Fluoroscopic Imaging Equipment Guidelines for Detector Input Dose Settings and Image Optimization Fluoroscopy Evaluation Patient Dose Image Quality kVp RELATIVE EERD vs 210% - 22 cm FOV Direct Shimadzu Bransist (Se/TFT) Flat Panel Detector FOV Indirect Flat Philips Xper - 22 cm Panel Detector (CsI(Tl)/TFT) 190% 180% 170% - 42 cm FOV Indirect Siemens Artis Zee (CsI(Tl)/TFT) Flat Panel Detector - 22 cm FOV X-ray Siemens Siregraph Image Intensifier Detector 160% 150% 140% Entrance Air Kerma for 22 cm FOV (Measure d at Intervent ional Ref Point) 140 120% 110% 100% Air Kerma Rate (mGy/min) 130% No Grid-70 k V 10R AF 2k W 4ms/ mic rof ocus 7ms Note: Thi s one at 40 nGy /pulse and 10 pps 100kV 10R A F 2kW smallf oc us 120 No Grid-70kV 10R A F 2kW 4ms /7ms mic rof ocus Note: Thi s one at 23 nGy /pulse and 30 pps 80kV 10R 0302 3kW smallfoc us Note: Thi s one at 36 nGy /pulse and 15 pps s o meas . Dos e was mul tiplied by 2 I nt erv ent ional IQ 2 100 90% 80% 40 50 60 70 80 90 100 110 kVp Phil Rauch Henry Ford Health Systems Detroit, MI Patient Entrance Rate Detector Exposure Normalized to Minimum 200% 80kV 10R 0302 2kW smallfoc us Note: Thi s one at 36 nGy /pulse and 15 pps s o meas . Dos e was mul tiplied by 2 70kV 10R A F 2kW 5ms/8ms mic rofoc us 80 80kV 10R A F 2kW smallf ocus 80kV 10R 02 3k W small 60 I nt erv ent ional focus IQ 1 (no DHHS) 70kV 10R A F 2kW microfocus 40 70kV Servic e 02 70kV 10R A F 3kW 6ms/9ms small focus 70kV 10R A F 3kW 8ms/12ms smallf oc us 20 70kV 10R 02 3k W small 70kV 10R 02 2k W small 0 focus focus 70kV 20R A F 3kW smallf ocus 4 6 70kV 10R A F 3kW smallf ocus 8 Nominal Phantom Thickness (inches) 10 70kV 20R A F 3kW 8ms/12ms smallf oc us 70kV Servic e 00 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Dose Reduction vs Image Quality Dose Reduction vs Image Quality Dose reduction depends on…. ….technology ….proper equipment design ….proper set up of equipment parameters ….proper utilization of the equipment ….knowledge and skill of the radiologist Image quality depends on…. ….technology ….proper equipment design ….proper set up of equipment parameters ….proper utilization of the equipment ….knowledge and skill of the radiologist 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 1 What is the Patient Dose? What is the Exam Protocol? What is the exam protocol? 2009 AAPM Meeting-Phil Rauch Entrance Air Kerma for 22 cm FOV (Measured at Interventional Ref Point) 140 Patient dose (and image quality) are highly dependent on the exam protocol setting 80kV 10R 0302 2kW smallfocus Note: This one at 36 nGy/pulse and 15 pps so meas. Dose was multiplied by 2 70kV 10R AF 2kW 5ms/8ms microfocus 100 80kV 10R AF 2kW smallfocus 80 Entrance Air Kerma for 22 cm FOV (Measured at Interventional Ref Point) No Grid-70 kV 10R AF 2kW 4ms/7ms microfocus Note: This one at 40 nGy/pulse and 10 pps 100kV 10R AF 2kW smallfocus No Grid-70kV 10R AF 2kW 4ms/7ms microfocus Note: This one at 23 nGy/pulse and 30 pps 80kV 10R 0302 3kW smallfocus Note: This one at 36 nGy/pulse and 15 pps so meas. Dose was multiplied by 2 Interventional IQ 2 80kV 10R 02 3kW smallfocus FDA Limit Interventional IQ 1 (no DHHS) 70kV 10R AF 2kW microfocus 60 70kV Service 02 70kV 10R AF 3kW 6ms/9ms smallfocus 40 70kV 10R AF 3kW 8ms/12ms smallfocus 70kV 10R 02 3kW smallfocus 70kV 10R 02 2kW smallfocus 20 70kV 20R AF 3kW smallfocus Patient Entrance Air Kerma Rate (mGy/min) Patient Entrance Air Kerma Rate (mGy/min) 140 120 2009 AAPM Meeting-Phil Rauch 120 No Grid-70 kV 10R AF 2kW 4ms/7ms microfocus Note: This one at 40 nGy/pulse and 10 pps 100kV 10R AF 2kW smallfocus No Grid-70kV 10R AF 2kW 4ms/7ms microfocus Note: This one at 23 nGy/pulse and 30 pps 80kV 10R 0302 3kW smallfocus Note: This one at 36 nGy/pulse and 15 pps so meas. Dose was multiplied by 2 Interventional IQ 2 Low doses for this protocol may be appropriate for pediatric imaging 80kV 10R 0302 2kW smallfocus Note: This one at 36 nGy/pulse and 15 pps so meas. Dose was multiplied by 2 70kV 10R AF 2kW 5ms/8ms microfocus 100 80kV 10R AF 2kW smallfocus 80 80kV 10R 02 3kW smallfocus FDA Limit Interventional IQ 1 (no DHHS) 70kV 10R AF 2kW microfocus 60 70kV Service 02 70kV 