2D Images KV CBCT Imaging Part I Rabih Hammoud, MS, DABR Henry Ford Health System, Detroit, MI Hamad Medical Corp. Doha, Qatar Learning objective: Detroit vs. Doha Outline 1. The objective of this educational session is to review KV-CBCT and MV-CBCT 2. imaging systems for daily localization 4. 3. 5. 6. Commissioning, image quality, dose, registration process, and acquisition modes Clinical integration QA, stability over time, and downtime Standard clinical applications Novel clinical applications Technology evolution and future directions 1 3D kV Imaging Siemens Artiste™ KVision Varian Trilogy OBI® Elekta Synergy™ VolumeView What’s CBCT? Novalis Tx™ On Board Imager CBCT acquires projections of a patient, Send them to a reconstruction application, and then returns a 3D image In the 3D/3D Match workspace, the CBCT images can be registered to the reference images, either manually or automatically, using a 3D mutual information algorithm. Elekta Axesse™ Why CBCT? Advances in treatment planning and delivery systems allow for higher doses to target and lower dose to normal tissue With the resulting steep dose gradients, motion management ( inter and intrafraction) becomes more critical Tumors often not visible in 2D images Role of CBCT is to help reduce interfractional motion and try to assess patient status ( tumor evaluation, adaptive planning, ..) Volume = 4/3 r3 Small margin reduction (5mm) decreases the volume of an orange by 1/2 2 CBCT work Flow Intra-fraction Prostate Motion What can we do to decrease intra-fraction motion? Data Preparation Planning CT with structure set Isocenter information Send via Record Verify System • Acquisition time (scan modes) Workflow • Reconstruction time Select patient Extend arms/gantry starting point Imaging parameters Acquire/reconstruct CBCT Align (bony, soft tissue, VOI) Apply shifts and record Post shifts? Treat • Matching (ROI, Intensity range, automatic/manual) • Treatment time (RapidArc?) Langen et al. IJROBP 71(4):1084-1090, 2008 Tolerated difference between the planning CT and CBCT depends on image quality, image registration, internal organ motion, margin definitions, user experience which are clinic and anatomical site dependent Elekta Synergy™ Half Scan vs. conventional x-ray tube mounted on a retractable arm extends from the accelerator gantry’s drum structure A 41×41 cm2 flat-panel x-ray detector is mounted opposite the kV tube at a gantry position of 590° in IEC coordinates The X-ray tube is powered by a highfrequency generator, operates ( 60-150 kVp) Full Scan Varian Trilogy OBI® Novalis Tx™ Full Fan Scan Half Fan Scan 3 Commissioning Commissioning of the OnOn-Board Imaging system is divided into 2 parts: Mechanical Calibration Imaging System Calibrations Mechanical Calibration Indexing the Arms: to establish a reference point for the motion of the joints of arms for KV source, KV and MV detectors using encoders with no absolute positions Indexing the Exact ArmArm-MV Detector Arm KV Source Mechanical Calibration ResultsResults-KV Source Mechanical Calibration Before After 4 Imaging System Calibration CBCT Mode Calibration To correct for non-uniform response of the image receptor and non-uniform intensity of the x-ray source, Reinsures the uniformity of receptor electronic readouts and updates pixel defect map (PDM), procedure includes: Dark field image (DF image): acquired without radiation & reflects any imperfection Flood field image (FF image): several images are acquired using radiation, it reflects field homogeneity pixel sensitivity and amplifier gain Geometry Calibration Imaging Calibration HU Calibration Dual Gain Calibration I0 Calibration Normalization Defective pixels locations are stored in PDM. When images are acquired the PDM is loaded and defective pixels are replaced with an average of neighboring pixels Geometry Calibration • Corrects for machine geo instability during scans Dual Gain Calibration Used to enhance the dynamic range of the imager for CBCT imaging Each projection is read twice ( at high sensitivity, and at low sensitivity) Determines the linearity of each pixel and the threshold above which the pixel values from the high sensitivity image are replaced by scaled pixels from the low sensitivity range image • Uses a phantom of known geometry (five needles) placed at the isocenter to estimate the kV/CBCT isocenter. • Many images of the phantom are acquired at different angles. • Since phantom/needles dimensions are known, one can estimate the location of the OBI isocenter • 3D/3D match of the CBCT uses the results from geo Cal to estimate the location of the kV isocenter Imaging Calibration I0 Calibration Needle Phantom It reduces ring artifacts in scans, stored in the reconstructor It’s repeated multiple times resulting in charge trapping similar to what occurs during CBCT acquiition 5 HU Calibration Quality Assurance Program Uses Catphan Phantom Calibrate HU for pixels in the reconstructed image based for known inserts Beam Hardening Correction: • Corrects for increase in energy as x-rays passes through the patient • Due to the polychromatic nature of the X-rays, lower energy components are attenuated as goes through the patient • Will increase the effective