3/18/2013 Medical Physics Support for Dental X‐ray Imaging : Technology and physics measurements for intraoral systems and accreditation of cone beam CT. Robert J. Pizzutiello, MS, FACR, FAAPM, FACMP Residency Program Director, Upstate Medical Physics, PC Senior Vice President, Imaging Physics LANDAUER Medical Physics Outline • Evolution of Dental imaging in Clinical Practice • Film based dental imaging – Dose, Image Quality and Consistency • Digital dental imaging, planar and CBCT • Dose, Image Quality and Consistency • Medical Physics Measurements for Dental Systems • Accreditation of Dental Imaging Facilities • Summary and Conclusion 1 3/18/2013 Learning Objectives • To understand the factors that affect dose and image quality in film and digital dental radiography • To understand the methods of performance evaluation of dental imaging systems • To understand the emerging importance of Cone Beam CT scanning in dental imaging • To understand the medical physicist’s requirements to support accreditation of dental practices by the Intersocietal Accreditation Commission CT program Film based Intraoral radiography • • Film, no screen, foil backing D-Speed, E-Speed 2 3/18/2013 RADIOGRAPHIC TRENDS OF DENTAL OFFICES AND DENTAL SCHOOLS ORHAN H. SULEIMAN et al JADA 1999 RADIOGRAPHIC TRENDS OF DENTAL OFFICES AND DENTAL SCHOOLS ORHAN H. SULEIMAN et al JADA 1999 3 3/18/2013 Typical Exposure Data, mR Dental Bite Wing for 2009 (N=7205) TYPE Total % of Total Average D 3483 55% 256 E 1001 16% 162 F 774 12% 148 T (digital) 1947 31% 105 SD 99 72 63 59 Highest 2150 1320 520 693 Dental Bite Wing for 2005 (N=6325) TYPE Total % of Total Average D 3084 49% 258 E 1118 18% 162 F 500 8% 140 T (digital) 1623 26% 106 SD 93 70 67 58 Highest 860 786 952 712 Dental Bite Wing for 2001 (N=8600) TYPE Total % of Total Average D 5127 81% 262 E 1674 26% 161 F 510 8% 148 T (digital) 1289 20% 110 SD 106 68 58 64 Highest 3308 675 650 560 Courtesy NYSDOH Courtesy Joel Gray, DIQUAD 4 3/18/2013 Collaborating to Bring a Unique Solution New York State Guidelines 240‐350 120‐170 100‐140 Based on the most recent FDA analysis of film based intraoral radiography, which of the following is a true statement? 20% 1. 2. 3. 4. 5. 20% 20% 2 3 20% 20% Newer, faster films can be processed in half the time previously required F speed film images use significantly lower dose than E-speed Film E and F speed film images use significantly lower dose than D speed film Approximately 90% of film based dental imaging is now performed using D speed film D Speed film images use significantly lower dose than Espeed Film 1 4 5 10 5 3/18/2013 Based on the most recent FDA analysis of film based intraoral radiography, which of the following is a true statement? 1. Newer, faster films can be processed in half the time previously required 2. F speed film images use significantly lower dose than E-speed Film 3. E and F speed film images use significantly lower dose than D speed film 4. Approximately 90% of film based dental imaging is now performed using D speed film 5. D Speed film images use significantly lower dose than E-speed Film Ref: Conference of Radiation Control Program Directors, Inc. (CRCPD), Publication E‐03‐6, "NEXT Tabulation and Graphical Summary of the 1999 Dental Radiography Survey", November 2003 Advantages of Digital Dental Radiography • immediate image production with solid‐state devices such as a charge‐coupled device (CCD) and a complementary metal‐oxide semiconductor (CMOS); • interactive display on a monitor with the ability to enhance image features and make direct measurements; • integrated storage, providing access to images through practice management software systems; • security of available backup and off‐site archiving; • wide exposure latitude Ref Farman et al. JADA 2008;139(6 supplement):14S‐19S 6 3/18/2013 Advantages of Digital Dental Radiography • perfect image duplicates to accompany referrals to other practitioners; • security mechanisms to identify original images and differentiate them from altered images; • ability to tag information such as a patient identifier, date of exposure and other relevant details; • interoperability of the (DICOM) file format, which enables practitioners with different equipment and software to view and enhance the same images. Ref Farman et al. JADA 2008;139(6 supplement):14S‐19S Typical Exposure Data, mR Dental Bite Wing for 2009 (N=7205) TYPE Total % of Total Average D 3483 55% 256 E 1001 16% 162 F 774 12% 148 T (digital) 1947 31% 105 SD 99 72 63 59 Highest 2150 1320 520 693 Dental Bite Wing for 2005 (N=6325) TYPE