CALIBRATING AUTOMATIC  EXPOSURE CONTROL FOR  DIGITAL RADIOGRAPHY

advertisement
CALIBRATING AUTOMATIC
EXPOSURE CONTROL FOR
DIGITAL RADIOGRAPHY
A. Kyle Jones, Ph.D., DABR
Instructor
Department of Imaging Physics
U.T. M. D. Anderson Cancer Center
Houston, TX
Introduction
ƒ AEC is an essential p
part of modern radiographic
g p
systems
à Consistent, reproducible exposures
à Compensation
C
for
f a wide
d range off factors
f
‚ Patient – size, anatomy imaged
‚ Technical factors – kVp,
p, time,, desired receptor
p dose
ƒ Calibrations may need to be adjusted for digital
receptors
à Energy response
ƒ Film O.D. can no longer be used to verify
calibration
A. Kyle Jones, Ph.D. AAPM 2009
AEC System
y
Diagram
g
Generator
AEC
Controller
Backup
Terminate?
kVp
Reference Voltage
‐3 ‐2 ‐1 0 +1 +2 +3
Amplifier
Comparator
Loosely based on Bushberg, Seibert, Leidholdt, and Boone, The Essential Physics of Medical Imaging.
A. Kyle Jones, Ph.D. AAPM 2009
Fundamental AEC performance
p
characteristics
ƒ Initial acceptance testing
à
à
à
à
à
à
Sensor selector/location
Density correction
Screen sensitivity adjustment
Sensitivity vs. speed setting
Reproducibility
AEC balance/field sensitivity
matching
à AEC sensitivity
à Patient thickness tracking
à kVp tracking
ƒ Ongoing QC testing
à
à
à
à
à
à
à
à
AEC sensitivity
Density correction
kVp tracking
Patient thickness tracking
Reproducibility
AEC balance
Backup timer
Minimum response time
‚ Beam quality correction curve
à
à
à
Backup timer
Cell mapping
Minimum response time
A. Kyle Jones, Ph.D. AAPM 2009
Fundamental AEC performance
p
characteristics
ƒ Initial acceptance testing
à
à
à
à
à
à
Sensor selector/location
Density correction
Screen sensitivity adjustment
Sensitivity vs. speed setting
Reproducibility
AEC balance/field sensitivity
matching
à AEC sensitivity
à Patient thickness tracking
à kVp tracking
ƒ Ongoing QC testing
à
à
à
à
à
à
à
à
AEC sensitivity
Density correction
kVp tracking
Patient thickness tracking
Reproducibility
AEC balance
Backup timer
Minimum response time
‚ Beam quality correction curve
à
à
à
Backup timer
Cell mapping
Minimum response time
AAPM 2008 – Extensive look at many fundamental performance characteristics
A. Kyle Jones, Ph.D. AAPM 2009
http://www.instablogsimages.com/images/2007/11/23/digital‐flat‐
panel‐x‐ray‐detector_28.jpg
≠
A. Kyle Jones, Ph.D. AAPM 2009
kVp
p correction
ƒ Many existing AEC systems were
1.00E+04
CsI
Gd2O2S
BaFBrI
1.00E+03
Absorption cross section (cm2/g)
calibrated for use with screen‐
film systems
ƒ The energy response of
gadolinium
d li i
oxysulfide
lfid (Gd2O2S)
screens is substantially different
from that of image receptors
used in digital radiography
ƒ Thus, to properly expose digital
radiographs, we must
recalculate the kVp correction
y
to
curve for our AEC systems
respond correctly considering
the image receptor
characteristics
1.00E+02
1.00E+01
1.00E+00
1.00E-01
0
20
40
60
80
100
120
140
160
Photon energy (keV)
A. Kyle Jones, Ph.D. AAPM 2009
kVp correction curves for DR
ƒ Doyle and Martin have
calculated theoretical
kVp correction curves
for both CR and iDR
detectors
ƒ Also note that the
addition of small
amounts of Cu
filt ti d
filtration
does nott
significantly affect the
calibration
Doyle, P and Martin, CJ, Calibrating automatic exposure control
devices for digital radiography, Phys. Med. Biol. 51:5475‐5485,
2006.
