Novel Endoscopes for Microscopic Assessment of Airway Clearance using Micro-Optical Coherence ARCHIES Tomography MASSACHUSETTS INSTITUTE OF TECHNOLOLGY by APR 14 2015 Carolin Isabella Unglert, PhD LIBRARIES Submitted to the Harvard-MIT Division of Health Sciences and Technology in partial fulfillment of the requirements for the degree of Master of Science at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY February 2015 Massachusetts Institute of Technology 2015. All rights reserved. Signature redacted A u th or ............ ................... Harvard-MIT Divi~s'6 Health Sciences and Technology Signature redacted C ertified by .......... Accepted by .... February 2, 2015 .................. Guillerm4. Tearney, MD, PhD Professor of Pathology, Harvard Medical School Thesis Supervisor Signature redacted.............. Emery N. Brown, MD, PhD Direc or, Harvard-MIT Program in Health Sciences and Technology/Professor of Computational Neuroscience and Health Sciences and Technology Novel Endoscopes for Microscopic Assessment of Airway Clearance using Micro-Optical Coherence Tomography by Carolin Isabella Unglert, PhD Submitted to the Harvard-MIT Division of Health Sciences and Technology on February 2, 2015, in partial fulfillment of the requirements for the degree of Master of Science Abstract The health of the human respiratory system depends critically on airway clearance via motile hair-like structures (cilia), which transport and eliminate unwanted particles trapped within mucus. Impairment of mucociliary clearance (MCC) can lead to lifethreatening airway narrowing and lung infections, and is a major cause of morbidity and mortality in patients with cystic fibrosis, primary ciliary dyskinesia and chronic obstructive lung disease. However, no tool for microscopic in-vivo visualization of ciliary function is currently available, limiting studies of disease pathogenesis, refined diagnosis and phenotyping, and the development of novel therapeutics. In this thesis, a novel, 1-pm resolution, optical interferometric imaging technique termed Micro-OCT was incorporated into miniaturized common-path endoscopes and mucociliary transport was visualized in vivo for the first time. The first-generation Micro-OCT probe had a rigid design with outer diameter of 4 mm and a two-prism configuration providing beam splitting and sample beam shaping into an annular profile. Image quality of the probe allowed visualization of the periodic pattern of ciliary beating, measurement of airway surface liquid depth (ASL) and visualization of mucociliary transport. Unaltered ciliary function was demonstrated in a living, spontaneously breathing swine model. Newer generation common-path endoscope designs were demonstrated that improve, among other limitations, the stability of the reference reflector position and provide greater potential for miniaturization. The presented work opens unprecedented avenues for studying MCC and the effect of novel therapeutics within the complexity of a living organism. Further, it lays the groundwork for the development of a human probe with the potential to revolutionize diagnosis, phenotyping, and therapy management for all patients with respiratory disease involving the mucociliary escalator. Thesis Supervisor: Guillermo J. Tearney, MD, PhD Title: Professor of Pathology, Harvard Medical School 3 4 Acknowledgments First and foremost I would like to thank my advisor Prof. Gary Tearney for the opportunity to work in his unique lab and for his continuous and generous support, which has enabled me to grow as a researcher and to pursue the work that I was passionate about. Not only have I found my calling as a translational scientist here, but also a second family and I would like to express my sincerest thank you to everyone I have had the chance to work and interact with. I am deeply indebted to Dr. Ken Chu for his most generous support. It was a true honor to work with Ken and his patient teaching, immense knowledge, and great sense of humor really enabled this thesis. I would also like to thank the other members of the Micro-OCT pulmonary group, Drs. Tim Ford and Kanwarpal Singh, for their teaching and support and for making the animal studies and lunch breaks most enjoyable. A critical work load of this project was achieved in collaboration with our fantastic engineering team. I would like to wholeheartedly thank Rob Carruth and Weina Lu for sharing their skills and for making the impossible possible on a regular basis. I could not be more grateful to have had the unique chance to work with and learn from Dr. Mireille Rosenberg and her clinical team. I owe Mireille everything I know about translational science and the regulatory landscape, and I am tremendously grateful for her mentorship, and career and life guidance. I would also like to particularly thank Drs. Michalina Gora and Aubrey Tiernan, who have been a critical help with understanding and advancing in the medical device design process. Absolutely critical to the success of my work in the lab and of this thesis were our assistants Valerie Madden and Princess Cruz. Val, my sincerest thank you for your smile and patience during all these years of scheduling and rescheduling meetings. Your organizational superpowers and personal warmth made the whole difference! Many, many others in the lab also made a big difference in my daily work and life and every contribution was absolutely essential. To just name a few: Dr. Mohini Lutchman continuously offered an open ear and generous treats to make life sweeter. 5 And I consider myself fortunate to have had the chance to be an advisor and mentor to Diana Mojahed, whose enthusiasm and scientific talent are truly enjoyable. In addition to the lab, I have had the chance to receive invaluable support from my department and fellow students. I would like to thank the HST faculty and HST academic office and in particular Profs. Elazer Edelman, Elfar Adalsteinsson, Martha Gray, Brett Brouma, and Julie Greenberg, as well as Laurie Ward, Traci Anderson, and Joe Stein. I have been continuously impressed by their open ears, helpful advice, and strong support. Last, but not least, I would like to express my sincerest thanks to the funding sources that have enabled my work and study, namely a generous gift from Air Liquide, a 9-month fellowship from HST, RA support from Prof. Gary Tearney and the NIH, and the SPIE Optics and Photonics Education Scholarship. 6 Contents 1 15 Introduction 21 2 Background Mucociliary clearance . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1 2.1.1 Spatial and temporal requirements for visualization of mucocil. iary function . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2.1 Primary ciliary dyskinesia 24 2.2.2 Cystic fibrosis . . . . . . . . 26 2.2.3 Chronic obstructive pulmonary disea se . . . . 27 Current methods characterizing ciliary func tion . . . 29 . . 29 2.3.2 Fluorescence confocal microscopy 29 2.3.3 High-speed video microscopy . . . 30 2.3.4 Radioaerosol clearance . . . . . . 30 2.3.5 Micro-OCT . . . . . . . . . . . . 30 2.3.6 Summary . . . . . . . . . . . . . 31 . 31 Micro-Optical Coherence Tomography 2.4.1 . . . . . Electron microscopy . . . . . . . . 2.3.1 Introduction to Optical Coherence Tomography (OCT) and Micro-OCT . . . . . . . . . . . . 31 2.4.2 Axial resolution . . . . . . . . . . 33 2.4.3 Tranverse resolution versus depth of focus in OCT . 7 . . . . . . 2.4 24 . 2.3 23 Diseases of mucociliary clearance . 2.2 22 34 Annular sample beam and aperture improve both lateral resolution and depth of focus in Micro-OCT 2.4.5 . . . . . . . . . . . . 35 Design requirements for a miniaturized Micro-OCT probe to visualize MCC in vivo . . . . . . . . . . . . . . . . . . . . . . 37 3 Miniaturized Micro-OCT probe for real-time visualization of mu41 3.1 M otivation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 3.2 Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 . . cociliary function in vivo 3.2.1 Common-path optical probe design with annular sample beam profile 3.2.2 Probe scanning . . . . . . . . . . . . . . . . . . . . . . . . . 43 3.2.3 Outer housing design characteristics . . . . . . . . . . . . . . 43 3.2.4 Fabrication of prototype . . . . . . . . . . . . . . . . . . . . 44 3.2.5 Experimental design of animal study . . . . . . . . . . . . . 45 3.2.6 Data processing . . . . . . . . . . . . . . . . . . . . . . . . . 46 R esults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 3.3.1 Miniaturized rigid probe . . . . . . . . . . . . . . . . . . . . 46 3.3.2 Lessons from the animal study . . . . . . . . . . . . . . . . . 47 3.3.3 Images of MCC in swine trachea 48 . . . . . . . . . 42 . . . . . . . . . . . . . . . . 3.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D iscussion . . . . . . . . . . . . . . . . . . . . 48 3.5 C onclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 . 3.4 57 4.1 M otivation . . . . . . . . . . . . . . . . . . . . . 57 4.2 M ethods . . . . . . . . . . . . . . . . . . . . . . 58 4.2.1 Integrated reference reflector design . . . 58 4.2.2 Semi-flexible probe design . . . . . . . . 59 4.2.3 Improvements to the outer housing . . . 59 4.2.4 Animal study . . . . . . . . . . . . . . . 60 Results . . . . . . . . . . . . . . . . . . . . . . . 60 4.3 . . . . . . 4 Novel common-path Micro-OCT probe designs . ............ 2.4.4 8 5 4.3.1 Second-generation Micro-OCT probe . . . . . . . . . . . . . . 60 4.3.2 First image of tracheal epithelium in living swine model . . . . 61 4.4 Limitations and future work . . . . . . . . . . . . . . . . . . . . . . . 61 4.5 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 67 Summary and Outlook 5.1 Thesis summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 5.2 N ext steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 5.3 Im pact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 9 10 List of Figures 1-1 Schematic illustrating common techniques for studying mucociliary clearance and the advantages of Micro-OCT. . . . . . . . . . . . . . . 17 1-2 Thesis summary schematic and outlook. 19 2-1 Qualitative comparison of axial intensities from the focus for a circular . . . . . . . . . . . . . . . . and annular apertures of different inner diameters. . . . . . . . . . . . 2-2 Qualitative comparison of radial intensities at the focus for a circular and annular apertures of different inner diameters. . . . . . . . . . . . 3-1 . . . . . . . . . . . . . . . . . . . . . . . . 52 Schematic illustration of main assembly steps of Micro-OCT optical probe subassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 37 Schematic and pictures of common-path configuration and components of first Micro-OCT probe. 3-2 36 53 Picture of a bronchoscopic image during balloon inflation for stabilization of the Micro-OCT probe during image acquisition with respect to the tissue region of interest. 3-4 . . . . . . . . . . . . . . . . . . . . . . . Picture of the optical probe subsystem, which is the scanning portion of the probe, and the donut-shaped beam profile. 3-5 . . . . . . . . . . . 54 Image of mucociliary clearance in freshly excised swine trachea obtained with first-generation Micro-OCT probe. . . . . . . . . . . . . . 3-6 54 55 Image of mucociliary clearance in vivo obtained with first-generation Micro-OCT probe in swine trachea. . . . . . . . . . . . . . . . . . . . 11 55 4-1 Schematic and pictures of common-path configuration and components of second-generation Micro-OCT probe. . . . . . . . . . . . . . . . . . 4-2 64 Schematic illustrating the knife edge test for evaluation of transverse resolution achieved with second generation Micro-OCT probe. .... 65 4-3 First in-vivo image obtained with second-generation Micro-OCT probe. 65 4-4 Schematic illustrating the three generations of Micro-OCT probe designs. 66 12 List of Tables 2.1 Design requirements of an imaging technique for visualization of mucociliary function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2 Comparison of governing parameters of spatial resolution and depth of focus in confocal microscopy, OCT, and Micro-OCT. 2.3 23 . . . . . . . . . 35 Design requirements for a miniaturized Micro-OCT probe to visualize mucociliary function in living swine trachea. . . . . . . . . . . . . . . 13 39 14 Chapter 1 Introduction The health of the human respiratory system depends critically on airway clearance via motile hair-like structures (cilia), which transport unwanted particles and infectious agents trapped within secreted mucus so that they can be swallowed or expectorated. Impairment of mucociliary clearance (MCC) can lead to life-threatening airway narrowing and lung infections, which are the primary cause of morbidity and mortality in patients with cystic fibrosis (CF) or primary ciliary dyskinesia (PCD) and potentially other diseases such as chronic obstructive pulmonary disease as well as in patients with acute infections [1, 2]. Great differences exist in patient presentation and likely in the exact defect and disease progression of any given patient, but no tool currently exists to evaluate mucociliary function on the ciliary level in a given patient or even in vivo. Diagnosing the precise mucociliary defect can therefore be difficult and even once a diagnosis has been made, the therapeutic management is often not targeted, but comparably uniform due to limited phenotyping and therapeutic options. In addition, researching defects of mucociliary function in animal models and the development of novel therapies is hindered by the inability to visualize MCC and the physiologic cause-effect relationships in these models in vivo. A tool that could visualize the micro-structure and function of MCC in vivo and in real-time could therefore significantly increase our understanding of disease pathogenesis and accelerate the development of new therapies, for example, in combination with the recently developed CF pig. In the clinic, such a tool could provide completely 15 new venues for diagnosis, phenotyping, and personalized therapeutic management. Currently available techniques to study MCC, however, are all limited to investigating excised tissue or cells, and typically provide only a single parameter of interest instead of a comprehensive picture of the interaction between mucus properties and ciliary function. Key parameters of mucociliary function include the periciliary layer (PCL), the airway surface liquid (ASL) depth, which is the combined PCL and mucus layer, ciliary beat frequency (CBF) and mucociliary transport (MCT) rate. Those main techniques, illustrated in Figure 1-1, include: 1. Electron microscopy, which can illuminate ciliary ultrastructure, periciliary, and airway surface liquid depth (PCL and ASL), but is limited to providing static images of processed cells and tissue [3, 4, 5]. 