I/I In vivo Nitrogen Measurement by use of (n,2n) method and PGNAA by Aiko Ishikawa B.S., Applied Physics (1996) Waseda University, Tokyo Submitted to the Department of Nuclear Engineering in partial fulfillment of the requirements for the degrees of Master of Science in Nuclear Engineering at the MASSACHUSETTS INSTITUE OF TECHNOLOGY June 1998 © 1998 Massachusetts Institute of Technology. All rights reserved. Signature of author: Nuclear Engineering Department May 15, 1998 Certified by : gro/ Jacqlyn C. Yanch, Thesis Supervisor ofessor of Nuclear Engineering Department N 1 Dr. Joseph J. Kehayias, Tesis Reaer USDA Human Nutrition R search Center on Aging at Tufts University Accepted by : _ ----rf.fi rence M. Lidsky Chairman, Department Comittiee on Graduate Students Chairman, Department Scier",. - OlGe In vivo Nitrogen Measurement by use of (n,2n) method and PGNAA by Aiko Ishikawa Submitted to the Department of Nuclear Engineering on May 15, 1998 in partial fulfillment of the requirements for the degrees of Master of Science in Nuclear Engineering Abstract. The technique of in vivo neutron activation analysis has been used for the determination of various elements in the human body. There are two methods of measuring nitrogen in vivo: (n,2n) method and prompt gamma neutron activation analysis (PGNAA). The first method utilizes the fast neutron reaction '4 N(n,2n) 13N and the second uses the thermal reaction 14N(n,y)1 N. Both methods are currently in use at several different facilities for evaluation of the nutritional status of surgical patients or metabolic study of normal human subjects. In order for the absolute determination of nitrogen, it is necessary to obtain the uniform sensitivity of detection in the subject. Experimental factors which influence the sensitivity are the activating neutron flux and the absolute detector efficiency. The distribution of the two factors in the human body is dependent on the spectral characteristics of the neutron source, the geometries of irradiation and counting systems and the size of the subject. In this thesis investigations of sensitivity and absorbed dose of the two nitrogen methods are made based on the results of Monte Carlo simulation. Two experimental facilities utilizing the (n,2n) method and the PGNAA method are simulated to assess the suitability and the feasibility of in vivo measurement of nitrogen. The potential of the use of an electrostatic tandem accelerator as a neutron source for nitrogen measurement is examined by simulating hypothetical experimental set-ups. Furthermore, those results obtained by the simulation are verified by performing accuracy measurements by the (n,2n) method. The counting system designed for human thigh measurement is described. The (n,2n) method delivers about half of the equivalent dose of the PGNAA method with same irradiation time. The simulation results showed 10 - 20 times higher sensitivity for the (n,2n) method than the PGNAA technique, depending on the patient position for the irradiation. The simulation results for the accelerator sources presented higher sensitivity than the 238Pu-Be radionuclide source conventionally used. Thesis Supervisor: Jacquelyn C. Yanch Title: Professor of Nuclear Engineering Department Acknowledgments I would like to thank Professor Jacquelyn C. Yanch for her advice for my thesis and great patience with a number of mistakes I have made. I would also like to thank Dr. Kehayias who taught me basic concepts needed for researching. I am grateful to members of LABA group who have been a great help throughout my research work. It has been my great honor to work with such competent and diligent scientists. Table of Contents 1. INTRODUCTION 2. TBN MEASUREMENT 2.1 (N,2N) METHOD 2.2 PGNAA METHOD 8 12 12 17 3. NITROGEN MEASUREMENT USING A D-T NEUTRON 25 GENERATOR 3.1 OBJECTIVE 25 3.2 IRRADIATION SYSTEM 3.3 COUNTING SYSTEM FOR HUMAN THIGH MEASUREMENT 3.4 EXPERIMENTAL PROCEDURE 26 3.4.1 NITROGEN MEASUREMENT 3.4.2 MEASUREMENT OF DETECTOR EFFICIENCY OF THIGH COUNTING SYSTEM 3.4.3 INTERFERENCE OF OXYGEN 3.5 DATA ANALYSIS AND RESULTS 3.5.1 3.5.2 3.5.3 3.5.4 COUNTING RESULTS OXYGEN INTERFERENCE ESTIMATION OF DETECTOR EFFICIENCY AT 511 KEV ESTIMATION OF NEUTRON OUTPUT 4. MONTE CARLO SIMULATION 4.1 MCNP CODE 4.2 FIGURES OF MERIT 4.2.1 DOSE 4.2.2 SENSITIVITY 4.3 NITROGEN MEASUREMENT SYSTEMS 4.3.1 D-T GENERATOR - (N,2N) METHOD 4.3.2 238Pu-BE RADIONUCLIDE - PGNAA METHOD 4.3.3 ACCELERATOR SOURCE 9BE(P,N) WITH LABA SET-UP - PGNAA METHOD 4.3.4 VARIOUS ACCELERATOR SOURCES WITH BNL SET-UP - PGNAA METHOD 27 29 29 31 31 32 32 33 33 35 36 36 37 37 41 47 47 53 62 68 4.4 RESULTS 70 4.4.1 DOSE 70 4.4.2 SENSITIVITY 4.5 SUMMARY 5. COMPARISON AND CONCLUSION 86 100 101 5.1 ACCURACY AND REPRODUCIBILITY 5.2 IRRADIATION AND COUNTING TIME 101 5.3 DOSE 106 5.3.1 INCIDENT DOSE 5.3.2 DOSE DISTRIBUTION 5.4 SENSITIVITY 5.4.1 SUPINE- AND PRONE-POSITION IRRADIATION FOR (N,2N) METHOD 5.4.2 ACCELERATOR SOURCES FOR PGNAA METHOD 5.4.3 (N,2N) METHOD AND PGNAA TECHNIQUE 102 106 106 108 109 109 110 5.5 CONCLUSION 5.6 FUTURE WORK 111 112 5.6.1 (N,2N) METHOD 113 5.6.2 PGNAA METHOD 114 6. REFERENCES 115 Table of Figures 12 Figure 2-1: (n,2n) method: Gamma-ray spectrum Figure 2-2 : PGNAA method: Gamma-ray spectrum obtained from a normal 18 human body 20 Figure 2-3 : Schematic illustration of the PGNAA facility employed at BNL 21 Figure 2-4 : The design of BOMAB phantom 26 Figure 3-1 : (n,2n) method: irradiation set-up 27 Figure 3-2 : (n,2n) method: counting set-up 28 Figure 3-3 : (n,2n) method: counting set-up with a patient Figure 3-4 : (n,2n) method: irradiation set-up for urea solution measurement 30 Figure 3-5: Nitrogen 511 keV y-spectrum obtained from a urea solution sample32 38 Figure 4-1 : The rectangular box phantom filled with a uniform solution 40 Figure 4-2 : Thigh model for MCNP simulation 50 Figure 4-3 : (n,2n) method: BCL irradiation system with the thigh model Figure 4-4 : (n,2n) method: BCL irradiation system with the rectangular box phantom Figure 4-5 : (n,2n) method: BCL counting system with the rectangular box 52 phantom Figure 4-6 : PGNAA method: BNL irradiation system with the thigh model for 57 calculation of dose distribution by MCNP 238 58 simulation the for Figure 4-7 : Pu-Be neutron yield data points used Figure 4-8 : PGNAA method: BNL irradiation system with the rectangular box 59 phantom for sensitivity calculations Figure 4-9 : PGNAA method: BNL detector positioning simulated by MCNP 60 Figure 4-10 : PGNAA method: one of the BNL detectors simulated by MCNP 61 Figure 4-11 : A tandem electrostatic accelerator installed in MIT's LABA is 64 shown with the ion source Figure 4-12 : Configuration of the LABA accelerator and experimental rooms 65 Figure 4-13 : PGNAA method: LABA irradiation system with the thigh model for 66 calculation of dose distribution by MCNP box rectangular the with system irradiation LABA Figure 4-14 : PGNAA method: 67 phantom for sencitivity calculations 69 sources accelerator various for Figure 4-15 : Neutron energy spectra 72 Figure 4-16 : Axial view of the thigh modelled by MCNP 74 model thigh in distribution Figure 4-17 : (n,2n) method: Total neutron dose Figure 4-18 : (n,2n) method: Activating neutron dose distribution in thigh model75 76 Figure 4-19 : (n,2n) method: Photon dose distribution in thigh model Figure 4-20 : PGNAA method with 238pu-Be source: Total neutron dose distribution in thigh model Figure 4-21 : PGNAA method with 238Pu-Be source: Activating neutron dose distribution in thigh model 77 78 Figure 4-22 : PGNAA method with 238Pu-Be source: Photon dose distribution in 79 thigh model Figure 4-23 : PGNAA method with 9 Be(p,n) neutron dose distribution in thigh model Figure 4-24 : PGNAA method with 9 Be(p,n) neutron dose distribution in thigh model Figure 4-25 : PGNAA method with 9 Be(p,n) dose distribution in thigh model Figure 4-26 : PGNAA method with 9 Be(p,n) neutron dose distribution in thigh model Figure 4-27 : PGNAA method with 9 Be(p,n) neutron dose distribution in thigh model Figure 4-28 : PGNAA method with 9 Be(p,n) dose distribution in thiah model source and BNL set-ups: Total source and source and source and source and source and 80 BNL set-ups: Activating 81 BNL set-ups: Photon 82 LABA set-ups: Total 83 LABA set-ups: Thermal 84 LABA set-ups: Photon 85 Figure 4-29 : (n,2n) method: distribution of number of 14N(n,2n)13N reactions in 88 the phantom 4-3015: PGNAA method with Figure 14 238Pu-Be N(ny) N reactions in the phantom source : distribution of number of 89 9Be(p,n) 9Be(p,n) 90 source and BNL set-ups: Figure 4-31 : PGNAA method with set-ups: LABA and source Figure 4-32 : PGNAA method with 1 14 91 in the phantom reactions 5N N(n,y) of number of distribution 93 Figure 4-33 : (n,2n) method: detector efficiency distribution 94 Figure 4-34 : PGNAA method: detector efficiency distribution Figure 4-35 : (n,2n) method: distribution of composite sensitivity in the phantom 96 for prone position irradiation Figure 4-36 : (n,2n) method: distribution of composite sensitivity in the phantom 97 for supine position irradiation source: distribution of composite Figure 4-37 : PGNAA method with 238Pu-Be 98 sensitivity in the phantom 9 distribution system: BNL Figure 4-38 : PGNAA method with Be(p,n) source and 99 of composite sensitivity in the phantom Figure 5-1 : Relationship between measurement time and figures of merit: 104 nitrogen counts and sensitivity 1. Introduction Determination of nitrogen levels in the human body provides an estimation of protein, a highly important and essential component of body composition [1-9]. Total body protein (TBP) has a good correlation with total body nitrogen (TBN) with the relationship TBP = 6.25 TBN [10]. Knowledge of TBP makes it possible to evaluate nutritional status of patients with progressive disease or metabolic disorders. Body nitrogen measurements are used to monitor their changing status and to assess the nutritional requirements for treatments [1, 2, 5, 9, 11-16]. Such disorders include cancer, protein malnutrition, renal failure, anorexia, cardiovascular disease, obesity, hyperthyroidism, liver and kidney desease, and growth deficiencies [1, 4, 16-19]. Diseases associated with muscle wasting (e.g. cystic fibrosis) can also be studied by the quantitative estimation of nitrogen [1, 4, 8, 17, 20-22]. Furthermore, estimation of total nitrogen allows the study of the effects of age, sex, race, and body size on body protein levels in normal subjects [16, 23]. The technique of in vivo neutron activation analysis (IVNAA) is currently used by an increasing number of biomedical research centers as one of the most effective methods for quantification of various elements in the human body. There are two nuclear methods that can be employed for determination of TBN by the use of the IVNAA technique; one is the '4N(n,2n) 13N method [5, 11, 24-26] and the other is prompt gamma neutron activation analysis (PGNAA)[2, 6, 8, 14, 20, 22, 27, 28]. Both methods take advantage of the fact that when a subject is irradiated with neutrons in a certain energy range, nitrogen in the subject will be converted into a radioactive isotope. The produced isotopes will decay at a characteristic rate and emit y-rays which are used to identify and determine the concentrations of nitrogen present in the subject. Fast neutrons are utilized for both methods, although the prompt gamma reaction 14 N(n,y)15 N occurs predominantly for neutrons of thermal energy which result from thermalization of the incident fast neutrons by the body tissues or by a separate moderator. The major restrictions with the IVNAA techniques for the absolute measurement of nitrogen in vivo are nonuniformity of neutron flux through the body and the dependence of counting efficiency on the position of nitrogen nuclei in the subject [2, 5, 7, 12, 16, 19, 22, 25, 26, 28-35]. The application of either IVNAA technique, therefore, had been limited to sequential, relative measurements until the late 1970's, when the PGNAA method was modified to make estimations of absolute quantities of nitrogen mass by Vartsky et al. [2, 12]. In this new method, total body hydrogen (TBH) is also measured as an internal standard based on the fact that the ratio of N/H counts, unlike TBN alone, is insensitive to body size. While sequential measurements of nitrogen are useful for surgical patients following various therapeutic protocols to monitor the change of body nitrogen level and to evaluate their response to the surgery, the absolute determination makes it possible to compare individual patients and thus to have an even broader range of clinical research application than we would have only by sequential measurements [12, 15]. An overview of TBN facilities currently in use is presented in Table 1-1 [8, 9, 16]. Facility Leeds, UK SURRC, Scotland Birmingham, UK BNL, US Auckland, New Zealand Swansea, UK Edinbugh, UK Toronto, Canada Sydney, Australia --- Neutron source References [11, 23, 36-38] PGNAA 14 MeV D-T generator 14 MeV D-T generator 10 MeV p-Li cyclotron 238Pu-Be PGNAA 238Pu-Be [2, 12, 13, 18, 35, 41, 42] [14] PGNAA 252Cf [28, 34] PGNAA 252Cf [7, 22] PGNAA 252Cf [3, 6, 20, 43] PGNAA 252Cf [8, 44] Method - (n,2n) or PGNAA (n,2n) (n,2n) PGNAA [25, 26, 39] [1, 27, 40] Table 1-1 : Overview of TBN measurement facilities currently in use There is another nuclear technique for TBN measurement other than the IVNAA, which is nuclear resonance absorption (NRA) [9]. This method, based on the gamma resonance absorption reaction 14N(y,p)13C, was first developed as a means of explosives detection in materials transported in the airport environment [45]. In this method, 9.17 MeV y-rays are resonantly, and uniquely, absorbed by nitrogen in the patient and thus nitrogen-specific resonant radiography can be achieved by imaging the transmission profile of the photons having 9.17 MeV energies. The feasibility of the use of the NRA method is currently under investigation. It is expected that this technique will offer the advantages of low dose, good activation uniformity, and simplification of the absolute measurement procedure. A major objective of the work described in this thesis is to assess the suitability and feasibility of the two IVNAA techniques: the (n,2n) method and the PGNAA method, for TBN measurement. Monte Carlo calculation is used to simulate the two experimental facilities which are currently in operation and to compare sensitivity and the absorbed dose of each system. The potential of the use of an accelerator as a neutron source for nitrogen measurement will be also investigated by simulating hypothetical set-ups. Experimental verification is examined by performing nitrogen measurements by the (n,2n) method and the work done to improve detector efficiency for measurement of the human thigh will be also described. 2. TBN measurement 2.1 (n,2n) method In the 1'4N(n,2n) 13N method, the body is irradiated with 14 MeV neutrons produced by a deuterium-tritium (D-T) generator [11, 24, 25, 29] or with those produced by a cyclotron [5]. This fast neutron reaction (cross section = 5.7 mb) has a high threshold energy of 11.3 MeV and the created radioisotope 13N decays to 13C by positron emission having a half-time of 9.96 min. The nitrogen mass in the human body can be estimated by the detection of the 511 keV annihilation quanta which are produced in the medium as a consequence of the positron decay of 13N. Figure 2-1 shows the y-ray spectrum obtained from a patient counted for 6 minutes following 14 MeV neutron irradiation [29]. 5 GAMMA- RAY SPECTRUM OF PATIENT' WIL 6 min POST IRRADIATION - q = 9 x 104 n cm-2 sec - I I 10 I I I I ' (ANNIHILATION PEAK) 0.51 MeV U2i I- 3CI :3 4 Z No (.64 MeV) "A 10.78M.V) 1.37 MOV 38CI 2.16 MeV 2.75 MV ) S103 10 20 30 40 50 60 70 80 90 I00 CHANNELS (33 keV/CHANNEL) Figure 2-1 : (n,2n) method: Gamma-ray spectrum obtained from a patient counted for 6 minutes following 14 MeV neutron irradiation [29]. 