Clinical Implementation of IMRT (TU-B2-01) Optimization and Delivery of IMRT • Optimization (Inverse Planning) • Delivery • Quality Assurance • Radiation Safety Chen-Shou Chui, Ph.D. Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York City, New York Treatment Planning Procedure Image Transfer Pre-optimization Compute Contouring compute dose to point-i from ray-j, for all i, j Beam Definition Optimization Dose Calculation a mathematical iterative optimization process Plan Evaluation • Examples of commonly used objective functions: 8target: (D-P) 2 8critical organs: (D-Dc)2, (D-Dc ) if D>Dc 8dose volume conditions: no more than p% of the volume to exceed dose q. • Reasonable and mathematically convenient. • No fundamental physical basis. Generation of leaf sequencing file Conventional planning steps • Could be any other forms such as |(D-P)m|. additional steps for IMRT Examples of Objective Functions Objective Functions based on Biological Indices organ at risk target w l(D-P l) Objective Functions based on Dose, Dose/Volume 2 (D-Dc) wu (D-Pu) 2 2 • Maximize TCP, minimize NTCPs. • Used in place of or in conjunction with dose-, dose/volume-based objective functions. • At present, clinical data are scarce and models not well-established yet. • Not available on commercial system at present. Pl Pu Dc 1 Global vs. Local Minimum Examples of Optimization Methods (example of local minima) Objective Function 1 • Deterministic: Uniform target dose (100% dose): 1 solution, beams 1&2 equally weighted. 8Gradient, Conjugate gradient. 8Maximum likelihood. target 2 • Stochastic: crit. org. 8Simulated annealing. 8Genetic algorithm. Global Minimum IMRT vs Conventional Conformal Plans Is it important ? no There is no issue. yes IMRT should do better (at least no worse), for there are more degrees of freedom. Dose Distributions Dose Volume Histograms IMRT vs Conventional Conformal Plans 105 Critical organs: no more than 1/2 to exceed 10% dose. 2 solutions, beams 1&2 weighted either (a,b) or (b,a) ⇒ local minima !! Dose Uniformity in the Target • In Principle: 8Stochastic methods can find global minimum. 8Deterministic methods may get trapped in local minima. • In Practice: 8It is not easily known if current solution is at global or local minimum, regardless of which methods are used. 8It may not be important if the solution is clinically satisfactory. 8If solution not satisfactory: ý for stochastic methods: change random selection parameters (e.g.annealing schedule), increase number of iterations. ý for deterministic methods: try different initial guess. ý is the new solution better? target Critical organs: no more than 1/2 to exceed 50% dose. 1 solution, beams 1&2 equally weighted. 100 IMRT vs Wedged Pair 100 70 IMRT 80 Volume (%) 45° 45° Wedged Pair 60 40 target 20 rest of volume 0 0 20 40 60 80 100 120 Dose (%) Wedged pair IMRT pair 2 Nasopharynx IMRT and Conformal Plans Nasopharynx Plan Comparison IMRT Constraints Beam arrangement Target Minimum: 100% Uniformity: < 20% Cord max: 40 Gy Brainstem max: 45 Gy Prescription – Gross disease: ≥ 70 Gy – Micro. Disease ≥ 54 Gy – Bid after 36 Gy 7700 7400 7000 6500 Conformal Plan 5000 Normal tissues – – – – 3500 Cord, brainstem Mandible Parotids Temp. Lobes 2000 LATS TO 36 Gy Plan for 34 Gy %VOLUME 100 100 PTV 80 60 40 40 C o n f. 20 CORD 1D Intensity Profiles 80 60 before & after smoothing C o n f. IMRT 20 IMRT 0 0 0 2000 4000 6000 8000 0 2000 4000 DOSE (cGy) 100 %VOLUME IMRT Conformal 6000 8000 DOSE (cGy) Savitzky-Golay (least squares fit) 100 MANDIBLE TEMP. LOBES 80 80 60 60 Conf. 40 40 IMRT Conf. 20 20 IMRT 0 0 0 2000 4000 6000 8000 0 2000 DOSE (cGy) 4000 6000 DOSE (cGy) 2D Intensity Distribution 8000 before smoothing after smoothing The ‘Skin-Flash’ Problem in Inverse Planning before & after smoothing Conventional: Margin added to field edge to allow for uncertainties. IMRT: PTV Intensity remains at 0 outside PTV. No skinflash before smoothing after smoothing 3 Skin-Flash for Intensity-Modulated Field Incorporation of Uncertainties in Inverse Planning Simple, flat extension • Each point in the region of interest (e.g. CTV) is