Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Update on Fluoroscopy Physics • Understand dosimetric concepts relating to interventional fluoroscopy • Characterize the dosimetric features and performance of a modern fluoroscope • Be able to set up a clinical dose recording and reporting policy that will meet clinical and JC requirements. Stephen Balter, PhD Columbia University Draft for MO-A-210A-1 2009 AAPM SB0907 UFP AAPM - 1 Educational objectives © S. Balter 2009 SB0907 UFP AAPM - 2 © S. Balter 2009 Outline Notation • ICRU diagnostic dosimetric quantities and their fluoroscopic extensions • Construction, dosimetric features, and performance characteristics of modern fluoroscopes • Extended QA protocols for compliance measurements, system characterization and clinical dosimetry. • Dose recording and reporting • The Joint Commission fluoroscopy sentinel event. • This presentation uses the notation given in ICRU-74 (2006) • The notation has been extended to encompass additional quantities specifically related to fluoroscopy • Organizations that have accepted or are in the process of accepting this notation include the IAEA, IEC, NCRP and NEMA. SB0907 UFP AAPM - 3 © S. Balter 2009 Air KERMA (Ka) © S. Balter 2009 From the lab to the clinic • Measurements are made under lowscatter conditions • Metric usually used by calibration laboratories to report instrument factors. • Unless the detector is air, there may be additional corrections needed to account for instrument construction and X-ray spectrum. SB0907 UFP AAPM - 5 SB0907 UFP AAPM - 4 © S. Balter 2009 • Ka,i is the air kerma at the location of the patient’s skin in the absence of the patient. • Ka,e is the air kerma at the location of the patient’s skin with the patient present. SB0907 UFP AAPM - 6 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Backscatter factors Ka,i → Ka,e Reference Point Air Kerma (Ka,r) • Reference point originally defined in IEC 60601-2-43 (2000) • Included in 2005 FDA fluoro regulations. • Intent is to approximate the location of the patient’s skin • Reported as Ka without scatter • IEC, FDA ± 35 % SB0907 UFP AAPM - 7 © S. Balter 2009 Reference Point Air Kerma (Ka,r) SB0907 UFP AAPM - 8 Reference Point ≠ FDA Point • Total air kerma accumulated at the reference point from the beginning of the procedure. (called “cumulative dose” in the regs) • Displayed to operators at the working position and in the control room. • Defined under low scatter conditions. • Table top and mattress attenuation? Image Receptor 30 cm © S. Balter 2009 Reference point ≈ skin Isocenter SID = Any 15 cm – Not standardized. – Propose measurement without attenuation unless the attenuators are always in the beam. SB0907 UFP AAPM - 9 © S. Balter 2009 IEC & FDA Ka,r Reference Point FDA “Dose” Compliance Point Focal Spot SB0907 UFP AAPM - 10 © S. Balter 2009 Conversion of Ka,r to Dskin • Dmax,skin is seldom numerically equal to Ka,r • For teaching purposes, assume that the beam does not move during the procedure. Ka,i = Ka,r * (SRD/SSD)2 where SRD is the source to reference point distance Ka,e = Ka,i * Backscatter f(kVp, filter, field size) Dskin = Ka,e * (µskin / µair) SB0907 UFP AAPM - 11 © S. Balter 2009 SB0907 UFP AAPM - 12 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Kerma Area Product (PKA) PKA is independent of distance! • Total KAP accumulated from the beginning of the procedure. • Displayed to operators at the working position and in the control room. • Can be used to estimate Ka,r • “Where should DAP be measured?” • Area = a * d2 • Output = k *d-2 (need to know field size at reference point) • Defined under low scatter conditions. • Table top and mattress attenuation? SB0907 UFP AAPM - 13 © S. Balter 2009 • PKA = a * d2 * k*d-2 =a*k SB0907 UFP AAPM - 14 © S. Balter 2009 Outline Geometry (generic system) • ICRU diagnostic dosimetric quantities and their fluoroscopic extensions • Construction, dosimetric features, and performance characteristics of modern fluoroscopes • Extended QA protocols for compliance measurements, system characterization and clinical dosimetry. • Dose recording and reporting • The Joint Commission fluoroscopy sentinel event. • • • • © S. Balter 2009 Radiation technique range • • • • SB0907 UFP AAPM - 16 © S. Balter 2009 Table-Top Outputs FP Ka,i range <1 – 4,000 mGy/min kVp 60 – 120 Fixed filtration 2 – 6 mm Al Added Filter 0.0 – 0.9 mm Cu 10000 Chamber 30 cm from FP Minimum SID 1000 – Brand A : Added Cu fluoro filter independent of kVp – Brands B,C: Added Cu fluoro filter tends to decrease as kVp increases – All Brands: Less or no Cu for acquisition modes • New systems do not provide physicist level control of basic X-ray factors 25NF 20NF 15NF 25CI 20CI 15CI mGy/min SB0907 UFP AAPM - 15 Min SSD = 38 cm (<30 cm) SID 80 cm – 125 cm SAD = 70 - 80 cm (focus to isocenter) Table top height - 30 to + 20 cm (relative to isocenter) • 110º LAO – 110º RAO • 45º CRA – 45º CAU • FS ≈ 0 – 2000 cm2 (at image receptor) 100 10 38 mm Al +0.5 mm Cu +1.0 mm Cu +1.5 mm Cu +2.0 mm Cu +3 mm Cu +4 mm Cu Attenuator SB0907 UFP AAPM - 17 © S. Balter 2009 SB0907 UFP AAPM - 18 6 mm Cu +10 mm Cu 0902D © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Observed values SB0907 UFP AAPM - 19 Dose rates are down; BUT!!! © S. Balter 2009 SB0907 UFP AAPM - 20 © S. Balter 2009 Conventional Wisdom Source to Image Receptor Tracking • FDA limits normal-mode table-top outputs to 87 mGy/min (10 R/min) • Using more magnification increases patient skin-dose-rate • Using more magnification increases limiting spatial resolution • FDA regulations limit normal mode fluoroscopy to less than 87 mGy/min at 30 cm from the image receptor. • Source to image receptor distance (SID) is not mentioned in the regulations. • Most interventional fluoroscopes limit output to less than 87 mGy/min at any SID • What happens at the table top when SID is varied? SB0907 UFP AAPM - 21 © S. Balter 2009 Minimum SID Table-Top Ka,i SB0907 UFP AAPM - 22 © S. Balter 2009 Maximum SID Table-Top Ka,i Minimum SID Maximum SID 30 cm Flu 75 mGy/m Flu 230 mGy/m Cin 1700 mGy/m 30 cm Flu 75 mGy/m Cin 1700 mGy/m SB0907 UFP AAPM - 23 © S. Balter 2009 SB0907 UFP AAPM - 24 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Low Table/Max SID Table-Top Ka,i Table Top Maximum Air Kerma Rate Maximum SID Relative Table Top Output 2.75 30 cm Flu 75 mGy/m Flu 270 mGy/m Cin 2000 mGy/m Table Top (chamber) fixed 55 cm from FS 2.25 Fluoroscopy Table Top Rate Inverse Square 1.75 Cine Table Top Rate 1.25 0.75 85 90 95 100 105 110 115 Source to Image Receptor Distance (cm) Minimum SSD SB0907 UFP AAPM - 25 © S. Balter 2009 Effect of zoom on EERII SB0907 UFP AAPM - 26 120 0907 Alpha © S. Balter 2009 Field of view considerations • An image intensifier zooms a variable size input field onto a constant output field. Unmodified: Dose per frame ≈ 1/ FOV2 Programmed: Dose per frame ≈ 1/ FOV • A first generation flat-panel electronically zooms a variable number of pixels onto a full size image display. Dose per frame ≈ constant Dose per frame ≈ 1/ FOV Dose per frame ≈ 1/ FOV2 SB0907 UFP AAPM - 27 © S. Balter 2009 6 Base 38 Al 5 Normalized Phantom Input (fluoroscopy) Program kVp Beam filter Dose Rate Constraint © S. Balter 2009 Image intensifier zoom Measured Phantom EER vs. f(FOV) • • • • SB0907 UFP AAPM - 28 4 0.5 Cu 1.0 Cu 2.0 Cu 4.0 Cu 8.0 Cu 1/r 1/r2 3 2 1 48 cm 42 cm 32 cm 22 cm 16 cm 11cm Field of View - Flat Panel 0805 BETA SB0907 UFP AAPM - 29 © S. Balter 2009 SB0907 UFP AAPM - 30 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Pixel size – image intensifier zoom DR – Indirect Detector X-ray Structured X--ray phosphor (CsI) Light S G + D Photodiode Storage capacitor Charge X-rays to light to electrons to electronic signal: a - Si TFT/ CsI phosphor SB0907 UFP AAPM - 31 © S. Balter 2009 Pixel size: flat panel zoom (1st gen) SB0907 UFP AAPM - 33 © S. Balter 2009 Pixel size: zoom comparison SB0907 UFP AAPM - 35 © S. Balter 2009 SB0907 UFP AAPM - 32 © S. Balter 2009 Smaller FOV = Fewer Pixels SB0907 UFP AAPM - 34 © S. Balter 2009 Rebinning SB0907 UFP AAPM - 36 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Display Outline • Display Processing • ICRU diagnostic dosimetric quantities and their fluoroscopic extensions • Construction, dosimetric features, and performance characteristics of modern fluoroscopes • Extended QA protocols for compliance measurements, system characterization and clinical dosimetry. • Dose recording and reporting • The Joint Commission fluoroscopy