2. Fluorescence confocal microscopy, which also provides static images and measures of PCL and ASL in excised tissue and cell cultures [5]. 3. High-speed videomicroscopy, which allows dynamic, qualitative assessment of ciliary beating patterns and quantitative assessment of ciliary beat frequency (CBF), but is also limited to analyzing excised tissue specimen or cell cultures [6, 7]. 4. Radioaerosol clearance [8] and CT particle tracking [9], which provide information about the global efficiency and localized pattern of mucociliary transport (MCT), respectively, but require exogenous contrast and do not illuminate the mechanism itself or further correlate MCT to mucus properties. Micro-OCT [10] is a novel, 1- tim resolution optical imaging technique, which was recently developed in our lab and is capable of simultaneous visualization of several key parameters governing the function of the respiratory epithelium including PCL, ASL, CBF and MCT [11, 12]. Thus, Micro-OCT is unique as it provides intrinsic registration in time and location of those parameters. Even though Micro-OCT can in principle be utilized in vivo, all studies with this technology to date have been limited to imaging excised tissue and cultured cells on a bench-top system. 16 1. Electron microscopy 2. Confocal microscopy 3. Video microscopy <KV Excised tissue or Cell cultures 4. Radioaerosol clearance test CT particle tracking & Alm Nasal brushing or Biopsy X -N 0- ASL 1 PCL 10 ASL 0 PCL 5. MicroOCT P Ciliary beat pattern P CBF l MCT D ASL P PCL o Ciliary beat pattern o CBF o Local MCT Figure 1-1: Schematic illustrating common techniques for studying mucociliary clearance and the advantages of Micro-OCT. Current techniques for studying ciliary function are limited to the study of cell cultures or excised tissue and most provide only a few parameters of ciliary behavior each. Micro-OCT is capable of providing several key parameters of ciliary function for a more comprehensive assessment of ciliary function. Example images illustrating electron microscopy [5], fluorescence confocal microscopy [5], video microscopy [6], radioaersol clearance [8], CT particle tracking [9], and Micro-OCT [11] are reproduced from the respective references. The goal of this thesis project was to build a miniaturized Micro-OCT probe and to demonstrate visualization of MCC in vivo. A 4-mm outer diameter (OD), 50- cm length rigid probe was built and ciliary activity and mucociliary transport were visualized for the first time within a living swine trachea. The swine model was chosen for the resemblance of its pulmonary anatomy to humans and to establish the technique as a, novel research tool in combination with the CF pig. A second objective of this thesis work was to lay the groundwork for the development of a clinical tool for two different applications: First, we envision to use our rigid 17 probe design for assessment of MCC in the human nasal cavity. Used similarly to currently used procedures, such as nasal brushings, access to the respiratory epithelium of the nasal cavity could provide insights into mucociliary function non-invasively and allow real-time assessment of a therapeutic response. Therefore, the probe for in-vivo use has to be stable, safe, and small enough to be comfortable. Secondly, we hope to extend our reach for assessment of MCC in lower airway generations. Therefore, the probe needs to be flexible and small enough to be inserted through the accessory channel of a commonly used bronchoscope. The scope of this thesis therefore encompasses co-development and fabrication of the first generation Micro-OCT probe, an experimental setup for in-vivo imaging of MCC, and development of a second-generation semi-flexible Micro-OCT probe design with improved stability and potential for further miniaturization. Figure 1-2 summarizes the project schematically. Chapter 2 provides a brief background on the clinical motivation, current state of the art in research methods of mucociliary clearance, an introduction to Micro-OCT, and design requirements for a miniaturized probe. The design and fabrication of the first-generation Micro-OCT endoscopes is explained in Chapter 3 and images obtained with the probe in swine trachea ex vivo and in vivo are shown. Limitations of the first-generation probe design are discussed. Chapter 4 illustrates improved designs incorporated into second- and third-generation probes as the groundwork for human studies and assessment of lower airway generations. Finally, Chapter 5 summarizes the work conducted to date and presents an outlook for future technology development and clinical investigations. 18 * PC Benchtop setup A Achieved Milestones cipp - Co-development and l Fabrication of miniaturized, hand-held Micro-OCT probe 4N EC --- -- WI4 l Real-time visualization of MCC in living swine ~IZIZThLI1 TZI Outlook Novel probe designs: - semi-flexible - improved reference reflector stability - potential for further miniaturization 4 lo Navigation into smaller airways within accessory channel of bronchoscope N Non-invasive assessment of MCC in human nasal cavity Figure 1-2: Thesis summary schematic and outlook: A miniaturized, hand-held, commonpath Micro-OCT probe was built that incorporates beam splitting, shaping, and focusing as well as scanning functionality. The probe was demonstrated for visualization of mucociliary clearance (MCC) in a swine model in vivo. Lastly, novel probe designs were developed and tested with improved stability, flexibility, and potential for further miniaturization. The achieved milestones lay the groundwork for further studies in the lower airways as well as in the human nasal cavity. MCC: Mucociliary clearance. 19 20 Chapter 2 Background The importance of mucociliary clearance (MCC) as a defense mechanism of the lung is easily understood by noting the life expectancy of cystic fibrosis patients with severely impaired MCC, which remains 30 or 40 years even with the most current therapies. Many other diseases are associated with impaired MCC as well, such as primary ciliary dyskinesia or COPD, a diagnosis with very high prevalence that affects an estimated 24 million people in the US alone . Current methods of studying MCC, however, are limited to the assessment of few isolated parameters each and to the evaluation of ex-vivo tissue and cells, which hinders effective research into the pathogenesis of the disease and the development of new therapies, and targeted therapeutic management based on clinical phenotype. Addressing the first limitation, our lab has recently developed a novel imaging technique, termed Micro-OCT, for improved and comprehensive characterization of MCC ex vivo. This thesis describes the integration of the Micro-OCT imaging technology into miniaturized imaging probes and the first visualization of MCC in large animal airways in vivo. This chapter outlines the background and motivation of this thesis work and summarizes: MCC and associated diseases, currently available techniques and their limitations, an introduction into Micro-OCT, and design requirements for the in-vivo probe. 21 2.1 Mucociliary clearance The mucociliary escalator is an important defense mechanisms of the lung and especially critical for clearing the smaller airways, where coughing is less effective than in the first few airway generations. It is comprised of motile hair-like structures (cilia), which transport unwanted particles and infectious agents trapped within secreted mucus so that they can be swallowed or expectorated. About 80% of the respiratory, pseudostratified columnar epithelium are ciliated cells (with the remaining 20% being goblet cells that produce mucus). Each ciliated cell has about 200 cilia to propel mucus and each cilium has a length of 5 to 7 Am in the trachea and 2 to 3 1 am in the seventh airway generation, at a diameter of 0.25 to 0.33 pm [2]. The cilia are surrounded by the so-called periciliary liquid (PCL), which is just about as thick as the length of the cilia and allows them to beat effectively. The thickness of the mucus layer can vary significantly in health and disease and is generally given as being on the order of 2 to 10 Am for healthy airways [13]. Together, the PCL and mucus layer thickness are often described and measured as the airway surface liquid depth (ASL). The ciliary stroke is comprised of two phases: the effective stroke, which propels the mucus and the recovery stroke, where the cilia return to their initial position by almost bending to their side. The frequency of one full stroke is a commonly measured parameter and named ciliary beat frequency (CBF). The CBF is on the order of 4 to 12 Hz in healthy humans. Patients with PCD have generally lower CBF, but higher frequencies have also been observed depending on the genetic mutation. Except for one mutation, where the maximum CBF value was found to be almost 25 HZ, other maxima were all below 13Hz [14]. CBF in other mammalian species, including the pig, were slightly higher and found to be in the range of 11 to 17 Hz at different locations within the respiratory tract [15]. 22 2.1.1 Spatial and temporal requirements for visualization of mucociliary function A cross-sectional imaging technology is the most advantageous for visualizing and measuring PCL and ASL as well as small changes in both disease and therapy. The technology should also provide best possible axial resolution, ideally better than 2 pm. Further, in order to resolve the complex ciliary beating pattern of effective stroke and recovery stroke, the lateral resolution should be finer than half the length of the cilium, at around 2 to 3 pm in the upper airways and as low as 1 pm in the 7th airway generation. Using Nyquist sampling as a metric for optimal data acquisition rates, in order to correctly quantify CBF, a frame rate above 34 Hz is sufficient for most cases and a frame rate above 50 Hz should capture all possible cases. Table 2.1 summarizes the design requirements of an imaging technique to visualize key parameters of mucociliary function. Table 2.1: Design requirements of an imaging technique for visualization of mucociliary function. Parameter of Interest Size/Frequency Design Requirement 5 to 7 /Lm (trachea) [2] Spatial resolution < 2.5 pim 2 to 3 ,um (7th generation) Spatial resolution < 1 pm ~ Ciliary length Axial resolution < 2.5/1 pm 2 to 10 pum Axial resolution < 1 pm = PCL + Mucus layer Axial resolution < 2 pm 4 to 12 Hz (human) Frame rate > 24 Hz 11 to 17 Hz (swine) Frame rate > 34 Hz Ciliary length PCL Mucus layer ASL CBF 23 2.2 Diseases of mucociliary clearance The classical diseases of mucociliary clearance (MCC) are primary ciliary dyskinesia (PCD) and cystic fibrosis (CF). However, MCC is also impaired in many others, such as chronic obstructive pulmonary disease (COPD), acute infections, or following exposure to hairspray and cigarette smoke [2]. Further, patients within each disease category present with a wide variety of symptoms and impairment in daily activities, which seems in contrast to the typically uniform therapeutic management. This discrepancy indicates that: * the pathogenesis of disease is insufficiently understood and patients are insufficiently phenotyped, e currently available therapies are limited and of limited efficiency, and * where therapies exist, their exact mechanism remains unclear. For example, only recently has there been evidence that different antibiotics, a cornerstone of the therapeutic management for many of these patient may influence ciliary function in different ways and is dependent on the patient's characteristics [16, 17]. An imaging tool that can visualize mucociliary function in vivo and in real time therefore has the potential to increase our understanding of pathogenesis, to allow better phenotyping of patients, to provide novel insights into the effect of therapeutic agents for accelerated development of new therapies and personalized therapeutic management. 