UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Page 1 of 43 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the DCA 2000 Analyser in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University of Malta. 2. Objective To describe the procedure for the set-up, calibration, operation, quality control, maintenance and troubleshooting of the DCA 2000 Analyser. 3. Definitions 3.1. Bar Code: A machine readable set of data storing information regarding the item it is attached to. 3.2. Bar Code Track: The area where a reagent cartridge, optical test cartridge, calibration card or control card are placed prior to scanning their bar code. 3.3. Calibration Card: A card that is provided with each reagent kit and used to calibrate the device. The calibration card for the HbA1c reagent kit has one bar code available whilst the calibration card for the microalbumin / creatinine reagent kit has 2 bar codes, one on each side of the card, that both must be scanned to ensure a successful calibration. 3.4. Capillary: Used to obtain a small volume of sample using capillary action. 3.5. Capillary Holder: Holds the capillary in place during sample acquisition. 3.6. Cartridge Compartment: The area which holds the reagent cartridge during an analysis and the optical test cartridge during an optical test. 3.7. Cartridge Return Spring: Part of the mechanism that allows a cartridge to be removed from the cartridge compartment. 3.8. Cleaning Kit: Contains a number of sponge swabs that are used to clean the cartridge compartment. 3.9. Control Card: Contains a specially designed control bar code to be used in conjunction with a control solution. The control card for the HbA1c test is double sided; one side for Normal results (C1) and the other for Abnormal Page 2 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval results (C2). The control card for the microalbumin / creatinine test is also double sided; one side for Low results (C1) and the other for High results (C2). 3.10. Control Solution: Used to perform a quality control test, where it is analysed in the same manner as a patient sample. Two control solutions are available for both the HbA1c and the microalbumin / creatinine tests i.e. Normal / Abnormal and Low / High respectively. 3.11. Display: Indicates date, time, error messages, test results as well as procedural commands. Screen saver is activated when the device is not in use for more than 5 minutes and any key can be pressed to return it to normal display before performing any other function. 3.12. Dot: A mark used to help identify the location of the bar code reader window. 3.13. Filter Holder: Holds a replaceable air filter in place to prevent dust contamination. 3.14. Haemoglobin A1c (HbA1c) Test: A quantitative test used to measure the percentage concentration of glycosylated haemoglobin in the blood, providing an indicator of the mean daily blood glucose concentration over the preceding two months. 3.15. Main Menu: Consists of 9 sub-menus i.e. [RECALL PREVIOUS TESTS?], [SET SEQUENCE NUMBER?], [RECALL CONTROL RESULTS?], [VIEW CALIBRATION STATUS?], [SET DATE/TIME?], [INSTRUMENT SETUP?], [SET CREATININE UNITS?], [INSTRUMENT TEST?] and [RUN CONTROL?] (SOP/PD/221_02/A1 and SOP/PD/221_02/A2). 3.16. Microalbumin Creatinine Ratio: A test used in diabetics to determine the occurrence and degree of kidney disease, where abnormal values indicate an increased risk of developing heart and blood vessel complications. 3.17. Optical Test Cartridge: A cartridge that is used to allow the user to verify the performance of the optical system of the device and to check that this was not damaged or contaminated during the cleaning of the cartridge compartment. 3.18. Program Card: Contains all the necessary data to operate the device. 3.19. Program Card Connector: The area on the device where the program card is inserted. Page 3 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 3.20. Quality Control Test: A test used to ensure that the device is giving appropriate results within acceptable limits. 3.21. Reagent Cartridge: Holds the sample during the analysis. Different reagent cartridges are available for each test i.e. HbA1c and microalbumin / creatinine. 3.22. Venipuncture: The process of puncturing a vein by a sharp rigid stylet, cannula or by a needle attached to a syringe or catheter for withdrawing a sample of blood, instilling a medication, starting an IV infusion or for injecting a radiopaque substance for radiologic examination of a part or system of the body. 4. Responsibility 4.1. The members of the Department of Pharmacy (staff and students) are responsible for following this SOP. 4.2. The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed. 5. Procedure 5.1. Diagram of DCA 2000 Analyser Keys Display Bar Code Track Dot Cartridge Compartment Air Filter Holder Page 4 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 5.2. Valid for: 2 years from approval Set-Up 5.2.1. 5.2.2. 5.2.3. Place the device on a firm, level surface. Check that the power switch is set to the Off (O) position. Insert the program card into the program card connector with its label facing upwards. 5.2.4. Connect the power cord to the power cord connector socket. 5.2.5. Plug in the power cord to an electricity outlet. 5.2.6. Set the power switch to the On (I) position. 5.2.7. Wait for a few minutes until the device warms up. 5.2.8. Press the [MENU] button to access the Main Menu. 5.2.9. Repeatedly press the [NEXT] button to scroll through the different submenus available. 5.2.10. Press the [ENTER] button to select the desired sub-menu. 5.2.11. See SOP/PD/221_02/A1 to set-up each sub-menu. 5.2.12. Press the [ESCAPE] button to exit the Main Menu. 5.3. Calibration 5.3.1. Calibrating device for HbA1c 5.3.1.1. 5.3.1.2. 5.3.1.3. 5.3.1.4. 5.3.1.5. 5.3.1.6. 5.3.1.7. 5.3.1.8. 5.3.1.9. 5.3.1.10. 5.3.1.11. 5.3.1.12. 5.3.1.13. 5.3.2. Set the power switch to the On (I) position. Wait for a few minutes until the device warms up. Check that no test is in progress before attempting to perform a calibration. Locate the dot on the device next to the bar code track. Locate the bar code on the calibration card. Hold the calibration card in a way that the bar code faces right. Insert the calibration card into the bar code track above the dot. Hold the calibration card against the right side of the bar code track. Quickly slide the calibration card down past the dot. Check that a beep is heard to ensure a successful scan. Repeat 5.3.1.4 to 5.3.1.9 if no beep is heard. Contact the supplier if a beep repeatedly fails to be heard. Set the power switch to the Off (O) position. Calibrating device for Microalbumin / Creatinine 5.3.2.1. Repeat 5.3.1.1 to 5.3.1.12. Page 5 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 5.3.2.2. 