Quality Assurance in Stereotactic diosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Shepard Ph.D. Ph D Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern Medical Center Dallas, TX Quality Assurance in Linac SRS/SBRT Outline Mechanical M h i l aspects • Linac • Frames eam data acquisition g of TP system y ommissioning nd-to-end evaluation maging and Image Fusion rameless l Radiosurgery di References and Guidelines Can we hit the target? n we putt the th dose d where h we wantt it? How accurate is radiosurgery? actic Radiosurgery, AAPM Report No. 54, 1995 er sources: MRI Distortion Image Fusion Relocatable frames Dosimetric Frames & CT Isocentric Accuracy: The Winston-Lutz Test Mechanical Uncertainties he projection of the all centered within the field? the projection of the all centered within the field? od results ≤ 0.5 mm Isocentric Accuracy: The Winston-Lutz Test daily Lutz test is extremely important because: he mechanical isocenter an shift over time he AMC board in Varian ouches can fail he cone or MLC may not be epositioned perfectly after ervice ore (right) and After (left) tively simple couch adjustment Aft After Before f A Lutz test with the MLC is also important because: The Cone Cone-based based Lutz test does not ell you anything about the mechanical isocenter of the MLC The MLC may not be repositioned perfectly after service Lutz test with 12 x 12 mm MLC field End-to-end localization evaluation Structure C li d Cylinder Cube Cone AP 00 0.0 20.0 -35.0 Phantom Specifications LAT VERT 00 0.0 30 0 30.0 -17.0 40.0 -20.0 40.0 iPlan Stereotactic Coordinates AP LAT VERT 10 1.0 04 0.4 30 8 30.8 20.8 -17.1 42.4 -34.6 -19.7 40.8 End-to-End Localization Accuracy End-to-End Localization Accuracy (Surely my vendor has checked this) Verification of MLC shapes and isocenter System Accuracy Simulate the entire procedures: Scan, target, plan, deliver Phantom with fil holder film h ld Pin denotes isocenter Resulting film provides measure of targeting accuracy … Offset from intended target … as well as falloff for a multiple arc delivery 1 0.8 0.6 0.4 02 0.2 0 -0.5 0 0.5 1 Phantom Hidden Target g Test scan,, p plan,, localize assess Lucy Phantom Imaging Uncertainties se CT for geometric accuracy se MR for target delineation RI contains distortions which pede direct correlation with CT ta at the level required for SRS SRS” ereotactic Radiosurgery – AAPM Report No. 54 ther References Sumanaweera, S JR Adler, Adl S Napel, N l ett al., l Characterization Ch t i ti off atial distortion in magnet resonance imaging and its implications 1.8 ± 0.5 mm shift of MR images relative to CT and d delivered d li d dose d Shifts occur in the f frequency encoding di direction Due to susceptibility artifacts between the phantom and fiducial markers of the Leksell localization box anabe, GM Perera, RB Mooij, “Image distortion in MRI-based er gel dosimetry of Gamma Knife stereotactic radiosurgery equency Encoding = L/R Frequency Encoding = A/P anabe, GM Perera, RB Mooij, “Image distortion in MRI-based er gel dosimetry of Gamma Knife stereotactic radiosurgery at do we do about MR spatial distortion? Use Image Fusion Fusion Verification ndard Diode Dosimetric Uncertainty Small field measurements can be challenging; Diodes and small ion chambers are well suited to SRS dosimetry, but their characteristics / response must be well understood. Stereotactic Diode Small field depth dose show familiar trends 0 6 12 18 24 30 36 42 60 80 100 50 100 150 200 Depth (mm) 250 300 350 400 Similar machines have similar characteristics 36A 12A 36 UNMC 12 0 50 100 150 200 Depth (cm) 250 300 350 400 Penumbra: Cones versus MMLC 3.5 Cones MMLC 3 2.5 2 1.5 1 0.5 0 0 20 40 60 80 100 Diode Warnings!!! Diodes exhibit enough energy dependence that atios between large and small field measurements re inaccurate at the level required for radiosurgery easure output factors ≤ 2x2 cm2 using diode ≥ 2x2 and ≤ 4x4 cm2 using diode & chamber ≥ 4x4 cm2 using chamber Diode response will drift over time Re-measure reference between each chance in ield size Reference diode output to intermediate field size utput actor = Reading (FS)diode Reading (Ref’)diode X Reading (Ref (Ref’) )IC Reading (Ref)IC Reference diode output to intermediate field size NO! YES! Reading (6 mm)diode Reading (100 mm)diode Reading (6 mm)diode Reading (24 mm)diode X Reading (24 mm)IC Reading (100 mm)IC adiosurgery beams exhibit a sharp decrease in output with decreasing field size Significant uncertainty Don’tt use high energy Don Need proof that beam data acquisition for small fields is difficult? Surveyed