AAPM Annual Meeting 2006 Abstract ID: 26-5971-9361 Digital Fluoroscopic Imaging: Acquisition, Processing and Display Barry Belanger, PhD Outline X-Ray Clinical Imaging Applications: Variety & Key Considerations Performance Metrics for Clinical Imaging Modern X-Ray Imaging System Design: Acquisition, Processing and Display Key Points to Consider in the Purchasing Process Disclosure: GE Healthcare Employee Implementing Quality Assurance & Pitfalls to Avoid AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Fluoro X-ray Systems under discussion AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger X-Ray Clinical Imaging Applications: Variety & Key Considerations: Anatomy/object of interest: radiographic attenuation, size, motion: Cardiovascular: R&F “fluoro” Bone: spine: vertebroplasty, needle placement GI tract with barium contrast Blood vessel or Heart chamber with iodine contrast Guidewire, catheter, stent • Background anatomy: attenuation variations, motion Thorax: large attenuation variations, rapid cardiac motion Abdomen: more uniform attenuation, less motion Cranium: large attenuation variations, little motion • Attenuation range: from pediatric to obese adult. AP thorax: 6cm (ped) to 32cm (large adult) and growing, equivalent to ~9HVLs of ~3cm, ~500:1 difference in attenuation. LAT thorax presents greater range and challenge Range of Needs and Conditions is huge: >> Need for Flexibility, Automation, Preset Protocols AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 1 Performance Metrics for Clinical Imaging: Performance Metrics: Clinical Image Quality & Dose Optimization Problem Statement: Benefit •Imaging Dose Efficiency: Cost Clinical Objective = Risk ~ = Clinical Image Quality Dose = Imaging Dose Efficiency Objectives: • Maximize Imaging Dose Efficiency: Best clinical image quality for any dose level clinician chooses to use, tailored to intended application. • Provide range of Dose/IQ selections to operator to fit application needs and preferences of clinicians. • Provide Dose Readout/Feedback: clinician awareness for decision making. • Automate the process as much as possible. X-Ray Source Efficiency Image Detector Efficiency •Dynamic Range •Temporal Response Caveat: • Dose = Dose Rate x Time, alternately Dose/Exposure x Number of Exposures. Need to consider impact of Image Quality on fluoro time, number of exposures required to achieve clinical goals •Image Artifacts A Comprehensive Set of Relevant IQ Measures AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Performance Metrics for Clinical Imaging: Requirement Clinical Benefits X-Ray Source Efficiency: High continuous power capability, range of spectral filters • High Dose Efficiency • High fluoro penetration capability for large patients Image Detector Efficiency (DQE): High XRay Conversion Efficiency, Low Readout Noise • High Image Quality: Object detectability • High Dose Efficiency Consistent with ALARA (As Low As Reasonably Achievable) Dose Principle AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Performance Metrics: XRT: Spectral Filtration Spectrum after 25 cm Water Spectrum at patient entrance 4.00E+02 7.00E+04 3.50E+02 3.00E+02 5.00E+04 Photon Nr Photon Nr 6.00E+04 4.00E+04 3.00E+04 2.00E+04 • Visualize contrast differences from thin to dense anatomy. • No Blooming in thin anatomy: robust, forgiving to difficulties with positioning, collimation, contour filters, etc. Temporal Response: Fast Readout, low lag, range of frame rates & exposure times •High definition of moving objects •Good rendering of dynamic events A Comprehensive Set of Relevant IQ & Dose Efficiency Measures AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 2.00E+02 1.50E+02 1.00E+02 1.00E+04 Wide Dynamic Range: High min/max signal range from image detector plus specialized image processing to present information effectively on display. 