j6 Hyperpolarized Noble Gas Magnetic Resonance Imaging of Human Lungs in 150G Magnetic Field by Adelaide Zhang Submitted to the Department of Electrical Engineering and Computer Science in partial fulfillment of the requirements for the degree of Master of Engineering in Electrical Engineering and Computer Science at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY August 2 00 BARKER ©2000 Adelaide X. Zhang. All rights reserved. The author hereby grants to M.I.T. permission to reproduce and Distribute publicly paper and electronic copies of this thesis And to grant others the right to do so. I MASSACHUSETT 8NSTITUTE OF TECHNOLOGY JUL 3 12002 LIBRARIES Department of lectrical Engineering and Computer Science August 14, 2000 A . . Certified by... -........................................ Mitchell Albert Thesis Supervisor .......................................... Frederick Bowman C ertified by.. /............ T)esis Supervisor Accepted by..... ............. .. ................ . . ....... Arthur C. Smith Chairman, Department Committee on Graduate Student Hyperpolarized Noble Gas magnetic Resonance Imaging of Human Lungs in 150G Magnetic Field by Adelaide Zhang Submitted to the Department of Electrical Engineering and Computer Science on August 14, 2000, in partial fulfillment of the requirements for the degree of Master of Engineering in Electrical Engineering and Computer Science Abstract For the past 40 years, Magnetic Resonance Imaging (MRI) has been the most common method used to obtain volume images of human organs. This technique involves the detection of proton nuclear spin in cells at 1.5 Tesla (T) magnetic field strength. However, conventional MRI is very costly and some organs in the human body, such as the lungs, do not contain sufficiently high concentrations of protons to produce high quality images. In 1994, Hyperpolarized Noble-Gas Magnetic Resonance Imaging (HPMRI) was introduced and is viewed today as the up and coming technology that will solve this image scanning dilemma [1]. Its special features include a signal to noise ratio (SNR) is independent of the static field strength and Noble gas hyperpolarization by optical pumping which is 100,000 times larger than that is obtainable at thermal equilibrium. Imaging has been done using 'He and 129 Xe at 1.5T magnetic field, but no images have been produced at a low magnetic field, even though the magnetization of HP-MRI is independent of the static field. To prove that HP-MRI theory can be used in practice and that it has potential as a lower cost technology, human sized coils were constructed for lung imaging using hyperpolarized noble gases at 150 Gauss (15 mT) magnetic field. In order to operate the HP-MRI, a superconducting magnet at 1.5 T was ramped down to 150 Gauss using a new set of broad-band electronics that can be used at different Larmor frequencies. Comparison tests between different human sized coils were done to optimize their function. Images of rat lung were produced in vivo using 3He and 129 Xe and human lung images were produced using 3He at 150 Gauss to suggest that the theoretical hypothesis is correct, and to opens possibilities for clinical HP-MRI in Very Low-field. Thesis supervisor: Mitchell Albert Title: Project Supervisor Thesis Supervisor: Frederick Bowman Title: MIT Thesis Supervisor 2 Acknowledgments I would like to thank my thesis supervisor at Brigham and Women's Hospital, Dr. Mitch Albert, who was always there to advise me and to provide supports. Thanks to Arvind Venkatesh, a Ph.D. student, who was so willing to help me to solve any technical problems and who was always performed the experiments with me. Also thanks to my lab-mates: Angela Tooker, Lyubov Kubatina, and Joey Mansour on helping the human experiments and coil building. We made a great team! My special thanks to Ralph Hashoian who taught me so much about coil technology and how to think like an engineer. I would also like to thank my MIT thesis supervisor, Dr. Frederick Bowman, who was always there to help me with paper work, and who made sure that I turned in everything on time. My appreciation to my beautiful, and tall roommate, Hua-yin Yu, who helped to polish up my thesis and gave me support through out my years at MIT. My cute, and adorable friend, Moksha Ranasinghe, who stayed up with me to do work for countless nights and offered me encouragement when I needed it. Finally, to my cool friend, Tina Jan, who always gave me great and logical advises. 3 Contents 1 Introduction......................................................................................9 2 Theory and Background of HP-MRI.....................................................11 2.1 Magnetic Properties of Atomic......................................................11 2.2 MR Imaging Theory................................................................16 2.3 Hyperpolarized Noble Gas MRI..................................................18 2.3.1 Optical Pumping.........................................................19 2.3.2 Magnetization Generated by Hyperpolarization....................20 3 Setup of HP-MRI at the Very Low-field..................................................22 3.1 Superconducting Magnet.............................................................24 3.2 Gradient C oils..........................................................................26 3.2.1 Z Gradient Coils.........................................................26 3.2.2 X and Y Gradient