Converting the Radiology Department from Film-Screen to Digital: Making the Transition S. Jeff Shepard, MS, DABR University of Texas M. D. Anderson Cancer Center Houston, Texas jshepard@di.mdacc.tmc.edu jshepard@di.mdacc.tmc.edu AAPM 2005 Continuing Education Course Radiographic and Fluoroscopy Physics and Technology Learning Objectives Understand the issues that a Medical Physicist is likely to face when supporting a clinical operation that is undergoing (or is about to undergo) a conversion from filmfilm-screen based radiography to digitally acquired images. AAPM 2005 Outline Image Acquisition • • • Getting started with CR Introducing direct digital radiography Digital Mammography Getting Started with CR Replace F/S receptors with CR • StraightStraight-forward • Cost effective Start in Portable XX-Ray • Traditional challenge in image quality Quality control • • • • AAPM 2005 – F/S is unforgiving – Over and under exposures are common Acquisition device QC Display device QC Operational QC Patient Identification MDACC Imaging Physics MDACC Imaging Physics • Wider dynamic range of CR yields quick improvements – Speed adjusts to technique automatically AAPM 2005 MDACC Imaging Physics 1 Getting Started with CR Different detector sensitivity and energy dependence • New technique charts Wide dynamic range • Consistent image quality over a wide range of exposures • Very beneficial (image consistency) • Potential downsides MDACC Imaging Physics AAPM 2005 Getting Started with CR UnderUnder- and OverOver- exposure • Image consistency (contrast and brightness) is maintained Fewer photons = More noise • Obscures lowlow-contrast details More photons = More signal strength • SignalSignal-toto-noise ratio improves • Beautiful images • High patient dose! Flexibility in technique selection can lead to higher patient doses (“ (“Exposure Creep” Creep”) Image Quality: Image Quality: 0.1 Range of Exit Exposure Data (VOI) 1.0 Unused Dynamic Range 10 Exposure Data Recognition Signal Strength Signal Strength Exposure Data Recognition Unused Dynamic Range MDACC Imaging Physics AAPM 2005 100 Incorrect VOI 0.1 Exposure (mR) System Dynamic Range 1.0 10 100 Exposure (mR) System Dynamic Range VOI recognition failure • Body habitus (Pediatrics and postpost-surgery) • Patient mis-position – air peak • Over-collimation, gonadal shields and prosthetics – Special processing algorithms – Fixed speed techniques AAPM 2005 MDACC Imaging Physics AAPM 2005 MDACC Imaging Physics 2 Getting Started with CR Technologist Feedback – Detector Exposure Indicators (E.I.) • • • • • • • Agfa CR - Lgm value (Logs available) Fuji CR - S Number (Logs available) Konica CR – S Number Kodak CR – EI (Exposure Index) Canon DR - REX Number Philips & Siemens DR - Speed value GE DR - ????? Getting Started with CR Technologist Feedback – Detector Exposure Indicators (E.I.) • Exposure to the detector • Accurate and consistent (reproducible) • Patient exposure index (DAP or ESE) – not the same! • Not editable!!! Proprietary algorithms – confusion AAPM (TG116), IEC and DIN are evaluating AAPM 2005 MDACC Imaging Physics AAPM 2005 Getting Started with CR Tuning the image quality • Verify QC console calibration conforms to DICOM PS3.14 first Getting Started with CR Tuning the image quality Convert each step to a %JND of the entire range of JND’ JND’s from Lmin to Lmax – Compare JND intervals in a test pattern to those on a PACS monitor % JND (step) = 100% x Set post processing to drive display to Lmin and Lmax Measure steps on the QC monitor with a photometer (OR(OR-3) Calculate JND at each step (DICOM 3.14, Table B1) DICOM is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications of medical information. MDACC Imaging Physics AAPM 2005 MDACC Imaging Physics JND(step)JND(step)-JND(Lmin) JND(Lmax) – JND(Lmin) Repeat on PACS monitor and compare at each step For DICOMDICOM-compliant PACS, calculate JND index at each step and test for linearity from Lmin to Lmax AAPM 2005 MDACC Imaging Physics 3 Display Matching 100% 90% Modality QC 80% PACS %JND 70% 60% 50% 40% 30% 20% 10% 0% 0 MDACC Imaging Physics AAPM 2005 AAPM 2005 