detector QC What is FFDM QC and why is it important? What to know before you start Eric A. Berns, Ph.D. eric.berns@gmail.com Overview and compare QC tests Denver Health Medical Center University of Colorado at Denver and HSC Denver, CO Slide 3 Key take home points Slide 4 Introduction 8,331 SFM Units July 1, 2008 8,832 Total Accredited Mammo Units 13,405 Certified Facilities with FFDM units 1,991 Accredited FFDM units 2,926 8,837 Difference % Units FFDM Lorad Selenia App: Oct ‘02 Lorad CCD App: Mar ‘02 +5 6% 13,450 Siemens Novation App: Aug ‘04 % Fac w Dig 10% 8% Total Certified Mammo Facilities Fuji FCRm App: Jul ‘06 12% Certified Statistics – Past Year July 1, 2007 FFDM vs. Date +45 Fischer Senoscan App: Sept ‘01 GE Essential App: Apr ‘06 4% GE 2000D App: Jan 2000 2% 3,366 0% Jul-98 5,119 GE DS App: Feb 2004 +1,375 Dec-99 Apr-01 Sep-02 Jan-04 May-05 Oct-06 +2,193 1 Slide 5 Slide 6 Introduction ScreenScreen-Film MQSA – Mammography Quality Standards Act ACR – American College of Radiology Slide 7 Slide 8 ScreenScreen-Film ScreenScreen-Film 2 Slide 9 Slide 10 ScreenScreen-Film Slide 11 ScreenScreen-Film Slide 12 ScreenScreen-Film FFDM In FFDM, the manufacturer designs and mandates their own QC program In FFDM, you must the manufacturers’ manufacturers’ QC program 3 Slide 13 Slide 14 Vs. RWS Slide 15 Slide 16 Quality Control: Why? Reduce exposure to patients and personnel Consistent image quality Detect and correct for potential problems, before they impact image quality What is Quality Control? Determination of what is “Normal” Normal” Detection of what is “Abnormal” Abnormal” Understanding of how to return to “Normal” Normal” from “Abnormal” Abnormal” In particular, in FFDM, how do you know what you are seeing is what it is supposed to be? 4 Slide 17 Slide 18 Before You Begin QC Before You Begin Golden Rules for QC – Must use manufacturer’ manufacturer’s QC procedures Obtain proper training & CE credits (8 hours) • Mandate action limits – HandsHands-on training on actual unit: – Manufacturers’ Manufacturers’ QC may refer to Monitor & Printer • Mechanics Manufacturers’ Manufacturers’ QC – Multimodality Workstations have own separate QC • Software – Printers may have their own QC • Artifacts – Most failures result in stopping clinical imaging • Learn vendor specific tests and tricks until failure can be corrected Slide 19 Slide 20 Before You Begin Must have proper continuing experience: – 2 facilities – 6 units – Within last 2 years Before You Begin Note: – For new unit: unit: Must use most current version – For renewal unit: unit: Can use older version (version used when tested previously) 5 Slide 21 Slide 22 Before You Begin Before You Begin ACR Accreditation - www.acr.org – Physics forms • http://acr.org/accreditation/mammography/mammo_qc_forms.aspx – – – – – GE Senographe 2000D, DS, and Essential Fischer Senoscan Lorad Selenia Siemens Novation Fuji FCRm Personnel Equipment Testing Tech QC Forms Equipment Evaluation Annual Survey Forms • Note: Note: if using unit for both screenscreen-film & CR, you must accredit for both FDA Accreditation – www.fda.gov/cdrh/mammography/ www.fda.gov/cdrh/mammography/ – Other vendors when approved Slide 23 Slide 24 Before You Begin Summary Forms GE Fischer Fuji Lorad Siemens Film 6 Slide 25 Slide 26 Before You Begin Before You Begin •ACR Accreditation – Equipment Evaluation Forms GE Medical Systems Report Date Survey Date Model Site Site Senographe 2000D Room ID Medical Physicist X-Ray Unit Manufacturer Medical Physicist's QC Tests ? 2277390-100 Rev 3, 2001 ? 