110 There are two major problems with this method which have been discussed since the early 1970's when the (n,2n) method was introduced as the first technique for nitrogen measurement in vivo. One problem comes from the fact there are no characteristic y-rays emitted by this reaction and the 511 keV photon peak represents not only nitrogen counts but also other positron emitters produced by fast neutrons in the human body, such as, oxygen, phosphorus, potassium, and chlorine [5, 11, 25, 26, 29, 31]. Furthermore, the y-rays emitted in nuclear reactions can undergo Compton scattering or pair production which will lead to additional contributions under the 511 keV peak. Table 2-1 shows nuclear reactions that might give rise to interference with the nitrogen measurement. Some of the interference can be reduced by making adjustments with the counting time according to the differences of their half-lives. For example, the irradiated subject can be counted long enough so that 30P, having a shorter half-life than '3 N, 14N(n,2n)' 3N Cross section [mb] 5.7 Half-life [min] 9.96 Threshold [MeV] 11.3 Principle emissions P+ O 160(p,C) 3 N 19-58 9.96 5.5 p+ P 31P(n,2n) 30 P 16.04 2.5 12.7 Cl 31P(n, c)2Al 35Cl(n,2n) 34mC1 121.2 2.82 2.3 32.0 2.0 12.9 37 Cl(n,y) 38 C1 433 1.6 530 102 1100 37.2 7.7 900 38.0 8.9 thermal 13.4 thermal 12 thermal Element Reaction N K Na Zn Ca 39K(n,2n) 38K 23Na(n,y) 24Na 64Zn(n,2n)63Zn 48 Ca(n,y) 49 Ca Table 2-1 : Reactions which might interfere with nitrogen measurement by (n,2n) method + y (1.78) p+, y (1.17, 2.12, 3.3 MeV) y (1.64, 2.17) p , y (2.17 MeV) y (1.37, 2.75) P+ y (3.08) contributes too little to the total counts. In that case, of course, the compensation which comes from nuclides having longer half-lives needs to be taken into account. A number of investigators have estimated the extent of interference in measuring nitrogen using phantoms containing each of the elemens listed in Table 2-1 and no significant interference from those products except oxygen was observed [11, 26, 29, 31, 32]. The interference from oxygen is induced by a knock-on proton from hydrogen in the subject. Since the product from this reaction is same as that from the nitrogen reaction, there is no way to distinguish these two elements. Contributions from oxygen can be estimated by comparing 511 keV photons counted from a phantom containing nitrogen and a nitrogenfree sample. 12 - 21 % of oxygen interference is obtained by some workers [11, 26, 31, 32]. Another complication that makes it difficult to use the (n,2n) reaction as a method for absolute measurement of nitrogen in vivo is the lack of uniformity of sensitivity within extended subjects such as the human body [29-31, 33]. Vartsky et al. defined composite sensitivity as the number of counts detected from the body / unit mass of element under investigation (N) / unit incident dose delivered to the body [2]. There are three factors that influence the spatial variation of sensitivity through the body: 1) attenuation of the activating neutron flux in the body, 2) self-absorption of photons within the body, and 3) detector efficiency. The distribution of fast neutrons within the subject is dependent on the type of neutron source and the irradiation set-up which characterize the energies of incident neutrons and determine the absorbed dose which is one of the components of composite sensitivity. The type of detectors and the geometry of the counting facility are other factors which affect the sensitivity because they determine the detector efficiency of the system. Also, the sensitivity is significantly affected by the size of the subject because of the influence on the escape probability of photons from the subject and the penetration of the activating neutrons. The sensitivity, therefore, is quite dependent on the whole experimental set-up employed. Estimations of the sensitivity have been made by some workers using 14 MeV neutron generators for bilateral irradiation and counting . Vartsky et al. reported uniformity of the sensitivity of 48 % within a 25 cm-thick box phantom, from a fast neutron fluence experiment with a 100 cm target-to-skin distance (TSD) [31 ]. The effects of photon attenuation and detector efficiency were roughly estimated according to geometrical assumptions. The variation of the sensitivity was expressed as the difference of maximum and minimum values from their mean. Other experimental results of sensitivity were obtained by two research groups, the Scottish Universities Research and Reactor Centre (SURRC, East Kilbride, Scotland) and the University of Leeds (UK). A 31 % uniformity for a phantom of 23.5 cm was reported by Leeds [11, 33, 36] while SURRC obtained a variation of 24 - 54 % for 15 - 30 cm-thick phantoms with approximately three times longer TSD and different counting systems [26, 39, 46]. Despite the limitations inherent to the (n,2n) method, two above facilities have established the technique for absolute nitrogen measurement [11, 25, 26, 36]. In the SURRC bilateral irradiation was achieved by placing two D-T neutron generators above and below the patient. The Leeds system had one D-T generator which irradiated the patient from either side by rotating the patient couch through 1800. Both irradiation systems allowed maintaining the patient in the supine position, which is needed for measurement of critically ill patients. Calibration procedure taken for both techniques used three anthropomorphic phantoms of different sizes which represent various habitus of the human body [25]. The method is based on the assumption that the distribution of nitrogen and interfering elements in the human body can be considered to be averaged over the whole subject and thus the relative counts of those elements are similar to those of the phantom. Williams et al. (SURRC) found that the reproducibility and absolute accuracy were dependent on the size of the subject, with results in the ranges of 1.4 - 3.2 % and 3.0 - 4.3 %, respectively [25]. These results were obtained by using three phantoms with a dose equivalent of 10 mSv. The Leeds group performed nitrogen measurements using two phantoms by the irradiation from both side of the subject and reported reproducibility of 1.3 - 1.6 % and accuracy of 1.2 - 2.4 % with a dose equivalent of 5 mSv, which was 10 times higher than that used for patient studies. 2.2 PGNAA method As another effective technique for body nitrogen measurement, the PGNAA method was introduced at Birmingham in the UK [27] soon after the emergence of the (n,2n) method. Unlike the (n,2n) method, 10.83 MeV y-rays produced from the neutron capture reaction 1aN(n, y)15N* (cross section = 0.075 b) are uniquely characteristic of nitrogen and thus can be readily identified by spectroscopic analysis. The specific y-rays are emitted in a very short time (life time = 10- 5 sec) by the de-excitation of 1N* to the ground state with about a 15 % branching ratio. The thermal neutrons that contribute to this reaction are produced by thermalization of fast neutrons in the medium of the subject or by the use of a moderator through which incident neutrons pass before interacting with the subject. The limitation of this method comes from the high background counts in the nitrogen peak of 10.83 MeV y-rays. The y-ray spectrum obtained from a normal human body for a 36 minute measurement is shown in Figure 2-2 [14]. High background is due to the fact that counting and irradiation are performed simultaneously for prompt gamma reactions and photons created from other elements existing in the whole system contribute to the total counts [2, 6]. Because of the high characteristic energy, however, large NaI(TI) crystals are needed to ensure high detector efficiency. The background is also affected by the size and shape of the subject and thus so is the nitrogen signal. Three types of neutron sources have been utilized for the PGNAA method: a cyclotron [1, 27, 40], a 23 8Pu-Be radionuclide [2, 14, 20], and a 2 52Cf spontaneous fission a 10 a. S102 A 2 4 6 8 Energy (MeV) 10 12 Figure 2-2 : PGNAA method: Gamma-ray spectrum obtained from a normal human body for 36 minute measurement [14]. source [6, 22, 28, 47]. Despite the needs of replacement, the two radionuclide sources have gained more acceptance than the cyclotron because of the lower cost for construction, the accessibility from the clinical environment, the simplicity of the operation, and the lower absorbed dose [17, 20, 48, 49]. A 252Cf neutron source provides lower nitrogen background than a 23 8Pu-Be radionuclide because the background in the nitrogen region is mainly contributed from random summing ofy-rays in the energy range of 4 - 7 MeV [43]. A 2 8Pu-Be source decays with photons of 4.4 MeV which may be coincidently detected and contribute to the nitrogen region (9.5 - 11.1 MeV), while no yrays in the energy range of 4 - 7 MeV are emitted from a 252Cf source. It was also reported that a 252Cf source provides a 40 %larger thermal neutron flux per incident dose than a 238Pu-Be source with comparable uniformity [44, 50]. Furthermore, the transportation regulations of a 252 Cf source are relatively less stringent than a 238Pu-Be radionuclide [43]. The short half-life, 2.65 years, of a 252 Cf source, however, entails the need of more frequent replacement of the neutron source than if a 238Pu-Be source is used [7]. A special feature of this technique which made it possible to determine the absolute amount of nitrogen in the human body is the use of total body hydrogen (TBH) as the internal standard, which was proposed in the mid 1970's by Vartsky et al. at the Brookhaven National Laboratory (BNL) [2, 12]. This technique takes advantage of the fact that the ratio of nitrogen to hydrogen counts is less dependent on body size than nitrogen counts alone. Hydrogen in the human subject can be counted simultaneously with the counting of nitrogen by the prompt gamma technique using the thermal neutron reaction 1H(n, y) 2H (cross section = 0.33 b). Characteristic y-rays of 2.23 MeV are produced with 100 % yield per neutron capture. Non-uniformity of the thermal neutron flux in the body can be reduced by normalizing the nitrogen signal relative to the hydrogen signal. The ratio of the N/H counts observed in the subject is then compared with that obtained from a similar-sized anthropomorphic phantom with known amounts of these elements. This method is employed in practice for absolute measurement of nitrogen in a number of clinical facilities[6-8, 14, 34, 35]. Calibration procedures are basically same, although every facility has its own way of measuring nitrogen and hydrogen background, deriving TBN and TBH using equations developed, and correcting data to alter the difference between the standard phantom and the subject, according to the facility construction materials and the phantoms they used. In this chapter, therefore, the calibration technique utilized at BNL [35, 42] is described. Figure 2-3 schematically shows the PGNAA facility currently used at BNL. The details of the composition, location and dimension of the collimator assembly and detectors will be provided in chapter 4. To experimentally determine absolute nitrogen concentration in the body, the patient is irradiated over five 20 cm-long sections along the length of the body, starting from the shoulder, for 200 sec for each section. This procedure is repeated for the prone and supine position to achieve a bilateral irradiation and counting and consequently the nitrogen gross counts of the patient are obtained. detector premoderator (D 2 0) collimator 238 Pu-Be source Figure 2-3 : Schematic illustration of the PGNAA facility employed at BNL Patient is irradiated over five 20 cm-long sections along the length of the body starting from the shoulder, for 200 sec for each section. This procedure is repeated for the prone and supine position to achieve bilateral irradiation and counting. Measurement of the nitrogen background in the energy range of 9.5 - 11.1 MeV that corresponds to the nitrogen peak is performed using a BOttle Mannequin ABsorber (BOMAB) phantom shown in Figure 2-4. Three BOMAB phantoms of different sizes which are filled with nitrogen-free tissue-equivalent solution are available. One of the three BOMAB phantoms which has a similar size to that of the subject is measured so that the nitrogen net counts of the patient can be obtained simply by subtracting the counts observed in the phantom from the nitrogen gross counts observed in the patient. The net hydrogen counts are obtained by trapezoidal subtraction in the hydrogen region O0 O0 Fat* 0 Figure 2-4: The design of BOMAB phantom Fat layers surround the sections of thorax, lumber and thighs [35, 51]. (2.06 - 2.47 MeV) and subsequent subtraction of interference from hydrogen in the facility from that value. The interference is assumed to be a fixed percentage of the hydrogen true counts, because significant dependence of the body size on the interference was not observed from the measurements of D2 0-filled BOMAB phantoms performed by the BNL research group. The N/H counts ratio obtained from the measurement are compared with that obtained from the same BOMAB phantom in which nitrogen-free tissue-equivalent solution is replaced by solution containing known amounts of these elements. In addition, measurement of BOMAB phantoms with fat layers of various thickness (Figure 2-4) showed that the ratio of N/H counts decrease with increasing fat layer thickness. The method of estimating the thickness of subcutaneous adipose tissue has not been reported yet. Corrections between the two values are made for the size and for the thickness of fat. The BOMAB phantom is irradiated and counted with the same procedure as the patient, i.e., 200 sec for each five sections from the shoulder for supine and prone positions. The corrected N/H ratios are averaged using the volume of each section as a weighting factor. The absolute amount of body nitrogen is obtained from the relationship TBN = C x <N/H> x TBH where C: A proportionality constant determined from irradiation of the BOMAB phantom that was assigned to the patient according to the size <N/H>: An average size- and fat-corrected N/H value over five sections TBH is determined from the relationship TBH = 0.11 TBW + 0.12 TBF + 0.07 TBP where TBW: total body water (obtained by dilution of tritiated water technique [52, 53]) TBF: total body fat = wt - (TBW + TBP + BMA) wt: body weight BMA: body mineral ash = TBCa (obtained by delayed gamma activation analysis) / 0.34 TBP: total body protein = 6.25 TBN TBW can be estimated by administration of a small quantity of 3 H2 0 to the human subject and the radioactivity of tritium is counted for blood sample of known volume using a liquid scintillator. The correlation factor of the nitrogen-protein relationship, 6.25, comes from the general assumption that nitrogen composes 16 % of protein as a rough approximation [10]. A reproducibility of 2.1 % was reported for measurement of the N/H count ratio by Stamatelatos et al. by repeating measurements of an anthropomorphic (Remcal) phantom ten times within one week [35]. Other facilities which are presently utilizing the PGNAA method for TBN measurement demonstrated reproducibility of 1.5 - 4.1 % using an anthropomorphic or a rectangular box phantom [6, 8, 14, 34, 35]. This chapter described the technical features and limitations of two different neutron activation analysis methods for in vivo measurement of nitrogen. The calibration procedure for the PGNAA methods employed by the BNL research group was also explained. In the following chapter, experimental procedures and results of the nitrogen measurement performed using the (n,2n) method will be described and comparison of the two techniques will be made based on the computer simulation in chapter 4. 