PFC/JA-83-4 A FAST NEUTRON SPECTROMETER FOR D-D FUSION NEUTRON MEASUREMENTS AT THE ALCATOR C TOKAMAK W. A. Fisher, S. H. Chen, D. Gwinn, R. R. Parker Plasma Fusion Center Massachusetts Institute of Technology Cambridge, MA 02139 May 1983 This work was supported No. DE-AC02-78ET51013. by the U.S. Department of Energy Contract Reproduction, translation, publication, use and disposal, in whole or in part by or for the United States government is permitted. By acceptance of this article, the publisher and/or recipient acknowledges the U.S. Government's right to retain a non-exclusive, royalty-free license in and to any copyright covering this paper. A Fast Neutron Spectrometer for D-D Fusion Neutron Measurements at the Alcator C Tokamak W. A. Fisher* and S. H. Chen Nuclear Engineering Department Massachusetts Institute of Technology Cambridge, MA 02139 and D. Gwinn and R. R. Parker Plasma Fusion Center Massachusetts Institute of Technology Cambridge, MA 02139 Abstract 3 He ionization chamber A neutron spectrometer using a high pressure has been designed and used to measure the neutron spectrum from an ohmically heated deuterium plasma. The resolution of the spectrometer at 2.45 ParMeV is determine to be 46 keV full width at the half maximum (FWHM). ticular attention has been paid to optimizing the detector shielding and collimation to reject thermal and epithermal neutrons scattered from the tokamak structure. As a result, measurements indicate that the ratio of the number of counts in the 2.45 MeV peak to the total number of detected For the 8 Psec amplifier time constant used, a neutron events is 1/67. per second has been achieved in the thermocounts 44 as as high count rate have been compared with calculated spectra observed The peak. nuclear Particle Transport code and they Neutral Carlo Monte MCNP spectra using the of neutrons produced from evidence is little There agreement. good show It has been possible to electrodisintegration. or photoneutron reactions is consistent with the peak thermonuclear of the shape the that confirm as determined from temperature ion the that and Gaussian shape predicted diagnostemperature ion C Alcator other with consistent the line width is broadline Doppler of theory the by tics, and follows the trends predicted ening. *Present Address: National Nuclear Corp. 1904 Colony St. Mountain View, CA 94043 Page 2 I. The fact the ion that the fusion neutron distribution energy Introduction of since the early days of presented a paper which most Unfortunately, in fusion a could be used thermonuclear research. In The work discussed here been plasma has known 1967 Lehner and Pohl knowledge been made of this of the difficulty in making high resolution measurements trons. to study [1] of the theory used today was detailed. use has very little spectrum represents a successful because of fusion neuattempt at such a measurement. Strachan et al. have [2] Torus (PLT) Princeton Large to 3 a used He the measure spectrum neutron at chamber ionization due the to a beam heated plasma, but were unable to resolve the D-D fusion peak for an ohmically heated plasma [3]. The measurement has been possible at Alca- the plasma density is an order of magnitude tor C because in PLT, resulting a ratio of fusion neutrons which is two orders of magnitude better than PLT. improved the ion chamber neutron thermal to fast count rate performance on the design shielding count and ratio, at the D-D peak. of the spectrometer. non-fusion to obtaining achieving a The emphasis of Additional details ments at Alcator C may be found in reference neutrons In addition, we have collimation thus greater than much a lower improved this paper is on the measure- [4]. The Alcator C fusion device is a high field compact tokamak located at and operated by the Massachusetts Institute of Technology for the U.S. Page 3 Department of Energy. Figure 1 shows a diagram of the machine. The de- sign parameters for the machine are displayed in table 1. An important result of a neutron spectrum measurement is that it is possible to identify the from its in an energy. Alcator Five physical processes C deuterium electrodisintegration of process are plasma known to be discharge: deuterium, thermonuclear fusion of deuterium, which generated the sources of photoneutron photodisintegration neutron neutrons reactions, of deuterium, and fusion of deuterium and tritium. These processes have been described in detail in references [3] and [5]. Photoneutron production, and photodisintegration of deuterium are pro- cesses which occur at the ring of molybdenum blocks used to define the plasma discharge (limiter). but has a cross-section Electrodisintegration which is several orders is a volume process of magnitude smaller than the D-D cross-section for the particle energies expected in Alcator C plasmas. All the processes except the thermonuclear D-D and D-T reac- tions are a result of some plasma conditions these 'run-away supra-thermal [6]. electrons' electrons which are produced for It is possible to limit the production of by controlling the plasma discharge condi- tions. Tritium produced in the D(d,p)T reaction to produce 14.7 MeV neutrons. reaction can also react via D-T These neutrons would be expected to cause a background of neutrons above the D-D peak due to down-scattering in the room and machine structure. Chrien [71 has measured the D-T rate to D-D rate ratio for PLT and found it to be between 10- Figure 2 shows spectrum. how each of the processes might affect 3 and 10- 4 . a neutron This figure is based mainly on a spectrum measured by Stra- Page 4 chan and Jassby 3 [3] using a He ionization of the same design as used here for an ohmic discharge at the Princton Large Torus PLT. sults were obtained at a plasma density roughly an order of magnitude lower than that used here tant. Their re- where non-thermonuclear processes are impor- As will be seen later, thermonuclear D-D reactions dominated the neutron spectra measured here. The D-D thermonuclear neutron rate is given by r n2 R = f() 2 where n is the functions of ylm2, Deuterium the and U(vrel) neutron spectrum f(y2 ) "(Yrel) d 3 v1 d3 v 2 lyrell (1) -'-' reacting D-D the nuclear can be f(ul),f(u 2 ) are density, pair, urel is A cross-section. obtained by the applying the distribution relative general the velocity, form condition for the that the 3 neutron have an energy - (0+Erel) in the center of mass coordinate 4 system. result of A general formulation evaluations of the is given in spectrum for Lehner the Maxwellian ion distribution function is used here [8]. case of However, the the isotropic [8,9,10,111. Williams [10], for example, found that the three-dimensional isotropic Maxwellian ion velocity plasmas, distribution, the expected would produce an approximately, distribution Gaussian energy for Alcator C spectrum with a mean energy given by 3 <En> kT bkT + =Q 4 2 2 2/3 7 (2) Page 5 and the line width FWHM in keV is given by AEn = 82.5/kT (3) where kT is the plasma temperature in keV as