Digital Mammography Update: Design and Characteristics of Current Systems July 27, 2009

advertisement
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Digital Mammography Update: Design
and Characteristics of Current Systems
Kalpana M. Kanal, Ph.D., DABR
Assistant Professor
Department of Radiology
University of Washington
Seattle, Washington
AAPM Annual Meeting 2009 – Anaheim, CA
Educational Objectives
™
™
™
™
Kanal
Kanal
Understand the physics of digital detector technology
Recognize that vendors use varying detector technology
in FFDM systems
Appreciate the advantages and disadvantages of digital
mammography systems
Understand briefly the image quality metrics such as
DQE and MTF
2
1
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Resources
™
™
™
™
™
Doug Pfeiffer,
Pfeiffer, MS, DABR
Digital Mammography: An
overview – Dr. Mahesh
(Radiographics
2004;24:17472004;24:1747-1760)
Fundamentals of Digital
Mammography Primer – Dr.
Smith (Hologic Inc)
Digital Mammography –
Pisano and Yaffe (Radiology
2005; 234:353234:353-262)
Bloomquist and Yaffe – Med
Phys 33 (3), 2006
Kanal
3
Introduction
Kanal
2
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Full-Field Digital Mammography (FFDM)
- Screen-Film
c.f. www.gehealthcare.com/rad/xr/education/dig_xray_intro.html
Kanal
5
Full-Field Digital Mammography (FFDM)
™
™
™
™
™
Kanal
Kanal
Wide dynamic range
(1000:1) compared with
SFM (40:1)
Dynamic image
manipulation
Ability to postpost-process
SoftSoft-copy read
accompanied by
computercomputer-aidedaideddiagnosis (CAD)
3D imaging
Radiographics 2004:24,1749
6
3
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Full-Field Digital Mammography (FFDM)
™
Advantage
™ Each component of the mammographic process can
be optimized with digital mammography
Radiographics 2004:24,1750
Kanal
7
SFM vs. FFDM
SFM: Half mAs, Automatic exposure control, Double mAs
FFDM: Same technique factors as SFM, W/L adjusted
Kanal
Kanal
Radiographics 2004:24,1750
8
4
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
SFM vs. FFDM
Kanal
Radiographics 2004:24,1751
9
FDA and Digital Mammography
Kanal
Kanal
Butler - http://www.acr.org/accreditation/mammography/rsna07presentation.aspx
http://www.acr.org/accreditation/mammography/rsna07presentation.aspx
10
5
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
MQSA Scorecard
™
™
™
™
™
Kanal
Certification statistics, as of July 1, 2009
Total certified facilities / Total accredited units
™ 8,714 / 12,875
Certified facilities with FFDM units / Accredited FFDM
units
™ 4,659 / 6,964
53% certified facilities with FFDM units
54% accredited FFDM units (38% last year)
http://www.fda.gov/cdrh/mammography/scorecardhttp://www.fda.gov/cdrh/mammography/scorecard-statistics.html
11
Technologies for FFDM
Kanal
6
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM
Indirect Capture
™
™
™
a scintillator such as
cesium iodide (CsI)
absorbs xx-rays and
generates a light
scintillation
detected by an array of
photodiodes or chargechargecoupled devices (CCDs)
Resolution degradation
http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf
Kanal
13
Technologies for FFDM
Direct Capture
™
™
Kanal
Kanal
X-ray photons are directly
captured by a
photoconductor such as
amorphous selenium (a(aSe), which converts the
absorbed xx-rays directly to
a digital signal
Spatial resolution limited
to pixel size
http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf
14
7
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Various Approaches in development of
FFDM systems
™
™
Indirect
™ A single flatflat-panel scintillator and an amorphous silicon
(aSi)
diode
array – GE
(a
™ Slot scanning with scintillators and CCD arrays – Fischer
Imaging, now Hologic
™ Photostimulable phosphor plates - Fuji
Direct
™ A flatflat-panel amorphous selenium (a(a-Se) array – Hologic,
Siemens
™ Tiled scintillators with fiberoptic tapers and mosaic CCD
arrays – Hologic Stereotactic
Kanal
15
Technologies for FFDM – Indirect Capture
GE
GE Senographe 2000D Full Field
Digital Mammography System on
01/28/00
GE Senographe DS Full Field
Digital Mammography (FFDM)
System - 02/19/04
GE Senographe Essential Full
Field Digital Mammography
(FFDM) System - 04/11/06 (larger
detector)
Kanal
Kanal
Fig. Ref: GE DS QC manual
16
8
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Indirect Capture
GE
™
™
™
In this system, the digital
detector array is constructed
from an aa-Si thinthin-film transistor
(TFT) matrix deposited on a
glass substrate
The CsI scintillator is deposited
on the aa-Si detector
Each lightlight-sensitive diode
element is connected by TFTs