5.3.2.3. 5.4. Valid for: 2 years from approval Repeat 5.3.1.4 to 5.3.1.12 within less than 5 minutes, using the other side of the calibration card. Set the power switch to the Off (O) position. Operation 5.4.1. Performing a Haemoglobin A1c test 5.4.1.1. 5.4.1.2. 5.4.1.3. 5.4.1.4. 5.4.1.5. 5.4.1.6. 5.4.1.7. 5.4.1.8. 5.4.1.9. 5.4.1.10. 5.4.1.11. 5.4.1.12. 5.4.1.13. 5.4.1.14. 5.4.1.15. 5.4.1.16. 5.4.1.17. 5.4.1.18. 5.4.1.19. 5.4.1.20. Set the power switch to the On (I) position. Wait for a few minutes until the device warms up. Take out a reagent cartridge pack from the refrigerator. Remove the reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature. Check that the capillary holder to be used has an absorbent pad, a glass capillary and a properly working latching mechanism. Hold the capillary holder at an angle. Touch the tip of the capillary to a small drop of blood on the finger, if the blood sample is going to be obtained from a finger prick. Touch the tip of the capillary to a blood sample present on the lid of a blood collection tube after mixing the blood sample well, if the blood sample is going to be obtained from a venipuncture sample. Wipe the outside surface of the capillary with a lint-free tissue. Discard the capillary holder together with the capillary in a biohazard container and start the test again if bubbles are present in the capillary. Position the capillary holder in the correct orientation for insertion into the reagent cartridge. Gently insert the capillary holder into the reagent cartridge until the capillary holder gently snaps into place. Locate the dot next to the bar code track. Locate the bar code on the reagent cartridge. Hold the reagent cartridge in a way that the bar code faces right. Insert the reagent cartridge into the bar code track above the dot. Quickly slide the reagent cartridge down past the dot. Check that a beep is heard to ensure a successful scan. Repeat steps 5.4.1.13 to 5.4.1.17 if no beep is heard. Contact the supplier if a beep repeatedly fails to be heard. Page 6 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 5.4.1.21. Wait for [LOAD CARTRIDGE] to be displayed. 5.4.1.22. Open the cartridge compartment door. 5.4.1.23. Hold the reagent cartridge in a way that the bar code faces right. 5.4.1.24. Insert the reagent cartridge into the cartridge compartment. 5.4.1.25. Check that the cartridge has snapped into place. 5.4.1.26. Pull the flexible plastic pull-tab completely out of the reagent cartridge. 5.4.1.27. Close the cartridge compartment door. 5.4.1.28. Wait for 5 seconds for a beep to be heard to indicate that the test has started. 5.4.1.29. Press the [ESCAPE] button and then the [ENTER] button if the test needs to be cancelled at any time. 5.4.1.30. Wait for the test to be performed. 5.4.1.31. Use the [MENU] button (SOP/PD/221_02/A1) to recall the test result if more than 15 minutes has elapsed from when the test was finished. 5.4.1.32. Record the displayed result. 5.4.1.33. Open the cartridge compartment door. 5.4.1.34. Push and hold down, with the right hand, the button situated on the right side of the cartridge compartment. 5.4.1.35. Gently push, with the left hand, the plastic tab on the reagent cartridge to the right to release the reagent cartridge. 5.4.1.36. Remove the reagent cartridge from the cartridge compartment. 5.4.1.37. Discard the reagent cartridge in a biohazard container. 5.4.1.38. Set the power switch to the Off (O) position. 5.4.2. Performing a Microalbumin / Creatinine Test 5.4.2.1. 5.4.2.2. 5.4.2.3. 5.4.2.4. 5.4.2.5. 5.4.2.6. 5.4.2.7. Repeat 5.4.1.1 to 5.4.1.4. Check that the capillary holder to be used has an absorbent pad, a glass capillary, a properly working latching mechanism and a starch plug that is not situated at the bottom of the capillary tube. Immerse the tip of the capillary tube in the urine sample to a level just above the capillary starch plug. Allow enough time for the urine sample to flow up the capillary tube and come into contact with the starch plug. Remove the capillary tube from the urine sample. Re-immerse the capillary tube in the urine sample, if the urine flows back down the capillary tube. Repeat 5.4.1.9 to 5.4.1.20. Page 7 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 5.4.2.8. 5.4.2.9. 5.4.2.10. 5.4.2.11. 5.4.2.12. 5.4.2.13. 5.4.2.14. 5.4.2.15. 5.4.2.16. 5.5. Valid for: 2 years from approval Wait for [LOAD CARTRIDGE, INSERT PLUNGER] to be displayed. Repeat 5.4.1.22 to 5.4.1.25. Insert the plunger into the hole found in the top of the capillary holder. Fully depress the plunger into the capillary holder to lock it in place. Press the [ENTER] button. Repeat 5.4.1.26 to 5.4.1.30. Use the [MENU] button (SOP/PD/221_02/A1) to recall the test result if more than 15 minutes has elapsed from when the test was finished. Record the displayed results for albumin, creatinine and the albumin / creatinine ratio. Repeat 5.4.1.33 to 5.4.1.38. Quality Control 5.5.1. Analysing control for Haemoglobin A1c 5.5.1.1. 5.5.1.2. 5.5.1.3. 5.5.1.4. 5.5.1.5. 5.5.1.6. 5.5.1.7. 5.5.1.8. 5.5.1.9. 5.5.1.10. 5.5.1.11. 5.5.1.12. 5.5.1.13. 5.5.1.14. 5.5.1.15. 5.5.1.16. Set the power switch to the On (I) position. Wait for a few minutes until the device warms up. Remove the 2 HbA1c control vials and a reagent cartridge from the refrigerator. Reconstitute the lyophilised controls using the solvent provided. Remove the reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature. Check that the capillary holder to be used has an absorbent pad, a glass capillary and a properly working latching mechanism. Locate the dot next to the bar code track on the device. Choose the correct side of the control card according to the control solution to be used first. Locate the bar code on the control card. Hold the control card in a way that the bar code faces right. Insert and hold gently the control card against the right side of the bar code track above the dot. Quickly slide the control card down past the dot. Check that a beep is heard to ensure a successful scan. Repeat steps 5.5.1.7 to 5.5.1.12 if no beep is heard. Contact the supplier if a beep repeatedly fails to be heard. Press the [ENTER] button. Page 8 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 5.5.1.17. Hold the capillary holder at angle. 