defined by the Maxwellian distribution. In equation (2), b is the coefficient in the Gamow cross- section, E0 a(Eo) -b/ (4) e a(Erel) Erel where EO is an energy at which the cross-section is known. gives the line width for a spatial and time invariant Equation (3) plasma source. average over In a real measurement the line width will represent the space and time of the plasma distribution. the maximum ion temperature temperature representing time is a more characteristic plasma parameter and ion However, in both space and is generally used. Thus a correction has to be made to obtain the peak central ion temperature from the temperature measured from the line width of the neutron spectrum. As will be seen in section IV it is necessary to integrate over a large portion of an Alcator plasma and also to sum the spectra of many plasma discharges (shots) represents was collected during from the about 11% total to obtain sufficient statistics in Figure 3 illustrates a typical plasma discharge. the neutron spectrum. The shaded area together the time during which the neutron the discharge. neutron rate The ion temperature can be deduced measurement over the shaded region. spectrum and is estimated to vary by However, the effect of averaging the Page 6 ion temperature from a measurement of the peak width is calculated yield a temperature which is 0.96 the peak ion temperature to or a 4% cor- rection. The ion temperature C is assumed to be poloidally and in Alcator toroidally constant but to vary with the minor radius as given by 2 r Ti(r) at e =Ti() (5) 2 at = 3 -a 2 qX is the limiter q [12]. - qy qo is central plasma safety factor and is the limiter radius, where a, q0 2 A computer program was written to numerically calculate the neutron spectrum for the Alcator C temperature and density Table profiles. qj. summarizes 2 the correction Ip/Bt indicating the correction slightly dependent of the major plasma parameters. can be understood where the neutron by production is essentially constant. However, examining Fig. function of the correction is quite constant. Note that over a wide range of qZ qZ is proportional to as a factor 4 and important, very The reason for this noting that r is factor < 6 cm, for the region the density is At 6.0 cm the density is 0.93 the central density. as can be seen in Table 2 the maximum neutron rate as a function of the minor radius does not occur at the center where the ion temperature is a maximum, but at about a minor radius of 2.3 to 4.0 cm (depending on qX) where the 0.92 the central Ti. ion temperature is essentially constant at about If the ion temperature profile broadens, the region Page 7 where the maximum production shifts with it while the maximum production region remains constant. derived from the half width of the neutron the temperature of Thus the ion temperature spectrum is insensitive to plasma profile effects. Equation 5 does not however, oscillation which temperature. causes both include spatial the effect and of the time variations 'sawtooth' in the ion The sawtooth oscillation is caused by a plasma instability. In a short period of time the ion temperature profile flattens as in Fig. 5a. Figure 5b shows the neutron production profile and neutron spectrum just after a sawtooth crash. When the effect of the sawtooth oscilla- tion is included the correction factor for the spatial effect is 0.84. Combining the spatial neutron spectrum and width is 0.81 vary as much as 10%, time effects, times the measured Ti the peak central Ti. for arbitrary time limits, from the This might but varied less than 5% for the measurements presented below. Table 3 summarizes values of the mean and dispersion for some plasma temperatures. Clearly, the shift of the mean energy is small com- pared to the change in the line width, and measurement of the line width is better suited to a temperature determination. Using the results of table 3 for Alcator C ohmically heated plasmas for which the ion temperatures are typically 1.0 keV, a detector resolution better than 2.5% is required. choice of the detector. General This requirement greatly limits the discussions can be found in references [131 and [14]. found to have a resolution better than 2%. of neutron spectrometers The ion chamber used here was While many other techniques Page 8 have been proposed to measure the widths for higher temperature plasmas, [2], [161, [151, [171, [18], [19], no other technique which has a sufficient resolution without a sacrifice several orders of magnitude. of spectrometer, 400 msec), As will be seen below, has been found of efficiency of the low efficiency the short duration of the plasma discharge, and the low neutron production rate make it (typically necessary to sum spectra from many plasma discharges to obtain a useful spectrum. [I. The Spectrometer System The spectrometer system shown in Figs. 6 and 7 consists of the high pressure 3 He ionization chamber [20] designed for the Alcator environment. ber has provided a thermal is in a shielding enclosure The manufacturer of the ion cham- neutron shield This assembly ion chamber. mounted surrounded which is mounted around the by a minimum of 1.0 cm of lead inside a double walled stainless steel box. The space between the walls of the Li box is filled with lightly packed 2 CO 3 powder access port has been provided on one end to view the plasma. and an The port is blocked by a 1 mm thick sheet of cadmium to reduce the incoming thermal neutron thickness of flux. 30 The stainless cm of Water steel box is Extended Polyester is a mixture of 40% water, and 60% resin, or L1 2 CO 3 used thermal neutrons. to capture surrounded by a minimum (WEP) The WEP with small amounts of boron A collimator, of WEP is used to limit the view of the spectrometer which pass through the "keyhole" [21]. also made only to neutrons of an Alcator C diagnostic port. This port allows a view of 4.4 cm by 26.0 cm through the 20 cm thick Bitter magnet used to provide the toroidal magnetic field. Figure 8 is a block diagram of the spectrometer electronics. The Page 9 ionization chamber has an integrally mounted FET charge amplifier. The preamplifier output is An 8 psec time shaping amplifier. routed sensitive pre- via a coax cable to a constant is used for pulse shaping. The amplifier output signal is direct coupled to a multi channel analyzSpectra are er (MCA). gain of 4096. of the collected into 2048 channels with a conversion A precision pulser provides a test signal onto the grid ion chamber. The capacitively coupled signal at the the ion chamber is monitored to indicate noise on the signal. litude of the pulser signal is adjusted anode of The amp- to place the pulser peak in a channel above channel 1900, well above the range of interest for neutron spectrum counts. vide archival The MCA is interfaced to a PDP-11/T55 computer to pro- of individual shots. Two high voltage supplies are used to supply the grid and anode voltages. The grid voltage was set to 847 volts and the anode voltage was set to 3000 volts as recommended by the manufacturer. 