10R AF 3kW 6ms/9ms smallfocus 40 70kV 10R AF 3kW 8ms/12ms smallfocus 70kV 10R 02 3kW smallfocus 70kV 10R 02 2kW smallfocus 20 70kV 20R AF 3kW smallfocus 70kV 10R AF 3kW smallfocus 0 70kV 10R AF 3kW smallfocus 0 4 6 8 Nominal Phantom Thickness (inches) 2009 AAPM Meeting-Phil Rauch 10 70kV 20R AF 3kW 8ms/12ms smallfocus 4 70kV Service 00 6 8 Nominal Phantom Thickness (inches) 10 70kV 20R AF 3kW 8ms/12ms smallfocus 70kV Service 00 2009 AAPM Meeting-Phil Rauch 2 Entrance Air Kerma for 22 cm FOV (Measured at Interventional Ref Point) Patient Entrance Air Kerma Rate (mGy/min) 140 120 100 No Grid-70 kV 10R AF 2kW 4ms/7ms microfocus Note: This one at 40 nGy/pulse and 10 pps 100kV 10R AF 2kW smallfocus Image Quality Assessment No Grid-70kV 10R AF 2kW 4ms/7ms microfocus Note: This one at 23 nGy/pulse and 30 pps 80kV 10R 0302 3kW smallfocus Note: This one at 36 nGy/pulse and 15 pps so meas. Dose was multiplied by 2 Interventional IQ 2 Settings optimized for low dose pediatric imaging may not be appropriate for adults 80kV 10R 0302 2kW smallfocus Note: This one at 36 nGy/pulse and 15 pps so meas. Dose was multiplied by 2 70kV 10R AF 2kW 5ms/8ms microfocus 80kV 10R AF 2kW smallfocus 80 80kV 10R 02 3kW smallfocus FDA Limit Interventional IQ 1 (no DHHS) 70kV 10R AF 2kW microfocus 60 70kV Service 02 70kV 10R AF 3kW 6ms/9ms smallfocus 40 70kV 10R AF 3kW 8ms/12ms smallfocus 70kV 10R 02 3kW smallfocus 70kV 10R 02 2kW smallfocus 20 70kV 20R AF 3kW smallfocus 70kV 10R AF 3kW smallfocus 0 4 6 8 Nominal Phantom Thickness (inches) 10 70kV 20R AF 3kW 8ms/12ms smallfocus 70kV Service 00 2009 AAPM Meeting-Phil Rauch Image Quality Assessment 2009 AAPM Meeting-Phil Rauch Image Quality Assessment Evaluation Descriptors (HVS) Contrast Edge Detection Noise Sharpness Pattern Recognition Saturation Comparison Artifacts Assessment Relevance Conclusion 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 3 Is Pulse Width Important? 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Fluoroscopic Image Quality me (Fra n Blur gratio inte ime) T Is Pulse Width Important? Exposure ON-time – Pulsed vs Cont Image Intensifier Flat Panel Pulsed Fluoro* (30 pulses per sec) *Displayed at 7.5 fps 2009 AAPM Meeting-Phil Rauch Continuous Fluoro* (30 video frames per sec) *Displayed at 7.5 fps 2009 AAPM Meeting-Phil Rauch 4 Fluoroscopic Image Quality me Fra ation gr inte ime t LAG (Sig rete nal ntion ) Image Intensifier Fluoroscopic Image Quality me Fra ation gr inte ime t LAG (Sig rete nal ntion ) Flat Panel Image Intensifier Flat Panel Rec ur filt sive s et er ting 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Blur versus Recursive Filtering Blur versus Recursive Filtering Continuous Fluoroscopy Pulsed Fluoroscopy Continuous Fluoroscopy Pulsed Fluoroscopy Without Recursive Filtering With Recursive Filtering Without Recursive Filtering With Recursive Filtering 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 5 Fluoroscopic Image Quality me Fra ation gr inte ime t LAG (Sig rete nal ntion ) Image Intensifier Perceptual Integration Where are the 3’s Flat Panel Rec ur filt sive set er ting l tua cep ion Per grat e int 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Perceptual Integration Perceptual Integration Where are the 3’s Where are the 3’s 30 Frames/sec 2009 AAPM Meeting-Phil Rauch 8 Frames/sec 2009 AAPM Meeting-Phil Rauch 6 Wedge Filter Acquisition Type Spatial Beam Shaping Collimation without Radiation Geometry Iris (Lead) Measurement Video Camera Last Image Hold SID SSD FOV Electronic Zoom Fluoroscopy ON-Time Generator Fluoroscopy Power Curves Normal Level FDA Max FDA 21CFR Patient Exposure High Level Minimum Filtration Patient Size Dose Mode EERD † Scatter to Primary Ratio Spectral Filtration X-ray Generator 120 Noise Anti-scatter Grid 110 Dynamic Range Detector Characteristics Spatial Resolution Pulsed Fluoroscopy Fluoroscopy Waveform Grid Control 100 Cont. Low - 30 (video) fps 90 MTF 80 Luminance Display Characteristics kV X-ray Tube DQE Image Quality Contrast Resolution Temporal Resolution kV-mA Power Curves 130 Normal Scan Slow Scan SNR Spectral Beam Shaping Generator Fluoroscopy Technique Iris (Optical) Image