2.50E+02 5.00E+01 0.00E+00 0.00E+00 15 25 35 45 55 65 75 85 15 25 35 Energy (keV) regular filtration 0.2 mm Cu additional filtration 45 55 65 75 85 Energy (keV) regular filtration 0.2 mmCu additional filtration • Softer radiation does not contribute significantly to image • Spectral filtration eliminates this radiation before it reaches the patient • Requires higher XRT power to be effective AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 2 Performance Metrics: Detector: Importance of DQE DQE=1.0 DQE = 0.5 DQE = 0.25 CNR = 5 CNR = 3.5 CNR = 2.5 Performance Metrics: Temporal Response: Value Pediatric Imaging Protocol 4 msec max pulse width 30 Frames/Sec Options enabled by higher DQE: Better Image, Same Dose Same Image, Lower Dose DQE (Detective Quantum Efficiency) is judged the best index of object detectability in contrast & dose-limited imaging, as in real clinical imaging. AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger X-Ray Source Assembly: X-Ray Tube & Collimator Modern X-Ray Imaging System Design • XRT/Source Assembly • Image Detector • Automatic Exposure Control X-Ray Tube: • 3200 Watt continuous fluoro capability • Metal envelope for reduced off focus radiation • 0.3, 0.6, 1.0 Focal Spot (0.3 for DSA) • Imaging modes: fluoro, dynamic record, Digital Spot, DSA, Rotational/3D Collimator with variable Spectral Filtration: • Image processing & Display • • Procedure Protocol Driven Customization and Control AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger • • Contour Filters 0.1, 0.2, 0.3, 0.6 & 0.9mm Copper filters Rectangular Blades AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 3 Detectors: Image Intensifier/TV >> Flat Panel DFP Benefits: “Birthright” Camera •200u Pixel 40x40cm Flat Panel •State of the Art 40cm Image Intensifier + 1k x 1k CCD • Brightness & MTF Uniformity Across Entire Image • No Veiling Glare • No Geometric Distortion • No Earth Field Distortion Image Intensifier Chain Image Optical Iris Intensifier Detector DQE: Flat Panel vs II/CCD Analog Processing & A/D Detective Quantum Efficiency 0.8 Innova 4100 0.7 II/CCD 0.6 Design-Dependent: • 5-10x Improvement in Dynamic Range: depends on imaging mode and DFP & Electronics Design • Radiographic DQE: Typically better • Fluoro DQE: Depends on Detector & Electronics Design Digital Flat Panel Detector AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger DQE 0.5 0.4 0.3 41cm FOV 20cm FOV 0.2 0.1 0 0 0.5 1 1.5 2 2.5 Spatial Frequency (cycles/mm) 40cm Flat Panel has Significant Advantage in DQE AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Detector DQE: X-Ray Detector/Imaging Chain Detector: DFP: Binning in Larger FOVs Innova 2100 Fluoro DQE @ 1 uR/Fr Native 200 x 200 micron photodiodes Summed 400 x 400 micron pixel (in blue) Binning Principle: Scan • Scan Adjacent Lines in Pairs • Digitally add adjacent columns in pairs amp amp amp • 4X signal/pixel: better SNR • Limiting Spatial Resolution 1.25 instead of 2.5 lp/mm (comparable to II/Video) • Good DQE at lower doses used in LFOV imaging, exceeding II/Video chain • Reduced image processing, storage and display requirements. amp Performance Implications: A/D A/D A/D A/D + + Binning Enables LFOV Imaging at Required Rates AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 0.80 To understand how well an image chain works we must consider: The DQE curve of the image chain The spatial frequency content of the object being imaged DQE(f) at fluoro level radiation exposure of 8 nGy/Fr (1 uR/Fr) 0.60 0.40 0.20 0.00 0.00 0.50 1.00 1.50 2.00 2.50 Spatial Frequency (cyc/mm) Spatial frequency content of a stent. Note that there is progressively less signal at higher frequencies, particularly beyond 2 lp/mm. • The spectrum of the stent extends over a wide range of spatial frequencies. • Therefore, the DQE response over this same range of spatial frequencies is important to determining the quality of the image. • Response at a single spatial frequency, e.g., spatial resolution limit, doesn’t tell the story. AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 4 Detector Dynamic Range: DFP Versus II/TV II/Video Chain Transfer Function Digital Detector Transfer Function Modern X-Ray Imaging System Design Saturation Limit Skin Line Lung Soft Tissue Noise Floor Spine, Diaphragm • XRT/Source Assembly Detector Signal Video Signal Saturation Limit • Image Detector • Automatic Exposure Control • Imaging modes: fluoro, dynamic record, Digital Spot, DSA, Rotational/3D Noise Floor X-Ray Intensity X-Ray Intensity • Image processing & Display • DFP Provides 5-10x Improvement in Dynamic Range of Captured X-ray Information. • Display Dynamic Range Must Be Addressed Independently (later in this presentation) • Higher Dynamic Range Can Reduce Retake Rates (no burnout/blackout) and is More