Informatics driven Quality Initiatives Using the Tools of Six Sigma William Pavlicek PhD AAPM, Charlotte, NC Aug 2, 2012 2:50PM ©2011 MFMER | slide-1 Outline of Talk • Six Sigma Concepts • PET Patient Dose reduction • CT Patient Dose reduction – positioning and Z axis coverage • Fluoroscopy – Lowering Patient Skin Dose • Questions ©2011 MFMER | slide-2 W. Edwards Deming • Taught engineering, physics in the 1920s, finished PhD in 1928 • Met Walter Shewhart at Western Electric • Long career in government statistics, USDA, Bureau of the Census W. Edwards Deming, 1900 – 1993 3 ©2011 MFMER | slide-3 1 W. Edwards Deming • Deming was asked by Japan to lecture on statistical quality control to management • Japanese adopted many aspects of Deming’s management philosophy • Deming stressed “continual never-ending improvement” • “If I had to reduce my message for management to just a few words, I’d say it all had to do with reducing variation”. (Deming, as quoted in The Deming Dimension, Henry R. Neave (SPC Press, 1990), p. 57) • Deming lectured widely in North America during the 1980s 4 ©2011 MFMER | slide-4 Deming energized “Statistical Thinking” • Integral aspect of Six-Sigma (3.4 per million) • Six Sigma is a ‘goal’ not often achieved! • Fundamentally different from Statistical Methods • The definition is broadened to include: • Operations Research tools • Discrete event “Simulation’ • Other elements of ‘lean process’ *ASQ Statistics Division, http://www.asqstatdiv.org/stats-everywhere.htm 5 ©2011 MFMER | slide-5 Defects Per Million Opportunities ©2011 MFMER | slide-6 2 How Variation Hurts … * From Montgomery, D. C. (2009), Introduction to Statistical Quality Control 6th edition, Wiley, New York 7 ©2011 MFMER | slide-7 The Motorola Six-Sigma Concept • The Motorola Six-Sigma concept is to reduce the variability in the process so that the specification limits are six standard deviations from the mean, or only about 2 parts per billion defective (Fig 111a). • When the Six-Sigma concept was initially developed, an assumption was made that when the process reached the Six-Sigma quality level, the process mean was still subject to disturbances that could cause it to shift by as much as 1.5 standard deviations off target (Fig 1-11b). • There is an apparent inconsistency in this. Process performance isn’t predictable unless the process behavior is stable. • No process or system is ever truly stable, and even in the best of situations, disturbances occur. The concept of a Six-Sigma process is one way to model this behavior. Like all models, it’s probably not exactly right, but it has proven to be a useful way to think about process performance, and has direct application to manufacturing.* * From Montgomery, D. C. (2009), Introduction to Statistical Quality8Control 6th edition, Wiley, New York. Figure 1-12. ©2011 MFMER | slide-8 The Generations of Six Sigma • Generation I • Focus on defect elimination • Motorola, 1987-1993 • Generation II • Focus on cost reduction • GE, Allied Signal/Honeywell, 1994-1999 • Generation III • Focus on value creation • DuPont, 2000-present 9 ©2011 MFMER | slide-9 3 Process Variation • Variation is present in all processes and every aspect of the workplace • Excess variation reduces process performance, decreases customer satisfaction, and has a negative impact on the bottom line • Applies to all aspects of a business • Statistical methods, an integral component of