100% 90% Modality QC 80% PACS 600 800 1000 QC console calibration is supported by few vendors • Use contrast and brightness settings to get as close as possible • Pressure vendors to add calibration capability to QC displays • Calibrate it yourself 70% %JND 400 Digital Driving Level MDACC Imaging Physics Getting Started with CR Display Matching 60% 50% 40% 30% 20% – Add 3rd party software – Replace display with selfself-calibrating 10% 0% 0 AAPM 2005 200 200 400 600 Digital Driving Level MDACC Imaging Physics 800 1000 AAPM 2005 MDACC Imaging Physics 4 Getting Started with CR Tuning the Image Processing • Start with vendor default settings on sample “raw” raw” images Getting Started with CR Portable XX-Ray • Start with chest • Add abdomen/pelvis • Add long bones – VendorVendor-supplied unprocessed image data of real patients for tweaking – Work with key technologists and radiologist(s) radiologist(s) to find optimal default settings – Change individual settings one at a time • Phantoms first for contrast and dose, then 5 patients • Adjust & repeat as necessary AAPM 2005 MDACC Imaging Physics AAPM 2005 Getting Started with CR General Radiography • Recalibrate AEC Direct Digital Radiography Introduction into the clinical operation: Same strategy as with CR • Verify AEC calibration • Verify QC console calibration • Start with vendor default settings on sample “raw” raw” images • Phantoms first, then patients • Adjust techniques and processing & repeat as necessary • Get processing established in one room • Migrate to other rooms when stable – Constant S/N ratio or detector exposure – not O.D. • Calibrate QC console • Use same strategy as with portable X-Ray to get techniques and postpostprocessing established in one room • Migrate to other rooms when stable AAPM 2005 MDACC Imaging Physics MDACC Imaging Physics AAPM 2005 MDACC Imaging Physics 5 Image Quality Moire patterns between the fixed grid lines and the and detector sampling matrix. Digital Mammography Acquisition Devices • FullFull-Field Digital Mammography (FFDM) – Scanning slot CCD with CsCs-I – Amorphous Silicon TFT with CsCs-I scintillator – Amorphous Selenium – CR (Pending FDA) • Usually seen in CR, but can happen in DR • Use high grid line frequency (> 4 lines/mm) • Some systems employ low pass filters – Decreases resolution – Inappropriate for smaller plates • Most DR’ DR’s use high line rate (~8 lines/mm) stationary or standard reciprocating grids AAPM 2005 MDACC Imaging Physics Digital Mammography Introduction into the clinical operation: • Same as DR and CR in general radiography • Verify AEC & QC console calibration • Tune the techniques in one room, then propagate to others Digital Mammography Connectivity • Huge data sets (up to 300 MB/study) require large local storage and fast hardware • Dedicated workstations using pointtopoint-topoint connections can impose crippling additional transmission times • Intermediate routing modules – kV selection may be limited • Image postpost-processing flexibility is restricted by FDA AAPM 2005 MDACC Imaging Physics MDACC Imaging Physics AAPM 2005 – Bottleneck if not very robust – Single point of failure AAPM 2005 MDACC Imaging Physics 6 Digital Mammography Workflow – What to store? • Unprocessed images (“ Processing”) (“For Processing” Digital Mammography Workflow – What to store? • Both? – Interpretation workstation must apply FDAFDAapproved processing – Unprocessed images are ugly – Processed for clinics, unprocessed for Radiologist’ Radiologist’s workstations – Doubles storage space on PACS – Doubles bandwidth requirements – Slows everything down • Processed images (“ Presentation”) (“For Presentation” – Image is processed prior to archival – Further processing is not possible (W/L only) – Images in clinics look good MDACC Imaging Physics AAPM 2005 QC - Acquisition Device and Operation AAPM Report #74 • • • • • • • • AAPM 2005 Chapter 1 – The QC process Chapter 2 – QC Instrumentation Chapter 3 – The Physics Report Chapter 4 – Repeat Rate Analysis Chapter 5 – Radiographic units Chapter 7 – Conventional Tomography Chapter 8 – Portable XX-Ray Chapter 14 – Photostimulable Phosphor Systems MDACC Imaging Physics MDACC Imaging Physics AAPM 2005 Quality Control Equipment selection • Match performance and configuration to clinical needs Equipment QC • Acquisition Devices • Display Devices Technical Operation Patient Data Integrity AAPM 2005 MDACC Imaging Physics 7 Quality Control … of the Equipment • Acquisition Device Quality Control … of the Equipment • Acquisition Device – X-Ray generator (AAPM Report 74) – Exposure index calibration – Exposure index dependence on beam quality – VOI recognition algorithms – Noise – Uniformity – Artifacts – Contrast sensitivity kV mA linearity Exposure time Collimation Focal Spots Grids Beam alignment – AEC (Constant raw SNR, or detector exposure) MDACC Imaging Physics AAPM 2005 AAPM 2005 Quality Control … of the Equipment • Acquisition Device MDACC Imaging Physics Quality Control … of the Equipment • Acquisition Device – DR systems – DICOM calibration at the QC console Detector sensitivity (SNR2 vs Exposure) Verify quarterly DR QC Workshop - J. Seibert, L. Goldman (1:15 PM WEWE-D-W-608608-1) Monthly Annual if selfself-calibrating – Mammography MQSA - Vendors QC Program – CR devices Erasure thoroughness AAPM Report #74, PSP System QC Siebert AJ, AAPM Monograph 20, 1994 Samei E, et al, Med Phys. 2001 AAPM 2005 MDACC Imaging Physics AAPM 2005 MDACC Imaging Physics 8 QC - Primary Displays AAPM OROR-3 Tests (Jerry Thomas and Mike Flynn, 8:30 tomorrow) • • • • • Luminance Uniformity Chromaticity Noise Luminance Response Diffuse and Specular Reflection • Pixel Defects AAPM 2005 Quality Control … of the Equipment • Film printers (AAPM Report #74) • Veiling Glare (CRT only) • Resolution (CRT only) • Geometric Distortion (CRT only) • Angular dependence of LR (@ +45 degrees from normal, LCD only) MDACC Imaging Physics – Density calibration (daily or as needed) – Density uniformity (monthly) – Geometric distortion (monthly) – Artifacts (monthly) – View boxes (annually) – SMPTE or TG18TG18-QC AAPM 2005 QC - Operation Repeat/Reject Rate Analysis and Exposure Index Logs • Expect same results as with Film/Screen • Films no longer available for counting • Software at console to track reasons for rejects and repeats – By tech – By anatomical view Easily accessible Formatted to facilitate interrogation AAPM 2005 MDACC Imaging Physics MDACC Imaging Physics QC - Data Integrity Modality Work List Management (MWLM) • Modality queries the RIS for a list of scheduled patients through a “Broker” Broker” • RIS returns the requested list (through the broker) with patient demographics • Automatically refreshes every 15 minutes • Tech selects patient from the list at the acquisition console • Exams are uniquely identified with “Accession Number” Number” for later pairing with reports AAPM 2005 MDACC Imaging Physics 9 QC - Data Integrity MWLM implementation • Opportunity for patient mismis-ID greatly diminished QC -Data Integrity Verification • Errors are still possible • Technologist or supervisor views EVERY exam on web viewer or QC workstation immediately after archival to verify status – Reiner B, et al, JDI, 2000 Overall patient ID failure rates decreased from 7.6% to 2% with introduction of MWLM in Baltimore VA’ VA’s CT operation – Missing images – Patient mismis-ID • Mandatory requirement for ALL modalities AAPM 2005 MDACC Imaging Physics MDACC Imaging Physics AAPM 2005 Summary Conversion to DR requires: • An inin-depth understanding of the technology behind CR, DR, PACS, RIS, workstations, and printers. • Understanding the impact of DR on workflow in exam rooms • Understanding the impact of DR on the QC program AAPM 2005 MDACC Imaging Physics Other Considerations • Reading room illumination • Prior, filmfilm-based comparisons with softsoft-copy • Primary interpretation display devices • • • • • AAPM 2005 Calibration QC Matrix size >3 MP? (Langer S, et al) Dead pixel definition M. Flynn - tomorrow, 8:55 MDACC Imaging Physics 10 Other Considerations • Detector latent image decay • DualDual-Energy subtraction, tomosynthesis, tomosynthesis, & image stitching • • • • Workflow? Optimization? Techniques? Latent