Coils..................................................27 3.3 RF Transmit and Receive Coils..................................................28 3.4 RF Pre-Amplifier and Transmit and Receive (T/R) Switching System........32 3.5 Hyperpolarizing System for Noble Gases.........................................34 3.5.1 Xenon Gas Hyperpolarizing System...................................35 3.5.2 Helium Gas Hyperpolarizing System....................................36 3.6 Animal Gas Delivery System.........................................................37 4 Coil Development in the Very Low-field....................................................39 4.1 Basic Coil Circuit Components Calculation.......................................41 4.2 Method of Coil Building...........................................................44 5 Results and Discussion...........................................................................47 5.1 Coil Analysis in Very Low-field..................................................47 5.1.1 Coil Optimization with Number of Turns............................47 5.1.2 Optimizing Coil Shape for Lung Imaging..............................52 5.1.3 Conclusion on Optimizing Lung Coils..............................53 4 5.2 Phantom Images in Very Low-field................................................55 5.3 Very Low-field Lung Images of Rat in vivo.......................................56 5.4 Very Low-field Human Lung HP-MRI.............................................57 5.4.1 Background Noise..........................................................57 5.4.2 Human Lung Image........................................................62 5.5 Summ ary ................................................................................ 63 5.6 Recommendations to Future Work...............................................63 Appendix.............................................................................................65 A ppendix A ............................................................................... 65 Appendix B ............................................................................... 66 References.........................................................................................67 5 List of Figures and Tables Figure 2.1: Direction of the magnetic field generated by the charged particle depends on the energy state level Figure 2.2: Magnetic dipole moment generated by the unpaired protons. Figure 2.3: Net magnetization can be generated when protons are placed in an external magnetic field Figure 2.4: In the absent of Bo-field, proton rotates along its axis, but when it is placed in an external Bo-field, it starts to precess along the external field axis. Figure 2.5: RF wave transmits into the body and exert a torque on the nuclear spin to generate NMR signal, which later receives by the receive coil. Figure 2.6: MRI signal generated in a homogeneous magnetic field, also called FID. Figure 3.1: Overview of HP-MRI setup at Very Low-field Figure 3.2: Cross section of a MRI superconducting magnet Figure 3.3: Z Gradient Coil configuration Figure 3.4: X Gradient Coil Configuration Figure 3.5: Y Gradient Coil Configuration Figure 3.6: The setup of transmit and receive coil which was used in all Very Low-Field experiments Figure 3.7: This figures shows the imaging area inside of the solenoid coil has homogeneous magnetic field, but the field strength for surface coil is nonuniform Figure 3.8: Overall diagram for the transmitting and receiving switch system Figure 3.9: The quarter-wave circuit also called the pi circuit. The values of inductance and capacitance depend on the Larmor frequency of the element. Figure 3.10: Block diagram of the Hyperpolarizing System 6 Figure 3.11: 12 9 Xe Hyperpolarization chamber and trapping glassware used to purify hyperpolarized xenon from helium and rubidium and to store the polarized gas for long periods of time. Figure 3.12: Animal Gas Delivery System Figure 4.1: Helmholtz Coil Figure 4.2: Circuit Diagram of the coil network Figure 4.3: Simple Circuit model of coil Figure 5.1: Human size proton coil inductance vs. number of turns Figure 5.2: Normalized R and Q vs. surface coils Figure 5.3: FID of Water Phantom using Human Size proton Surface Coils with different turns Figure 5.4: Possible lung coil designs in Very Low-field Figure 5.5: FIDs of Helmholtz, planar, and surface coils Figure 5.6: Spin echo water phantom image with Helmholtz Coil Dimension of the coil: 10.5"in diameter, 10" in height. Dimension of sample: 6" in diameter Figure 5.7: Gradient echo helium phantom image. The bright spot in the center is from the cell's Rubidium pull off finger. Dimension of Coil: 10.5" in diameter. Dimension of Sample: 30 mm in diameter Figure 5.8: 12 9Xe lung image of rat in vivo Figure 5.9: 3 He lung image of rat in vivo Figure 5.10: (a) Basic noise data with noise peaks (b) Hardware setup for the basic noise test Figure 5.11: (a) Basic noise data after cross diodes reduced noise peaks (b) Hardware setup with a pair of cross diodes Figure 5.12: (a) Basic noise data without connecting all active electronics to the same ground point. (b) Basic noise data after connecting all active electroinics to the same ground point. Figure 5.13: Circuit diagram of low-pass filter. It is used to eliminate noise that was generated by the gradient amplifier. Figure 5.14: (a) Basic noise data before using gradient amplifier noise filter. 