2.1 The Ionization Chamber The ionization He, in detail The chamber is a gridded type operating at a pressure of 291. 3 chamber has been described elsewhere 6 atm of argon and 0.5 atm of CH 4 . neutron reacts via the 3 positive reaction energy, 3 atm of In the detection process the He(n,p)T nuclear reaction. Q, [22- of 764.0 keV. This reaction has a The cross-section for this reaction is shown as a solid line and the competing elastic scattering cross-section is shown as a dotted line in Fig. Fig. 9 represent Franz et al. [26]. the spectrometer Note that the relative 9. The open circles in efficiency scattering as cross-section reported by is roughly three times larger than the reaction cross-section at 2.45 MeV. Further Page 10 the reaction cross-section increases as the inverse square root of the Thus the spectrometer is very sensitive to thermal neutron energy. and scatter- epithermal neutrons and will have contribution due the elastic ing cross-section at all neutron energies. The calibration of the spectrum shown cross-sections in Fig. 10 illustrates the effect response function. on the detector source The neutrons were produced by bombarding a deuterium target with a 140 keV This source spectrum has an estimated width of 50 keV at deuteron beam. There is a large component of ther- FWHM and a mean energy of 2.42 MeV. due mal neutrons walls and results the ion chamber, of thermalized zero thermal peak at a by a) from a neutron the room energy full their full which deposit from reactions b) neutrons characterized spectrum is The floor. peak which the return to energy in energy due to the large reaction cross-section for thermal energies, c) a recoil continuum from 0.75 times full energy the on the neutrons Hydrogen in the CH 4 neutron capture and hard gas, total energy deposited 3 He(n,p)T reaction, to 3 a recoil He gas d) e) and, x-rays keV, interacting and of the spectrum due to the y ray continuum which is a in the ion chamber 764 scattering elastic energy due neutron the in the ion is the incident due chamber. to The sum of the Q of the neutron energy. Thus the thermal peak is offset by 764 keV and the signal due to gamma rays appears below the thermal peak. reaction cross-section active volume produce but the do 2.4 MeV neutrons which react not continuum deposit between their full energy the full energy via the in the peak and the first recoil feature. Figure 11 shows a crossectional view of the known parts of the ion Page 11 chamber. [301. The operationr of a gridded ion chamber is described in Knoll In the commercial design the preamp is mounted at one end of the ion chamber to minimize pickup the preamplifier. filling the The other between the ion chamber and the FET of end of the chamber through a gas fill heater has been provided to eliminate chamber valve. provides a means of A calcium purifier and tritium from the ion chamber after extended use. The manufacturer has included a thermal neutron of 1 mm of shield consisting cadmium and 3 mm of boron nitride powder around the active This shield is claimed to have a shielding effec- ion chamber volume. tiveness of 200 by the manufacturer. 2.2 Shielding and Collimation In order shield the code [31] to assess scattered was the amount neutrons used to and type of the Monte Carlo study the neutron shielding Neutral transport required Particle for to (MCNP) the Alcator C machine geometry. The machine was modeled as a torus with the surround- ing Bitter plates. magnet tallies were made from the port. as a Figure A single diagnostic function 12 shows of little increase in distortion the 2.3 to of scattering from copper, The bulk of the scattered a positions cm from the plasma. the 2.35 MeV for was modeled and at and away the neutron spectrum calculated position at a diagnostic port 50 is very energy port 2.45 group major neutrons MeV would constituent source Note that there group. correspond of the are at energies for a to The small 180 degree Bitter magnets. less than 0.1 MeV. Because the ion chamber is very sensitive to thermal and epithermal Page 12 neutrons these results indicate that it is necessary to increase the amount of shielding over the manufacturers thermal neutrons shield. ther, in measurements Fur- done during plasma discharges with only the manu- facturers shield the detector was saturated by the thermal and epithermal The code predictions neutron flux. indicated that the greatest number of neutrons are in the epithermal range, where the capture cross-sections are still ineffective. Thus, the decision was made to design the shield system to moderate and then capture the thermal neutrons. A second consideration was to minimize the that the spectrometer viewed as little flux as possible. collimator opening of the machine scattered neutron The horizontal access to the plasma is limited by the width of "key-hole" of the diagnostic port to approximately 4.4 cm. vertical access so is limited to 26.0 cm. It was shown earlier The that the active length of the ion chamber is 14.9 cm and that the ion chamber's efficiency at detecting chamber orientation placing the ion neutrons as end opening opening 10.2 cm vertically. the ion as long MeV chamber neutron producing was This is 20 degrees without loss of designed vertical viewed a large part of the insensitive on configurations chamber at an angle of a minimum collimator The collimator 2.45 to be opening portion plasma. of the ion are avoided. By as shown in Fig. viewed, or 5.3 1.2 cm of percent the of total the 402 5.08 cm horizontally was large the About total cm 7, efficiency was obtained. enough radial ninety that of the percent of the The horizontal circumference neutron so and extent neutrons are produced inside a minor radius of 6.0 cm. limit allows to producing plasma to plasma. be The collimator opening is blocked by a 1 mm thick sheet of cadmium to remove thermal neutrons from the incoming neutron beam. The shield box was Page 13 designed so that there were no streaming paths. Since the chamber is sensitive to gamma rays Li2 CO3 powder was used While natu- as a thermal neutron capture material near the ion chamber. ral lithium in this form is not as effective as boron compounds the capture of a thermal neutron by boron leads to a capture gamma of 0.48 MeV while the lithium capture reaction does not have a gamma ray associated with it. The Li 2 C0 3 powder was lightly packed between the two stainless steel boxes. The packing density of the powder is 0.76 which is g/cc calculated to reduce thermal neutrons by at least two orders of magnitude. This thickness would correspond greater thermal shielding is to a half value layer at 0.5 MeV. required, Li 2 CO 3 enriched in 6 Li If could be used. Water Extended Polyester was chosen as the moderator because: 1. Shielding of any desired shape and size could be cast in house allowing a modular shield system to be used, 2. water which is trapped in resin cells, to a 1.0 atom percent. up to 40% of the weight