2277390-100 Rev 3 + Addendum 2354312-100, 2003 ? Mobile 2371698-100 Rev 0, 2003 ? Seno Advantage 2391082-100 Rev 1, 2003 PASS/FAIL/NA Slide 27 Report Date Survey Date Model SenoScan Room ID Signature Medical Physicist's QC Tests GE QC manual version(s) at facility (check all that apply): 1. Flat Field (as described in radiologic technologist tests) 2. Image Quality (Phantom) (as described in radiologic technologist tests) CNR (required) Change in CNR =0.2 (NA for Equipment Evaluations) Fibers Specks Masses Phantom IQ Test on AWS (* NA for units with Seno Phantom IQ Test on RWS – Left* Phantom IQ Test on RWS – Right* Advantage workstations) Phantom IQ Test on Printer 3. MTF Measurement (as described in radiologic technologist tests) 4. AOP Mode and Signal-to-Noise (SNR) (as described in radiologic technologist tests) 5. Collimation Assessment Deviation between X-ray field and light field is less than 2% of SID X-ray field does not extend beyond any side of the IR by more than 2% of SID Chest wall edge of compression paddle doesn't extend beyond IR by more than 1% of SID 6. Evaluation of Focal Spot Performance Measured performance within acceptable limits for large focal spot Measured performance within acceptable limits for small focal spot 7. Breast Entrance Exposure, Average Glandular Dose and Reproducibility Average glandular dose for average breast is below 3 mGy (300 mrad) Average glandular dose to a 4.2-cm-thick breast on your unit is mrad Exposure reproducibility (CV) for R and mAs must be less than 0.05 8. Artifact Evaluation and Flat Field Uniformity 9. Viewing Condition Check and Setting 10. Monitor Calibration 11. Image Quality – SMPTE Pattern 12. Analysis of Review Work Station (RWS) or Seno Advantage Screen Uniformity (required for Equipment Evaluations and as necessary) 13. kVp Accuracy and Reproducibility Measured average kVp within ±5%of indicated kVp kVp coefficient of variation =0.02 14. Beam Quality Assessment (Half-Value Layer Measurement) 15. Radiation Output Radiation output is =800 mR/s mR/s 16. Mammographic Unit Assembly Evaluation Fischer Date of Installation Medical Physicist Signature ? 2277390-100 Rev 1, 2000 ? 2371472-100 Rev 0, 2003 MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital – Lorad MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital - Fischer MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital – GE Medical Systems Site X-Ray Unit Manufacturer Date of Installation Fischer QC manual version at facility (check one) : 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. ? P-55943-OMIssue 1, Rev. 000 (2002) ? P-55943-OMIssue 1, Rev. 3 (July 2003) ? P-55943-OM Issue 1, Rev. 1 (April 2003) ? P-55943-OM Issue 1, Rev. 7 (March 2004) PASS/FAIL/NA X-Ray Field Size Alignment and Chest Wall Missed Tissue Checks X-ray field size aligns with field area indicated on breast support within =2% SID Chest wall edge of the digital image to the ruler reference mark of breast support is =8.5 mm Compression Paddle Alignment Compression paddle edges not visible within field of view Chest wall edge of compression paddle extends beyond image by =1% SID kVp Accuracy Test Measured average kVp within ±5% of indicated kVp kVp coefficient of variation =0.02 Linearity, Reproducibility and Accuracy Linearity <0.08 Reproducibility <0.035 for each technique Half-Value Layer and Output HVL =0.33 mm Al at 30 kVp, 100 mA Dosimetry – Average Glandular Dose and Output Average glandular dose for average breast is below 3 mGy (300 mrad) Average glandular dose to a 4.2-cm-thick breast on your unit is mrad Phantom Image Acquisition Test No obvious artifacts (ADU values required for both QC and Equipment Evaluations; P/F results are NA for Equipment Evaluations) Background StDev within +50/-0 ADU counts of baseline Background mean within ±100 ADU counts of baseline ADU level difference within ±300 ADU counts of baseline Image Quality Largest 4 fibers, 3 speck groups and 3 masses Fibers Specks Masses Phantom IQ Test on Review Work Station System Resolution/Scan Speed Uniformity Standard imaging mode @ 7 lp/mm: =5 transitions, =10% modulation High resolution imaging mode @ 11.1 lp/mm: =5 transitions, =5% modulation Flat Field Test Flat field test Deviations between corner ROIs and center within ±20% Geometric Distortion and Resolution Uniformity Automatic Decompression Control System Artifacts Image Display Monitor(s) Check (review daily log) Image Viewing Room Illuminance Test (=50 lux) X-Ray Unit Manufacturer Date of Installation Medical Physicist Lorad Report Date Survey Date Model Room ID Signature Selenia Medical Physicist's QC Tests ? June 11, 2002 ? 