Acknowledgements Monte Carlo-based Clinical Treatment Planning: Issues for Consideration • E. Heath, G. Stroian, K. Al-Yahya, W. Parker, F. Verhaegen • F. Deblois and A. Alexander: MMCTP GUI • Canadian Institutes of Health Research, National Cancer Institute of Canada and Natural Sciences and Engineering Research Council for grant funding & salary support (CIHR, NCIC, NSERC) Indrin Chetty and Jan Seuntjens Univerisity of Michigan, Ann Arbor, MI and McGill University, Montreal, QC AAPM Summer School 2006 Outline: “issues for consideration” • • Building blocks of a generic clinical MCTP system Verification issues in general • • • • CT calibration & artifacts Dose specification Absolute calibration, MU calculations Verification issues in detail • • • • • – Validation of relative output – Experimental verification of MCMC-based dose algorithms: beam modifiers (MLC) Experimental verification of transport within the phantom/patient phantom/patient – 2 Plan 1 in a lung cancer treatment Dose (Gy) __ 4-8 __ 8-12 __ 10-20 __ 20-30 __ 30-36 __ 36-38 __ 38-40 __ 41-42 __ 42- (a) (b) EqTAR corrected Monte Carlo MCMC-based treatment planning: comparisons of MC versus simple (correction(correction-based) and modelmodel-based algorithms Statistical Uncertainties in MCMC-based treatment planning MCTP and lung (NSCLC) Retrospective re-planning & outcome association Summary Difference map for PTV area Outcome? (c) AAPM Summer School 2006 treatment plan or treatment delivery CT slices composition of (e)density matrix 3 Beam information extractor Patient modifications: CT couch removal, bolus, CT artifact-cleanup Validation beam configuration Primary beam Beam modeling Beam calculation Phasespace or source model for each treatment field Absorbed dose to water calculation patient Monte Carlo MC-plan displayed in popular planning system Patient dose calculation AAPM Summer School 2006 Beam modifiers Reference and relative “output” in defined phantoms Relative “output” in patient geometries with defined materials 6 Validation - Wedge 100 Rel. Dose (%) 80 45 degrees wedge profile, 18 MV Photon Beam 140 3 cm (Measured) 3 cm (XVMC) 3 cm (dosxyz) 120 100 MC film CA24 60 40 20 0 -8 -6 -4 -2 0 2 4 6 8 Crossplane (cm) 80 100 60 80 Rel. Dose (%) 40 20 0 -15 -10 -5 0 5 10 Dynamic IMRT Delivery 15 MC Film CA24 60 40 20 Off-axis position (cm) 0 -8 Wedge factors -6 -4 -2 0 Inplane (cm) 2 4 6 8 Output factors 15x15 cm2 field size Wedge angle 15 30 45 Measurement 0.755 0.623 0.514 Simulation 0.754 0.621 0.514 Uncert. (%) 0.8 0.8 0.8 Difference (%) -0.1 -0.2 0.0 5x5 cm2 field size Wedge angle 15 30 45 60 Measurement 0.756 0.616 0.512 0.422 0.421 Simulation 0.765 0.609 0.518 Uncert. (%) 0.7 0.9 0.7 0.7 Difference (%) 0.9 -0.7 0.6 -0.1 Field size 5x5 10x10 15x15 20x20 measured 0.919 1.000 1.036 1.058 simulation 0.918 1.000 1.030 1.055 difference 0.001 0.006 0.003 AAPM Summer School 2006 Reference dose calibration Reference dose calibration (cont’d) DwMC (10 × 10,SSD = 100) = k MC [Gy / particle] particle Model-based system: fluence-based calibration Dwref (10 10,SSD = 100) = k [Gy /MU] U MC calculated reference dose to water per particle (fluence); Particle fluence at an accelerator reference point (usually upstream from monitor chamber) U: a monitor unit (quantity with dimension MU) Dw dose to water at the clinical reference point k = 0.01 Gy / MU. If accelerator tweaked in terms