7 (b) Basic noise data after using gradient amplifier noise filter. Figure 5.15: HP-MRI human lung image Table 5.1: Comparison of surface coils with different turns 8 Chapter 1 Introduction In today's medical field, Magnetic Resonance Imaging (MRI) has become the primarily technique used to obtain high quality volume images of the inside of the human body. MRI is based on the principle of Nuclear Magnetic Resonance (NMR), which is the interaction between radio waves and atomic nuclei. The most commonly used atom is hydrogen, which is ubiquitous throughout the human body, especially in the form of water and fat. However, water and fat are not very concentrated in some organs, which makes imaging difficult. In order to overcome this problem, fast imaging methods and contrast agents have been developed. Yet these techniques are unable to properly image the lung, a gas-filled, water-free space. The idea of using other atomic nuclei as spin probes for biological MRI studies has been proposed, but the concept is limited by the insufficient concentrations that can be achieved in living organisms and the consequent high signalto-noise ratio (SNR). Noble gases such as xenon and helium also have the same problem, but the nuclei of these gases can be hyperpolarized by optical pumping to achieve high magnetization to overcome the SNR problem. 9 In obtaining lung images, patients are asked to inhale hyperpolarized (HP) helium or xenon in order to enhance the NMR signal over one million times [4]. Since the NMR signals are much higher than before, the static magnetic field can be decreased to reduce the cost of the MRI scanner. The conventional 1.5 Tesla MRI scanner uses a high cost super-conducting magnet. If HP-MRI is used, the magnetic field can be dropped from 1.5 T to 150 Gauss, which requires only a low cost permanent magnet for MR imaging. The elimination of expensive superconducting magnets can lead to the development of vanbased imagers that could take this technology to patients whom are unable to travel to a hospital. Another advantage of using permanent magnets is that MRI technology can be used in the micro-gravity environments of spacecrafts and space stations to monitor the physical states of astronauts [5]. The new Hyperpolarized Noble Gas Magnetic Resonance Imaging (HP-MRI) technology, which utilizes a very low magnetic field, is currently being developed in the laboratory of Mitchell Albert at the Brigham and Women's Hospital in Boston, Massachusetts. In collaboration with Arvind Venkatesh, the project's goal is to build transmitting and receiving coils for human lung HP-MR imaging in the 150 Gauss magnetic field. 10 Chapter 2 Theory and Background of HP-MRI This chapter reviews the magnetic properties of the atoms, basic magnetic resonance imaging (MRI) theory, and the theories behind the hyperpolarized noble gas magnetic resonance imaging process. 2.1 Magnetic Properties of Atomic Nuclei Nuclear magnetic resonance (NMR), a technique that has been used for over 50 years to analyze chemicals, is the foundation on which MRI technology has been built. The frequency of Electromagnetic waves and the energy received in MRI are much lower than those of X-rays and of visible light. The energy of EM waves is directly proportional to it frequency: E = h*v. Due to these properties of MRI, a type of electromagnetic pulse called a radio frequency (RF) pulse is used to produce a signal. NMR theory is based on Felix Bloch's discovery that spinning charged particles create an electromagnetic field and that these particles have different energy levels. In I1 MRI, hydrogen nuclei are the most common source of protons [6]. This is because hydrogen is found in fat, which is in high concentrations in certain parts of the body, and in water, which comprises approximately 70% of human body masses. Following the Boltzmann statistics, charged particles have two energy states, -1/2 and +1/2, and spin about their axis to create a magnetic field (Figure 2.1). At room temperature, the number of spins in the lower energy level, N+, slightly outnumbers the number in the upper level, N-. This distribution can be represented by equation: AE -= e U (Eqn. 2.1). Where, E is the energy difference between the spin states, k is Boltzmann's constant, 1.3805x10-23 J/Kelvin, and T is the temperature in Kelvin. Because of the different energy states, protons spin in opposite directions (north and south) from one another and the resulting magnetic field is also generated in these two directions. Direction of magnetic field Direction of spin Direction of magnetic field Figure 2.1: Direction of the magnetic field generated by the charged particle depends on the energy state level. 12 If there was an even number of protons, then every proton that spins pointing north would be paired with another proton, which spins pointing south. The net magnetic field created by this pair of protons is zero. If there is an odd number of protons, then the unpaired proton causes a net magnetic field, known as a magnetic dipole moment (MDM), which is in the same direction as the proton's spin. (Figure 2.2) The occurrence of MDMs in some elements has lead scientists to utilize this property for imaging purposes. A problem arises in the fact that the magnetic fields of many individual protons cancel each other out, which results in a net field of zero. When an external magnetic field (BO) is applied to the protons, they line up along the BO, with approximately half pointing north and other half pointing south. (No magnetic field) Paired protons (Net magnetic field) Unpaired protons Figure 2.2: Magnetic dipole moment generated by the unpaired protons. 