9-500-0285, Rev. 004 (2004) Lorad QC manual version at facility (check one) : ? 9-500-0285, Rev. 003 (2003) ? 9-500-0285, Rev. 002 (2003) PASS/FAIL 1. Mammographic Unit Assembly Evaluation Autodecompression can be overidden to maintain compression (& status maintained) Manual emergency compression release can be activated in the event of power failure 2. Collimation Assessment Deviation between X-ray field and light field is less than 2% of SID X-ray field does not extend beyond any side of the IR by more than 2% of SID Chest wall edge of compression paddle doesn't extend beyond IR by more than 1% of SID 3. Artifact Evaluation Artifacts were not apparent or not significant 4. kVp Accuracy and Reproducibility Measured average kVp within ±5% of indicated kVp kVp coefficient of variation =0.02 5. Beam Quality Assessment - HVL Measurement Half-value layer is within acceptable lower and upper limits at all kVp values tested 6. Evaluation of System Resolution Measured performance within acceptable limits 7. Breast Entrance Exposure and Average Glandular Dose Average glandular dose for average breast is below 3 mGy (300 mrad) Average glandular dose to a 4.2-cm-thick breast on your unit is mrad 8. Radiation Output Rate Radiation output rate is greater than 800 mR/sec mR/sec 9. Phantom Image QualityEvaluation 5 largest fibers, 4 largest speck groups and 4 largest masses are visible Phantom image quality scores: Fibers Background Optical Density Disk Optical Density Specks Masses Disk Contrast Hard copy background density must be =1.20 (with operating level =1.40) Hard copy density difference over acrylic disk must be =0.35 (with operating level =0.40) 10. Signal-To-Noise Ratio and Contrast-To-Noise Ratio Measurement Signal-To-Noise Ratio should be equal or greater to 40 SNR Contrast-To-Noise Ratio should not vary by more than ±15% CNR 11. Viewbox Luminance and Room Illuminance Mammographic viewbox is capable of a luminance of at least 3000 cd/sq m (nit) Room illuminance (viewbox surface as seen by observer) is 50 lux or less Room illuminance (monitor surface) is 20 lux or less 12. Softcopy Workstation QC White level performance Black level performance Quality level performance Uniformity performance Slide 28 Before You Begin Mammo Unit Evaluation ScreenScreen-Film Mammo – Ensure locks, detents, angulation indicators, and mechanical support devices operate properly FFDM – Same 7 Slide 29 Slide 30 Collimation Collimation Fuji FCRm GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation Same as ScreenScreen-Film X X X X X Multiple target/filter combo’ combo’s X X ScreenScreen-Film Mammo – Assure XX-ray field aligns with the light field – Collimator allows full coverage of receptor – Chest wall edge of compression paddle aligns with chest wall of receptor Multiple paddle positions FFDM – Same – Multiple targe/filters targe/filters and/or multiple paddle positions – Chest Wall Missed Tissue – Compression Plate Overlap on the Chest Wall Side Slide 31 X X Chest Wall & Missed Tissue X X X X Compression Plate Overlap on Chest Wall X Slide 32 Limiting Spatial Resolution Spatial Resolution ScreenScreen-Film Mammo Fuji FCRm – LineLine-pair test tool GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation X X • Focal spot FFDM – LineLine-pair test tool • Focal spot • Detector & imaging chain Same as ScreenScreen-Film (Focal Spot Eval) Eval) Measure using LP phantom on top of acrylic MTF X X X X 8 Slide 33 Slide 34 GE 2000D, DS, & Essential GE 2000D MTF Measurement – Monthly – Raw Image Focal Spot Evaluation •Mo/Mo •Rh/Rh •Large & Small Spots (1.5 mag) mag) MTF (%) = (Std. Dev.) x 222 / (Mean Dark ROI – Mean Light ROI) MTF (%) @ 2 lp/mm lp/mm > 58% Slide 35 MTF (%) @ 4 lp/mm lp/mm > 25% Slide 36 GE 2000D 2000D Addendum: SubSub-system MTF – Optional