split into multiple points (typically 3) to account for the range of uncertainties. skin • Corresponding rays are brought into optimization. CTV • Performed in pre-optimization. Skin-Flash for Intensity-Modulated Field Incorporation of Uncertainties in Optimization Beam’s-Eye-View: Lateral Field Without skin flash With 2 cm skin flash skin Summary - Optimization Delivery of IMRT • Objective functions may be based on dose, dose/volume conditions, or biological indices. • Optimization methods may be stochastic or deterministic. • Local minima may exist, but there is no easy way to tell whether a solution is at a global or local minimum, regardless of which optimization method is used. • Optimized intensity distribution should give better dose distribution than conventional conformal plans. • Smoothing of intensity profiles may be useful for practical reasons. • Other clinical considerations, such as “skin flash” problem. • Compensator: less efficient (fabrication, re-entry into the room between fields). • Fixed field with conventional MLC: 8continuous leaf motion, 8step-and-shoot. • Rotational field: 8conventional MLC, 8NOMOS/MIMIC, Tomotherapy . 4 Delivery of IMRT with a Conventional MLC Delivery of IMRT with a Conventional MLC sliding window method radiation Continuous direction of motion Right-leaf Left-leaf Beam-on-Time T beam-on time Step-and-shoot T d Left-leaf d Left-leaf c c Right-leaf delivered intensity delivered intensity ∆t = 1 ≤ 1 ∆x v vmax a Right-leaf b b X P a X P P Clinac 2100C 15-MV X-rays Intensity Modulation by DMLC Delivery of IMRT with continuous leaf motion • Each leaf pair forms a window which slides across • Dose given through the window as function of MU ••• ••• 1 i N 1+...+ I +… +N • The final dose distribution is the summation from all “segments” Prostate Generating MLC Segments step-and-shoot Step-and-Shoot MLC Segments Field shapes and relative intensity weights Intensity Profile Intensity Grouping unconstrained intensity levels limit delivery to a few discrete intensity levels Beam Segments includes MLC constraints • A bitmap of the optimized beam intensity profile is converted into deliverable MLC fields, taking MLC positioning constraints into account. William Beaumont Hospital William Beaumont Hospital 5 Variation of Output with Field Shape/Size Backprojection to the Source Plane Total intensity at P : Intensity as a function of position from the leaf end beam Primary Source Scatter Source Effects of Rounded Leaf-End and Leaf Transmission Source plane T φ p = ∫ I(xr (t)-p) I(p-xl (t)) dt t=0 leaf end x(t) MLC opening left leaf MLC plane P beam-on time direct exposure 1 right leaf ε Right-leaf Left-leaf I(x) leaf penumbra leaf transm. Isocenter plane direction of motion x P P Iterative process to generate leaf paths Organ Motion Relative to Leaf Motion Desired fluence profile F(x) Assign Fwork (x) = F(x) T Beam-on-Time Modify working profile Fwork(x) = Fwork(x) - E(x) Parallel Perpendicular Calculate leaf paths using Fwork(x) Calculate generated profile Fg(x) taking all factors into account Calculate error E(x) = Fg(x) - F(x) d Left-leaf d’ Right-leaf c’ c b’ b Desired intensity a a’ error acceptable ? finish Effects of Intra-Fraction Organ Motion on Delivered Intensity X Y P P Intra-fraction breathing motion A = ±3mm, t = 4 sec. Total intensity at P : T φ p = ∫ I(xr (t)-p) I(p-xl (t)) dt t=0 Static : Intensity as a function of position from the leaf end beam leaf end x(t) Organ motion : f : motion function t: beam-on-time τ: period A: amplitude to : initial phase Averaged over 30 fraction I(x) P is constant. P = f(t;τ,A) + t o Single fraction direct exposure 1 ε leaf transm. leaf penumbra x Leaf pair # 14 6 Intra-fraction breathing motion Splitting a Large IMRT Field A = ±3mm, t = 4 sec. 105 103 50 IMRT Averaged over 30fx • Due to the design of MLC, the largest IMRT size that can be delivered without jaw repositioning may be limited (e.g. 14.5 cm on the current Varian MLC). • The most straight forward solution is to split the field at an intermediate position Xs, but it is less efficient (Tl-T r) and prone to field matching