Acknowledgements RapidArc: Clinical Implementation Fang-Fang Yin, PhD • Team efforts from staff at Duke Radiation Oncology, especially to Dr. J Chang, Dr. J O’Daniel for providing slide information • Technical and financial supports from Varian Medical Systems Q. Jackie Wu, PhD Duke University Medical Center Clinical Implementation of VMAT • Fundamentals for RapidArc • Infrastructure/Installation • Acceptance testing/commissioning • Planning • Delivery • Quality assurance Fundamentals for VMAT • Intensity Modulated arc therapy (VMAT) – An arc-based approach to IMRT – To be delivered on a conventional linear accelerator with a conventional MLC – During an arc, the leaves of the MLC move and dose rate changes continuously as the gantry rotates • RapidArc is one format of VMAT 1 The Principle of IMRT: Dose Painting Expected 3-field IMRT Conventional 3-field RT Static and Rotational IMRT VMAT Static gantry IMRT Typical dose distribution OAR PTV OAR Beam Profile PTV Multiple apertures At each angle The Format of Cone-Beam IMRT Static Gantry IMRT Rotational IMRT One aperture At each angle Volumetric Rotational IMRT Options • Existing Planning Systems • • • • Eclipse (Varian) – Duke choice ERGO++/Monaco (Elekta) Pinnacle SmartArc (Philips) Prowess (Prowess) • Existing Delivery Systems • RapidArc (Varian) – Duke choice • VMAT (Elekta) • Cone-beam Therapy (Siemens) (WIP) • Existing QA Systems Field1 @180o One of 7 fields One arc from 179o 181o Count-clockwise • • • • • Matrixx – Duke choice for routine QA Film – Duke choice for commission SunNuclear Delta 4 – Duke choice for future QA device Optical Scanner …… 2 Where Are We (Duke)? • Started investigation in June 2008 – A research RapidArc planning station from Varian • Clinical installation in October – Acceptance testing, commissioning, QA programs – Single arc, no couch rotation, partial arc • First patient treatment – December 2008 • New versions in August 2009 and May 2011 Clinical Implementation of RapidArc • Fundamentals for RapidArc • Infrastructure/Installation • Acceptance testing/commissioning • Planning • Delivery • Quality assurance – Allow multiple arcs, couch rotation, partial blocking, etc. Infrastructure/Installation • Staff (dedicated and trained) • Existing machines: Clinical Implementation of RapidArc • Fundamentals for RapidArc – 21EX machine with 120-leaf millennium MLC • Infrastructure/Installation – NovalisTx with 120-leaf HD MLC (SRS, SRT,SBRT) • Acceptance testing/commissioning • ARIA version 8.6 or above (v10 now) • Planning • Eclipse planning station (hardware and software) • Delivery • QA equipment • Quality assurance 3 Acceptance Testing • Machine readiness • Verification of installation against items included in the purchase order • Inspections of safety and quality of installation and components • VMAT performance • Testing of functionality of each component and system performance against specifications. Acceptance Testing Sample • Test 1.1: Gantry Angle Calibration – Tolerance: + 0.5° • Test 1.2: Isocenter Calibration – Tolerance: + 1 mm • Test 1.3: General Arc Dosimetry – Range: 0.2 MU/° to 5.0 MU/° – Tolerance: + 1% • End-to-end testing • Dry-runs for a few test case from simulation to delivery Acceptance Testing Sample Acceptance Testing Sample • Test 1.4 • dMLC Dosimetry • 0.5cm MLC slit sliding over 4 cm range • Gantry: 0°, 90°, 270°, 180° • Test 2.1: • Accuracy of dMLC position vs. gantry position • Tolerance: + 1 mm • Tolerance: + 2% (over mean value) 4 Acceptance Testing Sample Acceptance Testing Sample • Test 2.2: • Accuracy of dMLC position during arc • Tolerance: + 1 mm 2.1: Picket Fence vs. Gantry Angle (static) Acceptance Testing Sample Acceptance Testing Sample 2.1: Picket Fence vs. Gantry Angle (static) 5 Acceptance Testing Sample