PlantPen PRI 200 & NDVI 300 Instruction Manual P

advertisement
PlantPen
PRI 200 & NDVI 300
Instruction Manual
Please read this manual before operating this product
P
PSI, spol. s r. o., Drásov 470, 664 24 Drásov, Czech Republic
FAX: +420 511 440 901, TEL: +420 511 440 011, www.psi.cz
CONTENTS
Contents ........................................................................................................................................................... 2
Instruction Manual ........................................................................................................................................... 3
Technical Specifications .................................................................................................................................... 3
General Description .......................................................................................................................................... 6
Components of the PlantPen Series .................................................................................................................. 7
Description of the PlantPen Device- .................................................................................................................. 8
Operation Instructions .................................................................................................................................... 10
Control Menu Tree.......................................................................................................................................... 11
Connection of the Device via USB Pairing ........................................................................................................ 16
Bluetooth Pairing....................................................................................................................................... 18
Bluetooth Connection................................................................................................................................ 22
PlantPen Software .......................................................................................................................................... 24
Start up ..................................................................................................................................................... 24
Theory and References ................................................................................................................................... 29
PRI - Photochemical Reflectance Index ...................................................................................................... 29
NDVI - Normalized Difference Vegetation Index ......................................................................................... 29
References ................................................................................................................................................ 30
Software Update............................................................................................................................................. 31
GPS Module .................................................................................................................................................... 33
GPS Module Description and Setting .......................................................................................................... 33
First Time Installation of the GPS Module .................................................................................................. 34
GPS / PlantPen Operation .......................................................................................................................... 35
Data Download.......................................................................................................................................... 35
WARRANTY TERMS AND CONDITIONS............................................................................................................. 38
Troubleshooting and Customer Support .......................................................................................................... 39
2
INSTRUCTION MANUAL
TECHNICAL SPECIFICAT IONS
Measured Parameters:
PRI (Photochemical Reflectance Index)
PRI = (R525 - R592) / (R525 + R592)
NDVI (Normalized Difference Vegetative Index)
NDVI = (NIR760 – VIS635) / (NIR760 + VIS635)
Measuring Light:
PRI: Internal dual wavelength light source R525 = 525 nm (bandwidth 520 nm - 540 nm), R592 = 592 nm
(bandwidth 580 nm - 600 nm)
NDVI: Internal dual wavelength light source VIS = 635 nm (bandwidth 625 nm - 645 nm), NIR = 760 nm
(bandwidth 750 nm - 760 nm)
Detector Wavelength Range:
PRI: PIN photodiode with 500 to 600 nm bandpass filters
NDVI: PIN photodiode with 620 to 750 nm bandpass filters
Wavelenght Tolerance
± 5 nm
Communication:
Bluetooth, USB, or serial
PlantPen / PlantPen 1.0 Software:
Windows 2000, XP, or higher compatible*
Memory Capacity:
Up to 4 Mb
Internal Data Logging:
Up to 100,000 data points
Display:
2 x 8 characters LC display
Keypad:
Sealed, 2-key tactile response
3
Keypad Escape Time:
Turns off after 3 minutes of no use
Power Supply:
4 AAA alkaline batteries (single use or rechargeable)
Battery Life:
48 hours typical with full operation
Low Battery Detection:
Low battery indication displayed
Size:
120 mm x 57 mm x 30 mm
4.7" x 2.2" x 1.2"
Weight:
180 g, 6.5 oz
Sample Holder:
Mechanical leaf clip (non-destructive for plants, effective for sample pre-darkening)
