Outline a method you would use to calibrate a volumetric flask and/or pipette in your laboratory.
Where are your glassware records kept?
Photocopy a glassware calibration record form and attach it to this worksheet.
Complete the following question.
A laboratory technician, calibrates the fluoride meter on the 24th and 25th of July. The following is an
extract from his laboratory logbook.
Fluoride Calibration
240707
20.5o
250707
20.0 o
Worksheet 2 – Calibrate equipment
122.7/ 80/
122.6/ 78/
-55.81
-55.63
Page 1 of 3
1.
Enter the results accurately into the table below.
2.
Look at the readings for the 1.5g.L1 standard and note the trend (multiple choice
increasing/decreasing/staying steady)
3.
Select an initial course of action from the following list. Write your suggestion into the
comments box for the 25th July.
– Send meter back to the manufacturer
– Make up all new solutions
– Prepare fresh 1.5 mg.L-1 solution
– Perform 1 point calibration
Worksheet 2 – Calibrate equipment
Page 2 of 3
LABORATORY FLUORIDE ION METER CALIBRATION RECORD
Meter
(Make)
Electrode
(Make)
Date of
Calibration
21/07/07
22/07/07
23/07/07
TPS
Meter
(Model #)
ORION
Electrode
(Model #)
Type
(1point or
2Point)
2 point
2 point
2 point
Temp.
(oC)
20.5
20.0
21.0
Worksheet 2 – Calibrate equipment
0.50mg.L-1
F- solution
mV reading
122.7
122.6
121.5
900-1
Meter
(Serial #)
R 1427
9609 BN
Electrode
(serial #)
192 44
1.50mg.L-1
F- solution
mV
reading
86
84
82
Standard
Solution
Concn
Standard
Solution
Reading
Slope
Analyst
Print
Name
Sign
L Tech
-55.48
Lab
Technician
L Tech
-55.13
Lab
Technician
L Tech
-55.90
Lab
Technician
(mV/Decade
)
Comment/Action
Required
Page 3 of 3