Acceptance Testing Sample • Test 2.3: • Ability to accurately detect MLC position error • Criteria: detect submillimeter error in position 2.3: Picket Fence with Errors Acceptance Testing Sample • Test 2.4: • Accuracy of dose rate and gantry speed control during RapidArc • Tolerance: + 2% Acceptance Testing Sample • Test 2.5: • Ability to control leaf speed/position during RapidArc • Tolerance: + 2% 6 Acceptance Testing Sample 1.20 1.0 1.18 0.8 Percent Deviation (%) Relative Dose Acceptance Testing Sample 1.16 1.14 1.12 Off y-axis : -100 mm Off y-axis : 0 mm Off y-axis : 100 mm 1.10 1.08 Gantry speed vs. Dose rate (Tolerance 2%) 1.06 1.04 1.02 -150 -100 -50 0 50 0.4 0.2 0.0 -0.2 -0.4 Variation of gantry speed and dose rate (tolerance 2%) -0.6 -0.8 1.00 -200 Off y-axis: -100 mm Off y-axis: 0 mm Off y-axis: 100 mm 0.6 100 150 200 Off x-axis Position (mm) Commissioning • Validate that VMAT is capable of delivering radiation beams as good as SG-IMRT could • Define the limitations of planning optimization, gantry rotation, beam blocking, couch rotation, and leaf speed, collimator settings • Develop treatment process and documentation -1.0 -60 -40 -20 0 20 40 60 Off X-axis Position (mm) Workflow for RapidArc Treatment Case selection Immobilization/simulation Planning/prescription Quality assurance Target localization Treatment/validation 7 Clinical Implementation of RapidArc Planning - Process • Fundamentals for RapidArc • Infrastructure/Installation • Acceptance testing/commissioning • Planning • Delivery • Quality assurance SG-IMRT Planning - Preparation • Site-specific, for each site – 10 cases from previous IMRT RapidArc Planning - Strategy • Selected site-specific planning strategy – Develop RapidArc plan with different options – 1 arc: prostate only, prostate bed, prostate+SV, brain lesions – Comparison between IMRT vs. RapidArc – 2 arcs: prostate+SV+LN, spinal, brain lesions • Constraint/optimization – Optimization algorithm – Constraints sensitivity – Multiple arcs: Head and neck, brain lesions, analrectal – Partial arcs: partial breast, liver, … 8 Planning - Optimization Planning – Quality and Efficiency • Key challenge – interconnectivity of the beam shapes within consecutive VMAT gantry positions • Constraints – Target and normal structures – Mechanicals • Optimization algorithms • MLC segments – Number of segments: IMRT RapidArc – Plan quality is comparable if same segments – More segments: slow gantry rotation for RapidArc – More segments: long treatment time • Planner experience – More parameters – Understand algorithm – Aperture based objectives and algorithms – Understand limitations Planning - Logistics • Training – multiple people involved in the planning • Gaining experience Clinical Implementation of RapidArc • Fundamentals for RapidArc • Infrastructure/Installation • Acceptance testing/commissioning • Develop planning protocols • Planning • Last ~ 3 months before patients start • Delivery • Quality assurance 9 Delivery – Sample Parameters Delivery - Tolerances • MLC leaf motion • Leaf motion limit: 2.5 cm/s, 5 mm/degree • Dose rate and gantry rotation speed • One arc = one field • MU, doserate, gantry speed linked • Larger MU max doserate varying gantry speed • Smaller MU max gantry speed varying doserate • Middle varying gantry speed and doserate RapidArc Delivery Limits • Variable gantry speed – 0.5 – 5.6 degree/sec • Variable dose rate – 0 – 600 MU/min (0 -1000 Novalis Tx) • Variable dose per degree – 0.2 – 20 MU/degree • Variable MLC speed – 0 –2.5 cm/s Delivery - Efficiency • Plan quality • The quality of plan number of apertures • Single arc – 177 apertures • Complexity structures: • more apertures/arcs • multiple arcs • Delivery time • Single arc – less time (< 2 mins) but sometimes inferior