Operating Conditions:
Temperature: 0 to 55 ºC; 32 to 130 ºF Relative humidity: 0 to 95 % (non-condensing)
Storage Conditions:
Temperature: -10 to +60 ºC; 14 to 140 ºF Relative humidity: 0 to 95 % (non-condensing)
Warranty:
1 year parts and labor (see the last page of this Operation Manual for precise conditions)
* Windows is a registered trademark of Microsoft Corporation.
4
The Bluetoothmodule BlueNiceCom III is endowed with a declaration of conformity with the following norms:
EN 300 328 V1.6.1 (2004-11)
EN 301 489-1, -3 V1.5.1 (2003-12)
EN 50371 December 2002
EN60650
FCC Part 15.247
FCC Grantee Code: R7T
Bluotooth Qualified Product Notice:
GRA_013_04
Bluetooth Listing Identifier: B01572
PlantPen models (PRI 200-B, PRI 200-U, NDVI 300-B, NDVI 300-U) incorporate one of the two offered
communication modules: Bluetooth or USB cable. PlantPen 1.0 software provides visualization and data
transfer routines to a computer.
5
GENERAL DESCRIPTION
The PlantPen model PRI 200 measures Photochemical Reflectance Index (PRI) in two narrow wavelength
bands centered close to 525 nm and 592 nm. The photochemical reflectance index (PRI = (R525–
R592)/(R525+R592)) correlates with the epoxidation state of xanthophyll cycle pigments and photosystem
efficiency in many studies under diverse conditions, such as different 1. PRI is sensitive to changes in carotenoid
pigments that are indicative of photosynthetic light use efficiency, the rate of carbon dioxide uptake, or as a
reliable water-stress index. As such, it is used in studies of vegetation productivity and stress.
The PlantPen model NDVI 300 measures Normalized Difference Vegetation Index, which is an important
indicator of chlorophyll content in plants. The pigment in plant leaves, chlorophyll, strongly absorbs visible light
(from 0.4 to 0.7 µm) for use in photosynthesis. The cell structure of the leaves, on the other hand, strongly
reflects near-infrared light (from 0.7 to 1.1 µm). The differences in plant reflectance in the visible and nearinfrared wavelengths is used to calculate NDVI index as follows: NDVI=((NIR-VIS)/((NIR+VIS)), where VIS and
NIR stand for the spectral reflectance measurements acquired in the visible (red) and near-infrared regions,
respectively. NDVI is directly related to the photosynthetic capacity and hence energy absorption of plant
canopies 2. The NDVI 300 device compares reflected light at two distinct wavelengths, 660 and 740 nm.
PlantPen models (PRI 200-B, PRI 200-U, NDVI 300-B, NDVI 300-U) incorporate one of the two offered
communication modules: Bluetooth or USB cable. FluoPen 1.0 software provides visualization and data transfer
routines to a computer.
1
John Gamon, Josep Penuelas, and Christopher Field (1992). A narrow-waveband spectral index that tracks diurnal changes
in photosynthetic efficiency. Remote Sensing of environment, 41, 35-44.
2
Sellers, P. J. (1985) 'Canopy reflectance, photosynthesis, and transpiration', International Journal of Remote Sensing, 6,
1335-1372.
6
COMPONENTS OF THE PLANT PEN SERIES
Carefully unpack the carton. You should have received the following items, as described in table 1 and
shown in fig. 1.
Table 1. Components of the PlantPen Series package.
1.
PlantPen
2. 4 AAA Alkaline Batteries
3. Carrying Case
4. Textile Strap for Comfortable Wearing
5. Self-Adhesive Rubber Pads for Optics Protection
6. Install USB flash disc with FluorPen software
7. USB cable
8. Other Accessories or Optional Features (according to your specific order)
9. This Owner’s Manual (on a USB flash disc)
Fig.1 Components of the PlantPen series package. Numbers correspond to component number in table 1.
Note: If any item is missing, please, contact the manufacturer: support@psi.cz.
Please check the carton for any visible external damage. If you find any damage, notify the carrier and PSI
immediately. The carton and all packing materials should be retained for inspection by the carrier or
insurer. For customer support, please write to: support@psi.cz
7
DESCRIPTION OF THE PLANT PEN DEVICEPlantPen models (PRI 200-B, PRI 200-U, NDVI 300-B, NDVI 300-U) incorporate one of the two offered
communication modules: Bluetooth or USB cable. Depending on the version purchased design of the device
might slightly differ.PRI 200-U version of device is shown in figure 2.
Leaf clip
Optical window
Menu key
Set key
LC display
Textile strap holder
Connector for
communication cable
Fig.2 PlantPen series device with USB connector.
8
LEAF CLIP AND THE PROTECTIVE RUBBER PAD
The optical part of the PlantPen is covered with a self-adhesive rubber pad that protects the optics from dirt or
moisture. The damaged or dirty pad can be easily removed by tearing it off from the surface. To affix a new
one, just remove the sheeting and stick the pad on cleaned and dried surface. Be sure that you placed the pad
properly and that you did not cover the optical window.
BATTERIES
The PlantPen operates from four AAA single-use or rechargeable batteries. They may be easily replaced by
unscrewing the cover of the battery holder on the rear of the instrument. Battery life is approximately 48 hours
when the PlantPen is operated continuously.
9
OPERATION INSTRUCTIONS
The next chapter explains the operation software and the structure of the Main Menu and three Sub-Menus
with all their options. The explanations of the operation software are same for both versions of the PlantPen
devices (NDVI and PRI).