six sigma, are useful in shifting the process target to the desired level and reducing the variation around this target 10 ©2011 MFMER | slide-10 As has been said, “the mean never happens,” — a heroic 4-day delivery time on one order, with an awful 20-day delay on another, and no real consistency. The customers in this chart feel nothing. Their life experience hasn’t changed; one bit. These customers hear the sounds of celebration coming from within GE and ask, “What’s the big event; what did we miss?” The customer only feels the variance that we have not yet removed. … Variation is evil in any customer-touching process. * *Jack Welch, General Electric Company 1998 Annual Report 11 ©2011 MFMER | slide-11 Patients feel the variation! • Too long Waiting for their exam • Lack of proper prep means waiting, poss rescheduled exam • Repeated exam due to xyz • Multiple x-ray exams due to lack of available priors • Poss need of follow up exam due to confidence – due to IQ? ©2011 MFMER | slide-12 4 Why “Quality Improvement” is Important: • Patient goes for Out Patient Radiology exam: order, transport, waiting room, exam prep, tech/device performance, x-ray room availability, image quality, interpretation, communication to clinician. • The patient experience has 10 components - is 99% okay? ©2011 MFMER | slide-13 Propagation of Error -repeatability of the exam -reproduced with diff techs/equipment/rads ©2011 MFMER | slide-14 Repeatability (i.e. back to back Ca Scores, bone density, kVp 4X, etc) ©2011 MFMER | slide-15 5 Reproducibility: Different techs, devices, patients… Different Quality? ©2011 MFMER | slide-16 CT fluoro biopsy dose, breast compression, fluoroscopy dose by physician, etc. ©2011 MFMER | slide-17 Focus of Six Sigma is on Process Improvement with an Emphasis on Achieving Significant Impact • A process is an organized sequence of activities that produces an output that adds value to the organization • All work is performed in (interconnected) processes • Easy to see in some situations (manufacturing) • Harder in others • Any process can be improved • An organized approach to improvement is necessary • The process focus is essential to applying Six-Sigma methods 18 ©2011 MFMER | slide-18 6 Organized Approach: DMAIC (duy–may–ick) ©2011 MFMER | slide-19 ‘D’efine the Goal – What is important! (you are what you measure!) Project Charter – Admin Directive (like an internal contract with yourself) Problem Statement Goal Statement In/Out of Scope Team & Time Commitments Timeline / Milestone Estimate Financial Benefits Risks, Constraints & Compliance Issues Identified 20 ©2011 MFMER | slide-20 The Primary Six Sigma Tools • Process map • Cause and effect analysis • Measurement systems analysis* • Capability study* • Failure mode and effects analysis • Observational study (regression)* • Designed experiments* • Control charts and out-of-control-action-plans* * Statistical Tools 21 ©2011 MFMER | slide-21 7 Use of Reference Values Measure, Analyze and Improve! ©2011 MFMER | slide-22 General Comments • One doesn’t really achieve Six Sigma • Informatics with control charts are enormously helpful • Shining a light by sharing data is very useful! • Change can be difficult but fun! • Leadership is key: drive out fear! ©2011 MFMER | slide-23 ©2011 MFMER | slide-24 8 Joint Commission SE #47 ©2011 MFMER | slide-25 How to determine if doses are ‘out of range’ (reproducible)? ©2011 MFMER | slide-26 