replacement for MTF & Focal spot – No need for film • Use specific bar pattern & 4.5 cm acrylic block GE DS & Essential MTF and CNR Measurement - Weekly – Test consistency of CNR and ensure contrast is adequate • IQST Test tool • Use ROI tools to calculate subsub-system MTF • Repeat at 1.8 mag Action Limit: – MTF @ 2 lp/mm lp/mm > 58%, MTF @ 4 lp/mm lp/mm > 25% – Change in CNR must not exceed 0.2 9 Slide 37 Slide 38 Lorad & Siemens Fuji FCRm Evaluation of System Resolution System Resolution – 5-15 lp/mm lp/mm Test Pattern – Up to 10 lp/mm lp/mm Test Pattern The system limiting spatial resolution must be: – 4 cm acrylic > 7 lp/mm lp/mm – Parallel & Perpendicular to tube with 33-5 degree angle The system limiting spatial resolution must be: > 8 lp/mm lp/mm + 2 lp/mm lp/mm (6 to 10 lp/mm) lp/mm) Slide 39 Slide 40 AEC Performance AEC ScreenScreen-Film Mammo – Measure optical densities Fuji FCRm GE 2000D, DS, Essential Lorad Selenia Siemens Novation X X X X X X X X Fischer Senoscan Same as ScreenScreen-Film • Different thicknesses using clinical modes • Different density settings ((-2, -1, 0, +1, +2, etc.) FFDM – Measure resultant techniques – Measure signal, exposure, & SNR values Density Control Function X ACR Reproducibility X Image Mode Tracking X CNR X SNR 10 Slide 41 Slide 42 GE 2000D GE DS & Essential DS specific QC tests AOP Mode and SNR Check – Variable thicknesses of acrylic – Std, Auto – Evaluate: • Correct techniques? • Adequate SNR? Acrylic Thickness 2.5 cm TargetFilter Mo-Mo Selected kVp 27 kVp Selected mAs 20-60 4.0 cm Mo-Rh 28 kVp 35-90 6.0 cm Rh-Rh 32 kVp – AOP Mode and SNR Check • Use builtbuilt-in software for image acquisition 2.5 cm Acrylic • AOP: STD/Auto Exposure Parameters For AOP STD mode oly Acrylic Thickness (mm) 4.0 cm Acrylic Track/Filter mAs kV 25 Mo/Mo 2020-60 26 50 Rh/Rh 4040-90 29 60 Rh/Rh 4545-95 31 35-90 Each “raw” raw” image must have a measured SNR of at least 50 Slide 43 SNR must be > 50 6.0 cm Acrylic Slide 44 Lorad Selenia Automatic Exposure Control – AUTOAUTO-FILTER – 2, 4, 6, 8 cm Siemens Novation AEC Image Stability and Reproducibility and SNR – ACR Phantom – 4 cm ((-5 thru +5, or -3 to +4) – Use ROI to measure mean pixel value inside virtual AEC detector Performance Criteria – Pixel value should not vary by more than 10% of mean – Exposure compensation steps shall meet requirements in pixel value table – Mo/Mo, 28 kV, “H”, sensor 1 – Record mAs, mAs, SNR, Mean, Entrance Exposure (5 times) Action Limits: – Coef. Coef. of Var for mAS and X < 5% – SNR and Mean shall not vary by + 15% of Mean 11 Slide 45 Slide 46 Siemens Novation DR Fuji FCRm AEC Thickness Tracking Density Control Function – 2, 4, and 6 cm – “H” mode – 4 cm acrylic, ACR phantom technique – Max deviation = (Max difference / mean value ) * 100 – Repeat at -2, -1, 0, +1, +2 etc. density – Record mAs – mAs change should be 5 to 15% per step Slide 47 Slide 48 Fuji FCRm Fuji FCRm Reproducibility & Image Mode Tracking – 4 cm acrylic with clinical technique – Position ion chamber in beam CNR Per Object Thickness – Record mAs and X – Repeat 3 times – CNR using clinical technique for 2 cm – Repeat in each mode (small, large, mag, mag, no grid) – Action Limits: Limits: – Repeat for 4 and 6 cm • Coefficient of variation for Exposure or mAs must not exceed 0.05 • No significant difference in exposure between small and large bucky when using similar grids – Action Limits: Limits: • CNR of 2 cm relative to 4 cm must be > 100% • CNR of 6 cm relative to 4 cm must be > 75% 12 Slide 49 Slide 50 Uniformity of Screen Speed & Detector Performance ScreenScreen-Film Mammo Detector Performance GE Fuji FCRm – Measure if OD’ OD’s are consistent FlatFlat-Field Uniformity Detector Calibration Lorad Selenia X X X X X Siemens Novation X CR Reader Scanner Performance X Dynamic Range X – FlatFlat-field uniformity Primary Erasure X – Detector calibration InterInter-Plate Consistency X – Pixel correction test Fischer Senoscan Same as ScreenScreen-Film (Screen