x-ray energy which causes inconsistency in the reconstruction that assumes constant energy Safety and Functionality Check To check that the safety features are functioning properly and that the entire system is ready for clinical operations Daily Daily/Weekly Monthly or Quarterly Geometry Check To verify that the KV source and KV detector have maintained their geometric accuracy and stability 6 weekly geometric QA Image Quality To monitor the quality of radiographic and CBCT images over time Not comparable to conventional CT scanners QA tests adopted from diagnostic CT scanners using Catphan 504 phantom Based on establishing a baseline value Image Quality Tests include: a) b) c) d) Low contrast resolution Spatial resolution HU accuracy HU uniformity Some References for OBI QA… S. Yoo, G. Y. Kim, R. Hammoud, E. Elder, T. Pawlicki, H. Guan, T. Fox, G. Luxton, F. F. Yin and P. Munro, Med Phys 33 (11), 4431-4447 (2006) J. P. Bissonnette, D. Moseley, E. White, M. Sharpe, T. Purdie and D. A. Jaffray, Int J Radiat Oncol Biol Phys 71 (1 Suppl), S57-61 (2008) J. P. Bissonnette, D. J. Moseley and D. A. Jaffray, Med Phys 35 (5), 1807-1815 (2008) 7 Patient Dose from CBCT System Dependent kV/ mA Number of projections kV system properties (bow-tie) kV system field size Patient Dependent Size/shape of patient Body part What is our interest? Some Approaches of Characterizing CBCT Dose… Taking in vivo dose measurements in a Rando phantom and using the data as a predictor of patient dose Taking dose measurements directly on patients undergoing CBCT using TLD CT Dose Index (CTDI). Well defined conditions. Defining a dose ,metric for the cone beam dose index (CBDI) and taking dose measurements with a standard cylindrical risk/ benefit ratio needs to be considered Some Approaches of Characterizing CBCT Dose Taking in vivo dose measurements in a Rando phantom and using the data as a predictor of patient dose Taking dose measurements directly on patients undergoing CBCT using TLD Technique used: 125 kVp, 80 mA, 25 ms (2 mA s) AP skin doses ranged from 3-6 cGy for 20-23 separation Central dose was ~3.0 cGy The left hip received 10-11 cGy while the right received 6-7 cGy Wen N, Guan H, Hammoud R et. al. Dose delivered from Varian’s CBCT to patients receiving IMRT for prostate 2007 Phys. Med. Biol. 53(11): 2897-909 Another Approach Defining a dose metric for the cone beam dose index (CBDI) and taking dose measurements with a standard cylindrical CT phantom using both a 100 mm ion chamber and 0.6 cc farmer chamber to predict patient dose CBDIfW= 1/3 CBDIfCentre+ 2/3 CBDIfPeriphery (cGy) Adopting the same area averaging approximation used in conventional CT called CTDIW in which the CAX dose D0 (r = 0, where r is the distance from the centre of the CT phantom to the point of measurement) is weighted by 1/3 and peripheral axis doses Dp (r = R 5 1 cm, where R is the phantom radius) are weighted by 2/3 E. K. Osei et al., J. Radiol. Prot. 29 (1), 37-50 (2009) 8 CBCT Dose-OBI Advanced Imaging 16 cm phantom 40 mA, 25 ms (1 mA s) Full Fan No Bowtie • Amer et al.2007, reported 0.38 cGy/ 100 mAs for the Elekta Synergy System using 32 cm diameter phantom • Song et al. 2008, reported 0.43 cGy/ 100 mAs using 30 cm diameter phantom Full Fan with Bowtie 32 cm phantom 40 mA, 25 ms (1 mA s) Half Fan No Bowtie Half Fan with Bowtie CBDIfCentre (cGy) 7.47 4.65 2.85 2.01 CBDIfPeriphery (cGy) 9.34 5.06 6.34 2.97 CBDIfw (cGy) 8.72 4.93 5.18 2.65 CBDIfCentre (cGy/ 100 mAs) 1.15 0.72 0.44 0.31 CBDIfPeriphery (cGy/ 100 mAs) 1.44 0.78 0.98 0.46 CBDIfw (cGy/ 100 mAs) 1.34 0.76 0.80 0.41 AAPM TG for Imaging Dose Murphy, M.J., et al., The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. Med Phys, 2007. 34(10): p. 4041-63. Two assumptions are made in the current MLC bases IMRT process 1. Geometric sizes, shapes, and locations of the targets and organs are the same at the time of planning CT pCT 2. Delivered fluence maps are the same as the planned ones 9 This can trigger a replan for specific patients when warranted by significant changes Summary Commissioning of kV CBCT systems involves the characterization of the alignment of the kV CBCT system components with the linear accelerator isocentre No image Guidance Bony/Soft Tissue Commissioning work then typically focuses on establishing baseline values for image quality. Parameters assessed include spatial linearity (i.e., distance and scale), image uniformity, high and low contrast spatial resolution, and accuracy of CT numbers Finally, commissioning work involves characterization of the radiation dose to be used for IGRT. The literature reports point dose measurements involving Farmer chambers inserted in cylindrical water or acrylic phantoms 10 Recently, a parameter analogous to the commonly accepted CTDI has been adapted for the kV CBCT geometry, leading to the introduction of CBDI Though dose delivered from CBCT is small compared to the treatment dose, according to ICRP-60, the dose delivered to patients in medical imaging should be justified and optimized Justification is positional and target localization which may allow for dose escalation and potentially higher rates of tumor control and lower rates of complications Thank You 11