Forgiving to Errors, Hence it is Dose Conservative • Procedure Protocol Driven Customization and Control AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Automatic Exposure Control Systems Automatic Exposure Control Systems: “Trajectory” Table-Based Image Data from Detector: Possibilities Average Area-Location Weighted Histogram Detector Patient Auto Ex XRT Generator Exposure Control Algorithm: Possibilities Automatic selection of kVp, mA, pulse width, focal spot, spectral filter & detector dose as a function of image signal from detector and effective patient attenuation (calculated from xray technique, SID, other system parameters) Desirable Objectives: • Maximize Imaging Dose Efficiency: Best clinical image quality for any dose level clinician chooses to use, tailored to intended application, capitalizing on component characteristics. • Provide range of Dose/IQ selections to operator to fit application needs and preferences of clinicians. Exposure • Provide Dose Readout/Feedback: clinician awareness for On decision making. Off • Automate the process as much as possible. Skin Dose DAP Rate Total Total Total AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Effective Brightness is calculated from digital image Detector Patient Auto Ex XRT Generator kVp mA Pulse width Spectral Filter Focal spot Algorithm Calculates Effective Patient Thickness (EPT), based on: •Effective Brightness •kVp, mA, Pulse Width •Spectral Filter •FOV, SID, Grid In/Out Using new EPT, Determines Technique & Spectral Filter for next exposure from trajectory table, and sends to generator. Fluoro: 30 PPS, Normal Detail, 'Smart IQ' Trajectory EPT Technique Cm kVp pw mA pk Filt 15 73 8.0 8.3 0.2 20 73 8.0 27.4 0.2 25 77 11.8 36.6 0.1 30 95 12.4 34.3 0.1 Pat Dose R/min 0.41 1.29 4.16 10.00 Det Dose mR/min 3.6 3.6 3.6 3.6 Benefits: Flexibility, Optimization, Upgradeability AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 5 Automatic Exposure Control Systems: “Trajectory” Table-Based: Trajectory Design Process Automatic Exposure Control Systems: “Trajectory” TableBased: Typical Optimization Criteria Object Contrast: Difference in x-ray intensity due to the presence of the stent or other device, iodine in artery, barium in colon, etc. Background Noise: Normal statistical variation in background intensity due to finite number of x-rays used to image Contrast (more is better) = CNR Image Quality = Noise (less is better) Clinical Objectives Anatomy, Motion System Capabilities Regulatory Requirements Cost Physics Model Auto Exp Control Trajectories Patient Skin Dose Significant dose savings on small patients & shallow angles • Provide several levels of CNR (and therefore dose) to Operator: Fluoro: 30 PPS, Normal Detail, RDL Trajectory EPT Fluoro: 30 PPS, Low Technique Pat Dose Det Dose Detail, IQ Plus Trajectory CmEPT pw Detail,mA Filt R/min Technique Pat Dose mR/min Det Dose Cine: 30kVp PPS, Normal IQ pk Plus Trajectory pw mA pk Filt R/min mR/min 15 CmEPT 79 kVp 9.2 0.2 Trajectory 0.60 Technique Pat Dose6.0 Det Dose Fluoro: 15 PPS,8.0 Normal Detail, IQ Plus 20 15 Cm EPT 79 8.0 34.0 0.2 2.050.60 6.9 6.0 79 kVp 8.0 pw 9.2 mA pk 0.2 Filt R/min mR/min Technique Pat Dose Det Dose Fluoro: 30 PPS, Normal Detail, 'Smart IQ' Trajectory 25 20 15 Cm 92 EPT 10.98.0 8.0 38.4 0.1 9.94 9.4 79 79 kVp 34.0 0.2 2.05 6.9 6.0 pw mR/min 9.2mA pk 0.2 Filt 0.60R/min Technique Pat Dose Det Dose 30 25 20 120 12.5 32.1 0.0 10.00 7.1 92 10.9 38.4 0.1 9.94 9.4 79 8.0 34.0 0.2 2.05 R/min6.9 6.0 mR/min 15 Cm 79 kVp 8.0 pw 9.2mA pk 0.2 Filt 0.60 30 25 20 120 12.5 32.1 0.0 10.00 7.1 92 10.9 38.4 0.1 9.94 79 73 8.0 8.0 34.0 8.3 0.2 0.2 2.050.41 9.4 6.9 3.6 15 30 25 20 120 92 7312.510.9 8.032.138.427.40.0 0.1 0.210.009.941.29 7.1 9.4 3.6 30 25 120 77 12.511.8 32.136.6 0.0 0.1 10.004.16 7.1 3.6 30 95 12.4 34.3 0.1 10.00 3.6 Optimization Algorithm = Design Strategy: One of several possibilities: • Maintain near-constant CNR over range of patient thickness (within limits) Operator chooses level of IQ required for procedure: via protocols, tableside control, initial system setup (trajectory family) • Minimize patient dose required to achieve selected CNR by optimizing x-ray technique and filtration. • Pediatric: Limit Pulse width to 4 mSec max in fluoro and digital cardiac record Consistent with ALARA (As Low As Reasonably Achievable) Dose Principle AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Automatic Exposure Control Systems: “Trajectory” Table-Based: Sample Trajectories for One Operating Mode of Pediatric Angiography Automatic Exposure Control: “Trajectory” Table-Based: Trajectory “Family” Options Example 0.6 8 0.5 PW (ms) 100 80 60 40 0.3 4 10 20 thickness(cm) 30 40 0 0 10 20 thickness(cm ) 30 40 0 10 20 thickness(cm) 30 40 kVp (upper left), Pulse Width (upper middle), Spectral Filter thickness (upper right), mA (lower left) and patient entrance dose as a function of EPT in cm of PMMA for one fluoro AutoEx trajectory. 