of dose to tissue: then k = 0.0101 Gy / MU. AAPM Summer School 2006 10 11 AAPM Summer School 2006 12 Reference dose calibration (cont’d) Reference dose calibration (cont’d) MC ⎞ D MC (r) ⎛ k (r) Dtissue [Gy/MU] = tissue ⎜ [particle/MU]⎟ U particle ⎝ k MC ⎠ MC ⎞ D MC (r) ⎛ k (r) Dtissue [Gy/MU] = tissue ⎜ [particle/MU]⎟ B(x, y) U particle ⎝ k MC ⎠ The B(x,y) factor accounts for the change in output due to monitor backscatter. MC calculated reference dose to water per particle (fluence); Particle fluence at an accelerator reference point (usually upstream from monitor chamber) 1. Relatively small effect (<1% for 10x10 and smaller) 2. Effect reduces for higher energies. 3. Effect is smaller for electrons. 4. Varian series only. This approach ignores backscattering of moving components into the monitor chamber! AAPM Summer School 2006 treatment plan or treatment delivery CT slices AAPM Summer School 2006 13 composition of (e)density matrix Verhaegen et al 2000, Phys Med Biol 45, 3159 - 70 14 Photon beams: 6 MV, 15 MV, 250 kVp beam configuration Electron beam: 18 MeV No contaminant particles Beam information extractor Patient modifications: CT couch removal, bolus, CT artifact-cleanup Beam calculation Phasespace or source model for each treatment field CT scan patient Monte Carlo MC-plan displayed in popular planning system Exact geometry HU interval ρ interval Air [-1000 : -950] [0.001 : 0.044] Lung [-950 : -700] [0.044 : 0.302] ICRU tissue [-700 : 125] [0.302 : 1.101] ICRP cortical bone [125 : 2000] [1.101 : 2.088] Material Dose calculation: DOSXYZnrc Courtesy: Frank Verhaegen AAPM Summer School 2006 16 6 MV photons 2.1 100kV 120kV 140kV ctcreate ramp Catphan 100 kV bonefit 100kV -3 mass density (g.cm ) 1.9 ≈ 2% underdose in CTMC calculation 1.7 1.5 1.3 (5) 1.1 ( μen/ρ)w > ( μen/ρ)soft tissue (8) (3) 0.9 (10) (1) Region 1: large overestimation of ρ (9) (4, 6, 7) (2) -300 0.7 -100 100 300 500 700 900 HU Dose (Dexactin –CTMC DdefaultMap geometry )/ Ddefault Material (HU, ρ) relation AAPM Summer School 2006 Courtesy: Frank Verhaegen 17 AAPM Summer School 2006 Courtesy: Frank Verhaegen 18 CT artifacts, editing Conclusion: mis-assignment of media and ρ can cause significant dose errors Question: Does using only water with ρ derived from (HU,ρ) worse than risking mis-assignment of media? 6 MV 15 MV 250 kVp 18 MeVe default water only Catphan * default water only Catphan* default water only Catphan* default water only Catphan* μ (%) 0.61 0.30 0.77 0.59 0.25 0.73 0.64 0.75 0.66 0.58 0.17 0.79 σ (%) 0.26 0.36 0.32 0.30 0.33 0.33 0.95 2.0 0.91 0.72 0.92 0.99 Min (%) -8.5 -7.9 -9.0 -8.2 -6.0 -9.6 -45 -5.2 -45 -12 -10 -12 Max (%) +10 +5.9 +10 +8.3 +6.6 +10 +49 +78 +47 +9.7 +6.0 +34 • artifacts (dental fillings, etc) • couch issues (CT vs. treatment) • bolus or shielding editing onto CT data Not assigning media (other than water) is sometimes preferable over mis-assigning media! -> be critical about your CT calibration! -> be careful with how a manufacturer handles (HU - interaction coefficient) AAPM Summer School 2006 Courtesy: Frank Verhaegen 19 AAPM Summer School 2006 Dose specification The debate… • Assuming proper material specification MC dose calculations will provide absorbed dose specified to tissue, Dtissue In TP, do we really want the dose specified in terms of Dtissue? AAPM Summer School 2006 20 21 • Dtissue is the real quantity needed! • Reply: Yes, but life is not perfect! • Reply: Former algo’s are not accurate so they represent anything but Dw • Converting back to Dw is adding complexity • Dw is what is being used in former algo’s. – material specification is far from perfect – CT voxels are too large – soft tissues within bony matrix • Reply: The incremental complexity is not really worthwhile talking about. AAPM Summer School 2006 22 Validation Primary beam Beam modeling Siebers et al 2000 PMB 45, 983 Absorbed dose to water calculation Patient dose calculation AAPM Summer School 2006 23 AAPM Summer School 2006 Beam modifiers Reference and relative “output” in defined phantoms Relative “output” in patient geometries with defined materials 24 Internal consistency All tissues to water, plan 2 120 100 Left lung 80 60 CADplan xVMC 40 20 0 Monte Carlo – all tissues to water CADplan All tissues to water, plan 2 120 100 80 cord The conventional planning system uses a pencil beam algorithm once MLC is used as a beam shaping device 60 40 20 CADplan xVMC 0 0 500 1000 Dose (cGy) 1500 0 500 1000 1500 Dose (cGy) 2000 2500 AAPM Summer School 2006, Windsor, Canada Monte Carlo-based Clinical Treatment Planning: Issues for Consideration Indrin J. Chetty and Jan Seuntjens University of Michigan, Ann Arbor MI and McGill University, Montreal QC AAPM TG 105 report Many of the findings and recommendations presented here are reported in the AAPM Task Group No. 105 report: “Issues “Issues associated with clinical implementation of Monte CarloCarlobased treatment planning” submitted to Med Phys IJ Chetty, Chetty, B Curran, J Cygler, Cygler, J DeMarco, DeMarco, G Ezzell, Ezzell, B Faddegon, Faddegon, I Kawrakow, Kawrakow, P Keall, Keall, H Liu, CC-M Ma, DWO Rogers, D SheikhSheikh-Bagheri, Bagheri, J Seuntjens, Seuntjens, JV Siebers How should verification be performed for MC ? Outline Experimental verification of MCMC-based dose algorithms: beam modifiers (MLC) and transport within the phantom/patient B. Clinical treatment planning examples: MC vs. EPL and CS C. Statistical Uncertainties in MCMC-based treatment planning D. Conclusions A. Experimental verification of MCMC-based dose algorithms A Monte Carlo-based treatment planning system is just another treatment planning system and as such should be subjected to the same testing and verification requirements as any other planning system (such as provided in AAPM TG-53 – Fraass et al.) MC is known to be more accurate under nonCPE conditions, and should be verified under such situations MC modeling of beam modifying devices The BEAMnrc code system (NRCC) was developed for transport through detailed MLC geometries: component module developed by Heath and Seuntjens (PMB 48: 4045404564(2003)) for the Varian Millenium 120120-leaf MLC Target Collimator treatment head A. Vacuum Win Flattening Filter Ion Chamber Jaws MLC patient Courtesy of J Siebers, P Keall, MCV The BEAMnrc code system is not fast enough for routine clinical treatment planning (patient(patient-dependent calculations on the order of minutes) Department of Radiation Oncology • University of Michigan Health Systems 1 MC modeling of beam modifying devices Explicit MLC transport: BEAMnrc module To improve the efficiency of simulation through beam modifiers, investigators have developed other approaches: (a) Modified transport methods: first Compton scatter or multiple order scatter, ignoring Compton