quality • Multiple arcs – better quality but longer time 10 Delivery – Treatment Time Delivery – Partial Arc • Mechanical collision • Treatment time = Patient set-up time + Delivery time • If isocenter is close to center • Delivery time=beam-on time+between beam=on time • Patient setup time: no reduction • full arc • Beam-on time: 20-60% reduction (less MUs) • If isocenter is close to peripheral • Between beam-on time: • 100% saving for single arc • partial arc • 20-60% saving for multiple arcs Partial RapidArc for Liver SBRT • Partial blocking is also available Multiple Arcs vs. Single Arc Single arc RapidArc MU = 714 IMRT Total MU = 1572 Multiple arcs Time: 1.3 min 11 VMAT For Large Size PTV • Large field size -> sometimes IMRT beam has triple beam splits VMAT For Large PTV 3 Arcs, 1000 degree rotation 17cm Field Size << PTV in some directions • Multi-sections (parts) of the PTV, large variation of PTV shape, OARs and their constraints • Beam orientation selection is part of IMRT planning, is often not used for RapidArc (i.e. full arc) VMAT For Large Size PTV IMRT VMAT17cm VMAT For Large PTV IMRT VMAT17cm 12 VMAT For Large PTV IMRT VMAT26CM VMAT17cm VMAT For Large PTV VMAT 17cm Field Size Effect On VMAT Planning Quality PTV Hot Spot (D1%) VMAT23cm 160 Collimator 30 150 Collimator 45 140 D1% (%) VMAT17cm VMAT For Large PTV 130 120 110 100 0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 Projected Field Size X (cm) 13 VMAT For Head-and-Neck VMAT For Head-and-Neck VMAT IMRT VMAT Field Thru Shoulder Field Thru Shoulder VMAT For Head-and-Neck VMAT RapidArc For Head-and-Neck IMRT Rt Parotid Lt Parotid Pharynx Oral Cavity Cord 14 VMAT For Head-and-Neck Clinical Implementation of RapidArc • Fundamentals for RapidArc • Infrastructure/Installation • Acceptance testing/commissioning Segments Thru Shoulder • Planning • Delivery • Quality assurance Quality Assurance • The QA program for the VMAT is similar to SG-IMRT in principle but with different measurement approaches due to its dynamic nature that, during Quality Assurance • The QA program: validate the functionality and performance of accepted features VMAT delivery, – MLC leaves are moving • For each planned delivery – Gantry is rotating – Patient specific QA – Dose rate is changing – Machine specific QA 15 Quality Assurance • Machine specific QA – accuracy of the MLC leaf positions during VMAT delivery – ability of the system to accurately vary the dose rate and gantry speed during VMAT delivery – ability of the system to accurately vary the MLC leaf speed during VMAT delivery • Tolerances: Acceptance baselines Patient Specific QA • Hybrid QA technique – Plan to phantom – Dose measurement to phantom • Rotational nature – Not to single plan • Phantoms • Instruments – Ion chamber – 2-D array (ion chamber, diodes, film, …) Machine QA Chart • Daily – Standard linac QA – Standard MLC QA – Rotational delivery of dose to an ion chamber phantom • Monthly – Leaf motion – Gantry rotation – Dose output • leaf motion • gantry motion • dose rate changes Patient Specific QA • Data analysis – Multiple planes (axial, coronal, sagittal) – Profiles – Points – Gamma analysis • Collision check – Before patient on the couch – When patient on the couch 16 RapidArc QA vs. IMRT QA • More complex treatment delivery – Varying gantry angle, gantry speed, dose rate, and MLC leaf motion • ImSure does not calculate dose for RapidArc delivery • Current IMRT technique: Portal dosimetry • RapidArc technique: Ion chamber, film, and Matrixx QA Measurements • Ion chamber – Absolute dose: (meas – calc) / calc < 3% • Film (coronal, sagittal, and axial planes) – Optional • Matrixx (coronal and sagittal planes) • 3-6 hrs/patient “Gold Standard” QA Tools Ion Chamber • Equipment – 0.13cc or 0.01cc ion chamber