image affects? Other Considerations • Film Digitizers • Calibration – Linear? – Barten? Barten? • QC • DICOM • Conformance statements • Workflow enhancements • Detector calibration uniformity • Uncorrected flatflat-fields AAPM 2005 MDACC Imaging Physics – Storage Commitment – Performed Procedure Step AAPM 2005 MDACC Imaging Physics Other Considerations Other Considerations The Role of the Physicist in Planning and Design of Digital Image Management Systems (PACS) – D. Peck, M. Flynn (7:30 AM MOMO-A-I-609609-1) Overview of Digital Detector Technology - J. Seibert (8:55 AM MOMO-B-I-618618-2) Characteristics and Performance Evaluation of Digital Image Displays - H. Roehrig (7:30 AM TUTUA-I-609609-1) Evaluating Digital Mammography Systems - E. Berns (7:30 AM TUTU-A-I-618618-1) Digital Image Processing in Radiography - D. Foos, Foos, X. Wang (7:30 AM WEWE-A-I-609609-1) Recent Advances in Digital Mammography - M. Yaffe, Yaffe, R. Jong (7:30 AM WEWE-A-I-618618-1) Testing Flat Panel Imaging Systems – What the Medical Physicist Needs to Know - J. Tomlinson (8:30 AM WEWE-B-I-618618-1) Digital Image Displays – Resolution, Brightness and Grayscale Calibration - M. Flynn (8:55 AM WEWE-B-I-618618-2) Design and Performance Characteristics of Digital Radiographic Receptors - J. Seibert (7:30 AM THTH-A-I-609609-1) AAPM 2005 MDACC Imaging Physics AAPM 2005 MDACC Imaging Physics 11 Other Considerations Display Evaluation Demonstration Workshop - E. Samei (1:30 PM MOMO-D-W-608608-1) DR QC Workshop - J. Seibert, L. Goldman (1:15 PM WEWE-D-W-608608-1) Bibliography Digital Imaging and Communications in Medicine (DICOM), National Electrical Manufacturer’ Manufacturer’s Association (NEMA), Rosslyn, Rosslyn, VA, 2000, (http://www.nema.org (http://www.nema.org)) Freedman M, Pe E, Mun SK, Lo SCB, Nelson M. The potential for unnecessary patient exposure from the use of storage phosphor imaging systems. Proceedings of the International Society for Optical Engineering. 1993;1897:4721993;1897:472-479. Gur D, Fuhman CR, Feist JH, Slifko R, Peace B. Natural migration to a higher dose in CR imaging. Eighth European Congress of Radiology; September 1212-17, 1993; Vienna. Abstract 154. AAPM 2005 MDACC Imaging Physics AAPM 2005 Bibliography Honea R, Blado ME, Ma Y. Is reject analysis necessary after converting to computed radiography? J Digit Imaging, v15 #2, Suppl 1 (May), 2002, 4141-52. Langer S, et al, Comparing the Efficacy of 5 MP CRT vs 3 MP LCD in the Evaluation of Interstitial Lung Disease, J Digit Imaging, v17 #3 (September), 2004, 149149- 157. Reiner B, Seigel E, Kuzmak P, and Severance S, Transmission Failure Rate for Computed Tomography Exams in a Filmless Radiology Department, J Digit Imaging, v13 #2, Supp 1 (May), 2000, 7979-82 MDACC Imaging Physics Bibliography Samei E, Seibert JA, Willis CE, Flynn MJ, Mah E, Junck KL. Performance evaluation of computed radiography systems. Med Phys. 2001;28:3612001;28:361-371. Samei E, et al, Assessment of Display Performance for Medical Imaging Systems, AAPM OnOn-Line Report #3 (OR(OR-3), AAPM, 2005. Seibert JA. Photostimulable phosphor system acceptance testing. In: Seibert JA, Barnes GT, Gould RG, eds. Specification, Acceptance Testing and Quality Control of Diagnostic XX-ray Imaging Equipment. Woodbury, NY: American Association of Physicists in Medicine; 1994:7711994:771800. Monograph No. 20. Shepard S J, et al, Quality Control in Diagnostic Radiology, AAPM Report #74 AAPM, 2002. AAPM 2005 MDACC Imaging Physics AAPM 2005 MDACC Imaging Physics 12 Bibliography Willis CE, Mercier J, Patel M. Modification of conventional quality assurance procedures to accommodate computed radiography. SCAR '96 Computer Applications To Assist Radiology. Great Falls, Va: Va: Society for Computer Applications in Radiology; 1996:2751996:275-281 Willis CE. Quality assurance: an overview of quality assurance and quality control in the digital imaging department. In: Quality Assurance: Meeting the Challenge in the Digital Medical Enterprise. Great Falls, Va: Va: Society for Computer Applications in Radiology; 2002:12002:1-8. AAPM 2005 MDACC Imaging Physics 13