Uniformity) – Check for artifacts FFDM 2000D, DS, Essential Geometric Distortion X X Pixel Correction – CR Reader Scanner Performance Slide 51 Ghosting X Slide 52 Fuji FCRm Dynamic Range – To confirm the dynamic range of reader and plate – Use 2 & 4 cm acrylic Fuji FCRm CR Reader Scanner Performance – Establish that IP reader & optics do not exhibit scan or print jitter – Technique for 6 cm – NonNon-grid exposure – View image to determine if 3 regions are visible – Position rulers in “T” formation and discernable – Expose & process – Action Limit: Image must have smooth borders free from jagged edges or defects 13 Slide 53 Slide 54 Fuji FCRm Fuji FCRm Primary Erasure (Additive & Multiplicative) – To assess the erasure performance of reader and InterInter-Plate Consistency plates – To confirm xx-ray absorption & SNR consistency – Additive: Shoot phantom, process, wait 1 minute, – Expose plates using clinical technique with 4cm reprocess, change S value to 10X original, acrylic inspect for visibility of phantom image – Multiplicative: Shoot & process phantom, shoot 4 – Record mAS (must be within +10%) cm acrylic on same cassette, process and inspect – Calculate SNR (must be within +15%) for visibility of phantom image Slide 55 Slide 56 GE 2000D, DS, & Essential FlatFlat-field uniformity – Test to ensure detector performance acceptable • Measures detector uniformity (signal & noise) • Measures bad pixels – System automatically calculates pass/fail Lorad Selenia Detector Calibration – Weekly – 4 cm acrylic block – no paddle – Built in Calibration software – Follow onon-screen instructions • Review image for artifacts • Repeat until you have 8 images • End calibration QAP • Calibration is performed automatically 14 Slide 57 Slide 58 Siemens Novation DR Detector Uniformity Siemens Novation Detector Calibration Test – Objective: to determine if system – Make exposure: 28 kV, 50 mAS, mAS, was correctly gain calibrated Mo/Mo – Select ROI 128 x 128 pixels – Must be done weekly – Record measurements in corners (technologist) and center – Follow onon-screen instructions – Pixel value inside each of 5 ROI’ ROI’s – Acquire 8 images shall not differ by more than 10% – Images should be free of artifacts of mean Slide 59 Slide 60 Siemens Novation DR Pixel Correction Test – Calibration mode Artifacts ScreenScreen-Film Mammo – Create new pixel map – Follow onon-screen instructions – Evaluate cassettes – Acquire image – Evaluate image is free from FFDM collimator or debris artifacts – System compiles new pixel map – System display results – must meet requirements – Evaluate detector – CR: Imaging Plates 15 Slide 61 Slide 62 Artifacts Artifacts Fuji FCRm GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation X X X X X Same as ScreenScreen-Film Expose Using Acrylic Slide 63 Slide 64 Artifacts Image Quality Artifact evaluation ScreenScreen-Film Mammo – ACR Phantom Scores – Optical Density & Contrast FFDM – ACR Phantom Scores • Pass/fail requirements differ by vendor – SignalSignal-toto-Noise Ratio (SNR) – ContrastContrast-toto-Noise Ratio (CNR) Mo/Mo Mo/Rh Mo/Rh Rh/Rh 16 Slide 65 Slide 66 Image Quality Fuji FCRm Image Quality GE 2000D, DS, Essential Fischer Senoscan X X Lorad Selenia ACR Phantom Imaging Siemens Novation GE & Fischer & Fuji Lorad & Siemens Fibers 4 5 Specks 3 4 Masses 3 4 Same as ScreenScreen-Film Manual Techniques Clinical Technique X CNR X X Slide 67 X X Partial X X Slide 68 GE 2000D ACR Phantom Imaging – Manual technique (Mo/Mo, 26 kVp, kVp, 125 mAS) mAS) – Score the processed image GE 2000D ContrastContrast-toto-Noise Test (CNR) – To examine consistency of CNR ratio measured over time – Use the raw image – + 20% of baseline – Acquisition workstation – Each monitor of the RWS Background ROI Mass ROI – Laser imager CNR = (Mean (Meanbackground - Meanmass)/SDbackground 17 Slide 69 Slide 70 GE DS & Essential Lorad Selenia Phantom Image Quality Phantom Image Quality – Select clinical exposure mode (i.e. AUTOAUTO-FILTER) – Print film – Manual