3. Nitrogen measurement using a D-T neutron generator 3.1 Objective In order to assess the feasibility of the use of a 14 MeV neutron generator installed at the Body Composition Laboratory (BCL, United States Department of Agriculture, Human Nutrition Research Center on Aging, Tufts University, Boston) [54-57] for nitrogen measurement in vivo, measurements of a nitrogen sample are performed. Detector efficiency of the counting system, which was newly designed for human thigh measurements, is also measured. In addition to the evaluation of the system, the results of measurement of the nitrogen sample will be used for comparison with those obtained from Monte Carlo simulation. This verification of the simulation procedure is needed because the simulation will be also used for estimation of the sensitivity, detector efficiency and dose of both the (n,2n) method and the PGNAA method. Furthermore, oxygen interference which has been considered to be the major problem of the (n,2n) method [11, 26, 31, 32] will be investigated by experiments using the 14 MeV neutron generator. 3.2 Irradiation system An irradiation system using a small sealed D-T neutron source at BCL has been developed for the in vivo measurement of body oxygen, hydrogen, and carbon [54-57] and now the feasibility of the application for nitrogen measurement in vivo is under investigation. The main composition of this neutron source is a 13 cm-long MF Physics A-320 vacuum tube in which a deuterium accelerator is sealed. The ion beam is pulsed at the repetition rate of 4 - 10 kHz, where 103 - 104 neutrons per pulse are delivered. A scanning facility was also developed for the detection of O, H, and C though the system will not be used for nitrogen measurement. Figure 3-1 shows the irradiation system schematically. patient Aluminum bed Lead Steel D-T neutron generator Figure 3-1 :(n,2n) method: irradiation set-up The subject on the bed is irradiated by14 MeV neutrons produced from the D-T generator. The tube is located 64 cm beneath the scanning bed made of 3.2 mm-thick aluminum and 2 incident neutrons are collimated by a steel shield defining a 40.64 x 30.80 cm rectangular beam window. A 5.08 cm-thick lead comprises the upper part of the collimator, which reduces the the intensity of photons produced in the surrounding materials by neutron capture and inelastic scattering. 3.3 Counting system for human thigh measurement HV1 amplifier and MCA HV2 z NaI(TI) detector 1 x detector 2 y Figure 3-2 :(n,2n) method: counting set-up Each Nal(TI) detector in plastic pieces isconnected to the amplifier, the multi-channel analyzer and the high voltage supply. An aluminum sheet covers the detectors and the subject is placed on it. A whole body counting room, which is separate from the neutron generator, is used for counting 511 keV photons created by irradiation of the subject. They are countedby two NaI(TI) detectors (11 x 11 x 43 cm 3) which are positioned in a polyethylene holder and covered with a thin (< 1 mm) aluminum sheet. Figure 3-2 illustrates the whole counting system which was designed specially for measurement of nitrogen in the human upper thighs which contain relatively large amounts of nitrogen because of their large proportion of muscle. A patient sitting on the counting set-up is shown in Figure 3-3. Two NaI(T1) detectors Bed SIDE Figure 3-3 : (n,2n) method: counting set-up with a patient The Nal(TI) detectors are placed on the bed in the whole body counting room. FRONT 3.4 Experimental procedure 3.4.1 Nitrogen measurement 3.4.1.1 Preparation of sample Two polyethylene bottles (29.7 x 15.2 x 22.0 cm 3 for each) were filled with urea/water ([NH 2]2 CO) solution containing, in total, 893.4 g of nitrogen and 17.29 liter of water. The nitrogen amount is almost twice the amount of nitrogen in the upper thighs of reference man [58](TBN in the reference man is 1800 g. We made a rough assumption that the amount of nitrogen in both upper thighs is almost 1/4 of that in a whole body. 3.4.1.2 Irradiation The urea solution bottles were placed on the aluminum bed under which the small sealed Zetatron tube is located. Figure 3-4 shows the axial view of the (n,2n) irradiation set-up with the sample. The voltage was set 65 kV, and the sample were irradiated for 20 min. The samples were positioned 64 cm away from the 14 MeV neutron source and the thickness of the sample along the direction of irradiation was 15.2 cm. Two urea solution bottles Figure 3-4 : (n,2n) method: irradiation set-up for urea solution measurement A 14 MeV neutron generator is located beneath the aluminum bed (TSD = 64 cm). Two bottoles of urea/water solution were located on the aluminum bed for nitogen measurement. 3.4.1.3 Counting The whole body counting room was used for detection of 511 keV photons emitted from the 13N decays. Calibration of the counting system illustrated in Figure 3-2 was performed by placing 137 Cs and 60Co standard samples on a carton spacer which was located above the NaI(TI) detectors. The nitrogen background was measured before counting the urea sample. After the irradiation of the urea sample, the bottles were transferred to the counting system. Transfer time was 46 sec. The sample was counted for 20 min. 3.4.2 Measurement of detector efficiency of thigh counting system As a sample which provides estimation of the detector efficiency for the urea solution sample used for our experiments, we used 0.5218 ± 0.0136 ptCi. 137Cs 137Cs solution which has an activity of was chosen because the characteristic energy of y-rays detected is 662 keV, which is close to 511 keV of annihilation quanta emitted from nitrogen, and thus the error that comes from self-absorption in the sample will be minimized. The 137Cs solution is filled in a bottle identical to ones of urea solution. The sample is located above two NaI(TI) detectors and counted for 10.5 min. 3.4.3 Interference of oxygen Two identical polyethylene bottles were also filled with distilled and deionized water (18.86 liter in total) to simulate nitrogen-free samples. Irradiation and counting were performed in exactly same way as with the urea solution experiment except the transferring time was 39 sec instead of 46 sec for urea solution. 3.5 Data analysis and results 3.5.1 Counting results Gamma rays produced by the positron emission were counted in the range of photon energy of 432 to 560 keV. Figure 3-5 shows the y-spectrum obtained from urea solution samples. The nitrogen net counts were calculated by subtraction of the oxygen counts. The elements investigated and counts results are listed in Table 3-1. FULL SCALE: 8 - 2844 XeU (4 XeU/Channel), 8 - 32 Counts/Minute RUN654 DETECTOR A SPECTRUM RUN STATIC MINUTE 28 AIHO BOTTLES RUN TITLE: TUO UREA GROSS CNTS/MIN: 2482.95 GROSS COUNTS: 49659 **xx USER-DEFINED UINDOU ( 432 - 568 XeU ) **** GROSS CHTS/MIN INUINDOU: 576.45 GROSS COUNTS INUINDOU: 11529 NET CNTS/MIN INUINDOU: 219.225 4384.5 NET COUNTS INUINDOU: Figure 3-5 : Nitrogen 511 keV y-spectrum obtained from a urea solution sample H20 Counting time [min] 20 20 20 ' 37 Cs + H 2 0 10.5 H20 10.5 Element Compound (2 bottles) Nitrogen (2 bottles) Oxygen Background (2 bottles) [NH 2 ]2 CO + H2 0 H20 137Cs (1 bottle) Background (1 bottle) Gross c )unts 11529 _S107 6540 __81 6362 __80 607229 _ 779 3754 _61 Net counts 5582 ± 130 178 ± 114 603655 + 782 Table 3-1 : Counts measured from each sample The nitrogen count and its interference were obtained from the energy window of 432 to 560 keV.The nitrogen net counts was corrected by transferring time and oxygen fraction in the sample. The 137 Cs was counted in the range of 540 to 760 keV. 3.5.2 Oxygen interference According to the meaning of the interference mentioned by Leach et al. [31 ], the interference of oxygen was calculated as the fraction of counts contributed from oxygen alone out of the counts produced from 14N(n,2n)13N alone. Both nitrogen and oxygen were normalized to elemental composition of reference man [58]. As a result, the oxygen interference obtained from phantoms filled with nitrogen-containing and nitrogen-free solution was 4.07 ± 2.60 %. This result is much smaller than - 20 % reported by other investigators using rectangular shaped phantoms. 3.5.3 Estimation of detector efficiency at 511 keV The detector efficiency at 662 keV, S 6 62 for two bottles of samples was assumed to be same as that for one bottle and calculated to be 5.98 ± 0.16 % (error due to counting statistics). From this result, the estimation of the detector efficiency at 511 keV was made based on the following assumptions. * Photons emitted from the sample impinged at right on the detectors. * The path lengths of all photons in the sample were a half of the thickness of the sample (b = 7.6 cm) in average. * The attenuation in the urea and 137 Cs solution is the same as in water. * Attenuation in the air was negligible. 8511 '662 [(1-e 1- a- Ureasol. 1511 NL -1 e'')Na - 1- e Cssol. 662 where, a: the thickness of NaI(T1) along z-axis = 10.5 cm b: a half thickness of the sample = 7.6 cm From the assumptions made above, the detector efficiency at 511 keV for the two bottles of samples was estimated to be 5.74 ± 0.15 % (error due to counting statistics). The ratio of the detector efficiencies at 511 keV and 662 keV estimated on the assumptions mentioned above was confirmed by Monte Carlo simulation which provided detector efficiency at 511 keV of 5.50 + 0.16 % (error due to counting statistics). 3.5.4 Estimation of neutron output The neutron output of the D-T generator was calculated using the detector efficiency obtained in the last section and the results of nitrogen measurement. Assumptions: * The D-T generator was a isotropic point source of 14 MeV neutrons. * The flux of activating neutrons (11.3 MeV - 14 MeV) averaged over the sample is proportional to the 14 MeV neutron output of the D-T generator. The ratio of the activating neutron flux in the sample to the total neutron output was calculated from Monte Carlo simulation. According to the simulation results, the activating neutrons composes 11.8 % of total neutrons yielded in the sample. As a result, the neutron output was estimated to be 4.9891 x 107 n/sec (4.6 % error due to counting statistics). 4. Monte Carlo simulation Computer simulations of nitrogen experiments using a Monte Carlo code are performed for evaluation of the capabilities of the (n,2n) method and the PGNAA method. Sensitivity and dose of each method are investigated by simulating nitrogen measurement facilities currently in use. In the simulation several experimental problems inherent to the two methods can be eliminated so that the statistical error of the results are minimized. Various types of phantoms are designed for the simulation in order to adequately assess each system. The simulation results will be used for the comparison and the assessment of the suitability of the two methods for in vivo measurement of nitrogen. 4.1 MCNP code Monte Carlo N-particle (MCNP) is a general-purpose Monte Carlo transport code which has the capability to model the behavior of neutrons, photons, and electrons through statistical sampling processes based on extensive nuclear data collected from several sources [59]. MCNP allows the user to build three-dimensional configurations of experimental geometries and to calculate the neutron, photon and/or electron fluence averaged over a surface or a volume region ('cell') specified by the user based on highly dense, point-wise cross-section data. MCNP also has the capability of calculating the number of particles crossing the surface and the energy deposition throughout the cell. The MCNP simulation code has been under development at Los Alamos National Laboratory in Los Alamos, New Mexico since World War II. The use of MCNP for simulating the behavior of neutrons (and photons) has been validated by a number of investigators [44, 60-62]. Version 4b is used for the work of this thesis [59]. 4.2 Figures of merit 4.2.1 Dose The dose delivered to the body should be limited to the extent that the possible risks are negligible compared to enormous benefits caused from the radiation exposure. The lower absorbed dose is more desirable on the basis of the concept of as low as reasonably achievable (ALARA). The total dose equivalent reported by the PGNAA facilities is in the range of 0.14 - 0.75 mSv using the quality factor of 10 for the PGNAA method with a 238Pu-Be or a 252Cf neutron source for a 14 - 40 minute-scan [6-8, 14, 34, 42]. On the other hand, a dose of 0.5 mSv is reported for the (n,2n) method using a 14 MeV neutron generator for an 80 second irradiation time [11]. The dose normalized to unit irradiation time is 20 - 90 times higher for the (n,2n) method than the PGNAA technique. The dose equivalent for a 400 sec-bilateral irradiation at the PGNAA facility of BNL most recently upgraded was estimated to be 0.8 mSv using the radiation weighting factors currently recommended by ICRP [63], measured at the bed level by Eberline ESB2 and RO-2 monitors for neutron and y-rays, respectively [42]. 4.2.1.1 Equivalent dose In order to examine the eqivalent dose of nitrogen measurement systems with MCNP calculations, a 40 x 30 x 15 cm 3 box phantom is used. The phantom is filled with a uniform solution composed of 25 % adipose tissue and 75 % muscle which roughly approximates the composition of the human thighs. Table 4-1 shows the density and box thigh phantom density [g/cm 3 ] 1.025 H C [%] [%] 10.5 25.675 N [/ 2.725 O Na P S C1 K [%] [%] [%] [%] [%] [%] 60.2 0.1 0.15 0.25 0.1 0.3 Table 4-1 : Density and composition of the box thigh phantom The percentages are expressed as weight fractions [64]. ...... ............................ 7J Figure 4-1 : The rectangular box phantom filled with a uniform solution composed of 25 % adipose tissue and 75 % muscle. A 2 cm-diameter and 1 cm-thick cylinder cell is investigated for the maximum skin dose calculation. composition of the box thigh phantom used in the simulation. The maximum skin dose is investigated by calculating the dose delivered to a 2 cm-diameter and 1 cm-thick cell, which is included in the box phantom and located at the center of the neutron beam window on the bed level. Figure 4-1 shows the configuration used for the calculation of the maximum equivalent dose. Table 4-2 shows the weighting factors recommended by ICRP [63], which were used for the calculation of equivalent dose. Type and energy range Neutrons, energy < 10 keV 10 keV to 100 keV 100 keV to 2 MeV 2 MeV to 20 MeV > 20 MeV Photons, all energies Weighting factor 5 10 20 10 5 1 Table 4-2 : The weighting factors recommended by ICRP [63] 4.2.1.2 Dose distribution in human thighs The neutron and photon dose calculation in MCNP is attained by using tabulated data of kerma factors for each radiation as fluence-to-kerma conversion factors [62, 65]. The use of the fluence-to-kerma factors for neutron/photon dose calculation is valid because the dimensions of the cells under investigation are larger than the ranges of secondary charged particles (- 1 cm) and thus the condition of the secondary charged particle equilibrium is established [60]. As mentioned earlier, the calculation of neutron and/or photon fluence is based on the cross-section data provided in MCNP. In the simulation, neutron cross section data of natural elements are chosen from MCNP libraries except hydrogen, nitrogen and oxygen, for which the data of 1H, ' 4N and 160 are and crystalline used because of the lack of natural element data. The chemical binding to be effects of the thigh solution on the behavior of low energy neutrons are assumed data same as that of water which can be corrected by use of thermal neutron cross-section supplied in MCNP. While total dose delivered from a whole measurement is used as a crucial measure in the to assess the suitability of the system for human study, the distribution of the dose human body is also important. Knowledge of the spatial dose variation is essential is because the uniformity of dose contribution from activating neutrons in the subject adipose tissue 50 cm muscle 11 cm 13 cm femur 18 cm 20 cm Figure 4-2: Thigh model for MCNP simulation Three tissues; adipose tissue, muscle, and femur comprise the model. Tissue adipose tissue (at) muscle femur density H C [g/cm 3 ] [%] [% 0.95 1.05 1.33 11.4 10.2 7.0 59.8 14.3 34.5 0 Na %] [%] [%] 0.7 3.4 2.8 27.8 71.0 36.8 N 0.1 0.1 0.1 Mg [%] 0 0 0.1 P [%] 0 0.2 5.5 S Cl K Ca [%] [%] [%] [%] 0.1 0.3 0.2 Table 4-3 : Density and composition of adipose tissue, muscle, and femur which comprise the thigh model. The percentages are expressed as weight fractions [64]. 