to a control and a data line so
that charge produced in the
diode is read out in response to
light emission from the scintillator
Radiographics 2004:24,1753
Kanal
17
Technologies for FFDM – Indirect Capture
GE
Detector
size
Pixel size
Limiting
Spatial
Resolution
Image size
Bit Depth
Kanal
Kanal
2000D
DS
Essential
19.2 x 23.0
19.2 x 23.0
24.0 x 30.7
100 µm
100 µm
100 µm
5 lp/cm
5 lp/cm
5 lp/cm
1914 x 2294
1914 x 2294
2394 x 3062
pixels (9 MB) pixels (9 MB) pixels (14 MB)
14
14
14
Radiographics 2004:24,1753
18
9
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Indirect Capture
GE – Advantages and Disadvantages
™
™
™
™
™
™
Close bonding between CsI and aa-Si ensures minimal
light loss
Strong signal from the Si diode array yields higher
detective quantum efficiency
Detector is linear over a wide range (105) of exposures
Limiting factor is the large pixel size (100 μm)
Smaller pixel sizes improve spatial resolution but at the
cost of increased image noise and decreased SNR for
the same breast dose
Possibility of ghosting in images
Radiographics 2004:24,1753
Kanal
19
Technologies for FFDM – Indirect Capture
Computed Radiography – Fuji (7/10/2006)
Technology introduced in 1981
PSP
(Barium fluorobromide)
x-ray exposure
Base support
plate exposure:
create latent image
Kanal
Kanal
Pfeiffer
20
10
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Indirect Capture
Computed Radiography – Fuji
Technology introduced in 1981
laser beam scan
plate readout:
extract latent image
light erasure
plate erasure:
remove residual signal
Pfeiffer
Kanal
21
Technologies for FFDM – Indirect Capture
Computed Radiography – Fuji
™
Fuji FCRm, Dual-side reader
Detector
size
Pixel size
Image size
Spatial
Resolution
Dynamic
Range
Kanal
Kanal
18 x 24
24 x 30
50 µm
3328 x 4096
pixels (24 MB)
10 lp/mm
14 bits
http://www.fujimed.com/
22
11
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Fuji - Dual Side Readout
photodetector
Protective
layer
laser beam
optical
guide
mirror
Phosphor
layer
Transparent
support
emission
optical
guide
photodetector
Image courtesy: Doug Pfeiffer
Kanal
23
Technologies for FFDM – Indirect Capture
Fuji – Advantages and Disadvantages
™
™
™
™
™
™
Kanal
Kanal
Advantage of the CR mammo system is that filmfilm-screen
cassettes can be replaced by CR cassettes without replacing
the entire system
Both small and large cassettes can be accommodated by the
reader
Dual side reader, 50 μm pixel size
Disadvantage - Effective pixel size influenced by phosphor
thickness, light diffusion within phosphor, laser light scatter &
diameter of laser beam
Technologist time on processing of images
Noise associated with the low collection efficiency of emitted
light
24
12
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Indirect Capture
SenoScan; Fischer Imaging – 9/25/01
™
™
Kanal
A narrow slotslot-detector and
a narrow fan beam of xxrays are scanned
synchronously across the
full field of view to cover
the entire breast
System consists of
phosphor (thallium(thalliumactivated CsI) with a
fiberoptic coupling to a
CCD
Radiographics 2004:24,1752
25
Technologies for FFDM – Indirect Capture
SenoScan; Fischer Imaging (now Hologic)
Seibert, AAPM Meeting Handout
Kanal
Kanal
26
13
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Indirect Capture
(SenoScan; Fischer Imaging (now Hologic))
Advantages and Disadvantages
™
™
™
™
™
Compact detector that is less expensive compared to
others
Excellent scatter rejection due to small volume of breast
exposed at any time
No grid needed therefore less dose
Longer compression since scan times are longer
(approx. 6 sec)
Powerful tubes, elaborate signal readout and image
reconstruction required
Kanal
27
Comparison – Indirect Capture
Fuji FCRm
FCRm
PSP plate
18 x 24
24 x 30
3328 x 4096
pixels (24 MB)
50 µm
10 lp/mm
Kanal
Kanal
28
14
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Direct Capture
a-Se Array (Selenia, Hologic) – 10/2/01
™
™
™
a-Se, a good photoconductor is deposited directly onto
the aa-Si TFT substrate enabling direct capture
The aa-Se detector directly converts xx-rays to electronelectronhole pairs
The aa-Si TFT converts the electronelectron-hole pairs to
electronic signal
http://www.hologic.com/wh/digisel.htm
Kanal
29
Technologies for FFDM – Direct Capture
Flat-Panel a-Se Array (Selenia, Hologic)
Detector size 24.0 x 29.0
Kanal
Kanal
Pixel size
70 µm
Image size
3328 x 4096
pixels (24 MB)
Spatial
Resolution
> 7 lp/cm
Dynamic
Range
14 bits
http://www.hologic.com/wh/digisel.htm
30
15
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Technologies for FFDM – Direct Capture