13 However, about one in every million protons produces an extra northward spin, which can accumulate to generate a net magnetization (M) in the direction of BO. (Figure 2.3) The MDM grows exponentially over time, where the time constant of the curve is called the relaxation time, TI. M does not only depend on TI, but is also related to the mobile proton density or the spin density, N(H). Therefore, t M =N(H)*(1-e TI) (Eqn 2.2). :Net MF= 0 Bo off: BoBon: Figure 2.3: Net magnetization can be generated when protons are placed in an external magnetic field. When a proton spinning on its own axis is placed in a larger magnetic field, it begins to wobble or move along the axis of BO as a result of gravity. (Figure 2.4) The magnetic field exerts a force on the moving charge particle [7] can be described by equation: P = qv x BO (Eqn 2.3), 14 Where v is the velocity of the charge, and P is the force. From the force equation, three equations of motion can be derived, m( m( x dy d X )= qBO (-) -F(-) r dt dt dx_ dt Y) =-qBO(x) -F( 2 dt (Eqn 2.4) (Eqn 2.5) Y) r (Eqn 2.6). m( d)= -F() r dt - With the present of Bo, the equations of motion can be solved as: 2 2 M(d 2a =m( dd x )+im( d y2 ) m(2) dt dt2 dt- F da -()a dt r -iqBo (-) (Eqn 2.7), where a=x+iy, and i=V-1. Furthermore equation 2.6 can be derived into: F mk 2 +iqBok +-=0 r (Eqn 2.8). And the force exerted by BO is small compare to the central force, therefore, -i( q )BO (Eqn 2.9). i 2m rm Thus a = ecot (Q * el'st + R * e-1'wst) (Eqn 2.10), where Q and R are constants. From this equation, the rate which proton precesses around the BO is given by the Larmor Equation: 15 o=y*Bo (Eqn 2.11). where o is the angular precessional frequency of proton, normally expressed in units of Hertz (Hz), and y is the gyromagnetic ratio, expressed in MHz/T. For hydrogen protons, y(H)= 42.6 MHz/T. Bo OFF Bo ON Figure 2.4: In the absent of Bo-field, proton rotates along its axis, but when it is placed in an external Bo-field, it starts to precess along the external field axis. 2.2 MR Imaging Theory RF (radio frequency) waves and the resonant frequency of protons, also know as the Larmor frequency, is delivered from a RF coil to the human body to excite the hydrogen nuclei. (Figure 2.5) Coils are electrical devices generally composed of multiple loops of wires that either generate magnetic fields or detect the changes in magnetic fields. Typically from a stationary point of view, a proton spins around the magnetic field axis at the Larmor frequency. However, if the spinning system is viewed from the 16 frequency equal or close to the Larmor perspective of a coordinate system rotating at a becomes stationary [8]. In such a frequency, the perspective of the rotating proton is easier to follow during and after the coordinate system, the macroscopic magnetization excitation of a RF wave. RF Tran1smit Signal generated Receive exert Figure 2.5: RF wave transmits into the body and NMR a torque on the nuclear spin to generate signal, which later receives by the receive coil. with Larmor frequency are Commonly, high intensity and short-term RF waves, body. RF pulses exert a torque on pulsed into the coil to excite the protons in the human away from the direction of the external the magnetization (M), which causes the M to tip to precess around BO with the Larmor field, BO. However, the magnetization is forced an oscillating magnetic field, part frequency. Since this rotating magnetization represents is picked up by a receiving coil also of the energy that is associated with this oscillation is received from a homogeneous can be considered as an RF antenna. The signal that induction decay (FID), which looks like sample in a homogeneous field is called the free 17 a damped oscillation of a sine wave. (Figure 2.6) When a non-uniform external field is applied to the sample, a MR image can be obtained. I>L~ If Received signal Time Figure 2.6: MRI signal generated in a homogeneous magnetic field, also called FID. 2.3 Hyperpolarized Noble Gas MRI Conventional MRI is a water-proton-based technique that can effectively identify abnormalities in most soft tissues. However, it is not as effective in the lungs and other lipid bilayer membranes regions where proton based cells do not dominate the area. Albert and his group proposed the first solution to this problem: instead of using conventional MRI methods, hyper-polarized noble gases MRI (HP-MRI) can be used to solve the dilemma. 18 2.3.1 Optical Pumping HP-MRI requires that a patient inhale hyperpolarized xenon (129Xe) or helium (3He) gas to fill his/her lungs. The image of the lung region is then created from the signal that is obtained from the resonating nuclei of the isotope 12 9Xe or 3He, rather than that of protons. Normal thermal-equilibrium polarization of noble gases is only 0.04% of the proton in the human body, which is insufficient for imaging. However, when a laser optical pumping process is used to polarize the noble gases, the magnetization is enhanced by a factor of 105 [1]. This is achieved when hyperpolarized I2 9Xe or 3 He nuclei are produced by spin exchange with optically pumped rubidium (Rb), alkali-metal, atoms. Rb is an alkali metal which vaporizes at 85C. In an external magnetic field, Rb splits into different electron spin states under the excitation of a titanium sapphire laser. Depending on the direction of the circularly polarized laser beam, Rb moves into either a plus or minus state with respect to ground [10]. Ground state electrons can be excited out of these states by absorbing photons of 794.7nm wavelength. The polarization (P) of alkali-metal with nuclear spin I=0 can be represented by: P=NN1 -N 22 1 1 and N 1 +NiN_ =1 2'2 22 (Eqn 2.12). 