5.5.1.18. Mix well the contents of the control solution. 5.5.1.19. Touch the tip of the capillary to the control solution present on the lid of the control solution tube. 5.5.1.20. Repeat 5.4.1.9 to 5.4.1.20 5.5.1.21. Wait for [LOAD CONTROL] to be displayed. 5.5.1.22. Repeat 5.4.1.22 to 5.4.1.27. 5.5.1.23. Press the [ESCAPE] button and then the [ENTER] button if the test needs to be cancelled at any time. 5.5.1.24. Wait for the test to be performed. 5.5.1.25. Press the [ESCAPE] button to display the value of an out-ofrange control if [CONTROL OUT OF RANGE] is displayed. 5.5.1.26. Use the [MENU] button (SOP/PD/221_02/A1) to recall the test result if more than 15 minutes has elapsed from when the test was finished. 5.5.1.27. Record the displayed result and compare it with the result range present in the package insert of the control solution. 5.5.1.28. Repeat 5.4.1.33 to 5.4.1.37. 5.5.1.29. Repeat 5.5.1.5 to 5.5.1.28 using the second control solution. 5.5.1.30. Set the power switch to the Off (O) position. 5.5.2. Analysing control for Microalbumin / Creatinine 5.5.2.1. 5.5.2.2. 5.5.2.3. 5.5.2.4. 5.5.2.5. 5.5.2.6. 5.5.2.7. 5.5.2.8. 5.5.2.9. 5.5.2.10. 5.5.2.11. 5.5.2.12. 5.5.2.13. Set the power switch to the On (I) position. Wait for a few minutes until the device warms up. Remove the 2 Microalbumin / Creatinine control vials and a reagent cartridge from the refrigerator. Reconstitute the lyophilised controls using the solvent provided. Remove the reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature. Check that the capillary holder to be used has an absorbent pad, a glass capillary, a properly working latching mechanism and a starch plug that is not situated at the bottom of the capillary tube. Repeat 5.5.1.7 to 5.5.1.19. Repeat 5.4.1.9. to 5.4.1.20. Wait for [LOAD CARTRIDGE, INSERT PLUNGER] to be displayed. Repeat 5.4.1.22 to 5.4.1.25. Repeat 5.4.2.10 to 5.4.2.12. Repeat 5.4.1.26 to 5.4.1.27. Repeat 5.5.1.23 to 5.5.1.26. Page 9 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 5.5.2.14. Repeat 5.4.1.33 to 5.4.1.37. 5.5.2.15. Repeat 5.5.2.5 to 5.5.2.14 using the second control solution. 5.5.2.16. Set the power switch to the Off (O) position. 5.6. Maintenance 5.6.1. 5.6.2. 5.6.3. Check that the device is switched Off before cleaning its exterior surface. Clean the outside of the device with a moist lint-free cloth. If blood is present on the exterior surface of the device: 5.6.3.1. Expose the exterior surface to either a 0.5% sodium hypochlorite or a 2% glutaraldehyde solution for 10 minutes. 5.6.4. Allow the device to completely dry before attempting to switch it On. 5.6.5. Clean the cartridge compartment on a quarterly basis or depending on the duration of use of the device: 5.6.5.1. 5.6.5.2. 5.6.5.3. 5.6.5.4. 5.6.5.5. 5.6.5.6. 5.6.5.7. 5.6.5.8. 5.6.5.9. 5.6.5.10. 5.6.5.11. 5.6.5.12. 5.6.5.13. 5.6.5.14. Check that the device is switched Off. Open the cartridge compartment door as far as possible. Clean the inside surface with a moist lint-free cloth. Dry the inside surface with a dry lint-free cloth. Locate the cartridge return spring containing two holes at its periphery. Insert the tip of a straightened paper clip into the top hole of the spring. Gently pull this end of the spring towards the centre of the cartridge compartment to release one side of the spring. Insert the tip of a straightened paper clip into the bottom hole of the spring. Gently pull this end of the spring towards the centre of the cartridge compartment to release the other side of the spring. Pull the cartridge return spring completely out of the cartridge compartment. Clean the cartridge return spring with a moist lint-free cloth. Dry the cartridge return spring with a dry lint-free cloth. Remove any spilled liquid from the cartridge holder using a clean, dry sponge swab provided in the Cleaning Kit. Rotate the cartridge holder while the cartridge compartment door is partially closed to locate and remove any additional liquid that might still be present. Page 10 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 5.6.5.15. Clean the cartridge holder with a moist sponge swab and rotate the holder as necessary. 5.6.5.16. Locate the vertical grooves and the front and back slots inside the cartridge compartment. 5.6.5.17. Locate the leaf spring on one side of the cartridge return spring. 5.6.5.18. Orient the leaf spring towards the back of the device. 5.6.5.19. Hold the cartridge return spring from both sides near the holes. 5.6.5.20. Pinch both sides of the cartridge return spring and lower it into the device by sliding the sides of the spring between the vertical grooves in the compartment. 5.6.5.21. Check that the cartridge return spring can freely slide up and down within the cartridge compartment. 5.6.5.22. Gently push down one edge of the cartridge return spring and insert it into the slot. 5.6.5.23. Repeat 5.6.5.22 to attach the opposite side of the cartridge return spring to the cartridge compartment. 5.6.5.24. Check that the leaf spring of the cartridge return spring is oriented towards the back of the device. 5.6.5.25. Check that the cartridge return spring is securely attached via the grooves to the cartridge compartment. 5.6.6. Run an optical test after cleaning the cartridge compartment and whenever the performance of the optical system needs to be verified: 5.6.6.1. 5.6.6.2. Locate the bar code on the optical test cartridge. Hold the optical test cartridge in a way that the bar code faces right. 5.6.6.3. Insert the optical test cartridge into the bar code track above the dot on the device. 5.6.6.4. Quickly slide the cartridge down past the dot. 5.6.6.5. Check that a beep is heard to ensure a successful scan. 5.6.6.6. Repeat steps 5.6.5.1 to 5.6.5.4 if no beep is heard. 5.6.6.7. Contact the supplier if a beep repeatedly fails to be heard. 5.6.6.8. Press the [ENTER] button when [RUN STANDARD 1] is displayed. 5.6.6.9. Open the cartridge compartment door. 5.6.6.10. Hold the optical test cartridge in a way that the bar code faces right. 5.6.6.11. Insert the optical test cartridge into the cartridge compartment. 5.6.6.12. Check that the optical test cartridge has snapped into place. Page 11 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5.6.6.13. Close the cartridge compartment door. 5.6.6.14. Wait for 6 minutes for the test to be performed. 