0.1 0.1 0.1 0 0.4 0 0 0 12.9 needed for absolute measurement of nitrogen in the body. To simulate the human thigh measurement mentioned in chapter 2, a model of human upper thighs was designed. Figure 4-2 illustrates the model consisting of three tissues; adipose tissue, muscle, and femur [64, 65]. Density and composition of each tissue is listed in Table 4-3. 4.2.2 Sensitivity The uniformity of sensitivity to nitrogen detection in the subject is important for the absolute measurement of nitrogen and the investigation of this critical factor should be made to evaluate different neutron sources and the experimental set-ups. It should be noted again that the sensitivity is influenced by three factors: activating neutron flux, selfattenuation of y-rays detected, and detector efficiency. For sensitivity calculations a rectangular box phantom model will be used. The phantom is filled with the thigh solution composed of elements shown in Table 4-1. A 40 x 40 x 15 (depth) cm 3 phantom is divided into smaller cells so that the spatial variation of sensitivity within the phantom can be examined. The region under investigation of sensitivity variation is limited to the horizontal size corresponding to the neutron beam window of the nitrogen measurement system. 4.2.2.1 (n,2n) method The composite sensitivity can be obtained as follows. At the time irradiation stops, the activity of ' 3N; A 13 [#/min] is: - A 13 = R 13 x(1-e Ato (1) where R13: (n,2n) reaction rate [#/min] X: decay constant of 3N = ln2/T 1 /2 [min-l]; T1/ 2 = 9.96 [min] to: irradiation time = 20 [min] When the end of the irradiation is considered to be t = 0, the activity of 13N after the irradiation is: A 1 3 (t) = R 13 x (1 - e-Ato) x e - t (2) In the simulation, the transferring time i.e., the time between the irradiation and the counting, is assumed to be 46 sec which was taken in conducting nitrogen experiments at the BCL facility (chapter 3). The number of disintegrations during the counting time 20 min (= t 2 - ti); D13 [#] is: D13 = fA3(t)dt = R13 (1 - e )(e2t1 - e - 2) (3) where tj: time when the counting starts = 46 [sec] = 0.7667 [min] t2 : time when the counting ends = 46 [sec] + 20 [min] = 20.7667 [min] Net counts of 511 keV annihilation quanta; Cn2n [#] is: Cn2n = 6511 x e1 3 x D13 (4) where 8511: absolute detector efficiency [%] el 3: probability that a 511 keV photon is emitted per disintegration of 13N = 200 [%] Composite sensitivity; Sn2n [#/g/Sv] can be obtained such that: Cn2n Sn2 n -_n mNn2n x dn n 2 x (t 2 -t) '511 x e1 3 x R 13(I - eAto )(e-t - e-t2 (5) mNn2n x dn2n x to x A where mNr2n: amount of nitrogen for one cell of thigh solution box phantom [g] den: incident skin dose rate [Sv/min] The incident dose can be estimated from the results of the dose calculation. Obviously unknown values in estimating the sensitivity are only the (n,2n) reaction rate and the absolute detector efficiency, which will be calculated by MCNP simulation. Since 14N(n,2n) 13 N is a delayed gamma reaction, two MCNP runs are needed in order to obtain the sensitivity of one cell for the (n,2n) method. The first run provides the occurring rate of (n,2n) reactions in each cell (= RI 3 [#/min]), and the second run calculates the absolute detector efficiency supposing the cell to be a uniform source of photons of 511 keV(= Es11 [%]). Incorporation of these two products into the above equations results in the composite sensitivity for a unilateral nitrogen measurement at a given position in the phantom. In nitrogen measurement using the (n,2n) method, as shown in Figure 3-3, a patient will be asked to sit above the detectors so that the irradiated parts of the thighs will be detected from below. If the irradiation is conducted in a prone position, decreasing detector efficiency with the thickness of thighs will be compensated with increasing fast neutron flux delivered to the subject. The discrepancy in sensitivity measurement resulting from irradiation in different positions will be also examined. 4.2.2.2 PGNAA method Because of the prompt gamma reaction of nitrogen, the sensitivity measurement for the PGNAA method could be performed in one MCNP run by building both irradiation and counting systems. In this case, only the cell of interest would contain nitrogen with the rest of the phantom filled with a nitrogen-free solution. Such simulations, however, require too much computer time to obtain results with acceptable fluctuations due to the low cross section of 14 N(n,y) 1 N reaction and the relatively low detector efficieircy of the BNL counting system. Thus sensitivity of the PGNAA method will be obtained by separating the measurement into two steps as was done with the (n2n) method, i.e., irradiation and counting. Since there is no need to take into account the loss of the nitrogen counts during the irradiation and transferring time, R15 x t' = D15 (6) for the PGNAA, where the notation '15' indicates 15N, and t' is the measurement time. Using eq. (5), composite sensitivity for the PGNAA can be obtained as following: CPG mNPG x dpG x t' - 810.83 x e 1 5 x mNr R1 5 (7) xdPG The notations of 'PG' and '10.83' indicate the PGNAA method and 10.83 MeV which is the energy to be detected, respectively. e15 = 15 %;the branch ratio of the disintegration to the ground state. As with the (n,2n) method, the reaction rate of 14 N(n,y)'N (= R 15 [#/min]) and the absolute setecter efficiency (= 10.8 3 [%]) obtained by MCNP simulations leas to the composite sensitivity of the PGNAA technique. 4.2.2.3 Reaction rate MCNP calculates the total number of a requested type of reaction which occurs in all elements existing in a cell. The calculation is made in a way that the average neutron fluence over the cell is multiplied by the total number of atoms and the microscopic cross section of the reaction specified. The rate of 14N(n,2n)' 3N reactions occurring in the phantom can be obtained by specifying (n,2n) reactions because nitrogen is the only one nuclide for which the (n,2n) cross section data are available among all nuclides in the phantom. The estimation of nitrogen counts made by this method can be free from interference problems associated with the (n,2n) nitrogen measurement. On the other hand, it is not possible to calculate the occurring rate of 14N(n,y)'N reactions in the same way because MCNP provides the total number of the (n,y) reactions occurring from all other elements existing in the phantom. Instead, tabulated results of the (n,y) activity in each nuclide given after the simulation will be used to estimate the number of neutron absorption reactions from nitrogen for the PGNAA method. The high background contributed from the 238Pu-Be source will be excluded in the simulation by separating the measurement into two stages. 4.2.2.4 Absolute detector efficiency The absolute detector efficiency can be estimated by use of a "pulse height tally" provided by MCNP. This tally scores the number of photons which deposit energies within the range specified by the user, by all the tracks passing through a cell. The absolute detector efficiency of each cell in the thigh phantom is obtained by defining an isotropic photon source uniformly distributed in the cell. The energy of the photons is 511 keV for the (n,2n) method and 10.83 MeV for the PGNAA method. Therefore, the detector efficiency results obtained by the simulation are only determined by two factors: the counting geometry and the energy of the photon source. The photon energy range is set at 432 - 560 keV and 9.5 - 11.1 MeV for the (n,2n) method and the PGNAA method, respectively. 4.3 Nitrogen measurement systems Measurements of dose and nitrogen counts using four different experimental systems are simulated by the MCNP code. The four types of nitrogen measurement system are listed in Table 4-4. The details of the source characteristics and the geometric configurations defined in the simulations of one (n,2n) system and three PGNAA systems will be described in this section. As a neutron source for the PGNAA method, the feasibility of the use of an accelerator recently built in the Laboratory for Accelerator Beam Applications (LABA) at the Massachusetts Institute of Technology will be investigated. Method (n,2n) Neutron source D-T generator PGNAA 2Pu-Be PGNAA 7Li(p,n), Be(p,n), 9Be(d,n) PGNAA Li(p,n), Be(p,n), 9Be(d,n) Source of input data Experimental set-up currently used at BCL Experimental data Experimental set-up currently used at BNL, Reports in literature Experimental set-up practicable at LABA, Neutron spectrum experimentally measured by LABA Experimental set-up currently used at BNL, Neutron spectrum experimentally measured by LABA Table 4-4 : Four types of nitrogen measurement systems simulated by MCNP 4.3.1 D-T generator - (n,2n) method The method of nitrogen measurement employing the 14N(n,2n)13N reaction is described in chapter 3. The simulated irradiation system with the thigh model is shown in Figure 4-3. In the simulation the neutron source is defined as a 14 MeV isotropic point source and the neutron output of 4.9891E7 n/sec is used. The target-to-skin distance (TSD) is 64 cm. It is also assumed that the material under the irradiation system consisted of 100 % concrete. The composition of the materials simulated by MCNP is listed in Table 4-5. Figure 4-4 shows the irradiation set-up with a box thigh phantom used for sensitivity calculation. The region investigated is 40 (width) x 30 (length) x 15 (depth) cm 3 which corresponds to the size of the neutron beam window. The phantom is divided into 36 cells of 10 (width) x 10 (length) x 5 (depth) cm 3 for calculations of sensitivity distribution. The configuration of the counting system with the box thigh phantom is shown in Figure 4-5. Two rectangular-shaped NaI(TI) counters are placed in parallel in the polyethylene holder so that the human thighs are measured from below. The composition of NaI(TI) crystals is sodium and iodide at the atomic ratio of 1:1 and the thallium material density 3 [g/cm ] lead 11.4 steel 7.855 aluminum bed concrete 2.6989 2.3 compositions size [%, in weight] [cm] outer: 99.06 (x), 61.28 (y), 5.08 (z) Pb 100 inner: 40.64 (x), 30.80 (y), 5.08 (z) outer: 99.06 (x), 61.28 (y), 50.8 (z) Fe 70, Cr 19, Ni 11 inner: 19.69, 30.16, 40.64 (x) 30.80 (y), 5.08,10.16, 35.56 (z) 86.36 (x) x 193.04 (y) x 0.32 (z) Al 100 200 (x) x 200 (y) x 50 (z) (approximately) O 52.9, Si 33.7, Ca 4.4, Al 3.4, Na 1.6, Fe 1.4, K 1.3, 'H 1, Mg 0.2, C 0.1 Table 4-5 : Density, size, and composition of each component of the collimator assembly simulated for the BCL (n,2n) system activator is neglected in the simulation. Two detectors are separated by 2 cm. The parts of the photomultipliers are assumed to be 100 % polyethylene as is the composition of the polyethylene holder. The thin (< 1 mm) aluminum sheet covering the detectors is neglected in the simulation. Table 4-6 lists the density, size, and compositions of materials used in the simulation. material density [g/cm3 ] sodium-iodide crystal photomultiplier 3.67 polyethylene size [cm] 11 x 11 x 43 compositions [atomic fraction, except borated polyethylene] Na:I = 1:1 1 6 (dia), 13 (long) C:H = 1:2 1 55 (x), 52 (y), 13 (z) C:H = 1:2 Table 4-6 : Density, size, and compositions of each component of the detector and the shielding materials simulated for the BCL (n,2n) system o0 Sz s ... . 9: X oro nbed, o on A-* o on I *;-----; r--% ~ ; -' ~~:1~~~:~:~ ~-''X- o I i' ou, 1 ad ,rete; so0 -60 -40 -20 0 20 40 60 80 -80 -60 -40 -20 0 20 (b) lateral view (a)axial view Figure 4-3 : (n,2n) method: BCL irradiation system with the thigh model for calculation of dose distribution by MCNP 40 60 aluminum bed lead concrete.' T7 A -80 -60 -40 -20 (a) axial view 0 20 40 60 80 -80 -60 -40 -20 20 0 (b)lateral view Figure 4-4: (n,2n) method: BCL irradiation system with the rectangular box phantom filled with thigh solution for sensitivity calculation by MCNP 4 z t ' polyethylene holder ' Nal detectors 40 -40 -20 0 20 40 (a) axial view .(b) lateral view Figure 4-5 : (n,2n) method: BCL counting system with the rectangular box phantom filled with thigh solution for sensitivity calculations by MCNP 4.3.2 238Pu-Be radionuclide - PGNAA method The simulation of nitrogen measurement by the PGNAA technique employed at BNL is carried out using information supplied in published BNL papers [2, 35, 42]. The schematic configurations of the neutron irradiation set-up with the thigh model are shown in Figure 4-6. Because no detailed neutron spectrum data of the 238 Pu-Be radionuclide are provided [66], that of a 239Pu-Be neutron source is used for the energy range above 0.5 MeV [67] because of the similarity of the energies of ct-particles emitted from 239Pu. 2 38 Pu and The influence of this modification on the simulation results can be estimated negligible because the range of the neutron energy 0 - 11 MeV and the average neutron energy 4.5 MeV reported in Vartsky's paper [2] are very close to those used for the simulation (0 - 10.5 MeVand 4.6487 MeV, respectively). The neutron yield data used for commission the simulation are shown in Figure 4-7. The neutron source is assumed to be isotropic in the simulation and the source strength is estimated to be 2.08 x 108 n/sec [2]. The 239Pu-Be point source is modeled as being surrounded by a graphite reflector with a pyramid-shaped opening inside. The neutron source is positioned at the apex of the pyramid and TSD is 72 cm. The size of the rectangular beam aperture of the graphite is 28 (width) x 11.2 (length) cm 2 . In the simulation the graphite is assumed to consist of 100 % 12C for which corrections are made for the chemical binding effects on thermal neutron scattering. The graphite block is surrounded by bismuth, lithium-doped resin, and borated lead, in that order, and the entire assembly is capped with a thick bismuth block providing the beam aperture of 40 (width) x 16 (length) cm 2 . Density, size, and compositions of graphite density [g/cm3] 2.1 bismuth 9.8 material compositions size [cm] [%, in weight, except D20] C 100 72 (z) 30 (y) x outer: 30 (x) x inner: 11.2 (x) x 28 (y) x 42 (z) Bi 100 5 (thickness), 34 (z) (side) lithium-doped 1 28 (thickness), 53 (z) C 79.6, 1H 13.0, 6Li 1.1, 7Li resin 0.9, F 3.1, 160 2.3 borated lead 11.4 12 (thickness), 12 (z) Pb 95.0, '°B 5.0 bismuth 9.8 120 (x) x 120 (y) x 18 (z) Bi 100 68 (x) x 68 (y) x 2.5 (z) 7H 1 2 H:' 6 0 ( upp er ) heavy water 1.1034 1 = 1:1 (atomic fraction) Table 4-7 : Density, size, and composition of each component of the collimator assembly simulated for the BNL PGNAA system each component are listed in Table 4-7. Bi is an excellent shielding material because of its high density and low yield of (n,y) reactions. The Li-doped resin shields neutrons and minimizes the interference for hydrogen measurement because of the high neutron capture cross-section of lithium. Although no consideration for the hydrogen measurement is needed for the work of this thesis, this composition is also emulated because the component may affect the dose measurement. The borated lead is used to shield thermal neutrons and y-rays also because of the high cross-section of the thermal neutron reaction lB(n,ca) and the high density of the