electrons (b) Multiple source models parameterized for simulating square shapes defined by the jaws; requires explicit or empiricallyempirically-based model for transport through the MLC Explicit “approximate” MLC transport Heath and Seuntjens (PMB 48: 40454045-64, 2003) First Compton transport approximation Tongue-and groove effect maximized: Explicit approximate MLC transport In PEREGRINE Hartmann Siantar et. al. Med. Phys. 28 (2001) 120.0 1 2 3 4 5 6 7 8 9 10 11 12 13 0 1 Siebers et al. (PMB 47:322547:3225-49, 2002) Sensitivity of closed leaf profile to the rounded leaf tip (from Tyagi et al) 2 6 5 5 4 4 3 3 2 2 1 1 0 0 0 Multiple order Compton scatter approx. Delivered with even/odd leaves closed half the time, resp. rel. dose R=1cm R=2cm 9 9 8 8 7 7 6 100.0 10 R=3cm 11 10 MLC leaf tip 12 11 13 12 13 80.0 60.0 40.0 R=4cm Film 5cm Film DPM 8cm DPM 6 mm 20.0 0.0 -2 -1 0 1 2 3 X (cm) Tyagi et al. (Submitted to Medical Physics) x (cm) Department of Radiation Oncology • University of Michigan Health Systems 2 The importance of accurate vendor information ! 15 MV, 10x10 profile in water at 10 cm 1 FF w/ Cu design 0.5 0.25 0 -10.0 -7.5 -5.0 -2.5 Relative Dose 0.75 FF w/ Tungsten Patient Transport Experimental verification in phantom geometries Solid Line = Measurement Distance (cm) 0.0 2.5 5.0 7.5 10.0 Experimental verification in heterogeneous geometries water Verification – high density inhomogeneities water water water Arnfield et al. (MCV), Med. Phys, 27 (6) 2000 Ma et. al. Med. Phys. 26 (1999) Verification – Other geometries MCMC-based beam modifier transport: Summary • As we move MC closer to the clinic, we need more clinically realistic patient geometries ρ=0.31 ρ=1.0 tumor ρ=0.31 ρ=0.31 S 8 cm Solid water 8 cm Solid water 30 cm 30 cm Rice et al, IJROP, 15, (1988) Wang et al Med. Phys., 26 (1999) In summary, regardless of the methods used in modeling beam modifiers, appropriate experimental verification is necessary Experimental testing should include complex configurations designed to verify the improved accuracy expected with the use of the Monte Carlo method • Accurate measurements are difficult !!! Department of Radiation Oncology • University of Michigan Health Systems 3 Treatment Planning: The main dosimetric issue B. Monte CarloCarlo-based treatment planning Solid = MC , 100% Dashed = EPL, 100% Blue = PTV 6 MV oblique fields Underdosage of the PTV DVH for the PTV Percent Volume Comparison of MC vs. simple (EPL) and sophisticated heterogeneity correction methods (CS) in lung, head and neck cancer planning, and other treatment sites MC EPL Percent Dose 15 MV conformal lung plan 15 MV conformal lung plan Note the differences in the spatial dose and dose gradient due to penumbral broadening EPL 6 MV AP/PA lung plan: 95% IDL coverage % Volume MC % Dose MC CS EPL DVH - left lung MC % Volume EPL EPL DVH - CTV MC EPL % Dose 19 field (parotid sparing) head/neck forward plan MC CS EPL Department of Radiation Oncology • University of Michigan Health Systems 4 19 field (parotid sparing) head/neck forward plan Prostate (4 fields + 3 segments) MC EPL DVH – PTV MC EPL % Volume Convolution/Superposition % Dose Prostate DVH’s 100 DVH’s for the PTV and normal liver Difference map (MC-EPL) sagittal view 100 PTV 80 PTV 80 60 % Volume % Volume Liver MC EPL 60 rectum 40 < 2% 40 MC 20 20 % Dose 0 20 EPL 0 0 40 60 80 0 Blue/red colorwash < 5% 100 normal liver % Dose 25 50 75 100 beam penumbra effects Conventional (1 cm collimation) MC (1 cm) collimation Comparison of “model“model-based” algorithms The agreement between the MC and other modelbased methods (e.g. CS) will strongly depend on the particular implementation of the algorithm 120 CS (1) dose Radiosurgery CS (2) MC 100 Ion Chamber 80 60 40 Solberg et. al. Radiother. Oncol. 