and 3. could be Boric acid is soluble up Li 2 CO 3 powder could be added also but had to be suspended in the resin making casting much more difficult. The shielding effectiveness of a 0.3 m thick block was tested by measuring the neutron spectrum for PuBe and Cf252 sources with and without a shield block in place. The shielding effectiveness was found to be approximately 30 for neutron energies neutron energies. above 1 MeV and to increase to about 100 at thermal The WEP shield has a total weight of 1070 kg. The ion chamber is microphonic. To isolate the ion chamber approx- imately 50 kg of lead shielding was tightly coupled to the ion chamber, while the assembly was shock mounted on rubber cushions. One inch thick Page 14 pieces of soft piece of Cd pickup. foam rubber blocking the were used on each collimator When the WEP shield side of port to the further 1 mm thick reduce acoustic was in place around the shielding box no evidence of acoustic pickup was seen during acoustic noise tests simulating shot acoustics. Franz et al. Acoustic sensitivity was studied in detail by [261. III. Spectrometer Performance Energy Calibration with the D(D,n) 3 He Reaction 3.1 The energy calibration of the ion chamber was made with a D-D neutron generator. The accelerator is capable of to 140 keV with a beam current of 0.3 mA. accelerating Under these 5.0 x 107 neutrons are produced per second. deuterons conditions up to The spectrum was neasured at a beam to neutron emission angle of 99 degrees where the dependence of the neutron energy on the ion beam energy is at a minimum. angle of 99 degrees the energy of the neutrons is 2.42 MeV. At this The error in this energy is estimated to be 10 keV and due to error in the measurement of the angle. In order T, on the detector beam angle scan. In constant to determine the effect of Fig. is resolution 99 degrees. 13, one can a series Figures see of 13 and the decreased the response of the amplifier time spectra 14 qualitative function stant. measured the results effects. of As the at a this time in the vicinity of the peak becomes more asymmetric and the FWHM increases. fect as a plot of the increase show were constant, Figure 14 shows this ef- in the FWHM as a function of the time con- The data has been normalized to the 8 pisec value after which the Page 15 relative width appears to be constant. To maximize the possible count Here 8 psec the shortest possible time constant should be used. rate, has been used since in most data taken to date the 8 psec time constant has not resulted in excessive pile-up. Shorter time constants used if the loss of resolution is not important and if can be the asymmetry of the response function is corrected for. 3.2 Energy Resolution Calibration neutrons Calibration using the were which has 0 value of -1644 keV. provided via the 7 Li(p,n) 7 Be Reaction 7 Li(p,n) 7 Be A proton beam from the Lawrence Liver- more Laboratory's Tandem Van de Graaff Accelerator was used. trometer was placed mean energy of at a reaction beam angle of 0 degrees. 4130 + 25 keV and a dispersion of The spec- The protons had 3 key. a These condi- tions imply a neutron energy of 2450 + 25 keV for the full energy peak. The resolution of the spectrometer mounted in the primary shielding box was measured for neutrons incident on the spectrometer at angles of 20 degrees (parallel case) 70 degrees to allow measurement ciency of the spectrometer. perpendicular case. case) to the of the relative efficiency of the cases and an estimate of the intrinsic effi- parallel and perpendicular placed at a distance of (perpendicular The charge of the protons striking the target ion chamber anode wire. was collected and The active center of the spectrometer was 70.2 cm for the parallel case and 113 cm in the A LiF target of thickness 112 pg/cm2 (8 keV) on a 40 mil tantalum backing was used. Figure 15 shows the calibration spectrum for The data has been fitted with the function: the parallel case. Page 16 2 ch 1 P 2F(ch) = p1 2 e P c3 -0.707 + P 1P4W 1 4 P2 e ch P5 + P W 6 (P 1 W = - i Erfc - ch (6) - 2 3 para- Erfc is the Complementary Error Function and the p's are fit meters. Note that the background and exponential are cut off by a comfunction plementary error spectrum shown the full width Gaussian is 38.2 + 3 keV. keV. at the the of centroid For the of the component at half maximum (FWHM) The width of the fitted function is 45.6 + 3 The intrinsic efficiency for the parallel be 1.7+0.5 x 10-4. Gaussian. case was calculated to The ratio of the intrinsic efficiency for the paral- lel case to the perpendicular case was measured by normalizing the ion chamber count rate to correct for the different tances and total run time. source to detector dis- The ratio was found to be 0.98+0.05. Thus the orientation of the ion chamber does not seem to be critical implying that the detector passing through at either end, of efficiency. rate is the active dependent volume. only on However, the number because of of the neutrons structure the ion chamber end configurations will result in a loss Page 17 3.3 Figures of Merit Because good resolution requires a long time constant and tokamak plasma shots are typically less than one second in duration, portant to quantify the maximum allowable count rate. it is im- A useful figure of merit is the ratio of the number of counts in the full energy peak to the total number of counts in the neutron spectrum. Formally, the inte- gral of the number of counts between the full width at tenth maximum is compared with the integral of counts in the total spectrum. It should be noted that for all these comparisons an arbitrary cut off is used on the low energy side. This lower limit corresponds the lower level discriminator to the cutoff on the multi channel analyzer. from This cut off would correspond to an energy of less than 500 keV deposited in the The energy ion chamber. could be The typical value for this figure every 70 detected pulses recorded from gamma or neutron interactions. of merit is 1/70. in the MCA is Thus, a full one out of energy peak pulse. A second figure of merit is the total count tion must made count rate. between the The MCA count ionization chamber the facturer and various authors rate is 104 the MCA two rates. there can -be signifi- The effect charge and gamma ray level will be discussed further imum count rate and Because the number of gamma rays varies significantly with the type of plasma discharge, between count Here a distinc- rate does not include the pulses cut off by the lower level discriminator. cant differences rate. [2,22,24,26] counts per have second. type of below. dis- The manu- indicated that the max- However, tronic pulse system which limits the count rate. the it is the elec- If the time constant of the pulse shaping is reduced the count rate limitation will increase. Page 18 (At a cost of resolution, as was shown in Fig. rate is also set. sensitive The MCA count to where 13.) This maximum count the MCA lower level discriminator is rate