lead. The fast and epithermal neutrons collimated by the assembly are premoderated by heavy water before interacting in the phantom. The thickness is 2.5 cm, which is chosen by the BNL researchers as a compromise between the yields of nitrogen counts and the uniformity of composite sensitivity based on the experimental results. The chemical binding and crystalline effects of D20 are corrected as was done for light water. The size of the neutron beam window is 48 (width) x 19.2 (length) cm 2 . Figure 4-8 shows the irradiation system with the rectangular box phantom filled with thigh solution simulated for sensitivity calculations. The region measured for sensitivity calculations is 40 (width) x 20 (length) x 15 (depth) cm 3 and the variation of the sensitivity is investigated by dividing the phantom into 48 cells of 10 (width) x 5 (length) x 5 (depth) cm 3. material density [g/cm 3] sodium-iodide crystal photomultiplier 3.67 bismuth 9.8 borated carbide (side) borated carbide (front) boric acid borated polyethylene 15.2 x 15.2 x 15.2 compositions [atomic fraction, except borated polyethylene] Na:I= 1:1 5.2 (dia), 30.8 (long) C:H = 1:2 Bi = 1 2.52 behind: 2 (thick), 30.8 (long) front: 4 (thick), 18.7 (long) 5 (thick), 30.8 (long) 2.52 15.2 x 15.2 x 3.5 (thick) 1.435 behind: 7 (thick), 30.8 (long) front: 5 (thick), 18.7 (long) 28 (dia), 2.5 (thick) 1 1.067 size [cm] lB:1B:C = 0.85 : 3.15 : 1 °TB:B:C = 0.85 : 3.15 : 1 10B: "B: 1H: 16 0 = 0.2135 0.7865 : 3 : 3 l'B 0.99,11B 4.01, 'H 13.571, C 81.429 [%, in weight] Table 4-8 : Density, size, and compositions of each component of the detector and the shielding materials simulated for the BNL PGNAA system In the PGNAA method, the characteristic y-rays (10.83 MeV) emitted promptly by the neutron capture reaction in nitrogen are detected by two 15.2 cm-cube NaI(TI) crystal counters. Figure 4-9 shows the detectors positioned at 600 to the z-axis and at 200 at the x-axis in the simulation. The center of the detector window is 55 cm above the bed level. The location of the detectors is optimized by BNL researchers so as to minimize the nitrogen background [35] and the nonuniformity of composite sensitivity in the subject [42] for the absolute measurement of TBN. The detectors are shielded with several materials to reduce the nitrogen background and the hydrogen interference for the BNL calibration procedure. The simulated configuration of one of the two detectors is shown in Figure 4-10. Bismuth, which has a high density, shields the sides of the detectors from y-rays. Boron carbide surrounds the photomultiplier and covers the front of the detector. The outer layer of the detector consists of boric acid and a borated polyethylene slab is added on the front surface. '0 B contained in the last three materials, as Li was used in the collimator, reduces the number of H(n,y)D reactions which lead to interference in the hydrogen measurement. Table 4-8 lists the density, size, and composition of materials used for detector simulation. I . . I . . . . I . z . z bx -- borated lead D 20 y I i 0~ Li-doped resin I-60 -60 -40 -20 0 (a) axial view 20 40 60 80 -80 -60 -40 -20 0 20 40 60 -40 -20 0 20 40 60 (b) lateral view Figure 4-6 : PGNAA method: BNL irradiation system with the thigh model for calculation of dose distribution by MCNP A 238Pu-Be neutron source was positioned at the apex of the pyramid-shaped opening shown at the center of the collimator (TSD = 72 cm). 80 logarithmic scale 100 - A 10 4 . 01 1. .1 . I E 1.00E-07 I I 1.00E-06 1.00E-06 II 1.00E-04 I 1.00E-03 1.00E-02 1.00E-01 1.00E+00 1.00E+01 1.00E+02 Neutron Energy [MeV] linear scale 12Y 10. * . 8- * + 2* " 64 ** 2- 2 4 6 8 10 Neutron Energy [MeV] Figure 4-7 : 238 Pu-Be neutron yield data points used for the simulation Experimental data obtained using a 238Pu-Be were used for the energy range below 0.5 MeV[66], shown in the log scale (upper figure), and the data of a 239Pu-Be was used for the range above 0.5 MeV[67], shown in the linear scale (lower figure). 1 4-.. 1 I I ._ I , . I ~ I .' .1 I- z I ;s:i i I IJ I ,' borated lead D20 I Li-doped resin' i -80 -60 -40 -20 0 (a)axial view 20 40 60 -80 -60 -40 -20 0 20 (b) lateral view Figure 4-8 : PGNAA method: BNL irradiation system with the rectangular box phantom for sensitivity calculations A 238pu-Be neutron source is positioned at the apex of the pyramid-shaped opening shown at the center of the collimator. 40 60 60 7 i0 Y:, inn 0 L 1--0 0 - 50 so lO1 -inn (b) axial view (a) top view Figure 4-9 : PGNAA method: BNL detector positioning simulated by MCNP The shielded detectors are positioned at 600 to the z-axis and at 200 at the x-axis. 60 . 1 I . . . . A . . . . . . ... boric acid I - bismuth Nal detector -40 -20 0 20 40 -40 -20 Figure 4-10 : PGNAA method: one of the BNL detectors simulated by MCNP A 15.2 cm-cube Nal crystal is shielded with several shielding materials. 0 20 40 4.3.3 Accelerator source 9Be(p,n) with LABA set-up - PGNAA method Figure 4-11 shows the picture of a 4.1 MeV tandem electrostatic accelerator which has been developed and recently installed in MIT's LABA [68, 69]. The accelerator was designed to produce high-currents of charged particles and thus high intensity neutron beams for boron neutron capture therapy (BNCT) research. It will be possible to generate proton or deuteron beam currents up to 4 mA. Furthermore, the recent installation of a switching magnet allows the set-up of multiple experimental equipment simultaneously using five separate beam lines. Figure 4-12 illustrates the configuration of the LABA accelerator and the five endstations utilized for different types of experiments. Currently three reactions are under investigation as most likely effective neutron sources for BNCT: 7Li(p,n), 9Be(p,n) and 9Be(d,n). The neutron energy spectra produced from protons of various energies on a thick target were obtained by the LABA research group in collaboration with the University of Ohio [70, 71]. MCNP simulation is used to investigate the 9Be(p,n) reaction as a possible neutron source for nitrogen measurement employing the PGNAA method. The beryllium target is bombarded by protons accelerated up to 4 MeV through the aluminum beam tube. The neutron energy spectra experimentally measured are used to define the source in the simulation. The source is defined as a surface source with 2.307 cm-diameter. The Be target and Al tube are surrounded by heavy water and a thick graphite reflector. The produced neutrons are premoderated by the 15 cm-thick O)0. This configuration has been developed for the purpose of creating an epithermal beam which is needed for BNCT research. Figure 4-13 shows the moderator/reflector assembly with the thigh model. A patient is assumed to be irradiated in a standing position. The thigh model is located 16 cm away from the Be target. The characteristics of the geometry and the materials used in the simulation are listed in Table 4-9. The configuration with the box thigh solution model shown in Figure 4-14 is used for the measurement of the distribution of the number of 14N(n,2n) 13N reactions in the phantom. material aluminum tube graphite reflector heavy water density [g/cm 3] 2.6989 2.1 1.1034 size [cm] 29.65 (long) (od), 2.007 (id), 2.307 compositions [atomic fraction] Al 1 around Al tube: 20.307 (thick), 18 (long) around D2 0: 18 (thick), 15 (long) 9.23 (dia), 15 (long) 12C 1 H: 6 = 1:1 Table 4-9 : Density, size, and composition of each component of the detector and the shielding materials simulated for the LABA PGNAA system ... .... ... .... .. ... .. ... ..... ... . . . ;*:; _ )r ::-:' Figure 4-11 : A tandem electrostatic accelerator installed in MIT's LABA is shown with the ion source next room and Protons or deuterons are accelerated through the beam tube extended to the 9 bombarded on the thick Be or Li target [68]. The feasibility of the use of the Be(p,n) as a neutron source for nitrogen measurement is investigated. 64 4 Figure 4-12 : Configuration of the LABA accelerator and experimental rooms The installation of a switching magnet provideded the five separate beam lines [68]. i |, 1 ::r:^ I,,. .... i~~ ; s- . ~~~:;ai-~~~ ~1 .I... ! ;-;-; :~i i I I I I y ; ~8~::~ -;; ,,,,, fl _ : i--; o- :Z ;*-ri:" o | -- "- - ~ -- 40 -- 0 , -" ~~ -2 - r- -- -- -- --- i: ;i - i ;Isl_ .3~ ":-~; " ~~ o I . : ;:: I o ! ....: i: i -4 - ..: --**: " 0 (a) axial view 10 20 30 -30 -20 -10 0 10 20 (b) lateral view Figure 4-13 : PGNAA method: LABA irradiation system with the thigh model for calculation of dose distribution by MCNP The Be target is bombarded with 4 MeV protons accelerated through the aluminum tube surrounded by the graphite reflector. The produced neutrons are moderated by 15 cm-thick D20. 30 40 -40 -30 -20 -10 0 10 20 30 40 - (a) axial view -30 -20 -10 0 10 20 (b) lateral view Figure 4-14 : PGNAA method: LABA irradiation system with the rectangular box phantom for sencitivity calculations 67 30 40 4.3.4 Various accelerator sources with BNL set-up - PGNAA method The 238pu-Be neutron source used at the BNL facility is replaced by various accelerator sources in the MCNP simulation. The reactions and the proton/deuteron yield and energies investigated in the simulation are listed in Table 4-10. The neutron energy range for each neutron source measured at LABA is also listed. Experimentally determined neutron energy spectra for those sources are shown in Figure 4-15. The Reaction 7Li(p,n) 9Be(p,n) 9Be(p,n) 9Be(d,n) 9Be(d,n) Proton energy [MeV] 2.5 3.0 4.0 1.5 2.6 Neutron yield [n/min/pA] 4.86E07 1.44E10 6.00E10 n/a n/a Neutron energy range [MeV] 0- 1.0 0-1.0 0 - 2.0 0-6.3 0-7.3 Table 4-10 : LABA accelerator neutron reactions used for the MCNP simulation composition and the positioning of the surrounding materials including the counting of the BNL system are defined exactly in the same way as those used in the simulation the 238Pu-Be system. By investigating this hypothetical system, the comparison between and the accelerator sources can be made, regardless of the effects of the experimental assemblies for nitrogen measurement, to assess the suitability as neutron sources for the PGNAA method. Furthermore, proton currents of the charged particles which will give the same 238 which level of incident dose as the Pu-Be are calculated to estimate the source strength BNL. would make the accelerator neutron sources as effective as the PGNAA system at 0 1 2 3 4 5 Neutron Energy [MeV] Figure 4-15 : Neutron energy spectra for various accelerator sources 6 7 8 4.4 Results 4.4.1 Dose 4.4.1.1 Incident skin dose The incident skin dose was calculated at the center of the neutron beam window of the irradiation systems for the (n,2n) and PGNAA techniques. Calculated results of equivalent dose per unit irradiation time for BCL and BNL systems are presented in Table 4-11. Neutron yields of the D-T generator and the 238Pu-Be radionuclide source used for the simulation are also shown. Neutron yield [n/sec] Equivalent dose per unit time [mSv/sec] (error bar) BCL - (n,2n) method, 4.9891E7 1.6490E-3 D-T generator BNL- PGNAA 2.0300E8 (3.01 %) 3.0464E-3 Method and facility method, 2 38Pu-Be (4.47 %) Table 4-11 : Neutron yield and equivalent dose per unit irradiation time for two nitrogen systems: BCL ((n,2n)) and BNL (PGNAA). The proton currents for the LABA accelerator source, 4 MeV 9Be(p,n), which would give same incident skin dose as the BNL 238Pu-Be source are listed in the Table 4- 12. The calculations were made for the collimator/reflector assemblies of BNL and LABA. Table 4-13 lists the results of proton currents for other types of (p,n) reactions. The BNL collimating assembly was used for all different neutron sources and the simulation of the 2.5 cm-thick D20 moderator was neglected for this case. Neutron source 4.0 MeV p-Be w/ BNL system 4.0 MeV p-Be w/ LABA system Neutron yield [n/min/pA] 6.00E10 Proton current [pA] 0.2703 ± 0.0178 6.00E10 0.0982 ± 0.0069 Table 4-12 : Proton current used for a 4 MeV 9Be(p,n) reaction, which will give same dose as the 238pu-Be for the PGNAA irradiation systems. 4.0 MeV p-Be Neutron yield [n/min/pA] 6.00E10 Proton current [pA] 0.1937 + 0.0111 3.0 MeV p-Be 1.44E10 1.3823 ± 0.0831 2.5 MeV p-Li 4.86E 10 0.2088 + 0.0110 Neutron source Table 4-13 : Proton current which will give same dose as the 238 Pu-Be for the PGNAA irradiation systems. No D20 moderator. 4.4.1.2 Dose distribution Figure 4-1 - 4-3 shows the dose distributions, which are contributed from neutrons, fast neutrons (11.3 - 14 MeV), and photons, respectively, along the thigh length. The five plots shown in the figure correspond to the dose delivered to adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur, indicated in Figure 4-15. In the same way, the dose distributions of neutrons, thermal neutrons, and photons are shown in Figure 4-4 - 4-12 for three different irradiation systems utilizing the PGNAA method: the BNL PGNAA facility with the 238pu-Be source, the LABA facility with the accelerator source, and the combined system of the BNL facility and the accelerator source. As the accelerator source, the 9Be(p,n) reaction was used for both experimental set-ups of the BNL and the LABA. The proton beam accelerated up to 4 MeV at a current of 0.2703 pA was used to provide the same incident maximum dose as the 238Pu-Be source. Table 4-14 lists the each radiation dose averaged over the thigh length for five sections in the thighs. adipose tissue 3 muscle 3 muscle 3 femur femur - adipose tissue 1 : neutron beam Figure 4-16 : Axial view of the thigh modelled by MCNP Doses given in adipose tissue 1,3, muscle 1,3, and femur were plotted along the length of thigh. Measurement system D-T generator (BCL) 23 8Pu-Be (BNL) 4 MeV, .0982pA 9Be(p,n) (LABA) 4 MeV, .2703pA 9Be(p,n) (BNL) Type of radiation total neutron activating neutron (11.3-14) total photon total neutron activating neutron (0- 1E-6) total photon total neutron activating neutron (0- 1E-6) total photon total neutron activating neutron (0- 1E-6) total adipose tissue 1 [ptGy/min] 1.7E+1 muscle 1 femur muscle 3 [ tGy/min] 1.2E+1 [pGy/min] 2.6E+1 [ptGy/min] 6.4E+0 adipose tissue 3 [ptGy/min] 6.1E+0 9.4E+O 6.8E+0 1.6E+1 4.OE+O 4.OE+O 2.OE+O 2.OE+O 7.2E+O 1.2E+O 9.1E-1 3.lE+1 2.1E+1 4.1E+l 9.5E+O 8.4E+0O 5.1E-3 2.9E-2 7.6E-2 1.OE-2 6.0E-4 2.OE+O 2.2E+O 8.1E+O 1.2E+O 8.5E-1 1.8E+O 8.7E-1 7.5E-1 6.2E-2 1.5E-2 7.8E-2 3.OE-1 4.2E-1 3.3E-2 1.4E-3 1.7E+O 1.8E+O 4.7E+0O 5.8E-1 3.6E-1 2.1E+1 1.1E+l 1.3E+1 2.1E+0O 1.5E+0 1.2E-2 6.4E-2 1.6E-1 2.OE-2 1.1E-3 3.3E+0 3.8E+0 1.4E+1 2.OE+0 1.2E+O photon Table 4-14 : Activating neutron dose averaged over the thigh length for five sections in the thighs: adipose tissue 1,3, muscle 1,3, and femur. Relative error ranges from 0.07 to 2.10 % which represents statistical precision of simulated data. (n,2n) method - D-T generator at BCL Neutron Dose 1 I I 1 I 1 I ( I I I I 1 ) I I I 1 I 1 I 1 I I ( I 1 I 1 I I I I L~( I 1 1 I I I I I I 4]b mcnp 05/02/98 17:2 8:37 .~.---i ., iI----i- ---~---i- tally - I- N- 114 n 10000 00 nps bin normed mctal - mc/ddth ------1 ------4....-----.C------''' 4-------I-------------------------... -- Z f cell 1 d flag/dir 1 u user 1 s segment 1 m mult 1 c cosine 1 e energy 3 t t time 1 a.t. 1 - . ----------- -- -- . a.t. 3 ___-----+'-:--___- muscle 1 muscle 3 femur 2 4 y-axis 6 8 1 : 3-8 inside beam window Figure 4-17 : (n,2n) method: Total neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. (n,2n) method - D-T generator at BCL 11.3 - 14 MeV Neutron Dose . .. . I .I I . I I I I ,, )II I, 11 , I , I I,. tI III , , I . , 4b mcnp 05/02/98 17:28:37 114 tally n /I\ ----- / 1 a 1000000 nps / "\ bin normed \I / 0 metal = mc/ddth f cell *d flag/dir user e in a 0 segment mult c o }---------- --.......----- . - .... .. .. cosine energy ... . t time a.t. 1 a.t. 3 muscle muscle I 'I ' ' I ' ' ' I ' 4 2 y-axis 1 I I 6 . . . . . . . . . 