49 (1998) Electron disequilibrium always exists whenever the field size is smaller than the range of the secondary electrons ρ=0.25 20 ρ=1.0 depth (cm) 0 0 4 8 12 16 20 24 28 32 CS (1): 2 rep. component energies CS (2): 1 rep. energy averaged over spectrum Both CS (1) and (2) agree in water-based tests Department of Radiation Oncology • University of Michigan Health Systems 5 Summary It’s not just the technology The differences found in comparing the MC method with other algorithms will be highly dependent on the beam arrangements, field sizes, beam energies, and tumor size and location Situations where there is a lack of CPE (small field sizes, high energies, low density tissues) may pose a challenge even for CS algorithms because these algorithms do not explicitly transport electrons It’s how you drive it ! C. Statistical uncertainties in MC-based treatment planning “The only certainty is that nothing is certain” Pliny the Elder Roman scholar & scientist (23 AD - 79 AD) Implementation is critical ! Courtesy: Solberg (UNMC) Statistical uncertainties Probability “Jittery” isodose lines due to the stochastic nature of the MC method are quite different from dose distributions computed with conventional (deterministic) algorithms σ ~ 1/√N, 8.0 N= total no. of 0.05 Gy particles simulated 6.0 σ 4.0 2.0 μ 0.0 0.8 0.9 Dose (Gy) 1.0 1.1 1.2 In tx planning, Relative uncertainty = σ/ μ ~ 1/√dose Statistical uncertainties • Two sources of uncertainty: treatment head simulation (latent uncertainty – term coined by Sempau) and the patient simulation • Latent uncertainty in the phase space is systematic in nature, while the uncertainty in the phantom dose is random • The statistical uncertainty in calculated dose will approach (as a function of 1/√N, where N is the number of simulated particles), the finite, latent uncertainty associated with the phase space, regardless of the number of times the phase space is sampled Department of Radiation Oncology • University of Michigan Health Systems 6 Statistical uncertainties: previous work • Several investigators have published on statistical uncertainties in Monte Carlo dose calculation (see the chapter for references) 3F lung plan (RT_DPM): relative uncert. uncert. rel. uncert. = (1σ/μ)x100 % 10 million particles AP 55% 9% 25% • Clinical planning studies show that a statistical uncertainty of 2% or less does not significantly affect isodose lines, DVHs, or biological indices L LAT 7% 20% GTV PA Clinical plan: one sigma % uncertainty Clinical plan: one sigma % uncertainty 1.5 billion particles 150 million particles 20% 15% 5% 5.5% 9% 3.5% 0.5% 4% 1.8% 1.5% 0.5% rel. uncert. = (1σ/μ)x100 % rel. uncert. = (1σ/μ)x100 % Effect of uncertainties on the 95% IDL Effect of uncertainties on DVHs 150 million 10 50million million 1.5 billion 100 10 milion 150 million 1.5 billion Volume (%) 80 60 40 20 Dose (%) 0 90 95 100 105 DVH for a plan w/given uncertainty is derived by convolving the DVH for the “0%”uncert. plan with the given random