5.6.6.15. Record the displayed results for transmittance, standard deviation and drift. 5.6.6.16. Open the cartridge compartment door. 5.6.6.17. Push and hold down with right hand the button situated on the right side of the cartridge compartment. 5.6.6.18. Gently push the plastic tab on the cartridge to the right with the left hand to release the optical test cartridge. 5.6.6.19. Pull the optical test cartridge out of the cartridge compartment. 5.6.6.20. Compare the results obtained with any previously obtained results. Results must be within the limits mentioned in 6.8 to 6.10. 5.6.7. Change the air filter on a quarterly basis or depending on the duration of use of the device: 5.6.7.1. 5.6.7.2. 5.6.7.3. 5.6.7.4. 5.6.7.5. 5.6.7.6. 5.7. Completely rotate the filter holder anticlockwise. Pull off the filter holder. Discard the old air filter. Place a new air filter into the filter holder. Place the filter holder back onto the device. Completely rotate the filter holder clockwise until it stops turning. Troubleshooting Error Message/s HbA1c result preceded by [<] or [>] Cause/s Concentration is below the lower limit or above the upper limit Albumin and/or Creatinine result preceded by [<] or [>] [CANCELLED: DISCARD TEST] displayed Albumin and/or Creatinine result is under or over range [DOOR OPEN ERROR Cartridge compartment door was opened during a test or [ESCAPE] button was pressed Cartridge compartment door Page 12 of 43 Solution/s Report result as being less than 2.5% HbA1c - Report result as being more than 14.0% HbA1c N/A - Remove cartridge and repeat test with a new reagent cartridge Close door and wait for UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER CLOSE DOOR] displayed [OUT OF DATE CONTROL] displayed [OUT OF DATE REAGENT] displayed [SCAN CALIBRATION CARD] displayed was opened while testing was in progress Bar code for an expired control has been read Bar code for an expired reagent cartridge has been read Calibration card for reagent cartridge in use was not scanned [PLEASE WAIT] to be displayed Discard expired control Discard expired reagent cartridge [E-1] displayed [E-2] displayed Internal error has occurred Error detected in the device’s non-volatile memory [E-3] displayed - Wrong program card inserted Check that correct calibration card is scanned (check name and lot number of cartridge in use) Contact supplier Switch power switch to the Off [O] position and contact supplier - Insert correct program card - No program card present - Insert correct program card - Program card contacts require cleaning or are defective - - Wrong program card inserted - Clean contacts with a pencil eraser and wipe any left-over particles with a lint-free tissue. If problem persists, contact supplier. Insert correct program card - Program card is not completely inserted - - Program card contacts require cleaning or are defective - [E-4] displayed [E-5] displayed Defective program card Page 13 of 43 Correctly insert program card Clean contacts with a pencil eraser and wipe any left-over particles with a lint-free tissue. If problem persists, contact supplier Use a replacement program card. If problem persists, contact supplier. UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER [E-6] displayed Error in the system timing [101] or [301] displayed - No cartridge is present - Optical alignment problem Condensation on cartridge - - Cartridge not inserted correctly - Check that cartridge is inserted correctly - Buffer tab pulled off prematurely or not pulled at all - Check that buffer tab is pulled off after inserting cartridge into holder - Cartridge defect in optical window or cartridge optical window is blocked or dirty Condensation on cartridge - If problem persists, contact supplier - Allow at least 10 minutes after cartridge is removed from refrigerator before starting test - Particulate contamination - - Capillary was not adequately filled or air bubble was present - If problem persists, contact supplier Check that capillary has no air bubbles - Blood dried in capillary - [102] or [302] displayed [103] or [303] displayed [104] displayed - - Page 14 of 43 Switch power switch to the Off [O] position and contact supplier - Repeat test with cartridge - If problem persists, contact supplier Allow at least 10 minutes after cartridge is removed from refrigerator before starting test No more than 5 minutes should elapse between filling capillary and starting UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER [106] displayed test Repeat test with a capillary holder in place - No capillary holder inserted - - Controls were improperly reconstituted - Check that the controls were properly reconstituted - Haemoglobin < 7g/dL - Perform test by another method - Buffer tab not pulled or buffer not released from tray Excess blood on capillary holder - Pull buffer tab just before closing door. [105] displayed - [107] displayed - Blood not lysing either due to an irregularity in patient’s red blood cells or because cartridge was not given enough time to warm up - Haemoglobin > 24g/dL (patient has abnormally high Hb or very high triglycerides) Particulate contamination [108], [304], [306] or [309] displayed [109], [110], [111] or [112] displayed - Cartridge exposed to excessive temperature and/or humidity [113], [114], [116] or [313] displayed - Cartridge exposed to excessive temperature and/or humidity Page 15 of 43 Check that capillary is wiped before inserting into cartridge - Freeze/thaw sample before use if patient is found to have such an irregularity or else allow at least 10 minutes after cartridge is removed from refrigerator before starting test - Perform test by another method If problem persists, contact supplier - Check that reagent kits are being properly stored - Open foil package just prior to using cartridge - Check that reagent kits are being properly stored UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER - [115] displayed [305] displayed [307], [308], [310] or [311] displayed [315] displayed [316] displayed Particulate contamination Open foil package just prior to using cartridge - If problem persists, contact supplier Blood was left too long in No more than 5 minutes capillary should elapse between filling capillary and starting test Urine sample turbid or highly - Centrifuge sample pigmented before test if sample is very turbid or cloudy - Obtain a fresh sample if sample contains visible amounts of blood or is highly