2 2 The optical pumping exchange rate between two states can be described by equation: dN(± })L2)_=) [ FsDy 2 dt ]N(-+) T FSDN(+-L) (Eqn. 2.13). 19 Where y, is total rate per atom of pumping out of one state into another, and FsD is the possible relaxation of electron spin polarization [12]. If polarized gaseous Rb is brought together with 12 9Xe or 3 He, the collisions between Rb and 129Xe or 3He atoms result in the transfer of angular momentum from Rb valence electrons to the 129Xe or 3He nuclei. The steady-state of the polarization after the spin exchange can be expressed as: P = N N N (Eqn. 2.14) N is the total number of spins. The spin exchange process increases the noble gas spin population by twenty five percent over the equilibrium state, and therefore enhances the NMR signal up to 105 times the thermal equilibrium value. If the polarized gas is kept in a chilled magnetized charm, the polarization can last for hours, even days. 2.3.2 Magnetization Generated by Hyperpolarization The magnetization of hyperpolarized gas can be derived from the general magnetization equation: M -N y (Eqn. 2.15). 4kT Now, substitute the polarization equation into the equation 2.14 to form the magnetization of hyperpolarized gas [13], M Hyperpolarized - y 2 (Eqn. 2.16) 20 The substitution of P is derived the Boltzmann equation, N- e( k 1+ - AE N+ (Eqn. 2.17). kT Let's manipulate the equation 2.16 by subtracting 1 from both sides, then the new equality becomes: N_-N N+ -AE kT (Eqn. 2.18), then dividing both sides by N and multiplying N+ to produce: N_ -N N -AE N kT N (Eqn. 2.19), Substitute h o> for AE, and N for N+, by assuming the spins in one state comprises approximately half of the total spin population, N_-N, N -- 2 kT N (Eqn. 2.20) Final, replacing the left side of equation by P, we obtain the equality of P= h 2kT (Eqn. 2.21). Equation 2.16 shows that magnetization of hyperpolarized gas is independent of the external magnetic field strength, which means Very Low-field MR image is theoretically possible. In the later chapter, this concept was also proven experimentally. 21 Chapter 3 Setup of HP-MRI at the Very Low-field The schematic representation of the Very Low-field system with which hyperpolarized 129Xe and 3He images were acquired is shown in figure 3.1. The prototype Very Low-field MRI imager was a 60 cm diameter superconducting magnet with a magnetic field ramped from 1.5 T down to 15 mT (150 Gauss). Within the magnet were Techron gradient coils for producing a gradient in the BO in the X, Y, and Z directions with maximum strength of 0.8 Gauss/cm. Inside the magnet core, a shield made from a 0.1 mm thick copper sheet was installed to reduce electrostatic noise. In order to acquire MRI signals, a self-designed RF coil was placed inside the scanner to transmit RF pulses and to receive NMR signals. Depending on the type of experiment, different RF coils were used. The scan room was surrounded by an RF shield, which was made from copper sheets. The shield prevented the surrounding RF pulses to enter the scan room. It especially blocked various RF signals from television and radio stations which can by detected by the imager. 22 Figure 3.1: Overview of HP-MRI setup at Very Low-field The magnet, gradient coils, and RF coil were interfaced with a Resonance Instruments Maran Ultra imaging spectrometer, which was operated from a Gateway computer that controlled all components on the imager. The spectrometer signaled the pulse software to send sine wave pulses at the 129 Xe and 3 He frequencies. The software program caused the shaping of the RF pulses into apodized sinc pulses. After the RF source sent out the pulse, the amplifier augmented the pulses' power from milli-Watts to kilo-Watts before it reached the RF coil. The computer also controlled the gradient pulse program, which set the shape and amplitude of each of the three gradient fields. A 23 gradient amplifier increased the power of these pulses to a level that was sufficient to drive the gradient coils. Commands to the MRI system were given to the computer through a control console. Imaging sequences were selected and customized from the console according to each experiment. The image and free induction decay (FID) signals were displayed on the console, and further evaluations and analyses were done using the MATLAB software program. Besides the normal setup that is required for all MRI systems, HP-MRI also requires a different gas hyperpolarizing system for the 12 9Xe and 3 He experiments. For animal experiments, Sprague Dawley white rats are used. A gas delivery system is set up to help the rats inhale the hyperpolarized gas into their lungs. For human experiments, gas is delivered in a Tedlar bag to the subject. In the following sections of this chapter, the function of individual hardware, hyperpolarizaing system, and the gas delivery system of the Very Low-field MRI will be described in greater details. 