Partitions (mm) Slabs (mm) Windows (mm) 1.5mm (3mm for

advertisement
Lead in Doors
Lead on Floor
Partitions (mm)
& Control
Slabs (mm)
Windows (mm)
Radiographic Fluoroscopic Room
1.5mm (3mm for
1.5 mm (3 mm
None on 100
Special Procedures Diagnostic and Tele-Fluoroscopy
Dark Room Side)
Room
Normal Weight
Deep Therapy Room: Conventional
Shielding as Determined by a Licensed Physicist
(Ceiling Mounted X-Ray)
on Partition on
for Door to Dark
Cobalt
Height of Lead
None
2100 mm (7")
Slabs (mm)
in Partitions
mm (min.)
Concrete
(Ceiling Mounted X-Ray)
2
Lead on Ceiling
CADD DETAIL NO. : SD134900-01.DWG
1
Lead in
Betatrons
Linear Accelerators
None
Dark Room Side)
Room
Weight Concrete
4
Animal X-Ray, Animal Research,
Radiography and Fluoroscopy
for Door to Dark
1.5mm (3mm for
1.5mm (3mm for
Dark Room Side)
Dark Room Side)
1
on Partition on
on Partition on
5
Dental X-Ray Area
Oral Surgery Room
General Treatment Operatories
1
6
Dental X-Ray Room (Portable 50/65 KVP)
None
None
7
Fluoroscopic Rooms, Vertical or Tilt Type,
1.5
Gastroenterology (No Radiography)
1.5 mm (No Control
Booth Required)
Operating Room Including Animal Operating
None
None
(Except 3 mm on
Dark Room Side)
(Except 3 mm on
Dark Room Side)
8
(Portable X-Ray), Cardiac and Neurosurgical
Operating Room (Portable X-Ray)
9
Cardiac Catheterization in cardiopulmonary
Clinic (Fixed Equipment)
Operating Room (Fixed Equipment)
10
Ultrasound & Thermography
mm
mm
100
mm
1.5 mm (3 mm for
Door to Dark Room
None
None
None
on
100
mm
None
None
Weight Concrete
2.* LEAD SHIELDING REQUIREMENTS IN PARTITIONS ADJACENT TO, FACING, OR PROXIMITY
TO PHOTOGRAPHIC LABS SHOULD BE REVIEWED BY A REGISTERED PHYSICIST.
on
100
mm
None
2100 mm (7")
None
2100 mm (7")
Weight Concrete
Room
ALL RADIOLOGY FACILITY PROJECTS WILL BE REVIEWED BY THE VA HEALTH PHYSICS
None
(min.) Normal
Dark Room Side)
PROGRAM OFFICE FOR APPROPRIATE APPLICATION OF PROTECTIVE STANDARDS.
2100 mm (7")
(min.) Normal
None
None
None
None
None
1.5 mm (3 mm
on Partition on
None
Weight Concrete
1.5mm (3mm for
GENERAL NOTES:
1.