pigmented Cartridge exposed to - Check that reagent kits excessive temperature and/or are being properly humidity stored - Open foil package just prior to using cartridge Device has not been adjusted to run a Microalbumin / Creatinine test - Normalisation cartridge has not been properly inserted Contact supplier - Check that cartridge has been inserted completely - - If problem persists, contact supplier Normalisation cartridge is dirty or defective Page 16 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 5.8. Flow charts 5.8.1. Set-Up Start Place device on a firm, level surface Check that power switch is set to the Off (O) position Insert program card into program card connector with its label facing upwards Connect power cord to power cord connector socket Plug in power cord to an electricity outlet Set power switch to On (I) position Wait for a few minutes until device warms up Press [MENU] button to access Main Menu Repeatedly press [NEXT] button to scroll through the different sub-menus available Press [ENTER] button to Select desired sub-menu See SOP/PD/221_02/A1 to set-up each sub-menu Press [ESCAPE] button to exit Main Menu End Page 17 of 43 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5.8.2. Calibration – Calibrating device for HbA1c Start Set power switch to On (I) position Wait for a few minutes until device warms up Check that no test is in progress before attempting to perform a calibration Locate dot on device next to bar code track Locate bar code on calibration card Hold calibration card in a way that bar code faces right Insert calibration card into bar code track above dot Hold calibration card against right side of bar code track Quickly slide calibration card past the dot No Beep is heard No Beep repeatedly fails to be heard Repeat procedure Yes Yes 1 Contact supplier Scan was successful End Page 18 of 43 1 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5.8.3. Calibration – Calibrating device for Microalbumin / Creatinine Start Set power switch to On (I) position Wait for a few minutes until device warms up Check that no test is in progress before attempting to perform a calibration Locate dot on device next To bar code track Locate bar code on calibration card Hold calibration card in a way that bar code faces right Insert calibration card into bar code track above dot Hold calibration card against right side of bar code track Quickly slide calibration card past the dot No Beep is heard No Beep repeatedly fails to be heard Repeat procedure Yes Yes 1 Contact supplier Scan was successful Repeat procedure using the other side of calibration card End Page 19 of 43 1 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5.8.4. Operation – Performing a Haemoglobin A1c test Start Set power switch to On (I) position Wait for a few minutes until device warms up Take out reagent cartridge pack from refrigerator Remove reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature Check that capillary holder has an absorbent pad, a glass capillary and a properly working latching mechanism Hold capillary holder at an angle No Blood sample obtained from finger prick Mix blood sample well Yes Touch tip of capillary to blood sample present on lid of blood collection tube Touch tip of capillary to a small drop of blood on the finger Wipe outside surface of capillary with lint-free cloth No Bubbles present in capillary Position capillary holder in correct orientation for insertion into reagent cartridge Discard capillary holder together with capillary in a biohazard container Gently insert capillary holder into reagent cartridge until capillary holder gently snaps into place Yes 1 Page 20 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 1 Locate dot next to bar code track Locate bar code on reagent cartridge Hold reagent cartridge in a way that bar code faces right Insert reagent cartridge into bar code track above dot Quickly slide reagent cartridge down past the dot No Beep is heard No Beep repeatedly fails to be heard Repeat procedure Yes Yes 3 Contact supplier Scan was successful 2 Wait for [LOAD CARTRIDGE] to be displayed Open cartridge compartment door Hold reagent cartridge in a way that bar code faces right Insert reagent cartridge into cartridge compartment Check that reagent cartridge has snapped into place Pull flexible plastic pull-tab completely out of reagent cartridge 4 Page 21 of 43 3 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 4 Close cartridge compartment door Wait for 5 seconds for a beep to be heard to indicate that test has started No Wait for test to be performed Need to cancel test Yes Press [ESCAPE] button More than 15 minutes has elapsed from when test was finished Press [ENTER] button Yes Use [MENU] button (SOP/PD/221_02/A1) to recall test result Open cartridge compartment door Push and hold down with right hand the button situated on right side of cartridge compartment Gently push down with left hand the plastic tab on reagent cartridge to the right to release reagent cartridge Remove reagent cartridge from cartridge compartment Discard reagent cartridge in biohazard container Set power switch to Off (O) position 2 End Page 22 of 43 No Record displayed test result UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5.8.5. Operation – Performing a Microalbumin / Creatinine test Start Set power switch to On (I) position Wait for a few minutes until device warms up Take out reagent cartridge pack from refrigerator Remove reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature Check that capillary holder has an absorbent pad, a glass capillary, a properly working latching mechanism and a starch plug that is not situated at the bottom of capillary tube Immerse tip of capillary tube in urine sample to a level just above the capillary starch plug Allow enough time for urine sample to flow up the capillary tube and come into contact with starch plug Remove capillary tube from urine sample Urine flowed back down capillary tube No Wipe outside surface of capillary with lint-free cloth Yes No Bubbles present in capillary Position capillary holder in correct orientation for insertion into reagent cartridge Discard capillary holder together with capillary in a biohazard container Gently insert capillary holder into reagent cartridge until capillary holder gently snaps into place Re-immerse capillary tube in urine sample Yes Page 23 of 43 1 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 1 Locate