Size kVp Beam Alignment 70 Contrast 60 Pulse Width Gamma 0.0 0.5 Continuous Fluoroscopy † EERD = Entrance Exposure Rate to Detector 2.0 2.5 3.0 3.5 4.0 2009 AAPM Meeting-Phil Rauch Generator Fluoroscopy Technique Generator Fluoroscopy Power Curves Generator Fluoroscopy Power Curves 130 130 120 120 110 110 Cont. Norm - 30 (video) fps kVp kVp 1.5 mA Generator Fluoroscopy Technique 100 1.0 Pulse Rate 90 100 Pulsed - 30 pps Cont. Norm - 30 (video) fps 90 Cont. Low - 30 (video) fps Cont. Low - 30 (video) fps 80 80 70 70 60 60 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 0.0 0.5 mA 2009 AAPM Meeting-Phil Rauch 1.0 1.5 2.0 2.5 3.0 3.5 4.0 mA 2009 AAPM Meeting-Phil Rauch 7 Generator Fluoroscopy Technique Generator Fluoroscopy Technique Generator Fluoroscopy Power Curves Generator Fluoroscopy Power Curves 90 90 80 80 kVp 100 kVp 100 70 70 60 60 H ig h D o s e P u ls e d - 3 . 7 5 p p s / 0 .2 m m C u S p e c t ra l F ilt e r N o rm a l D o s e P u ls e d - 3 . 7 5 p p s / 0 . 2 m m C u S p e c t ra l F ilt e r N o rm a l D o s e P u ls e d - 3 . 7 5 p p s / 0 . 2 m m C u S p e c t ra l F ilt e r 50 50 0 5 10 15 20 25 30 35 40 45 0 5 10 15 20 mA 25 30 35 40 45 mA 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Generator Fluoroscopy Technique Generator Fluoroscopy Technique Generator Fluoroscopy Power Curves Generator Fluoroscopy Power Curves 90 90 80 80 kVp 100 kVp 100 70 70 60 60 H ig h P u ls e d - 15 p p s / 0 . 2 m m C u S p e c t ra l F ilte r N o rm a l D o s e P u ls e d - 15 p p s / 0 . 2 m m C u S p e c t ra l F ilt e r N o rm a l D o s e P u ls e d - 15 p p s / 0 . 2 m m C u S p e c t ra l F ilt e r 50 50 0 5 10 15 20 25 30 35 40 45 0 mA 2009 AAPM Meeting-Phil Rauch 5 10 15 20 25 30 35 40 45 mA 2009 AAPM Meeting-Phil Rauch 8 Generator Fluoroscopy Technique Generator Fluoroscopy Technique Generator Fluoroscopy Power Curves Generator Fluoroscopy Power Curves 100 90 90 80 80 kVp kVp 100 70 70 60 60 High Dose Continuous / 0.2mm Cu Spectral Filter) Normal Dose Continuous / 0.2mm Cu Spectral Filter Normal Dose Continuous / 0.2mm Cu Spectral Filter 50 50 0 5 10 15 20 25 30 35 40 45 0 5 10 15 20 mA 25 30 35 40 45 mA 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Entrance Exposure Rate to Detector EERD Test Conditions RELATIVE EERD vs kVp 210% 200% Detector Exposure Rate Normalized to Minimum Solid state detectors must be positioned outside the automatic dose rate measurement field 190% 180% 170% Siemens Siregraph - 22 cm FOV X-ray Image Intensifier Detector 160% 150% 140% 130% 120% 110% 100% 90% 80% 40 50 60 70 80 90 100 110 kVp 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 9 EERD Test Conditions EERD Test Conditions RELATIVE EERD vs kVp RELATIVE EERD vs kVp 210% 210% 200% Detector Exposure Rate Normalized to Minimum Detector Exposure Rate Normalized to Minimum 200% 190% Siemens Artis Zee - 42 cm FOV Indirect Flat Panel Detector (CsI(Tl)/TFT) 180% 170% 160% Siemens Siregraph - 22 cm FOV X-ray Image Intensifier Detector 150% 140% 130% 120% 110% 100% 190% 170% Siemens Artis Zee - 42 cm FOV Indirect Flat Panel Detector (CsI(Tl)/TFT) 160% 150% Siemens Siregraph - 22 cm FOV X-ray Image Intensifier Detector 140% 130% 120% 110% 100% 90% 90% 80% 80% 40 50 60 70 80 90 100 40 110 50 60 70 80 90 100 110 kVp kVp 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch EERD Test Conditions EERD Test Conditions EERD for Different Attenuators RELATIVE EERD vs kVp 250 210% Acrylic (inches) 190% EERD (micro-R per sec) 200% Detector Exposure Rate Normalized to Minimum Philips Xper - 22 cm FOV Indirect Flat Panel Detector (CsI(Tl)/TFT) 180% Shimadzu Bransist - 22 cm FOV Direct Flat Panel Detector (Se/TFT) 180% Philips Xper - 22 cm FOV Indirect Flat Panel Detector (CsI(Tl)/TFT) 170% 160% Siemens Artis Zee - 42 cm FOV Indirect Flat Panel Detector (CsI(Tl)/TFT) 150% Siemens Siregraph - 22 cm FOV X-ray Image Intensifier Detector 140% 130% 120% 110% 200 Aluminum (inches) Copper (mm) 150 100 50 100% 90% 0 80% 40 50 60 70 80 90 100 110 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 kVp 2009 AAPM Meeting-Phil Rauch Attenuator Thickness 2009 AAPM Meeting-Phil Rauch 10 Image Quality: What is it? “Image quality depends only on intrinsic, objective physical characteristics of an imaging system, and can be measured independently of an observer” Detector Input Dose Rates (FP) Image Intensifier Detector Metrics Flat Panel Definitions courtesy Ralph Schaetzing, Agfa Corp. 