The blue color represents the Main Menu and its Options.

The yellow color represents the first-level Sub-Menus and their Options.

The green color represents the second-level Sub-Menus and their Options.

Full-line arrows are used for the SET key.

Dashed-line arrows are used for the MENU key.
To START the device please press the SET key for 1 second.
To CONTROL the operation of the device use the MENU key to scroll through sequential menu options on the
digital display.
Use the SET key to select a menu option based on cursor (>) position.
Follow the description of the CONTROL MENU TREE in next section for graphical explanation of the device
operation.
10
CONTROL MENU TREE
11
12
13
14
15
DATA TRANSFER AND VISUALISATION
Please follow the next steps for data transfer from the PlantPen device and data visualisation with the FluorPen
1.0 software.
Please note that the device always comes with just one communication module - either USB or Bluetooth. For
USB connection you need to have the USB driver installed in your PC. You find the driver on the installation disk
(USB driver folder). If you check the Device Manager in Windows you should see the USB serial port in the
device tree. In case of missing driver you may download it from the following link:
http://www.psi.cz/ftp/FluorPen/USB_Driver_Setup.exe. When the driver is installed correctly you should be
able to connect to the device in the FluorPen software menu Setup->Device ID.
CONNECTION OF THE DEVICE VIA USB PAIRING
Please follow next steps if you have PRI 200-U /NDVI 300-U type of device.
Before you set up the USB connection between the PlantPen and PC, make sure you have these components:

USB enabled PlantPen

USB cable (fig. 3)

Installed FluorPen 1.0 software to your PC
Fig.3 NDVI-200-U with USB connector and highlighted connector for USB cable.

Connect the USB cable to a computer and PlantPen as shown in fig.4. Then Switch ON the PlantPen
(press and hold the SET key). On computer run the program Fluorpen 1.0. Please refer to page 23 for
detail description of the software.
16
Fig.4 NDVI-200-U with USB connector (left panel) and the connection in USB port of the PC.
17
CONNECTION OF THE DEVICE VIA BLUETOOTH
Please follow next steps if you have PRI 200-B /NDVI 300-B type of device.
Before you set up the USB connection between the PlantPen and PC, make sure you have these components:

Bluetooth enabled PlantPen with built-in Bluetooth capabilities.

Bluetooth enabled PC
The PC used must have Bluetooth wireless technology, either built-in or through a Bluetooth card.
Make sure that the PC's Bluetooth setting is "discoverable" (meaning that it shows up when other
devices search for nearby Bluetooth connections). Consult the user guide for your PC or Bluetooth
card for more information.

Bluetooth configuration software properly set up on PC
Before you can exchange files with your PC, you will need to set up the Bluetooth software that was
provided together with your PC, or your PC's Bluetooth card. This software varies by manufacturer.
Please consult your PC's Bluetooth documentation for more information.

Switched ON Bluetooth in visible mode on both devices
To pair the PlantPen with another Bluetooth device, such as a computer, you will need to ensure that
Bluetooth is switched on visible on both devices.
BLUETOOTH PAIRING
Please follow next steps to establish Bluetooth connection for your PlantPen device.
1.
Enable Bluetooth in the PlantPen

Switch on the PlantPen (press and hold the SET key).

Scroll to the “Setting” menu (press the MENU key twice, then press the SET key once).

