Quality Assurance for image-guidance technologies J.-P. Bissonnette, D. Moseley, T. Purdie, M. Sharpe, D. Jaffray IGRT with Cone-beam CT • Introduction • Geometric QA • Image quality • Patterns of failure Cone-beam CT: Infrastructure QA • Safety • Geometric • System stability • Image quality • System infrastructure • Dose Image-guidance: MV Geometry • Imaging and treatment beams coincide Image-guidance: MV/kV Coincidence • Treatment is orthogonal to imaging kV/MV Calibration Concept BB (Reconstruction Iso-centre) MV Mechanical isocentre kV y MV Radiation isocentre x Calibrated isocentre z Geometric Calibration • Analogous to the Winston-Lutz test used for brain stereotactic QA – Lutz, Winston, & Maleki, IJROBP 14, pp. 373-81 (1988) 1. MV Localization (0o) of BB; collimator at 0 and 90o. 2. Repeat MV Localization of BB for gantry angles of 90o, 180o, and 270o. 3. Analyze images and adjust BB to Treatment Isocentre (± 0.3 mm) +1mm θgantry θgantry u -1mm -180 v θgantry +180 Reconstruction 4. Measure BB Location in kV radiographic coordinates (u,v) vs. θgantry. 5. Analysis of ‘Flex Map’ and Storage for Future Use. 6. Employment of ‘Flex Map’ During Routine Clinical Imaging. MV/kV Calibration Procedure v u Flexmap • A plot of the apparent travel of a point as a function of gantry angle. • Removes the effect of component flexes and torques prior to reconstructions. • Ties the 3D image matrix to the radiation isocentre of the accelerator. Flexmap Elekta • Flexes are corrected for in reconstruction algorithm. Varian • Flexes are compensated for by robotic arm motion. Flexmap Elekta + “Burns” radiation isocentre into CBCT dataset - Vulnerable to file corruption Varian + No software vulnerability - Robotic arm motions need monitoring Results for Six Units 5 4.5 4 3.5 Absolute U displacement (mm) 3 2.5 2 1.5 Unit A 1 Unit B 0.5 Unit C 0 Unit D -0.5 Unit F -1 Unit G -1.5 -2 -2.5 -3 -3.5 -4 -4.5 -5 -180 -135 -90 -45 0 Gantry angle (degrees) 45 90 135 180 Residual Error 1.50 Residual U displacement (mm) 1.25 1.00 0.75 0.50 Unit A Unit B Unit C Unit D Unit F Unit G 0.25 0.00 -0.25 -0.50 -0.75 -1.00 -1.25 -1.50 -180 -135 -90 -45 0 45 Gantry Angle (degrees) 90 135 180 Results for Six Units 5 4.5 4 3.5 3 Absolute V displacement (mm) 2.5 2 1.5 Unit A 1 Unit B 0.5 Unit C 0 Unit D -0.5 Unit F -1 Unit G -1.5 -2 -2.5 -3 -3.5 -4 -4.5 -5 -180 -135 -90 -45 0 Gantry angle (degrees) 45 90 135 180 Residual Error Residual V displacement (mm) 3 2 Unit A Unit B Unit C Unit D Unit F Unit G 1 0 -1 -2 -3 -180 -135 -90 -45 0 45 Gantry Angle (degrees) 90 135 180 Long-term Stability: Flexmap 1.25 1 R e s id u a l d is p la c e m e n t (m m ) 0.75 v 0.5 0.25 u 0 -0.25 -0.5 12 calibrations over 28 months -0.75 95% confidence interval = 0.25 mm -1 -1.25 -1.5 -180 -135 -90 -45 0 Gantry angle (degrees) 45 90 135 180 Effect of Incorrect Calibration Lessons Learned: Geometric Calibration • Takes ~ 2 hours to perform per unit • Flexmaps are stable on Elekta – Accruing long term data on Varian platform • Residual flexmaps are comparable on both platforms – Ultimately affects geometric accuracy of volumetric imaging • Considering changing test frequency from monthly to semi-annually Daily Geometry QA • Align phantom with lasers • Acquire portal images (AP & Lat) & assess central axis • Acquire CBCT • Difference between predicted couch displacements (MV & kV) should be < 2 mm http://www.modusmed.com/igrt.htm Daily Geometry QA • Align phantom with lasers • Acquire portal images (AP & Lat) & assess central axis • Acquire CBCT • Difference between predicted couch displacements (MV & kV) should be < 2 mm 1. Shift BB embedded in cube from isocentre. 2. MV Localization of BB for gantry angles of 0o and 90o. θgantry Reconstruction 3. kV Localization with cone-beam CT 4. Compare kV and MV localizations; tolerance is ± 2 mm 5. Use automatic couch to place BB to isocentre; verify shift with imaging Compare Portal Image & DRR 0.25 X Y Z Deviation from isocentre (cm) 0.20 0.15 0.10 0.05 0.00 -0.05 -0.10 -0.15 -0.20 -0.25 A B C Unit D G Lessons Learned: Daily QA • In a single set-up, therapists: – Warm-up tube – Software/hardware integrity – Sufficient disk/storage space – Lasers, ODI, field size indicators, etc. – QC of image-guided process • Takes ~ 20 mins in the morning Cone-beam CT: QA of a Device • Safety • Geometric • System stability • Image