dot next to bar code track Locate bar code on reagent cartridge Hold reagent cartridge in a way that bar code faces right Insert reagent cartridge into bar code track above dot Quickly slide reagent cartridge down past the dot No Beep is heard No Beep repeatedly fails to be heard Repeat procedure Yes Yes 3 Contact supplier Scan was successful 2 Wait for [LOAD CARTRDIGE, INSERT PLUNGER] to be displayed Open cartridge compartment door Hold reagent cartridge in a way that bar code faces right Insert reagent cartridge into cartridge compartment Check that cartridge has snapped into place Insert plunger into hole found in the top of the capillary holder Fully depress plunger into capillary holder to lock it in place 4 Page 24 of 43 3 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 4 Press [ENTER] button Pull flexible plastic pull-tab completely out of reagent cartridge Close cartridge compartment door Wait for 5 seconds for a beep to be heard to indicate that test has started No Need to cancel test Wait for test to be performed Yes Press [ESCAPE] button More than 15 minutes has elapsed from when test was finished Press [ENTER] button Yes Use [MENU] button (SOP/PD/221_02/A1) to recall test result Open cartridge compartment door Push and hold down with right hand the button situated on right side of cartridge compartment Gently push down with left hand the plastic tab on reagent cartridge to the right to release reagent cartridge Remove reagent cartridge from cartridge compartment Discard reagent cartridge in biohazard container 2 Set power switch to Off (O) position End Page 25 of 43 No Record displayed test result for albumin, creatinine and albumin/creatinine ratio UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 5.8.6. Valid for: 2 years from approval Quality Control – Analysing control for Haemoglobin A1c Start Set power switch to On (I) position Wait for a few minutes until device warms up Remove 2 HbA1c control vials and a reagent cartridge from refrigerator Reconstitute lyophilised controls using solvent provided Remove reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature 4 Check that capillary holder has an absorbent pad, a glass capillary and a properly working latching mechanism Locate dot next to bar code track on device Choose correct side of control card according to control solution to be used first Locate bar code on control card Hold control card in a way that bar code faces right Insert and hold gently the control card against right side of bar code track above the dot 1 Page 26 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 1 Quickly slide control card down past the dot No Beep is heard No Beep repeatedly fails to be heard Repeat procedure Yes Yes 3 Contact supplier Scan was successful 2 Press [ENTER] button Hold capillary holder at an angle Mix well contents of control solution Touch tip of capillary to control solution Wipe outside surface of capillary with lint-free cloth Bubbles present in capillary No Position capillary holder in correct orientation for insertion into reagent cartridge Yes Discard capillary holder together with capillary in a biohazard container Gently insert capillary holder into reagent cartridge until capillary holder gently snaps into place 4 Locate dot next to bar code track Locate bar code on reagent cartridge Hold reagent cartridge in a way that bar code faces right 5 Page 27 of 43 3 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5 Insert reagent cartridge into bar code track above dot Quickly slide reagent cartridge down past the dot No Beep is heard Repeat procedure Scan was successful Beep repeatedly fails to be heard Yes 6 No 6 Yes Wait for [LOAD CONTROL] to be displayed Contact supplier Close cartridge compartment door 2 Hold reagent cartridge in a way that bar code faces right Insert reagent cartridge into cartridge compartment Check that cartridge has snapped into place Pull flexible plastic pull-tab completely out of reagent cartridge Close cartridge compartment door No Wait for test to be performed Need to cancel test Yes No Press [ESCAPE] button [CONTROL OUT OF RANGE] is displayed Press [ENTER] button Yes 7 8 Page 28 of 43 9 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 9 8 No More than 15 minutes has elapsed from when test was finished Press [ESCAPE] button to display out-of-range control Record displayed result and compare it with result range present in package insert of control solution Yes Use [MENU] button (SOP/PD/221_02/A1) to recall test result 7 Open cartridge compartment door Push and hold down with right hand the button situated on right side of cartridge compartment Gently push down with left hand the plastic tab on reagent cartridge to the right to release reagent cartridge Remove reagent cartridge from cartridge compartment Discard reagent cartridge in biohazard container Repeat whole procedure using the second control solution Set power switch to Off (O) position 2 End Page 29 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 5.8.7. Valid for: 2 years from approval Quality Control – Analysing control for Microalbumin / Creatinine Start Set power switch to On (I) position Wait for a few minutes until device warms up Remove the 2 Microalbumin / Creatinine control vials and a reagent cartridge from refrigerator Reconstitute lyophilised controls using solvent provided Remove reagent cartridge from its foil package and let it stand for about 5-10 minutes at ambient temperature 4 Check that capillary holder to be used has an absorbent pad, a glass capillary, a properly working latching mechanism and a starch plug that is not situated at the bottom of capillary tube Locate dot next to bar code track Choose correct side of control card according to control solution to be used first Locate bar code on control card Hold control card in a way that bar code faces right Insert and hold gently control card against right side of bar code track above dot 1 Page 30 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 1 Quickly slide control card down past dot No Beep is heard Repeat procedure Scan was successful Beep repeatedly fails to be heard Yes 3 No Yes Press [ENTER] button Contact supplier Hold capillary holder at an angle 2 Mix well contents of control solution Touch tip of capillary to control solution Wipe outside surface of capillary with a lint-free tissue No Bubbles present in capillary Position capillary holder in correct orientation for insertion into