Select “BT_On” to enable Bluetooth (press the SET key).
NOTE: Please note that the PlantPen turns off automatically after about 3 minutes of no action. Turning
off the PlantPen will always turn off Bluetooth.
2.
Starting the Bluetooth Application on Your PC
NOTE: Be aware that this description is a general example; some of the steps may be different on your
PC.

Select: Start>All Programs>My Bluetooth Places (fig. 5)
18
Fig.5 Selection of my Bluetooth places function.
You may also start your Bluetooth application via the Control Panel:

3.
Select: Start>Settings>Control Panel>Bluetooth Devices.
Opening the Add Bluetooth Device Wizard

Select: “Add” to start the wizard (fig.6)
NOTE: Before starting to use the wizard, be sure that the PlantPen is in discoverable mode (see Step 1).
Fig.6 Add function of Bluetooth device.
4.
Searching for a New Bluetooth Device

Mark the following box: “My device is set up and ready to be found” as shown in fig. 7

Go to Next step.
19
Fig.7 Searching of Bluetooth device.
5.
Selecting the PlantPen

Select: Bluetooth PlantPen icon (PlantPen FluoroMeter PSI) (fig. 7)

Select: Next.
Note: Please note that your display may show more Bluetooth Device icons.
Fig.8 Selection and adding of Bluetooth device.
6.
Starting the Pairing Process

Select: “Let me choose my own passkey.” (Fig. 9)

Enter: 0000 (four digits).

Select: Next.
20
IMPORTANT NOTE: Your Bluetooth Pairing Passkey is 0000.
Fig.9 Adding of the own passkey.
7.
Completing the PlantPen Pairing

Select: Finish. (fig.10)
Fig.10
21
BLUETOOTH CONNECTION
1.
Installing the PlantPen Software

Install the PlantPen software to your PC. Use the CD that came enclosed with your PlantPen.*
* See page 23 of this Operation Manual for complete information on PlantPen software.
2.
Enabling Bluetooth in the PlantPen

Switch on the PlantPen (press and hold the SET key).

Scroll to the “Setting menu” (press the MENU key twice,

Select “BT_On” to enable Bluetooth (press the SET key).*
then press the SET key once).
* Keep in mind that the PlantPen turns off automatically after about 3 minutes of no action. Turning off the
PlantPen always turns off Bluetooth.
3.
Starting Connection

Start the PlantPen software on your PC.

Select: Setup>Device ID (Ctrl+I).

If properly connected, the message “Device: PlantPen” appears on the bottom part of the screen.
22
4.
Registering the PlantPen Software

Select: Help>Register.

Enter your serial (registration) number.*

Select: OK.
* You will find your serial (registration) number in the file SN.txt on the enclosed CD.
Important Note: How to Reconnect Bluetooth
Disconnection can occur, either when the Bluetooth feature has been turned off in one or both of the devices,
or when the units move outside their operating range.
IF THE DEVICES HAVE BEEN TURNED OFF, SIMPLY TURN THEM ON AND ENABLE BLUETOOTH IN THE
PLANTPEN AGAIN.
IF THE PLANTPEN HAS BEEN MOVED OUTSIDE THE BLUETOOTH OPERATIONAL RANGE, BRINGING IT BACK
INTO RANGE WITHIN 90 SECONDS WILL ALLOW IT TO RECONNECT AUTOMATICALLY. IF MORE TIME
ELAPSES, SIMPLY TURN THE PLANTPEN ON AND ENABLE BLUETOOTH AGAIN.
23
PLANTPEN SOFTWARE
This Chapter applies to users of the PlantPen NDVI 300 and to users of the PlantPen PRI 200.
START UP
1. STARTING UP
Please follow next steps to start the FluorPen software for visualisation and data transfer of data from PlantPen
devices.
Note: Please note that measurements with the PlantPen devices can be performed also when directly
connected to the PC via USB cable. Data download can be performed only when performed measurement is
finished.
1. Switch on the computer.
2. Switch on the PlantPen and enable Bluetooth.*
3. Make sure that your PC and the PlantPen are properly paired.*
4. For PlantPens supplied with the USB communication module, use provided USB cable to connect your device
to a PC.
5. Start the PlantPen program.
* Points 2 and 3 apply only to users of the PlantPen NDVI 300-B and PlantPen PRI 200-B. See page 17 of this
Operation Manual for complete information on Bluetooth pairing and connecting.
6. Connect device. Select: Setup>Device ID (Ctrl+I).

