AAPM Task Group 106 Accelerator Beam Data Commissioning Equipment and Procedures New Measurement Dosimetry – New Developments John P. Gibbons, Jr., Ph.D. Med Phys 35 (9): 4186-4210 (2008) Relative Dose Measurements Relative Dose Measurements Beam Scanning Beam Scanning Beam Scanning System Setup: • Water Phantom Scanning Arm Tilt – Water conditions • Water temperature is equilibrated • Periodically check for evaporation – Leveling • Level (horizontal) scanning arm, not tank • Can verify visually, or scan in gradient region (e.g., R50) of low-energy electron beam • Verify vertical (depth) scans are parallel to central axis Das et al, TG-106 1 Relative Dose Measurements Relative Dose Measurements Beam Scanning Beam Scanning Gantry Tilt - Photons Gantry Tilt - Electrons Das et al, TG-106 Das et al, TG-106 Relative Dose Measurements Relative Dose Measurements Beam Scanning Electrometer amplification Comparison of chamber and bias Beam Scanning System Setup: • Scanning System 110 100 6 MV bad chamber, incorrect gain – Range of motion sufficient (Offset tank scans for 40 x 40 cm2 scans) – Sufficient scatter material for field sizes and depths measured – Detectors positioned appropriately – Detection electronics adjusted for measurement conditions 6 MV bad chamber, correct gain 15 MV good chamber 15 MV bad chamber, incorrect gain 80 15 MV bad chamber, correct gain Percent Depth Dose (TG40: +1mm over 300 mm movement) 6 MV good chamber 90 – Detector motion calibrated 70 60 50 40 30 20 10 0 0 5 10 15 20 25 30 Depth(cm) Das et al, TG-106 2 Relative Dose Measurements Relative Dose Measurements Beam Scanning - Photons Beam Scanning - Photons Percent Depth Dose scanning: • Corroborate with integrated measurements • Adjust for effective point of measurement • Should ensure sufficient scatter for measured depths • Detector alignment should be verified for wedged fields Percent Depth Dose scanning: • Corroborate with integrated measurements • Adjust for effective point of measurement • Should ensure sufficient scatter for measured depths • Detector alignment should be verified for wedged fields Relative Dose Measurements Relative Dose Measurements Cylindrical chamber shift Beam Scanning - Photons Percent Depth Dose scanning: • Corroborate with integrated measurements • Adjust for effective point of measurement • Should ensure sufficient scatter for measured depths • Detector alignment should be verified for wedged fields Almond et al., AAPM TG-51, Med Phys 26:1854 (1999) 3 Relative Dose Measurements Relative Dose Measurements Detector alignment for wedged beams Beam Scanning - Photons Beam profile scanning: • Should ensure sufficient scatter for measured field sizes • Penumbra dependent on detector size • Scanning speed should not disturb surface • Careful with electronic amplification for gradient fields CAX Scanning direction Relative Dose Measurements Relative Dose Measurements Beam Scanning Beam Scanning - Photons Scanning Speed Beam profile scanning: • MLC-defined fields: – Detector volume centered under leaf center – Avoid scanning abutted or close leaves • Diagonal profiles: – Confirm calibration of detector movement – Scatter material more important – Consider detector orientation • Special fields (e.g., blocks, compensators) – Avoid scanning through mounting tray holes, slots, etc. Das et al, TG-106 4 Relative Dose Measurements Relative Dose Measurements Beam Scanning - Electrons Beam Scanning - Electrons Percent Depth Dose scanning : • Ionization chamber data require corrections (stopping power ratio, prepl) • Depth accuracy important Determination of Surface Position – R50 determines energy and calibration – Effective point of measurement more critical – Water ripple has more effect Das et al., Phys Med Phys 43(11) : 3419 (1998) Relative Dose Measurements Beam Scanning - Electrons Relative Dose Measurements Beam Scanning - Electrons Effect of Scanning Speed - Electrons Das et al, TG-106 5 Relative Dose Measurements Relative Dose Measurements Integrated Measurements - Photons Integrated Measurements - Photons General Considerations: • Choice of Normalization Depth – All quantities should be measured at this depth • Periodically repeat normalization readings Output factor measurements (Scp): • Phantom large enough for larger fields • Detector small enough for smaller fields • Should verify minimal stem effect, especially for smaller volume, non-farmer type chambers Relative Dose Measurements Relative Dose Measurements Integrated Measurements - Photons Integrated Measurements - Photons Sc Collimator Exchange Effect In-Air output ratio measurements (Sc): • Mini-Phantoms recommended • Collimator exchange effect: – Defined as Sc(a,b) Sc(b,a) – Verify small (<2%) for range of clinical field sizes Lam et al., MU Calculations for Photon and Electrons Fields, 2000 6 Relative Dose Measurements Relative Dose Measurements Integrated Measurements - Photons Integrated Measurements - Electrons Transmission factor measurements (WF, TF, etc.): • Should be measured at normalization depth Output factor measurements (Se): • Function of applicator size and jaw setting, so record and periodically verify both. • Should verify both wedge orientations, especially for steeper wedges • Careful with wedge factor normalization. • Account for the effective point of measurement of the detector • Should be measured at R100, which may vary with applicator size. • If depth changes between measured and reference applicator, should convert ionization to dose for each measurement. (S/p is a function of depth) – TPS may require WF = Dw(r)/Do(10x10) • Wedges may have depth and/or field size dependence Conclusions • TG-106 gives specific recommendations for determining relative doses via scanning or integrated measurements •A variety of simple tests should be performed on the beam scanning systems prior to each measurement to ensure correct setup. • Integrated measurements should be performed using an appropriate detector and at the correction depth. 7