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Image Detector Pixel Pitch – Estimated Resolution Indirect Conversion FPD Pixel Pitch Detector Safire (a-Se, a-Si TFT) Philips, Siemens (CsI(Tl), a-Si TFT) - Angio Philips, Siemens (CsI(Tl), a-Si TFT) - Cardiac GE, Revolution (CsI(Tl), a-Si TFT) 2009 AAPM Meeting-Phil Rauch Mfg. Shimadzu Trixell Trixell GE Pixel Nyquist Nyquist x 2^(.5) Pitch (lp/mm) (lp/mm) 0.150 3.33 4.71 0.154 3.25 4.59 0.184 2.72 3.84 0.200 2.50 3.54 Mesh Lines per Inch at 45-degrees 120 117 98 90 2009 AAPM Meeting-Phil Rauch 11 Actual Resolution Image Quality Assessment Detroit Phantom Archived 512 x 512 Archived 1024 x 1024 2009 AAPM Meeting-Phil Rauch Image Quality Assessment 2009 AAPM Meeting-Phil Rauch Image Quality Assessment FOV Coverage Motion blur Resolution Contrast to noise Comparison with other fluoroscopy equipment Rotational Angiography 3D Reconstruction Recursive Filter effects Image processing effects Pixel matrix effects 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 12 Image Quality Assessment FOV Coverage Motion blur Resolution Contrast to noise Comparison with other fluoroscopy equipment Rotational Angiography 3D Reconstruction Recursive Filter effects Image processing effects Pixel matrix effects 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Rotational Angiography Digital Radiograph 2009 AAPM Meeting-Phil Rauch Image Quality Assessment Rotational Angiography Image Quality Assessment Digital Radiograph Image Quality Assessment 3-D Reconstruction Digital Radiograph 2009 AAPM Meeting-Phil Rauch 13 Image Detector (Fluoro LIH) Image Detector (Fluoro LIH) Indirect Conversion FPD Can you match the fluoroscopy LIH image to the detector? A. Image Intensifier B. Indirect Flat Panel C. Direct Flat Panel ? 2009 AAPM Meeting-Phil Rauch Image Detector (Fluoro LIH) Car Moving 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Image Detector (Fluoro LIH) Car Stationary 2009 AAPM Meeting-Phil Rauch 14 Resolution Resolution Can you see a 50- moving wire using a flat panel detector with a pixel pitch of 150-? With a pixel pitch of 200-? Direct Capture, 150- pixel pitch Indirect Capture, 200- pixel pitch Wire size numbers are located below the corresponding wire 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Image Detector (Fluoro LIH) Detector Input Dose Rates (XRII) Indirect Conversion FPD Cannot rely on reported detector metrics alone. Must understand and appropriately utilize optimum technical settings and image processing 2009 AAPM Meeting-Phil Rauch Dose Rate Selection Field of View Selection 2009 AAPM Meeting-Phil Rauch 15 Detector Input Dose Rates (XRII) Detector Input Dose Rates (XRII) Dose Rate Selection Dose Rate Selection Low Norm High Low Field of View Selection Field of View Selection 40 cm 30 cm 22 cm 17 cm 30 cm Continuous Fluoroscopy: Input dose per frame of fluoroscopy? 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Detector Input Dose Rates (XRII) • • • Low Dose Mode Cont. Fluoroscopy 30 cm FOV Detector Input Dose/Fr? A. B. C. D. E. F. • • • Low 0.37 µR (3.2 nGy) 0.75 µR (6.6 nGy) 30 cm 1.5 µR (13.1 nGy) 3.0 µR (26.3 nGy) ……values assume FDA minimum beam filtration 6.0 µR (52.6 nGy) No Clue 2009 AAPM Meeting-Phil Rauch Detector Input Dose Rates (XRII) Low Dose Mode Cont. Fluoroscopy 30 cm FOV Detector Input Dose/Fr? A. B. C. D. E. F. Low 0.37 µR (3.2 nGy) 0.75 µR (6.6 nGy) 30 cm 1.5 µR (13.1 nGy) 3.0 µR (26.3 nGy) ……values assume FDA minimum beam filtration 6.0 µR (52.6 nGy) No Clue 2009 AAPM Meeting-Phil Rauch 16 Detector Input Dose Rates (XRII) • • • Low Dose Mode Cont. Fluoroscopy 30 cm FOV • • • Detector Input Dose/Fr? A. B. C. D. E. F. Low Dose Mode Cont. Fluoroscopy 30 cm FOV Detector Input Dose/Fr? 1x 1.5x 3.0x 0.37 µR (3.2 nGy) 0.75 µR (6.6 nGy) 1.5 µR (13.1 nGy) 3.0 µR (26.3 nGy) ……values assume FDA minimum beam filtration 6.0 µR (52.6 