will not include pulses which fall below the lower level discriminator and the number of pulses increases sharply in the lower energy channels. Thus the count rate at significantly with the lower level discriminator the MCA setting. can vary Further the number of gamma rays and hard x-rays varies significantly with the type of plasma discharge, and the pulse amplitudes for photon interactions are generally below the discriminator level, so there can be significant variation due to the gamma ray background. Thus, the MCA count rate is always less than the rate at which radiation interacts in the ion chamber. The authors above have all used settings and amplifier time constants. pile-up on the peaks in the spectrum. the MCA discriminator level. the pile-up effects different lower discriminator More important is the amount of This effect is insensitive of For the discriminator level we have chosen, are minimized and the MCA dead time is below 2%. This would correspond to a total count rate of about 4 x 103 per second for the 8 psec time constant used. IV. Measurements at Alcator C A typical plasma discharge or shot for Alcator C is shown in Fig. 3. For the shot shown the field was 8 T and the plasma kA. current was 375 The maximum electron density was 3 x 102 0 m-3 . The neutron rate signal is derived from a bank of 14 one inch diameter 3 He proportional operational amplifier. counter tubes whose current is summed through an The shaded which the spectrometer is gated on. region indicates the time during Page Such a plasma discharge produces 1010 neutrons. 19 Unfortunately only about one out of every 7 x 109 of the total number of neutrons produced in the plasma contributes to the full energy peak due to solid angle losses and the low detection efficiency of the ion chamber. Neutron spectra were collected for individual plasma discharges and stored on magnetic media for later analysis. Typical 2.45 Mev peak counts rates were only a few per shot requiring a large number of similar shots summed together to be to obtain useful spectra. The best single analyzed to date had a peak rate of 2 x 1011 neutrons per second. shot During the 250 msec on time there were 11 neutrons collected in the full energy peak or a peak count rate of 44 per second. This is very close to the maximum rate at which the spectrometer can operate for the 8ysec amplifier time constant used. The dead time was estimated to be shot and the thermal peak was not distorted. time constant and using a pile-up rejecter 2% on that By reducing the amplifier a count rate as high as a 100 per second appears to be an upper limit on the count rate. In practice, it is necessary to sum a large number of similar plasma discharge spectra together In Fig. 16, to deduce useful information 303 plasma discharges have been summed together in order to understand overall properties of the spectrometer calibration spectra and a sum of a single day, shown. on the plasma. performance. 62 plasma shots are also It is clear from Fig. 16 that there is a peak at 2.45 MeV. ther there is little Two Fur- or no contribution to the spectrum from neutrons pro- duced by photonuclear reactions, electrodisintegration, photodisintegration or neutron down-scattering tively, the number of from the machine counts in an integral structure. containing Quantita- the peak to an Page 20 for both the integral of counts in the lower channels has been compared calibration spectrum and the shot data. In the first case, a D(D,n) 3 He calibration, the shield this comparison. including the WEP was assembled in the accelerator reaction but without the WEP moderator. to the high level of has one meter thick concrete The comparison walls. of The accelerator The nearest these WEP is to the shield effectiveness. In the second The fifty fold increase in the room return. vault wall was only 200 cm The source to detector distance away and the floor was only 60 cm away. was 96 cm. room. spectrum were taken using the D-D case in table 2 the integrals for a ratio is due the results of Table 4 shows tests how important indicate the A further test of the shield effec- tiveness was performed by blocking the collimator with a 10 cm polyethylene block during typical plasma discharges. A total of 11 neutron counts in the entire spectrum were recorded compared to an average of 86 counts recorded for the unblocked case. Assuming a mean free path of 2.5 cm for fast neutrons in the polyethylene block a 6% transmission is expected. calibration case the WEP moderator was not used, false floor For the Li(p,n)Be This. is consistent with the test results. allowing the spectrometer to be but the facility has a at least 500 cm from the nearest wall or floor while the source to detector distance was Thus the spectrum would be expected to be dominated 70 cm. by the direct flux. The small peak near 2.0 MeV is due to an excited state of Be. The inte- grals for a series of spectra sums are similar to the calibration results and indicate that the fusion spectra are dominated by 2.5 MeV neutrons. V. Determination of the Ion Temperature To deduce the ion temperature following procedure was used: from the line width of the peak the Page 21 1. The thermal and pulser peaks were examined on an individual shot basis to reject bad shots. Plasma spectra measured under 2. similar conditions were summed to ob- tain sufficient statistics. 3. The pulser and thermal peaks were fitted using a Gaussian function to determine the noise during the set of summed shots. 4. The full energy peak was fitted and the error of the fit was deter- mined as described below. 5. The fitted width was corrected for the detector resolution and for noise broadening of the peak. 6. The average ion temperature was determined using equation 3 and corrected as described earlier to obtain the central peak ion temperature. The primary criteria for rejecting a plasma shot was a loss of more than 4 counts from a normal 20 counts in the pulser peak during the 200 msec spectrum measurement of a plasma shot. The most probable cause of the loss of pulser counts was the pile-up due to hard x-rays and photoneutrons produced at the limiter where the most difficulty with pileup was encountered. A ring of molybdenum blocks is used to limit the extent of the plasma minor radius at two toroidal locations at Alcator C. level of photoneutrons gamma and rays is a strong function of how the gas is added to the plasma discharge at the start of the discharge. ther, in preliminary tests at a non-limiter port, The Fur- very few shots had to be rejected due to pile-up of gamma and photoneutron counts. The measure- ments reported below were done at a port with a limiter and great care was taken in the gas programming to avoid high photo-neutron and hard xray levels. Page 22 The summing of the data was accomplished by simply adding spectra on a channel to channel basis. This procedure was allowed only because the absolute channel drift was less than one channel per day and less than 3 channels per