Lots of Software Solutions! ©2011 MFMER | slide-27 9 RIMS System Architecture DICOM DICOM, HL7 Gateway DICOM SR HL7 Exam Information Database Dose Tracking System WAN Knowledge Base Dosimetry Analyzer Reporting & Alerting Module Emails Texts ©2011 MFMER | slide-28 #1. PET-CT Six Sigma driven lower dose ©2011 MFMER | slide-29 PET Scanner ‘Blocks’ BGO - PMT ©2011 MFMER | slide-30 10 PET F-18 γ Coincidence ? γ Register Yes Event ©2011 MFMER | slide-31 Types of Coincident Events True Scatter Randoms Gamma ray (Singles) Line of Response (LOR) Prompts = True + Scatter + Randoms ©2011 MFMER | slide-32 2D PET – SEPTA! All events used to reconstruct a slice are detected within that slice. ©2011 MFMER | slide-33 11 3D Volume Acquisition Events which cross several transaxial slices are collected. These events are used in the reconstruction of each of the slices crossed. ©2011 MFMER | slide-34 NECR kCounts per second 100000 Practical Clinical Range for Oncology 2D 80000 60000 40000 3D 20000 kBq/cc 0 10 10 20 20 30 40 30 50 60 40 70 50 80 60 70 80 mCi ©2011 MFMER | slide-35 Why wouldn’t one lower dose from 15-20 to 10 mCi (or lower) for 3D? • What if patient is late? • What if patient is early? • What if patient has low uptake? • What if patient is non-compliant? Variation of time/dose has greater effect with smaller injected dose! ©2011 MFMER | slide-36 12 Solution: Bulk F-18 Infusion System No patient to patient variability of dose! ©2011 MFMER | slide-37 Problem: Uptake Room Restroom Tech gets patient to scanner ‘on time’? After 45min ©2011 MFMER | slide-38 Solution: Monitor Uptake Restroom After 45min ©2011 MFMER | slide-39 13 Automated Report Summary by Room/Day ©2011 MFMER | slide-40 Problem: Muscle uptake when walking Tool: Spaghetti Diagram ©2011 MFMER | slide-41 Solution: Minimum distance! ©2011 MFMER | slide-42 14 Six Sigma Results • Variation in administered dose removed! • Variation in ‘uptake time’ removed! • Variation in transport uptake removed! • These are ‘monitored’! • Dose is LOWERED! • Patients still provide unpredictable variation ©2011 MFMER | slide-43 Variability still present! PET Muscle uptake (variability of counts in lesions) ©2011 MFMER | slide-44 #2. CT Dose Index Reduction Define Measure Analyze Improve Control ©2011 MFMER | slide-45 15 Goal: AAPM Notification Values ©2011 MFMER | slide-46 ©2011 MFMER | slide-47 DMAI “C”: Control Chart ©2011 MFMER | slide-48 16 Variation: Non-Size-Adjusted DLP vs #CT Exams for MCA Scanners January 2011 (0018,1030) protocol name: 6300_AbdPel_uro_w_fur (0008,103e) series description: 10m AbdPel a300 Standardized: Mean - 2SD 288 mGy-cm Mean 705 mGy-cm Mean + 2SD 1122 mGy-cm N um be r o f C T Ex am s 10 8 8 6 5 4 4 5 3 3 3 1 2 1 1 00 0 50 0-54 9 55 0-59 9 0 ??????? 2 2 2 5 4 4 4 3 2 1 22 2 1 1 11 0 1 1 0 0 0 0 0 0 0 0 0 0 1 1 0 0 0 0 0 ≥ 1300 12 50-1 299 12 00-1 249 11 50-1 199 11 00-1 149 10 50-1 099 10 00-1 049 95 0-99 9 90 0-94 9 85 0-89 9 80 0-84 9 75 0-79 9 70 0-74 9 65 0-69 9 60 0-64 9 45 0-49 9 40 0-44 9 35 0-39 9 30 0-34 9 25 0-29 9 DLP w/o patient size adjustment (mGy-cm) ©2011 MFMER | slide-49 Analyze - Why Variation of DLP? • Really Large Patient? • Really Tall Patient? • Elevated CTDIvol prescription? • Other cause(s)? ©2011 MFMER | slide-50 ©2011 MFMER | slide-51 17 Same patient 30 days apart Can you see the mAs/centering? ©2011 MFMER | slide-52 Not Centered Centered 23 mSv 13 mSv 40 cm Projected SCOUT 36 cm Projected