Total % of Total Average D 3084 49% 258 E 1118 18% 162 F 500 8% 140 T (digital) 1623 26% 106 SD 93 70 67 58 Highest 860 786 952 712 Dental Bite Wing for 2001 (N=8600) TYPE Total % of Total Average D 5127 81% 262 E 1674 26% 161 F 510 8% 148 T (digital) 1289 20% 110 SD 106 68 58 64 Highest 3308 675 650 560 Courtesy NYSDOH 7 Ref Farman et al. JADA 2008;139(6 supplement):14S‐19S 3/18/2013 Compared with dental film radiography, for digital intraoral radiographs: a) Lower dose is always delivered b) c) d) e) 20% 20% 20% b) c) 20% to the patient than with film techniques Image processing of digital intraoral radiographs is comparable to processing time needed for table top automatic film processors Digital dental systems incorporate new technology to alert the operator when preset dose levels have been exceeded Digital dental imaging systems have superior spatial resolution Operators of digital intraoral radiographs can produce images over a wide range of doses a) d) 20% 10 e) 8 3/18/2013 Compared with dental film radiography, for digital intraoral radiographs: a) Lower dose is always delivered to the patient than with film techniques b) Image processing of digital intraoral radiographs is comparable to processing time needed for table top automatic film processors c) Digital dental systems incorporate new technology to alert the operator when preset dose levels have been exceeded d) Digital dental imaging systems have superior spatial resolution e) Operators of digital intraoral radiographs can produce images over a wide range of doses Ref: Farman and Farman, Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005;99:485‐9 Toshiba Aquilion 64 MDCT and NewTom vGi CBCT Scanner (Dental) A Comparison of Maxillofacial CBCT and Medical CT. Christos Angelopoulos 9 3/18/2013 Basics of CBCT • CBCT imaging is performed using a rotating platform to which an x‐ray source and detector are fixed. • A divergent pyramidal‐ or cone‐shaped x‐ray source is directed through the middle of the RO • Transmitted, attenuated radiation is projected onto an area x‐ ray detector on the opposite side • The x‐ray source and detector rotate around a fulcrum, the center of the final volume imaged • During the rotation, multiple sequential planar projection images covered by the detector or the field of view (FOV) are acquired in an arc of 180o or greater • These single projection transmission images constitute the raw primary data and are individually referred to as basis, frame, or raw images Basics of CBCT • Basis images appear similar to cephalometric radiographic images • There are usually several hundred projection‐basis images that are reconstructed an image volume • The complete series of images is referred to as the projection data or volumetric dataset 10 3/18/2013 Advantages of CBCT • Irradiated field of view (FOV) reduced to region of interest • Mechanical lead collimation or electronic masking • Only collimation reduces volume of tissue irradiated Collaborating to Bring a Unique Solution MDCT vs. CBCT X-ray source 80-140 kVp 20-100 kW 80 – 120 kVp Stationary Anode Pulsed beam Focal Spot 0.5 – 1.2 mm 0.5 – 1.2 mm Detector MD arrays 64 – 256+ rows Flat panel Tl:CsI, Tr:GdOS Spatial Res 0.5 – 0.625 mm Contrast Res ~1 HU 0.4 mm (20 cm FOV) ~10 HU A Comparison of Maxillofacial CBCT and Medical CT. Angelopoulos et al 11 3/18/2013 DOSIMETRY OOOOE (Palomo, et al. & Ludlow, et al.) Principles of CBCT – Dosimetry Conclusions: The Kodak 9000 3D provides doses that are substantially lower than previously reported doses produced by medium and large FOV CBCT units. The digital panoramic mode provides a low dose alternative for panoramic examinations of the jaws using the same unit. Level s 2 3 4 5 6 7 9 Ludlow JB: Dosimetry of Kodak 9000 3D Small FOV CBCT and Panoramic Unit, UNC School of Dentistry, Chapel Hill, NC, 2008. 12 3/18/2013 Ludlow JB: Patient risk related to common dental radiographic examinations JADA 2008 Ludlow JB: Patient risk related to common dental radiographic examinations JADA 2008 13 3/18/2013 Ludlow JB: Patient risk related to common dental radiographic examinations JADA 2008 Principles of CBCT – Dosimetry Ludlow JB: Dosimetry of Kodak 9000 3D Small FOV CBCT and Panoramic Unit, UNC School of Dentistry, Chapel Hill, NC (2008) 14 3/18/2013 Adapt CT measurement techniques • Follow MDCT Methods + • Dosimetry – Head only • Challenges – Image Quality phantom size – Manufacturer provided, CATPhan, old RMI – AAPM Head dosimetry OK – Use of detector entrance exposure? – Positioning phantoms – not trivial! 