2.2.1 Primary ciliary dyskinesia Primary ciliary dyskinesia (PCD) is a rare genetic (autosomal recessive) disease characterized by immotile or inefficiently functioning cilia. A typical patient may suffer from neonatal respiratory distress in the first days of his life and go on to having recurrent respiratory and ear infections leading to bronchiectasis and progressive decline of lung function. Males are often infertile and females subfertile with an increased risk 24 of ectopic pregnancies. Other associated conditions are situs inversus, polysplenia, congenital heart defects, and hydrocephalus. While the disease progression cannot be stopped today, an early diagnosis is the best predictor for sustained lung function and therefore crucial to the disease management [18, 19]. However, diagnosing PCD is difficult due to the wide variety of symptomatic presentations, likely resulting from a wide variety of structural and functional defects underlying the disease [20, 21]. Today, a definitive diagnosis can be made on the basis of genetic testing, which is still at an early stage and does not yet provide sufficient information on phenotype. The most important diagnostics remain electron microscopy, ciliary beat pattern and frequency analysis using high-speed video microscopy [6, 22, 19]. Slow CBF raises suspicion for PCD, but normal CBF does not preclude a diagnosis of PCD. Electron microscopy may reveal a lack of dynein arms indicating a diagnosis of PCD, but many sections need to be looked at, subtle defects may be missed, and the patient can still suffer from inefficient MCC in spite of normal ultrastructure. Further, all of these techniques are limited to excised tissue or cell samples and therapeutic effects cannot be monitored in real time. Other methods for diagnosing PCD are therefore intensely studied such as measuring the levels of nitric oxide produced by the respiratory epithelium [22]. While all methods in addition to the patients medical history are helpful in establishing a diagnosis, none elucidate the exact functional defect on a ciliary level in an individual patient. As a result, PCD is believed to be largely underdiagnosed, diagnosed too late, and as a result, the disease is undertreated [23]. In addition, there are currently no therapies specifically targeted at PCD or at the individual defect with which a patient may present [1]. Current therapeutic options are typically extrapolated from other diseases, such as CF, and are limited to antibiotics, chest physical therapy, exercise, mucolytics, immunisation, and reduction of exposure to cigarette smoke and environmental pollutants. The effects of some therapies, such as nebulised rhDNase, normal or hypertonic saline, and prophylactic antibiotics remain controversial [19]. A method to visualize ciliary microfunction non-invasively and in real-time such as 25 developed in this thesis could therefore contribute to better diagnosis, phenotyping, and therapeutic management of patients with PCD. 2.2.2 Cystic fibrosis Cystic Fibrosis (CF) is an autosomal recessive genetic disease currently affecting about 30, 000 people in the US. It is caused by mutations of an anion transporter of chloride and bicarbonate named cystic fibrosis transmembrane conductance regulator (CFTR). This ion transport defect most severely affects the respiratory and digestive system and results in thick, sticky mucus build-up in the lungs and obstruction of the pancreas. A patient with CF is often diagnosed shortly after birth or within the first few years of life. At birth, he/she may for example be unable to pass meconium and require surgical removal thereof. Later, he/she is likely to present with shortness of breath, recurrent lung infections or a failure to grow. The compromised respiratory system is typically the reason for major morbidity and early mortality in CF patients. Most US patients are now diagnosed at birth if they present with meconium ileum, or through a newborn screening including a sweat test and genetic test. None of these diagnostic criteria, however, allow precise assessment of the MCC defect at the ciliary and mucus level, which in turn could help guide a finer phenotyping of patients for improved therapeutic management. Similarly, monitoring options throughout life are limited to indirect and unspecific measures, such as lung function (spirometry) and oxygen saturation tests, which do not provide information about the mechanistic defect necessary for optimal therapeutic management - with the exception of microbiologic surveillance of respiratory secretions enabling targeted and if needed chronic antibiotic treatment. In addition to unsatisfactory phenotyping and monitoring of CF patients, current therapeutic options are insufficient. Most therapies do not treat the underlying cause of CF, such as is the case for chest physical therapy, inhaled mucolytics and/or hypertonic saline or are effective in only a minority of patients, as is the case for the recently developed CFTR potentiator named ivacaftor (brand name Kalydeco). Ivacaftor rescues some of the functionality of a defective CFTR if the CFTR is present 26 at the cell surface. This is the case for only about 5% of CF patients with very specific mutations. Thus, optimal CF diagnosis and monitoring as well as therapeutic management and the development of novel therapies is hindered by the lack of a tool that allows assessment of mucociliary clearance and any therapeutic effects on the level of ciliary micro-anatomy and function in vivo. Further, such a tool would provide new avenues for studying CF pathogenesis and novel therapeutics in animal models and humans and could be particularly valuable to assess the early pathogenesis before secondary changes due to lung infections occur. A method to visualize ciliary micro-function non-invasively and in real-time has therefore the potential to significantly increase our understanding of CF pathogenesis - in particular if it can be adapted to the study of the lower airways of young animal models and small children, and would be uniquely well suited to quantitatively assess the therapeutic effect of any new compound in a given patient or patient population for accelerated development of new therapies and personalized therapeutic management. 2.2.3 Chronic obstructive pulmonary disease Chronic obstructive pulmonary disease (COPD) is characterized by a fixed (nonepisodic and irreversible despite bronchodilator use) airflow limitation, currently affecting about 6 % of the US population. COPD usually develops as a result of cumulative exposure to direct or indirect cigarette smoke, occupational dust or chemicals, and air pollution, being additionally influenced by genetic effects. The diagnosis of COPD is currently based on a patient's medical history and lung function tests, which do not provide any insights into a specific pathogenesis and are only slightly correlated with the impact of the disease for the patient, which in turn is predictive of mortality. Further, the current literature distinguishes in general between emphysema and chronic bronchitis as the primary, not-mutually exclusive, causes for the symptoms in these patients - a distinction without predictive value and little consequence for therapeutic management. Thus, COPD is an umbrella diagnosis encompassing 27 very different patients with varying pathogenesis. A patient with COPD is typically beyond his/her 40s complaining about shortness of breath during daily activities such as climbing stairs or holding a shower head above his/her head. The patient may be hypoxemic and present with cough, productive sputum, and distinct breath sounds, or at the other extreme may have normal blood gases at rest, no cough, and only distant breath sounds with possibly an overinflated, "barrel" chest. Despite large differences in patient presentation, state-of-the-art therapeutic management is fairly uniform and includes smoking cessation, bronchodilators, inhaled or oral corticosteroids, and supplemental oxygen in severely hypoxemic patients. The mismatch between the complexity of the disease and the uniformity of therapy is suboptimal. Better phenotyping of patients and novel therapies are needed to decrease morbidity and mortality in COPD. For example, it has recently been hypothesized that COPD patients associated with chronic bronchitis may have an acquired CFTR dysfunction caused by cigarette exposure leading to reduced ASL, increased mucus expression, and reduced MCT and thus could benefit from a CFTR potentiator such as ivacaftor [24]. So far, studies to assess the efficacy of ivacaftor for COPD have only been performed in cell cultures. A method to visualize and quantify ASL and MCT non-invasively, over time and in real-time in a given patient could then provide new insights into individual disease pathogenesis, allow refined patient selection prior to administration of ivacaftor or any novel therapy targeting mucus stasis in COPD, and illustrate the functional efficacy of such therapy at the ciliary and mucus level. Such an imaging advance could therefore contribute to better phenotyping and therapeutic management of patients with COPD. 28 2.3 Current methods characterizing ciliary function Current techniques for studying mucociliary clearance are all limited to studying excised tissue or cells and most provide only a few parameters of interest each. (Figure 1-1 illustrates the main techniques schematically.) 2.3.1 Electron microscopy Electron microscopy is indispensible in the study of ciliary ultrastructure [3, 4] and remains the "gold" standard for diagnosing PCD. However, it is limited to providing static images and prone to tissue processing artifacts. It may therefore provide measures of ASL and PCL within the limits of these artifacts, but is unsuitable for the evaluation of dynamic ciliary function. 2.3.2 Fluorescence confocal microscopy Fluorescence confocal microscopy has been used previously to measure PCL and ASL in cell cultures [5]. Due to the point-by-point sampling of the technique, however, it would be technically challenging to provide real-time visualization of ciliary function within a miniaturized endoscope. Further, the fluorescent dye remains an exogenous contrast and it is unclear how its introduction affects the native fluids. Newer generations of regular confocal microscopy, such as the high-speed technique spectrally encoded confocal microscopy (SECM) [25], have recently been demonstrated within human imaging probes by our group, and could potentially provide sufficient contrast for the visualization of respiratory cilia. However, at this time, even the best compromise between probe size, depth of field, spatial and temporal resolution for SECM would be inferior to the high, isotropic resolution and cross-sectional imaging capabilities of Micro-OCT. 29 2.3.3 High-speed video microscopy High-speed video microscopy (HSVM) is the current gold standard for the evaluation of ciliary function. Both qualitative parameters, such as ciliary beating pattern and coordination can be studied, as well the quantitative ciliary beat frequency (CBF) [6, 7]. However, videomicroscopy does not allow optical depth sectioning and is therefore not suitable to study mucociliary clearance in vivo. This technique therefore remains limited to analyzing excised tissue specimen or cell cultures. 2.3.4 Radioaerosol clearance The radioaerosol clearance technique requires the research subject to inhale a radioactively labeled aerosol, the clearance of which is then observed over the following hours. It can therefore be used to investigate the efficiency of mucociliary transport (MCT) but due its global character, cannot further illuminate the mechanism itself [8]. In a similar technique, the saccharin test, a saccharin tablet is placed in the nasal cavity of the subject and the time is recorded until the subject reports its sweet taste. Subject to the same limitations as the radioaerosol clearance test, the saccharin test is in addition prone to artifacts due to the patients behavior, therefore even less suitable to studying mucociliary function and should be considered obsolete in the diagnosis of PCD [22]. 2.3.5 Micro-OCT Micro-OCT is the only currently available technique to meet all of the design requirements of an imaging technique for the evaluation of MCC as summarized in table 2.1. It has been shown to provide co-localized, simultaneous, quantitative characterization of ASL, PCL, CBF, MCT and the full ciliary stroke pattern [11] within cell cultures and excised swine trachea using a bench-top system. 30 2.3.6 Summary Commonly used techniques to study mucociliary function are incapable of measuring four key features of MCC (ASL, PCL, CBF, MCT) or ciliary beating patterns within a same test or reasonable correlation between several necessary tests. Micro-OCT is a novel technique and overcomes this limitation. However, all previous Micro-OCT applications have been limited to studying excised tissue or cell cultures. 2.4 Micro-Optical Coherence Tomography Noninvasive imaging techniques such as X-ray computed tomography (CT), magnetic resonance tomography (MRI), and ultrasound are not capable of directly visualizing ciliary motion and mucociliary microfunction due to their relatively coarse spatial resolution (100 pm - 1mm). Optical reflectance techniques can provide spatial resolution on the order of 1 - 10 pm and do not rely on transmission of the signal through the sample, but detect the backscattered light, such that their spatial resolution and its decay with increasing penetration depth in the tissue is independent of the sample size. Micro-Optical Coherence Tomography (Micro-OCT) is a novel high-resolution optical reflectance technique capable to meet the previously discussed requirements of spatial (<2pm) and temporal (>40 fps) resolutions for visualization of mucociliary function as demonstrated recently on a bench-top system [11]. The following paragraphs give some background on the technique, which is based on conventional (spectral-domain) OCT. 2.4.1 Introduction to Optical Coherence Tomography (OCT) and Micro-OCT Optical Coherence Tomography (OCT) [26] is a high-speed cross-sectional optical imaging technique that can relatively easily be incorporated into miniaturized probes and is therefore uniquely well suited for in-vivo imaging of tissue microstructure or "optical biopsy" [27] without the use of contrast agents or ionizing radiation. Today, 31 OCT is in wide-spread clinical use in opthalmology and is gaining popularity in the evaluation of coronary artery disease, gastrointestinal pathologies, and maladies of the respiratory tree [27]. OCT can be considered the optical analogue of ultrasound, as it measures the echo time delays of backscattered light at changes in refractive index within the tissue. The consequence of using light instead of sound is that OCT can provide significantly better spatial resolution (on the order of 10 [tm) and imaging speed (> 100 fps) compared to ultrasound at the cost of lower penetration depth (< 2 mm in scattering tissue). Because the frequency of light (> 200 THz for visible and near-infrared light) is beyond the limit of electronic detection, OCT employs an interferometric technique to detect the echo time delay between the various reflecting and backscattering structures within the sample [27]. OCT acquires a 1-dimensional image (A-line) along the entire imaging depth with each analysis of the interference signal and mapping of the intensity of backscattered light from the different reflecting and backscattering structures along the axis of illumination onto a grey- or colorscale to create the image. In order to obtain a two- or three-dimensional image, the OCT beam needs to be scanned in the lateral directions - for example in a raster scan pattern along perpendicular lateral dimensions or in a spiral pattern within a luminal organ. Different OCT technologies can be further specified according to employed strategies for detecting and processing the interference signal. In spectraldomain OCT as used in this thesis, a broadband light source illuminates the sample and the entire spectral interference signal is acquired simultaneously, as a function of wavelength using a spectrometer. Rescaling the spectrometer output from wavelength to wave number and then Fourier transforming the interference signal provides the depth location of the scattering events [27, 28]. In order to visualize ciliary microstructure, the spatial resolution of typical OCT systems (on the order of 10 pm) is, however, insufficient. In these systems, the transverse resolution is intentionally kept low to provide sufficient depth of focus and enable fast, cross-sectional imaging. Similarly, the axial resolution may be kept low in favor of imaging speed and/or cost and availability of detectors and light sources. 