300 1000 regdose (mGy/min) 250 PeakmA Target Values: • kVp: 60-80 kVp for Iodine imaging • Pulse Width: 4 mSec max for peds 200 150 100 50 0 0 10 20 thickness (cm) Normal : 91.2 Low : 32.4 Normal : 293 Low : 146 Normal : 2.0 Low : 1.0 • Mode Selection 2 Normal : 25.2 Low : 10.4 Normal : 91.2 Low : 35.2 Normal : 293 Low : 146 Normal : 2.0 Low : 1.0 • Mode Selection 3 Normal : 14.9 Low : 5.7 Normal : 91.2 Low : 32.4 Normal : 293 Low : 146 Normal : 2.0 Low : 1.0 • Mode Selection 4 Normal : 18.7 Low : 7.7 Normal : 32.4 Low : 14.3 Normal : 293 Low : 146 Normal : 1.0 Low : 0.5 • Mode Selection 5 Normal : 9.7 Low : 4.0 Normal : 32.4 Low : 14.3 Normal : 293 Low : 146 Normal : 1.0 Low : 0.5 0.1 0 0 Cardiac Normal : 21.8 Low : 9.4 Dyn 5 fps DSA 0.2 2 20 0 Fluoro 30fps • Mode Selection 1 0.4 6 30 40 Note log scale Loweer Dose Patient Thickness Control Range: • From 0 cm (bare table) to >40cm of PMMA effective patient thickness (EPT) 10 120 Spectral filter Within limits of: •XR Tube: filament, power, etc. •Generator: Power, control ranges •Regulations Typical Dose Levels: Fluoro (mGy/min), Cardiac & Innova Chase (µGy/fr) and DSA (mGy/fr) 140 kV Independent control of: • kVp • mA • Pulse Width (primary & Grid) • Spectral Filter • Focal Spot 100 Test conditions: 10 1 0.1 0 5 10 15 20 0.01 thickness(cm) AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 25 30 35 40 • Patient Entrance Dose • 20 cm FOV • 20 cm PMMA phantom • Reference: Operator Manual • Fluoro Levels : 5:1 range in dose rate, @ 30 fps: (20:1 with 15 and 7.5 fps) • Mode 2 and Mode 5 Trajectories have 5R/min limit in Fluoro Low Mode. • Cardiac Record Levels : 6:1 range in dose/frame • Dynamic 5 fps: 2:1 range in dose/frame • DSA: 4:1 range in dose/frame AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 6 Automatic Exposure Control: Options for Dose Change with Frame Rate Two Options Seen on Current Systems: 1. Balanced Dose/IQ : Follows “Aufrichtig” or “Square Root” Scale: Maintains static object detectability. 2. Minimize Dose: Dose rate drops directly with fluoro frame rate. Automatic Exposure Control: Options for Dose Change with Field of View: DFP vs II Output Phosphor II: As FOV is reduced, II sensitivity decreases, because a smaller input area of the scintillator is being projected to the same output phosphor area. Required II entrance dose goes up inversely with area, or… Compensations & compromises must be made (video camera iris), and … Electron Trajectories Smaller FOV’s may operate at higher than optimal entrance doses Scintillator Frame Rate 30 15.0 7.5 Balanced Dose/IQ 100% 75% 50% Minimize Frame Rate Dose 30 100% 15.0 50% 7.5 25% Balanced Dose/IQ 10.4 7.8 5.2 Minimize Dose 10.4 5.2 2.6 Data for GE Innova 2100 in mGy/min Scintillator FOV: 40 FOV: 17 Photodiode Array FOV: 40 FOV: 17 DFP: Flat panel Sensitivity is independent of FOV, because it depends only on the size of the detector elements, which is fixed: Patient Patient X-Rays X-Rays Free to choose the optimal dose for each FOV. Best choice may depend upon application and user: Select lowest frame rate tolerated by user for application Select lowest usable dose at selected frame rate No compromises with FP AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Automatic Exposure Control: Options for Dose Change with Field of View Automatic Exposure Control: Options for Dose Change with Field of View Illustration of FOV Change without Dose Change 12 cm Illustration of FOV Change with Dose Change 30 cm 20 cm 16 cm 12 cm Dose Increase from 30 cm 16 cm 20 cm 6x 30 cm Inc Dose ea 1/Ar with rease Dose Increase with 1/Length Between Dose Con sta 2.5x the two nt 1x How much does IQ improve with magnification