uncertainty distribution 110 Department of Radiation Oncology • University of Michigan Health Systems 7 Uncertainty volume histograms (UVHs (UVHs)) Volume (%) 100.0 Direct UVHs for the CTV cumulative UVH % vol. 50.0 150E6 80 500E6 60 1500E9 cumulative DVH 75 50 25 0 0 20 0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 Relative uncertainty (%) Effect of uncert.: uncert.: parallel organs (lung) No. of % σ/μ histories (min.) % σ/μ (max.) % σ/μ (mean dose) MLD (Gy) % NTCP 66 Gy 10E6 6.5 100.0 46.0 6.5 0.51 150E6 1.7 41.0 18.9 6.8 0.55 500E6 0.9 23.8 10.5 6.8 0.55 1500E9 0.5 17.9 6.1 6.8 0.55 Statistical uncertainties: 4D planning Exhale Intermediate states Inhale 0 20 40 DE 80 100 Relative uncertainty (%) 60 IJ Chetty et al. “Reporting and analyzing statistical uncertainties in Monte Carlo–based treatment planning” in press Red Journal Effect of uncert.: uncert.: serial organs (cord) No. of histories % σ/μ (min.) % σ/μ (max.) % σ/μ (mean dose) 10E6 12.6 97.3 42.2 150E6 3.3 28.0 12.9 3.7 500E6 1.8 15.2 7.1 2.0 1500E6 1.0 8.8 4.2 1.2 % σ/μ (D>0.5 Dmax) 14.3 Statistical uncertainties: 4D planning Inhale (mapped onto exhale CT) Exhale 5.5% 4% 1.8% Dint 60 40 dose (cGy) 20 0.0 See: Rosu et al. Med Phys 32:248732:2487-95 (2005) 100 40 25.0 Deformable organ dosimetry % vol. 10E6 150E6 500E6 1500E9 100 10E6 75.0 UVH’s/DVH’s for normal lung tissue 0.5% 6% 0.6% 0.5% 2% 0.5% DI Cumulative (exh+inh) 3.5% Dosevoxel = wE DE + wintDint + wI DI W = weighting factors derived from a breathing function: Lujan et al. Med Phys (Med. Phys. 26, 715-20 (1999)) 1.4% 0.45% 0.4% Ref: IJ Chetty et al. “Reporting and analyzing statistical uncertainties in Monte Carlo–based treatment planning” in press Red Journal Department of Radiation Oncology • University of Michigan Health Systems 8 σ D >0.5 D max Statistical uncertainties – Summary • AAPM TG-105 report recommendations: Report uncertainties to volumes rather than individual points, i.e. σD>0.5Dmax, σPTV, or σPRV σ D >0.5 D • Point doses (ex. Dmax points) are subject to large statistical fluctuation (See Kawrakow and Fippel, PMB: 45, 2163-83(2000)) max • Point (individual voxel) dose uncertainties may be a concern for serial organs and requires further investigation • Prescribing doses to a line may obviate the statistical issues with point doses D. Conclusion • A properly commissioned MC-based dose algorithm will improve dose calculation accuracy in 3D-CRT and IMRT treatment planning and is likely to improve dose-effect correlations • Clinical implementation of MC-based systems must be performed thoughtfully and physicists must understand the differences between MC-based and conventional dose algorithms • Successful implementation of clinical MC algorithms will require strong clinician support and an understanding of the paradigm shift with MC algorithms Acknowledgements Neelam Tyagi Mihaela Rosu Eduardo Acosta Martha Coselmon Jean Moran Dale Litzenberg Daniel McShan Randall Ten Haken Bruce Curran Alex Bielajew Feng Ming (Spring) Kong Benedick Fraass Grant Support: NIH P01-CA-59827 and R01 CA106770 AAPM-TG 105 co-authors Department of Radiation Oncology • University of Michigan Health Systems 9 Outline: “issues for consideration” • • Building blocks of a generic clinical MCTP system Verification issues in general • • • • CT calibration & artifacts Dose specification Absolute calibration, MU calculations Verification