32 Other optical reflectance techniques, such as confocal microscopy, which could provide finer spatial resolution (often on the order of 2 pm for in-vivo imaging systems) are not suitable either, in particular for in-vivo imaging. Their lateral resolution and optical depth sectioning capabilities are coupled via the numerical aperture of the lens and thus, they typically require point scanning in the depth direction, which is technically challenging, comparably unstable, and time-consuming. Micro-OCT combines the advantages of OCT (speed, cross-sectional imaging, decoupling of lateral and axial resolution) with an order of magnitude higher spatial resolution than conventional OCT systems. Micro-OCT differs from most conventional OCT systems primarily by 1) using a particularly large bandwidth source for higher axial resolution, 2) using a high numerical aperture (NA) objective lens for higher lateral resolution, and 3) shaping of the imaging beam and aperture to an annular profile that provides a usable, extended depth of focus in addition to the high lateral resolution. The following paragraphs describe these differences in more detail and Table 2.2 summarizes qualitatively the different factors governing axial and transverse resolution as well as depth of focus in confocal microscopy, OCT, and Micro-OCT. 2.4.2 Axial resolution The axial resolution in OCT is primarily governed by coherence gating and is proportional to the ratio of the center wavelength A, to the bandwidth A A of the sourcedetector combination, A (equation 2.1 [29], derived for a Gaussian source spectrum). Thus, to improve axial resolution, one can image at shorter wavelengths and increase the bandwidth of the system. The current implementation of our Micro-OCT system operates within the visible to near-infrared spectrum with particularly large bandwidth, namely in the approximate wavelength ranges from 650 to 950 nm (FWHM of Gaussian spectral envelope), such that A, = 800 nm and AA = 300 nm. The theoretical diffraction-limited axial resolution in air (n=1) according to equation 2.1 [29] 33 is therefore 0.9 pm. 1 6z = 2.4.3 21n2 A 2 7r C nAA (2.1) Tranverse resolution versus depth of focus in OCT The transverse resolution in OCT is, similar to confocal microscopy, primarily governed by the numerical aperture (NA) of the imaging lens and is proportional to the ratio n. A large numerical aperture allows focusing of the beam to a smaller spot and collecting backscattered light from smaller structures (increased spatial frequency) and therefore improves resolution. Depth of focus, however, decreases simultaneously with increasing diameter of the numerical aperture and is proportional to the ratio N. Since typical OCT systems acquire the one-dimensional depth image without scanning in depth, they require the depth of focus to be at least the desired imaging depth.2 Therefore, even if a desired lateral resolution can be obtained, the resulting depth of focus may be insufficient. For example, in order to achieve a theoretical, diffraction limited 2 pm lateral resolution at A, = 800 nm using a fiber-based illumination/collection OCT probe, a NA on the order of 0.18 would be required according to equation 2.2 [31, 28]. The resulting depth of focus at this wavelength and NA, however, would only be on the order of 16 pm according to equation 2.3 [32, 28]. Although on the order of the combined height of the ciliary (7 pm) and mucus layer (10 pm), such a depth of focus would be insufficient: the epithelial cell bodies could not appear in focus at the same time, the short axial range would be a major challenge for any probe design, and there would be no margin to accommodate tissue irregularities or the presence of cardiac, respiratory, and operator motion. 1Note that in addition to the coherence effect, the use of a high numerical aperture (NA) lens can improve axial resolution further in systems with higher center wavelength and lower bandwidth [30, 28]. This effect is not considered here as it small in most commonly used OCT systems and negligible in the considered Micro-OCT system. 2 In reality, the depth of focus is often maximized to not only cover the imaging depth but also potential surface variations in the tissue and operator/scanning motion. 34 6x = 0.44 = (2.2) NA 2 A~ A 7r N A2 (2.3) Table 2.2: Comparison of governing parameters of spatial resolution and depth of focus in confocal microscopy, OCT, and Micro-OCT. Ac: Center wavelength, AA: Bandwidth, n: Refractive index, NA: Numerical aperture Confocal Axial Resolution/Depth Sectioning A2 NA A cLNA Lateral Resolution DethofFocus Dept ofNA 2.4.4 oc OC 2 Ax OCT Micro-OCT oc L cxnAA _C_ nA c -A NA 0c NA A cxNAA2 0 cxNA' 2 Annular sample beam and aperture improve both lateral resolution and depth of focus in Micro-OCT Micro-OCT uses an annular sample beam and collection aperture, also referred to as "donut beam", to provide extended depth of focus in addition to fine lateral resolution and inspite of using a relatively high NA. W. T. Welford first showed in 1960 the benefit of obstructing a significant central fraction of a circular aperture in order to increase depth of focus [33]. For the case of photography, he gives equations 2.4 with 2.5 and 2.6 with 2.7, which describe the intensity distribution on axis away from the focus and in the radial direction at the focus, respectively. Figures 2-1 and 2-2 show these intensity distributions in a graphical way, for a qualitative understanding of the effect of the fractional central obstruction c: 1) In order to increase the depth of focus significantly, a large portion (greater 0.6) of the aperture radius should be blocked. 2) In doing so, the Airy pattern 35 in the radial direction becomes narrower as well, however at the cost of higher order Airy rings with increased intensity. Thus, lateral resolution is slightly improved as the depth of focus is extended, but the image may be prone to more artifacts from the increased side lobes, and "ringing" around point objects may be observed. In the case of Micro-OCT, not only the collecting aperture is annular, but the sample beam is of annular profile and focused onto the tissue of interest. a4 I(z, 0) = A 2 R 2 { sin[Ip(1 - E2)] 2 (2.4) 1 is the intensity along the axis, where p = 7ra2z/AR 2 (2.5) and a is the diameter of the aperture, A the wavelength of light, R the focal distance of the lens, e the fractional central obstruction, and z the distance from the focus along the axis. -epsion=0.3 0.6 -0.9 - 0.9 - o 0.8 - L E - 2 0.7 U- z 0.5 0.3 -- E0.2 -- - 0.6 - 0 0 0.02 0.04 0.12 0.14 0.16 0.06 0.8 0.1 Axial Distance From Focus [mm] 0.18 0.2 Figure 2-1: Qualitative comparison of axial intensities from the focus for a circular and annular apertures of different inner diameters. Factor epsilon is the fractional radius of the obstructed center portion as described in [33]. Computed from equation 2.4 with 2.5 using: a=0.672mm, A=0.08mm, R=2.8mm 36 I(0, p) = a42J A2 R 2 V v E)2(2.6) is the radial intensity at the focus, where (2.7) v = 27rap/AR and a is the diameter of the aperture, A the wavelength of light, R the focal distance of the lens, E the fractional central obstruction, and p the radial distance from the axis at the focus. -epsilon=0 -0.3 0.9 -0.6 . -0.9 . C 0.6 4) - A-_ <0.7 -o ~0 04 0.1 01 0 1 2 4 3 5 Radial Distance From Focus [mm] 6 7 x103 Figure 2-2: Qualitative comparison of radial intensities at the focus for a circular and annular apertures of different inner diameters. Factor epsilon is the fractional radius of the obstructed center portion as described in [33]. Computed from equation 2.6 with 2.7 using: a=0.672mm, A=0.08mm, R=2.8mm 2.4.5 Design requirements for a miniaturized Micro-OCT probe to visualize MCC in vivo The main goal of this thesis was to provide visualization of MCC in vivo enabled by a miniaturized Micro-OCT probe. A swine model was chosen for the resemblance of its pulmonary anatomy to the human lower respiratory system and for its relevance with respect to the existing genetic CF pig model [34, 35, 36] for future studies of CF 37 pathophysiology. Further, the probe should be adaptable to imaging the human nasal cavity and smaller airways (down to about 7 generations) in the future. A commonpath probe design was chosen that combines the beam shaping and focusing function of the probe with integrated splitting and recombination of sample and reference arm beam to minimize dispersion. It can be seen from the overview of governing parameters of axial and lateral resolution as well as depth of focus for Micro-OCT in table 2.2 that the axial resolution requirement is mainly provided by the choice of the source-detector combination and thus need not be considered in the probe design. In addition, both the lateral resolution and depth of focus depend on the numerical aperture of the focus, which is created within the probe. Other design requirements to consider for the probe include its dimensions (diameter and length) to reach the tissue of interest, scanning speed to allow quantification of CBF, motion stability, and sufficient working distance to avoid touching the mucus, as well as a general noncontact design with respect to the region of interest, such that the natural, unaltered mucus flow can be visualized. 38 Table 2.3: Design requirements for a miniaturized Micro-OCT probe to visualize mucociliary function in living swine trachea. ET: endotracheal. CBF: ciliary beat frequency Requirement Comment Lateral Resolution < 2.5 pm half the ciliary length in trachea Depth of Focus > 200 pm covering working distance, epithelial Parameter thickness, surface variations, and perpendicular motion during scanning Diameter allowing < 4 mm side-by-side bronchoscopic guidance within ET tube Length > 40 cm to access respiratory epithelium beyond ET tube Frame rate > 34 Hz twice the CBF within swine trachea and as fast as possible to limit motion artifacts mechanical and/or software algorithms to limit/correct for motion artifacts Motion stability Working distance 20 pm from outer most sur- minimum required to observe unaltered ASL and mucus transport, the larger entire the better (within the limits of the depth of focus) > face of probe 39 40 Chapter 3 Miniaturized Micro-OCT probe for real-time visualization of mucociliary function in vivo 3.1 Motivation Our lab has recently introduced a novel imaging technology for advanced study of microscopic mucociliary anatomy and function, termed Micro-OCT [11]. Micro-OCT was demonstrated to provide several key parameters of mucociliary clearance, simultaneously, in real-time, without the use of exogenous contrast. The quantitative assessment of airway surface liquid depth (ASL), periciliary liquid depth (PCL), ciliary beat frequency (CBF), and mucociliary transport rate (MCT) was validated against standard techniques - none of which can provide all four parameters at once. Further, the high resolution of Micro-OCT was capable of visualizing the ciliary forward and recovery stroke as well as glandular secretion of mucus. Like current techniques, however, these parameters were studied in tissue cultures and excised tissue requiring an invasive procedure and precise control of experimental conditions. Further, it remains unclear how the observed ciliary behavior compares to their natural behavior in vivo and the exact effect of therapeutic interventions or 41 insults cannot be studied in their full complexity within the living host. We have therefore developed a miniaturized probe (4 mm outer diameter) for Micro-OCT imaging and demonstrate its capability to visualize real-time mucociliary function of tracheal epithelium in a living swine model. 