15 3/18/2013 Dosimetry and Image Quality Evaluation of a Dedicated Cone‐beam CT System for Sinus and Temporal‐Bone Applications Lifeng Yu, PhD, Thomas J. Vrieze, Michael R. Bruesewitz, James M. Kofler, PhD, Cynthia H. McCollough, PhD 16 3/18/2013 Reports 17 3/18/2013 18 3/18/2013 For dental CBCT systems, measurement of dose and image quality 20% 20% 20% b) c) 20% 20% a) Can follow the methods b) c) d) e) described in the ACR CT accreditation program QC Manual (2012) Can only be performed using TLD’s in a Rando head phantom Require adaptation of standard CT testing methods to the unique challenges of CBCT systems Are not necessary because all new systems are self-calibrating Can be performed by the service engineer a) d) 10 e) 19 3/18/2013 For dental CBCT systems, measurement of dose and image quality a) Can follow the methods described in the ACR CT accreditation program QC Manual (2012) b) Can only be performed using TLD’s in a Rando head phantom c) Require adaptation of standard CT testing methods to the unique challenges of CBCT systems d) Are not necessary because all new systems are selfcalibrating e) Can be performed by the service engineer Ref. Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit, Ludlow, Daview‐Ludlow and Brooks. Dentopmaxillofascial Radiology (2003) 32 229‐234 20 3/18/2013 Correct for Geometry to Image Receptor Distances, from CJ, IMTEC FS to detector FS to tube cover (for CTDP) FS to axis of rotation Patient (axis of rotation) to detector FS to detector active depth Full Field Reduced Field Horizontal inches 31 mm 787.4 V: Approx CTDP 40.5mm 6.347 24 161.2 609.6 H: V2: 51mm 29mm 7 177.8 31.607 802.8 CTDP at tube half Exposed length of pencil chamber (assuming point cover (mm) length source) 40.5 20.25 76.6 29 51 14.5 25.5 54.8 Regarding assessment of beam collimation in CBCT systems: a) Scanning of GafChromic media images may be used, corrected for geometry b) CR or film based imaging systems must be used to detect the x-ray beam c) This is not necessary because CBCT systems collimate automatically d) Is not possible to be performed in the field by a medical physicist 21 3/18/2013 Regarding assessment of beam collimation in CBCT systems: a) Scanning of GafChromic 25% 25% 25% b) c) 25% media images may be used, corrected for geometry b) CR or film based imaging systems must be used to detect the x-ray beam c) This is not necessary because CBCT systems collimate automatically d) Is not possible to be performed in the field by a medical physicist a) d) 10 Regarding assessment of beam collimation in CBCT systems? a) Scanning of GafChromic media images may be used, corrected for geometry b) CR or film based imaging systems must be used to detect the x-ray beam c) This is not necessary because CBCT systems collimate automatically d) Is not possible to be performed in the field by a medical physicist Ref: Application of Gafchromic film in the study of dosimetry methods in CT phantoms. Martin CJ, Gentle DJ, Sookpeng S, Loveland J., J Radiol Prot. 2011 Dec;31(4):389‐409 22 3/18/2013 IAC CT Accreditation for Dental CT Mary Lally, MS, RTR (MR) IAC Deputy Chief Executive Officer Improving health care through accreditation Improving health care through accreditation Who is the IAC? • Accreditation is IAC’s only business for 22 years • Over 10,000 sites accredited • One non profit 501(c)(6) organization with seven divisions – Five diagnostic imaging modalities (vascular testing, echo, nuclear/PET, MR and CT) – Two therapeutic (carotid stent; vein center) • Overall organization directed by IAC board – Composed of 2 representatives from each division – Responsible for strategic direction and financial oversight • Primary division responsibility – review and revise modality specific standards Improving health care through accreditation 23 3/18/2013 Improving health care through accreditation IAC Sponsoring Organizations American Academy of Dermatology (AAD) American Academy of Neurology (AAN) American Academy of Orthopaedic Surgeons (AAOS) American Academy of Otolaryngology — Head and Neck Surgery (AAO-HNSF) American Association of Physicists in Medicine (AAPM) American Academy Of Oral and Maxillofacial Radiology (AAOMR) American Association of Oral and Maxillofacial Surgeons (AAOMS) American College of Cardiology (ACC) American College of Phlebology (ACP) American College of Surgeons (ACS) American College of Nuclear Medicine (ACNM) American Institute of Ultrasound in Medicine (AIUM) American Society of Echocardiography (ASE) American Society of Neuroimaging (ASN) American Society of Neuroradiology (ASNR) American