technique: Rh/Rh, Rh/Rh, 29 kVp, kVp, 56 mAs – Measure background OD and density difference – MTF and CNR Measurement Plot on tech worksheets – Use IQST test tool • Background must be > 1.20 OD + 0.20 – Use builtbuilt-in software for image acquisition • DD must be > 0.40 + 0.05 – Score on each Soft Copy Workstation – Manual technique: Rh/Rh 30 kVp, kVp, 56 mAs • 5 fibers – Results are automatically displayed (pass/fail) • 4 speck groups – Same action limits as 2000D Slide 71 • 4 masses Slide 72 Siemens Novation DR Lorad & Siemens Phantom Image Quality – Position phantom 1 cm over chest wall edge – Select: 28 kV, AECAEC-Auto, Mo/Mo – Score on RWS or Film SNR and CNR Measurements – SNR at least > 40 – CNR should stay within ±15% of baseline – Fiber > 5 – Specks > 4 • Obtained during acceptance testing – Masses > 4 18 Slide 73 Slide 74 Fuji FCRm kVp ScreenScreen-Film Mammo ContrastContrast-toto-Noise Test (CNR) – Assure that actual kVp is – To examine consistency of CNR ratio measured over time – Use 4 cm acrylic & 0.2 mm Al • Accurate (+ (+ 5% of indicated) – Manual technique (Mo/Mo, 26 kVp, kVp, 125 mAs) mAs) • Reproducible (Coeff (Coeff.. Var. < 2%) – Calculate CNR using software FFDM – + 20% of baseline – Same Slide 75 Slide 76 HVL ScreenScreen-Film Mammo – Assure HVL is adequate to minimize patient dose – HVL shall meet FDA’ FDA’s Performance Standards FFDM Dose ScreenScreen-Film Mammo – Dose for single CC view of ACR phantom shall not exceed 3.0 mGy per exposure FFDM – Same – Same Note the lead sheet 19 Slide 77 Slide 78 Viewing & Viewing Conditions Radiation Output ScreenScreen-Film Mammo ScreenScreen-Film Mammo – System shall be capable of producing a – Luminance (> (> 3000 cd/m2) minimum output of 7.0 mGy/sec mGy/sec @ 28 kVp – Room illuminance (< 50 lux) lux) Mo/Mo over 3.0 second period – Viewbox cleaning FFDM – Same FFDM – S value confirmation – Same – Exposure Linearity Slide 79 Slide 80 Assessing Tech QC ScreenScreen-Film Mammo Film Processing ScreenScreen-Film Mammo – Measure optical densities – Assess site’ site’s QC program – Check that QC tests are properly performed – Independent checks of tech QC tests • Density difference, midmid-density, base+fog • Measures consistency FFDM – Manufacturers’ Manufacturers’ recommendations FFDM – Same – Some refer to printer manufacturers’ manufacturers’ recommendations – Typically identical to ACR SFM Manual 20 Slide 81 Slide 82 Film Processor QC Fuji FCRm Follow Printer Manufacturers QC If Not, Use Theirs GE 2000D, DS, Essential Fischer Senoscan X X X X X X Follow FFDMs’ FFDMs’ QC Film Processing Lorad Selenia X MidMid-Density Siemens Novation X Wet Processing Laser Printer Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Mid-Density Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Mid-Density 2.30 1.70 MD Step=12 2.20 1.60 2.10 1.50 2.00 1.40 1.90 1.30 1.20 1.80 Mid-Density Upper Control Limit = 1.48 1.10 1.70 Mid-Density Daily Value 1.00 Mid-Density Upper Control Limit = 2.06 1.60 Mid-Density Daily Value Operating Level = 1.33 Mid-Density Daily Value Mid-Density Lower Control Limit = 1.18 0.90 Mid-Density Daily Value Operating Level = 1.91 1.50 Mid-Density Lower Control Limit = 1.76 0.80 1.40 6 6 2 00 9, 20 06 er 11, 20 06 er 12, 20 ob 06 er 13, O ct 20 ob 06 er 16, O ct 20 ob 06 er 1 O 7, ct 20 ob 06 er 18, O ct 20 ob 06 er 19, O ct 20 ob 06 er 20, O ct 20 ob 06 er 23, O ct 20 ob 06 er 24, O ct 20 ob 06 er 25, O ct 20 ob 06 er 26, O c to 20 ber 06 27 O c to ,2 00 ber 6 30 ,2 00 6 er er 10, ob ob ob ob 6 Date Date Slide 83 O ct O ct O ct O ct 2 00 6, er ct O 6 6 2 00 2 00 2 00 2 00 3, 2, 5, 4, er er er er ob O ct ob ob ob ob ct ct ct ct O O O 6 6 ar06 23 -M ar2 4- 06 M ar -0 27 6 -M ar2 8- 06 M ar -0 29 6 -M ar06 30 -M ar3 1- 06 M ar -0 6 ar -0 ar06 21 -M ar -0 22 6 -M 20 -M 6 4 06 3 /1 ar -0 ar0 ar- M M ic e rv 1 7- 16 -M 1 5- se 6 6 M ar 1 0 -0 6 -M ar06 13 -M ar -0 14 6 -M ar- 06 9- 6 06 ar -0 M 8 -M 06 ar- ar0 ar -0 M M ar7- 3- 6 -M 2- 1 -M O 1.30 6 0.70 Slide 84 RWS QC