issues in detail – Validation of relative output – Experimental verification of MCMC-based dose algorithms: beam modifiers (MLC) Experimental verification of transport within the phantom/patient phantom/patient – • Retrospective lung treatment planning • Patients from a Phase I/II multicentre clinical trial (stage IIIA + B NSCLC, concomitant chemo) • Treatment planning: – plan I (PTV1 dose up to 40 Gy) – plan II (PTV2 boost to 60 Gy) – Typically 3-fields per plan, MLC shaped, wedged (CADplan, no heterogeneity corrections) MCMC-based treatment planning: comparisons of MC versus simple (correction(correction-based) and modelmodel-based algorithms Statistical Uncertainties in MCMC-based treatment planning MCTP and lung (NSCLC) Retrospective re-planning & outcome association Summary • • • • AAPM Summer School 2006 • Delivered dose simulated from delivered MU’s AAPM Summer School 2006 1 Patient 3 2 Patient 3 Patient 2 Patient 2 Gy Gy Gy Gy Gy PTV2 PTV2 Gy Gy EqTAR Monte Carlo Gy EqTAR Monte Carlo EqTAR underestimates the volume that receives a dose between 95% and 100% of the dose 120 100 30 80 PTV 20 Patient 3 - PTV2 60 10 40 Eqtar 20 xVMC 0 0 20 40 60 80 -10 0 -20 0 2000 4000 Dose (cGy) 6000 8000 -30 -40 % of planned PTV dose 100 120 Mean PTV Dose Maximum PTV Dose Patient EqTAR MC Diff EqTAR MC Diff 1 62.0 60.8 -1.9% 67.9 64.8 -4.6% 2 60.8 59.9 -1.5% 63.6 62.8 -1.3% 3 61.9 60.3 -2.5% 67.0 63.7 -5.0% 80 4 60.1 58.6 -2.5% 62.9 62.8 -0.2% 60 5 60.7 60.2 -0.8% 65.5 64.7 -1.2% 6 61.3 60.8 -0.8% 66.1 64.9 -1.8% 7 61.5 59.6 -2.2% 63.5 60.9 -3.1% 8 63.1 59.3 -6.0% 67.9 63.7 -6.2% 9 60.1 57.1 -5.1% 63.3 62.1 -1.9% 10 58.9 57.2 -2.9% 63.0 61.2 -2.8% 11 60.4 61.6 2.0% 65.0 65.3 0.4% 12 60.9 59.1 -2.6% 64.2 63.1 Average -2.2% 120 Patient 3 - co 100 EqTAR xVMC 40 20 0 0 1000 2000 3000 4000 Dose (cG 5000 6000 -1.7% -2.5% EqTAR underestimates volume receiving 10-20% of planned dose 120 Patient 3 contralateral lung 100 15 EqTAR xVMC 80 contralateral lung 10 5 60 0 0 40 -5 20 -10 20 40 60 80 100 120 -15 0 0 2000 4000 6000 Dose (cGy) 120 % of planned dose 8000 EqTAR overestimates volume receiving 0-10% of planned dose Ipsilateral lung Patient 3, ipsilateral lung 100 15 80 EqTAR xVMC 60 10 5 40 0 20 -5 0 20 40 60 80 -10 0 0 2000 4000 Dose (cGy) 6000 8000 -15 % of planned dose 100 120 100 80 EqTAR xVMC 60 40 20 0 0 2000 4000 6000 8000 Heart 4 2 0 -2 0 20 40 60 80 100 120 100* ( V_M C- V_EqTAR) / V_t (V_EqTAR - V_MC)/V_total *100 Patient 3, he 120 -4 -6 -8 -10 Dose (cG % of planned dose 100*(MC – EqTAR)/EqTAR Cord Tumour bearing lung Contralateral lung heart Patient 1 Patient 2 Patient 3 6.2% 6.3% 19.1% -4.2% -2.2% -7.5% 20.3% 15.9% 6.1% -4.5% -6.1% -7.7% Patient 4 -1.6% 2.2% 8.7% -2.2% Patient 5 Patient 6 Patient 7 Patient 8 Patient 9 Patient 10 Patient 11 Patient 12 5.3% 6.6% 2.6% 0.0% 1.7% 2.2% 7.4% 7.3% 1.9% 9.9% 0.3% 3.6% 4.8% 6.1% 4.6% -1.0% 15.7% 14.4% 2.8% 6.7% 11.1% 18.4% 8.2% 13.5% 35.8% 41.4% 17.3% 32.6% 4.6% 24.2% 5.8% 11.1% 1.5% (4.8%) 11.8% (5.4%) e-scatter 12.7% (17.3%) Average St.Dev. 5.3% (5.3%) Beam model How can an improved dose calculation algorithm be useful in relation to outcome? ¾ Revise known dose-response or dosecomplication relations ¾ Study associations of outcome with difference maps Beam model Mean lung dose (normalized using 2 Gy/ fraction scheme) Radiation Pneumonitis 25.000 A main complication for lung cancer radiation therapy Bio-model indicators for radiation pneumonitis: • Mean lung dose (MLD) • Vdose (V20 or V30) • NTCP 20.000 15.000 10.000 y = 1.3637x 0.9264 5.000 0.000 0.000 5.000 10.000 15.000 20.000 25.000 MLD (Gy, CadPlan) 25 20 need for large-scale replanning calculations! 