Outline Basic Physics of PET and PET/CT Frederic H. Fahey DSc Children’s Hospital Boston Harvard Medical School frederic.fahey@childrens.harvard.edu Positron Emission • • • • • • • Basics of PET Scanner Design Data Acquisition Scintillation Materials Reconstruction Methods PET/CT Time-of-Flight (TOF) PET MicroPET Detector Ring 18F + 511 keV e511 keV 1 Detector Blocks (GE Advance NXi) PET Sinograms Sinogram PET Sinograms Brain Projection View Slice Angle Point Source Image for Each Angle Image for Each Slice Note: Sinograms and projection views are different ways or showing the same data. 2 True, Scatter and Random Coincidence Detections 18 17 16 15 14 13 True 12 11 10 9 8 7 6 Scatter 5 4 3 2 1 * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * 1 Random 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Span of 3 Michelogram (R = 2 τ N1 N2) 2D Detector * * * * * * * 17 * * * * * * * 16 * * * * * * * 15 * * * * * * * 14 * * * * * * * 13 * * * * * * * 12 * * * * * * * 11 * * * * * * * 10 * * * * * * * 9 * * * * * * * 8 * * * * * * * 7 * * * * * * 6 * * * * * 5 4 3 2 1 * * * * 1 2 3 4 5 6 7 8 9 10 11 12 13 14 * * * * * * 15 * * * * * 16 Span of 7 Michelogram Scintillator Crystals * * * * PMTs 17 18 Axial Direction * * * * * * * 18 End Shields Septa 511 kev photon Courtesy of M. Graham, M. Madsen, U Iowa 3 16 15 14 13 Acquisition Modes 12 11 10 9 8 7 6 5 2D 4 3 2 1 * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * 1 3D septa are removed * * * * * * * * * * * * * * * * 2 * * * * * * * * * * * * * * * * 3 * * * * * * * * * * * * * * * * 4 * * * * * * * * * * * * * * * * 5 * * * * * * * * * * * * * * * * 6 * * * * * * * * * * * * * * * * 7 * * * * * * * * * * * * * * * * 8 * * * * * * * * * * * * * * * * 9 * * * * * * * * * * * * * * * * 10 * * * * * * * * * * * * * * * * 11 * * * * * * * * * * * * * * * * 12 * * * * * * * * * * * * * * * * 13 * * * * * * * * * * * * * * * * 14 * * * * * * * * * * * * * * * * 15 Segment 2 16 Segment 1 Segment 3 GE 3D Projection view and Michelogram 11 * * * * * * * * * * * * * * * * * * 18 * * * * * * * * * * * * * * * * * 17 * * * * * * * * * * * * * * * * 16 * * * * * * * * * * * * * * * 15 * * * * * * * * * * * * * * 14 * * * * * * * * * * * * * 13 * * 10 * 9 * 8 * 7 * 6 * 5 * 4 * 3 * 2 * 1 * 12 11 1 2 3 4 5 6 7 7 9 * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * 8 * * * * * * * * * * * * * * * * * * 9 10 * * * * * * * * * * * * * * * * * * 11 5 * * * * * * * * * * * * * * * * * * 12 * * * * * * * * * * * * * * * * * 13 * * * * * * * * * * * * * * * * 14 3 * * * * * * * * * * * * * * * 15 3D PET 1 * * * * * * * * * * * * * * 16 * * * * * * * * * * * * * 17 * * * * * * * * * * * * 18 2 4 6 8 10 • • • • • Sensitivity drops off towards edges 4-5X increased sensitivity overall Increased scatter (15% to 40%) Increased randoms from out-of-field activity Rebinning algorithms to apply 2D reconstruction • Some devices can acquire in 2D or 3D whereas some can only acquire in 3D • 3D in Brain, 2D (or 3D) in Whole Body 4 Criteria for Scintillation Material Crystal Identification • Detection Efficiency (Stopping Power) – High Effective Z – High Density • Light Output – Good energy resolution – Good crystal identification • Decay Time – Reduction of random coincidences – Time-of-Flight PET Feedback Loop New Detector Materials SCINTILLATOR Rel. Light Output NaI(Tl) 100 BGO LSO GSO 15-20 75 20-25 Peak Wavelength (nm) 410 480 420 440 Decay Constant (ns) 230 300 12,42 30-60 Density (g/mL) 3.67 7.13 7.40 6.71 Effective Z 51 75 66 59 Index of Refraction 1.85 2.15 1.82 1.85 Hygroscopic ? Yes No No No Current Estimate Actual Projections Simulated Projections Use to improve current estimate Compare Backprojection Error Courtesy of Jerold W. Wallis, M.D. 5 Maximum Likelihood Reconstruction (ML-EM) aij is the probability that a photon emitted from pixelj is detected at projection bini. pixelj • Maximize the likelihood that the estimated activity distribution in the body (the reconstructed transaxial slices) would lead to the measured projections • Use the expectation maximization (EM) algorithm to iteratively estimate the activity distribution projection bini ML-EM Algorithm EM-ML Reconstruction • aij can contain physical information (effects of spatial resolution, scatter, attenuation…) • EM-ML algorithm takes into account the nature of the noise (quantum mottle) in the projection data • Can yield a more accurate reconstruction λ’(k)j = λ’(k -1)j Σ [aij {d/ Σ [aij] λ’(k -1)j }] Σ [aij] λ’(k)j is the newest estimate of the object pixel value λ’(k -1)j is estimate of the object at last iteration d is the projection data d’ = Σ [aij] λ’(k -1)j is the calculated projection data 6 ML-EM Algorithm (added corrections) Advantages of Iterative Reconstruction λ’(k)j = λ’(k -1)j Σ [aij {d/ (Σ [aij] λ’(k -1)j+acj(rj nj + cj) }] Σ [aij/nj acj] • Correct physics can be included in the reconstruction and Poission statistics • Attenuation correction • Reduction of streak artifact • Overall quality ac is photon attenuation data r is random data n is normalization data c is scatter data Accurate characterization of the point spread functions (PSFs) throughout the files of view allows for an accurate estimation of the “aij” matrix leading to improvement in spatial resolution and signalto-noise. Improved Image Quality with Accurate Specification Of the PSF Throughout the Field of View FBP Courtesy of Siemens Medical Solutions HD Courtesy of Siemens Medical Solutions 7 Disadvantages of Iterative Reconstruction OS-EM Algorithm • Ordered-subset expectation maximization • At each step, project and backproject at only some angles (i.e. a subset) • Perform the steps in an ordered way to include all angles • Data start to converge even before the 1st iteration is complete • Convergence achieved in 3 - 10 iterations • Computation time of a few minutes • Slow convergence to the desired solution (e.g. tens - hundreds of iterations) • Computationally demanding - number of iterations and inclusion of the physics Ways to Increase the Rate of Convergence • Stopping rules • Regularization • OS-EM OSEM Iterative Reconstruction OSEM Iterative Reconstruction First Subset Assume 80 angles (rows) Divided into 8 subsets Thus, 10 rows/subset. Angle Angle Assume 80 angles (rows) Divided into 8 subsets Thus, 10 rows/subset. Second Subset 8 OSEM Iterative Reconstruction Angle Assume 80 angles (rows) Divided into 8 subsets Thus, 10 rows/subset. Third Subset By the time one has processed all of the projections (rows) once, the estimate of the object has been updated 8 times. Single Slice Rebinning (SSRB) OSEM Iterative Reconstruction • Filtered Back-Projection – Fast – Robust – Subject to noise & streaks • OSEM – Almost as fast – Handles noise & streaks Example: 28 subsets (12 projections per subset) and 2 iterations) Fourier Rebinning • More accurate approach to rebinning • Better estimate of determining into which parallel plane oblique data should be placed • Based on the frequency-distance relationship (Value of Fourier transform of a sinogram receives contributions mainly from sources at a fixed distance t=-k/ ) 9 GE Discovery ST PET/CT PET/CT • State-of-the-art PET combined with state-of-the-art CT (up to 64 slice) • Anatomical correlation • CT-based attenuation correction CT PET Attenuation Correction PET PET-CT Attenuation Correction P2 = e-µ(L-x) X P1 = e-µx L PTOT = P1 x P2 = e-µL 10 PET-CT Attenuation Correction PET-CT Attenuation Correction • • • • • Dose from CT of PET-CT Acquire CT Scan and reconstruct Apply energy transformation Reproject to generate correction matrix Smooth to resolution of PET Apply during reconstruction Quality of CTAC CTADIvol (160 m A, 0.8 s , 1.5:1 pitch) 3000.00 CTADIvol (mrad) 2500.00 New Born 2000.00 1 Y ear Old 1500.00 5 Y ear Old 10 Y ear Old 1000.00 Med Adult 80 kVp 10 mA 0.5 s/rot 1.5:1 140 kVp 160 mA 0.8 s/rot 1.5:1 500.00 0.00 70 90 110 130 Tube V oltage (k V p) 150 ED from 14 mCi of FDG 1 rad 11 Effect of Patient Size 80 kVp, 10 mA, 0.5 s/rotation Quality of CTAC Accuracy of Atte nuation Correction w ith Patie nt Size 0.0920 80 kVp 10 mA 0.5 s/rot 1.5:1 140 kVp 160 mA 0.8 s/rot 1.5:1 Linear Attenuation Coefficient (cm-1) 0.0900 New born 0.0880 1 YO 5 YO 0.0860 10 Y O 15 Y O 0.0840 Small A dult Med Adult 0.0820 Large A dult 0.0800 0.0780 80/10/.5 PET-CT Scanners • • • • Siemens Biograph GE Discovery ST