nGy) ……can use 2x these values if minimum 0.2 mm Cu No Clue beam filtration is utilized 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Detector Input Dose Rates (XRII) • • • Detector Input Dose Rates (XRII) Low Dose Mode Cont. Fluoroscopy 30 cm FOV Detector Input Dose/Fr? Video Frame Rate What is the video frame rate in the U.S.A.? Ans: 30 video frames per second 1x 1.5x 3.0x 40 cm 30 cm 22 cm 17 cm Detector dose scales inversely as either the ratio of the FOV, or as the ratio squared 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 17 Detector Input Dose Rates (XRII) Rauch's 30-30-30 Rule: For a nominal 30 cm image intensifier field of view (FOV), Pulse rate of 30 pps, Set the EERD to 30 3R/sec For other FOV's, scale by either the ratio of the FOV (auto optical lens aperture) or the square of the ratio For spectral beam filtration… 0.1 mm Cu equivalent, multiply by 1.41 0.2 mm Cu equivalent, multiply by 2 Detector Input Dose Rates (FP) Image Intensifier Detector Metrics (Fluoroscopy Doses) Flat Panel Adapted from Koch, A., Macherel, J.M., Wirth, T., de Groot, P., Ducourant, T., Couder, D., Moy, J.P., & Calais, E. (2001). Detective quantum efficiency of an X-ray image intensifier chain as a benchmark for amorphous silicon flat panel detectors, Proc. SPIE, 4320, 115-20 For flat panel, special consideration is required!! 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Detector Input Dose Rates (FP) Detector Metrics (Fluoroscopy Doses) Pixel Pitch – Air Kerma per Pixel Image Intensifier Pixel Pitch Flat Panel “The minimum operating dose level is defined by the magnitude of the noise arising in the AM array and readout electronics. This typically sets the lower operating limit for detector dose at 20-50 nGy.” Cowen AR et al., Solid-state, flat-panel, digital radiography detectors and their physical imaging characteristics, Clin Radiol (2008) 2009 AAPM Meeting-Phil Rauch Detector Safire (a-Se, a-Si TFT) Philips, Siemens (CsI(Tl), a-Si TFT) - Angio Philips, Siemens (CsI(Tl), a-Si TFT) - Cardiac GE, Revolution (CsI(Tl), a-Si TFT) Mfg. Shimadzu Trixell Trixell GE Pixel Pitch 0.150 0.154 0.184 0.200 For the transition from XRII to Flat Panel, the required EERD can be estimated by determining the air kerma required for the XRII when it is operating at the same pixel pitch at that for the flat panel. 2009 AAPM Meeting-Phil Rauch 18 Pixel Pitch vs Input Dose Pixel Pitch (mm) Equivalent XRII Diam (cm) for 1024 Pixels Diam^2 Pixel Pitch vs Input Dose Ratio wrt 30 cm EERD EERD FOV (µR) (nGy) XRII Pixel Pitch (mm) Equivalent XRII Diam (cm) for 1024 Pixels Diam^2 XRII 0.293 30.00 900.00 1.00 1.00* 8.76 0.293 15.36 15.77 18.84 20.48 235.93 248.68 355.01 419.43 3.81 3.62 2.54 2.15 3.81 3.62 2.54 2.15 33.42 31.70 22.21 18.80 0.150 0.154 0.184 0.200 Flat Panel 0.150 0.154 0.184 0.200 30.00 900.00 1.00 1.00* 8.76 15.36 15.77 18.84 20.48 235.93 248.68 355.01 419.43 3.81 3.62 2.54 2.15 3.81 3.62 2.54 2.15 33.42 31.70 22.21 18.80 Flat Panel *Per Rauch’s 30-30-30 rule *Per Rauch’s 30-30-30 rule 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Pixel Pitch vs Input Dose Pixel Pitch (mm) Ratio wrt 30 cm EERD EERD FOV (µR) (nGy) Equivalent XRII Diam (cm) for 1024 Pixels Diam^2 Pixel Pitch vs Input Dose Ratio wrt 30 cm EERD EERD FOV (µR) (nGy) XRII 0.293 Equivalent XRII Diam (cm) for 1024 Pixels Diam^2 Ratio wrt 30 cm EERD EERD FOV (µR) (nGy) XRII 30.00 900.00 1.00 1.00* 8.76 0.293 15.36 15.77 18.84 20.48 235.93 248.68 355.01 419.43 3.81 3.62 2.54 2.15 3.81 3.62 2.54 2.15 33.42 31.70 22.21 18.80 0.150 0.154 0.184 0.200 Flat Panel 0.150 0.154 0.184 0.200 Pixel Pitch (mm) 30.00 900.00 1.00 1.00* 8.76 15.36 15.77 18.84 20.48 235.93 248.68 355.01 419.43 3.81 3.62 2.54 2.15 3.81 3.62 2.54 2.15 33.42 31.70 22.21 18.80 Flat Panel *Per Rauch’s 30-30-30 rule *Per Rauch’s 30-30-30 rule 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 19 Pixel Pitch vs Input Dose Pixel Pitch (mm) Equivalent XRII Diam (cm) for 1024 Pixels Diam^2 Pixel Pitch vs Input Dose Ratio wrt 30 cm EERD EERD FOV (µR) (nGy) XRII 0.293 Equivalent XRII Diam (cm) for 1024 Pixels Diam^2 Ratio wrt 30 cm EERD EERD FOV (µR) (nGy) XRII 30.00 