A. Kyle Jones, Ph.D. AAPM 2009
kVp Tracking Unit #1
800
120
700
100
80
500
400
60
300
Noise/SNR
R
Pixel value
P
600
Pixel value
Noise
SNR
40
200
20
100
0
0
60
70
80
90
100
110
120
130
kVp
A. Kyle Jones, Ph.D. AAPM 2009
225
80
220
70
215
60
210
50
205
40
200
30
195
20
190
10
185
0
50
70
90
110
Noise/SNR
Pixel value
kVp Tracking - Unit #2
Pixel value
Noise
SNR
130
kVp
A. Kyle Jones, Ph.D. AAPM 2009
AEC Balance
ƒ Field
e d se
sensitivity
s t ty matching
atc g
ƒ To achieve consistent exposures for all views,
AEC cells should be balanced
ƒ Manufacturers of x‐ray systems use several
different schemes for AEC balancing
ƒ Thus, a one‐size‐fits‐all test will not be valid if
you work with a variety of systems
A. Kyle Jones, Ph.D. AAPM 2009
AEC Balance
ƒ Test AEC balance during
g acceptance
p
testing
g to set a
baseline standard for balance
ƒ Ask your service engineer about the
calibration/balancing procedure
à Many manufacturers today use AEC systems that are not
serviceable for balance – they can only be replaced
à In this case you are stuck with manufacturer
manufacturer’ss balance scheme
à Other manufacturers still have tunable AEC cells
‚ Pots in generator
‚ Software interface
‚ Pots in detector housing
ƒ Published guidelines/recommendations
à ± 5% across all combinations (AAPM 14)
14)*
A. Kyle Jones, Ph.D. AAPM 2009
AEC Balance
ƒ Cells must also be matched when used in combinations
ƒ AEC systems may terminate the exposure when the most
sensitive cell reaches the required current, or may average
the signal between the cells being used
ƒ Same criteria apply
C
L
R
L+R
L+C
R+C
All
Unit 1
6.32
7.23
7.16
7.16
6.73
6.7
6.83
Unit 2
46.6
69.9
70.4
69.9
55.7
55.9
59.8
Unit 3
4.5
4.32
4.44
4.39
4.39
4.44
4.39
See also ‘X‐ray generator and automatic exposure control device acceptance testing’ by Raymond P. Rossi, M.S., in
Specification, Acceptance Testing and Quality Control of Diagnostic X‐ray Imaging Equipment, Proceedings of the
1991 AAPM Summer School
A. Kyle Jones, Ph.D. AAPM 2009
AEC SENSITIVITY
Methods for calibration
ƒ Noise‐based method
ƒ Ka‐based methods
à Use a CR cassette with a
cutout for a detector
à Solid‐state detector behind
grid
à Pre‐detector Ka and primary
transmission
i i through
h
h grid
id*
ƒ Exposure indicator (EI) –
based methods
à EI
à Pixel value
Doyle, P and Martin, CJ, Calibrating automatic exposure control devices
for digital radiography, Phys. Med. Biol. 51:5475‐5485, 2006.
‚ Characteristic function
à SNR
A. Kyle Jones, Ph.D. AAPM 2009
Noise‐based AEC calibration
ƒ For each kVp,
p, an S number can be associated with
the standard deviation of pixel values that produce
the desired noise characteristics
ƒ Recommend use of an SNR‐based threshold for
choosing noise level
à Example: SNRthreshold of 5 → Signal difference of 50 can be
seen with σ = 10
à Still some determination to be done
ƒ Setting up AEC in Semi‐automatic EDR mode
yields clinically valid results in Automatic EDR
mode
Christodoulou, EG, Goodsitt, MG, Chan, H, and Hepburn, T,
Phototimer setup for CR imaging, Med. Phys. 27:2652‐2658, 2000.
A. Kyle Jones, Ph.D. AAPM 2009
Ka‐based methods
ƒ CR cassette with cutout
cutout*
ƒ S ~ 200/E, E = exposure (mR) for Fuji CR
ƒ Cutout machined into IP cassette to accept
solid state dosimeter
à Introduces realistic scatter* from cassette,
cassette
overcomes cassette sensing
Doyle P, Gentle D, Martin CJ, Optimising
g automatic exposure
control in computed radiography and the impact on patient dose,
Rad. Prot. Dosim. 114:236‐39, 2005.