sentinel event. – Zoomed FP images are digitally transformed from the detected pixel matrix to the full display matrix. – All images are highly processed to enhance perception of detail. • Dose per frame is programmed by the system to standardize the perception of noise. • These measures are designed to maximize perception – High Contrast Objects (small interventional devices < 1 mm) – Low Contrast Objects (relatively large tissue blush > 5 mm) • Different acquisition and display optimization is needed for different clinical procedures. SB0907 UFP AAPM - 37 © S. Balter 2009 What instrument? SB0907 UFP AAPM - 38 © S. Balter 2009 Reference Point Locations • If you are super-fussy (and want to try to stay below the ICRU-74 tolerance of 7%) the functionally closer to a free air chamber the better. • For most work, energy compensation of solid-state detectors is adequate Image Receptor 30 cm SID = Any Absolute accuracy should improve when appropriate calibration beams are defined and implemented. • Consider the influence of any instrument on the fluoroscope’s ADRC. • Caution: You, your service engineers, and your regulators may be using different types of instrumentation. SB0907 UFP AAPM - 39 © S. Balter 2009 SB0907 UFP AAPM - 41 IEC & FDA Ka,r Reference Point FDA “Dose” Compliance Point Focal Spot SB0907 UFP AAPM - 40 © S. Balter 2009 Compliance Equipment placed on table top Attenuators 15 cm above ion chamber Minimum SID Adjust table height for 30 cm between center of chamber to entrance surface of image receptor • Collimate to just inside smallest FOV • Test ‘common’ modes at all FOVs. • • • • 50% 30 cm minimum SID Compliance Setup Isocenter 15 cm © S. Balter 2009 SB0907 UFP AAPM - 42 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Attenuator protocol • • • • • Acquisition modes • There is reluctance to test acquisition modes at full output due to concerns about damaging the fluoroscope. 19 mm Al (children) 38 mm Al (small adult) 38 mm Al + 0.5 mm Cu (medium adult) 38 mm Al + 2.0 mm Cu (large adult) 38 Al + 2 Cu + 3 mm Pb (maximum) or 38 mm Al + 8-10 mm Cu (maximum) This is reflected in regulatory requirements • Modern systems are actually clinically operated at or near full output a significant fraction of the time. Maximum outputs are often more than double the 38 Al + 2 Cu outputs SB0907 UFP AAPM - 43 © S. Balter 2009 SB0907 UFP AAPM - 44 Clinical kVp in one laboratory © S. Balter 2009 Typical compliance data Compliance Image Intensifier - 17 cm FOV FDA Geometry Cine n = 1856 Fluoro n = 6594 1000 mGy/min 110 FL FN CL CN 100 100 kVp 10 fKVP cKVP 90 Display = Cine actual/2, fC actual mGy/min 10000 120 1000 100 10 cine S fluoro 1 0 1 38 Al + 0.5 Cu + 1 Cu + 2 Cu + 4 Cu + 8 Cu + 8Cu+3Pb Attenuator 2 4 6 8 10 38 mm Al + X mm Cu 80 Flat Panel Detector Image Intensifier 70 60 1% 5% 10% 20% 25% 30% 40% 50% 60% 70% 75% 80% 90% 95% Both have variable Cu beam filters 99% Percentile SB0907 UFP AAPM - 45 © S. Balter 2009 “Dose” instrumentation SB0907 UFP AAPM - 46 © S. Balter 2009 Dose displays in the lab • Locations – Integrated – Interfaced – External • Measurement technologies – None – Calculated from settings – PKA chamber + calculation – Ka,r chamber + calculation – PKA & Ka,r dual channel chamber SB0907 UFP AAPM - 47 The same RPDose will display as 5678 on all three systems The same KAP will display as 12345 on GE and SIEMENS and 123450 on PHILIPS © S. Balter 2009 SB0907 UFP AAPM - 48 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Verification of displayed “dose” Stability of Ka,r instruments • Accuracy of display AVE – IEC ± 50% (RPDose &KAP) – FDA ± 35% (RPDose) – Stability usually better A B C DA DB E F GB GA • Usually validated at factory • Seldom validated by installers • When verified as part of QA constancy should be ± 5% SB0907 UFP AAPM - 49 © S. Balter 2009 Ka,i and PKA at isocenter SB0907 UFP AAPM - 51 0.84 0.85 0.85 0.87 0.88 0.91 0.98 1.17 1.18 SB0907 UFP AAPM - 50 © S. Balter 2009 Field size measurement © S. Balter 2009 Outline SB0907 UFP AAPM - 52 © S. Balter 2009 Skin dose map and beam position • ICRU diagnostic dosimetric quantities and their fluoroscopic extensions • Construction, dosimetric features, and performance characteristics of modern fluoroscopes • Extended