Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30st, 2012 Learning Objectives • Understand proton beam dosimetry characteristics and compare them to photon beams • Familiarize with proton dosimetry QA tools • Understand challenges in proton therapy QA 2 Clinically operating proton therapy facilities WA MT ME ND OR VT MN NH ID WI SD MI NY MA CT RI WY PA IA OH IL UT IN CO OK AZ MD DE UPENN WV VA MO KS Loma Linda University NJ NE NV CA Massachusetts General Hospital KY Midwest Proton Radiotherapy TN Institute AR NC Hampton University SC NM AL GA MS TX LA FL MD Anderson University of Florida Proton Therapy Institute 3 1 Multi-room Facilities FBTR1 IBTR2 IBTR3 GTR4 4 In-room Design Inclined Beam 2 Inclined Beam 1 Gantry Fixed Beam 5 Inside Treatment Room Three major elements of QA: • Imaging System • Positioning System • Beam delivery 6 2 Beam Delivery Techniques Beam Spreading Techniques Passive Scattering Active Scanning vs Single Scattering Uniform Scanning (US) Double Scattering (DS) Pencil Beam Scanning (PB) 8 PB US DS Beam Delivery Techniques 9 3 Beam Characteristics at Depth Dosimetric Advantage of PT 11 Coverage at depth: Protons vs Photons Target Y. Zheng 12 4 Anatomy of a Spread-Out Bragg Peak (SOBP) M distal end lateral penumbra distal penumbra R 40 60 Ref 0 20 Dose (%) 80 90 100 proximal end 0 50 100 Depth (mm) ICRU 78 150 Tolerances Flatness within 5% Symmetry within 3% Range within 1.5 mm Modulation within 5 mm 13 Lateral penumbra at depth 10 MV Uniform Scanning beam data, ProCure - OKC 14 Distal penumbra at depth Uniform Scanning beam data – ProCure -OKC 15 5 Proton vs. Photon PDDs in presence of heterogenieties Photons Loss in Fluence (attenuation) SAME ENERGY Protons Loss in Range (Energy) (degradation) SAME FLUENCE 16 How to manipulate the SOBP beam? detector M BEAM x BEAM x x + y =M BEAM x+y 17 surface What can you get from a SINGLE delivery? beam Get creative with compensator design Get creative with array housing Ding et al …. 2012 18 6 QA of Patient Devices 10 cm snout 18 cm snout IBA Universal Nozzle Nozzle & Snout Design 20 Patient Devices 21 7 Distal end shaping - no compensator Target Proton Beam Aperture Inhomogeneity (Air Pocket) 22 Distal end shaping – with compensator Target Compensator Aperture Inhomogeneity (Air Pocket) 23 Patient Device QA thick for tissue, thin for bone 24 8 Improving QA equipment 25 Output factor measurements D = 11 cm Ref. Beam Patient Aperture 10 cm Aperture R16M10 26 Output factor dependencies Other factors: Field size, snout position, phantom material, dose rate 27 9 Beam QA with 1D Arrays 1D Arrays – How do they compare for PDD measurements? vs 29 Zebra PDDs 30 10 Monthly Range Trend IBL3 31 Beam QA with 2D Arrays Measurements of Flatness & Symmetry Monthly QA Sheet, IBL2 – Jan 2012 33 11 Monthly Flatness Trend – reference beam Flatness Trend - July 2010 to June 2011 3.5 3 Flatness X Flatness (%) 2.5 Flatness Y 2 1.5 1 0.5 0 -0.5 Month IBL3 34 Monthly Symmetry Trend – reference beam Symmetry Trend - July 2010 to June 2011 3.5 3 Symmetry (%) 2.5 Symmetry X 2 1.5 1 0.5 0 -0.5 Month IBL3 35 ProCure Morning QA Device rf Daily QA3 Irradiation area + fiducials ProCure Machine Shop Xiaoning 36 Ding, PhD 12 Imaging QA: Comparing DRR with X-ray Image X Images DRR 37 Morning QA Procedure One setup, One device, One beam to get the following: 1. 2. 3. 4. 5. Output consistency check Range consistency check Symmetry consistency check Imaging vs mechanical alignment check In-room laser check 38 Output Factor Morning QA Trends 1.04 1.02 1 0.98 0.96 6/1/10 8/16/10 10/31/10 1/15/11 4/1/11 6/16/11 8/31/11 11/15/11 Range (mm) 161.5 161 160.5 160 159.5 159 Symmetry (%) 158.5 7/5/11 4 3 2 1 0 -1 -2 -3 -4 8/20/11 8/5/11 8/30/11 9/9/11 9/5/11 9/19/11 10/6/11 9/30/11 10/10/11 10/20/11 39 13 Temporal tracking of PPS correction vector Y X Z 40 Colinearity Test • Purpose: to check that imaging isocenter coincides with radiation isocenter to within 1 millimeter. Proton Iso Imaging Iso 41 Daily Checks Monthly Checks Annual Checks • • • • Imaging vs mechanical alignment Output Range Software Communication • • • • Proton-imaging isocentricity Flatness & Symmetry Ranges and Modulations Mechanical • • • • PPDs + Modulations Combinations of field sizes and gantry angles X-ray source & detector image characteristics Dose rate dependencies 42 14 QA Challenges in PT QA challenges in PT • Proton delivery modes & control systems are complex-more things to check • Lack of methodology or forum to exchange ideas that improves QA processes – very few clinical proton physicists • PT systems are not robust yet – few years of operations, many bugs to resolve (software & hardware) • QA programs highly depend on vendor’s system specs 44 QA Challenges in PT – cont. • There are currently no task group recommendations for proton beam QA. Where relevant we follow guidelines from the following sources: – – – – – – IAEA TRS 398 ICRU 59 ICRU 78 TG 40 TG 142 Journal publications • Lack of dedicated commercial QA devices for PT –adaptation of photon QA devices is necessary 45 15 QA Challenges in PT – cont. • It takes time to switch, tune, and deliver beam in every room –QA tasks takes longer compared to linac systems • Current PT centers have 3-5 rooms with sequentially beam delivery – beam sharing is necessary • Cost of proton specific QA equipment • Multi vendor software/hardware – lack of true integration 46 Anatomy of a linac head • • • • • • • Carousel (scatterers) Magnets Jaws (primary) Jaws (tertiary) Ion chamber MLCs Light field • OUTPUT – Electrons (4-6 energies) – Photons (1-3 energies) 47 Anatomy of a Nozzle • • • • • • • • • • • • Compensator Aperture(s) Snout with variable positions Lollipops Modulator wheels (multiple tracks) Multiple ion chambers Collimators (X-Y) X-Y magnets (3 scanning fields) Range verifier X-ray source Scatterers Light field • OUTPUT IBA Universal Nozzle – Modulation (very large combinations) – Range (very large combination) 48 16 Summary • Proton Therapy Systems are complex and requires specialized equipment to measure various beam parameters • It is imperative to make use of commercially available 1D & 2D arrays and adapt them to PT to check routinely for • • • • Beam parameters (R,M, Symmetry, Flatness, Output) Imaging System Robotic positioning System Standardization of QA procedures for PT is essential in establishing tolerance limits 49 Contributors Yuanshui Zheng Xiaoning Ding Anthony Mascia Eric Ramirez Yixiu Kang Wen Hsi 50 Thank you 51 17