8 10 3-8 inside beam window Figure 4-18: (n,2n) method: Activating neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. femur (n,2n) method - D-T generator at BCL Photon Dose I.. . .. . t. i . . . . . . . . . . . . .. 0innp mcnp . . . 4b 05/02/98 17:28:37 -. --- tally \, bin normed .i. metal f Sd l' \ u 0 V ------ ----- - mc/ddth - cell * flag/dir 1 user 1 s segment 1 m mult 1 c cosine 1 e energy 1 time 1 t ------ - -- . -------- 414 a.t. 1 a .t . 3 -------I --- muscle 1 ~----------------- 2 femur .----------- . 4 6 --- muscle 3 8 10 y-axis : 3-8 inside beam window Figure 4-19: (n,2n) method: Photon dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. PGNAA - 239Pu-Be source with BNL setup Neutron Dose . . . .. . . . . . . mcnp .. 4D 01/28/98 17:58:59 114 tally A-----.{ 10000000 nps bin normed mectal = mc/pdth i / / /ii/,, :. --"............. .,, 1 cell * d flag/dir 1 u user 1 s segment 1 mult 1 c cosine 1 e energy 3 t time 1 m %, \ i / f I' a.t. 1 . .- "- - -.-.- --.--- . a.t. 3 muscle 1 -* € " . w L I . . __. _ muscle 3 . I . . . . . . . . . •• . I i . l l . l .l l . . l I ] l l l l y-axis : 4-7 inside beam window Figure 4-20 : PGNAA method with 238Pu-Be source: Total neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. femur PGNAA - 239Pu-Be source with BNL setup Thermal Neutron Dose . . . 1I . . .., , . . , , , ,. i I ., . . . . . . . . . . . . ., . . . ., 4b mcnp 01/28/98 17:58:59 tally 114 n 10000000 nps bin normed ,/\ / mectal = mc/pdth /'\ /"\ // / / /1 \ 1 / ./" \ ------------ // 1 >._ f cell d flag/dir U user a segment m mult / - / / / / " - "o o O ' /" / -\ .. . . . . . N c cosine e energy t time a.t. 1 "" - "- . . ." , a.t. 3 muscle muscle femur 4 6 8 10 y-axis : 4-7 inside beam window Figure 4-21 : PGNAA method with 238Pu-Be source: Activating neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. PGNAA - 239Pu-Be source with BNL setup I I I I I I I I Photon Dose .. . . I I , . I .I ., . I I I I I I 4b mcnp I 01/28/98 17:58:59 414 tally P 10000000 nps bin normed / 1 metal - mc/pdth iN / / /' " ' \ X f cell d flag/dir U u / //\//" // user segment mult /"" //\ /° o. - -, -.. . C cosine e energy t time a.t. 1 a.t. 3 muscle muscle femur 2 4 6 8 y-axis : 4-7 inside beam window Figure 4-22: PGNAA method with 238Pu-Be source: Photon dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. 10 PGNAA - Be(p,n) with BNL setup Neutron Dose S .... .... .... ... . . .. 4b mcnp 01/21/98 08:27:31 tally o 114 n 30000000 nps bin normed mctal = mc/padth so ./ 0 .------ ... ....-- /- o S' f cell * d flag/dir 1 u user 1 a segment 1 m mult 1 c cosine 1 e energy 3 time 1 t ,, ------- a.t. 3 3 a.t. ---------------. muscle 3 ------ 2 6 4 y-axis : 4-7 inside beam window 8 10 Figure 4-23 :PGNAA method with 9Be(p,n) source and BNL set-ups: Total neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. fem xr t PGNAA - Be (p, n) with BNL setup Thermal Neutron Dose *e . S . . . , . . , , , ~Ii . , | , . 4b mcnp I 01/21/98 08:27:31 114 tally n / /\ 30000000 nps bin normed mc/lpadth mctal f cell d flag/dir u user s segment m -_------------ /-/ mult c cosine e energy t time \ a.t. 1 a.t. 3 muscle muscle femur 2 4 6 8 10 y-axis : 4-7 inside beam window Figure 4-24: PGNAA method with 9 Be(p,n) source and BNL set-ups: Activating neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. 81 PGNAA - Be(p,n) with BNL setup Photon Dose - . i , . . . . , I ... . . . ..I . . ,.. I . . . . . . . I 4b mcnp 01/21/98 08:27:31 414 tally P nps h----- 30000000 bin normed / mctal = mc/padth ". f cell d flag/dir u user s segment /'"/' /' \ /\ ,/~ / '" '. /\ m /\ /', /' _,l,/S c cosine e energy t time a.t. 1 ",-, c--I;---- -- mult .,, a.t. 3 muscle I muscle 2 ----------femur 4 6 8 10 y-axis : 4-7 inside beam window Figure 4-25: PGNAA method with 9Be(p,n) source and BNL set-ups: Photon dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. 82 PGNAA - Be(p,n) with LABA setup Neutron Dose . * . , II , . . . .i.. i . . . . . . . . . . . . . . . . mcnp 4b 03/30/98 19:00:30 tally 114 n nps 20000000 bin normed cmetal ; o ae S------- - mc/adth2 * f cell d flag/dir 1 u user 1 a segment 1 m mult 1 c cosine 1 e energy 3 time 1 t a.t. 1 ------- a.t. 3 muscle 1 muscle 3 --------2 4 8 6 10 y-axis : 4-7 inside beam window Figure 4-26 : PGNAA method with 9Be(p,n) source and LABA set-ups: Total neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. 83 -femur t PGNAA - Be(p,n) with LABA setup Thermal Neutron Dose 4b mcnp 03/30/98 19:00:30 114 tally n 20000000 nps / / bin normed / / mctal = mc/adth2 I ------- ---- / / / / I I f cell d flag/dir u user s segment m mult c cosine e energy t time / / / I / -- a.t. 1 ------ \\% a.t. 3 muscle - %-\ % , . I . , . . . . -.-I .-.--.-- - - -. - - . - .- . . --. -- . ' . . . . . . .4 y-axis : 4-7 inside beam window 9 Figure 4-27 : PGNAA method with Be(p,n) source and LABA set-ups: Thermal neutron dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. 84 muscle femur PGNAA - Be(p,n) with LABA setup Photon Dose .... * 4b mcnp .. . 03/30/98 19:00:30 414 tally P 20000000 aps bin normed " . _/ mectal = mc/adth2 \ / / / / / f cell * d flag/dir 1 u user 1 s segment 1 mult 1 c cosine 1 e energy 1 t time 1 S/ S .- / a.t. 3 -------- a t. 3 muscle 1 ------------- -2 8 6 10 y-axis : 4-7 inside beam window Figure 4-28 : PGNAA method with 9Be(p,n) source and LABA set-ups: Photon dose distribution in thigh model Five plots correspond to the dose given adipose tissue 1, adipose tissue 3, muscle 1, muscle 3, and femur. 85 muscle 3 femur 4.4.2 Sensitivity As mentioned in 4.2.2, composite sensitivity is proportional to the product of the reaction rate and the absolute detector efficiency. The variation of those two factors will be also investigated. 4.4.2.1 Number of reactions Reaction Neutron source D-T generator, (BCL) 238 Pu-Be (BNL) 4 MeV Be(p,n) (LABA) 4 MeV Be(p,n) (BNL) 14 N(n,2n) 3N 14 N(n,Y)s15N 14N(n,y) 15N 14N(n,y)15N Average number of reactions [#/g/mSv] (error bar) 9.7418E1 (0.28 %) 1.3527E5 (0.49 %) 9.5823E5 (0.66 %) 2.1248E5 (0.53 %) Uniformity [%] (# of data points) 41.40 ± 0.78 (36) 54.66 ± 1.21 (48) 103.58 + 1.56 (48) 70.78 ± 1.21 (48) Table 4-15 : Average number of reactions and their uniformity in the thigh box phantom measured for irradiation systems used at BCL, BNL and LABA and the combination of the accelerator source and the BNL collimating system. One of the factors which will affect the composite sensitivity is the occurring rate of reaction: 14N(n,2n)13N or 14N(n,y)15N. The reaction rates which were normalized to the amount of nitrogen and the incident dose are listed in Table 4-15. The irradiation time for the (n,2n) method was assumed to be 20 minutes. The uniformity is expressed as the coefficient of variation (standard deviation / mean value) of 48 data points in the phantom. The variation of the reaction rate in the phantom for each system is shown in Figure 4-13 - 4-16. Neutron source Average number of reactions [#/g/mSv] (error bar) 1.1755E5 (0.50 %) 1.8166E5 (0.52 %) 3.1216E5 Uniformity 14N(n,y) 15N 238Pu-Be 4.0 MeV p-Be 3.0 MeV p-Be [%] (48 data points) 57.84 ± 1.22 66 .02 ± 1.20 71 .35 ± 1.20 (0.53 %) 1.5 MeV d-Be 2.6 MeV d-Be 2.5 MeV p-Li 1.3502E5 (0.52 %) 1.6720E5 (0.51 %) 2.3025E5 (0.52 %) 66 .80 + 1.24 63 .86 + 1.20 66 .14 ± 1.19 Table 4-16 : Average number of 14 N(n,y)15N reactions and their uniformity in the thigh box phantom measured for the irradiation system used at BNL and the combination of the accelerator source and the BNL collimating system. No D20 moderater. Also the average numbers of 14N(n,y) 15N reactions in the phantom calculated using the BNL collimating facilities are shown in Table 4-16. Simulated neutron sources are a 238Pu-Be source and various types of accelerator sources. For comparison, all simulations were performed without the 2.5 cm-D 2 0 moderator. Number of (n,2n) reactions using a D-T generator (BCL) E 150 C 100S50 E 20 z -lO -20 y-axis [cm] x-axis [cm] 150 - S100 50 E 1020 y-axis [cm] o E -20 8850 20 x-axis [cm] 10 10 20 "00 -20 y-axis [cm] x-axis [cm] Figure 4-29 : (n,2n) method: distribution of number of 14N(n,2n)13N reactions in the phantom The phantom was divided into three 5 cm-thick layers along the depth. From the upper to the lower, the investigated region in the phantom gets closer to the D-T neutron generator source used at BCL. The horizontal region examined in the simulation is 40 cm (x) x 30 cm (y). The phantom was divided into 36 cells in total and the number of (n,2n) reactions was calculated for each cell. 88 Number of (n,g) reactions using a 238Pu-Be source (BNL) x 10 s ............ 0 0 E10 z -10 -10 -20 y-axis [cm] C,) uc2 0 x-axis [cm] x 10s '' E L-0 " 0 zv.oc a) oc) 3 y-axis [cm] C,, cn2 x-axis [cm] x 10 s E M zz C u,2 o0 5. Ea 01 E -, 5 0 ......... ..... . z8 -5 -5 -10 -20 y-axis [cm] -10 1 x-axis [cm] Figure 4-30 : PGNAA method with 23 8Pu-Be source: distribution of number of 14 N(n,y) 15N reactions in the phantom The phantom was divided into three 5 cm-thick layers along the depth. From the upper to the lower, the investigated region in the phantom gets closer to the 238Pu-Be neutron source used at BNL. The horizontal region examined in the simulation is 40 cm (x) x 20 cm (y). The phantom was divided into 48 cells in total and the number of (n,y) reactions was calculated for each cell. 89 10.. x1 x 10s Number of (n,g) reactions using a 4 MeV p-Be accelerator source (BNL) >, (14 Do 5 E -5- z -1o -20 y-axis [cm] x-axis [cm] x 105 C', E " o} C cc2 o10 E zM y-axis [cm] x-axis [cm] x 10 s C 4 0 .0 E z -5 y-axis [cm] -10 -20 -10 0 10 x-axis [cm] Figure 4-31 : PGNAA method with 9 Be(p,n) source and BNL set-ups: distribution of number of 14N(n,y)1sN reactions in the phantom The phantom was divided into three 5 cm-thick layers along the depth. From the upper to the lower, the investigated region in the phantom gets closer to the neutron source. The horizontal region examined in the simulation is40 cm (x) x 20 cm (y). The phantom was divided into 48 cells in total and the number of (n,y) reactions was calculated for each cell. xNumber of (n,g) reactions using a 4 MeV p-Be accelerator source (LABA) x 10 (I,O3 U) 28 ............... . ... ........ ..............: .... :........ :...... 0 .00 E Z -10 -10 -20 y-axis [cm] Cn E 02 't x-axis [cm] 1" - i, 106... cc ................ . 10 ._10CD E z y-axis [cm] x-axis [cm] Co 3. 7... ...... U) . . as ...... ...... -10 x 0: ."10 E z 10 0 -5 -10 -20 y-axis [cm] [c0 x-axis [cm] Figure 4-32: PGNAA method with 9Be(p,n) source and LABA set-ups: distribution of number of 14 N(n,y)15 N reactions in the phantom The phantom was divided into three 5 cm-thick layers along the depth. From the upper to the lower, the investigated region in the phantom gets closer to the neutron source. The horizontal region examined in the simulation is 40 cm (x) x 20 cm (y). The phantom was divided into 48 cells in total and the number of (ny) reactions was calculated for each cell. 91 4.4.2.2. Detector efficiency Figure 4-17, 18 show the distribution of the absolute detector efficiency within the thigh phantom. These results were obtained by simulating counting systems of two methods: those employed at BCL and BNL. Because the detectors are positioned below the sample for the BCL counting system, the detector efficiency decreases with increasing depth. The BNL system also shows higher detector efficiencies for the cells closer to the detectors. Table 4-17 presents the average detector efficiency and its uniformity in the thigh phantom for the counting geometries used for the (n,2n) and the PGNAA methods. Facility BCL - (n,2n) method BNL - PGNAA method Average detector efficiency [%] 9.07 ± 0.033 0.58 ± 0.002 Uniformity [%] (# of data points) 69.45 ± 0.74 (36) 20.57 + 1.73 (48) Table 4-17 : Average detector efficiency and their uniformity in the thigh phantom for nitrogen measurement counting systems used at BCL ((n,2n)) and BNL (PGNAA). Thigh counting system at BCL - Detector efficiency: (n,2n) reaction 0.20.15 0o.o05 01- . 0 0 y-axis [cm] 0.2 10 20 -10 z-20 x-axis [cm] S................... y-axis (cm] x-axis [cm] -0.15- 0.05 . .. 0I -20 Z y-axis [cm) -10 0 10 -20 x-axis [cm] Figure 4-33.: (n,2n) method: detector efficiency distribution From the upper to the lower, the investigated region inthe phantom gets closer to the Nal(TI) detectors used at BCL. The horizontal region examined in the simulation is 40 cm (x) x 30 cm (y). The phantom was divided into 36 cells in total and the detector efficiency for 511 keV y-rays was calculated for each cell. Nitrogen counting system at BNL - Detector efficiency : (n,g) reaction C 0o C0 10 ... . 20 10 x-10 0.01 -20 y-axis [cm] ~~~~~..... - - x-axis [cm] -- i ' ' 0.005 0 -10 C B 4 -20 y-axis [cm) x-axis [cm] 0.01 0.005 0 l C 0 z y-axis [cm] -10 -20 x-axis [cm] Figure 4-34: PGNAA method: detector efficiency distribution From the upper to the lower figure, the investigated region in the phantom gets closer to the neutron source. Two Nal(TI) detectors are diagonally located in the plus y-region. The horizontal region examined in the simulation is 40 cm (x)x20 cm (y). The phantom was divided into 36 cells in total and the detector efficiency for 10.83 MeV y-rays was calculated for each cell. 4.4.2.3. Sensitivity The composite sensitivity was calculated as explained in 4.2.2. Those results are shown in Table 4-18. The uniformity of sensitivity for the bilateral measurement is also shown. To achieve the bilateral measurements for the (n,2n) method, the patient was assumed to be irradiated by two identical D-T neutron generators and detected by identical counting systems from both sides of the patient, although this measurement method is not practical for the current system. The irradiation time and counting time were both assumed to be 20 minutes. The distributions of sensitivity in the phantom for unilateral measurements are presented in Figure 4-19 - 4-22. Method and neutron source (n,2n) : D-T generator - prone irradiation (n,2n) : D-T generator - supine irradiation PGNAA : 2 38 Pu-Be PGNAA : 4 MeV p-Be Uniformity for bilateral irradiation [%] (# of data points) 29.46 ± 3.27 (36) Average composite sensitivity for unilateral irradiation [#/g/mSv] 5.0355 (0.74 %) Uniformity for unilateral irradiation [%] (# of data points) 33.92 ± 2.47 (36) 8.4779 (0.86 %) 97.12 ± 1.82 (36) 107.25 (0.81 %) 42.34 ± 2.31 (48) 27.83 ± 4.17 (48) 191.34 (0.92 %) 56.15 ± 2.21 (48) 27.65 + 3.74 (48) Table 4-18 : Average composite sensitivity and their uniformity in the thigh box phantom measured for the nitrogen measurement systems Sensitivity distribution : (n,2n) method - prone irradiation 3 c- ' ,, ii I: P-' -10 .20 10 0 -- 2 -A - 3 ; x-axis [cm] Sensitivity distribution : (n,2n) method - prone irradiation /9 L-31 '4- -2 -31 A -5 -15 5 15 y-axis [cm] Figure 4-35 : (n,2n) method: distribution of composite sensitivity in the phantom for prone position irradiation Investigated regions get closer to the D-T generator as '1' -> '3'. ' Sensitivity distribution : (n,2n) method - supine irradiation -A - -2 - 2 --- ---- ~---~ ' '~-~~r-~~-'- -10 -20 0 10 x-axis [cm] Sensitivity distribution : (n,2n) method - supine irradiation 25 U) EE20, 20 is- 15 o E 10- o -- ..-- "" ' '""" "" " -Imm 0 -15 -5 5 15 y-axis [cm] Figure 4-36 : (n,2n) method: distribution of composite sensitivity in the phantom for supine position irradiation Investigated regions get closer to the D-T generator as '1' -+ '3'. - 2 -- A - 3 Sensitivity distribution using a 238Pu-Be source : PGNAA I r 160 140120 100 - U-r -I- - - - - II - -m - 3 2 80 60 4020 S1 0 -20 x-axis [cm] Sensitivity distribution using a 238Pu-Be source : PGNAA - 180 160 140 120 100 -- -A- 80 60 40 20 0 -10 -5 0 5 10 y-axis [cm] Figure 4-37 : PGNAA method with 238Pu-Be source: distribution of composite sensitivity in the phantom Investigated regions get closer to the source as '1' -+ '3'. 