8/2/2011 Imaging Educational Course - Radiography III Understanding Exposure and Deviation I di Indices David L. Leong1,2 Marie-Louise Butler2 1 2 Analogic Corporation, Peabody, MA, USA UCD School of Medicine and Medical Sciences, University College Dublin, Dublin, Ireland AAPM 2011 WE-A-110-1 Learning Objectives 1. Understand AAPM recommendations, 2. Understand the IEC standard, 3. Understand factors that affect Exposure and Deviation Index values. 4. Understand DICOM support for Exposure and Deviation Index values. 2 1 8/2/2011 State of Confusion • Manufacturer Specific Indices – – – – – – – – Rex S lgM EI DEI EXI EVI f# • A range of units and relationships to exposure. • Calibration conditions ranged from 70 to 80 kVp with different filtration selections 3 AAPM Task Group 116 • TG116 was started in July 25, 2004 • Report published June 2009 • 29 individuals participated in the Task Group • Full report available on AAPM Website – http://www.aapm.org/pubs/reports/ • Executive summary published in Medical Physics – S. J. Shepard et al. "An exposure indicator for digital radiography: di h AAPM Task T kG Group 116 (Executive (E ti Summary)," Med. Phys. 36, 2898-2914 (2009). – http://www.aapm.org/pubs/reports/ 4 2 8/2/2011 IEC Working Group 43/62b • Preparatory work in the German (DIN) standardization organization 2004-2006 • Proposal for standard by the German National Committee June 2006 • WG43 was started in September 2006 • Experts from Germany, Belgium, Canada, China, Japan, Netherlands, United States – 16 members, members 6 people overlapped with TG116 • First meeting October 2006 in Hamburg, Germany • IEC 62494-1 published August 2008 • Stability date: 2012 5 Keys Parts • Exposure Index • Target Exposure Index • Deviation Index • Calibration Condition 6 3 8/2/2011 Exposure Index IEC 62494 AAPM TG116 Name Exposure Index (EI) Indicated Equivalent Air Kerma (Kind) Units Unitless μGy Not specified 3 significant Precision • Different definitions, but substantially the same. • Kind typically less than 1 as it is μGy • EI is designed to be greater than 1 7 Exposure Index Formulas • IEC: VC AL f K C AL K CAL f 1 E I c 0 g V 100μG y -1 f V C A L 1 g V f 1 V V • TG116: K K Q K 1000 log 10 ST D 1000 log 10 ST D 0.001 K0 K IND K 0 10 Q 1000 , where K 0 0.001 Gy 8 4 8/2/2011 Target Exposure Index • Target values: KTGT (TG116) and EIT (IEC) • EIT – expected value of EI when the detector is properly exposed • Manufacturer / Site Specified – Needs to be adjustable by user – Can be established by professional societies • Body part and view specific targets (recommended by TG116) 9 Deviation Index IEC 62494 AAPM TG116 Name Deviation Index (EI) Deviation Index (EI) Units Unitless Unitless Not specified 1 decimal place Precision • TG116 and IEC are the same: EI D I 10 log EIT • More important clinically than EI 10 5 8/2/2011 Action Levels (TG116) Table 2. Exposure Indicator DI Control Limits for Clinical Images DI Range Action > +3 +3.0 0 Excessive patient radiation exposure Repeat only if relevant anatomy is clipped or “burned out” Require immediate management followup. +1 to +3.0 Overexposure: Repeat only if relevant anatomy is clipped pp or “burned out” −0.5 to +0.5 Target range Less than −1.0 Underexposed: Consult radiologist for repeat Less than −3.0 Repeat 11 Calibration IEC Based on RQA5 from IEC 61267 AAPM • kVp: 66 – 74 • kVp: 70 ± 4 • HVL: 6.8 ± 0.3 mm Al • HVL: 6.8 ± 0.25 mm Al – Added filter of either 21 mm aluminum or 0.5 mm copper and 2 mm aluminum – Adjustment of the kVp from 66 to 74 is permitted to achieve the target HVL – Added filter of either 21 mm aluminum or 0.5 mm copper and 0 to 4 mm aluminum*, – Adjustment of the kVp and filter permitted to achieve the target HVL *Type 1100 Document the added filter and X-ray tube voltage used for calibration 12 6 8/2/2011 Calibration - Recommendation • Start as close to 70 kVp as possible – Measure non-invasively without added filtration – If your meter support Practical Peak Voltage (PPV), use it • Use at least 1 meter Source to Detector distance • Adjust added filtration to achieve 6.8 mm HVL – No less than 0.5 mm Cu or 18 mm Al – No more than 0.5 mm Cu + 4 mm Al or 23 mm Al – Raise or lower kV as a last resort • Record the final filter thickness and generator kVp – On repeat calibrations, start with the previous filter thickness 13 Summary IEC 62494 AAPM TG116 Exposure Indicator Name Exposure Index (EI) Indicated Equivalent Air K Kerma (Kind) Exposure Indicator Units Unitless μGy Not specified 3 significant Exposure Indicator Precision Deviation Index formula Deviation Index Units Deviation Index Precision Beam Conditions EI D I 10 log EIT K DI 10*log IND KTGT Unitless Unitless Not specified 1 decimal place 6.8 ± 0.3 mm Al 6.8 ± 0.25 mm Al 14 7 8/2/2011 Factors that affect EI • Calculation of EI based on anatomical region of interest. (ROI) - Relevant Image Region • Method to determine ROI and area of ROI affect EI value • Either define a specific procedure or method 15 ROI and Exposure Variation 16 8 8/2/2011 30 Exposures – No Changes Bone and Mostly Bone Bone Only Soft Tissue Mean 398.1 277.9 231.2 Std 2.164 0.712 0.774 17 Chest Exposures All Image Anatomy Mean 466.00 462.17 Std 2.280 2.483 18 9 8/2/2011 ROI Effects on DI Bone and Mostly Bone Bone Only Soft Tissue 4 mAs -1.0 -1.0 -1.0 6 mAs +1.1 +1.1 +1.1 19 DICOM Changes • CP1024 written to support new Exposure Indices • Became “Final Text” June 21, 2010 • Use is not mandatory 20 10 8/2/2011 DICOM Changes • Legacy Exposure Indices Attribute Name Relative X-Ray Exposure Tag (0018,1405) Sensitivity (0018,6000) Type Attribute Description 3 Relative x-ray exposure on the plate. Meaning of values is implementation specific. May be used to adjust the dynamic range of the plate digitizer (scanner). Note: This value is manufacturerspecific. DICOM specifies standard Attributes in Table 10.x-1 ‘Exposure Index Macro’ which are recommended. 3 Read out sensitivity. Note: This value is manufacturerspecific. DICOM specifies standard Attributes in Table 10.x-1 ‘Exposure Index Macro’ which are recommended. 21 DICOM Changes • New Exposure Indices Attribute Name Exposure Index Tag (0018,1411) Target Exposure Index (0018,1412) Deviation Index (0018,1413) Type Attribute Description 3 Measure of the detector response to radiation in the relevant image region of an image acquired with a digital x-ray imaging system as defined in IEC 62494-1. Notes: 1. A string rather than binary Value Representation is used for this Attribute, in order to allow the sender to control the precision of the value as suggested in the report of AAPM Task Group 116. 2. This index value is scaled as defined by IEC 62494 62494-1. 1 3 The target value used to calculate the Deviation Index (0018,1413) as defined in IEC 62494-1. 3 A scaled representation of the difference of the Exposure Index compared to the Target Exposure Index as defined in IEC 62494-1 and the report of AAPM TG 116. 22 11 8/2/2011 Recommendations • Insist that all new CR and DR equipment support IEC 62494-1 – Both EI and DI • Push your PACS vendor to support and display DI on all workstations (CP1024) • Attempt to upgrade older CR and DR systems to add support for IEC 62494-1 • Develop a set of Target Exposure Values • Setup a training program on the use of EI, DI and action limits 23 Learning Objectives 1. Understand AAPM recommendations, 2. Understand the IEC standard, 3. Understand factors that affect Exposure and Deviation Index values. 4. Understand DICOM support for Exposure and Deviation Index values. 24 12 8/2/2011 Final Comment “It is i essential ti l th thatt all ll aspects t off good d clinical technique be adhered to and an appropriate DI value should not be interpreted as proof of good work.” TG116 Report Section X 25 Thank You UCD School of Medicine & Medical Science Scoil an Leighis agus Eolaíocht An Leighis UCD 13