Ion Chamber Film • Calibration – 10x10cm2, 100SSD, depth = dmax, 200MU – cGy/nC correction factor • RA delivery – Center of 30cm x 30cm x 20cm solid water phantom – Compare to Eclipse calculation • 39 VMAT plans 17 Verification with “Gold Standard” QA Film • Equipment – Kodak EDR2 film – OmniPro ImRT • Calibration – 12 2cm x 2cm squares – 0 – 300cGy • RA delivery – MultiCube (IBA dosimetry) – Coronal, sagittal, and axial planes – Compare to Eclipse calculation • Ion chamber: – Median: +1.7% – Range: -0.9% - 2.8% • Film – 24/24 > 93% passing rate – 23/24 > 95% passing rate – 20/24 > 97% passing rate – Gamma analysis • 8 MAT plans Ion Chamber vs. Eclipse Result: Axial Film vs. Eclipse Ion chamber array vs. Eclipse Eclipse 3%, 3mm DTA, 5% threshold Film 18 Film vs. Eclipse Film vs. Eclipse Film Film vs.vs. Eclipse Eclipse 100 100 95 85 80 75 70 Coronal Sagittal Axial 65 3%, 3mm DTA, 5% threshold % Pixels Passing Gamma (<1) 90 85 80 75 70 Coronal Sagittal Axial 65 3%, 3mm DTA, 5% threshold Sagittal and Coronal, 40% threshold Axial 3%, 3mm, axial thresholds up to 40% 1 2 3 14 4 15 5 23 6 24 7 32 8 Plan # Validation of 2D Ion Chamber Array 60 1 2 3 – 0.07 cm3 sensitive volume – 0.4cm diameter • 24 x 24 cm2 grid • 7.6mm spacing • Automatic temperaturepressure correction 15 5 23 6 24 7 32 8 Plan # Angular Dependency: Future solution from IBA: Gantry angle sensor Apply correction factor to each ion chamber based on angularity +2.2% 15x • MatriXX Evolution (IBA Dosimetry) • 1020 ionization chambers 14 4 +3.0% 6x +2.0% 15x +2.8% 6x Back 60 Front % Pixels Passing Gamma (<1) 95 90 -1.8% 15x -3.5% 6x Measured/Calculated (%) Herzen et al., PMB 2007; 52: 1197-1208 +3.2% 15x -1.7% 15x +3.7% 6x -2.0% 6x 19 Results Patient Specific QA Matrixx QA %Pixels Passing (Gamma <= 1) Matrixx vs Eclipse 100 98 96 94 92 90 MatrixxCoronal MatrixxSagittal 3%, 3mm, axial thresholds up to 5% 88 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 RapidArc (Plan #) Film vs. Ion Chamber Film vs. Ion Chamber ArrayArray 1: IonEclipse Chamber vs. Eclipse Matrixxvs. vs. Eclipse Ion Chamber Fig. vs. & and ICA Eclipse 100 6% IC IC Array %Difference from Eclipse Calculation 2% 0% 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 -2% %Pixels Passing Gamma (<1) 98 4% 96 94 92 90 Coronal Sagittal -4% Brain Prostate Bed Prostate -6% Plan # Prostate + SV Spine Gamma = 3%, 3mm, 5% threshold 88 0 1 1 2 33 4 14 155 6 23 7 24 8 32 9 Plan # 20 Effective vs. Efficient • Stage 1: Intensive QA – Ion chamber, film in 3 planes, ion chamber array in 2 planes, 3D polymer gel dosimetry • Stage 2: Rigorous QA – Ion chamber, ion chamber array in 2 planes • Stage 3: Effective and Efficient QA – Ion chamber, ion chamber array in 1 plane – Ion chamber array only Effective vs. Efficient • When implementing a new technology – Perform intensive patient-specific QA for the first group of patients – Rely on QA “gold standards” – 3D QA very useful – Use “gold standard” QA technology to transition to newer QA devices • Goal: Effective QA Effective vs. Efficient Preparation Delivery Ion chamber – 1st 15 15 Analysis 5 Film – 1st 15 20 20 5 Ion chamber array – 1st 15 15 Ion chamber – additional 15 7 5 Film – additional 15 10 10 Ion chamber array – additional 15 7 5 IC + 3 Film + 2 ICA 90 77 55 IC + 2 ICA 45 37 15 IC + ICA 30 30 10 Conclusion • RapidArc is one format of rotational IMRT for dose painting • Implementation of RapidArc requires careful planning, testing, and verifications. • Thoroughly testing and commissioning are necessary prior to patient treatment • QA is a critical step, always compare with static field IMRT plan in the early phase • RapidArc should be judged by its accuracy, safety, efficiency, applicability, integration , and adaptation 21 Thank you for your attention 22