alone? AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger How much does IQ improve with magnification + dose? AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 7 Automatic Exposure Control: Options for Dose Change with Field of View Modern X-Ray Imaging System Design DoseRateChangewithFOV 35.0 mGy/min 30.0 25.0 NoChange 20.0 1/Area Change 15.0 1/Length Change 10.0 Actual Change • XRT/Source Assembly • Image Detector 5.0 0.0 0 10 20 30 40 FOV (length in cm) Different Philosophies: 1/Area: Assume that smaller FOV indicates need for better CNR and keep perceived Contrast/Noise Constant No Change: Accommodate Clinicians’ Vision Only 1/Length: Split the difference between the two above • Automatic Exposure Control • Imaging modes: Fluoro, Dynamic Record, Digital Spot, DSA, Subtracted Bolus Chase, Rotational Angio/3D/CT • Image processing & Display • Procedure Protocol Driven Customization and Control GE Philosophy is Close to 1/Length. Most manufacturers change less than 1/Area with FOV AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Rotational Angio/3D/CT on Flat Panel Angiography Systems Rotational Angio/3D/CT on Flat Panel Angiography Systems MIP Volume Rendering Sagittal cross-section Vascular, bone and soft tissue 3D imaging in the intervention suite 3D Images Cross-sections Rotational Acquisition Automatic 3D Recon 3D Display In Several Formats AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Oblique cross-section Axial cross-section Coronal cross-section Both 3D views and cross-sections Can be automatically created at 3D volume loading AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 8 Rotational Angio/3D/CT on Flat Panel Angiography Systems Oblique cross-section Rotational Angio/3D/CT on Flat Panel Angiography Systems Acquisition and Reconstruction Protocols Axial cross-section • • • • • • “Classical” “3D” Mode ~5 second acquisition One or two spins (for subtracted) ~200o at 35-40o/s ~150 frames 3D recon in 2563 and/or 5123 Display: MIP, VR, MPVR, etc. Vascular, bone and soft-tissue 3D imaging Sagittal cross-section Coronal cross-section Full screen axial cross-section Both 3D views and cross-sections Can be automatically created at 3D volume loading AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger “CT” Mode • • • • • • ~5, 10 & 20 second acquisition One single non-subtracted spin ~200o at ~40, 20 and 10o/s ~150, 300 or 600 frames 3D recon in 2563 or 5123 Display: MPVR ( + MIP, VR, etc. ) Increased contrast senstivity for bone and soft-tissue imaging Evolution toward better “CT” Capabilities: Longer effective exposure times than CT in some cases Higher spatial resolution capability than CT Low contrast sensitivity less than CT AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Rotational Angio/3D/CT on Flat Panel Angiography Systems Modern X-Ray Imaging System Design Typical Reconstruction Algorithm Elements and Attributes: • Full cone-beam tomographic reconstruction • Based on Feldkamp algorithm • Compensation/calibration for “Open Gantry” dynamic mechanical behavior • Processing of truncated views • Beam hardening correction • Ring artefacts removal • Parallel implementation • Reconstruction time depends primarily on matrix, e.g., 2563, 5123. • XRT/Source Assembly • Image Detector • Automatic Exposure Control • Imaging modes: fluoro, dynamic record, Digital Spot, DSA, Rotational/3D • Image processing & Display • Procedure Protocol Driven Customization and Control Your CT knowledge should be transferrable! AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 9 Image Processing: Dynamic Range Management Digital Spot Example: Simple Display LUT Digital detector dynamic range covers the entire range of exposure levels; more than can be displayed without processing! Image Brightness Classic Contrast/Latitude Tradeoff Image Processing: Dynamic Range Management Desired Behavior Breaking the Contrast/Latitude Constraint Image Processing: Dynamic Range Management Digital Spot Example: Multiresolution Processing Transform •Select Weights Raw Input Lung, Lung, Skin Line Line Skin Lung, Skin Line Skin Skin Line Line Simple Display LUT Can Mimic … or many different looks suchLook as a of …or a Bone Look Film...look soft-tissue But one LUT can’t do it all at once. AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Image Processing: Dynamic Range