Technical Note of Apparatus of Spin-dependent Short-range Forces Experiment Ping-Han Chua,∗ a Triangle Universities Nuclear Laboratory and Department of Physics, Duke University, Durham, North Carolina 27708, USA 1. Measurement The free induction decay (FID) is applied to measure signals of nuclear magnetic resonance(NMR) from 3 He precession. The system is controlled by a pulse generator. It sends triggers to a function generator, a computer DAQ card, and preamplifiers, sequently. After receiving a trigger, the function generator generates a pulse with few cycles (usually we use 3 cycles with voltage 10 volts) to a rf coil. The rf coil generates an oscillating field around the Larmor frequency and tip the spin of 3 He with a small angle. The trick is to get signals big enough for NMR measurement but small enough so that FID signals don’t decay too quickly. Once the spin is away from the holding field direction, the spin starts to precess and induces an emf in the pickup coil. The induced current will go through a preamplifier and be recorded through the DAQ card. The more details of operation are described in [1]. The parameters for the preamplifiers are really empirical. Usually we set the filter range to cover the Larmor frequency. The gains are set so that the signals of two pickup coils have similar heights and best signal-to-noise ratios. Usually we set the gain around 103 to 104 . 2. Electronics I briefly describe the electronics used in this system. The block diagram of the electronics is shown in Fig. 1 [2]. The function generator is Tektronix Corresponding author at : Triangle Universities Nuclear Laboratory (Bldg), Duke University, Durham, North Carolina 27708, USA. Tel.: +1 9196602950 Email address: pinghan@gmail.com (Ping-Han Chu) ∗ Preprint submitted to Elsevier October 9, 2014 AFG 3021. The pulse generator is Quantum Composers Model 9614. The preamplifier is Stanford Research System Model SR560. The DAQ card is NI PCI-5122. Figure 1: Block diagram of electronics 3. Magnetic Fields The magnetic field system is composed of several coils in order to provide a uniform magnetic field, as shown in Fig. 2 [3]. The red Helmholtz coil with diameter of 60 inches provides the holding field along the ẑ-axis. Three square Helmholtz coils, called the compensation coils, with length of 84 inches along x̂, ŷ and ẑ-axis, respectively, form a cube [4]. One square Helmholtz coil can generate a maximum magnetic field more than 600 mG with current I=3 amp [6]. This assembly of the compensation coils can compensate the background fields, which is dominated by Earth’s fields, so that the magnitude of the residual background field is less than 10 mG within a 2 cubic space of length of 10 cm [5]. Together with magnetic shielding systems with the shielding factor 1000, the magnitude of residual background fields, in principle, can be reduced to less than 0.01 mG. Once the absolute magnitude of the field becomes small, the gradient becomes small as well. Currently, we do not have magnetic shielding systems. Without shielding, Cx and Cy do not play roles to improve T2 . Two types of gradient coils are used in this system, Maxwell coil and Golay coil (in the page 17 of [4]). The Maxwell coil, as shown in Fig. 3, can provide gradients of dBz /dz. The Golay coil, composed of saddle coils, as shown in Fig. 4 can provide gradients at transverse direction. This system includes two sets of Maxwell coils with radius of 7 inches (a = 7′′ ) along ẑ-axis, and two sets of Golay coils with radius of 7 inches (a = 7′′ ) providing gradients of dBz /dx and dBz /dy, respectively. The detailed studies are described in [7, 8, 9]. The compensation coils are fixed with respect to the holding field and the supportive table. The gradient coils and the cell holder are one set and can be moved together. We can adjust the height and the location of the assembly of the gradient coils with respect to the supportive table. Then the height and the location can be fixed using nuts and plastic threaded rods or screws. Each gradient coil can be re-connected in different ways to create not only linear gradients but also quadratic gradients. At each end of coil, there is a label on it. Here we list the combination of each coil. The dash, “−”, means the connection between two ends of coils. • Compensation coil Cx : Cx 1 − Cx 2 − Cx 3 − Cx 4, quadratic gradient • Compensation coil Cy : Cy 1 − Cy 2 − Cy 3 − Cy 4, quadratic gradient • Compensation coil Cz : Cz 1 − Cz 2 − Cz 3 − Cz 4, quadratic gradient • Gradient coil dBz /dx : x1 − x2 − x3 − x4, linear gradient • Gradient coil dBz /dy : y1 − x2 − x3 − x4, linear gradient • Gradient coil dBz /dz 1: z1 − z2 − z4 − z3, linear gradient • Gradient coil dBz /dz 2: z1 − z2 − z3 − z4, quadratic gradient 3 Figure 2: Magnetic field assembly, including the holding field, the compensation fields and gradient fields. 