Flat-Panel a-Se Array (Selenia, Hologic)
Advantages and Disadvantages
™
™
™
™
Advantage is that the detector response function
maintains its sharpness even with increasing thickness
High MTF and DQE can be achieved
Potential weaknesses are the need for high biasing
voltage, drifting of the dark signal and cost of detector
Inherent sharpness of detector may also increase the
severity of aliasing artifacts associated with
undersampling on any digital detector
Kanal
31
Technologies for FFDM – Direct Capture
a-Se Array (Mammomat Novation, Siemens – 8/20/04)
Detector size
24.0 x 29.0
Pixel size
70 µm
Image size
3328 x 4096
pixels (24 MB)
Spatial
Resolution
Dynamic
Range
Kanal
Kanal
> 7 lp/mm
14 bits
http://www.medical.siemens.com
32
16
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Comparison – Direct Capture
Kanal
33
Technologies for FFDM
Fuji/Kodak/
Agfa/Philips
Kanal
Kanal
Fischer
(Hologic)
GE
Hologic
Siemens
http://www.hologic.com/oem/pdf/DROverviewR-007_Nov2000.pdf
34
17
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
MTF, DQE, DOSE
Kanal
35
MTF
™
MTF is a measure of signal transfer over a range of
frequencies and quantifies spatial resolution
Type 1 – CsI (TFT)
Type 2 – CsI (CCD)
Type 3 – CR
Type 4 – a-Se
Yaffe - Radiology 2005:234,353
Kanal
Kanal
http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf
36
18
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
MTF
™
Bloomquist et al - DMIST trial
Bloomquist – Medical Physics 2006:33 (3), 719
Kanal
37
DQE
™
Kanal
Kanal
Detective Quantum Efficiency (DQE) measures SNR
transfer through the system as a function of spatial
frequency and is a good measure of dose efficiency
http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf 38
19
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
FFDM – Radiation Dose
™
Kanal
Bloomquist et al - DMIST trial
Bloomquist – Medical Physics 2006:33 (3), 719
39
Storage and Economics
Kanal
Kanal
40
20
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Storage of Digital Images
Kanal
Radiographics 2004:24,1755
41
Display of Digital Images
Kanal
Kanal
Radiographics 2004:24,1757
42
21
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Advantages and Disadvantages
™
Advantages
™ Optimize postpost-processing of images
™ Permit computercomputer-aided detection to improve the
detection of lesions
™ Storage of images easier
™
Disadvantages
™ Image display and system cost
™ Limiting spatial resolution is inferior to film, 55-13
lp/mm vs. 20 lp/mm
Kanal
43
Economics of FFDM
™
™
™
Reimbursement rates approved by Medicare are
decreasing every year
™
Plain Film Screening
Digital
Screening
Diagnostic
CAD
™
™
Kanal
Kanal
SFM systems cost well under $100,000
FFDM systems cost in the range of $300,000 - $450,000
2008 - $83.03, 2009 - $81.51
2008 - $133.69, 2009 - $129.84
2008 - $151.21, 2009 - $152.56
2008 - $14.86, 2009 - $12.26
44
22
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Benefits
Kanal
45
Expected Benefits of FFDM
™
Kanal
Kanal
The costs of FFDM systems should be compared along
with the inherent benefits of the digital technology prior
to the purchase:
™ Reduced recall rates
™ Increased patient throughput
™ Increased early detection of breast cancer
™ Decreased falsefalse-negative biopsy results
™ Decreasing film and processing costs
™ Increasing the caseload of each mammography room
46
23
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Clinical Trials and Phantom Studies
™
™
Large screening study screened 49,500 women
Digital Mammographic Imaging Screening Trial (DMIST),
funded by NCI and conducted by ACRIN
(http://www.acrin.org/6652_protocol.html)
Kanal
47
Future Developments
Kanal
Kanal
48
24
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Future Developments
™
™
™
™
™
Dual energy
CAD integration
Telemammography
Tomosynthesis
Contrast enhanced digital mammography
Kanal
49
Future Developments
Tomosynthesis
Kanal
Kanal
Radiology 2005:234:353:362
50
25
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
Future Developments
Contrast Enhanced Digital Mammography
Radiology 2005:234:353:362
Kanal
51
TAKE HOME POINTS
™
™
™
™
™
™
Kanal
Kanal
Different technologies exist for digital systems – indirect and
direct
Commercially available FFDM systems vary in technology
Many advantages exist for FFDM in comparison to FSM
MTF and DQE are typically used to characterize FFDM
systems
Dose is lower with FFDM compared to FSM
Advanced applications of FFDM include tomosynthesis,
contrast enhanced mammography etc
52
26
AAPM 2009 Breast Imaging Course Digital Mammography Update
July 27, 2009
THANK YOU
Kanal
Kanal
53