900.00 1.00 1.00* 8.76 0.293 15.36 15.77 18.84 20.48 235.93 248.68 355.01 419.43 3.81 3.62 2.54 2.15 3.81 3.62 2.54 2.15 33.42 31.70 22.21 18.80 0.150 0.154 0.184 0.200 Flat Panel 0.150 0.154 0.184 0.200 Pixel Pitch (mm) 30.00 900.00 1.00 1.00* 8.76 15.36 15.77 18.84 20.48 235.93 248.68 355.01 419.43 3.81 3.62 2.54 2.15 3.81 3.62 2.54 2.15 33.42 31.70 22.21 18.80 Flat Panel *Per Rauch’s 30-30-30 rule *Per Rauch’s 30-30-30 rule 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Detector Input Dose Rates Dose Per Pulse (<30 pps) For pulse rates lower than 30 pps, use Aufrichtig* perceptual pulsed fluoroscopy scale factor to adjust dose per frame for different frame rates 1x 1.5x 3.0x (Dose/Pulse)2 = (Dose/Pulse)1*SQRT[(Pulse Rate)1/(Pulse Rate)2] Example: Change from 30pps to 7.5 pps What about pulsed fluoroscopy? (Dose/Pulse)2 = (Dose/Pulse)1*SQRT[30/7.5] = (Dose/Pulse)1*2 *Perceptual comparison of pulsed and continuous fluoroscopy, Richard Aufrichtig, Med. Phys. 21 (2), February 1994 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch 20 Dose Per Pulse (<30 pps) C-arm Dose Rates Mobile C-arm Operational Settings For pulse rates lower than 30 pps, use Aufrichtig* perceptual pulsed fluoroscopy scale factor to adjust dose per frame for different frame rates For pulse rate of 5 per second or lower, use the Aufrichtig calculated value for 5 pps (The retina can only retain an image for 200 ms, so there is no integration of noise for pulse rates of 5 pps or lower) 2009 AAPM Meeting-Phil Rauch Operational Mode Max Dose Rate (mGy/min) 6 37 Low dose Boost Continuous 9 55 Default dose Non-boost Continuous 12 67 Default dose Boost Continuous 22 148 Low dose Non-boost Pulsed 6 82 Low dose Boost Pulsed 37 2833 Default dose Non-boost Pulsed 12 132 Default dose Boost Pulsed 74 3143 Acquisition Modes 1 2 3 2009 AAPM Meeting-Phil Rauch EERD 1 (nGy/pulse) Low dose Non-boost Continuous Operator can store 0 to 30 frames/sec (maxed at selected pulse rate 15, 30 pps Ability to select “Low dose” Ability to use the two foot pedals All frames recorded 15, 30 pps Ability to select “Low dose” Ability to use the two foot pedals All frames recorded C-arm Dose Rates EERD and Max Dose Rate (FDA) Mobile C-arm Operational Settings – Left: Normal Fluoro – Right: Boost Fluoro – Left: Normal Fluoro – Right: Boost Fluoro Operator can store 0 to 30 frames/sec (maxed at selected pulse rate 2009 AAPM Meeting-Phil Rauch C-arm Dose Rates Cont., 8, 4, 2, 1 pps Ability to select “Low” dose Two foot pedals: Cont., 8, 4, 2, 1 pps Ability to select “Low” dose Two foot pedals: Acquisition Modes *Perceptual comparison of pulsed and continuous fluoroscopy, Richard Aufrichtig, Med. Phys. 21 (2), February 1994 XRII Imaging System (31cm FOV) Filtration equiv to 0.1 mm Cu Fluoro modes: XRII Imaging System (31cm FOV) Filtration equiv to 0.1 mm Cu Fluoro modes: EERD per pulse is the same for all pulse rates Data for 8pps & 4pps; scales with pulse rate below 4pps Data for 30pps; scales with pulse rate, except 8pps and 4pps have the same max rate due to change in pulse width EERD and Max Dose Rate (FDA) Dose Rate Analysis Operational Mode EERD 1 (nGy/pulse) Max Dose Rate (mGy/min) Low dose Non-boost Continuous 6 37 Low dose Boost Continuous 9 55 Default dose Non-boost Continuous 12 67 Default dose Boost Continuous 22 148 Are these values acceptable? Values exceed the FDA limits of 176 mGy/min for high dose fluoroscopy EERD dose not change with pulse rate, so the same detector dose is delivered at 1pps as at 30pps Appears to be pulsed fluoroscopy…. ….but the 15pps and 30pps are labeled “Cine”, and all frames are recorded. The FDA limits do not apply 1 2 3 2009 AAPM Meeting-Phil Rauch Low dose Non-boost Pulsed 6 82 Low dose Boost Pulsed 37 2833 Default dose Non-boost Pulsed 12 132 Default dose Boost Pulsed 74 3143 EERD per pulse is the same for all pulse rates Data for 8pps & 4pps; scales with pulse rate below 4pps Data for 30pps; scales with pulse rate, except 8pps and 4pps have the same max rate due to change in pulse width 21 Entrance Exposure Rate to Detector EERD and Max Dose Rate (FDA) EERD Analysis Using Rauch 30-30-30 rule, adjustment