QA protocols for compliance measurements, system characterization and clinical dosimetry. • Dose recording and reporting • The Joint Commission fluoroscopy sentinel event. SB0907 UFP AAPM - 53 STDEV Jul-Dec 05Jan-Jun 06Jul-Dec 06Jan-Jun 07Jul-Dec 07 0.03 0.85 0.82 0.80 0.84 0.89 0.04 0.82 0.82 0.84 0.89 0.89 0.04 0.85 0.80 0.85 0.88 0.89 0.03 0.85 0.82 0.86 0.90 0.90 0.04 0.85 0.86 0.86 0.93 0.92 0.01 0.89 0.91 0.91 0.93 0.92 0.06 0.89 0.99 0.96 1.04 1.02 0.05 1.09 1.16 1.21 1.19 1.18 0.04 1.14 1.13 1.22 1.21 1.18 © S. Balter 2009 Illustration courtesy of Siemens SB0907 UFP AAPM - 54 © S. Balter 2009 © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. What can be measured in the lab? Dose displays in the lab • Fluoroscopy time • Number of acquisition – Runs – Frames • • • • Kerma Area Product PKA Reference Point Air Kerma Ka,r Peak Skin Dose Dose maps SB0907 UFP AAPM - 55 The same RPDose will display as 5678 on all three systems The same KAP will display as 12345 on GE and SIEMENS and 123450 on PHILIPS © S. Balter 2009 In the control room SB0907 UFP AAPM - 56 © S. Balter 2009 SIR Standard of Practice • Record all available data for every interventional procedure • Order of Precedence – – – – Skin dose map (not currently available) Reference Point Air Kerma Ka,r Kerma Area Product PKA Fluoro time and number of images • Uses for the data – – – – SB0907 UFP AAPM - 57 © S. Balter 2009 Patient pre discharge instruction Dose reconstruction when necessary Clinical QA Physics QA against guidance levels SB0907 UFP AAPM - 58 © S. Balter 2009 A New IHE Profile Manual log IHE Radiation Exposure Monitoring Profile Integration of systems reporting dose and systems which receive, store, or process those reports Modalities, PACS, RIS, Workstations, Registries • Technician hand writes data to log • Data retyped into clinical records or dose data base • Time lags and overflow can be problems • Process is error prone SB0907 UFP AAPM - 59 © S. Balter 2009 Facilitate compliance with Euratom 97/43, ACR Guidelines, etc. Directly based on DICOM Dose Reports Creation, Collection, Distribution, Processing IHE 2008 Webinar Series © S.Balter 2009 Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D. Outline JC Radiation Sentinel Event • ICRU diagnostic dosimetric quantities and their fluoroscopic extensions • Construction, dosimetric features, and performance characteristics of modern fluoroscopes • Extended QA protocols for compliance measurements, system characterization and clinical dosimetry. • Dose recording and reporting • The Joint Commission fluoroscopy sentinel event. SB0907 UFP AAPM - 61 © S. Balter 2009 MedPhys list – May 2007 • “A facility without a full-time physicist needs a quick, easy number that they can refer to, so we use fluoro time, as archaic and inaccurate as it might be. • If you want to know if your patient received more than the threshold for skin injuries, look at the patient's skin, not the KAP meter or fluoro time!” Etiology not identified for more than one year SB0907 UFP AAPM - 63 An unexpected occurrence involving death or serious physical or psychological injury, or the risk thereof. Serious injury specifically includes loss of limb or function. The phrase “or the risk thereof” includes any process variation for which a recurrence would carry a significant chance of a serious adverse outcome. Such events are called “sentinel” because they signal the need for immediate investigation and response. • Prolonged fluoroscopy with cumulative dose >1500 rads to a single field or any delivery of radiotherapy to the wrong body region or >25% above the planned radiotherapy dose • Can you demonstrate to the surveyors that Sentinel Events did not occur? • How to be accurate enough? (not easy) SB0907 UFP AAPM - 62 © S. Balter 2009 What was covered Fluoro 35 minutes Cine 12 minutes © S. Balter 2009 • ICRU diagnostic dosimetric quantities and their fluoroscopic extensions • Construction, dosimetric features, and performance characteristics of modern fluoroscopes • Extended QA protocols for compliance measurements, system characterization and clinical dosimetry. • Dose recording and reporting • The Joint Commission fluoroscopy sentinel event. SB0907 UFP AAPM - 64 © S. Balter 2009 © S.Balter 2009