GE Discovery STE Philips Gemini PET-CT Scanners Detector Dimension (mm) # of PET Detectors PET Detector Material Spatial Resolution 2D/3D Atten Corr Detector Dimension (mm) # of PET Detectors PET Detector Material Spatial Resolution 2D/3D Atten Corr GE Discovery STE 4.7 x 6.3 x 30 13,440 BGO 5.0 2D/3D CT GE Discovery ST 6.2 x 6.2 x 30 10,080 BGO 6.1 2D/3D CT Philips Gemini 4 x 6 x 20 17,864 GSO 4.9 3D CT&Cs-137 Siemens Biograph LSO Siemens Hi-Rez LSO 6.5 x 6.5 x 25 4 x 4 x 20 9,216 23,336 LSO LSO 6.3 4.6 3D 3D CT CT 12 Time-of-Flight PET Time-of-Flight PET x = c t/2 Speed of Light Time (ns) Distance (cm) c=3x 1010 cm/s Where x is the time-of-flight spatial uncertainty and t is the timing resolution. 0.1 0.5 1.0 5.0 t (ns) 0.1 0.3 0.5 1.0 3 15 30 150 x (cm) 1.5 4.5 7.5 15.0 Time-of-Flight PET Time-of-Flight PET Assume t of 0.5 ns => x of 7.5 cm 13 Philips Gemini TF PET/CT scanner Time-of-Flight PET SNR Gain from Time-of-Flight PET D/1.6 x 2 D/ 1.6 c t PET scanner CT scanner 70-cm bore 18-cm axial FOV Brilliance 16-slice LYSO crystals where D is the diameter of the object D (cm) 20 30 40 SNR Gain 1.6 2.5 3.3 Intrinsic Energy resolution: 11.5% fwhm, Timing resolution: 585 ps NEMA NU-2 Spatial resolution: 4.8 mm at 1 cm, 5.2 mm at 10 cm Sensitivity: 6.6 cps/kBq Scatter fraction (at 440 keV): 27% for 20-cm x 70-cm Peak NEC: 125 kcps @ 0.42µCi/ml Courtesy of Joel Karp, PhD, Univ of Penn Measurements: 35-cm lesion phantom TOF non TOF Gemini TF 6-to-1 contrast 300s similar noise and scan time 180s TOF higher CRC 60s Courtesy of Joel Karp, PhD, Univ of Penn Heavy-weight patient study 13 mCi 2 hr post-inj 3 min/bed Colon cancer 119 kg BMI = 46.5 MIP LDCT non-TOF TOF Courtesy of Joel Karp, PhD, Univ of Penn 14 non-Hodgkin’s lymphoma 136 kg (45 BMI) Courtesy of Joel Karp, PhD, Univ of Penn nonTOF 30 min TOF 30 min TOF CHB MicroPET 10 min TOF tumor contrast (SUV) higher than non-TOF by 1.5 • • • • Installed 8/16/06 Siemens Focus 120 1.5 mm spatial resolution 7 % Sensitivity TOF tumor contrast superior to non-TOF for 10 min as well as 30 min scan Summary • Modern scanners designed for oncologic imaging • Practically all PET sales are PET/CT scanners • New scintillation crystals combine excellent detection efficiency with short decay times • Appropriate choice of reconstruction algorithm can lead to better image quality • Shorter decay times leads to possibility of time-offlight PET. • MicroPET scanners can provide very high spatial resolution with high sensitivity in a small foot print and easy access to the research animals. Now for the SAM questions!! Acquiring a PET scan in 3D mode leads to 0% 0% 0% 0% 1. 2. 3. 4. Shorter reconstruction times Less scatter and random coincidences Improved spatial resolution Improved sensitivity 10 15 Time-of-flight PET improves image quality, particularly in Answer 4. Improved sensitivity 0% 0% Reference: Cherry SR, Sorenson JA, Phelps ME, “Physics in Nuclear Medicine, Saunders, Philadelphia PA. page 351-352. Explanation: 3D PET has approximately the same spatial resolution as 2D PET with increased scatter and random coincidences. It is also more complicated to reconstruct. However, the removal of the inter-plane septa leads to a large increase in the lines-of-response (LORs) and thus a 4 to 5 fold increase in sensitivity. 0% 0% 1. 2. 3. 4. Fast scans Larger patients Small animals Time sequences of scans 10 Answer 2. Larger Patients Reference: Suleman Surti, Austin Kuhn, Matthew E. Werner, Amy E. Perkins, Jeffrey Kolthammer and Joel S. Karp, “Performance of Philips Gemini TF PET/CT Scanner with Special Consideration for Its Time-of-Flight Imaging Capabilities”, J Nucl Med 48: 471-480, 2007. Explanation: Time-of-flight PET involves both data acquisition and reconstruction. It allows one to locate the annihilation event along the line-of-response which leads to reduced noise and higher contrast in the image data. With current technology, the event can be localized to within about 7-8 cm which leads to better image improvement in larger patients but not for small patients or small animals. It is not at all related to the duration of the scan or whether the study is acquired as a time sequence. 16