Validating Monte Carlo as a clinical tool for commissioning the UCLH proton therapy centre Presented by: Vasilis Rompokos, Malet Place - UCL, 22nd of April 2015 (work by: Ilina Vasileva Korn) Proton Therapy Monte Carlo Both sexy Reality Monte Carlo is neither accurate nor precise Project aim: The primary and most important aim is to treat patients by 2018! Project aim: • Check how bad Geant4 is doing and optimise it • Simulate the nozzle commissioning data that are required by the treatment planning system • Commission the treatment planning system • Create the local protocols Background: Effect of range cut and step size on 230MeV protons Grevillot L et al. “Optimization of GEANT4 settings for proton pencil beam scanning simulations using GATE” Nucl Instrum Methods Phys Res B 268:3295-3305 Background: Effect of precompound Nuclear models (Physics-lists) Grevillot L et al. “Optimization of GEANT4 settings for proton pencil beam scanning simulations using GATE” Nucl Instrum Methods Phys Res B 268:3295-3305 Background: Nuclear halo Dose: - core (primary p) - halo ~ 1/3 beam range (secondary p) - aura (neutrons, γ) Gottschalk B et al. “Nuclear halo of 177MeV proton beam in water” Med Phys submitted 2014 Geant4 code The Monte Carlo code uses Geant4 v10.0.02. The code is installed on the plus1 HEP cluster at UCL. Beam parameters position, energy, energy spread, spot shape and size, direction and angular dispersion Phantom Water, voxels Output Scored absorbed dose in voxels, text and root files Test the physics list and dose scoring algorithms Phantom Water tank * - 30x30x30 cm, ionization potential=75eV (ICRU37 and 49) - rows of 85 voxels: 0.4x0.4x0.35 cm (sensitive volume of Extradin T1) at 10 transverse distances (0-10 cm) Beam parameters * - Position: at the water tank wall - Mono-energetic beam of 177 MeV, no energy spread - Radius: 5.948 mm - Angular dispersion: 31.55 mrad * “Nuclear halo of a 177 MeV proton beam in water: theory, measurement and parametrization” (arXiv:1409.1938v1, 2014, MGH Burr Proton Therapy Center) Physics list G4VUserPhysicsList - G4EmStandardPhysics_option3 - G4HadronPhysicsQGSP_BIC - G4EmExtraPhysics * - G4HadronElasticPhysics - G4StoppingPhysics - G4IonBinaryCascadePhysics - G4NeutronTrackingCut * - G4DecayPhysics Geant4 Hadrontherapy advanced example physics list (QGSP_BIC_EMY). The Hadrontherapy example simulates the passive transport beam line installed at Laboratori Nazionali del Sud (INFN) in Catania, Italy. * TOPAS, Paganetti GATE, Physics List for Proton Therapy Applications Scoring absorbed dose Dose in 10 transverse distances in 85 voxels scored in two ways: 1) G4ScoringManager and 2) G4UserSteppingAction G4UserRunAction - The maximum step size is 0.1 mm – distance to the next interaction or to the next geometrical boundary (5% of voxel thickness) - Production thresholds for e-, e+, γ is 0.1 mm * * “Optimization of Geant4 settings for proton pencil beam scanning simulations using GATE”( Nuc.Inst. and Methods in Phys.Research B 268, 2010) Simulation vs Measurement (Preliminary results) Simulation of 106 protons Simulation vs Measurement (Preliminary results) Simulation vs Measurement (Preliminary results) Simulation vs Measurement (Preliminary results) Simulation vs Measurement (Preliminary results) Simulation vs Measurement (Preliminary results) Future work Implementation: - study the physics list - more variables (LET, fluence, dose per particle type, dose per physics process) - implement sensitive detector Validation: - understand the 177 MeV measurements - data from the PBS nozzle at the Hospital of the University of Pennsylvania PBS data from the Hospital of the University of Pennsylvania 100 MeV – 226.7 MeV PBS data from the Hospital of the University of Pennsylvania Take home message: If User <> “theoretical background understanding” Then If CodeSetup <> “optimised for the specific use” Then MonteCarlo <> “Accurate” MonteCarlo <> “Precise” End If End If Questions? Email: vasilis.rompokos@uclh.nhs.uk Monte Carlo data Solid line: Nuclear reactions turned-off {Proton Therapy Physics (Harald Paganett - p.54)} Moyers MF et al “Calibration of a proton beam energy monitor”, Medical Physics, 2007 34(6): 1952-1966 {Although there are small differences in various correction factors used to calculate the Janni and ICRU stopping power tables, the primary difference is the mean excitation energy or I value.}