4. Optimization The goal is to improve the sensitivity of the precession frequency. The sensitivity depends on two factors, the signal-to-noise ratio and the transverse relaxation time, T2 . The signal-to-noise ratio depends on the particle density, the polarization, the geometry of the pickup coil, and the resonance RLC box. The particle density is a default input when constructing cells. Larger particle density should have stronger signals; however, larger particle density as well as larger pressure gives shorter transverse relaxation time, decreasing the sensitivity. There is a trade-off between larger particle density and longer transverse relaxation time. Currently, we don’t have a good method to optimize particle density. People (T. Chupp’s group at Michigan and Walsworth’s group at Harvard) using SEOP for precision measurements use the particle density of 1 amg. Duke group used a cell with the particle density of 7 amg for the phase II and made new cells with the particle density 4 Figure 3: Maxwell Coil of 3 amg for the phase III. The polarization depends on the laser frequency and power, the oven temperature and the longitudinal relaxation time. Both needs to be optimized before experiments. In order to tune the laser frequency, a spectrum meter (SM240/SM440 Compact CCD Spectrometer) is put behind the cell and the oven. The absorption spectrum shows double peaks with a dip at the center if the laser is absorbed by rubidium atoms at a right frequency. The laser power is limited by the laser itself. The oven temperature should be optimized by checking the amplitude of FID signals versus temperature. For a rubidium only cell, the temperature is about 160-180 ◦ C. For a hybrid cell (rubidium-potassium mixture), the temperature is usually higher than 200 ◦ C. The longitudinal relaxation time depends on the property of the cell, especially the surface properties. It is beyond the discussion of this note. Several factors can affect the performance of pickup coils: the number of turns, the size of the coil and the inductance which depends on the first two factors. Duke group uses commercial pickup coils with diameter about 1 inch with inductance of 2 mH and resistance of 7 Ω (made by Ablecoil http://www.ablecoil.com/). The resonance RLC box is to change the natural frequency of the circuit. 5 Figure 4: Golay Coil Once the inductance of the pickup coil is known, it is straightforward to add a capacitor in the circuit to change the natural frequency close to the Larmor frequency. It needs an empirical test to determine the capacitor in series or in parallel in the circuit. Another critical noise is the rf noise. We observed strange rf noise in the frequency spectrum; the signal became very noisy with many spikes. After surrounding every electronics, including preamplifiers, function generator, pulse generator, etc., with aluminum foil, the problem has been reduced. The optimization of the transverse relaxation time is tricky. We have done field mapping using a fluxgate magnetometer with stepping motors [6, 10]. We can scan the background field, the holding field, the compensation fields and the gradient fields, respectively. With a careful tuning, in principle, the location of the cell should have quite uniform magnetic fields at the level of less than 0.1mG/cm [11]. However, the T2 is usually quite short without further tuning. There are three possible reasons. First is the background fields changing with time because of the activities around. It is not easy to avoid the background changing without magnetic field shields. Second is the power supply stability. We have observed the time-dependent drift of magnetic fields. Each coil is controlled by a power supply (E3646A 60W 6 Dual Output Power Supply). Many power supplies increase the instability of the magnetic fields. Third is the field generated by polarized atoms [12]. In the page 8 of [12], we estimated a spherical cell with radius of 1 cm with pressure 3 atm and polarization 40% can generate a gradient of 2.66 mG/cm. It can destroy any optimization of fields based on the result of field mapping. Therefore, the optimization of the transverse relaxation is quite empirical. We found out that the configuration of the magnetic field and spin directions can be quite important for the transverse relaxation time in [13] and phase II. The reason is not clear. It seems that it is easier to cancel out gradient fields for a certain configuration, especially when the magnetic field direction is inverse of the polarization direction. Polarized atoms can play a role to compensate gradient fields. We strongly suggest that people should try different configurations for the optimization of transverse relaxation. Once the proper configuration is determined, the following procedure is applied. The cell is moved around until the best T2 is approached. In principle, we should scan every points in the space step by step. However, it is not practical to scan the lattice of space; we usually optimize the vertical direction first because it is not easy to move the cell along this direction. Once the height of the cell is optimized, we