Dose and Image Quality in C-arm CT Rotational Angiography R Fahrig, E Girard-Hughes , A Ganguly, A Tognolini , N Kothary Collaborators/Support o The Physics Gang:: ‘Bob’ R.L.Dixon, ‘Tom’ J. Payne, ‘Rick’ R.L. Morin o A. Ganguly, N. Strobel and T. Moore o L. Hoffman, N. Kothary, D. Sze, M. Marks, H. Do o Technical support : A. White, N.R. Bennett, J. Kneebone, M. Lozada-Parks, W. Baumgardner o Siemens Medical Solutions o NIH R01 EB003524 o Lucas Foundation Department of Radiology, Stanford University Introduction o C-arm CT for visualization in 3D of vasculature and other high-contrast structures has become commonplace o The transition from XRIIs to digital flat panels opened the doors to the possibility of lowcontrast 3D CT imaging in the interventional suite o What doses are likely? How do current settings compare to clinical CT doses? o How is low-contrast visibility affected by choice of kVp? o How does the AEC system on the C-arm affect dose and low-contrast visibility? C-arm System :: CT System half scan, area detector vs. full scan/narrow detector Creating 3D Images in the Interventional Lab 1) Rotational Angiography Run 4) In-room Display Topics of Discussion o First question : Dose vs. Image quality : Neuroimaging • Standard dosimetry with 16-cm head phantom • image quality comparison as a function of dose o Second question : Dose vs. Image quality : Body Imaging 3) Reconstruction and Visualization 2) Image transfer Dose Measurement o small 0.6cc Ion Chamber o measuring maximum dose at center of z-extent of the scanning range in an appropriate phantom o See AAPM TG111 Report for full protocol o Note that Farmer chamber calibration is typically carried out using Co-60; special correction factors are required for low-HVL dosimetry • Automatic Exposure Control with custom phantom • FOV reduction (slab imaging) improves conspicuity o Third question : Does 3D information Increase or Decrease dose in clinical use? Farmer calibration o non-standard calibration provided by Radcal Corporation kVp 70 81 109 125 HVL(mmAl) 2.9 3.2 3.4 4.4 Dose Measurement o CTDI phantom (16cm diameter, 15cm long) o Dose measured at center and eight peripheral positions for : (30x40) cm detector format based on 543 views o Beam Size (iso-center): Width: 26.67cm Height: 20.00cm 16 cm NEUROIMAGING 32 cm Dose Measurement o CTDI phantom (16cm diameter, 15cm long) o Dose measured at center and eight peripheral positions for : (30x40) cm detector format based on 543 views o Beam Size (iso-center): Width: 26.67cm Height: 20.00cm 16 cm 32 cm Dose Measurements: 81kVp 2/3 mean peripheral + 1/3 central Measured Doses : ‘Medium-High’ dose requested Dose Measurements: 81kVp D (0) = (1 / 3) D0 + ( 2 / 3) D p 2/3 mean peripheral + 1/3 central source side kVp Total mAs Peak Dose (mGy) 86 Center Dose (mGy) 34 “CTDIw” (mGy) 1167 Detector dose (uGy/view) 0.46 70 81 608 0.44 63 28 37 109 310 0.70 66 31 40 125 260 0.92 76 38 46 Variation in CTDIw in spite of AEC. The EU guidelines for routine head CT scans specify a CTDIw of 60mGy. detector side Visibility vs. Dose Visibility of Low-Contrast Objects? Visibility Chart (81kVp, 543view s) Nominal Contrast 100 U) (3H 90 80 9mm 0% 6mm 0 (1 ) HU Detail Diameters [mm] (2, …, 9, 15) 70 Visibility (%) 15mm 0.5% (5HU) % 0 .3 1. o Catphan Module CTP515 used as image quality phantom (20cm housing) o Acquired 543 views over 20sec at various dose and kVp settings, Zoom 0 o Reconstructed soft tissue segment (smooth kernel, 10mm slice width) o Analyzed visibility of (outer) 5HU insets 48 60 50 40 CTDIw = 19.93mGy CTDIw = 26.31mGy 30 20 CTDIw = 37.01mGy CTDIw = 54.12mGy 10 0 0 