If properly connected, the message “Device: PlantPen” appears in the bottom part of the screen.
24
Menu: Device
Download
Downloads data from the PlantPen to your PC.
Erase Memory
Erases data from the PlantPen memory.
Online Control
Used to control device which is connected – used for setting the sound, time,
protocols which can be measured and specification of measurement settings such as
intensity of. measuring pulses or saturating pulses used for chlorophyll kinetic
measurements (Note: This option is not included in the NDVI and PRI pen).
Note: Online Control settings differ depending on the PlantPen series type.
Fig.11 Online control.
7. To visualise the stored data use the download function to transfer your data from the PlantPen device to
your PC. Select icon on left side of screen (as shown below) or go to Device and select Download. After
successful download you will see table with all recorded data as it is shown in fig. 12. All data stored are logged
with time when measurement, experiment respectively, was performed. Experiment description function can
be used to name the measurement with respective label. Please note that if there are no data stored in the
memory, the download function is not functional.
Download
8. To further work with the data downloaded from the PlantPen device to the PC please use the File window in
the FluorPen software toolbar. Downloaded data files can be visualised, modified by labelling the individual
measurements or used for exporting the data into .txt files. For detail explanation of individual functions please
refer to the description below.
25
Time of experiment
Measured parameter
File name
Save experiment
Measured and
calculated values
Space for
written notes
Experiment description
Fig. 12 Visualisation of stored data with the FluorPen software after data download.
Menu: File
Load
Loads previously saved data files.
Save
Saves data to hard disc.
Export
Exports data in .txt format.
Close
Closes the current experiment.
Close All
Closes all running experiments.
Exit
Exits the program.
Note: Please note that if more data files are loaded they will be automatically visualised as the last one in the
visualisation window if the Add to open function is set in the Settings window. Use the scroll bar to find the
data in the right edge of the window. All measured parameters will be visualised.
Export data – If you want to export the measured data from the FluorPen software to .txt file, you have to
select only one parameter out of the parameters measured to be exported per one time. If more parameters
were measured you need to export them one by one to separate .txt files.
26
Selected only
Export only selected data. Use the right mouse
click to select the data to be exported.
Source data
Export source data, e. g. for OJIP protocol all
time points from the curve are exported.
Computed values
Export calculated values (Fig.12), e.g. for OJIP
measurement measured values as Mo, area,
Fv, Fm, Fj are exported.
Measuring by protocol
Export data measure by protocol (e.g. OJIP,
Light curve 1, 2, 3 etc.)
MURPI
Export Multiple reflectance index
N
Export data measure by N – Pen (measuring of
nitrogen)
Note: Export data window and options differ depending on the PlantPen series type.
9. Setup window is used for setting the connection of the device, update firmware and settings set-up of the
device.
Menu: Setup
Device ID
Detects the Bluetooth or via USB cable connected device.
Update Firmware
Used for software updates. Note: For more information on software updating, see
page 31 of this Operation Manual.
Settings
Used for modification of the program settings. For detail explanation of individual functions
please refer to the description below.
After Download - Memory Erase
If the box is checked, the PlantPen memory is erased after each data download.
Data - Inverted
If the box is checked, the polarity of data is inverted, e.g., multiplied by -1.*
Data – Add to Opened
If the box is checked, the downloaded data are added to that of the current opened experiment.
27
Graf - Single
If the box is checked, all measured data are visualized in one graph, i.e., the value of each new measurement is
added to the currently used graph window.
If the box is not checked, a new graph is opened for every new measurement.
* This feature can be helpful for a certain type of experiment when the measured data are undesirably
interpreted as negative values.
10. Help function is used for basic information about the program and registration of PlantPen.
Menu: Help
About
Register
Offers basic information about the program.
Used for the PlantPen software registration.*
* See Chapter 7C of this Operation Manual for more information on
PlantPen software registration.
28
THEORY AND REFERENCES
PRI - PHOTOCHEMICAL REFLECTANCE INDEX
The Photochemical Reflectance Index or PRI is a reflectance measurement. The PRI is sensitive to changes in
carotenoid pigments (e.g. xanthophyll pigments) in live foliage. Carotenoid pigments are indicative of
photosynthetic light use efficiency, or the rate of carbon dioxide uptake by foliage per unit energy absorbed. As
such, it is used in studies of vegetation productivity and stress. Because the PRI measures plant responses to
stress, it can be used to assess general ecosystem health using satellite data or other forms of remote sensing.
Applications include vegetation health in evergreen shrublands, forests, and agricultural crops prior to
senescence. PRI is defined by the following equation using reflectance (ρ) at 531 and 570 nm wavelength:
The values range from -1 to 1.
NDVI - NORMALIZED DIFFERENCE VEGETATION INDEX
Normalized Difference Vegetation Index or NDVI is an important indicator of chlorophyll content in plants.