A. Kyle Jones, Ph.D. AAPM 2009
Ka‐based methods
ƒ Solid‐state dosimeter with lead backing can be
used to reduce the effect of backscatter
ƒ Not possible for all DR systems
http://www.rpi.edu/dept/radsafe/public_html/RADIA
TION%20DETECTORS_files/MOSFET.jpg
http://www.unfors.com/products.php?catid=168
Doyle P, and Martin CJ, Calibrating automatic exopsure control
devices for digital radiography, Phys. Med. Biol. 51:5475‐85, 2006.
A. Kyle Jones, Ph.D. AAPM 2009
Pre‐detector Ka
ƒ Simple measurement to make
ƒ Eliminates backscatter
ƒ Necessitates measurement of Tp or B of anti‐
anti
scatter grid
ƒ Some inaccuracy
inacc rac introduced
introd ced due
d e to beam
hardening by grid which is not considered in
this measurement
A. Kyle Jones, Ph.D. AAPM 2009
EI‐based
ƒ Calibrate using
g the EI p
provided byy the
manufacturer
ƒ Many caveats to consider
à Calibration conditions
‚ AEC
‚ Detector
‚ EI
à Practicality
à Usefulness of EI
ƒ Basically the same as Ka‐based methods
à Dependence of EI on beam conditions,
conditions grid,
grid etc
etc.
A. Kyle Jones, Ph.D. AAPM 2009
Three different methods…
ƒ GE
ƒ Siemens
Si
à Variable
à 0.6 mm Cu
PMMA
thickness (vs
kVp) to
calibrate
lib t EI
à 20 mm Al
phantom to
p
calibrate AEC
and detector
à 80
8 kVp
kV
filter to
calibrate
AEC, EI, and
detector
à 70 kVp
ƒ Fuji
F ji
à 80 kVp bare
beam to
calibrate EI
à We use 0.5
mm Cu + 1
mm Al
à They don’t
calibrate the
AEC for
retrofitted
systems
A. Kyle Jones, Ph.D. AAPM 2009
To what value should I
calibrate?
ƒ CR – Calibrate AEC to deliver the desired Ka at
the image plane
à Vary Ka by body part and view using exposure
indicator targets
‚ E.g.
E g S = 200 general,
general S = 100 extremity
ƒ DR – Calibrate AEC to manufacturer’s
specifications
à E.g. 2.5 µGy @ 400 sensitivity setting
à Use sensitivity
y setting
g to varyy Ka byy bodyy part
p and
view
A. Kyle Jones, Ph.D. AAPM 2009
35
120.00
mAs
30
SNR
80.00
20
60.00
15
SNR
mAs
25
100.00
40.00
10
20.00
5
0
0
200
400
600
800
1000
1200
1400
1600
0.00
1800
Sensitivity setting
S1
mAs 2 = mAs1 ×
S2
A. Kyle Jones, Ph.D. AAPM 2009
SNR vs. Sensitivity setting
120.00
PMMA
100.00
Aluminum
SNR
80 00
80.00
60.00
40.00
20.00
0.00
0
200
400
600
800
1000
1200
1400
1600
Sensitivity
A. Kyle Jones, Ph.D. AAPM 2009
CR
ƒ Test tool (e.g.
(e g Radchex)
Simplest
p
ƒ CR plate, Sensitivity (semi‐EDR),
calibrated reader (use EI)
ƒ Measure X or Ka at detector
plane
Most Complex
A. Kyle Jones, Ph.D. AAPM 2009
Test tools
ƒ A common test tool is
the CR Radchex meter
ƒ Consists of a sensor that
mimics the
h response off a
CR screen
ƒ Insert in Bucky,
Bucky make
exposure, get reading
ƒ Tables provided to
convert CRLU to EI or Ka
http://disccorp.mb.ca/products/Wired_CRRadchex.html
A. Kyle Jones, Ph.D. AAPM 2009
Radchex meter
Cons
Pros
ƒ Fast
ƒ Does not require plate or
reader
d
ƒ Easy
ƒ Defeats cassette sensing
ƒ May not exactly simulate
ƒ
ƒ
ƒ
ƒ
response of CR plate
M not exactly
May
l simulate
i l
clinical situation
Expensive
p
Lack of patient‐like
attenuator
Transmitted spectrum
through 1.5 mm Cu –
CRLU to EI conversion
may not be correct
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
CR p
plate and reader
ƒ Use a patient
patient‐equivalent
equivalent attenuator and
expose a CR plate in the Bucky
à ‘Test’
Test plate confirmed to be uniform with clinical
stock
à Reader calibration must be verified
ƒ Exposure indicator is the test quantity
à ‘Test’ mode – e.g.