Scoring Question: What size “5HU” objects can you see? 3 6 9 Detail Diameter [mm] 12 15 d ref = Normalized Visibility : D ( 0) ⋅d Dref (0) Visibility vs. kVp 120 D (0) ⋅d Dref (0) 120 100 Visibility [%] 100 80 Visibility [%] d ref = 60 80 60 40 Average (70kVp) 40 Cat: ex 38.41mGy Average (81kVp) 20 Cat: 56.63mGy Average (109kVp) Average (125kVp) Cat: ex 81.73mGy 20 Average (56.63mGy) 0 0 2 4 6 8 10 12 14 16 Normalized Diameter [mm] 0 0 2 4 6 8 10 12 14 16 9mm scoring objects with contrast of 5HU visible in over 90% of all cases Normalized Diameter [mm] MTF: 100 µm steel wire Intracranial Imaging C-arm CT Clinical CT 1.0 In vivo pig model, Autologous blood, NO iodine contrast Sharp 0.8 Normal MTF Smooth 0.6 ARTIFACT: 0.4 Beam hardening Scatter Conebeam 0.2 0.0 0.0 0.5 1.0 1.5 2.0 2.5 Spatial Frequency (cycles/mm) M. Marks, H. Do et al. Conclusions - Neuro Materials Image Quality o Place the ‘Siemens 16cm image quality ConeBeam’ phantom inside a body-shaped phantom o 168 negative contrast objects in three slices : 3, 5, 10, 15, 20, 25, 30, 45, etc. HU, 32, 16, 8, 4 and 2 mm diameter Image quality slices of interest BODY IMAGING Body Phantom Design 50 cm 37 cm Circumference = 104cm 16 cm 26 cm o Better visibility at lower energy (kVp) offers potential for image acquisition protocol optimization o We can detect 9mm scoring objects (nominal contrast “5HU”) in over 90% of all cases (70kVp through 125kVp) o The dose applied to obtain this image quality performance is close to the EU guideline for CT head scans (60mGy) o The in-plane spatial resolution of the system is excellent. • IQ 16-cm conebeam phantom was centered inside the body phantom and two additional 50cm solid sections added to provide adequate scatter Dose Measurement Insert • 13.1-mm diameter holes in 16-cm center (same geometry as CTDI phantom) • six holes at the half way points • the eight holes at the edge of the green outer ring are placed such that distance from center of the hole to the edge of phantom = 10 mm AEC kVp for 109 kVp Request Experiment o Imaged at 3 different collimations : 28 cm (22.2 cm @ isocenter) 17.5 cm (14 cm @ isocenter) 10.5 cm (8.4 cm @ isocenter) o Smallest collimation still included the full dominant for automatic exposure control o DynaCT, 0.5 degrees per step, large focus, 8-s scan, 396 projections/reconstruction o kVp = 109 or kVp = 125 kVp o dose request = 0.36 uGy, 0.54 uGy, or 0.81 uGy per image at the detector AEC mAs Impact of Scatter on Image Quality o Images of a 16-cm ‘medium contrast’ insert in the abdomen phantom, 0.8 mm slice thickness o Objects are : -20, -25, -30 and -45 HU 32, 16, 8, 4 and 2 mm diam 109 kVp 125 kVp Impact of Scatter on Image Quality o Images of a 16-cm ‘low contrast’ insert in the abdomen phantom, reconstructed with 5.0 mm slice thickness o Objects are : -3, -5, -10 HU and -15 HU 32, 16, 8, 4 and 2 mm diam Detectibility @ 125 kVp o Sum up the number of visible objects in each of the three slices : high-contrast, medium-contrast and low-contrast : recon 0.5x0.5x1.0 mm o Three ‘group’ reads of each dataset at 125 kVp as a fn. of collimation as a fn. of dose requested Conclusions – Body Imaging o Need to calibrate AEC behavior for a range of body sizes; our phantom represents a very un-circular (slim?) case o Body imaging pushes the x-ray tube limits : • Higher kVp needed to increase tube output per mAs and to increase patient penetration CLINICAL DOSES o Slab imaging increases detectibility and decreases average dose in max. slice (as well as decreasing total vol. irradiated) TACE Rationale Chemoembolization: Procedure/Technique o Ischemia to tumor • Susceptible to therapy o Increase concentration compared to IV infusion o Increased dwell time o Reduced systemic toxicity o Aortogram o Visceral arteriography • Delineate vessels (i.e. cystic artery) o Selective catheterization of target artery supplying tumor L. Hoffman et al. Augmented by 3D Chemoembolization: A clinical study o Acquisition: 419 images, 8 second scan through 210°°, 512x512 reconstruction, 0.36 µGy per frame o Injection : 50-50 mix of iodinated contrast and saline o technically successful in 93/100 procedures (93%) o provided information not available by DSA in 30 patients (35.7%) o resulted in a change in diagnosis, treatment planning or treatment delivery in 24 patients (28.6%) L. Hoffman et al. Methods- Imaging protocol Methods- Imaging protocol DSA study arm (reference standard): C-arm CT study arm: o DSA groin run at 42 cm FOV Aortogram 30cc @ 15 cc/sec o Non-contrast CACT if patient had residual ethiodol from previous chemoembo o AP and 30º RAO, 21 ml @ 3ml/sec from proper/common hepatic artery (~ 3 frame/s, w/ late parenchymal phase) POWER INJECTION ONLY. o Proper hepatic CACT 12cc from proper/common hepatic (4 sec X-ray delay)- 8 sec rotational scan (~ 200°) with image acquisition every 0.5° (419 images) o Additional DSA as required o Additional oblique DSA as needed o Single DA image of the RUQ and completion post-embolization angiogram (hand) o Additional selective CACT if needed for problem solving o Post-treatment non-contrast CACT Methods: C-arm CT Navigation Methods- data collection Methods- additional data collection Radiation dose endpoints (dose report): o Total Dose Area Product (DAP: µGym²) o DAP for guidance • from 5F catheter in common/proper hepatic to superselective positioning of microcatheter o Fluoro time (min) for total procedure o Fluoro time for guidance Time endpoints (procedure log sheet): o Total procedure time (min) o Time for guidance o For DSA arm • Contrast enhanced CACT performed, however images not reconstructed till microcathter positoned • Post chemoembolization non-contrast CACT (instead of unenhanced helical CT) to ensure complete geographic uptake in the tumor Methods- DSA study arm Methods- CACT study arm Total DAP: 14965.6 µGym² Guidance DAP: 4425.6+860.3= 5285.9 µGym² Total DAP: 35812.2– (6559.6+6499.3)= 22753.3 µGym² Guidance DAP: 3944.6+6260.9+1967.6= 12173.1µGym² Results - DAP (µGym²) CACT study arm Mean: 16513.9 Guidance Median:14690.3 (range: 5012-33628) Total dose (includes fluoro ) DSA study arm Mean: 15143.5 9.5% Median:13406.6 (p= 0.588) (range: 2567-30684) Mean: 36567.4 Mean: 27417. 2 Median: 31945.1 Median: 25134.1 (range:14716-65270) (range:4717-65786) (*includes w/ CACT post) % dose increment (median) 27% (p= 0.069) (mean DAP w/CACT 11% post: 32917.2) (p=0.4) Results - Total and Guidance Time CACT study arm DSA study arm % increase (median) Total time (min) Mean: 100 Median:90 (range:62171) Mean: 80 Median:76 (range: 30-141) 18.4% (p= 0.08) Total time from CHA to completion (min) Mean: 65 Median:58 (range: 32- 108) Mean: 57 Median:54 (range: 27-98) 7.4% (p= 0.19) Guidance time (min) Mean: 33 Median: 26 (range: 7-51) Mean: 31 Median: 25 (range: 9-74) 4% (p= 0.54) Results : Fluoro time/Contrast volume CACT study arm DSA study arm % increase (median) Total fluoro time (min) Mean: 26 Median: 23 (range: 9-65) Mean: 22 Median: 20 (range:10-43) 15% (p= 0.17) Fluoro time guidance (min) Mean: 8 Median: 6 (range: 2-20) Mean: 20 Median:17 (range: 9-40) -64% (p= 0.0001) Contrast- P. Injector (ml) Mean: 48 Median: 45 (range:27-75) Mean: 86 Median: 82 (range: 63-120) -45% (p= 0.0001) Contrasttotal (includes hand runs) Mean: 78 Median: 75 (range:72-120) Mean: 95 Median: 100 (range: 60-130) -25% (p= 0.0001) Conclusion o C-arm CT does not significantly add to the radiation dose, procedural time or catheterization time o C-arm CT potentially can reduce operator dose by decreasing fluoroscopy time o C-arm CT reduces overall contrast volume • Particularly important in this group of patients that are predisposed to nephrotoxicity due to underlying cirrhosis Context CARDIAC IMAGING o Single sweep in the interventional suite provides geometry of cardiac chambers for guidance of RF ablation o Contrast injected into the chamber of interest during acq. o System lateral • tube rotates from +10º above horizontal, around patient back to +10º above horizontal Automatic Registration o pre-registration of EAM system and C-arm CT FOV could shorten RF procedures by eliminating timeconsuming point-by-point registration as is done currently with prior CT or MR images Monte Carlo-based Dose o Excellent paper : “Effective dose analysis of three-dimensional rotational angiography during catheter ablation procedures” • J-Y Wielandts, K Smans, J Ector, S De Buck, H Heidbuchel and H Bosmans Phys. Med. Biol. 55 (2010) 563–579 o Monte Carlo used to calculate BMI-specific organ doses using recorded (and variable!) kVp/mAs and assumed geometry o 5-s sweep with 249 projections, 60 fps Monte-Carlo Results Multi-sweep Gated… From Phys. Med. Biol. 55 (2010) 563–579 o Specific organ doses that are of interest due to highest weights • Breasts, lung, bone marrow (sternum), lung o Timing the return of each rotation properly provides sufficient data for a reconstruction of ¼ of the cardiac cycle e.g. in diastole 4x5s Sweep, ECG gating Large (30x40 cm) Detector Patient 6: with ECG gating o 4 sweeps, 5s per sweep, ~250 projections per sweep o Total scan time ~24 s (including time for C-arm turn around) : total breath hold ~ 27 s o 12-15 s delay between start of injection and start of imaging o Injection of ~140 ml contrast (3:1 dilution) at 4 ml/s into inferior vena cava Patient 6: no ECG gating Comparison of Protocols Rando Phantom Organ Dose? Multi-sweep Cardiac Dose? o Measured organ doses according to standard prescriptions for dosimeter locations (solid state MOSFET dosimetry system calibrated for diagnostic energies) but for our imaging protocol o Compared against Organ thyroid esophagus lung heart MC simulations seems to be Rando 0.44 5.71 5.93 6.80 significantly lower (mGy) Monte o Based on 4xMCavg Carlo 0.94 13.05 23.35 16.6 for 82 kg person (mGy) our gated cardiac scans would give Scale 0.47 0.44 0.25 0.41 22 mSv o Based on 4xMCavg for 82 kg person our gated cardiac scans would give 22 mSv o Clearly individual organ scaling factors indicate the dose will be lower o Clinical un-gated CT cardiac scans are between 10-16 mSv o Our ‘constant energy’ system setting was 0.18 kJ vs. MC-paper value of 0.25 kJ o Need to expand Rando measurements to ‘large’ patients to stress the system From Phys. Med. Biol. 55 (2010) 563–579 C-arm CT vs. Clinical CT Image quality • Streak artifacts due to remaining motion • Contrast brightening and darkening due to beam hardening and scatter o o o o o o o Intra-procedural Single circle scan 1000 slices Rotation time 5-10s Volume in 5-10s See better than 10 HU AEC! o o o o o o o Diagnostic Spiral scan 64 slices Rotation time 0.3s Volume in 5-10s See better than 3 HU mA modulation? Conclusions o Dosimetry for rotation C-arm CT angiography is complicated by AEC o Careful evaluation of dose vs. image quality for a particular application is required o We have the necessary tools…