TG-104: In-room kV Computed Tomography for Image-guidance D.A. Jaffray, Ph.D. Radiation Medicine Program Princess Margaret Hospital/Ontario Cancer Institute Professor Departments of Radiation Oncology and Medical Biophysics University of Toronto AAPM’10 Acknowledgements • Princess Margaret Hospital – Douglas Moseley, Michael Sharpe, Elizabeth White, J.P. Bissonnette, Bern Norrlinger, Jason Smale, D. Letourneau, Winnie Li Disclosure Research agreements funded with the following commercial entities: Elekta, Philips, Raysearch The presenter has financial interest in two technologies reported here: (i) cone-beam CT imaging technology, and (ii) a phantom for daily QA. AAPM’10 Learning Objectives • Become aware of guidance documents for IGRT QA. • Understand in-room kV computed tomography IG technologies. • Familiar with the level of targeting performance these systems can achieve. AAPM’10 Volumetric kV CT Systems Siemens PRIMATOM™ Elekta Synergy™ Siemens MVision™ Varian OBI™ Siemens Artiste™ kV CT Approach kV and MV Cone-beam CT Approach AAPM’10 kV Fluoroscopic, Radiographic, and CT Functionality Fluoroscopic Radiographic Tomographic AAPM’10 AAPM’10 TG-104 The role of kV imaging in patient setup. Description. IGRT Workflow. Specific Processes. Manpower. AAPM’10 IGRT is Quality Assurance • Provides measurement of patient position in treatment position. – – – – Quantitative, accurate, repetitive Minimally invasive Large field-of-view Markers, bone, soft-tissue, skin-line • Verify consistency of planned and actual geometry – Provides a critical data source for rational margin design AAPM’10 PTV Design + IGRT • IGRT is part of a system-wide QA activity – Geometric quality assurance / uncertainty management • Treatment localization reduces, but does not eliminate, geometric uncertainties • Reduced uncertainties may allow reduced PTV margins • PTV margin selection relies on assumptions about stability of guidance performance. AAPM’10 AAPM’10 From “Modern Technology of Radiation Therapy – Supplement (Vol. 2)” Ed: J. van Dyk IG Systems, Processes, and QA for CBCT Systems Jaffray ,Craig, Bissonnette -2003 Yoo et al. 2006 • Quality Assurance program specific to Varian On-Board Imager (OBI) • Includes specific mechanical tests associated with robotic arms. • General concepts applicable to all vendors Med. Phys. 33 (11) November 2006 AAPM’10 AAPM’10 * Recommendations for Imaging System QA AAPM’10 * Recommendations for Imaging System QA AAPM’10 * Recommendations for Imaging System QA In-room Conventional CT for IGRT Memorial Sloan Kettering Cancer Center AAPM’10 Siemens CT-on-Rails System AAPM’10 CT – Linac system with RTRT Yamaguchi University Hospital, Japan AAPM’10 In-room Conventional CT for IGRT Kuriyama et al. Int.J.Rad.Onc.Biol.Phys. 55(2) Feb 2003 AAPM’10 In-room Conventional CT for IGRT Positional Accuracy: 0.2 mm (LAT) 0.18 mm (VERT) 0.39 mm (LONG) Onishi et al. Int.J.Rad.Onc.Biol.Phys. 56(1) May 2003 AAPM’10 Varian CT-on-Rails at MDACC AAPM’10 Central guide rail Magnetic encoder strip Side rail to provide balance Fig. IIA.