mm
on
(min.) Normal
for Door to Dark
None
on
100
mm
None
(min.) Normal
None
None
None
2100 mm (7")
None
NONE
1.5 mm (3 mm
on Partition on
SCALE :
1.5mm (3mm for
LINING IN X-RAY ROOMS
Cystoscopic Room
DETAIL TITLE / LEAD
3
DATE ISSUED :DECEMBER 2008
Room
MILLI-
DECIMALS
FRACTIONS
LBS. PER
METERS
OF AN INCH
OF AN INCH
SQ. FT
LEAD GLASS EQUIVALENT
0.396
0.0156
1/64"
1
1 PLY APPROX. 1/4"
0.792
0.0312
1/32"
2
1 PLY APPROX. 1/4"
1.000
0.0391
5/128"
2 1/2
1 PLY APPROX. 1/4"
1.191
0.0469
3/64"
3
1 PLY APPROX. 1/4"
1.587
0.0625
1/16"
4
1 PLY APPROX. 1/4"
1.984
0.0781
5/64"
5
1 PLY APPROX. 1/4"
2.380
0.0937
3/32"
6
1 PLY APPROX. 1/4"
3.175
0.1250
1/8"
8
2 PLY APPROX. 1/2"
3.967
0.1562
5/32"
10
2 PLY APPROX. 1/2"
4.763
0.1875
3/16
12
3 PLY APPROX. 3/4"
5.555
0.2187
7/32"
14
3 PLY APPROX. 3/4"
6.350
0.2500
1/4"
16
4 PLY APPROX. 1"
7.142
0.2812
9/32"
18
4 PLY APPROX. 1"
8.465
0.3333
1/3"
20
5 PLY APPROX. 1 1/4"
9.525
0.3750
* 3/8"
22 1/2
6 PLY APPROX. 1 1/2"
10.160
0.4000
2/5"
24
6 PLY APPROX. 1 1/2"
10.317
0.4062
13/32"
25
6 PLY APPROX. 1 1/2"
11.112
0.4375
* 7/16"
28
7 PLY APPROX. 1 3/4"
12.700
0.5000
30
7 PLY APPROX. 1 3/4"
14.267
0.5625
34
8 PLY APPROX. 2"
15.876
0.6250
5/8"
38
9 PLY APPROX. 2 1/4"
16.934
0.6667
2/3"
40
9 PLY APPROX. 2 1/4"
19.050
0.7500
3/4"
44
10 PLY APPROX. 2 1/2"
20.637
0.8125
48
11 PLY APPROX. 2 3/4"
22.225
0.8750
7/8"
52
12 PLY APPROX. 3"
25.400
1.000
1"
60
14 PLY APPROX. 3 1/4"
1/2"
* 9/16"
* 13/16"
GENERAL NOTES:
1.
SHEET LEAD IN THIS TABLE GIVEN IN COMMERCIAL THICKNESSES EXCEPT AS NOTED.
2.* THESE SPECIAL THICKNESSES MAY BE ROLLED IF IN SUFFICIENT QUANTITIES
DETAIL TITLE / LEAD
SCALE :
SHEET THICKNESS & WEIGHT
NONE
DATE ISSUED :DECEMBER 2008
CADD DETAIL NO. :
SD134900-02.DWG
[100mm]
10"
4'-0"
4"
[100mm]
[1670mm]
[250mm]
[1200mm]
[100mm]
METAL STUDS
RACEWAY
DOUBLE 4" [100mm]
STUD WALL
D
FLOOR
1" [25mm] SPACE
RACEWAY
RACEWAY ABOVE CEILING
SEE NOTE 2 THIS DWG.
LINED PARTITION
.
SECTION E-E
LEAD LINED PARTITION
.
1.50 mm LEAD
4" [100mm]
E
METAL STUD
.
WALL
8'-6"
E
[2590mm]
10"
5/8" (16mm)
GWB
[250mm]
3 1/2" [90mm] VERTICAL RACEWAY
16'-0"
CORRIDOR SIDE
[4880mm]
PLAN
STUD WALL
5/8" (15mm) GYPSUM
LEAD LINED
FROM VERTICAL
TRANSITIONAL SECTION
PARTITION
1 1/2"
GENERAL NOTES:
1.
VERTICAL & HORIZONTAL RACEWAYS REQUIRE A 6" [152mm] DEPTH WHERE THERE IS A 3.5"
[115mm] TOPPING & DEPRESS THE SLAB 2.5" [65mm]. WHERE THE SLAB DOES NOT HAVE A
TOPPING, DEPRESS THE SLAB 6" [152mm].
2. RACEWAY CORNERS ARE BEVELED AT 45 DEGREES TO FACILITATE CONDUIT INSTALLATION.
3.
6" [152mm] LEAD LINED PARTITION AT THE ENTIRE PERIMETER.
FINISHED FLOOR
[40mm]
LEAD LINING
CONCRETE
IF REQ'D.