The device compares reflected light at two distinct wavelengths, 660 and 740 nm.
The NDVI is calculated from these individual measurements as follows:
where VIS and NIR stand for the spectral reflectance measurements acquired in the visible (red) and nearinfrared regions, respectively These spectral reflectances are themselves ratios of the reflected over the
incoming radiation in each spectral band individually, hence they take on values between 0.0 and 1.0. By
design, the NDVI itself thus varies between -1.0 and +1.0. It should be noted that NDVI is functionally, but not
linearly, equivalent to the simple infrared/red ratio (NIR/VIS). The advantage of NDVI over a simple
infrared/red ratio is therefore generally limited to any possible linearity of its functional relationship with
vegetation properties (e.g. biomass). The simple ratio (unlike NDVI) is always positive, which may have practical
advantages, but it also has a mathematically infinite range (0 to infinity), which can be a practical disadvantage
as compared to NDVI. Also in this regard, note that the VIS term in the numerator of NDVI only scales the
result, thereby creating negative values. NDVI is functionally and linearly equivalent to the ratio NIR / (NIR+VIS),
29
which ranges from 0 to 1 and is thus never negative nor limitless in range.But the most important concept in
the understanding of the NDVI algebraic formula is that, despite its name, it is a transformation of a spectral
ratio (NIR/VIS), and it has no functional relationship to a spectral difference (NIR-VIS).
REFERENCES
Below references to scientific publications are listed where PlantPen NDVI and PRI models were used for
various research purposes.
Zarco-Tejada, P.J., González-Dugo, V., Berni, J.A.J. (2012). Fluorescence, temperature and narrow-band indices
acquired from a UAV platform for water stress detection using a micro-hyperspectral imager and a thermal
camera, Remote Sensing of Environment, 117, 322-337
Zarco-Tejada, P.J., Guillén-Climent, M.L., Hernández-Clemente, R., Catalina, A., González, M.R., Martín, P.
(2013). Estimating leaf carotenoid content in vineyards using high resolution hyperspectral imagery acquired
from an unmanned aerial vehicle., Agricultural and Forest Meteorology, 171-172, 281-294.
Cody Chytyk,Pierre Hucl,Gordon Gray (2011). Leaf photosynthetic properties and biomass accumulation of
selected western Canadian spring wheat cultivars. Canadian Journal of Plant Science, 91(2): 305-314.
30
SOFTWARE UPDATE
Very important!
The PlantPen memory is erased during the software update!
Before starting any software update, export all your data from the PlantPen memory into your computer!
Step 1: Starting Update
Select: Setup>Update Firmware
Step 2: Selecting .bxn File
Find: Binary file (with the extension .bxn)
Select: Open
31
Step 3: Finishing Upload
Select: “OK” to start uploading of the update.
The bottom bar indicates the upload progress.
Press: “OK” to finish upload.
32
GPS MODULE
Please follow next steps if your device is purchased together with the GPS module.
IMPORTANT INFORMATION:
For proper GPS reading, the TIME in your PlantPen and in your computer must be synchronized! Preset time
and time zone must correspond to GPS time (time zone) in your location.
GPS MODULE DESCRIPTI ON AND SETTING
Step 1: AA Batteries Installation
The device operates on two AA batteries. You can use alkaline, NiMH or lithium batteries. Use NiMH of
lithium batteries for best results.
1.
Turn the D-ring counter clockwise and pull up to remove the cover.
33
2.
3.
Insert the batteries, observing polarity.
Replace the battery cover and turn the D-ring clockwise.
4.
5.
6.
Hold
Select Setup->System->Battery Type
Select Alkaline, Lithium or Rechargeable NiMH
Step 2: GPS Device setting
1.
Hold
2.
After the device is on, it begins acquiring satellite signals. The device may need a clear view of the
sky to acquire satellite signals.
3.
You can check current position and accuracy in Satellite Page
4.
For better precision you can enable GPS+GLONASS mode in Setup->System
5.
->Satellite System->GPS+GLONASS and turn WAAS/EGNOS On.
6.
Make sure that USB Mode is set to Mass Storage (Setup->System->USB Mode)
7.
For more information please check Garmin eTrex manual.
FIRST TIME INSTALLATION OF THE GPS MODULE
Step 1: Enabling GPS Module in the PlantPen Software
Select: Setup>Settings
34
Select: Settings>GPS visible
Step 2: Accepting GPS Setting
Close and restart the PlantPen program to accept GPS setting.
GPS / PLANTPEN OPERATION
Step 1: Time Synchronization
Synchronize the PlantPen time with the time of your PC. Time must be set correctly witch respect to your
time zone.
Step 2: GPS Positioning
Switch the GPS module on and wait until the GPS position is fixed (GPS green LED indicator flashes 1s ON
and 2s OFF while tracking the position).
Step 3: Operation
Be aware that while performing field experiments, the PlantPen and the GPS module must be kept close to
each other.
DATA DOWNLOAD
Step 1: Enabling Communication
Switch on the computer, PlantPen, and GPS module and set your computer to PlantPen communication
(enable Bluetooth or connect to serial or USB port).
35
Step 2: Downloading PlantPen Data
Start: PlantPen program.
Connect: PlantPen device.
Download: Measured data from the
PlantPen to your PC.
Be aware that no GPS coordinates are
visible at this moment.
Step 3: Connecting GPS Module
Connect the GPS Module to your PC. Communication is set properly if the hardware is recognized by your
PC.
36
Step 4: Downloading GPS Data
Select: Device>Attach GPS file to download data from the GPS module.
Step 5: Completing the Download
Successfully
downloaded
GPS
coordinates paired with PlantPen
data.
37
WARRANTY TERMS AND CONDITIONS