g Sensitivity,
y, Semi‐EDR
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
CR p
plate and reader
Pros
Cons
ƒ Cheap
ƒ Reader must be
ƒ Simple if reader
calibrated
ƒ Takes more time than
Radchex meter
calibration is known to
be good
ƒ Exactly simulates
response of CR plate
ƒ Simulates clinical
situation
ƒ Defeats cassette sensing
à Multiplied with more
exposures
à Wait period
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
Measure Ka
ƒ Use ionization
o at o cchamber
a be in Bucky
uc y to measure
easu e
Ka
ƒ Doyle
y et al. p
proposed
p
that an old cassette be
modified to hold a solid state device to more
accurately simulate the clinical case
ƒ A solid state dosimeter can also be used
à Behind AEC chambers
à Pb‐backed,
Pb b k d so lless b
backscatter
k tt concern
‚ Does not integrate scatter from 2π in front
à Still helpful to “mount”
mount to keep centered
Doyle P, Gentle D, Martin CJ, Optimising automatic exposure control in computed radiography and the impact on patient dose, Rad.
Prot. Dosim. 114:236‐39, 2005.
A. Kyle Jones, Ph.D. AAPM 2009
Measure Ka
Pros
Cons
ƒ Cheap†
ƒ Does not simulate
ƒ Accurately simulates
response of CR plate
plate*
ƒ Time consuming
clinical situation
ƒ Direct measurement of
Ka
ƒ Independent
verification – does not
rely on plates or reader
à Setup
p
ƒ Requires additional
equipment
q p
(modified
cassette)
A. Kyle Jones, Ph.D. AAPM 2009
Measure Ka
Pros
Cons
ƒ Cheap†
ƒ Does not simulate
ƒ Accurately simulates
response of CR plate
plate*
ƒ Time consuming
clinical situation
ƒ Direct measurement of
Ka
ƒ Independent
verification – does not
rely on plates or reader
à Setup
p
ƒ Requires additional
equipment
q p
(modified
cassette)
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
Comparison
p
of methods
Scenario
mAs
S#
µGy
CRLU/Speed #
CR cassette w/screen
72.5
196
8.53
CR cassette w/o screen
86.1
15 cc chamber in wood
78 7
78.7
11 39
11.39
15 cc chamber in wood w/Pb
73.4
13.74
Nothing in Bucky
87.7
Unfors Xi
91.1
9.95
Radchex w/PMMA*
82.2
11.5
15.7/127
Radchex w/1.5 mm Cu*
22.4
11.6
18.6/107
Target Ka in Bucky at detector plane = 8.76 µGy
8.76 µGy to plate under calibration conditions (0.5 Cu/1 Al) yielded S# of 191
A. Kyle Jones, Ph.D. AAPM 2009
iDR and dDR
ƒ Use exposure
p
indicator to determine
Ka
Simplest
p
à Must verify accuracy of indicator
‚ TG 116
ƒ Measure pre‐detector Ka and
calculate Ka at detector plane
à Need
N d to
t know
k
TP off grid
id
à Depends on calibration conditions
ƒ Measure Ka at detector p
plane
ƒ Measure CF for clinical conditions
and use it to calculate Ka
Most Complex
A. Kyle Jones, Ph.D. AAPM 2009
Ideally
y
ƒ Verify
y EI at calibration conditions and use the EI
to calibrate the AEC
ƒ Manufacturer methods
à GE: 2.5
2 5 µGy to detector @ 400 sensitivity,
sensitivity measured,
measured
factors applied for distance and table/cover
‚ Without grid if removable
‚ “Grid
Grid factor
factor” used for dedicated chest systems
‚ 20 mm Al phantom, 80 kVp
 Unfortunately, EI conditions are different
à Siemens
Siemens: Detector dose = 1000/Sensitivity