3.2 3.2.1 Methods Common-path optical probe design with annular sample beam profile The principles behind Micro-OCT have been described in section 2.4 of this thesis and prior publications [10, 11]. The Micro-OCT probe developed in this thesis integrates both the sample and reference arm of our previous bench-top system within a common-path configuration to minimize dispersion mismatch between the sample and reference arms (Figure reffig:ProbeASchematic). Compared to most conventional OCT probes, the common-path design is particularly critical in Micro-OCT, as both the lower wavelength (A, = 800nm versus 1300 nm) and order of magnitude higher spatial resolution (1 pm versus 10 jtm) make Micro-OCT more sensitive to dispersion artifacts. Beam splitting into sample and reference arm within the probe is achieved through use of two serially assembled prisms (Figure reffig:ProbeASchematic). The first is coated with gold on its long side except for a circular opening in the center, thus reflecting the sample arm light at 90 degrees and letting the reference arm light through. Reflection of the reference arm light is achieved through a reflective surface mounted to match the sample arm length at its approximate focus. Reflection of the sample arm light is provided by the tissue structures. In addition to beam splitting, the two-prism configuration provides a change of the sample beam profile from an approximate Gaussian to an annular shape. Because the opening within the gold coating is circular on the long side of the prism and the beam is focused towards the surface at 90 degrees, the obstructed center of the beam is oval shaped. Figure 3-la illustrates the optical probe components in more detail. A single mode 42 fiber is centered within a stainless steel housing (driveshaft, 2 mm OD) with help of a ferrule (1 mm OD, BK7). The light exiting the fiber is then allowed to diverge within a glass spacer of center length 2.9 0.05 mm. The exit surface of the light from the fiber as well as the spacer surface are polished to 0.3 pm roughness and at 8 degrees to minimize Fresnel back reflections at this first interface between the driveshaft and spacer. The diverging light is then focused by a 5-mm long gradient index (GRIN) lens (Grintech, 2 mm OD, pitch 0.23) towards two right triangular prisms. The first rectangular prism face, which is oriented at 45 degrees to the beam is coated with a gold reflective surface excluding a circular opening in its center of about 250 pm diameter. The outer portion of the beam is thus reflected at 90 degrees to focus on the tissue in a side-viewing configuration and with a beam profile resembling an oval shaped donut. The center portion of the beam is allowed to continue along the axis of the probe through the uncoated, second right triangular prism. This center portion will then subsequently be reflected by a reference reflector along the axis of the probe and serve as the reference arm. Returning light within both arms is recombined and the interference signal is transported to the Micro-OCT console. All serial components of the optical probe are attached using UV-curing adhesive (NOA 65, Norland Products, NJ) with nearly perfect transmission in the visible and near-infrared wavelength ranges. 3.2.2 Probe scanning In order to create a two-dimensional image over time, the optical probe subsystem is scanned over a 400 pm distance at 40 frames per second (fps) using a rigid steel driveshaft connected to a Piezo stack actuator. 3.2.3 Outer housing design characteristics The stationary outer housing of the endoscopic probe is comprised of a cylindrical steel hypotube (4 mm OD) incorporating a thin imaging window and an end-cap to provide distance rails and a seal against body fluids. The window was realized as a cut-out 43 within the hypotube overlaid by polyolefin heat shrink. The end cap was custom designed and 3D printed and features distance rails to facilitate placing the probe's focus with respect to the tissue surface. The working distance of the probe from the center of the optics is on the order of 2.8 mm, or about 0.8 mm from the imaging window. A hand piece (custom design, 3D printed) stabilizes the outer housing as well as the piezo scanner and facilitates manipulation of the endoscopic probe. Figure 3-1b through e illustrates the assembly and appearance of the components of the outer housing. 3.2.4 Fabrication of prototype During fabrication, it is critical to assure minimal back-reflections from the surfaces between two serially assembled components. Such back-reflections would be lossy and could act as additional undesired reference reflectors. Therefore, all optical components and adhesives should be index matched. For the same reason, the fiber tip and subsequent spacer surface were angle polished at 8 degrees. Further, all serial optical components should be aligned precisely to minimize the diameter of the optical probe subsystem and to avoid clipping of the beam at any surface. Precise positioning of the gold-coated prism is important, such that the hole within the coated surface obstructs the center portion of the beam and leaves a rotationally symmetric outer beam profile. Lastly, the reference reflector should be adjusted for maximized, non-saturating reference reflector power. Figure 3-2 illustrates the assembly steps of the optical probe subsystem. First, the shuttle tube should be threaded over the driveshaft as placing it later poses a risk of damaging the assembled optics. A single mode fiber (SMF) is threaded through a side opening towards the proximal end of the probe and advanced through the inside of the drive shaft until it exits the drive shaft. The exiting end is then stripped to reveal several mms of bare fiber, coated with epoxy, and a 1-mm outer diameter (OD) ferrule is threaded over it. The epoxy is allowed to cure at room temperature over night. By pulling on the fiber, the ferrule is then carefully placed and epoxied within the drive shaft in a similar manner. The drive shaft assembly is polished at an 8 degree angle 44 and to a 0.3-micron surface quality. The spacer is polished at an 8 degree angle and to length (2.9 mm center length). The spacer and subsequently the GRIN lens (5 mm length) are then aligned to the subassembly with help of a custom made holder and attached using UV-curing Norland Optical Adhesive (NOA) 65. The custom made holder piece allows the components surfaces to contact above a cut-out to prevent attaching the assembly to the holder. The subassembly is then oriented vertically and connected to a handheld diode laser for alignment. The coated, first prism is placed on top of the subassembly on a dot of NOA and its position is optimized by centering the diffraction pattern within the reflected light spot before being UV-cured. The uncoated, second prism is then attached and the desired donut beam can be observed. The shuttle tube, which is tightly fit on a nylon bearing, is put in place and the nylon bearing is attached to the driveshaft with epoxy. A polished, polycarbonate (PC) set screw is positioned within the shuttle tube to provide optimized reference path length compared to a sample at the focus. The optical probe subsystem is now fully assembled and connected to the scanner. A stationary outer housing is placed over the optical probe subsystem for protection. 3.2.5 Experimental design of animal study All animal experiments were approved and carried out in accordance with the regulations set forth by the Massachusetts General Hospital Institutional Animal Care and Use Committee (IACUC). Three healthy female Yorkshire swine (40 to 50 kg) were intubated with an 8.5 endotracheal (ET) tube and breathing spontaneously throughout the study while anesthesia was maintained on 2% isoflurane. The ET tube had been precut to approximately 23 cm length and a 7-fr Arndt Bronchial Blocker (Cook Medical) accompanied the ET tube on its outside during intubation. The endoscopic probe was then inserted through the straight section of a Y-adapter, with a pediatric size bronchoscope (Olympus, 2.7 mm OD) inserted through the angled section to guide the probe and provide visual feedback for probe placement. Once a suitable imaging location within the trachea was identified, the bronchial blocker balloon was positioned behind the probe and inflated for several seconds during image acquisition 45 behind the probe's imaging window to stabilize the outer housing of the probe with respect to the tissue region of interest. Figure 3-3 shows an example bronchoscopic image during balloon inflation. Following the first in-vivo imaging experiment, the trachea was further freshly excised and images of beating cilia were obtained with the probe in the ex-vivo tissue. 3.2.6 Data processing Each resulting three-dimensional image (cross-sectional image over time) was motion corrected and analyzed using ImageJ [371. Computation of CBF: The image stack was re-sliced to yield a stack of en-face images, then the en-face images were visually evaluated to identify periodic patterns of ciliary beat. Once such a pattern was identified, a line profile over the pattern was computed and CBF was calculated as the number of peaks in 100 frames, corresponding to 2.5 seconds. ASL measurement: The cross-sectional images were displayed and ASL was measured manually using the line tool. MCT visualization: The transport of particles on the mucus surface was visualized using a time-lapsed color image. 13 frames corresponding to approximately 0.32 s of transport were displayed in a single two-dimensional image, where the maximum intensity in each of the 13 images was represented by a color on a continuous color scale from blue (first image) to white (last image). The motion of a bright particle on the mucus surface can therefore be seen as its color changes from blue to red to white. 3.3 3.3.1 Results Miniaturized rigid probe A miniaturized, 4 mm OD, 50 cm driveshaft length, common-path Micro-OCT probe was built where beam apodization and beam splitting was provided by serial ar46 rangement of a gold-coated and uncoated prism, where the coated surface contained a circular uncoated region to provide passage of the reference arm light. The length of the probe was chosen long enough to reach tracheal epithelium unaffected during intubation and as short as possible for convenience during fabrication and handling. Figure 3-1 shows the full probe assembly schematically. Figure 3-4 shows the resulting donut beam. The probe performance was evaluated qualitatively only and validated as providing similar quality and detail to the previously used bench-top system. 3.3.2 Lessons from the animal study Three in-vivo swine studies were performed, and only the third study provided the desired in-vivo image as shown in figure 3-6. Initially challenging was the probe placement within the trachea, which was blind in the first two studies. Thus, if no tissue was within the ranging depth of the Micro-OCT image, it could not be identified whether the probe was too close or too far from the tissue and in what way the position should be altered. The addition of bronchoscopic guidance of the probe in the third study then allowed visually guided adjustment, thus easier and faster trouble shooting, finally yielding the desired image. A second challenge was the encountered respiratory and cardiac motion. In the initial study, the tissue could not be kept consistently enough within the ranging depth of the Micro-OCT image. This issue was solved through addition of an inflatable balloon that pushed the probe towards the tissue as well as distance rails on the outer housing of the probe that assured a minimum distance between tissue and probe optics. Further, the design of the imaging window, which was a polyolefin heat shrink placed over the window cut-out of the outer housing, increased experimental time significantly. It was easily removed by normal handling during the experiment such as insertion and retraction of the probe and had to be replaced a few times. Overall, finding a good position within the trachea with visible ciliary motion remains time-consuming and challenging, such that only a few percent of the obtained images from each experiment provide visual ciliary motion and mucociliary transport. 47 3.3.3 Images of MCC in swine trachea The image quality provided by our miniaturized probe design allowed visualization of beating cilia and mucociliary transport within tracheal tissue. Figure 3-5 demonstrates the periodic ciliary beating pattern, which allows computation of CBF, observed using the probe within freshly excised tracheal swine tissue. CBF was calculated to be 8 Hz, which appears low for swine tracheal epithelium, but remains in the expected range. Figure 3-6 shows an in-vivo image of swine tracheal respiratory epithelium that clearly visualizes ASL. ASL was measured to be 18 pm, which is within the expected range. Also, a color-coded time-lapsed image is shown that is averaged over about 1/3rd of a second (13 frames at 40 fps) or approximately one to three ciliary strokes and qualitatively visualizes MCT. 3.4 Discussion We have demonstrated a first miniaturized micro-OCT endoscopic probe and visualization of mucociliary micro-anatomy and function within swine trachea in vivo. Current limitations and future developments include: e Further miniaturization: The current endoscope is 4 mm in outer diameter and thus within the range of the larger, current pediatric bronchoscopes and rhinoscopes. This size is therefore suitable for imaging applications in swine as well as potential future clinical applications. However, simultaneous navigation using a standard bronchoscope is difficult since the Micro-OCT probe is too large to be guided through the accessory channel of the scope and inserting both alongside through the endotracheal (ET) tube significantly narrows the ET tube diameter and increases resistance and work of breathing, which can be limiting in a fragile or diseased population, such as children. Blind placement of the micro-OCT probe is feasible, but undesirable even in a swine study, as it prolongs the search for a 48 suitable imaging location, increases uncertainty about the exact placement and limits probe placement to locations that do not require navigation in further airway generations. Further miniaturization of the probe is therefore merited to extend the application space of the novel micro-OCT technology and probe and a readily realizable solution could be to preserve the 4-mm OD at the very tip of the probe to provide sufficient space for the bearing and shuttle tube, but to reduce the more proximal diameter of the outer housing. However, to provide navigation of the probe within the accessory channel of a bronchoscope in the future, the design of the optical probe subsystem should also be reviewed and miniaturized further. o (Semi-) flexible probe design: Secondly, a rigid endoscope design was chosen for this first generation probe for ease of scanning the imaging beam, where the rigid driveshaft could precisely transmit a translating motion provided by a piezo scanner. Rigid endoscopes are clinically used and can be preferable in certain cases, for example because they allow