for filtration, and Aufrichtig scale, expect 11.6 nGy per pulse for 30 pps† Measured default non-boost values are in agreement for 30pps However, for pulsed mode the dose does not change as the pulse rate is reduced (Does not follow the Aufrichtig scale) Result is too low a dose for the lower pulse rates, which may discourage operators from using the lower rates Operational Mode Max Dose Rate (mGy/min) Operational Mode EERD 1 (nGy/pulse) Max Dose Rate (mGy/min) Low dose Non-boost Continuous 6 37 6 37 9 55 Low dose Boost Continuous 9 55 Default dose Non-boost Continuous 12 67 Default dose Non-boost Continuous 12 67 Default dose Boost Continuous 22 148 Default dose Boost Continuous 22 148 Low dose Non-boost Pulsed 2 2009 AAPM Meeting-Phil Rauch 6 82 Low dose Non-boost Pulsed 6 82 Low dose Boost Pulsed 37 2833 Low dose Boost Pulsed 37 2833 Default dose Non-boost Pulsed 12 132 Default dose Non-boost Pulsed 12 132 Default dose Boost Pulsed 74 3143 Default dose Boost Pulsed 74 3143 EERD per pulse is the same for all pulse rates Data for 8pps & 4pps; scales with pulse rate below 4pps Data for 30pps; scales with pulse rate, except 8pps and 4pps have the same max rate due to change in pulse width Entrance Exposure Rate to Detector EERD and Max Dose Rate (FDA) Analysis Operational Mode EERD 1 (nGy/pulse) Max Dose Rate (mGy/min) Low dose Non-boost Continuous 6 37 Low dose Boost Continuous 9 55 Default dose Non-boost Continuous 12 67 Default dose Boost Continuous 22 148 Low dose Non-boost Pulsed 6 82 Low dose Boost Pulsed 37 2833 Default dose Non-boost Pulsed 12 132 Default dose Boost Pulsed 74 3143 1 2 3 2009 AAPM Meeting-Phil Rauch Expect low dose mode to be about 1/2 to 2/3 of the default dose Measured non-boost values are in agreement Low dose Boost Continuous 3 Expected values for default boost pulsed would be 2x to 3x the nonboost values Measured default boost pulsed values are 6x larger than nonboost EERD and Max Dose Rate (FDA) Analysis Low dose Non-boost Continuous 1 †8.76nGy*(30/31)2*SQRT(30/30)*1.41 EERD 1 (nGy/pulse) Entrance Exposure Rate to Detector EERD per pulse is the same for all pulse rates Data for 8pps & 4pps; scales with pulse rate below 4pps Data for 30pps; scales with pulse rate, except 8pps and 4pps have the same max rate due to change in pulse width 1 2 3 2009 AAPM Meeting-Phil Rauch EERD per pulse is the same for all pulse rates Data for 8pps & 4pps; scales with pulse rate below 4pps Data for 30pps; scales with pulse rate, except 8pps and 4pps have the same max rate due to change in pulse width Entrance Exposure Rate to Detector EERD and Max Dose Rate (FDA) Analysis Using Rauch 30-30-30 rule, adjustment for filtration, and Aufrichtig scale, expect 28.4 to 56.8 nGy per pulse for boost at 5 pps or lower† Measured default boost value is much larger than this estimate Furthermore, the dose does not change as the pulse rate is increased (Does not follow the Aufrichtig scale) Result is much larger than necessary doses for the higher frame rates Operational Mode EERD 1 (nGy/pulse) Max Dose Rate (mGy/min) Low dose Non-boost Continuous 6 37 Low dose Boost Continuous 9 55 Default dose Non-boost Continuous 12 67 Default dose Boost Continuous 22 148 1 †8.76nGy*(30/31)2*SQRT(30/5)*2 2 3 2009 AAPM Meeting-Phil Rauch Low dose Non-boost Pulsed 6 82 Low dose Boost Pulsed 37 2833 Default dose Non-boost Pulsed 12 132 Default dose Boost Pulsed 74 3143 EERD per pulse is the same for all pulse rates Data for 8pps & 4pps; scales with pulse rate below 4pps Data for 30pps; scales with pulse rate, except 8pps and 4pps have the same max rate due to change in pulse width 22 Rauch’s 30-30-30 Rule Use the previous type analysis to determine how the vendor specified or measured EERD compares with the expected values Use the rule to compare measured values with the expected change in EERD whenever there are variations in the field of view, filtration, or pulse rate Image Quality: What is it? “Image quality depends only on intrinsic, objective physical characteristics of an imaging system, and can be measured independently of an observer” “Image quality is whatever the observer says it is (i.e., it is a