uncertainties. T T Time IMRT Planned 100 x2 Field width x2-x1 > 14.5 cm Splitting a Large IMRT Field (cont’d) L2 T Left leaf Ls Ts L1 R2 Rs Right leaf F(x) F2(x) F1(x) x2 xl xr Split at a beam-on-time 0 x 1R 1 T L2 Left leaf A R2 Right leaf Tb L1 R F1(x) 0 x1 1 x B b F(x) F2(x) xa Rs R 0 x 1 1 X R2 Right leaf F(x) xs Split at a position x2 Summary - Delivery • Another way is to split the field at an intermediate path, with each segment satisfying the field size limitation. It provides a more general solution but is less efficient in beam-on-time. Ta Ls L1 0 x 1 • The most efficient way is to split the field at an intermediate time Ts, if each segment meets the field size limitation. It is most efficient (same beam-ontime) and is insensitive to field matching uncertainties. Left leaf Tr Right leaf F(x) Transverse View L2 Tl Left leaf x2 • Intensity-modulated field can be delivered with a conventional MLC, either in continuous or step-and-shoot mode. • Need to account for leaf transmission, rounded leaf end, and head scatter. • Increased beam-on-time comparing to conventional treatment: - prostate, 2-3 times. - head & neck, 3-4 times. - breast, about the same. • Need to account for intra-fraction organ motion, if present: single fraction & multiple fractions. • Split large field into 2 or more segments, if necessary. Split with an intermediate path Gap error → Dose error Quality Assurance 20.0 Range of gap width • Machine Performance : • Patient Dosimetry: 8Record & verify, file check-sums (each fraction) 8Independent MU check, portal image (each patient) 8Log file analysis, chamber measurement, film dosimetry (periodically or new technique, new site) 15.0 % Dose error 8Film test (semi-weekly) 8Dosimetry test (monthly) 8Drift test (monthly) Gap error (mm) 10.0 2.0 1.0 5.0 0.5 0.2 0.0 0 1 2 3 4 5 Nominal gap (cm) 7 Film test 1 mm bands errors introduced DMLC Output vs Time 0.170 0.168 DMLC / 10x10 ← ← - 0.5 mm ← ← - 0.2 mm ← ← + 0.2 mm 0.2 mm 0.166 0.164 0.162 245 ← ← + 0.5 mm 0.160 8 9 8 7 8 8 7 r-9 r-9 un-9 g-9 c-9 c-9 p-9 Ma Se De Ma J Au De 445 Date Independent Monitor-Unit (MU) Calculation • Required by regulations, traditionally hand-calculation by the user, 1.03 1.03 not part of the treatment planning system. 1.02 1.02 • For IMRT, such hand calculation is difficult. Independent program 1.01 1.01 is needed to perform MU check. 1.00 0.2 mm 1.00 0.99 0.99 0.98 0.98 245 0.97 8 7 8 7 8 98 t-9 v-9 n-9 pr-9 n-98 ug- ct-9 Oc No Ja A Ju A O Date • Uses the Beam-on-Time (MU) and leaf sequence file (DVA) as input, calculate dose distributions in a flat phantom. • Takes into account rounded leaf-ends, extended source, and leaf 445 0-0 0.97 90-0 270-0 8 7 8 8 7 98 t-9 v-9 an-9 pr-9 un-9 ug- ct-98 180-0 Oc No J A J A O 90-90 270-90 Date transmission. • Agreement between calculation & measurement, generally < 2% in dose or < 2mm • Serves as independent check to treatment planning system. Nasopharynx - Left Lateral Measured vs Calculated Sum of 5 IMRT fields for prostate treatment Sup 1.03 6 MVx 1.02 Ant Dmeas / Dcalc Relative output DMLC Output vs Gantry / Collimator Angles Post 1.01 1.00 0.99 0.98 0.97 0.96 Inf Film Plan mean = 0.993, s = 0.008 0.95 0 245 100 445 200 300 400 Patient # 8 Beam Stability: Dose Rate Step-and-Shoot 1.04 Beam Stability: Dose Rate SL20 - 18MV u For > 3MU, dose rate is within +/-2% (2s). u u 1.03 Relative Dose Rate u With step-and-shoot delivery, there is the potential for short irradiation times (MUs). Dose rate stability influences the treatment precision. Measure dose per MU versus total MU. Check short, and long term stability. u Initial 2 weeks later 6 months later, (new monitor chamber) 1.02 1.01 1.00 0.99 1 10 MU 100 William Beaumont Hospital William Beaumont Hospital Beam Stability: Flatness, Symmetry Beam Stability: Flatness, Symmetry u Stability of flatness and symmetry affects dose rate for small fields directed off the central axis. u F o r a n o p e n 2 0 x 2 0 c m 2 field, measure profiles for irradiation ranging from 1 to 100 MU. Symmetry 0.10 u Symmetry is less than +/-3% if more than 3MU delivered. Flatness – Sun Nuclear