3.1 Superconducting Magnet The magnet is the component of the magnetic resonance imaging system which created the static field for the imaging. Depicts the actual magnet used in these experiments, most conventional clinical magnets are 1.5 Tesla superconducting magnets. However, the one we used was ramped down to 150 Gauss. A superconducting magnet 24 is an electromagnet made of superconducting wire. This wire has a resistance approximately equal to zero when it is immersed in liquid helium that is cooled to a temperature close to absolute zero (-273.15 C or 0 K). Once a current induced in the coil, it will continue to flow as long as the coil is kept at liquid helium temperatures. Some losses might occur over time due to very small resistance that might be exhibited by the coil. These losses are on the order of a ppm of the main magnetic field per year. Figure 3.2 shows a cross sectional view of a superconducting magnet. The length of superconducting wire in the magnet is typically several miles long. The wire coil is kept at a temperature of 4.2 K by immersion in liquid helium in a large dewar. The dewar is surrounded by a liquid nitrogen dewar at a temperature of 77.4 K which acts as a thermal buffer between the exterior temperature (298K) and the liquid helium temperature. Vacuum Liquid Helium -E Liquid Nitrogen MEE Container &Support ME Superconducting Coil Figure 3.2: Cross section of a MRI superconducting magnet 25 The helium and liquid nitrogen used to cool the superconducting magnet and to keep it running are costly. However, with only a 150 Gauss magnetic field strength, the expensive superconducting magnet can be replaced with the relatively cheap permanent magnet. This is one of the greatest advantages of Very Low-field MRI. 3.2 Gradient Coils The gradient coils produced the linear gradient change in the homogeneous static magnetic field, BO, which decoded spatial information from the signal and localized it in a given space. The variation in the gradient magnetic field is several orders of magnitude smaller than that of the static magnetic field, but it is significant enough to allow spatial encoding. Unlike imaging magnets, the gradient was kept at room temperature and the configuration of the coils created the desired gradient in the X, Y, and Z directions. 3.2.1 Z Gradient Coils Even though the imaging magnet was ramped down to 150 Gauss, the Very Lowfield system was still standardized like other MRI systems. A gradient in BO in the Z direction is achieved with an Anti-Helmholtz type of coil (Figure 3.3). Current in the two coils flow in opposite directions, creating a magnetic field gradient between the two coils. The B field at one coil adds to the BO field while the B field at the center of the other coil subtracts from the BO field. 26 Z Gradient Coil Gz B B X zY Figure 3.3: Z Gradient Coil configuration 3.2.2 X and Y Gradient Coils The X and Y gradients in the Bo field are created by a pair of eight-shaped coils. In the X axis, the direction of the current through the eight-shaped coils create a gradient in BO in the X direction (Figure 3.4). The Y axis figure-eight coils generate a similar gradient in Bo along the Y direction as the X-gradient coils (Figure3.5). X Gradient Coil Gx x Y B Figure 3.4 X Gradient Coil Configuration 27 Y Gradient Coil X Figure 3.5 Y Gradient Coil Configuration 3.3 RF Transmit and Receive Coils RF coils used on humans to obtain MRI signals are also called transmit/receive coils. The coils can be divided into three general categories: receive coils, transmit coils, and both transmit and receive coils. A transmit coil is used to send RF pulses to interact with protons inside of the human body. A receive coil is used as an antenna to pick up the signal that is sent out from the body. For all the experiments contacted in the HypX lab, the same coil acts as both transmitter and receiver to reduce the interference in the magnetic field. (Figure 3.6) The RF source transmits pulsed waves to the RF coils, which creates the magnetic field B, that is perpendicular to the static BO field. The coils also detect the transverse magnetization as it precesses in the XY plane. 28 T/R Switch ftM~fl:: Transmission ~~~Line Receiving Line Figure 3.6: The setup of transmit and receive coil which was used in all Very Low-Field experiments A transmit only coil is used to create the B, field and a separate receive only coil is used in conjunction with it to detect or receive the signal from the spins in the imaged object. However, the separate transmit and receive coils have to be placed inside of the BO field together, which means the two coils will interfere with each other and alter the behavior of the B, therefore distorting the signal from nuclear spin. This is the main reasons why a single coil is used in experiment to transmit and receive RF waves. The ones used in the experiments described are the transmit and receive coils. These types of coils serve as the transmitter of the B1 field and receiver of the RF energy from the imaged object. All RF coils must resonate, and efficiently store energy at the Larmor frequency in order to obtain the small NMR signal that is produced by the spin in the atoms. Therefore the coils are resonated with capacitor circuits. Conventional MRI also uses inductors in conjunction with the capacitors in the circuit. However, in the 150 Gauss field, the Larmor frequencies of proton, helium, and xenon are all below 1 MHz which is much lower than the Larmor frequencies at 1.5 Tesla. As we know the resonant frequency, o, is related to inductors (L) and capacitor (C) circuit by 29 LC (Eqn 3.1). For RF coil building, C is the total capacitance in the circuit. In