week. The pulser and thermal peaks were fitted in a least squares fashion with a single Gaussian function. The contribution to the peak width from ion chamber noise pick-up during the shot was typically 40 keV for a sum of shots. As mentioned above, spectra were recorded -into 2048 channels, approximately 70 channels across the full energy peak. procedure is to collapse the data until only or While the normal four to five contain the half width of the peak and then fit a function minimizing the Chi Square statistic, Cash if the original It [32,33] has suggested un-collapsed data is that more fitted information using the 'C' is gained statistic. is necessary to abandon the Chi-square statistic because of the re- quirement that the standard deviation of the number of counts channel be equal to the square root of the number of counts. in any This is not true when the number of counts is less than 10 as is the case here. The 'C' statistic is simply N C = 2 [ei - ni kn(ei)] where n is the number of counts in a channel, N is the number of channels and e is the expected number of counts from the fitted function. Confidence intervals are generated around the best fitted value. (7) Confi- by varying the parameter of interest The value of C for the minimum case is then subtracted from the C found for varied value. Cash has shown that Page 23 this difference is distributed statistically as a Chi square distribution with one degree of freedom. AC. Thus a confidence can be generated from the Figure 17 shows the confidence of fit as a function of Gaussian com- ponent sigma using the technique August, 1982. of Cash for data taken on the 6th of Such a curve can be used to determine the range of a about a best estimate of a given a required confidence. A 68% confidence in- terval has Using been used for error analysis here. this shots with as few as 20 counts in the peak have been analyzed. it technique However, was necessary to have at least 150 counts in the peak to obtain 20% error bars on the ion temperature deduced from the width. In order to assess whether the width could be used to deduce the ion temperature, two sets of plasma shots were analyzed. tions are summarized in Table 6. ature (low Ti) 800 eV and case the peak 1050 eV in The plasma condi- Note that in the low plasma ion temper- ion temperature the high Ti case. is estimated to be about These temperature estimates are based on the total neutron rate and neutral particle charge exchange Both diagnostics. uncertainties on these diagnostics the order of are 10%. reported The shot to have to shot measurement variation of 9 channels of these derived temperatures was on the order of 10 to 15%. Figure 18 shows the two data sets after groups of data have been added together for presentation purposes.- The functional form of the fitting function is equation (6) p5 fixed from the calibration. normalized to the this data which resolution, peak height has not been The optimal fit from the 'C' is shown for the low Ti case. corrected for the high Ti case has temperature is above with parameters p4 and the noise a greater width. or statistic Even in spectrometer The calculated shown plotted against the total neutron ion rate and charge Page 24 exchange derived temperatures ion measurements that Note agreement. represent perfect line The dotted 19. Fig. in would consistent are with the theoretically expected result. VI. The neutron Summary here has system described spectrometer measure the neutron spectrum at Alcator C. been used to The spectrometer resolution has been measured to be 46 keV FWHM and the absolute energy calibration By utilizing the structure of the Alcator C mag- is accurate to 10 keV. net in the collimator and shielding design an effective rejection of unwanted neutron and gamma ray counts has been accomplished. Furthermore, system is modular allowing it to be assem- the shielding and collimator bled at any Alcator C horizontal diagnostic port with normal laboratory All lifting equipment. materials primary shielding material, WEP, used are readily available the can be cast and machined with relative The ion chamber used is also commercially available. ease. While the spectrometer has very good energy resolution it be count rate limited to less than 100 counts second. [2]. and in the 2.45 appears to MeV peak per This is an improvement over the measurements of Strachan et al. the count For the measurements done here, rate was limited by the production rate from the Alcator C plasma and many plasma shots had to be summed to obtain a usable spectrum. rate in the plasma is increased it least several seconds of 'on' time. discharges would be required. several ion relaxed. chambers are used will Even if the neutron production be necessary to accumulate at Thus, at least 10 typical Alcator C If discharge durations are extended or if in parallel this requirement could be Page 25 The measurements done to date at Alcator C indicate that the speceasy to use in the tokamak environment. trometer is quite was magnetic shielding 50 P coaxial standard routed over incorporated cable and design, special signals the preamplifier from were to the the plasma noise pickup during Nevertheless, shaping amplifier. the in No shot contributed only an equivalent of 40 keV to the pulser and thermal peak widths and could be Table 5 provides corrected for. a summary of the spectrometer performance. By summing shots the temperatures from together widths it of has the been possible thermonuclear ion to determine This peaks. is the case known to us in which the ion temperature has been determined first from the peak width for an ohmically heated discharge and also the first systematic test of the technique against other ion temperature diagnos- The spectrometer is expected to be extremely useful for the study tics. of plasmas with non-thermal ion energy distributions. We would like to thank the support staff of the MIT Alcator C fusion group, Francis Bitter Magnet laboratory, for their assistance. the Lawrence Livermore the ion chamber, and MIT Reactor machine We would also like to thank Dr. Laboratory for his help in the D. shop Slaughter at calibration of and Frank Chmara of Peabody Scientific for his help in conjunction with the ion chamber. This work was supported by U.S. DE-AC02-78ET51013. Department of Energy Contract No. Page 26 REFERENCES 1. G. Lehner and F. Pohl, "Reaktionsneutronen als Hilfsmittel der Plasmadiagnostik", Zeitschrift fUr Physik 207 (1967) 83-104. 2. J. D. 3. J. D. Strachan and D. 4. W. A. Strachan, Fisher, et al., Nature (London) L. Jassby, Trans. 279 Am. Nucl. (1979) 626-628. Soc. 26 (1977) 509. "D-D Fusion Neutron Spectra Measurements and Ion Tem- perature Determination at Alcator C". M.I.T. Plasma Fusion Center Report PFC/RR-83-3. 5. "Studies of the Photonuclear Neutron Emission Phase of the Alcator C Tokamak", D. During the Start-up S. Pappas, R. Furnstahl, Nucl. Fusion 19 (1979) G. P. Kochanski, PFC/RR-81-22, May 1981. 