can move the cell at the horizontal plane, forward and backward, rightward and leftward until the T2 is optimized. It is easier if there are three people working on this process together. The next step is to tune the gradient coils. In this system there are seven coils we can tune. So far we have only used four of them, dBz /dx, dBz /dy, dBz /dz 2 and Cz because of the limitation of the number of power supply. However, without magnetic shielding systems, Cx and Cy do not play roles to improve T2 . We try four gradient coils independently and fix the one which can provide the best T2 . Then we try other three coils and repeat this process. The optimization process may go back and forth between several coils. Usually a T2 of two seconds is quite common for our current setup. However, it is possible to get T2 longer than ten seconds for two pickup coils with careful tuning(like the page 7 in [19]). A very special case that T2 ∼ 80 seconds for one pickup coil has been obtained(in the page 8 of [13]). 5. Comments In order to succeed the experiment, a cell with strong signals and a long longitudinal relaxation time is necessary. We require the signal to noise ratio close to 10. As mentioned before, this can be done by reducing noise and 7 increasing polarization, etc.. Besides, a long longitudinal relaxation allows that the signal amplitude won’t decay quickly while applying FID. Another thing is the transverse relaxation time. The transverse relaxation does not depend on the cell itself. We have tried several cells including TINY, Pablum, KellyK #1, KellyK #2, and Test Cell #5(in the page 25 of [14], [15], [16]) and all of them give similar T2 result. The only trick to improve T2 is to carefully tune the gradient fields and the cell location. A T2 of at least 10 seconds for both pickup coils is possible and necessary. In a long run, a system like the one in Walsworth’s group at Harvard is necessary; more details are described in D. Bear’s thesis [17, 4, 18]. Magnetic field shields and a solenoid are needed to provide more uniform magnetic field and improve T2 . The holding fields and oven temperature need to be stabilized. In this note, we focus on the improvement of the frequency sensitivity. We can see the success of this experiment is possible. However, a cell with good polarization and a long T1 is necessary for reliable signals. Currently, a new cell being constructed with a short length (5 cm) and thinner windows (100 µm) gives us a good chance to improve sensitivity. Recently, we observed signals with a very long T2 (longer than 50 sec) using the cell Pablum [20]. Pablum is a cell with T1 longer than 10 hours and high density of 9 amg, and has stable polarization. This test demonstrates the possibility to push T2 longer than 10 seconds using gradient coils if the cell has long T1 and stable polarization. However, current magnetic field is not extremely stable so that the NMR signal may change between cycles. Sometimes the signal increases and then decreases (like the page 8 in [20]). In fact, it is similar with the node point we observed before in a shorter time scale because of non-uniform gradient fields. If the measurement time can be extended up to 100 seconds, we should be able to observe decaying NMR signals with node points. However, the DAQ card limits the measurement time because its maximum record length is 2 × 106 points. It is possible to lower the holding field as well as the precession frequency so that the measurement time can be extended. It requires the construction of new RLC boxes and the retuning of gradient fields. References [1] Y. Chang, http://www.phy.duke.edu/~ pc102/MeasurementProcedure.docx [2] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20120904.pdf 8 [3] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20131108.pdf [4] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20130828.pdf [5] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20140207.pdf [6] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20140124.pdf [7] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20140227.pdf [8] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20140307.pdf [9] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20140321.pdf [10] P.-H. Chu, http://mep.phy.duke.edu/index.php/Field_Mapping [11] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20140402.pdf [12] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20130824.pdf [13] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20140620.pdf [14] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20140326.pdf [15] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20140411.pdf [16] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20140523.pdf [17] D. Bear, Ph.D. thesis (2000) http://walsworth.physics.harvard.edu/publications/2000_Bear_HUPhDThesis.pdf [18] P.-H. Chu, http://www.tunl.duke.edu/~ pc102/20131122.pdf [19] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20140628.pdf [20] P.-H. Chu, http://www.phy.duke.edu/~ pc102/20141006.pdf 9