reagent cartridge Discard capillary holder together with capillary Gently insert capillary holder into reagent cartridge until capillary holder gently snaps into place Yes 4 Locate dot next to bar code track Locate bar code on reagent cartridge Hold reagent cartridge in a way that bar code faces right 5 Page 31 of 43 3 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5 Insert reagent cartridge into bar code track above dot Quickly slide reagent cartridge past the dot No No Beep repeatedly fails to be heard Beep is heard Yes Yes 6 Contact supplier Scan was successful 2 Wait for [LOAD CARTRIDGE, INSERT PLUNGER] to be displayed Open cartridge compartment door Hold reagent cartridge in a way that bar code faces right Insert reagent cartridge into cartridge compartment Check that cartridge has snapped into place Insert plunger into hole found in the top of capillary holder Fully depress plunger into capillary holder to lock it in place Press [ENTER] button Pull flexible pull-tab completely out of reagent cartridge Close cartridge compartment door 7 Page 32 of 43 6 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 7 No Test needs to be cancelled Wait for test to be performed Press [ESCAPE] button [CONTROL OUT OF RANGE] displayed Yes No More than 15 minutes has elapsed from when test was finished Yes Press [ENTER] button Yes Press [ESCAPE] button to display result Use [MENU] button (SOP/PD/221_02/A1) to recall result Record displayed result and compare it with result range present in package insert of control solution Open cartridge compartment door Push and hold down with right hand the button on right side of cartridge compartment Gently push with left hand the plastic tab on cartridge to the right to release reagent cartridge Remove reagent cartridge from cartridge compartment Discard reagent cartridge in a biohazard container Repeat whole procedure using the second control solution Set power switch to Off (O) position 2 End Page 33 of 43 No UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 5.8.8. Maintenance 1 Start No Need to clean outside surface Need to clean cartridge compartment Yes Yes Check that device is switched Off Check that device is switched Off Clean with a moist lint-free cloth Open cartridge compartment door as far as possible No Blood present on exterior surface Allow device to completely dry before attempting to switch it On Clean inside surface with a moist lint-free cloth Yes 1 Expose exterior surface to either 0.5% sodium hypochlorite or 2% glutaraldehyde solution for 10 minutes Dry inside surface with a dry lint-free cloth Locate cartridge return spring containing two holes at its periphery Insert tip of straightened paper clip into top hole of spring Gently pull this end of spring towards centre of cartridge compartment to release one side of the spring Insert tip of straightened paper clip into bottom hole of spring Gently pull this end of spring towards centre of cartridge compartment to release other side of the spring Pull cartridge return spring completely out of cartridge compartment Page 34 of 43 3 No 2 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 3 Clean cartridge return spring with a moist lint-free cloth Dry cartridge return spring with a dry lint-free cloth Remove any spilled liquid from cartridge holder using a clean, dry sponge swab provided in Cleaning Kit Rotate cartridge holder while cartridge compartment door is partially closed to locate and remove any additional liquid that might still be present Clean cartridge compartment holder with a moist sponge swab and rotate holder as necessary Locate vertical grooves and front and back slots inside cartridge compartment Locate leaf spring on one side of cartridge return spring Orient leaf spring towards back of device Hold cartridge return spring from both sides near the holes Pinch both sides of cartridge return spring and lower it into device by sliding the sides of the spring between the vertical grooves in compartment 4 Page 35 of 43 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval DCA 2000® ANALYSER 4 Check that cartridge return spring can freely slide up and down within cartridge compartment Gently push down one edge of cartridge return spring and insert it into slot Gently push down opposite edge of cartridge return spring into other slot 2 No Check that leaf spring of cartridge return spring is oriented towards back of device Need to perform an optical test 5 Yes Check that cartridge return spring is securely attached via the grooves of cartridge compartment Locate bar code on optical test cartridge Hold optical test cartridge in a way that bar code faces right Insert optical test cartridge into bar code track above dot of device Quickly slide cartridge down past dot No Beep is heard Repeat procedure Scan was successful Beep repeatedly fails to be heard Yes No Yes Contact supplier 6 Page 36 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 6 Press [ENTER] button when [RUN STANDARD 1] is displayed Open cartridge compartment door Hold optical test cartridge in a way that bar code faces right Insert optical test cartridge into cartridge compartment Check that optical test cartridge has snapped into place Close cartridge compartment door Wait for 6 minutes for test to be performed Record displayed result for transmittance, standard deviation and drift Open cartridge compartment door Push and hold down with right hand the button situated on right side of cartridge compartment Gently push plastic tab on cartridge to the right with left hand to release optical test cartridge Pull optical test cartridge out of cartridge compartment 7 Page 37 of 43 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER 7 Compare results with any previously obtained results. Results must be within limits mentioned in precautions section No Need to change air filter 5 Yes Completely rotate filter holder anticlockwise Pull off filter holder Discard old filter Place new air filter into filter holder Place filter holder back onto device Completely rotate filter holder clockwise until it stops turning End Page 38 of 43 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 6. Precautions 6.1. 6.2. 6.3. 6.4. 6.5. 6.6. 6.7. 6.8. 6.9. 6.10. 6.11. 6.12. 6.13. 