6. H. 7. R.E. Chrien "Measurement of Fusion Reactions from a Tokamak Plasma", Knoepfel and D. A. Spong, 785-829. Ph.D. Thesis, Dept. of Astrophysical Sciences, Princeton University, 1981. 8. G. Lehner in Reactions under Plasma Conditions, Vol II, edited by M. Venugapalan, John Wiley and Sons, 1971, 509. 9. T. Elevant, Nucl. Instrum. Methods 185 (1981) 313-320. 10. H. Brysk, Plasma Phys. 15 (1973) 611. 11. M. M. R. Williams, J. Nucl. Energy 25 (1971), 489-501. 12. Fairfax et al., "Energy and Particle Confinement in the Alcator Toka- Page 27 maks", IAEA-CN-3/N-6. 13. Calvert and Jaffee in Fast Neutron Physics, Part II edited by Marion and Fowler, John Wiley and Sons, 1980. 14. G. Knoll, Radiation Detection and Measurement chapt. 15, John Wiley and Sons, 1980. 15. D. R. 16. H. W. Slaughter, IEEE Hendel et al., Trans. Nucl. Sci. NS-26 (1979) 802-808. "Tokamak Fusion Test Reactor Fusion-Reaction- Products Diagnostics" presented at the Fourth ASTM-EURATOM Symposium on Reactor Dosimetry, NBS, March 1982. 17. M. 18. TFR Group, in 8th Conf. Chatelier et al., Nucl. Instrum. (Unpublished) Methods 190 (1981) Controlled Fusion and Plasma Physics, 107-118. 1 (1977) 1. 19. R. A. Lerche et al., Appl. Phys. Lett. 31 (1977) 645. 20. Model FNS-1, Seforad-Applied Radiation, LTD., Israel. 21. Manufactured by Ashland Chemical, Ohio. 22. Instrument Manual for Seforad Fast Neutron Spectrometer, Model FNS-1 Seforad-Applied Radiation LTD. Emekhayarden, Israel. 23. J. Cuttler, et al., Trans. Am. 24. A. E. Evans, J. D. Nucl. Soc. 14 (1971). Brandenberger, "High Resolution Fast Neutron Spec- trometry without Time-Of-Flight", LA-UR 78-2562 (1978). Page 28 25. S. Shalev, J. M. Cuttler, Nucl. Sci. Eng. 51 (1973) 52-66. 26. H. 27. W. Franz C. et Sailor, al., S. G. Nucl. Methods Instrum. Prussin, Nucl. 144 (1977) 253-261. Instrum. Methods 173 (1980) 511- 515. G. Prussin et al., Nucl. Instrum. Methods 173 (1980) 28. S. 29. G. Rudstam, Nucl. Instrum. Methods 177 (1980) 529-536. 30. G. Knoll Radiation Detection and Measurement, p.178, 505-509. John Wiley and Sons, 1980. 31. MCNP - A General Monte Carlo Code for Neutron and Photon Transport, Version 2B, LA-7396-M revised April, 1981. 32. W. Cash, Astron. Astrophys. 52 (1976) 307-308. 33. W. Cash, Astrophys. J. 228 (1979) 939-947. Page 29 Machine Parameters to date Design unit Major radius 64 58 to 71 cm Minor radius 16.5 10 to 16.5 cm Toroidal field 140 30 to 130 kG Current 1000 100 to 750 kA Density 2x10 2 lx1019-1,3x1021 M-3 1 Aux heating 4.6 Ghz Lower Hybrid 4000 0 to 750 kW 150-200 Ghz Ion Cyclotron 4-10 0 MW Table 1. Alcator C Operating parameters. Page 30 Ti(rmax) Ti from width qlim rmax 2.00 3.5 0.928 0.863 2.50 3.2 0.924 0.860 3.00 2.9 0.926 0.858 3.50 2.7 0.925 0.856 4.00 2.6 0.921 0.855 4.50 2.4 0.924 0.855 5.00 2.3 0.922 0.854 5.50 2.2 0.922 0.854 6.00 2.1 0.923 0.853 Table 2. Correction factor versus qlim. Page 31 plasma temperature Table 3. mean energy width FWHM (keV) (keV) (keV) 0.5 2453 58.2 1.0 2455 82.5 1.5 2456 100.8 2.0 2458 116.4 Mean neutron energy and peak line width versus the plasma temperature. Page 32 Notes Case, File 19aprc.270 D(D,n) 3 He integral integral 1250-1400 800-1250 ratio 1.44 2896 4166 252 13107 16139 29946 1.86 846 1174 1.39 101 114 1.13 full shield 31janc.253 D(D,n) 3 He 52.0 no WEP, high scatter 18augc.282 Li(p,n)Be no moderator, low scatter super6.un1 sum of 303 plasma discharges 06augs.um5 sum of .62 plasma discharges 8-6-82 Table 4. Ratio of counts in peak integral to below peak integral. Page 33 resolution is 46 keV FWHM at 2450 at keV. 1. The energy 2. The energy 3. The contribution to the peak width due to noise pickup during calibration is accurate 2450 keV. to 10 keV at a plasma shot can be measured, allowing the peak width to be corrected for the noise A typical contribution. value at FWHM is 40 keV . 4. The in the ratio of counts full energy peak to total ion chamber counts is 1/70. 5. The maximum count rate to date is 44 counts per second in the full energy peak. count rate is The ultimate achievable D-D neutron peak estimated to be less than 100 counts per second. 6. The intrinsic efficiency of the ion chamber is measured to be 1.7±0.5x10~4 at 2.45 MeV. 7. The collimator magnet in its and shielding design. The design system is utilizes modular the in Bitter design, allowing the spectrometer to be used at any horizontal port. The shield weight is approximately 1 metric ton. Table 5. Summary of Spectrometer Performance. Page 34 "Low" Ti case number of shots plasma current toroidal field line averaged "High" Ti case unit 169 38 350-400 550-625 kA 110 kG 2.3x10 20 M-3 80 2.5x10 2 0 density Ti from neutron 780 1050 eV rate and charge exchange Table 6. Plasma conditions for "low" and "high" Ti cases. Page 35 FIGURES 1. Fig. The plasma is confined in The Alcator C fusion device. vacuum chamber by the effect of the toroidal magnetic field and the ohmic heating current. A typical plasma limiter consisting a of ring of molybdenum blocks is also shown. Fig. which 2. The neutron spectrum which would exhibits all the significant observed for a result neutron plasma producing reactions in an Alcator deuterium plasma discharge. Fig. 3. A typical high Alcator C plasma discharge as illustratTop, plasma current, ed by three plasma diagnostics. density from laser interferometer, average sents the Fig. spectrum time 4. during which expected The neutron spectrum was measured. Neutron production at a given plasma radius and neutron after integrating neutron production, solid line. the neutron produc- The shaded region repre- Toroidal field was 8.1 Tesla. tion rate. bottom, middle, line ion line, temperature is over the entire plasma. ion temperature assumed profile, Top, dotted to be zero at the plasma limiter, r = 16.5 cm. Bottom, obtained by subtracting 2450 keV from the calculated neutron energy. neutron spectrum. The width is Fig. 5a. Model of the plasma ion temperature distribution before and after a sawtooth after crash. crash. Solid line, before crash, dotted line rs is the surface at which the plasma q = 1. Fig. 5b. Neutron production rate versus radius and corresponding neutron spectrum for a plasma just after a sawtooth crash. Page 36 view of neutron spectrometer Cross-section is through the 6. Fig. device. Top of mid-plane plasma the total 1% of approximately this geometry, next Alcator the to In machine. the C volume is viewed. Fig. 7. Side view of the neutron spectrometer. Fig. 8. Spectrometer electronics. 9. 