FILL TOPPING
5"
5"
[125mm]
[125mm]
SECTION F-F
STRUCTURAL
FLOOR SLAB
DATE ISSUED :DECEMBER 2008
[760mm]
5/8" [15mm] GWB
NONE
X-RAY ROOM
2'-6"
4" [100mm] OR
6" [150mm]
SCALE :
PARTITION
PLAN AND SECTIONS
LEAD LINED
[1525mm]
SECTION D-D
DETAIL TITLE / DOD / VA UNIVERSAL X-RAY
(R-F) ROOM
4'-0"
5'-0"
LINED PARTITION
METAL STUDS
]
[2950mm]
1" [25mm] SPACE
1.50 mm LEAD
N.
mm
.
RACEWAY
4" [100mm] OR
6" [150mm]
[2130mm]
D
VERTICAL RACEWAY
12" [305mm]
MI
9'-8"
"
35
F
[1930mm]
10" [250mm]
RACEWAY
5/8" (16mm) GWB
7'-0" HIGH
6'-3"
1'-5" [430mm]
ON VERTICAL
0
CONF. RM.
10" [250mm]
MAIN BREAKER
-6
[4
"
6
m]
'
3
6m
COVER
OFFICE OR
F
135°
12" [305mm]
PLATE
RACEWAY
[1625mm]
[1
VERTICAL
[1220mm]
.
.
SUPPLY
LOCATION
.
10" [250mm]
RACEWAY
5'-4"
1'
1'-5" [430mm]
4" [100mm]
METAL STUDS
METAL STUD
POWER
.
10" [250mm]
PARTITION
.
12" [305mm]
LEAD LINED
5/8" [16mm] GWB
CADD DETAIL NO. : SD134900-03.DWG
[250mm]
5'-6"
WALL
[6100mm]
4"
4"
MTL. STUD
20'-0"
10"
SD134900-04.DWG
7'-0" HIGH PARTITION
CADD DETAIL NO. :
AT CONTROL BELOW
1/2" [12mm] DIA.
HANGER RODS FOR
CEILING
PANELS
PANELS
CEILING SUPPORT
GRID
ATTACH CLNG.
EDGE ANGLE
TO CHANNELS
A
OR
REFLECTED CEILING PLAN
UNIVERSAL X-RAY ROOM
PROVIDE SUSPENDED GRID OF STRUT CHANNELS TO SUPPORT X-RAY MACHINE RAILS.
SHALL DESIGN SUPPORT SYSTEM TO SUIT BUILDING STRUCTURE.
SUPPORT RAILS FROM
A/E
STRUCTURAL BEAMS OR JOISTS (PROVIDE INTERMEDIATE FRAMING AS NECESSARY). DO NOT
SUPPORT FROM METAL DECK, CONCRETE SLABS, OR WALK-ON (INTERSTITIAL) DECKS. BRACE
SUPPORT SYSTEM FOR LATERAL LOADS FROM MOVING EQUIPMENT AND SEISMIC.
LIGHTING AND HVAC PER VA DESIGN MANUALS.
WITH OPTIONAL CEILING
CHANNELS
TILE FLANGES
A
GENERAL NOTES:
PROVIDE
K-SERIES CHANNELS
SERIES
SUPPORT STRUT DETAILS LAY-IN CEILING
DATE ISSUED :DECEMBER 2008
ACOUSTICAL
ACOUSTICAL
CEILING
NONE
TYP.
SCALE :
CEILING PANELS 2x2
MOUNTED X-RAY EQUIPMENT
SYSTEM FOR CEILING
LAY-IN ACOUSTICAL
DETAIL TITLE / SUPPORT
26 TO 26-5/8" O.C.
EQUALLY SPACED STRUTS
FINAL LEVELING
SD134900-05.DWG
ABOVE CEILING
ABOVE CEILNG
CADD DETAIL NO. :
RACEWAYS
C
RACEWAY
IN FLOOR
RACEWAY
R/F ROOM SCHEME 1
USE WITH VA HOSPITAL
BUILDING SYSTEM (VA HBS)
1 1/2"
[40mm]
[225mm]
SECTION B
RACEWAY
[225mm]
10"
[300mm]
IF REQ'D
[140mm]
NOTE 2
NOTE 2
12" WIDE COVER
PLATE FOR RACEWAY
STEEL REMOVABLE COVER
1 1/2"
[40mm]
LEAD LINING
5 1/2"
10"
[90mm]
3 1/2"
[305mm]
VCT OVER 1/4" [5mm]
FIELD INSTALL
WHERE REQ'D.