2 3 3 Sensitivity distribution using a 4 MeV p-Be accelerator source : PGNAA 350 300 250 I-' 200 U-'- I- 150 -- 2 -- 3 - 2 --- 3 100 0 x-axis [cm] Sensitivity distribution using a 4 MeV p-Be accelerator source : PGNAA 350 360 - -- I " 300- -- 250 " 200- S / 150100 4r 50- -10 -5 IIs 0 5 10 y-axis [cm] Figure 4-38 : PGNAA method with 9Be(p,n) source and BNL system: distribution of composite sensitivity in the phantom Investigated regions get closer to the source as '1' -> '3'. 4.5 Summary In this section MCNP simulations of nitrogen measurements were conducted using various types of experimental systems. In the simulation the maximum skin dose delivered to the subject and the sensitivity to the nitrogen detection were calculated for in vivo measurement systems of the (n,2n) and the PGNAA methods. The measurement geometries and characteristics for all simulations were described and the results of dose and nitrogen sensitivity were presented. The investigation was made not only for the nitrogen measurement facilities in operation, but also for hypothetical accelerator systems, in order to examine the potential of using different types of neutron sources than those presently used. 100 5. Comparison and conclusion This chapter discusses the simulation results presented in chapter 4 and demonstrates how experimental parameters determined by investigators, such as irradiation and counting time, affect figures of merit of the two nitrogen measurement systems: the BCL (n,2n) system and the BNL PGNAA facility. Possible improvements which can be made on those systems are also described for future work. 5.1 Accuracy and reproducibility Table 5-1 presents a comparison between the nitrogen net counts of both methods which were calculated by MCNP simulation using a thigh box phantom (40 x 20 x 15 cm 3 ) in which 1.4414E25 nitrogen nuclei were contained. An irradiation time of 6 minutes and a counting time of 20 minutes were assumed for the (n,2n) method while a 200 sec-measurement was assumed for the PGNAA method. Method and facility BCL - (n,2n) method, D-T generator BNL - PGNAA method, 238Pu-Be Neutron yield [n/sec] 4.9891E7 2.0300E8 Nitrogen counts [#] (error bar) 5.1801E3 (3.14 %) 1.9494E4 (0.64 %) Table 5-1 : Nitrogen net counts for the (n,2n) method and the PGNAA technique which were obtained by MCNP calculation. As mentioned earlier, the background problem was excluded in the MCNP simulation for both nitrogen measurement methods. However, the signal-to-background ratio is a critical factor in evaluating accuracy and reproducibility of a measurement system because the standard deviation of the counting results is calculated as G +B where G is the gross counts and B is the background. Although the nitrogen counting for the (n,2n) method is performed in the whole body counting room well-shielded with steel and lead, the nitrogen peak at the low energy of 511 keV contains significant background contributions, which mainly come from natural radiation sources such as cosmic ray, terrestrial radiation, and radioactivity of the constituent materials of the measurement facility. Such high background in the low energy region is shown in Figure 3-5. On the other hand, the contribution from natural sources to the 10.83 MeV photopeak region for the PGNAA technique is negligible, while the dominant background contribution from the 238Pu-Be radiation source itself (due to the random summing of 4.4 MeV photons emitted from the 238Pu-Be source). 102 5.2 Irradiation and counting time The relationships of the net counts of nitrogen and the composite sensitivity with the measurement time can be obtained as shown in Figure 5-1, by use of following equations derived in chapter 4. Cn2n 6511 n2n=511 x Sn2n CPG =10.83 SPG =10.83 t )(e-t - e- t2 x R 13(1 - e 2 to )(e-2t - e mNn2n x dn2n x t o xA t2 X e 13 X e 13 - (8) 3 X e1 5 x R15 x t' (9) (10) x 1e 5 x R15(11) mNPG x dPG where C: nitrogen net count for the (n,2n) method ('n2n') and the PGNAA method ('PG') S: composite sensitivity for the (n,2n) method ('n2n') and the PGNAA method ('PG') s: detector efficiency at 511 keV and 10.83 MeV e: emission probability of photons of radioactive products; '3 N ('13') and 15N('15') R: reaction rate of 1'4N(n,2n)13N ('13') and 14N(n,y)' 5N ('15') m: nitrogen amount d: incident dose X: decay constant of 13N 103 (n,2n) method at BCL PGNAA at BNL 6000 180 160 5000 140 120- S 4000 0 o 100 o Ti 3000 80 60 40 22000 z I | --* irradiation (c.t.=20 min) p-C counting ( i.t.=20 min) 1000 ! u- 0 40 20 30 0 10 Irradiation / Counting time (i.t./ c.t.) [min] 10 20 30 40 Measurement time [min] PGNAA at BNL (n,2n) method at BCL 108.5 7, 6 108 P55 * E N -4 ,, . O 12 107.5 A A AA A 107 / 0'' /c co 2 1- A 106.5 I - * irradiation c.t.=20 min - counting (i.t.=20 min 106 0 10 20 30 40 Irradiation / Counting time (i.t./ c.t.) [min] 0 | I I 10 20 30 Measurement time [min] Figure 5-1 : Relationship between measurement time and figures of merit: nitrogen counts and sensitivity Nitrogen net counts are related to the accuracy and reproducibility of the system. 104 40 to: irradiation time for the (n,2n) method tl: counting starting time for the (n,2n) method t 2 : counting ending time for the (n,2n) method t': measurement time for PGNAA method The nitrogen counts are related to the accuracy and the reproducibility of the system. To determine the irradiation time for the (n,2n) method, a compromise between the net counts and the composite sensitivity needs to be made since they are related with the irradiation time in the opposite way. On the other hand, the longer counting time provides a greater gain in both the nitrogen counts and the composite sensitivity. It is important, however, to remember that the body elements which have long half-lives could present a major interference for the nitrogen measurement when the counting is conducted for a long time. It is observed that the effects of the irradiation and counting time above 20 minutes on these factors are not significant. For the PGNAA method, the composite sensitivity is unaffected by the measurement time while the nitrogen counts are proportionally related. One should note, however, that the nitrogen background increases proportionally with the measurement time as well and thus the nitrogen signal-to-background ratio is theoretically constant. The PGNAA system at BNL employs a 400 second-bilateral irradiation for one section of the body to obtain sufficient nitrogen counts for the estimation of nitrogen level in the subject. 105 5.3 Dose 5.3.1 Incident dose The equivalent dose of 0.7 mSv was calculated by the MCNP simulation for the PGNAA method from the simulation of a 400 second-bilateral irradiation of a 15 cmthick phantom using the weighting factors recommended by ICRP. This result is close to the 80 mrem reported by BNL researchers [42], although the method of calculating the equivalent dose is not explained in detail. As can be seen in Table 4-11, the MCNP simulation results showed that in vivo nitrogen measurement by the (n,2n) technique delivers about half of the skin equivalent dose delivered by the PGNAA method with the same irradiation time. The total equivalent dose is proportionally dependent on the irradiation time. Therefore a roughly 6 minute-unilateral irradiation for the (n,2n) method delivers the total dose of same level as the current PGNAA technique. 5.3.2 Dose distribution Figure 4-17 - 22, 24, 25, 27, 28 showed the highest dose in femur among three kinds of thigh tissues, confirming the fact that radiation is absorbed in denser tissues more efficiently. The exceptions were total neutron dose for the accelerator source shown in Figure 4-23, 26 which was most absorbed at the surface of the thighs. It can be 106 observed that the neutron spectrum of the 4 MeV 9Be(p,n) source is much softer than the 238Pu-Be source or the 14 MeV monoenergetic fast neutron source from Figure 4-15. These simulation results suggest that the relatively low energy neutrons (100 keV to 1 MeV) from the 4 MeV 9Be(p,n) source are attenuated primarily near the surface leaving predominantly thermal neutrons to interact in the inner medium. On the other hand, fast neutron components of the 238Pu-Be source and the 14 MeV neutron source produce high neutron flux within the subject which lead to high dose in condensed tissues. It is observed from Table 4-14 that, excluding the highest dose in femur, the (n,2n) system delivers higher fast neutron dose at the surface while the inner region is more affected by thermal neutrons for the PGNAA method. Comparing the maximum and minimum dose among three tissues in the thighs, the fast neutron dose decreased only by a factor of 4 from the surface nearest to the neutron beam to the furthest for the (n,2n) method, whereas the other three PGNAA systems showed 2 orders of magnitude difference. Now it can be assumed that the variation of activating neutron dose correlates with the activating neutron flux because the kerma factors is almost constant in the energy range defined for activating neutrons. Therefore these results suggest that the sensitivity in the human thighs is more uniform for the (n,2n) method than the PGNAA technique for unilateral irradiation. The use of the (n,2n) method, therefore, may be more suitable than the PGNAA method for absolute nitrogen measurement of human thighs. Table 4-14 also shows that the dose from y-rays for the accelerator source relative to total neutron dose is high compared to the other two fast neutron sources: the D-T generator and the 238Pu-Be radionuclide. Since the neutron energy spectrum of the 4 MeV 107 9Be(p,n) is much 'softer' than the other two sources, the high photon dose can be considered due to secondary y-rays from neutron capture reactions arising from the elements existing in the surrounding materials. Such thermal neutron reactions include 'H (2.23 MeV), 209Bi (0.32, 0.16, 4.17, 4.05 MeV), 7Li (2.03 MeV), and 'oB (4.43 MeV). Photons in the energy range of 4 - 7 MeV may interfere with nitrogen measurement because of random summing (See chapter 2.2). Comparing the BCL (Figure 4-17 - 19) and BNL systems (Figure 4-20 - 25), a superior collimating effect is obvious for the BNL PGNAA facility which has been designed and developed for the irradiation of sectioned parts of a human subject. As mentioned in chapter 3, the BCL system was originally developed for the in vivo measurement of carbon, hydrogen and oxygen. In these measurements a patient is scanned from the shoulders to the knees, where the collimating effect is not a determining factor for the evaluation of the measurement system. Considering the partial body irradiation for the thigh measurement, however, improvements need to be made for the current (n,2n) collimating system in order to reduce dose to the rest of the patient. 108 5.4 Sensitivity 5.4.1 Supine- and prone-position irradiation for (n,2n) method The supine-positioned irradiation showed about 70 % higher average sensitivity than the prone-position irradiation (Table 4-18). This is because the nitrogen detection is most sensitive in the region closest to the D-T generator and Nal(TI) detectors for the supine measurement. These data result in about three times lower uniformity of detection in the subject compared to the prone irradiation for unilateral measurement. The choice between prone or supine irradiation should be made according to the purpose of the measurement. When sequential measurements are intended rather than the absolute determination of nitrogen, the magnitude of sensitivity may be more important than the uniformity in order to acquire a desirable reproducibility of the measurements. However, when the absolute amount of nitrogen in the subject needs to be measured, the uniformity of sensitivity is a critical factor which should be fulfilled as well as the sensitivity yield. 5.4.2 Accelerator sources for PGNAA method Table 4-16 showed that all of the accelerator neutron sources investigated in the simulation provided higher yields of the number of 14N(n,2n)'SN reactions than the 2 38 pu- Be source with comparable uniformity using same irradiation geometries. Particularly the 3 MeV 9Be(p,n) source presented about three times greater nitrogen activating effect than 109 the 238Pu-Be source. Consequent calculations also showed higher sensitivity for the 4 MeV 9Be(p,n) accelerator source by a factor of 2 than the conventional source (Table 418). These results are very promising for the use of the LABA accelerator as a neutron source for in vivo nitrogen measurement. Nonuniformity of the number of '4 N(n,y)1SN reactions observed in the accelerator results is due to the relativaly large size of the phantom compared to the size of the collimator/reflector assembly. Better uniformity of the thermal neutron fluence in the subject may be achieved by using a moderator of adequate thickness and/or modifying the size of collimating systems for human study. 5.4.3 (n,2n) method and PGNAA technique Comparing the simulation data shown in Table 4-15 for the BCL and BNL irradiation systems, the number of 14N(n,2n)15N reactions are less by 3 orders of magnitude than 14N(n,y)' 5 N with the incident dose of same level. Considering that the number of reactions is proportional to the activating neutron fluence and the cross-section of the reaction, it can be estimated that thermal neutrons delivered to the subject by the BNL PGNAA system is by 2 orders of magnitude higher than fast neutrons delivered by the BCL (n,2n) facility. However the (n,2n) counting system designed for the thigh measurement provides about 20 times higher absolute detector efficiency than the PGNAA system (Table 4-17), because of the low energy photons which will be more efficiently detected. These simulation results provided about 10 - 20 times higher 110 sensitivity for the PGNAA method than the (n,2n) method, depending on the irradiation position of the patient, as presented in Table 4-18. Now it may be informative to examine the effects of the experimental parameters, such as measurement times, which were determined by investigators. For the (n,2n) method, those sensitivity results presented in chapter 4 for the 20 minute-irradiation and the 20 minute-counting could be theoretically improved by shortening the irradiation time and increasing the counting time, according to the relationships shown in Figure 5-1. However the sensitivity increases only by a factor of 2.5 even with an irradiation time of 40 minutes and a counting time of 1 minute with considerable loss in the accuracy and the reproducibility, which results in at the best 20 % of the sensitivity for the PGNAA technique. The hypothetical bilateral measurement by the BCL (n,2n) method showed sensitivity uniformity comparable to the PGNAA method. The use of the (n,2n) method for the absolute measurement of nitrogen may be possible and worth the effort to examine the feasibility as carried out by some investigators in other (n,2n) facilities [11, 25, 26, 36]. 