one-handed navigation. However, the rigidity of the design also prevents navigation of the probe within the accessory channel of a common bronchoscope, which is required to enable precise navigation, access to smaller airways, and future translation into a clinical bronchoscopy setting. Moreover, even in this study, we project that a more flexible probe design would have facilitated stabilization of the probe within the airway for better image quality and easier extraction of the parameters of interest, because in the employed stabilization method the balloon was found to act more by pulling the tissue towards the probe than by fixing the probe position within the airway. The development of a flexible probe is thus a further goal of future research. o Stable reference reflector position: A clear practical limitation of the current probe is the placement of the reference reflector by means of a set screw. In this design, the reference reflector position was not stable enough and as it moved during the experiment, image quality is 49 significantly degraded or the image was lost. Readjusting the reference reflector was difficult, as the optical probe subsystem (OPS) was protected by the outer housing during the experiment. Readjusting therefore required extraction of the probe and disassembly and re-assembly of the outer housing from the OPS. A novel probe design should therefore provide a reference reflector position that is set once without further readjustments. " Shortening distance between sample beam exit and probe tip: Further, the distance between the exit of the imaging beam and the tip of the probe was prone to limiting imaging locations in this study. For example, when the tip was in good contact with a bent portion of the tissue, the region of interest could be outside of the working range of the probe. We therefore aim to shorten this distance in a subsequent probe design. " Permanent imaging window: The current imaging window covering the cut-out section in the outer housing was comprised of polyolefin heat shrink that was applied before the experiment and would provide a suitably thin, non-birefringent, transparent seal. However, this window design was fragile and did not withstand multiple extraction and repositioning of the probe within one experimental setting. It therefore had to be regularly re-placed in a given setting, which was inconvenient and time consuming. A more permanent imaging window is desirable for a second-generation probe. " Finer navigation control: Lastly, the hand piece should be redesigned to provide finer navigation control with one hand. For example, it should be able to be held and navigated similarly to a pen and common standard rigid bronchoscopes and rhinoscopes. 50 3.5 Conclusion In this chapter, we demonstrated a first Micro-OCT rigid endoscopic probe, which incorporates both the sample and reference arm of the previous bench-top probe within a miniaturized, 4 mm outer diameter, common-path design. Using this probe in a bronchoscopically guided swine imaging procedure, we demonstrated the first visualization of mucociliary micro-anatomy and function in vivo. Further improvements of the probe design are merited to increase the research application space and facilitate translation into the clinical setting. 51 a Fiber Ferrule Shuttle Bearing i Uncoated Prism I PC Driveshaft Spacer GRIN Lens Coated Prism Set Screw b / Endcap Outer housing C Polyolefin window Distance Rails Handpiece stotionary Piezo Scanner Electrical Connection d translating Optical Connection e Endcap Handpiece Piezo ,Scanner Figure 3-1: Schematic and pictures of common-path configuration and components of first Micro-OCT probe. a. Optical probe subsystem (OPS). The sample beam is focused on the tissue via a GRIN lens and a two-prism assembly, the first of which is coated with a thin layer of gold on its long side, with the exception of a central circular opening. A shuttle tube is placed over the optics to hold a reference reflector in place at the focus of the reference beam. b. OPS within the outer housing. The outer housing remains stationary with respect to the sample and at a pre-defined distance corresponding to the height of the distance rails on the endcap. A polyolefin heat shrink is placed over the outer housing and provides the imaging window. c. The OPS is translated for scanning of the beam via a piezo scanner. The outer housing and piezo scanner are manipulated by the user and held stationary with respect to the tissue with a handpiece. d. and e. Pictures of the fully assembled probe. 52 a Drive Shaft, 7 F-- - Shuttle Tube L--I=-- 'SMF b Ferrule cI Epoxy d 4- 4 f UV-Light e GRIN NOA Spacer h g PC Set Screw Coated Prism _HC Prism rl hEi o - - -U7 Figure 3-2: Schematic illustration of main assembly steps of Micro-OCT optical probe subassembly. a. The shuttle tube was threaded over the driveshaft before the assembly of any optics. A single mode fiber (SMF) was threaded through the lumen of the driveshaft. b. A 1-mm outer diameter (OD) ferrule was threaded over the bare fiber and held in place with epoxy. c. By pulling on the fiber, the ferrule was carefully placed and epoxied within the drive shaft. d. The drive shaft assembly was polished at an 8 degree angle. The spacer was polished at an 8 degree angle and to length (2.9mm). The spacer and GRIN lens (5mm length) were then carefully aligned to the subassembly and attached using a custom-made holder and Norland Optical Adhesive (NOA) 65. e. NOR was cured with UV-light. f. The subassembly was oriented vertically and connected to a handheld diode laser for alignment. The coated prism was placed on top of the subassembly and its position was optimized by centering the diffraction pattern within the reflected light spot. g. The uncoated prism was placed and the desired donut beam could be observed. h. A polished, polycarbonate (PC) set screw was positioned within the shuttle tube to provide optimized reference path length compared to a sample at the focus. The optical probe subsystem was then connected to the scanner and was protected by a stationary outer housing. 53 Tracheal epithelium Blocker Balloon Probe Outer Housing Figure 3-3: Picture of a bronchoscopic image during balloon inflation for stabilization of the Micro-OCT probe during image acquisition with respect to the tissue region of interest. Figure 3-4: Picture of the optical probe subsystem, which is the scanning portion of the probe, and the donut-shaped beam profile. 54 Yt x x 4-, z Gray Value Figure 3-5: Image of mucociliary clearance in freshly excised swine trachea obtained with first-generation Micro-OCT probe. The square in the cross-sectional (x-z) image indicates the position of the time-lapsed image (x-t), in which the periodic pattern of ciliary beat can be observed and measured from the line profile. CBF was calculated here as 8 Hz (single cilium). X t x x _A k z z P Figure 3-6: Image of mucociliary clearance in vivo obtained with first-generation MicroOCT probe in swine trachea. Left: One cross-sectional frame showing airway surface liquid depth (ASL), measured to be 18 pm. Right: Time-lapsed color image showing 13 frames corresponding to about 1/3rd of a second of observed mucociliary transport. The transport of particles on the mucus was visualized as the particle's color changes from blue (first frame) to white (last frame). MCT could be quantified by measuring the traveled distance over the observed time. 55 56 Chapter 4 Novel common-path Micro-OCT probe designs 4.1 Motivation In the previous chapter, we demonstrated the first generation Micro-OCT probe which enabled unprecedented images of mucociliary micro- structure and function in vivo. However, the ease of use was limited due to various factors, such as an unstable reference reflector position, long distance between the imaging beam and probe tip, suboptimal window and hand piece design, rigid driveshaft, and limited potential for further miniaturization. This chapter describes our second generation probe design addressing the above mentioned limitations for improved stability and usability and with the potential to enable the assessment of mucociliary clearance in smaller airways. The major change compared to the previous design is the shift from the two prism approach and a reference reflector held in position by an additional shuttle tube to an integrated reference reflector design. This new design eliminates the need for a second prism and additional shuttle tube for improved stability and ease of use and is a promising approach and important step towards further miniaturization of the probe. Further, a semi-flexible driveshaft is provided by switching from the stainless steel to copper driveshaft. The outer housing is improved through a more permanent poly57 carbonate imaging window and new hand piece design allowing manipulation that is similar to manipulating a pen and clinically available bronchoscopes and rhinoscopes. 4.2 4.2.1 Methods Integrated reference reflector design Two major limitations of the first generation Micro-OCT probes were an unstable reference reflector position requiring time-consuming readjustment during the experiments and potential loss of position of the probe at a desired imaging location, and a relatively long distance between the imaging beam and tip of the probe limiting potential locations that could be assessed. Further, a shuttle tube was required in addition to the drive shaft and outer housing with the sole purpose of holding a reference reflector in place, increasing the diameter of the probe. This second generation probe design is based on the idea that the reference reflection could be provided by the back-surface of a glass spacer, which is an integral part of the probe optics. This design element eliminates the need for and instability of a separate reference reflector. Compared to the first generation probe design, the second prism here is a spacer polished at 45 degree and still provides an on-axis passage of the reference reflector light, but its length (1.5 mm center length) is adjusted such that the glass-air interface at its back surface provides the reference reflection at equal optical path length compared to the sample beam. Thus, the position of the reference reflector is fixed and stable, the shuttle tube can be eliminated with the potential to decrease the probe diameter, and the distance from the sample beam to the tip of optical probe subsystem (OPS) (and thus to the probe tip itself) is minimized as the reference reflection is located to the very last surface of the OPS. Fabrication: The glass spacer was initially polished on the 45 degree angled side to a surface quality of 0.3 pm and then to length on the opposite side. Once attached to the probe optics, the flat surface of the spacer was repolished to be approximately perpendicular to the axis of the reference arm light. 58 Figure 4-1, parts a to d illustrate the new design of the OPS: The sample beam is focused on the tissue of interest via a GRIN lens and a prism-spacer assembly. The prism is coated with a thin layer of gold on its long side, with the exception of a central circular opening. The reference arm light passes through this opening and is reflected from the back-surface of the spacer, the first surface of which is polished at 45 degrees to fit the prism and the back-surface of which is polished perpendicular to the axis of the OPS. 4.2.2 Semi-flexible probe design As a first step towards a more flexible probe design, the steel driveshaft was replaced by a semi-flexible copper driveshaft. The copper driveshaft remains rigid enough to transmit linear scanning motion initiated by a piezo scanner, but has the potential for better stabilization by the bronchial blocker balloon and can be inserted into the accessory channel of a bronchoscope as long as the bronchoscope is navigated using small angles (<30 degrees). 4.2.3 Improvements to the outer housing Two major improvements were made to the outer housing: 1. The imaging window was provided by a 125 pm-thick polycarbonate sheet that was epoxied to the steel outer housing at minimum distance from the OPS. This design is permanent, withstands the forces and manipulations during an experiment and subsequent disinfection of the device for reuse. This design therefore reduces overall preparation and experimental time and increases the success rate of a given procedure. The improvement is therefore convenient for current animal studies and critical in view of potential future clinical applications. 2. An ergonomically shaped hand piece was designed and 3D-printed that allows the imaging probe to be held and navigated similarly to a pen and to clinically used rigid rhinoscopes. Navigation and placement of the probe is therefore 59 facilitated, refined, and approaches a clinical standard of care in view of future clinical testing of the Micro-OCT probe. Figure 4-1, parts c and e show the hand piece. 4.2.4 Animal study A single in-vivo swine study was performed according to the same procedures as described in the previous chapter, namely under bronchoscopic guidance of the secondgeneration Micro-OCT probe and stabilization with an intermittently inflated balloon for stabilization. 4.3 4.3.1 Results Second-generation Micro-OCT probe Figure 4-1d and e show pictures of the second-generation Micro-OCT probe with improved reference reflector stability, semi-flexible driveshaft, permanent imaging window and novel hand-piece design. Figure 4-2 shows a performed "Knife Edge Test" on an image of a glass slide suspended in air to test the lateral resolution provided by the probe. The mean gray value of pixels representing air was calculated to be 40, the mean gray value of maximum intensity (reflection from the glass slide) 142. The transition from low to high intensity characterized at the 10% to 90% points (gray values of 50 and 132, respectively) was 5.5 pm which would correspond to a 1/e 2 spot size radius of a Gaussian beam of 4.3 pm according to equation 4.1 [38]. WO = XIO(4.1) 1.28 60 4.3.2 First image of tracheal epithelium in living swine model Figure 4-3 shows an example image from a preliminary swine study. The study set up was equivalent to imaging MCC with the first generation Micro-OCT probe as described in the previous chapter. Layers of the respiratory epithelium could be visualized, but no ciliary motion could be observed. It can be seen that, in this first prototype, the reference power provided by the reflective surface was insufficient compared to the reflective power of the mucus surface. In other words, the autocorrelation signal was stronger than the desired cross-correlation signal, the signal from surface of interest saturated the detector and the features of interest could not be visualized. 