SCOUT ©2011 MFMER | slide-53 Variation in DLP seen with variation in centering ©2011 MFMER | slide-54 18 ANALYZE: Variation due to Scout Order (180 mAs) 8.09 mGy 9.45 mGy 9.1 mGy 8.09 mGy ©2011 MFMER | slide-55 ANALYZE new example CT of Abdomen/Pelvis 1/16/2011 120kVp 180 ref. mAs CT of Abdomen/Pelvis 4/15/2011 120kVp 180 ref. mAs 14.9mGy 18.4mGy 23% increase in dose between these scans on the same patient. ©2011 MFMER | slide-56 CT Center FOV Positioning ©2011 MFMER | slide-57 19 % Dose Increase Due to Position %mGy Increase Isocenter:Displaced 0.8% -9.0% 0% -1.0% 2.5% 8.9% 4.5% Position changes mAs lookup table. ©2011 MFMER | slide-58 Same Patient CT of Abdomen/Pelvis 1/16/2011 120kVp CT of Abdomen/Pelvis 4/15/2011 120kVp 14.9mGy 18.4mGy 23% increase on the same patient. 180in ref.dose mAs between these scans 180 ref. mAs Dose Affecting Parameters: 1. Sliderboard 2. X-Y Position 3. Body Width 4. Z-Axis Attenuation 5. Metal Monitor ©2011 MFMER | slide-59 Pareto Graph: Rank Change in CTDIvol 60% 50% 40% 30% 20% 10% 0% Board Attenuation Patient Too Low Patient Wider Z Axis Coverage Different ©2011 MFMER | slide-60 20 Variation: Z axis: Abdomen & Pelvis ©2011 MFMER | slide-61 New ‘standard’ Z axis coverage ©2011 MFMER | slide-62 ©2011 MFMER | slide-63 21 Height/Weight with AP and Lateral Size Use Random Forest Model estimates potentially ‘excessive’ Z axis coverage ©2011 MFMER | slide-64 Six Sigma Routine AbdomenPelvis N= 300+ pts Overscan superior 1 cm inferior .4 cm Top 10 Overscans Superior (cm) 6.3 4.5 4.2 4.2 4.2 4.2 3.9 3.9 3.6 3.6 Inferior (cm) 11 7.4 6.5 5.3 4.2 3.8 3.5 3.2 3.1 2.9 ©2011 MFMER | slide-65 CT Colonography OK to exclude lung bases, just include colon Good scan coverage ©2011 MFMER | slide-66 22 CT Colonography N= 20 pts superior inferior Overscan 6.4 cm 2.3 cm Top 10 Overscans Superior (cm) 22.12 13.6 9.1 8.3 8.3 8 7.9 7.5 5.9 5.6 Inferior (cm) 9.5 8 6.5 4.2 4.6 2.6 2.6 2 2 1 ©2011 MFMER | slide-67 Six Sigma Results • Standardize Protocol Names • Standardize CT’s Techniques • Standardize CT Radiograph • Standardize Patient Centering • Standardize Z Axis coverage • Control Charts ©2011 MFMER | slide-68 #3 Six Sigma DMAIC with Fluoroscopic Procedures ©2011 MFMER | slide-69 23 NEW NORMAL! • New arterial and endoscopic procedures save open surgery! • Patients are having multiple episodes of fluorsocopy • 18% of currently scheduled patients have had more than 1 fluoroscopy! • Patients are bigger • Most fluoroscopy is outside of Radiology ©2011 MFMER | slide-70 Technical vs Behaviorial ©2011 MFMER | slide-71 5 ALARA Technical Tasks for reproducibility • Standardize Protocol Names! • Standardize Fluoro/Acquistions! • Default with Low(est) Fluoro pulse rate? • Default with Low(est) DSA/Cine frame rate? • Default with Low(est) dose per pulse? • Always use Copper (0.1mm is 40%!) ©2011 MFMER | slide-72 24 15 cine OldFPS Dose 7.5 FPS ½ cine Current Dose (70-75 bpm, II) (70-75 bpm, flat panel) ©2011 MFMER | slide-73 Use copper filter! vary thickness Lucite 15 cm to 35 cm 20% Iodine solution for CNR/mGy measures ©2011 MFMER | slide-74 ~ 10% Lower Contrast to Noise for 0.1 mm Copper CNR vs mm Cu Acquisition 160 140 120 15cm Lucite CNR 100 20cm Lucite 25cm Lucite 80 27.5cm Lucite 30cm Lucite 60 35cm Lucite 40 20 0 0 0.2 0.4 0.6 0.8 1 mm Cu ©2011 MFMER | slide-75 25 40% Lower Dose with 0.1mm Copper! Differential Dose Acquisition P e rc e n t C h a n g e in D o s e fo r A d d itio