This Limited Warranty applies only to the PlantPen series devices. It is valid for one year from the date
of shipment.

If at any time within this warranty period the instrument does not function as warranted, return it and
the manufacturer will repair or replace it at no charge. The customer is responsible for shipping and
insurance charges (for the full product value) to PSI. The manufacturer is responsible for shipping and
insurance on return of the instrument to the customer.

No warranty will apply to any instrument that has been (i) modified, altered, or repaired by persons
unauthorized by the manufacturer; (ii) subjected to misuse, negligence, or accident; (iii) connected,
installed, adjusted, or used otherwise than in accordance with the instructions supplied by the
manufacturer.

The warranty is return-to-base only, and does not include on-site repair charges such as labor, travel,
or other expenses associated with the repair or installation of replacement parts at the customer's
site.

The manufacturer repairs or replaces faulty instruments as quickly as possible; the maximum time is
one month.

The manufacturer will keep spare parts or their adequate substitutes for a period of at least five years.

Returned instruments must be packaged sufficiently so as not to assume any transit damage. If
damage is caused due to insufficient packaging, the instrument will be treated as an out-of-warranty
repair and charged as such.

PSI also offers out-of-warranty repairs. These are usually returned to the customer on a cash-ondelivery basis.

WEAR & TEAR ITEMS (such as sealing, tubing, padding, etc.) are excluded from this warranty. The
term WEAR & TEAR denotes the damage that naturally and inevitably occurs as a result of normal use
or aging even when an item is used competently and with care and proper maintenance.
38
TROUBLESHOOTING AND CUSTOMER SUPPORT
In case of troubles and for customer support, please, write to support@psi.cz or contact your local distributor.
Manual Version: 2014/02
© PSI (Photon Systems Instruments), spol. s r.o.