quality • System infrastructure • Dose Image Quality CatPhan 500 phantom Scale • Geometric calibration to tie isocentre to centre of volumetric reconstruction • Scale to relate all pixels to isocentre 5 cm Scale • Geometric calibration to tie isocentre to centre of volumetric reconstruction • Scale to relate all pixels to isocentre Linearity of CT Numbers Linearity of CT Numbers 2000 1800 Measured Hounsfield unit 1600 1400 Unit 7 Unit 8 1200 Unit 9 Unit 10 1000 Unit 12 Unit 16 800 Unit 16 with annulus Unit 17 600 400 200 0 0 200 400 600 800 1000 1200 Theoretical Housfield unit 1400 1600 1800 2000 Linearity of CT Numbers Mean Elekta Mean Varian Identity 2000 1800 Measured Housfield Units 1600 1400 1200 1000 800 600 400 200 0 0 200 400 600 800 1000 1200 Theoretical Housfield Units 1400 1600 1800 2000 Add Scatter IEC standard 61675-1 Linearity of CT Numbers Mean Varian Identity Varian with annulus 2000 1800 Measured Housfield Units 1600 1400 1200 1000 800 600 400 200 0 0 200 400 600 800 1000 1200 Theoretical Housfield Units 1400 1600 1800 2000 Linearity of CT Numbers • Fairly linear (χ2 > 0.99) for all systems • Beam hardening • Scatter conditions • Non-standard metric; use only as a baseline Spatial Resolution 1.2 Unit A Unit B Unit C Unit D Unit E Unit F Unit G Unit H Unit I Unit J 1.0 MTF 0.8 0.6 0.4 0.2 0.0 0 2 4 6 8 -1 Spatial frequency (cm ) Droege, Radiology 146, pp. 244-246 (1983) 10 Spatial Resolution 1.2 Mean Elekta Mean Varian 1.0 1.4 mm MTF 0.8 0.6 1.2 mm 0.4 0.2 0.0 0 2 4 6 8 -1 Spatial frequency (cm ) 10 12 Effect of Scatter on MTF Acceptance Narrow x-ray field Window & level identical Spatial Resolution 1.2 Mean Elekta Mean Varian Unit E small field 1.0 0.8 mm MTF 0.8 0.6 1.2 mm 0.4 0.2 0.0 0 2 4 6 8 -1 Spatial frequency (cm ) 10 12 Image Quality: Lessons Learned • Scale is accurate • CBCT affected by beam hardening and scatter – CT linearity and MTF curves differ for individual units – Reasonable changes in scatter conditions affect curves • Track as baseline; CBCT not ready for quantitative CT. Patterns of Failure: Truncation Truncation artefact, Zhang, IJROBP 2005 Patterns of Failure: Ring Artefacts Patterns of Failure: Capping Artefact Patterns of Failure: Streaking Patterns of Failure: Motion Artefacts Conclusions • Geometric calibration can be within 0.25 mm • Daily QA tolerance ± 2 mm • Value if image quality QA? – Debatable, except for scale – Stick to baseline values for now – should improve in the future Conclusions • Spatial resolution, contrast & uniformity – Depend on unit and imaging conditions • Scatter environment • Imaging technique • Beam hardening • Artefacts can hint at defective components or reveal limits of the CBCT physics Daily CBCT QA Program Dimension Procedure Tolerance Detector stability Dark image calibration Geometry Localising lasers < 1 mm MV/kV/laser alignment ± 2 mm Accuracy of shifts ± 2 mm Safety Warm-up Interlocks: interrupts or prevents irradiation Functional Warning lights Functional Generator operation Functional Detector operation Functional Detector signal Collimator operational Clinical process issues Database integrity Storage space availability Within expected range Functional Monthly CBCT QA Program Dimension Procedure Tolerance Imaging system Gain stability Replace or refresh performance Defect maps Replace or refresh Image quality Scale and distances ± 0.5 mm CT number linearity & stability Baseline Image uniformity Baseline High contrast spatial resolution Baseline Artefacts Absence Geometric Geometric calibration Accuracy of couch shifts Clinical process issues Review of daily test results Replace / refresh < 1 mm Annual CBCT QA Program (service) Dimension Procedure Tolerance X-ray generator kVp accuracy Baseline stability mAs linearity Baseline Radiation quality (HVL) Baseline Accuracy of mA & mAs Baseline Geometry Couch scales 1 mm Couch motion accuracy (manual or remote) 1 mm Detector tilt Baseline Detector skew Baseline Detector scale Baseline Annual CBCT QA Program (upgrades) Dimension Data transfer Procedure Link to treatment planning Tolerance Functional and accurate Long term and short term storage Functional Dosimetry Axial and skin doses Baseline Clinical process issues Database integrity and maintenance Baseline Documentation of imaging procedure Up-to-date Review of daily and monthly test results Completeness