‚ 2.5 µGy @ 400 sensitivity, no grid, 70 kVp
‚ EI conditions are the same
A. Kyle Jones, Ph.D. AAPM 2009
VERIFYING THE EI
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
GE
Siemens
A. Kyle Jones, Ph.D. AAPM 2009
Technical mode
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
GE EI verification
DEI
Measured
pre‐grid
(µGy)
Post grid
Post‐grid
(Tp) (µGy)
55.59
59
1.32
3
55.53
53
55.53
53
5.84
5.84
1.46
5.48
5.48
90 kVp
kV
no grid
5.07
4.74
1.21
4.46
4.46
70 kVp
grid
4.28
3.41
1.45
8.86
4.34
80 kVp
grid
3.21
2.23
1.00
5.48
2.74
90 kVp
grid
2.82
1.91
0.82
4.46
2.19
Scenario
UDExp
(µGy)
CDExp
(µGy)
70 kVp
no grid
id
4.95
4
95
80 kVp
no grid
Post grid
Post‐grid
(B) (µGy)
1.69
A. Kyle Jones, Ph.D. AAPM 2009
Siemens EI verification
Measured pre
pre‐
grid (µGy)
Post grid (Tp)
Post‐grid
(µGy)
2.24
2.24
2.20
2.20
496
6
2.27
2.27
60 kVp grid
358
4.28
2.77
70 kVp grid
389
4 18
4.18
2 74
2.74
81 kVp grid
439
4.52
3.00
Scenario
EXI
60 kVp no
grid
g
354
70 kVp
no grid
401
81 kVp
8
p no
o
grid
Calculated from
EXI (µGy)
2.65
2 57
2.57
A. Kyle Jones, Ph.D. AAPM 2009
Effect of table or cover
ƒ Attenuation by the table may be significant
à Manufacturer 1: 0.80 to 0.82 from 60 – 81 kVp
à Manufacturer 2: 0.91 at 80 kVp
ƒ Detector covers had very little influence
A. Kyle Jones, Ph.D. AAPM 2009
Other considerations
ƒ Measuring
g Ka at detector p
plane mayy not be
possible
ƒ Removal of the anti‐scatter grid may alter AEC
response
ƒ Exposure indicator is usually provided
ƒ Would like to verify
y it’s performance
p
under
calibration conditions (hopefully the same as
AEC conditions) and with presence of grid
à kVp correction – if EI is correct under calibration
conditions for AEC, rely on kVp correction/SNR
tracking instead of worrying about EI and kVp
p
dependence
A. Kyle Jones, Ph.D. AAPM 2009
Clinical simulation
ƒ
ƒ
ƒ
ƒ
ƒ
8” PMMA phantom
p
10 x 10 in field size at tabletop
AEC,, center cell
GE: 80 kVp
Siemens : 70 kVp
mAs
UDExp (µGy)
CDExp (µGy)
DEI
11 29
11.29
3 53
3.53
2 45
2.45
11
1.1
mAs
EXI
Calculated (µGy)
27.76
284
1.88
A. Kyle Jones, Ph.D. AAPM 2009
Other questions
q
ƒ Dependence of AEC calibration on field size
ƒ Batch‐to‐batch PMMA variation
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
A. Kyle Jones, Ph.D. AAPM 2009
Batch‐to‐batch PMMA
ƒ I tested two different batches of PMMA,
PMMA and
found:
à 3% variation in mAs
à 1% variation in EI
ƒ n
n=1,
1, variation in batches/sources of PMMA
was not a concern
A. Kyle Jones, Ph.D. AAPM 2009
One
http://www.toolandtechnology.com/images/vernier%20calipers.jpg
A. Kyle Jones, Ph.D. AAPM 2009
Huge
learnwithmst.wordpress.com
A. Kyle Jones, Ph.D. AAPM 2009
Problem
http://www.ricardodiaz.com/wp‐content/uploads/2009/01/axe.jpg
A. Kyle Jones, Ph.D. AAPM 2009
Image
g receptor
p
dose
ƒ Relatively
y simple
p task with screen/film imaging
g g–
ƒ
ƒ
ƒ
ƒ
achieve O.D. in linear portion of H+D curve
Digital imaging is not contrast‐limited, but noise
limited
How can you set up your AEC system to deliver
the desired amount of noise in an image?