4.4 Limitations and future work During fabrication of the device, a major limiting, time-consuming, step was found to be the repolishing of the reflective spacer surface after assembly to assure perpendicularity to the reference beam axis and maximize optical path distance match with respect to the sample beam. The assembly could only be held at the height of the driveshaft rather than as close as possible to the surface-to-polish, which did not allow minimizing the forces exerted on the assembled spacer. Thus, polishing had to be performed extremely carefully. Secondly, the exact length of the spacer had to be tested in a reiterative fashion - polishing off a few tens of micrometers was alternated with tests of the reference position using the Micro-OCT bench top system. A more precise method of alignment of the spacer before assembly could eliminate the need for repolishing and should be found. Further, the reference power provided by the reflective surface was insufficient in the first prototype. Potential options to increase reference reflector power could include more precise alignment/repolishing to assure precise straight back-reflection of the reference arm light and/or the addition of a reflective coating. Lastly, the usefulness of the semi-flexible driveshaft remains to be demonstrated. For the preliminary animal study, no copper outer housing was available and steel 61 was used as in the previous chapter. An important advantage of the integrated reference reflector design is its potential for significant miniaturization of the OPS. The 90 degree reflection of the sample beam, which in the current first (chapter 3) and second-generation probe (Figure 4-1) designs is assured by a gold-coated prism surface, could be achieved without any coating and solely through total internal reflection if the 45-degree spacer had a diameter corresponding to the desired central obstruction of the sample beam. As a side effect, the obstructed portion of the resulting donut beam would then also be circular instead of oval further improving image quality. Figure 4-4 illustrates this design idea (c) named third-generation Micro-OCT probe, in comparison to the two previous generation probes. 4.5 Summary We have shown a second generation common-path Micro-OCT probe with integrated reference reflector, semi-flexible driveshaft, and improvements to the outer housing. Advantages compared to the first generation are: * Stable reference position, which is additionally advantageous for further miniaturization as well as flexible probe designs * Minimized distance from sample beam to probe tip " Elimination of the shuttle tube simplifying fabrication and adjustment of the OPS and further facilitating future miniaturization * Semi-flexible copper drive shaft for better stabilization with respect to the tissue of interest * Permanent polycarbonate window within the outer housing reducing preparation and experimental time * Novel ergonomic hand piece design for finer probe navigation control 62 The major current limitations observed in the first prototype of the secondgeneration Micro-OCT probe include lower than expected lateral resolution inspite of dispersion compensation and lower reference reflector power, which was insufficient in a first imaging experiment. Future work should focus on facilitating angled placement and alignment of the spacer with respect to the prism and other options such as surface coating such that reference reflector power can be significantly increased. In the discussion, we further presented a third-generation probe design idea that is a further development from our second-generation. It therefore is prone to similar challenges to the second-generation design, but provides great potential for further miniaturization of the Micro-OCT probe. 63 Fiber Engineered Surface Bearing Uncoated Spacer Ferrule - -- -- Brass Driveshaft Spacer GRIN Lens - -- Coated Prism b Outer housing C PC window Distance Rails Handpiece stationary Piezo Scanner Electrical Connection d translating Optical Connection e Handpiece Piezo Scanner Electrical Connection Optical Connection Figure 4-1: Schematic and pictures of common-path configuration and components of second-generation Micro-OCT probe. a. Optical probe subsystem (OPS). The sample beam is focused on the tissue via a GRIN lens and a prism-spacer assembly. The prism is coated with a thin layer of gold on its long side, with the exception of a central circular opening. The reference arm light passes through this opening and is reflected from the back-surface of the spacer. b. OPS within the outer housing. The outer housing remains stationary with respect to the sample and at a pre-defined distance corresponding to the height of the distance rails on the endcap. A 125-pm thick polycarbonate (PC) sheet is epoxied on a cut-out in the outer housing and provides the imaging window. c. The OPS is translated for scanning of the beam via a piezo scanner. The outer housing and piezo scanner are manipulated by the user and held stationary with respect to the tissue with a pen-like ergonomic handpiece. d. Picture of the distal probe optics. e. Picture of the proximal probe. 64 ........... ........... --------------- ..................... x x a C b 140 90% 140 130 120 120 1.28wo 110 100 100 90 TO80 80 10%, 60 60 40 50 - - - 70 40 Figure 4-2: Schematic illustrating the knife edge test for evaluation of transverse resolution achieved with second generation Micro-OCT probe. a. Micro-OCT image of cover glass edge. b. Color-coded equivalent to image a demonstrates corresponding gray values. c. Gray values profile plot corresponding to image b illustrates conversion from the 10% and 90% knife edge response of the Micro-OCT probe into a measure of spot size radius wO. Here, a knife edge response over 5.5 pm corresponds to a spot size radius of 4.3 pm. t x Auto-correlation image Tissue surface Cross-correlation image Z Figure 4-3: First in-vivo image obtained with second-generation Micro-OCT probe. The tissue surface provided a stronger reflection than the reference reflector resulting in a stronger auto-correlation than cross-correlation image. 65 Figure 4-4: Schematic illustrating the the three generations of Micro-OCT probe designs. a. First-generation Micro-OCT probe optics based on the two-prism design with separate reference reflector. An example donut beam is shown at a significant distance from the focus to exaggerate misalignment and imperfections. b. Second-generation Micro-OCT probe optics based on a prism-spacer design of integrated reference reflector. The star indicates that the sample beam profile picture was taken before assembly of the spacer and thus illustrates the diffraction pattern caused by centrally placed cut-out within the gold coating. c. Third-generation Micro-OCT probe optics with great potential for miniaturization. Based on a prism-fiber design employing total internal reflection at the prism surface and thus eliminating the need for gold coating. The resulting donut beam here is shown after accidental misalignment of the fiber with respect to the beam center. However, the round profile of the central obstruction can be observed in comparison to the oval-shaped central obstruction of design a and b. 66 Chapter 5 Summary and Outlook Mucociliary clearance (MCC) is an absolute critical defense mechanism of our respiratory system. Its importance is underlined by the fact that patients with cystic fibrosis (CF), and thus impaired MCC, used to die in their childhood and only over the last decades with more advanced and aggressive therapy their life span has extended to an average of 30 or 40 years. In addition, the prevalence of mucociliary impairment extends far beyond the classical rare genetic diseases such as CF and primary ciliary dyskinesia (PCD). Recently, there has been an increasing understanding that patients with chronic obstructive pulmonary disease (COPD) exhibit mucociliary impairment and that MCC is altered in any lung infection and may be influenced by antibiotic use [1, 2, 16, 17]. In spite of its significance and high prevalence, there has been no tool available to study the complexity of MCC at the level of the cilia and mucus in real time in vivo. This thesis filled the gap as it demonstrated a first and unique endoscopic probe for visualization of ciliary micro-function in vivo. The probe employs a novel optical imaging technique termed Micro-OCT, which had previously been shown to be superior to commonly used techniques for studying MCC. This chapter summarizes the achieved milestones as initially illustrated in Figure 1-2 and discusses next steps and the projected impact. 67 5.1 Thesis summary We demonstrated an innovative tool for visualization of mucociliary micro-anatomy and -function in vivo based on a novel optical imaging technique termed Micro-OCT. First images of MCC in living swine trachea were shown and novel probe designs are introduced as the groundwork for future translation of the Micro-OCT probe into a clinical endoscopic prototype. A miniaturized rigid endoscopic Micro-OCT probe was shown that provides the high spatial (< 2 pm) and temporal (> 34 Hz) resolution necessary for the evaluation of ciliary micro-function. The common-path probe design minimizes dispersion artifacts and combines the functionality of both the sample and reference arm of the previously used bench-top system within a 2-mm diameter optical assembly for a total probe diameter of 4 mm, which corresponds to the size of commonly used clinical rhinoscopes and bronchoscopes. This first-generation probe enabled unprecedented visualization of mucociliary function in healthy swine trachea in vivo. The probe was guided alongside a pediatric bronchoscope for visual identification of a region of interest. Motion artifacts were minimized by an intermittently inflatable balloon stabilizing the probe with respect to the tissue. Longitudinally placed distance rails on the tip of the probe positioned the probe and focusing optics such that the tissue surface appeared in focus while keeping the mucus flow undisturbed. CBF was calculated from an ex-vivo image, ASL was measured and MCT visualized from the in-vivo image. In order to lay the groundwork for research studies in further generations of the lower airways and clinical translation, newer generation probe designs were tested with improved reference reflector stability, semi-flexible driveshaft and a greater potential for miniaturization among other improvements. The presented results open new avenues for studying MCC in health and disease, in pre-clinical and clinical research settings and have great potential to eventually revolutionize diagnosis, phenotyping, and therapeutic management for all patients 68 with respiratory disease. 5.2 Next steps The results presented in this thesis open new avenues for studying MCC. Interesting next steps include: " Study of mucociliary modulators: Novel therapies to improve mucociliary function are intensely searched and studied, and the application of recently developed CFTR potentiators is extended from CF to COPD patients [39, 40, 41]. While the therapeutic efficacy can be demonstrated in clinical studies, the exact mechanistic effect and its influence on ciliary function versus mucus properties remains incompletely understood and cannot be quantified in these studies. With the developed endoscopic Micro-OCT probe, modulators of MCC can now be studied in a complex invivo system, the exact effect on cilia and mucus can be observed from onset to weaning and can be quantified in real time. Different modulators with different mechanism of action can be compared with respect to their effect on a common parameter, such as the resulting mucociliary transport (MCT). Similarly, novel therapeutic compounds can now be tested for accelerated development and translation into the clinic. * Non-invasive evaluation of MCC in the human nasal cavity: The developed rigid probe should be translated for non-invasive assessment of MCC within the human nasal cavity. Size, materials and single-handed manipulation are close to existing rigid rhinoscopes and the respiratory epithelium of the inferior turbinate within the human nasal cavity is commonly evaluated for diagnosing PCD or studies of CF and easily accessible. Remaining challenges for the adaption of the current probe into a clinical tool include assurance of electrical safety and appropriate navigation to the region of interest. Such a tool, which provides non-invasive evaluation of ciliary beating patterns in vivo 69 could then provide a novel and more sensitive way of diagnosing patients with primary ciliary dyskinesia - a disease which is thought to be largely underdiagnosed with current methods. Further, since the respiratory epithelium is continuous from the upper to lower airways, it would be interesting to show that the pathophysiology observed in the human nasal cavity is representative of the defects that would be observed in the lower airways. If the nose could serve as a window to the lung, studies of MCC and any therapeutic effects would be non-invasive and could be easily performed on each individual patient and large patient populations with the potential to transform the diagnosis, phenotyping and treatment management of patients with respiratory disease. * Evaluation of the early CF defect in CF piglets and children: The demonstrated probe should be adapted into a fully flexible design and be further miniaturized to allow navigation within the accessory channel of a bronchoscope. Such a design would enable assessment of the smaller airways in swine and humans and could significantly enhance our understanding of CF disease pathogenesis. It could elucidate the unclear relationship between the micro-anatomy and function of the respiratory epithelium and the mucus properties in CF piglets [34, 35, 36] and children before potentially confounding secondary changes occur due to lung inflammation. Observing the natural history of disease in children could finally prove or revoke the hypothesis that CF lung disease begins in the lower airways - consequently, this may allow more targeted therapeutic management. Similarly, longitudinal studies in CF piglets could lead us to a better understanding of the initial events in the onset of disease and its underlying cause to guide the development of novel therapies. 