conventional MRI, L is the total inductance for the coil and the inductor, whereas in the Very Low-field, L only represents the coil inductance. Inductors are not recommended for use in the circuit for 150 G magnet field. This is because L is inversely proportional to o, which is very small, and therefore for the same capacitor values, L is relatively large for the Very Low-field. Adding more inductors to the circuit would only increase the inductance and the resistance in the coil, which leads to a decrease in the coil's energy storing efficiency. Further descriptions of coil building and development will be described in chapter four. One of the requirements of an RF coil is that the produced B, field must be perpendicular to the Bo magnetic field. Therefore, solenoid coils were used for the animal experiments, and surface coils were used for the human experiments. The magnetic field created by the solenoid coil is given by the equation, B =#JOiN (Eqn 3.2). The magnetic field created by the surface coil is given by the equation, B - 0 i 0N 21r (Eqn 3.3). where go is the permeablity constant, i is the current flowing through the wire, N is the number of loops of wire that constitute the coil, and r is the distance from the coil. By comparing these two equations, it is clear that the magnetic field in the solenoid is homogenous, and that the field in a surface coil decreases with the distance, r (Figure 30 must be placed 3.7). In lung imaging, solenoid coils are preferred, but the solenoid field. Therefore a horizontally in the external magnet in order to create the perpendicular the external magnet human cannot fit in the coil, because a human cannot fit in horizontally. B0 B0 A T/R Switch Solenoid Coil Figure 3.7: This figures shows the imaging area inside of the solenoid coil has homogeneous magnetic field, but the field strength for surface coil is non-uniform. useful only to certain Each coil generates its own magnetic field pattern which is in the Very Low-field lungs part of the body. Only solenoid and surface coils are used of coils, such as birdcage imaging, but convention field generation utilizes other types coil of choice for imaging the coils and single turn solenoid coils. The birdcage coil is the field around head. The head and the brain, because the coil can create a homogeneous turn solenoid is similar to the shape of the magnetic field that is generated by the single for the detection of shape of the breasts, therefore this type of coil is often used abnormality in the breasts. 31 The 150 Gauss coil building technique is based on the equations and formulas that are used in 1.5T, but alterations have been made. Development of coils in low field will be discussed in detail in Chapter Four. 3.4 RF Pre-Amplifier and Transmit and Receive (TR) Switching System The Resonance Broadband RF Pre-amplifier and TR switch system generate frequencies from 0.1MHz to 30MHz. Pulse sequence controls regulate the amplitude and duration of the pulse (Figure 3.8). The pulse is then transmitted into a circuit that switches between the transmit and receive states. Receiver COIL ProbeL C C2 O Transmitter G Figure 3.8: Overall diagram for the transmitting and receiving switch system 32 The RF wave is sent down to the transmission port in the pre-amp system and then through a pair of cross diodes to the RF coil. The cross diodes are used to reduce the noise from the transmission line when the MRI spin signal is received from the coil. A quarter wave circuit set at the desired noble gas frequency is connected to the coil to serve as an TR switch (Figure 3.9). If one end of the quarter wave (1/4) circuits is open or closed, then the other end of the circuit is being inverted to closed or open. At a 50Q load on one end, the circuit will also see a 50K at the other end. L Output Input Figure 3.9: The quarter-wave circuit also called the pi circuit. The values of inductance and capacitance depend on the Larmor frequency of the element. When the RF power is transmitted into the coil, the circuit path is closed at one end of the quarter wave, and the circuit connected to the pre-amplifier at the other end is open. When the circuit is in the receive mode, the quarter wave sees 50K2 at the coil end. This results in the transmission of the NMR signal through the k/4 circuit to the first stage MITEQ pre-amplifier. The NMR signal is then filtered through a band-pass filter at 33 the desired noble gas nuclear frequency before it is sent to the pre-amplifier (Appendix A). The band-pass filter serves an important role in the circuit by eliminating noise that is added to the system by other sources, and allowing only the desired frequency to pass through. In the circuit, a passive filter is preferred over an active filter. A passive filter contains only passive elements such as capacitors and inductors. However an active filter needs to be activated with a power source. In general, the noise level of the power source is as high as the acquired NMR signal, making the active filter undesirable for use in the circuit. 3.5 The Hyperpolarizing system for Noble Gases The 3 He and 129Xe gas flow-through system, also known as the hyperpolarizing system, was designed with a 150 Gauss portable HP-MRI in mind for the future. The entire system was built on a cart with pneumatic casters for accessibility [14]. The hyperpolarizing system takes mixed gas from a compressed gas tank and delivers it into a purifying manifold via " stainless steel tubing. After purification, the elutriated gas flows into the hyperpolarization chamber to be polarized by a laser and is then collected in a glass cell for final use. When the experiment is ready to proceed, the collected polarized and purified gas is transferred in a Tedlar bag for human inhalation, or to a gas delivery system for animal use (Figure 3.10). 