Follow monitor manufacturers QC If Not, Use FFDM’ FFDM’s Fuji FCRm GE 2000D, DS, Essential X X X X X X AGFA IMPAX MA3000 Fischer Senoscan Lorad Selenia Siemens Novation – – – – – – – Clean monitor surface Measure display white & black Measure quality level Measure uniformity Calibrate displays View SMPTE pattern Viewing conditions GE Seno Advantage Use FFDM manufacturers’ manufacturers’s QC X X – – – – Viewing conditions check and setting Monitor calibration Image quality – SMPTE pattern Analysis of screen uniformity Siemens MammoReport Plus – – – – – – – Geometric distortion Reflection Luminance Response Luminance Uniformity Resolution Noise Overall Evaluation 21 Slide 85 Slide 86 New FDA Hot Topics Barco – – – – I-Guard Check Quality Level Display White Calibration Settings Check Tungsten w/ Silver or Rhodium – Calibration: Calibration: FDA requires all air kerma measuring Eizo – – – – instruments be calibrated to NISTNIST-traceable Pattern Check Luminance Check Grayscale Check Uniformity Check standard for at least one targettarget-filter combination in the mammo x-ray range… range…, Planar – – – – – Monitor Cleaning Viewing Conditions Check Monitor Calibration Check Image Quality – SMPTE Visual Inspection for Display Defects Slide 87 – Do not require such a calibration for W/Ag or W/Rh W/Rh – May require corrections for solid state instruments Slide 88 New FDA Hot Topics Tungsten w/ Silver or Rhodium – kVp Measurements: Measurements: • Scenario 1: 1: If using kVp meter certified for W/Ag or W/Rh W/Rh and a mammo calibrated ion chamber, then ok • Scenario 2: 2: If not, you need to include a correction factor in kVp measurements obtained from kVp meter manufacturer, or, by working with service engineer and comparing measurements • Scenario 3: 3: If not W certified and using a separate New FDA Hot Topics Tungsten w/ Silver or Rhodium – HVL Measurements: Measurements: • Scenario 1: 1: If using integrated solidsolid-state instrument that is manufactured for W/Ag or W/Rh W/Rh,, then ok. • Scenario 2: If using integrated solidsolid-state instrument that is not manufactured for W/Ag or W/Rh W/Rh,, then obtain necessary correction factors. solidsolid-state probe, then follow scenario 2. 22 Slide 89 Slide 90 New FDA Hot Topics New FDA Hot Topics Use of Foam Cushions on FFDM QC testing for Printers – FDA recommends using pads for QC – Follow FFDM manufacturer’ manufacturer’s QC Manual if used clinically – Then, follow printer’ printer’s QC manual – Required if manufacturer’ manufacturer’s QC – To determine the appropriate testing manuals specifies the pad be present during QC Slide 91 Slide 92 ACR FFDM QC Manual Project The current ACR QC Manual was written for screenscreen-film – Most does not apply to digital Subcommittee on QA is writing new manual – Chair - Eric Berns, Berns, Ph.D. – Members – Radiologists, technologists, medical physicists, MITA reps ACR FFDM QC Manual Project QC tests – Will apply to all manufacturers and models – Fewer tests and written to be “tech friendly” friendly” – New phantom to be more applicable to digital (but usable with screenscreen-film) – When ready, draft will be sent to manufacturers for their input before it is final Standardize QC tests, performance criteria and When final, ACR will apply for FDA alternative frequencies across all systems standard Clinical section specific to digital issues Hope to have draft completed by end of 2008 23 Slide 93 Slide 94 Suggestions Archive QC images periodically on a CD Perform flatflat-field imaging before phantom Troubleshooting QC Repeat test – Clean plexiglass & move slightly Questions to ask yourself: – Are you using the correct image? Raw vs. Processed? – Are the ROI’ ROI’s in the correct location? imaging tests Perform testing in the morning Have a system for monitoring the “Auto ReRe-boot system Print/Auto Push” Push” functions Call service – ReRe-calibration solves most image quality problems Slide 95 Slide 96 Key Take Home Points Obtain relevant handshands-on