6.14. Do not expose the device to extremes of temperature and humidity, direct sunlight or air current. Do not block the ventilation panels found on the back and right side of the device. When performing an HbA1c test, the device should be at an ambient temperature of 15-32oC. When performing a Microalbumin / Creatinine test, the device should be at an ambient temperature of 18-30oC. Do not use normal cotton swabs to clean the cartridge holder since any cotton fibres left can interfere with the optical system of the device. Do not allow any liquid to spill into the device whilst cleaning the cartridge holder since this can interfere with the optical system. The mean transmittance of the optical system should be within the range of 0.9500 to 1.0500, and should not vary by more than /- 0.01000. The standard deviation of the optical system should be less than 0.00150. The drift of the optical system should be less than 0.01400. Ensure that the program card is never inserted while the power switch is in the On [I] position since this can damage the card. Calibrate the device each time a new lot number of reagent cartridges are to be used. Do not use scissors to open a reagent pack since this could damage the reagent or its plastic pull-tab. When handling a reagent cartridge, do not touch its optical window since this can lead to erroneous results. Do not use a reagent cartridge if: 6.14.1. 6.14.2. 6.14.3. 6.14.4. Physically damaged Flexible plastic pull-tab is loose or missing Desiccant sack is missing Desiccant particles are found loose inside the package 6.15. Do not allow blood or urine to make contact with any of the plastic parts of the capillary holder. 6.16. When using a blood sample obtained from a venipuncture, do not attempt to fill the capillary by touching it to the blood in the collection tube since this will most often lead to the contamination of the capillary holder. 6.17. When wiping the outside surface of the capillary, do not allow the tissue to touch its open end since this could result in loss of sample. 6.18. Once the capillary is filled with the sample, analysis must begin within 5 minutes. Page 39 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval 6.19. Run a quality control test: 6.19.1. 6.19.2. 6.19.3. 6.19.4. For each new lot of reagent kits After calibration To help train new analysts When results do not coincide with the patient’s clinical signs or symptoms. 7. References Bayer Corporation. DCA 2000 Analyser Operating Manual. USA: Bayer Corporation; 1997. Mosby’s dictionary of medicine, nursing and health professions. 7th ed. USA: Mosby Elsevier; 2006. Venipuncture; p. 1948. 8. Appendices SOP/PD/221_02/A1 – Set-Up Menu Table SOP/PD/221_02/A2 – General Set-Up Sub-Menu Table 9. Revision History Version Number 01 02 Amendments/ Reasons for change Initial Release Inclusion of subheading titles of Procedure section in Table of Contents Standardisation of text when referring to appendices and updating version number Page 40 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval SOP/PD/221_02/A1 – Set Up Menu Table Sub-Menu Function - To recall the latest 16 RECALL PREVIOUS TESTS? test results - SET SEQUENCE NUMBER? An optional function that allows the user to develop a numbering system for the test results - RECALL CONTROL RESULTS? To recall the latest 16 control test results VIEW CALIBRATION STATUS? To recall the latest 2 calibration statuses for each reagent test - - - SET DATE/TIME? To set the date and time - - Page 41 of 43 Set-Up Repeatedly press [NEXT] button to scroll through the latest 16 test results Press [ESCAPE] button Press Up or Down arrow buttons to select a number for the first digit. Press [NEXT] button to move the cursor to the next digit Repeat previous steps to select a number for second and third digit Press [ENTER] button to confirm sequence number Press [ESCAPE] button Repeatedly press [NEXT] button to scroll through the latest 16 control test results Press [ESCAPE] button Repeatedly press [NEXT] button to recall the status of each calibration record Press [ESCAPE] button Repeatedly press Up and Down arrow buttons until desired day is displayed Press [NEXT] button Repeatedly press Up and Down arrow buttons until the desired number of the month is displayed Press [NEXT] button Repeatedly press Up and Down arrow buttons until the desired year is displayed Press [NEXT] button Repeatedly press Up and Down arrow buttons until desired hour is displayed Press [NEXT] button Repeatedly press Up and Down UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER INSTRUMENT SETUP? To set-up the device with respect to Time Format (T), Date Format (D), Labels displayed with results (L), Controls (C) and Computer Port Configuration (P) SET CREATININE UNITS? To select between ‘mg/dL’ or ‘mmol/L’ INSTRUMENT SETUP? RUN CONTROL? To determine any existing problems with keyboard, display, bar code reader, printer port, memory and computer port functions To set a desired control number before attempting to start a control test Valid for: 2 years from approval arrow buttons until desired minute is displayed - Press [ENTER] button if a 24 hr format time is selected - Press [NEXT] button and then Up or Down arrow button if an AM/PM format time is selected. Press [ENTER] button. - Repeatedly press [NEXT] button until cursor is displayed beneath the desired letter abbreviation - Press Up or Down arrow button to scroll through the different numbers (each number represents a setting, see SOP/PD/221_02/A2) - Press [NEXT] button to move the cursor to the next letter abbreviation - Press [ESCAPE] button when finished from this sub-menu - Press Up or Down arrow buttons to select between [mg/dL] and [mmol/L] - Press [ENTER] button to confirm - Press [ESCAPE] button To be used only by authorised supplier Available only if selected previously in the INSTRUMENT SETUP submenu (when control is different from a DCA 2000) Page 42 of 43 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/221_02 STANDARD OPERATING PROCEDURE DCA 2000® ANALYSER Valid for: 2 years from approval SOP/PD/221_02/A2 – General Set Up Sub-Menu Table General Set-Up Setting Number 1 Setting Definition AM/PM T – TIME FORMAT D – DATE FORMAT L – Labels displayed with results 2 1 24 hour format Month/Day/Year 2 Day/Month/Year 3 0 Year/Month/Day No labels 1 Sequence number that resets daily 2 Continuous sequence number 3 Time assay began 4 Time assay began; sequence number that resets daily 5 Time assay began; continuous sequence number Only DCA controls are used 0 C – CONTROLS P – Computer Port Configuration 1 0 Any other control can be used Computer port is OFF 2 Computer port is ON Page 43 of 43