3 (n,p) NIM Standard have electronics been used. Fig. elastic, 1e of ion chamber. efficiency and relative cross-sections [24]. Efficiency data taken from Franz et al. Cross-section curves from ion chamber users manual [20]. Fig. 10. produced from target. where the MeV. The Ion striking a deuterated beam keV deuterium a 140 is 2.42 0.05 of about with a dispersion MeV titanium to the beam direction spectrometer was at an angle of 99* neutron energy were neutrons Source function. response char-ber The pulser and thermal peak features have been reduced in magnitude for presentation. Fig. chamber 11. view Crossectional structure which ion of use the inhibits of the illustrating chamber in an chamber the ion ion end-on configuration. monoenergetic MeV 2.45 Carlo Neutral Particle) spectrum a spectrum neutron Fig. 12. Calculated source using Alcator C distributed code position is for 13. Qualitative for the 50 cm the MCNP geometry. from the plasma, a from resulting at (Monte The a diagnostic port. Fig. effect of the pulse amplifier shaping time Page 37 were Spectra constant. with a taken 2.42 MeV and an energy dispersion of about 50 keV beam on deuteron deuterated titanium a energy source neutron 140 a using The gain target. of keV of the amplifier varies with the time constant. 14. Fig. the peak FWHM normalized to the 8 psec time constant of width the The effect of the source width (-50 case. constant on of the amplifier time Quantitative effect keV) has not been removed and is roughly equal to the detector resolution isec case. for the 8 Fig. 15. Calibration peak and fit due to the Li(p,n)Be reaction. A thin Li target was used so that the soltrce dispersion is less than 25 keV. An 8 psec amplifier time constant was used and the FWHM = 46 key. Fig. 16. Neutron spectra. reaction. left, Bottom Bottom, of the 62 discharges plasma The small peak in the Li(p,n)Be case is due to an a single day. excited state of Be. that right, D(D,n)T Top, right, a due to Li(p,n)Be reaction. sum of 303 plasma discharges. a from Top left, spectrum neutrons indicating Note the similarity of the spectra from an Alcator plasma are dominated by the D-D 2.45 MeV neutrons. Fig. 17. Gaussian Confidence the fit component of the fit. been used here. by using of the Standard values (one of a at for a given o (width) The method outlined by W. sigma) rhe 68% uncertainty percent can be confidence of the Cash has obtained level. Page 38 Fig. Ti cases. 18. Nine channel collapse of peak data for low Note the greater width for the high Ti case (top) even in data which has not been corrected for tion or is the system fit and noise. obtained The using curve the on the C-statistic the high evident detector resolu- low Ti data, (bottom) minimization, normal- ized for collapsed data magnitude. Fig. 19. Reduced width of the 2.45 MeV fusion peak and corres- ponding temperature plotted determined from total neutron rate the exchange diagnostics. against the central ion temperature and neutral particle charge The dotted line shows the expected dependence. The scale on the right can be used to convert the width to a central ion temperature. CP 0- cr 0 Co 0 O LU Eo cc C0 -+- cic 4-Q - 97Mo (yn) 96 MO D (y,n)H and D (e,e',n)D dR dEn D(Dn) 3 He T(D,n) Down Scatter of D- T Neutrons 2I I 3 4 5Ne 12 I2I 13 I PFC- 8355 FIGURE 2 I 14 15 16 17 D, k He / n) ron Energy (MeV) Neut D 4 SPECTROMETER ON SHOT 62 P lasm a Current ea k Va I ue 3 7 5 kA ne Average D ensity ak Value W O Ym 3 N eutrons R elative S ca le ALCATOR C B 8.1 T 6/22/82 .................... .. . .. ... .. .. ... .. .... .. .. ............. ......................... ... .. .. . . .. . I . .. .. . .. .. .. .. . . ... . .. ... ... ........ ........ ....................... .............. ............................ ........................... ...................... .................... ................... ................. .................... ...................................................... ....... .... ........................................... .. .. .. .... ..... .. .. ... ... .... ... .. .... .,*... ........................................ ......................... ..... .. .. .. ... ...... ... ... .... ........................................ .............. .......... ............. .................................... ...................... .................. .......................... X. ................. 1:,.*.,.,.,.*.,......-.-.-.-.-..........-....-,...,............, ... . .. ... . . .. ... .. .. .. ... .... .. . ... .. .. ...... ........................................ ....... ............. .................... ....................................... ... ... ... ... ... .*..... ... ... .. .. .. .... ............ ........ ................... .............. .......................... ................... ..... ............ ....... .. .... ...... . ................. ............. . ..... ............ ............... ........................................ ............ ........................................ .. ...................................... .... ..................... .................... ........................... .................... '7*:*7 --7-- '7 .-7 (155) (355) TIME (msec) PFC-8118 FIGURE 3' d Rate/ d radius for qlim 3 I 4 C 0 4- Ti (r) a) 4- -o 0 2 I- 11 -. 20 15 10 5 plasma radius (cm) 1.0 Neutron Spectrum for qlim 3 0 0 10 C: CL) 0 0.I 0.0 0 20 40 width 60 80 100 (keV) PFC -81/6 FIGURE 4 rs rf - r2,n 2 Before Crash Tr N=T ( ) et 2 Ti (0) (16.5) -1.5 at qo fT (0) -- Af ter Crash 0 f lat Profile Out to rf, magnitude fTi(0) Plasma Minor Radius PFC - 8355 FIGURE 5a dRate /d radius for qlim 3.95 I I I 10 15 41 C: 0 0 :3 -0 0 4- 0 a:: 2 L.. O C 5 20 plasma radius (cm) Neutron spectrum qlim 3.95 80 4- C 0 0 -o 0 'I) 0 4- C-) 0 0. 'C w a:: 60 40 - - 20 0' C 20 40 60 80 100 width (keV) PFC FIGURE 5b - 6352 E0 .0-0 00 -oo -o E ta. 0 qJ *0 0) VIhm 0.4-to W>h ~~0 4- 0 . plm 0 '0 0 C, 0) 0) a:: w 0 w z z L)0I CL a- aa- U) Ul) w (0 w 0 0 0 z _I__LU o w a- Li C) -z 11 ________________________ (9 a-. w a.- --IT U) z 0 D 0- a: w z w Q9 T LL 0 w H .:i 1.2 2.4 -n 0 2.0 r 0.9 0.8. E S-1.6 1.2 uj 0.6:11 *5-0.0 ijO0.3 0 00 0 0.2 - Relative 0 1ii1 1 11 . 0.4 0 Efficiency 1 - 0 1 11 1 1 1 101 2000 3000 NEUTRON ENERGY (keV) 1000 PFC-81/3 FIGURE 9 0 0 1 1000 Thermal Peak 800 Pulser Peak +20 +20 <600 - 4-HORci He 3 Recoil Full Energy Peak 0 400 Proton 200 -20 0 Recoil 500 1000 1500 Channel PFC- 8354 FIGURE 10 2000 w >< c 0 cr (D z z w (\j w '4- -LJ y i- (I) z -J 0 D i0 C,) 4 (1) I WL z D -J >0 0 --N (9I IE hLL J z LJJ LL. II II E < II wEw F- q - II w 0 II C') II Sa:z w T w cn It 0- w cc z aw a: CL 10- 7 - I io- 8- Z) -I II 0 0.01 0.1 I 4.- 2.3 2.35 24 2.45 1.0 MeV FIGURE 12 PFC-8350 120 , 140 100 - 8 /isec 120 100 U) 80 F- z 80 60- 60 0 400 40 1250 20- 1 20 0 120 0 120 I 100- 4 psec - 4 /.sec 100 5sec (n 80- 80 z 60 7 40 60- 0 400 20 20 0. 1050 -I 1250 1150 CHANNEL 1350 0 1050 1250 1150 CHANNEL 1350 PFC-8/22 FIGURE 13 I I @4 - N~ ~4. 04 00 C) .4. .4 4. 4 co rNO00 HIGIM BAIiV-1JH wi LL- I 600 1 . Li (p,n) Be 500400300 46 keV FWHM 200100 0 2300 2350 2400 2450 2500 2550 NEUTRON ENERGY(keV) PFC-8/55 FIGURE 15 0~ U)cI co 0 U) 0 0 E 00 0 0 0O - 0 0 cc -0 0 00 roC0 SI~noo 0 1.0 * ILjj 0.8 -* 0 z wOi .6 0z 0 zO4 .0 -- 0 * 00 0 _ 0 0 18 19 0 15 16 17 20 21 22 23 o- IN CHANNELS PFC -8/24 FIGURE 17 24 40 I I I 30H C Co 0 20 I C) 10 -IF~ 0 2200 50 I 2400 0 0 30 / 20 74 10 "II m4f 1< 0 2200 Thml~b 2600 I i 40H Co I I 2400 Neutron Energy 2600 (keV) PFC FIGURE 18 - 6356 o z (D o --. I co o( o o0 K 0 U. 0 N CL 0 0 0 a) LU 0 0co 0 0 0 0 (D Wmi 0 00 N -0o (A9M) o ) VJHMd I oo o 0 (0 . 00 LqlP!M peoflp98 t- 4- C a) 0