-3 There are three rails in this moving-gantry CT scanner. The central rail contains positional sensor and drive mechanism; and the two side rails provide level and balance during movement. Fiducial Transfer Method AAPM’10 kV Cone-Beam CT for IGRT Robust 2D Detector Elekta Synergy (XVI) Varian iX (OBI) Feasible Reconstruction Method kV IG Systems: Front of Mind • Baseline Radiographic/Fluoroscopic Performance • Stability of Geometric Calibration/MV-kV Correspondence • Dosimetry of Imaging Techniques • Baseline Performance of Image Quality • Image-guidance Performance (Phantom Orientation, Residuals) • Automated Couch Adjustment Performance AAPM’10 kV IG System Commission/Test X-ray Tube kVp Linearity Timer accuracy mR/mAs HVL CBCT Image Quality Uniformity Spatial Resolution CNR, Low-Contrast CT # (linearity) AAPM’10 Automatic Couch Accuracy Reproducibility Flat-Panel Linearity [ADU/mR] Defects Dark noise Lag MV/kV Iso-centre Coincidence Stability kV/MV Calibration Concept Winston-Lutz et al, IJROBP 14 1988 y MV Mechanical Iso-centre kV MV Radiation Iso-centre x z AAPM’10 Calibrated Iso-centre Reconstruction Iso-centre kV/MV Calibration Concept – Elekta Flex Maps: Synergy (XVI) Units FOV: 26 cm Rot. Dir: CW XVI1 XVI2 XVI3 XVI4 XVI5 26 cm CCW 40 cm CW 40 cm CCW 50cm CW 50 cm CCW All Flex Maps: Synergy (XVI) Units AAPM’10 OBI1 Approach – “Residual” AAPM’10 Accept if within specified tolerance. Sensitivity of Cone-beam CT Performance to Geometric Miscalibration • Shift panel 5 mm in “u” • E.g. Nasopharynx v u AAPM’10 Sensitivity of 2D kV Radiographic Localization Performance to Miscalibration • Image quality is not sensitive to detector/source shifts • No physical reference marks – i.e. cross-hair or field-edge • Electronic field size/isocentre * Vendor Notices AAPM’10 Clinical Assessment of Remote Couch for IGRT Background • The XVI system is linked to the Remote Figure 2: Frequency Histograms for Translational x, y, z & Rotational x, y, z Automatic Table Movement (RATM). • The patient is shifted per the imageguidance system via the remote couch interface. Sarcoma 9% Lymphoma 3% • The stability of the system and residual error is measured through verification scans acquired following a table shift. Upper GI 21% Head and Neck 26% Lung 41% Methodology Figure 1: Patient Population by Tumor Site • Collection Period Oct 19th ‘06 – Nov 17th ‘06 • Patients with repeat (verification) scans were measured and matched using an Automatic Algorithm • 34 patients AAPM’10 with 135 scans W Li et al. J Appl Clin Med Phys. 2009 Oct 7;10(4):3056 IG Performance: Connectivity,Orientation/Scale Checks AAPM’10 Image Source: GE and Philips CT Planning: Philips Pinnacle v7.4 IG Performance: Connectivity,Orientation/Scale Checks Residual Error • • • • • • Anthropomorphic phantom 4 Orientations Target bony anatomy Arbitrary initial shifts Plot residual error 5 XVI units AAPM’10 IG Performance: Connectivity,Orientation/Scale Checks Residual Absolute Error [cm] 0.18 0.16 0.14 0.12 L/R S/I A/P 0.10 0.08 0.06 0.04 0.02 0.00 1 All XVI’s AAPM’10 Daily Geometry QA • Align phantom with lasers • Acquire portal images (AP & Lat) & assess central axis • Acquire CBCT • Difference between predicted