Beam Data from 40 identical treatment units: Percent Depth Dose Relative Scatter Factors Absolute Dose-to-Monitor Unit CF Reference Condition Applied statistical methods to compare data Percent Deptth Dose servations of some treatment units: 6 mm x 6 mm MLC 110 100 90 Institution A 80 InstitutionAB Institution Institution C 70 60 50 40 30 20 10 0 0 50 100 150 200 250 Depth (mm) 300 350 400 450 10 00 90 Institution A Institution B 80 6.0 % 70 10.3 % 60 50 40 30 20 10 0 0 50 100 150 200 250 Depth (mm) 300 350 400 450 bservations of some treatment units: 5 mm collimator 22 11 Institution A A Institution Institution B 88 66 44 22 00 00 50 50 100 150 Depth (mm) Depth (mm) 200 250 300 servations of some treatment units: 15 mm collimator .2 .2 Institution A Institution A Institution B 11 0.8 .8 0.6 .6 0.4 .4 0.2 .2 00 00 50 50 100 100 150 150 Depth (mm) Depth (mm) 200 200 250 250 300 300 Relative Output Factor: 6 mm x 6 mm MLC Outp put Fac ctor ~45% Outp put Fac ctor Relative Output Factor: 42 mm x 42 mm MLC ~6% Institution Commissioning your system: Does calculation agree with measurement? nd-to-end testing simetric uncertainty Calculation Calculation arc-step = 10o arc-step = 2o ocenter ocenters 4 field box Dynamic Conformal Arcs End-to-end testing Dosimetric Do i et i uncertainty Absolute Dosimetry Independent MU Calculations -to-end dosimetric evaluation RPC SRS Phantom es not include MRI imaging es not include image fusion ovides only a spherical target Hidden Target ecludes l d iimageimage-guided id d positioning iti i int dosimetry limited by TLD accuracy mited information on dose distribution st results require return of phantom Conventional Localization Absolute Dosimetry Lucy Prototype Imaging Insert R contouring and osimetry based on R contours Circular volumes for image fusion MR Fusion MR based contouring, treatment plan and dosimetry What about “Frameless Systems?” “frameless” stereotactic system provides calization accuracy consistent with the safe livery of a therapeutic dose of radiation ven in one or few fractions, without the aid an external reference frame, and in a anner that is non-invasive. ameless stereotaxis is inherently image guided o required: i d mobilization – need not be linked to localization ereo)photogrammetry the principle behind rameless l technologies t h l i Photogrammetry is a easurement technology in which the threemensional coordinates f points on an object are determined by easurements t made d in i two or more photographic images t k taken ffrom different diff t positions Stereophotogrammetry in Radiotherapy Optical otogrammetry Spatial Resolution: 0.05 mm Temporal Resolution: 0 0.03 03 s Localization Accuracy: 0.2 mm Stereophotogrammetry in Radiotherapy X-ray Stereophotogrammetry Calibration of Frameless SRS Systems Establish spatial dimensions Specify the isocenter How do we know the system is targeting properly? nd-to-end evaluation that mimics a patient procedure X-ray entify target & plan DRR Set up p in treatment room Irradiate Results of Phantom Data ( (mm) ) L t Lat. L Long. V t Vert. 3D vector t Average -0.06 -0.01 0.05 1.11 Standard Deviation 0.56 0.32 0.82 0.42 • Sample size = 50 trials (justified to 95% confidence level,, +/- 0.12mm) ) Comparison in 35 SRS patients and 565 SRT fractions 1.2 Lateral Sup/Inf Ant/Post Single Fraction 1 0.8 0.6 04 0.4 0.2 0 -6.0 Single -4.0 -2.0 0.0 2.0 (mm) AP Lat Average 0.05 -0.06 0.06 4.0 6.0 8.0 Axial 3D vector 0.26 1.01 1.2 Lateral S /I f Sup/Inf Ant/Post Multiple Fractions 0 1 0.8 0.6 04 0.4 0.2 0 -6.0 ultiple -4.0 -2.0 0.0 2.0 (mm) AP Lat Average 0 17 0.17 0 17 0.17 4.0 6.0 8.0 Axial 3D vector 0 47 0.47 2 36 2.36 1.2 uperior / Inferior Multiple Fraction Single Fraction 1 08 0.8 0.6 0.4 0.2 0 -6.0 -4.0 -2.0 0.0 2.0 4.0 6.0 8.0 eless localization appears equivalent to frame frame-based based rigid fixation Localization using mplanted fiducials Localization using implanted fiducials Radiosurgery Guidelines TRO/AANS / Consensus Statement on stereotactic radiosurgery d quality improvement, 1993 OG Radiosurgery QA Guidelines, 1993 PM Task Group Report 54, 1995 opean Quality Assurance Program on Stereotactic Radiosurgery, 1995 N 6875-1 (Germany) Quality Assurance in Stereotactic Radiosurgery/Radiotherapy, 2004 PM Task Group 68 on Intracranial stereotactic positioning systems, 2005 R Practice Guidelines for the Performance of Stereotactic Radiosurgery 2006 Radiosurgery, R Practice Guidelines for the Performance of Stereotactic Body Radiation Therapy, 2006 PM Task T kG Group 101, 101 St Stereotactic t ti Body B d Radiotherapy, R di th 2008