subjective perception of the image, ‘in the eye of the beholder’)” Definitions courtesy Ralph Schaetzing, Agfa Corp. 2009 AAPM Meeting-Phil Rauch 2009 AAPM Meeting-Phil Rauch Fluoroscopic Image Quality Biliary endoscopic catheterization Mobile C-arm Gastroenterology clinic (No X-ray techs) 1024 x 1280 x 12 bit image Secondary capture of fluoro frame Human Visual Perception 2009 AAPM Meeting-Phil Rauch 23 Observations Mottled Blurred, probably due to long pulse width Low dynamic range Lack of detail Fluoro Frame 8-Bit Grayscale 512 x 512 Matrix Flat Detector Modality Image downsized from 10242 Inappropriate anatomical setting No x-ray technique or image processing information available in the DICOM metadata Fluoro Frame 12-Bit Grayscale 1024 x 1024 Matrix Flat Detector Modality Store Monitor Image Fluoro Frame 8-Bit Grayscale 768 x 576 Matrix XRII Digital Video Screen Capture via DVR Fluoro Frame 12-Bit Grayscale 1024 x 1024 Matrix Flat Detector Modality Store Monitor Image Fluoro Frame 8-Bit Grayscale 768 x 576 Matrix XRII Digital Video Screen Capture via DVR Fluoro Frame 8-Bit Grayscale 884 x 884 Matrix XRII Screen Capture via Paxport 24 Dynamic Range 2009 AAPM Meeting-Phil Rauch Fluoroscopy – Vascular Mode Digital Radiography Trixell Pixium 4700 (0.154 µ; 2480 x 1910 matrix Trixell Pixium 4600 (0.143 µ; 3001 x 3001 matrix Video Bandwidth •Three display monitors provide with installation (Live fluoroscopy, Roadmap, Reference) •A single video graphics display card is used to simultaneously paint all three images and graphics Fluoroscopy – Vascular Mode Fluoroscopy – Cardiac Mode Trixell Pixium 4700 (0.154 µ; 2480 x 1910 matrix Trixell Pixium 4700 (0.154 µ; 2480 x 1910 matrix •Result is a loss of resolution when compared with the display of single images on a modality workstation or on PACS 2009 AAPM Meeting-Phil Rauch 25 Test Your Knowledge Test Your Knowledge Question: Name two ways the vessel contrast can appear as in this image Answer: 1. Start injection of contrast well before starting the run 2. Change the mask to a frame after the contrast has filled some vessels 2009 AAPM Meeting-Phil Rauch Test Your Knowledge Why are these image so different? 2009 AAPM Meeting-Phil Rauch Test Your Knowledge Why are these image so different? 72 kVp, 250 mA, 8 ms 1024 x 1024 x 8 bits 2009 AAPM Meeting-Phil Rauch 65 kVp, 15 mA, 13 ms 512 x 512 x 8 bits 2009 AAPM Meeting-Phil Rauch 26 Image Quality Analysis Spectral filtration Bolus filters Generator power curves Pulse width Grid vs Non-grid Collimation Detector dose per image Anatomical program selection Recursive filter setting Edge enhancement Harmonization (Dynamic Range( Pixel matrix and bit depth for acquisition Pixel matrix and bit depth stored Pixel matrix and bit depth archived 2009 AAPM Meeting-Phil Rauch Image Interpretation Successful interpretation depends on… Image Attributes Human Observer (education, experience) Human Visual Perception Pattern Recognition Clarity of the imaging task Minimal external distractions Level of confidence 2009 AAPM Meeting-Phil Rauch Image Quality: What is it? “Image quality depends only on intrinsic, objective physical characteristics of an imaging system, and can be measured independently of an observer” “Image quality is whatever the observer says it is (i.e., it is a subjective perception of the image, ‘in the eye of the beholder’)” “Image quality is defined by an observer's ability to achieve an acceptable level of performance for a specified task” Definitions courtesy Ralph Schaetzing, Agfa Corp. 2009 AAPM Meeting-Phil Rauch Conclusions Image quality is not just detector metrics Image quality is not just a subjective impression Image quality for fluoroscopy must be assessed with moving objects in order to get a complete understanding of performance Use pulsed fluoroscopy and the Aufrichtig scale to produce better images, not solely to reduce dose 2009 AAPM Meeting-Phil Rauch 27 Final Imaging Task How can this be true? Thank you Below the four parts are reordered Phil Rauch philr@rad.hfh.edu The four parts are exactly the same as those used above Why is there an extra square? 2009 AAPM Meeting-Phil Rauch 28