Society of Nuclear Cardiology (ASNC) American Society of Radiologic Technologists (ASRT) American Venous Forum (AVF) Improving health care through accreditation International Society for Musculoskeletal Imaging in Rheumatology (ISEMIR) American Association of Neurological Surgeons Neurocritical Care Society (NCS) Society for Cardiovascular Angiography and Interventions (SCAI) Society for Cardiovascular Magnetic Resonance (SCMR) Society for Clinical Vascular Surgery (SCVS) Society for Vascular Medicine (SVM) Society for Vascular Surgery (SVS) Society for Vascular Ultrasound (SVU) Society of Cardiovascular Computed Tomography (SCCT) Society of Diagnostic Medical Sonography (SDMS) Society of Interventional Radiology (SIR) Society of NeuroInterventional Surgery (SNIS) Society of Nuclear Medicine and Molecular Imaging (SNMMI) Society of Nuclear Medicine and Molecular Imaging Technologist Section (SNMMITS) Society of Pediatric Echocardiography (SOPE) Society of Radiologists in Ultrasound (SRU) Society of Vascular and Interventional Neurology (SVIN) World Molecular Imaging Society (WMIS) 24 3/18/2013 Improving health care through accreditation Improving health care through accreditation 25 3/18/2013 Improving health care through accreditation http://www.intersocietal.org/ct/standards/IACCTStandards2012.pdf Improving health care through accreditation 26 3/18/2013 Dental CT Standards • Dental staff training and experience requirements • Patient safety and confidentiality policies • Equipment QC • Radiation oversight and safety adherence • Report content • Quality assurance program • Image quality review Improving health care through accreditation Medical Physics Responsibilities • Reports follow Guidance Documents for – Radiation Protection Survey (one time) – Image Quality and Dose reports • Phantom Images from Annual MP Survey • Participate in Quarterly QC Committee Meetings • 3 hour CT Radiation Safety Training (quiz=100%) for operators who are not RT’s • Standards definition of medical physicist – “…board certified or in states where licensed, registered, or otherwise state approved” Improving health care through accreditation 27 3/18/2013 Improving health care through accreditation Improving health care through accreditation 28 3/18/2013 Improving health care through accreditation Dental CT Standards • Dental staff training and experience requirements • Patient safety and confidentiality policies • Equipment QC • Radiation oversight and safety adherence • Report content • Quality assurance program • Image quality review Improving health care through accreditation 29 3/18/2013 For IAC CT accreditation, facilities must submit: a) Dosimetry provided by the manufacturer 20% 20% 20% b) c) 20% 20% b) TLD’s mounted to the surface of the ACR accreditation phantom c) Images of the ACR CT accreditation phantom, reconstructed at the minimum slice thickness d) Reports and image quality assessments performed by a medical physicist or other individual licensed, registered, or otherwise state approved to perform these measurements in CT facilities e) The standard ACR dosimetry report for an adult abdomen a) d) 10 e) For IAC CT accreditation, facilities must submit: a) Dosimetry provided by the manufacturer b) TLD’s mounted to the surface of the ACR accreditation phantom c) Images of the ACR CT accreditation phantom, reconstructed at the minimum slice thickness d) Reports and image quality assessments performed by a medical physicist or other individual licensed, registered, or otherwise state approved to perform these measurements in CT facilities e) The standard ACR dosimetry report for an adult abdomen Ref: IAC CT Standards. http://www.intersocietal.org/ct/reaccreditation/ct_standards.htm 30 3/18/2013 Growth of Dental CBCT • Initially used by specialists for implants • Increasingly found in larger general dentistry offices • Current estimates about 10,000 Dental CBCT • Growing rapidly • American Academy of Oral and Maxillofacial Radiologists • American Dental Association – Council on Dental Benefits – Council od Dental Specialties • Support for accreditation Summary • Evolution of Dental Imaging in Clinical Practice • Film based dental imaging – Dose, Image Quality and Consistency • Digital dental imaging, planar and CBCT • Dose, Image Quality and Consistency • Medical Physics Measurements for Dental Systems • Accreditation of Dental Imaging Facilities 31 3/18/2013 Conclusion • The medical physicist can make a significant contribution to dental imaging facilities through careful measurements , education and consultation 32