12"
X-RAY TABLE BASE
RWT-60 PARTITION
RACEWAY
SECTION C
GENERAL NOTES:
1. REFER TO STANDARD DETAILS 13 49 00-1 & 13 49 00-2 FOR LEAD LINING
REQUIREMENTS.
2. INSTALL DIVIDERS IN RACEWAYS IF REQUIRED.
3. LOCATE RACEWAY ON THE UNDERSIDE OF THE VA HBS WALK-ON DECK.
3 1/2" [90mm]
5"
CONC. FILL TOPPING
[125mm]
5"
[125mm]
STRUCTURAL
FLOOR SLAB
6"x6"x1/4" STEEL PLATE
[150mm x 150mm x 6mm]
SECTION A
1" [25mm] DIA. I.D. STL.
PIPE SLEEVE BONDED
W/1mm LEAD & 5/8"
[15mm] THRU BOLT.
NONE
R/F ROOM SCHEME 2
(CONVENTIONAL
CONSTRUCTION)
IN FLOOR
SCALE :
B
SCHEME 1 & 2
B
DETAIL TITLE /
A
A
DOD / VA UNIVERSAL X-RAY
(R-F0) ROOM RACEWAY LAYOUT
C
DATE ISSUED :DECEMBER 2008
RACEWAYS
SEE
P1000
A
B
C
OR
D
P1000
4'-0"
[1219mm]
[1219mm]
10"[230mm] MAX.
P2473
DIST. FROM WALL
P2346
P1000
P2346
C
P5500
[124mm]
P1001
5-1/16"
[127mm]
4-7/8"
[127mm]
P5500
P1001
P1000 WITH ANCHORS
INSERTED IN CONCRETE
SLAB.
LAY-IN CEILING
R/F ROOM SCHEME 1
P2346
TRACK(S) BY EQUIPMENT SUPPLIER
P1000
P3184 F WHERE P1001 IS EXPOSED
D
CLEARANCE FOR RECESSED LIGHTS AND CEILING
SECTION THROUGH SUPPORT SYSTEM
NOTES:
ALL SUPPORTS AND HANGERS SHALL BE ARRANGED
L
BENT P
1/4"X4"X4"X6"
METAL PLATE
[6mmX100mmX100mmX150mm]
AS SHOWN ON ARCHITECTURAL DRAWINGS AND SHALL
TERMINATE APPROX. 6" [150mm] FROM FACE OF WALLS
OR PARTITIONS, EXCEPT FOR MEMBERS FLUSH WITH
CEILING.
THE NUMBER SHOWN ON DRAWING WITH A "P"
PREFIX ARE UNISTRUT CATALOG NUMBERS USED AS A
CONVENIENCE FOR IDENTIFYING DESIRED SHAPES AND
THICKNESS OF MEMBERS.
NOTE FOR ARCHITECT:
DETAILS A, B, C & D SHOW ANCHORAGE SUPPORT
ANCHOR BOLTS THROUGH
INTO VARIOUS TYPES OF SLAB CONDITIONS. ACTUAL
SUPPORT MUST BE DESIGNED TO SUIT THE
CONCRETE SLAB
PARTICULAR SITUATION.
P2346
P1000
25" [625mm]
APPROX.
B
NONE
4'-0"
EXPANSION SHIELDS (MIN.
FACTOR OF SAFETY OF 4).
(PROVIDE ADEQUATE INSPECTION
OR TESTING FOR PULL OUT).
SCALE :
ANCHOR BOLTS WITH
A
CADD DETAIL NO. :SD134900-06.DWG
D
DOD / VA UNIVERSAL X-RAY (R-F) ROOM SUPPORT
OR
SYSTEM FOR CEILING MOUNTED X-RAY EQUIPMENT
C
DETAIL TITLE /
DIAGONAL BRACING MAY BE REQUIRED
B
P2473
5-1/16"
PENDING THIS DIMENSION
A
SEE
DATE ISSUED :DECEMBER 2008
P2346
P2346