Profiler (46 diodes, 10 profiles/sec). u Flatness is less than +/-3% if more than 5MU delivered. GUN-TARGET DIRECTION CROSS-PLANE DIRECTION 0.05 0.00 -0.05 0.10 0.05 0.00 -0.05 1 Radiation Safety Considerations • Whole Body Dose: (for more details, see Followill et al. IJROBP.38: 667-672, 1997.) To estimate the total body dose delivered to patients by photons and neutrons outside the radiation fields when intensity modulated beams are used for treatment. These estimates are then used to compute the risk of induction of secondary cancers as a sequella of the radiation therapy. • Room Shielding Requirement: (for more details, see TU-D2-5) To calculate the shielding required as a result of increased beamon-time due to delivery of intensity-modulated beams. 10 100 Total M U s Delivered William Beaumont Hospital William Beaumont Hospital X-ray and Neutron whole-body dose equivalent (mSv) per unit calibration dose (cGy) Radiation Type X-ray Beam Energy 6 MV 18 MV 25 MV x-ray 8.0 x 10-2 6.5 x 10-3 1.0 x 10-2 neutron 0.0 4.6 x 10-2 7.6 x 10-2 9 MU/cGy to deliver conventional and modulated beam intensity radiotherapy Conventional Beam Energy Total whole-body dose equivalent (mSv) for delivered dose of 7000 cGy at isocenter Beam Intensity Modu lated unwedged wedged Varian MLC Nomos MLC (MU/cGy) (MU/cGy) (MU/cGy) (MU/cGy) 6 MV Conventional MLC M odul ated 6 MV 1.2 2.4 3.4 9.7 18 MV 1.0 1.5 2.8 8.1 25 MV 1.0 1.5 2.8 8.1 Tomotherapy 18 MV 25 MV no wedges wedges no wedges wedges no wedges wedges 67. 190. 543. 326. 911. 2637. 602. 1686. 4876. 134. ------- 488. ------- 903. ------- Estimated percent likelihood of a fatal secondary cancer due to a 7000 cGy course of IMRT 6 MV Conventional MLC M odul ated Tomotherapy 18 MV 25 MV no wedges wedges no wedges wedges no wedges wedges 0.3% 0.8% 2.2% 1.3% 2.0% 3.6%` ---10.5% ---- 2.4% 3.6% 6.7% ---19.5% ---- 0.5% ------- Conclusion: Careful consideration should be made of the implications associated with secondary whole body radiation before implementation of beam intensity modulated conformal therapy using x-ray energies greater than 10 MV. Shielding Primary Barrier • IMRT = relatively inefficient delivery • New formalism required • Evaluate • Standard: – workload = TD = MU (50-100K/week) = dose equivalent • DMLC – workload = TD (not MU) = dose equivalent – Primary – Scatter – Leakage • Tomotherapy – workload = TD * N * F = dose equivalent • Decouple the workload from MUs Mallincrodt Inst . of Radiology • N = average number of indexes • F = fraction of arc subtended by primary beam Mallincrodt Inst . of Radiology 10 Secondary Barrier Secondary Barrier • Leakage – Standard • Patient Scatter • workload = TD = MU (50-100K/week) = dose equivalent – Standard: • workload = TD = MU (50-100K/week) = dose equivalent – DMLC • workload = TD * E = MU = dose equivalent – DMLC – E = efficiency (approx 1.7) • workload = TD (not MU) = dose equivalent – Tomotherapy – Tomotherapy • workload = TD (not MU) * α = dose equivalent • workload = TD * N * E = MU = dose equivalent – N = average number of indexes – E = efficiency (approx 2.5) Mallincrodt Inst . of Radiology Mallincrodt Inst . of Radiology Conclusions Neutron Shielding • Optimized IMRT plans can produce better dose distributions than – Standard • workload = TD = MU (50-100K/week) – DMLC conventional conformal plans, in terms of both target coverage and normal organ sparing. • Delivery of IMRT with a conventional MLC is practical, either in continuous or in step-and-shoot mode. Delivery with MIMIC is also practical, but in general requires longer beam-on-time. • workload = TD * E = MU – E = efficiency (approx 1.7) • Comprehensive QA is needed, for machine performance and – Tomotherapy patient-specific dosimetry. • workload = TD * N * E = MU • Radiation safety issues need to be considered. • PROCEED WITH CAUTION ! – N = average number of indexes – E = efficiency (approx 2.5) Mallincrodt Inst . of Radiology Acknowledgment MSKCC MD Anderson Spiridon Spirou David Followill Jie Yang Stanford Margie Hunt William Beaumont Di Yan John Wang Art Boyer Linda Hong Tom LoSasso Mallincrodt Dan Low Sasa Mutic 11