n o f C o p p e r 0% -10% 0.0 0.2 0.4 0.6 0.8 1.0 -20% 15cm Lucite 20cm Lucite 25cm Lucite -30% -40% 27.5cm Lucite 30cm Lucite 35cm Lucite -50% -60% -70% -80% -90% mm Cu ©2011 MFMER | slide-76 Use at least 0.1 mm Copper! • Every fixed room without exception • Every protocol!!!! (especially obese pt technique!) • Portable C-arms (check with vendor) • ALARA -100% exams with copper. ©2011 MFMER | slide-77 8 ALARA Behaviors for reproducibility • Proper collimation • Minimize use of Hi Mag • Return to Normal FOV Fluoro prior to DSA • Use of Fluoro Save (with LIH) • Optimize patient positioning • Table distant from tube • DSA and Tap Lightly • Announce cumulative Gy levels ©2011 MFMER | slide-78 26 Collimation Is Key: improves IQ, lowers PSD < 1 Gy 1 -2 Gy 2 - 3 Gy 3 - 4 Gy 4 - 5 Gy 5 - 6 Gy 6- 7 Gy 7-8 Gy 8-9 Gy 9-10 Gy 79 ©2011 MFMER | slide-79 Skin Dose Rate (mR/min) Behavior: choosing small FOV Result: High dose fluoro AND high DSA 6000 5000 Low Normal 4000 High 3000 2000 1000 0 19 17 13 10.5 8 7 6 Diagonal Size (in) ©2011 MFMER | slide-80 Single frame of II fluoro – fluoro save 0.7R/m 8in Fluoroscopy 1.2 R/m 7in Fluoroscopy 2 R/m 6in Fluoroscopy ©2011 MFMER | slide-81 27 II Image Quality and small FOV Spot Normal 8in Single Shot Normal 7in Single Shot Normal 6in Single Shot ©2011 MFMER | slide-82 Control: Are Fluoro Saved images in PACS? Informatic tools can monitor XA radiation events ©2011 MFMER | slide-83 ©2011 MFMER | slide-84 28 - + 5/6/2011 5/13/2011 5/20/2011 5/27/2011 6/3/2011 6/10/2011 6/17/2011 6/24/2011 7/1/2011 7/8/2011 7/15/2011 7/22/2011 7/29/2011 8/5/2011 8/12/2011 8/19/2011 8/26/2011 9/2/2011 9/9/2011 9/16/2011 9/23/2011 9/30/2011 10/7/2011 10/14/2011 10/21/2011 10/28/2011 11/4/2011 11/11/2011 11/18/2011 11/25/2011 12/2/2011 12/9/2011 12/16/2011 12/23/2011 12/30/2011 1/6/2012 1/13/2012 1/20/2012 1/27/2012 2/3/2012 2/10/2012 2/17/2012 2/24/2012 3/2/2012 3/9/2012 3/16/2012 3/23/2012 3/30/2012 4/6/2012 4/13/2012 4/20/2012 4/27/2012 5/4/2012 5/11/2012 5/18/2012 5/25/2012 SET ALARA I and II Levels (AK is ok) 16000 IR Radiation Dose in mGy - EVAR 14000 12000 10000 Sentinel Event (15,000 mGy) 8000 6000 4000 2000 0 Date of IR Procedure Radiation Dose in mGy Too much! Yellow Red Sentinel Event ©2011 MFMER | slide-85 Fluoroscopy Positioning Review Excellent Skin below IRP with Fluoro! ©2011 MFMER | slide-86 Peak Skin Dose Calculator ©2011 MFMER | slide-87 29 ©2011 MFMER | slide-88 Fluoroscopy Service OLD Exposure Rate CURRENT Exposure Rate GI Endoscopy/ERCP 3 ½ Pulses per second 3 ½ Pulses per second Urology 15 Pules per second 3 ½ Pulses per second * IR Room 1 Radiology IR Room 1 Vascular IR Room 1 DSA Aorta, C-E 10 pulses per second 10 pulses per second 4 Frames per second 7 ½ Frames per second * 7 ½ Frames per second * 2 Frames per second ** IR Room 2 Radiology IR Room 2 Vascular IR Room 2 DSA Aorta, C-E 15 Frames per second 15 Frames per second 4 Frames per second 7 ½ Frames per second * 7 ½ Frames per second * 2 Frames per second ** Cardiac Cath Rm 2 and 3 15 Pulses persecond FLUORO 15 Fr/second CINE 7 ½ Pulses per second FLUORO 7 ½ Fr/second CINE ** Electrophysiology Lab Venograms 15 Fr/second FLUORO 30 Fr/second CINE 4 Fr/second FLUORO * 4 Fr/second CINE *** ©2011 MFMER | slide-89 Thank You! • Six Sigma is a way of problem solving • DMAIC • Drive out variation – measure reproducibility • Processes are the problem, not people! • Team effort ©2011 MFMER | slide-90 30