AEC system must be calibrated to deliver the
desired Ka to the image receptor
Fundamental difference from S/F: Wide range of
Ka will yield usable images, must decide on
acceptable noise level in images
A. Kyle Jones, Ph.D. AAPM 2009
Rong XJ, Shaw CC, Liu X, Lemacks MR, Thompson SK, Comparison of an
amorphous silicon/cesium iodide flat‐panel digital chest radiography system
with screen/film and computed radiography systems — A contrast‐detail
phantom
h
study,
d Med.
M d Ph
Phys. 28(11):2328‐35,
8( )
8
2001.
Use of scored images of contrast‐detail phantom (CDRAD) under clinical
image processing conditions to attempt to determine detector exposure
required
i d tto achieve
hi
similar
i il detectability
d t t bilit for
f small
ll low‐contrast
l
t t targets
t
t
Conclusion: 0.43 mR required to achieve similar LCD for 0.5 mm object compared
to 2.09 mR for the same screen‐film image (21%)
A. Kyle Jones, Ph.D. AAPM 2009
Huda W, Slone RM, Belden CJ, Williams JL, Cumming WA, Palmer CK, Mottle
on computed radiographs of the chest in pediatric patients, Med. Phys.
199:242‐252, 1996.
Use of a rating system based on degrees of mottle to determine what
sensitivity number of a Fuji AC series would be needed to match the mottle
in a 600‐speed s/f system, then use the actual sensitivity numbers
measured
d to
t compare CR tto s/f.
/f
Conclusion: To achieve comparable
p
mottle to 600‐speed
p
s/f system
y
a two‐fold
increase in exposure is needed as compared to mean sensitivity number. An
exposure consistent with a 200‐speed s/f system would be needed to achieve
negligible mottle.
A. Kyle Jones, Ph.D. AAPM 2009
Liu X and Shaw, CC, a‐Si:H/CsI(Tl) flat‐panel versus computed radiography for
chest imaging applications: image quality metrics measurement, Med. Phys.
331(1):98‐110,
( )9
, 2004.
4
Compare indirect digital radiography and cassette‐based digital
radiography systems in terms of fundamental image quality metrics.
Conclusion: Indirect digital
g
radiography
g p y detector has significantly
g
y higher
g
DQE
and lower NPS than cassette‐based digital radiography (CR) detector. The
magnitude of the differences is illustrated in the graphs from the manuscript.
A. Kyle Jones, Ph.D. AAPM 2009
Give radiologists a visual idea of what the impact of dose to the image
receptor is
i on iimage quality
li – CDRAD phantom
h
exposed
d under
d PMMA
Can also be done with other phantoms – ACR R/F phantom, anthropomorphic
phantom, or even add noise to a patient image for comparison
A. Kyle Jones, Ph.D. AAPM 2009
Caveats
ƒ Image processing has a large impact on noise
(low‐contrast detectability) and high‐contrast
resolution,, and thus AEC sensitivityy should be
configured with this in mind.
ƒ Also, pixel size has an impact on noise in
images – this is especially important for
digital
g
receptors
p
where p
pixel size is variable,,
such as PSP systems
A. Kyle Jones, Ph.D. AAPM 2009
Conclusions or
ƒ Several methods can be used for AEC sensitivity
y
calibration
ƒ Other AEC performance characteristics should
be monitored and evaluated
ƒ Exposure conditions have a tremendous impact
on AEC sensitivity calibration
ƒ Clinical exposures are made using clinical
exposure conditions
à Grid
à Scatter
à Field size
A. Kyle Jones, Ph.D. AAPM 2009
Questions left unanswered
Q
ƒ How
o does o
onee dec
decide
de what
at dose to tthee
receptor is needed?
ƒ How does one evaluate the impact
p of
increased or reduced detector dose?
à Dose – fairly easy
à Image quality – hard
‚ Clinical images – anatomical noise
‚ Image processing
‚ Not round objects and nice orderly line pairs
‚ This is a really hard question
A. Kyle Jones, Ph.D. AAPM 2009
Download