couch displacements (MV & kV) should be < 2 mm AAPM’10 Daily Geometry QA • Align phantom with lasers • Acquire portal images (AP & Lat) & assess central axis • Acquire CBCT • Difference between predicted couch displacements (MV & kV) should be < 2 mm AAPM’10 Full kV/MV Calibration - Monthly AAPM’10 AAPM’10 AAPM’10 Compare Portal Image & DRR AAPM’10 Iso_ODI = Q/A Phantom = Lasers/ODI E2 (CBCT) E1 (EPID’s) Iso_kV MV Mechanical Isocenter E3 MV Nomimal Isocenter z y x Iso_MV MV Radiation Isocenter Current* Action Level on E 3: 2 mm (in any one direction) AAPM’10 * Due to large observer variability in MV alignment 2006 XVI Daily QA Results (E3 Error) XVI1 XVI2 [mm] M XVI3 L/R XVI4 S/I A/P (-0.04, -0.72, 0.52) Ave(σ) (0.96, 0.82, 1.04) ~ 6 Months AAPM’10 of Daily QA Testing XVI5 2007 OBI Daily QA Results (E3 Error) OBI1 [mm] OBI2 L/R OBI3 S/I OBI4 A/P M = (-0.08, -0.19, 0.27) Ave(σ) = (0.51, 1.02, 0.71) AAPM’10 TG-104: QA Summary • Guidance documents are available to assist in the establishment of QA programs for IGRT technologies. • Published literature demonstrate that these systems can be accurate, precise, and reliable. – Compare your results to others. • Maintenance of IGRT performance is central to confidence in appropriate PTV margin. • An integrated daily check for IG system consistency has been implemented into routine clinical use with a 15 minute time penalty. AAPM’10 AAPM’10 kV Conebeam CT Dose AAPM’10 Patient Dose Estimation: kV-CBCT Dose depends on • Beam Quality: HVL (kVp, filtration) • Tube output: Reference (mR/ mAs) • Scanning Geometry: SAD, FOV, No. of projections • Technique settings: mAs • Patient Size (Body , Head…) AAPM’10 Scanning Geometry and Technique Offset Detector Asymmetric Col. 26 cm X1=6.0 cm, X2=20 cm 41 cm x 41 cm 155 cm Elekta Precise: XVI AAPM’10 Bench top CBCT System AAPM’10 CBCT: Radial Dose Phantom: 16 cm dia. 2.4 2.2 Imaging Technique: 2.0 100 kVp 20 ms 330 Projection Depth 660 mAs Dose (cGy) 100 mA 1.8 1.6 1.4 FOV: 5 cm x 26cm FOV: 10 cm x 26cm FOV: 15 cm x 26cm FOV: 26 cm x 26cm 1.2 1.0 0.8 0.6 0 2 4 Depth (cm) AAPM’10 6 8 CBCT: Radial Dose 2.4 FOV: 5 cm x 26cm FOV: 10 cm x 26cm FOV: 15 cm x 26cm FOV: 26 cm x 26cm 2.2 2.0 1.8 Dose (cGy) Phantom: 30 cm dia. Imaging Technique: 120 kVp 100 mA 20 ms 330 Projection 660 mAs Depth 1.6 1.4 1.2 1.0 0.8 0.6 0 2 4 6 8 Depth (cm) AAPM’10 10 12 14 Dose Comparison: Dose at Isocenter Dose in cGy/ 640 mAs; FOVx= 26 cm Phantom FOVz Size: dia. 30 cm 16 cm AAPM’10 100 kVp 120 kVp 10 cm 0.58 1.07 26 cm 0.80 1.5 10 cm 1.52 2.68 26 cm 1.9 3.2 CBCT Imaging Dose Experimental Setup Dose vs. Field Size Total Dose (cGy) 2 D2cm 1.5 1 Dcenter 0.5 0 0 10 20 30 Field Size (z) (cm) Nuclear Enterprises Free-Air Chamber (0.6 cc) 32 cm “Body” Phantom 330 projections at 2 mAs / proj AAPM’10 Offset geometry (FOV 40 cm) Islam et al., Med. Phys. 2006 Deciding Upon the Necessary Image Quality for the Application Variation of image quality with lens dose (cGy) mAs/ Projection 2 1.0 2.0 0.5 1.0 2.0 0.25 0.5 1.0 4.0 1 0.5 80 160 320 Number of Projections 16x Reduction Low Dose (1.5 mGy) Pediatric Imaging for Routine On-line IGRT AAPM’10