Management Realtime Processing in Peripheral Runoff Sequence Bolus Chase: Unsubtracted •Spatially filter A1 Soft Tissue Dense Soft Tissue Spine, Spine, Spine, Breast Spine, Diaphragm Diaphragm Diaphragm Diaphragm X-ray Detected AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Decompose Tissue Soft Typical Breast Mass Soft Tissue Reconstruct •No contour/bolus filters used •Weighted Addition •Manually pan table to follow contrast bolus. Processed Image •AutoEx automatically adjusts technique on the fly. •DRM handles wide dynamic range of image data from detector A2 ... An Wide Dynamic Range + DRM: Get all the desired information in one image AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger • More diagnostic info in one run • Robust: fewer retakes? AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 10 Image Processing: Fluoro Noise Reduction Advanced Fluoro Noise Reduction (FNR) Image Processing: FNR Motion Compensation Impact of Motion Compensation at 30 fps What is it? 16 frame recursive averaging 16 frame recursive averaging with motion compensation no motion compensation none •Temporal averaging of image data to reduce quantum noise (fluoroonly application) What is advanced? •Motion compensation is integrated into the algorithm: reduces lag effects in the presence of anatomical motion •Level of filtering is matched to frame-rate Results: •Significant improvement in clincial image quality > opportunity to reduce dose. •Sufficient noise reduction to allow (in certain cases) use of a lastimage hold in lieu of acquiring a spot image AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Modern X-Ray Imaging System Design Motion Compensation is a Clear Benefit AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Procedure Protocol-Driven Customization and Control Auto Exposure Control Feedback signal •Average •Location Weighted •Histogram Protocol Selection: •Acquisition Mode •Low/Norm/High Dose •Trajectory “Family” •Frame Rate •FOV Algorithm Positioner: •SID •Grid In/Out • XRT/Source Assembly • Image Detector detector • Automatic Exposure Control • Imaging modes: fluoro, dynamic record, Digital Spot, DSA, Rotational/3D • Image processing & Display FOV kVp, mA, PW, spectral filter, focal spot patient table collimation spectral filter • Procedure Protocol Driven Customization and Control X-ray tube kVp, mAs, Focal spot X-ray generator Acq Mode Dose Traj FR, FOV DRM Contrast FNR Strength Edge Enhance Operator Controls: •Low/Norm/High Dose •Frame Rate •Fluoro/Record Mode Realtime Digital Image Processing: DRM FNR Edge Enhance Image Processing Can Use Protocol & Exposure Information for Optimization: Beware of Interactions when Testing AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 11 Procedure Protocol-Driven Customization and Control Key Points to Consider in the Purchasing Process System Features for Dose Reduction Pre-defined Procedure Protocols (fit dose & technique to procedure) Automatic Spectral Filters Pulsed, Variable Frame Rate Fluoro (linked to Procedure Protocols) Virtual Collimation, Virtual Panning Adjustable Contour (Wedge) Filters in Collimator Last Image Hold Fluoro Loop Replay & Store (with DICOM & DVD Record) • Applications breadth required by clinical department, both today and looking out ~5 years. • Procedure/protocol capability/flexibility/performance provided by equipment • Imaging/Dose performance provided by equipment Auto or Manual Grid removal Patient Contouring (Automatic Minimization of patient-detector distance) Dose reporting The More Automatic/Programmable, the Better AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger Implementing Quality Assurance & Pitfalls to Avoid • Prioritize by key clinical procedural needs. • Work with Supplier’s Field Service Engineer and Applications Specialists through installation and customer training to fully understand capabilities and optimize to clinical needs and preferences. Thank You! • Perform full verification test set at installation •Consider “standardized” procedures: •Manufacturer’s •IEC 61223-3-1 •SCA&I?/NEMA XR-21 (Cardiac) • Perform key functional & performance checks at periodic intervals. AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger AAPM 2006: Digital Fluoroscopic Imaging: Acquisition, Processing and Display. Abstract ID: 26-5971-9361. Belanger 12