5.5 Conclusion In this thesis comparisons between two methods for in vivo nitrogen measurement were made on the basis of MCNP simulation results, some of which were verified by nitrogen measurements performed at the BCL (n,2n) facility. In nitrogen measurement using the BCL system, only 4.1 % oxygen interference was found as opposed to the results obtained by other investigators (- 20 %). Incident dose delivered to a patient per unit irradiation time was about a factor of 2 higher for the PGNAA method employed at BNL than the BCL (n,2n) method. The sensitivity results, however, showed that the PGNAA method provides the nitrogen sensitivity yield which is 10 - 20 times higher than that of the (n,2n) method, depending on the patient position during the irradiation. Prone position irradiation for the (n,2n) method showed the excellent uniformity of sensitivity within the subject for unilateral irradiation (34 %) compared to the PGNAA method (42 %). The simulations with the BNL irradiation system using the proton accelerator source spectra showed high potential to employ the LABA accelerator as an alternative neutron source for nitrogen measurement in vivo. 112 5.6 Future work 5.6.1 (n,2n) method In order to improve sensitivity to nitrogen detection, some modifications can be made with the current BCL measurement system. Although the counting system which was constructed for the human thigh measurement provided significantly higher detector efficiency compared to that of the PGNAA detector arrangement, the counting system needs to be still more improved since the composite sensitivity is increased with the detector efficiency. The detector efficiency is increased by using multiple detectors. The use of two more NaI(TI) crystals identical to the two detectors currently employed is under consideration, each of those to be located by the side of the patient's thighs. This configuration is also expected to provide better uniformity of detector efficiency in the subject. For absolute determination of nitrogen further interference investigation should be made. Although the oxygen interference was estimated to be much smaller than expected by other investigators, the effect of differences in distribution of nitrogen and oxygen in the human body needs to be still examined since a homogenous phantom would be used to determine a calibration factor which converts the nitrogen level in the phantom to that in the patient. Furthermore, the bilateral irradiation system is imperative for absolute measurement although it seems difficult to install another D-T neutron generator above a patient with bulky collimator/reflector material in addition to the present system. 113 5.6.2 PGNAA method After modified and upgraded several times by dedicated researchers, there remains very little which can be done to improve the performance characteristics of BNL PGNAA facility. Although the better uniformity of the composite sensitivity is desired for the thigh measurement, further modifications of the system, such as changing the thickness of D2 0 moderator, would not necessarily improve the sensitivity uniformity for other parts of the body. It should be noted, however, that the graphite collimator which defines the neutron beam window may be a major interference for nitrogen measurement, because 4.43 MeV photons emitted by the inelastic reaction 12 C(n,n')' 2C (threshold = 4.8 MeV, a = 0.2 b) cause random summing which may contribute to 10.83 MeV nitrogen background. Despite the advantages of the low cost, relatively light weight, and good moderating and reflecting properties (a high elastic and inelastic neutron cross section), the use of graphite as a part of the nitrogen measurement system may need to be reconsidered. The use of an accelerator as a neutron source for the PGNAA method may be desirable not only for because sensitivity may be possible, as demonstrated in this thesis, but also because a number of neutron spectra can be generated with neutron energies below the threshold energy of the inelastic reactions of carbon. 114 6. References [1] Ettinger K V, Biggin H C, Chen N S, Ettinger K V, Fremlin H, Morgan W D, Nowotny R and Chamberlain M J 1975 In vivo neutron activation analysis of nitrogen using capture gamma rays Kerntechnik 17 89-92. [2] Vartsky D, Ellis K J and Cohn S H 1979 In vivo measurement of body nitrogen by analysis of prompt gamma from neutron capture JNucl Med 20 1158-65. [3] McNeill K G, Mernagh J R, Jeejeebhoy N, Wolman S L and Harrison J E 1979 In vivo measurement of body protein based on the determination of nitrogen by prompt y analysis Am J Clin Nutr 32 1955-61. [4] Cohn S H, Gartenhaus W, Sawitsky A, Rai K, Zanzi I, Vaswani A, Ellis K J, Yasumura S, Cortes E and Vartsky D 1981 Compartmental body composition of cancer patients by measurement of total body nitrogen, potassium, and water Metabolism 30 222-9. [5] Spinks T J, Coode A W, Ranicar A S O and Steere E 1984 Measurement of total body nitrogen and oxygen by irradiation with cyclotron neutrons and 'delayed' gamma-ray counting Phys Med Biol 29 385-94. [6] Krishnan S S, Mcneill K G, Mernagh J R, Bayley A J and Harrison J E 1990 Improved clinical facility for in vivo nitrogen measurement Phys Med Biol 35 489-99. [7] Mackie A, Cowen S and Hannan J 1990 Calibration of a prompt neutron activation analysis facility for the measurement of total body protein Phys Med Biol 35 613-24. [8] Baur L A, Allen B J, Rose A, Blagojevic N and Gaskin K J 1991 A total body nitrogen facility for pediatric use Phys Med Biol 36 1363-75. [9] Vartsky D and Goldberg M B, "Body composition studies - neutron activation analysis vs. nuclear resonance absorption," presented at International conference: Neutrons in research and industry, Crete, Greece, 1996. 115 [10] Fruton J S and Simmonds S, "General biochemistry,", 2nd ed. New York: John Wiley, 1959. [11] Oxby C B, Appleby D B, Brooks K, Burkinshaw L, Krupowicz D W, McCarthy I D, Oldroyd B and Ellis R E 1978 A technique for measuring total-body nitrogen in clinical investigations using the 14N(n,2n)13N reaction Int JAppl RadiatIsot 29 205-11. [12] Vartsky D, Prestwich W V, Thomas B J, Dabek J T, Chettle D R, Fremlin J H and Stammers K 1979 The use of body hydrogen as an internal standard in the measurement of nitrogen in vivo by prompt neutron capture gamma ray analysisJ RadioanalChem 48 243-52. [13] Ellis K J, Yasumura S, Vartsky D, Vaswani A N and Cohn S H 1982 Total body nitrogen in health and disease: effects of age, weight, height, and sex JLab Clin Med 99 917-26. [14] Beddoe A H, Zuidmeer H and Hill G L 1984 A prompt gamma in vivo neutron activation analysis facility for measurement of total body nitrogen in the critically ill Phy Med Biol 29 371-83. [15] Chettle D R and Fremlin J H 1984 Techniques of in vivo neutron activation analysis Phy Med Biol 29 1011-43. [16] Cohn S H and Parr R M (ed): Report of an Advisory Group of IAEA 1985 Nuclear-based techniques for the in vivo study of human body composition Clin Phy Physiol Meas 6 275-301. [17] Cohn S H 1980 The present state of in vivo neutron activation analysis in clinical diagnosis and therapy Atom Ener Rev 18 599-660. [18] Vartsky D, Ellis K J, Vaswani A N, Yasumura S and Cohn S H 1984 An improved calibration for the in vivo determination of body nitrogen, hydrogen and fat Phys Med Biol 29 209-18. [19] Ellis K J 1993 In vivo activation analysis: present and future prospects J RadioanalNucl Chem 169 291-300. [20] Mernagh J R, Harrison J E and McNell K G 1977 In vivo determination of nitrogen using Pu-Be sources Phys Med Biol 22 831-5. 116 [21] Allen B J, Blagojevic N, McGregor B J, Parsons D E, Gakin K, Soutter V, Waters D, Allman M, Stewart P and Tiller D 1987 In vivo determination of protein in malnourished patients In vivo body composition studies 77-82. [22] Mackie A, Hanna W J, Smith M A and Tothill R 1988 Apparatus for the measurement of total body nitrogen using prompt neutron activation analysis with 252 Cf Med Eng [23] Technol 12 152-9. Burkinshaw L, Morgan D B, Silverton N P and Thomas R D 1981 Total body nitrogen and its relation to body potassium and fat-free mass in healthy subjects Clin Sci 61 457-62. [24] Battye C K , Knight V , Marshall T O, Knight A and Godfrey B E 1971 Use of 14 MeV neutrons for activation analysis of five elements in man-like models Nucl Instr Meth 92 601-7. [25] Williams E D, Boddy K, Harvey I and Haywood J K 1978 Calibration and evaluation of a system for total body in vivo activation analysis using 14 MeV neutrons Phys Med Biol 23 405-15. [26] Haywood J K, Williams E D, McArdle F J and K Boddy 1980 Reliability of absolute and relative measurement of total body nitrogen by the 14N(n,2n)13N reaction Phys Med Biol 26 591-602. [27] Biggin H C, Chen N S, Ettinger K V, Fremlin H, Morgan W D, Nowotny R and Chamberlain M J 1972 Determination of nitrogen in living patients Nature New Bio 236 187-8. [28] Ryde S J S, Morgan W D, Evans C J and Dutton J 1987 A clinical instrument for multi-element in vivo analysis by prompt, delayed and cyclic neutron activation using 252 Cf Phys Med Biol 32 1257-71. [29] Cohn S H and Dombrowski C S 1971 Measurement of total body calcium, sodium, chlorine, nitrogen and phosphorus in man by in vivo neutron activation analysis JNucl Med 12 499-505. [30] Vartsky D, Thomas B J and Prestwich W V 1976 Comparison of the spatial uniformity of sensitivity of neutron activation techniques for whole body nitrogen measurements Kerntechnik 18 304-7. 117 [31] Leach M O, Thomas B J and Vartsky D 1977 Total body nitrogen measured by the 14N(n,2n) 13N method: A study of the interfering reactions and the variation of spatial sensitivity with depth Int JAppl Radiat Isot 29 263-9. [32] Williams E D, Boddy K and Haywood J K 1977 Measurement of neutron fluence in total body in vivo neutron activation analysis Phys Med Biol 22 1003-6. [33] Elliott A, Holloway I, Boddy K, Haywood J K and Williams D 1978 Neutron uniformity studies related to clinical total body in vivo neutron activation analysis Phys Med Biol 23 269-81. [34] Ryde S J S, Morgen W D, Evans C J, Sivyer A and J Dutton 1989 Calibration and evaluation of a 252Cf-based neutron activation analysis instrument for the determination of nitrogen in vivo Phys Med Biol 34 1429-41. [35] Stametelatos I E, Dilmanian F A, Ma R, Lidofsky L J, Weber D A, Pierson Jr R N, Kamen Y and Yasumura S 1993 Calibration for measuring total body nitrogen with a newly upgraded prompt gamma neutron activation facility Phys Med Biol 38 615-26. [36] McCarthy I D, Shafari A, Oxby C B and Burkinshaw L 1980 The accuracy of total body nitrogen determined by neutron activation analysis Phy Med Biol 25 849-63. [37] Sharafi A, Pearson D, Oxby C B, Oldroyd B, Krupowicz D W, Brooks K and Ellis RE 1983 Multi-element analysis of the human body using neutron activation Phys Med Biol 28 203-14. [38] MacFie J and Burkinshaw L 1987 Body composition in malignant disease Metabolism Sci Eng 36 290-4. [39] Boddy K, Holloway I and Elliott A 1973 A simple facility for total body in vivo activation analysis Int JAppl Rad Isot 24 428-30. [40] Harvey T C, Dykes P W, Chen N S, Ettinger K V, Jain S, James H, Chettle D R, Fremlin J M and Thomas B J 1973 Measurement of whole-body nitrogen by neutron-activation analysis Lancet 2 395-9. [41] Dilmanian F A, Weber D A, Yasumura S, Kamen Y, Lidofsky L J, Heymsfield S B, Pierson Jr R N, Wang J, Kehayias J J and Ellis K J 1990 Performance of the 118 delayed- and prompt-gamma neutron activation systems at Brookhaven National Laboratory Advances in Body Composition Studies 309-15. [42] Dilmanian F A, Lidofsky L J, Stametelatos I E, Kamen Y, Yasumura S, Vartsky D, Pierson Jr R N, Weber D A, Moore R I and Ma R 1998 Improvement of the prompt-gamma neutron activation facility as Brookhaven National Laboratory Phys Med Biol 43 339-49. [43] McNeill K G, Borovinicar D J, Krishnan S S, Wang H, Waana C and Harrison J E 1989 Investigation of factors which lead to the background in the measurement of nitrogen by IVNAA Phys Med Biol 34 53-9. [44] McGregor B J and Allen B J 1987 Monte Carlo study of neutron and gamma ray transport in the AAEC total body nitrogen facility Aust Phys Eng Sci Med 10 15561. [45] Vartsky D, Engler G and Goldberg M B 1994 A method for detection of explosives based on nuclear resonance absorption of gamma rays in 14N Nucl Instr Meth Phys Res A 348 688-91. [46] Boddy K, Elliott A, Robertson I, Mahaffy M E and Holloway I 1975 A high sensitivity dual-detector shadow-shield whole-body counter with an 'invariant' response for total body in vivo neutron activation analysis Phys Med Biol 20 296304. [47] Vartsky D, Ellis K J, Wielopolski L and Cohn S H 1983 The spatial shape and the components of the background in in vivo neutron capture prompt gamma ray analysis Nucl Instr Meth 213 437-46. [48] Cohn S H 1981 In vivo neutron activation analysis: State of the art and future prospects Med Phys 8 145-54. [49] Stamatelatos I E M, Chettle D R, Green S and Scott M C 1992 Design studies related to an in vivo neutron activation analysis facility for measuring total body nitrogen Phys Med Biol 37 1657-74. [50] Morgan W D, Vartsky D, Ellis K J and Cohn S H 1981 A comparison of 252 Cf and 238Pu, Be neutron sources for partial-body in vivo activation analysis Phys Med Biol 26 413-24. 119 [51] Yasumura S, Ma R, Weber D A, Dilmanian F A, Zhao M, LoMonte A F, Moore R I, Skelany R and Pierson Jr R N 1997 A modified bottle manikin phantom for in vivo neutron activation analysis Health Phys 72 443-9. [52] Vaughan B E and Boling E A 1961 Rapid assay procedures for tritium-labeled water in body fluids J Lab & Clin Med 57 159-64. [53] Reid J T, Breidenstein B C, Stonaker H H, Hansard S L and Zobrisky S E, Body composition in animal and man. University of Missouri, Columbia: National Academy of Sciences, 1967. [54] Kehayias J J, Ellis K J and Cohn S H 1987 Use of high repetition rate generator for in vivo body composition measurements via neutron inelastic scattering Nucl Instr Meth Phys Res B 24/25 1006-9. [55] Kehayias J J, Heymsfield S B, LoMonte A F, Wang J and Pierson Jr R N 1991 In vivo determination of body fat by measuring total body carbonAm J Clin Nutr 53 1339-44. [56] Kehayias J J and Zhuang H 1993 Use of the Zetatron D-T neutron generator for the simultaneous measurement of carbon, oxygen, and hydrogen in vivo in humans Nucl Instr andMeth B 79 555-9. [57] Kehayias J J, Zhuang H, Dowling L, Ma R and Moore R I 1994 Choice of detectors for in vivo elemental analysis by counting natural and neutron-induced gamma rays for medical applications Nucl Instr Meth Phys Res A 353 444-7. [58] ICRP-23, "Report of the task group on reference man," Pergamon Press, Oxford 1975. [59] Briesmeister J F, "MCNP - A general Monte Carlo N-particle transport code,", 4B ed: Los Alamos National Laboratory, 1997. [60] ZamenhofR G, Murray B W, Brownell G L, Wellum G R and Tolpin E I 1975 Boron neutron capture therapy for the treatment of cerebral gliomas. I: Theoretical evaluation of the efficacy of various neutron beams Med Phys 2 47-60. [61] Deutsch O L and Murray B W 1975 Monte Carlo dosimetry calculation for boron neutron capture therapy in the treatment of brain tumors Nucl Tech 25 320-39. 120 [62] Yanch J C, Zhou X-L and Browell G L 1991 A Monte Carlo investigation of the dosimetric properties of monoenergetic neutron beams for neutron capture therapy Rad Res 126 1-20. [63] ICRP-60, "1990 Recommendations of the international commission on radiological protection," Pergamon press 1991. [64] ICRU-44, "Tissue substitutes in radiation dosimetry and measurement," International comission on radiation units and measurement 1989. [65] ICRU-46, "Photon, electron, proton and neutron interaction data for body tissues," International comission on radiation units and measurement 1992. [66] Griffith R V, Slaughter D R, Patterson H W, Beach J L, Frank E G, Rueppel D W and Fisher J C, "Multi-technique characterization of neutron fields from moderated 252 Cf and 238PuBe sources," presented at International Symposium on National and international Standarization of Radiation Dosimetry, Atlanta, 1978. [67] Geiger K W and Van der Zwan L 1975 Nucl Instr Meth 131 315-. [68] Yanch J C, Shefer R E, Klinkowstein R E, Howard W B, Song H, Blackburn B and Binello E 1997 Research in boron neutron capture therapy at MIT LABA Appl Acce Res Ind 1281-4. [69] Shefer R E, Klinkowstein R E, Yanch J C and Howard W B 1994 Tandem electrostatic accelerators for BNCT Proceedingsof thefirst international workshop on accelerator-basedneutron sourcesfor boron neutron capture therapy, INEL CONF-940976 89-97. [70] Yanch J C, Zhou X-L, Shefer R E and Klinkowstein R E 1992 Accelerator-based epithermal neutron beam design for neutron capture therapy Med Phys 19 709-21. [71] Howard W B, Yanch J C, Grimes S M, Massey T N, AI-Quraishi S I, Jacobs D K and Brient C E 1996 Measurement of the 9 Be(p,n) thick target spectrum for use in accelerator-based Boron Neutron Capture Therapy Med Phys 23 1233-5.