training Must perform manufacturer specific QC tests Artifacts – most problems can be seen here ReRe-booting and/or rere-calibrating fixes most problems Laser Printer – Dmax at least 3.5 OD – MidMid-density about 1.5 OD Key Take Home Points Review workstation monitors – look at the clinical images! – Do they match? – Appropriate dark and light levels Be aware of separate QC for printers & monitors Know, and be involved with, your site QC program 24 Slide 97 Slide 98 SAM Questions Digital Mammography QC In digital mammography, who mandates the pass/fail criteria for site QC? 20% 1. The American College of Radiology 20% 2. The FDA 20% 3. The FFDM unit manufacturer 20% 4. NEMA 20% 5. MQSA Eric Berns, Berns, PhD 10 Slide 99 Slide 100 Answer: 3 - The FFDM unit manufacturer References MQSA Regulations 900.12(e)(6) http://www.fda.gov/CDRH/MAMMOGRA PHY/frmamcom2.html#s90012 Answer 3: The FFDM unit manufacturer Explanation “the quality assurance program shall be substantially the same as the quality assurance program recommended by the image receptor manufacturer, except that the minimum allowable dose for screenscreenfilm systems in this section” section” 25 Slide 101 Slide 102 How do you accredit a Fuji CR Mammography system which uses both CR and screenscreen-film on the same x-ray system? 25% 1. Answer: 2 – As 2 mammography units References As 1 mammography unit? http://www.acr.org/accreditation/mammography/ mammo_faq/mammo_faq_mamac.aspx#8.0.1 As 2 mammography units? You cannot use CR and film on the same unit. 25% 4. CR is exempt from accreditation 25% 2. 25% 3. 10 Slide 103 Slide 104 What are the minimum passing ACR phantom scores for the Lorad Selenia for weekly QC? Answer: 2 - As 2 mammography units Explanation As of November 15, 2006 facilities using both screenscreen-film and CR on the same mammography units must accredit these 2 systems as 2 separate units. 20% 1. 20% 2. 20% 3. 20% 4. 20% 5. 5 Fibers, 4 Speck Groups, 3 Masses 4 Fibers, 3 Speck Groups, 3 Masses 5 Fibers, 4 Speck Groups, 4 Masses 4 Fibers, 4 Speck Groups, 3 Masses 4 Fibers, 4 Speck Groups, 4 Masses 10 26 Slide 105 Slide 106 Answer 3: 5 Fibers, 4 Speck Groups, 4 Masses Answer 3: 5 Fibers, 4 Speck Groups, 4 Masses References Explanation Hologic. Hologic. Lorad Selenia Quality Control Manual MANMAN-00093, Revision 003, Bedford (MA): Hologic, Hologic, 2005 The Lorad Selenia QC manual states that the ACR Phantom minimum passing scores are as follows: 5 Fibers 4 Speck Groups 4 Masses Slide 107 Slide 108 To meet the FDA requirement for continuing experience, how many mammography facilities and mammography unit surveys must be performed within the previous 24 months? 20% 1. 20% 2. 20% 3. 20% 4. 20% 5. 6 Facilities and 2 mammography units 4 Facilities and 12 mammography units 2 Facilities and 6 mammography units 1 Facilities and 6 mammography units 2 Facilities and 12 mammography units 10 Answer 3: 2 Facilities and 6 Mammography Units References http://www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/ START.HTM 900.12(a)(3)(iii)(B): Continuing experience. Following the second anniversary date of the end of the calendar quarter in which the requirements of paragraphs (a)(3)(i) and (a)(3)(ii) of this section were completed or of April 28, 1999, whichever is later, the medical physicist shall have surveyed at least two mammography facilities and a total of at least six mammography units during the 24 months immediately preceding the date of the facility’ facility’s annual MQSA inspection or the last day of the calendar quarter preceding the inspection or any date in between the two. The facility shall choose one of these dates to determine the 2424-month period. No more than one survey of a specific facility within a 1010-month period or a specific unit within a period of 60 days can be counted towards this requirement. 27 Slide 109 Answer 3: 2 Facilities and 6 Mammography Units Explanation The FDA requires: 2 Facilities 6 Mammography Units Within past 24 months 28