70 5.3 Impact The presented work is only the first step in the development and translation of a completely novel technique for the assessment of ciliary function in general and mucociliary clearance in particular. We expect our Micro-OCT probe to become a cornerstone technology for studying MCC and novel therapies in the pre-clinical and clinical setting. The new capabilities provided by our endoscopic Micro-OCT probe will greatly increase our understanding of the pathogenesis of all diseases of MCC and to accelerate the development and translation of therapeutics. In PCD, the technique could become the standard of care for the diagnosis as ciliary beating patterns could be analyzed in vivo in a rapid, non-invasive way. In COPD, the technique could entirely replace spirometry for phenotyping and therapeutic management. Patients could be categorized depending on the extent of impairment of MCC or reversal thereof rather than non-specific parameters such as impairment of FEVI or other spirometry results, which includes patients with very different presentations and morbidity. Such improved phenotyping would further greatly facilitate the development of novel therapeutics as their efficacy can be more easily be established in a patient population of similar pathophysiology. Further, the technique could be adapted to evaluate other organ systems with ciliated respiratory epithelium, such as the fallopian tube, and potentially even to elucidate the often insufficiently understood function of stereo cilia in organs such as the inner ear or kidneys. In summary, a clinical Micro-OCT endoscope has the potential to provide noninvasive improved and personalized diagnosis, phenotyping and therapeutic management in CF, PCD, COPD and all respiratory disease and could provide new avenues for studying other organ systems with motile and stereo-cilia. 71 72 Bibliography [1] Michael R Knowles, Leigh Anne Daniels, Stephanie D Davis, Maimoona A Zariwala, and Margaret W Leigh. Primary Ciliary Dyskinesia. Recent Advances in Diagnostics, Genetics, and Characterization of Clinical Disease. American Journal of Respiratory and Critical Care Medicine, 188(8):913-922, October 2013. [2] E Houtmeyers, R Gosselink, G Gayan-Ramirez, and M Decramer. Regulation of mucociliary clearance in health and disease. European Respiratory Journal, 13(5):1177-1188, May 1999. [3] B A Afzelius, R Dallai, S Lanzavecchia, and P L Bellon. Flagellar structure in normal human spermatozoa and in spermatozoa that lack dynein arms. Tissue & cell, 27(3):241-247, June 1995. [4] B A Afzelius. Role of cilia in human health. Cell motility and the cytoskeleton, 32(2):95-97, 1995. [5] Hirotoshi Matsui, Barbara R Grubb, Robert Tarran, Scott H Randell, John T Gatzy, C William Davis, and Richard C Boucher. Evidence for Periciliary Liquid Layer Depletion, Not Abnormal Ion Composition, in the Pathogenesis of Cystic Fibrosis Airways Disease. Cell, 95(7):1005-1015, December 1998. [6] Jean-Frangois Papon, Laurence Bassinet, Gwena811e Cariou-Patron, Francoise Zerah-Lancner, Anne-Marie Vojtek, Sylvain Blanchon, Bruno Crestani, Serge Amselem, Andre Coste, Bruno Housset, Estelle Escudier, and Bruno Louis. Quantitative analysis of ciliary beating in primary ciliary dyskinesia: a pilot study. Orphanetjournal of rare diseases, 7(1):78, 2012. [7] Wendy A Stannard, Mark A Chilvers, Andrew R Rutman, Chris D Williams, and Chris O'Callaghan. Diagnostic testing of patients suspected of primary ciliary dyskinesia. American Journal of Respiratory and Critical Care Medicine, 181(4):307-314, February 2010. [8] June Kehlet Marthin. Pulmonary Radioaerosol Mucociliary Clearance in Diagnosis of Primary Ciliary Dyskinesiaixref rid=" AFF1" isupZ*i/sup i/xrefZ. CHEST Journal, 132(3):966, September 2007. [9] Mark J Hoegger, Maged Awadalla, Eman Namati, Omar A Itani, Anthony J Fischer, Alexander J Tucker, Ryan J Adam, Geoffrey McLennan, Eric A Hoffman, 73 David A Stoltz, and Michael J Welsh. Assessing mucociliary transport of single particles in vivo shows variable speed and preference for the ventral trachea in newborn pigs. Proceedings of the National Academy of Sciences of the United States of America, 111(6):2355-2360, February 2014. [10] Linbo Liu, Joseph A Gardecki, Seemantini K Nadkarni, Jimmy D Toussaint, Yukako Yagi, Brett E Bouma, and Guillermo J Tearney. Imaging the subcellular structure of human coronary atherosclerosis using micro-optical coherence tomography. Nature medicine, 17(8):1010-1014, August 2011. [11] Linbo Liu, Kengyeh K Chu, Grace H Houser, Bradford J Diephuis, Yao Li, Eric J Wilsterman, Suresh Shastry, Gregory Dierksen, Susan E Birket, Marina Mazur, Suzanne Byan-Parker, William E Grizzle, Eric J Sorscher, Steven M Rowe, and Guillermo J Tearney. Method for Quantitative Study of Airway Functional Microanatomy Using Micro-Optical Coherence Tomography. PLoS ONE, 8(1):e54473, January 2013. [12] Susan E Birket, Kengyeh K Chu, Linbo Liu, Grace H Houser, Bradford J Diephuis, Eric J Wilsterman, Gregory Dierksen, Marina Mazur, Suresh Shastry, Yao Li, John D Watson, Alexander T Smith, Benjamin S Schuster, Justin Hanes, William E Grizzle, Eric J Sorscher, Guillermo J Tearney, and Steven M Rowe. A functional anatomic defect of the cystic fibrosis airway. American Journal of Respiratory and Critical Care Medicine, 190(4):421-432, August 2014. [13] Bruce K Rubin. Physiology of airway mucus clearance. 47(7):761-768, July 2002. Respiratory care, [14] Johanna Raidt, Julia Wallmeier, Rim Hjeij, J6rg GroBe Onnebrink, Petra Pennekamp, Niki T Loges, Heike Olbrich, Karsten Hiffner, Gerard W Dougherty, Heymut Omran, and Claudius Werner. Ciliary beat pattern and frequency in genetic variants of primary ciliary dyskinesia. European Respiratory Journal, 44(6):1579-1588, December 2014. [15] S Joki and V Saano. Ciliary beat frequency at six levels of the respiratory tract in cow, dog, guinea-pig, pig, rabbit and rat. Clinical and experimentalpharmacology & physiology, 21(5):427-434, May 1994. [16] Roel Mallants, Mark Jorissen, and Patrick Augustijns. Beneficial effect of antibiotics on ciliary beat frequency of human nasal epithelial cells exposed to bacterial toxins. The Journal of pharmacy and pharmacology, 60(4):437-443, April 2008. [17] Takashi Kido, Kazuhiro Yatera, Kei Yamasaki, Shuya Nagata, Yasuo Choujin, Chiyo Yamaga, Kanako Hara, Hiroshi Ishimoto, Masanori Hisaoka, and Hiroshi Mukae. Two cases of primary ciliary dyskinesia with different responses to macrolide treatment. Internal medicine (Tokyo, Japan), 51(9):1093-1098, 2012. 74 [18] A Ellerman and H Bisgaard. Longitudinal study of lung function in a cohort of primary ciliary dyskinesia. European Respiratory Journal, 10(10):2376-2379, October 1997. [19] A Barbato, T Frischer, C E Kuehni, D Snijders, I Azevedo, G Baktai, L Bartoloni, E Eber, A Escribano, E Haarman, B Hesselmar, C Hogg, M Jorissen, J Lucas, K G Nielsen, C O'Callaghan, H Omran, P Pohunek, M-P F Strippoli, and A Bush. Primary ciliary dyskinesia: a consensus statement on diagnostic and treatment approaches in children. The European respiratory journal, 34(6):1264-1276, December 2009. [20] Margaret W Leigh, Christopher O'Callaghan, and Michael R Knowles. The challenges of diagnosing primary ciliary dyskinesia. Proceedings of the American Thoracic Society, 8(5):434-437, September 2011. [21] Mieke Boon, Anne Smits, Harry Cuppens, Martine Jaspers, Marijke Proesmans, Lieven J Dupont, Francois L Vermeulen, Sabine Van Daele, Anne Malfroot, Veronique Godding, Mark Jorissen, and Kris De Boeck. Primary ciliary dyskinesia: critical evaluation of clinical symptoms and diagnosis in patients with normal and abnormal ultrastructure. Orphanet Journal of Rare Diseases, 9(1):1-10, January 2014. [22] June Kehlet Marthin. Nasal nitric oxide and pulmonary radioaerosol mucociliary clearance as supplementary tools in diagnosis of primary ciliary dyskinesia. Danish medical bulletin, 57(8):B4174, August 2010. [23] C E Kuehni, T Frischer, M-P F Strippoli, E Maurer, A Bush, K G Nielsen, A Escribano, J S A Lucas, P Yiallouros, H Omran, E Eber, C O'Callaghan, D Snijders, A Barbato, and ERS Task Force on Primary Ciliary Dyskinesia in Children. Factors influencing age at diagnosis of primary ciliary dyskinesia in European children. The European respiratoryjournal, 36(6):1248-1258, December 2010. [24] Peter A Sloane, Suresh Shastry, Andrew Wilhelm, Clifford Courville, Li Ping Tang, Kyle Backer, Elina Levin, S Vamsee Raju, Yao Li, Marina Mazur, Suzanne Byan-Parker, William Grizzle, Eric J Sorscher, Mark T Dransfield, and Steven M Rowe. A Pharmacologic Approach to Acquired Cystic Fibrosis Transmembrane Conductance Regulator Dysfunction in Smoking Related Lung Disease. PLoS ONE, 7(6):e39809, June 2012. [25] Nima Tabatabaei, Dongkyun Kang, Tao Wu, Minkyu Kim, Robert W Carruth, John Leung, Jenny S Sauk, Wayne Shreffler, Qian Yuan, Aubrey Katz, Norman S Nishioka, and Guillermo J Tearney. Tethered confocal endomicroscopy capsule for diagnosis and monitoring of eosinophilic esophagitis. Biomedical optics express, 5(1):197-207, December 2013. 75 [26] D Huang, E A Swanson, C P Lin, J S Schuman, W G Stinson, W Chang, M R Hee, T Flotte, K Gregory, and C A Puliafito. Optical coherence tomography. Science (New York, N.Y.), 254(5035):1178-1181, November 1991. [27] W C Warger II, M J Gora, C I Unglert, B E Bouma, and G J Tearney. Optical Coherence Tomography. In Pathobiology of Human Disease, pages 3859-3889. Elsevier, 2014. [28] Carolin I Unglert. Quantitative Optical Coherence Tomography for Imaging Healthy and Emphysematous Pulmonary Alveoli. PhD thesis, March 2013. [29] S Yun, G Tearney, Johannes de Boer, N Iftimia, and B Bouma. High-speed optical frequency-domain imaging. Optics express, 11(22):2953-2963, November 2003. [30] A Dubois and A Boccara. Full-Field Optical Coherence Tomography. In Optical Coherence Tomography: Technology and Applications, pages 565-591. Springer, 2008. [31] W C Warger II, C A DiMarzio, and M Rajadhyaksha. Confocal Microscopy. In Handbook of Biomedical Optics, pages 517-541. 2011. [32] Lihong V Wang and Hsin-I Wu. Biomedical Optics. Principles and Imaging. 2007. [33] W T Welford. Use of annular apertures to increase focal depth. JOSA, 1960. [34] Christopher S Rogers, David A Stoltz, David K Meyerholz, Lynda S Ostedgaard, Tatiana Rokhlina, Peter J Taft, Mark P Rogan, Alejandro A Pezzulo, Philip H Karp, Omar A Itani, Amanda C Kabel, Christine L Wohlford-Lenane, Greg J Davis, Robert A Hanfland, Tony L Smith, Melissa Samuel, David Wax, Clifton N Murphy, August Rieke, Kristin Whitworth, Aliye Uc, Timothy D Starner, Kim A Brogden, Joel Shilyansky, Paul B McCray, Joseph Zabner, Randall S Prather, and Michael J Welsh. Disruption of the CFTR gene produces a model of cystic fibrosis in newborn pigs. Science (New York, N. Y.), 321(5897):1837-1841, September 2008. [35] Lynda S Ostedgaard, David K Meyerholz, Jeng-Haur Chen, Alejandro A Pezzulo, Philip H Karp, Tatiana Rokhlina, Sarah E Ernst, Robert A Hanfland, Leah R Reznikov, Paula S Ludwig, Mark P Rogan, Greg J Davis, Cassie L Dohrn, Christine Wohlford-Lenane, Peter J Taft, Michael V Rector, Emma Hornick, Boulos S Nassar, Melissa Samuel, Yuping Zhang, Sandra S Richter, Aliye Uc, Joel Shilyansky, Randall S Prather, Paul B McCray, Joseph Zabner, Michael J Welsh, and David A Stoltz. The AF508 mutation causes CFTR misprocessing and cystic fibrosis-like disease in pigs. Science translationalmedicine, 3(74):74ra24-74ra24, March 2011. 76 [36] Michael J Welsh, Christopher S Rogers, David A Stoltz, David K Meyerholz, and Randall S Prather. Development of a porcine model of cystic fibrosis. Transactions of the American Clinical and Climatological Association, 120:149-162, 2009. [37] Caroline A Schneider, Wayne S Rasband, and Kevin W Eliceiri. NIH Image to ImageJ: 25 years of image analysis. Nature methods, 9(7):671-675, July 2012. [38] Gaussian Beams and the Knife-Edge Measurement, March 2009. [39] Marcus A Mall and Dominik Hartl. CFTR: cystic fibrosis and beyond. European respiratoryjournal, 44(4):1042-1054, October 2014. The [40] Frank J Accurso, Steven M Rowe, J P Clancy, Michael P Boyle, Jordan M Dunitz, Peter R Durie, Scott D Sagel, Douglas B Hornick, Michael W Konstan, Scott H Donaldson, Richard B Moss, Joseph M Pilewski, Ronald C Rubenstein, Ahmet Z Uluer, Moira L Aitken, Steven D Freedman, Lynn M Rose, Nicole Mayer-Hamblett, Qunming Dong, Jiuhong Zha, Anne J Stone, Eric R Olson, Claudia L Ordofiez, Preston W Campbell, Melissa A Ashlock, and Bonnie W Ramsey. Effect of VX-770 in persons with cystic fibrosis and the G551D-CFTR mutation. The New England journal of medicine, 363(21):1991-2003, November 2010. [41] Bonnie W Ramsey, Jane Davies, N Gerard McElvaney, Elizabeth Tullis, Scott C Bell, Pavel Drevinek, Matthias Griese, Edward F McKone, Claire E Wainwright, Michael W Konstan, Richard Moss, Felix Ratjen, Isabelle Sermet-Gaudelus, Steven M Rowe, Qunming Dong, Sally Rodriguez, Karl Yen, Claudia Ordofiez, J Stuart Elborn, and VX08-770-102 Study Group. A CFTR potentiator in patients with cystic fibrosis and the G551D mutation. The New England journal of medicine, 365(18):1663-1672, November 2011. 77