15 10 y = 1.1421x 0.9606 5 0 0 5 10 15 MLD (Gy, EqTAR) 20 25 Gy How can an improved dose calculation algorithm be useful in relation to outcome? ¾ Revise known dose-response or dosecomplication relations ¾ Study associations of outcome with difference maps Monte Carlo Conventional Correlating late complications with dose Post-treatment complications (patient 1) Automatic fibrosis segmentation using: ¾ Automatic lung volume segmentation on planning and diagnostic CT images ¾ Calculated Pre/Post RT tissue density changes corresponding to Pre RT lung volume ¾ Mean tissue density changes corresponding to physician-identified radiographic fibrosis grades*: Grade 1 fibrosis: from 0.123 to 0.279 g/cc Grade 2 fibrosis: from 0.279 to 0.546 g/cc Grade 3 fibrosis: from 0.546 to 0.799 g/cc 9 9 9 +5 to 6 Gy I. Rosen, T. Fischer, J.A. Anatolak et al., Radiology, 221:614-622, 2001 Post-treatment CT (+ 1y) AAPM Summer School 2006 Dose-fibrosis correlation Fibrosis segmentation through image registration Post1 (RT+20d) Pre (RT-1d) DoseDose-response curves for the RTRT-induced fibrosis in the ipsilateral lung Post1 - Pre RT+82d RT+20d All grades fibrosis (ips) 100 CADPLAN 80 60 40 20 0 80 60 40 20 0 0 20 40 Dose [Gy] 60 80 Post3 - Pre 20 40 Dose [Gy] 60 80 60 40 20 0 0 RT+188d 40 Dose [Gy] 60 80 MC 80 100 CADPLAN 60 40 20 0 20 40 Dose [Gy] 20 Ipsilateral (RT) lung CADPLAN 80 0 20 40 100 MC Fibrosis probability [%] Fibrosis probability [%] Post3 (RT+243d) MC 60 All grades fibrosis (ips) 100 CADPLAN CADPLAN 80 0 0 All grades fibrosis (ips) Grade 3 fibrosis Grade 2 fibrosis Grade 1 fibrosis MC Fibrosis probability [%] Post2 - Pre All grades fibrosis (ips) 100 MC Fibrosis probability [%] Fibrosis probability [%] CADPLAN RT+133d All grades fibrosis (ips) 100 Post2 (RT+144d) 22 G. Stroian et al 2006 60 80 0 20 40 Dose [Gy] RT+237d 60 80 Percent volume [%] Dose difference map - planning CT MC 80 60 40 20 0 0 AAPM Summer School 2006 G. Stroian et al 2006 23 AAPM Summer School 2006 5 10 15 20 25 30 35 40 45 50 55 60 65 Dose [Gy] 24 Dose-fibrosis correlation (cont’d) Conclusions DoseDose-response curves for the RTRT-induced fibrosis in the contralateral lung 80 60 40 • beam model accuracy • heterogeneities 60 40 • MC planning presents specific clinical issues in addition to the issues one is already familiar with. • The new dose information can be used retrospectively in two ways: 20 20 0 0 20 40 Dose [Gy] 60 0 80 5 10 15 20 25 30 35 40 45 50 55 60 65 Dose [Gy] Contralateral (LT) lung All grades fibrosis (contr) 20 CADPLAN CADPLAN – revise dose - response relationships – to correlate to complications MC 100 MC 15 P e r c e nt v olum e [% ] Fibrosis probability [%] MC 80 Percent volume [%] Fibrosis probability [%] CADPLAN MC 0 Patient #2; RT+362d – The devil is in the details – Dosimetric differences are due to two components: 100 CADPLAN Patient #4; RT+82d • MC planning = raising the bar Contralateral (LT) lung All grades fibrosis (contr) 100 10 5 80 60 40 20 0 0 0 10 20 30 Dose [Gy] 40 AAPM Summer School 2006 50 60 0 5 10 15 20 25 30 35 40 45 50 55 60 65 Dose [Gy] G. Stroian et al 2006 25 AAPM Summer School 2006 26