34 Gas m ure _Manifold! ufcon mixtuepuifictionchamber Glassware/ hyperpolarizing Hyperpolarized gas Figure 3.10: Block diagram of the Hyperpolarizing System 3.5.1 Xenon Gas Hyperpolarizing System In the case of the 129Xe gas flow-through system, the gas mixture consists of 1% xenon, 98% helium and 1% other impurities. The helium in the gas mixture acts as a catalyst to increase the collision rate between rubidium and xenon during optical pumping. The gas mixture flows into the purifying manifold and then enters the hyperpolarizing chamber, where gas is stored in a glass cell that is coated with rubidium alkali-metal. When the cell is heated to 85C in a glass oven, the rubidium vaporizes in the cell and mixes with the 129Xe gas. Then a laser is shined on the cell to induce the two gases to shift in their energy states and begin the optical pumping process (Appendix B). After the gas mixture is polarized, it flows into capture/transfer glass cells to separate the pure hyperpolarized 129Xe gas. The first transfer glass cell uses cold water to condense and capture all of the rubidium and helium atoms, and only allows hyperpolarized 35 129 Xe to move to the second transfer glass cell. The pure hyperpolarized ' 29Xe is condensed and collected by using liquid nitrogen to freeze the cell. (Figure 3.11) This method is able to keep the polarization of the gas for days. When the gas is needed, the liquid nitrogen is removed and the polarized gas starts to flow into the Tedlar bag or the animal delivery system. Hyperpolarization Chamber To Tedlar Oven First Transfer/Capture Glass-ware Second Transfer/Capture Glass-ware Figure 3.11: 12 9Xe Hyperpolarization chamber and trapping glassware used to purify hyperpolarized xenon from helium and rubidium and to store the polarized gas for long periods of time. 3.5.2 Helium Hyperpolarizing System The method used to polarize 3He follows four steps similar to that used to polarize 129 Xe (figure 3.13) except with slight differences. The helium mixture contains 99% helium, and 1 % impurities. The purifying manifold eliminates most impurities that are introduced during the manufacturing process. Unlike the 3He 129Xe polarization process, the system does not require the transfer/capture glass-cell. The polarized gas does not need to be frozen or separated from other molecules. Instead, the only glass cell that is 36 needed is the inside of the oven where gas can be stored and polarized by the laser. The optical pumping process is the same as the ones described for the I29Xe system. (Appendix B) The 3He must be pumped until the experiment is ready to proceed, when it can be transferred into a Tedlar bag. 3.6 Animal Gas Delivery System The animal gas delivery system simulates the gas breathing process for the Dawley rats. Computer controls the ventilation, gas injection and triggers scanner through LabVIEW software. The SAR-830 series ventilator regulates the gas that go in and out of the animal by connected to a series of tubes and valves that adjoined the hyperpolarizing system and the animal [9]. (Figure 3.12) SAR-830 serdes Ventilator PC with TimIng Board & Labvlew PC conls %VnNadon,Gas Nnjbon 0-d TRACHEAL PRECISION Frigae prue s N plu " RESTRICTOR TB PNEUMAIC UAT Figure 3.12: Animal Gas Delivery System 37 The gas is stored in a Tedlar bag contained in a rigid acrylic box. By pressuring the box, the gas flows out of the bag and enters the breathing circuit. The gas injection to the rat is synchronized to the computer controlled animal ventilation which connected to animal's airway through a set of non-metallic valves. The animal ventilator gates a controlled gas flow into the animal during the inspiration phase of the respiration. The airflow is set by a rotameter-type regulator, and the timing is controlled by the LabVIEW software program. Expiration is passive and occurs when the animal's airway is opened to the atmosphere via the solenoid valves. 38 Chapter 4 Coil Development in the Very Low-field From the MRI theory, the Larmor frequency is proportional to the external magnetic field strength. In the 150 Gauss Very Low-field, the Larmor frequencies for the different elements' nuclear spins are 634.87 KHz for protons, 483.64 KHz for 3He, and 175.60 KHz for 129Xe. In order to achieve resonant frequencies lower than 1 MHz and to generate a large signal to noise ratio (SNR), the RF coils need to have a high quality factor, Q. QoL The Q factor for the coil is describe by the relationship: (Eqn 4.1) R where R is the loss in the coil, coo is the Larmor frequency, and L is the inductance of the coil. In order to have a large Q, wo is given for 150 Gauss, and for a small R, L has to be large. The only way to achieve large inductance, multiple loops of litz wires are used to build the RF coil. Litz wire is proven to be better than copper wire by Y.J. Yang at Korea University. In his study, it showed that Litz wire gives approximately 2.6dB (30%) better 39 SNR for frequencies around 200 KHz, but at 1 MHz, the difference between two wires is only 0.5dB (5%)[15]. After deciding what type of wire to use for building the RF coil, the next major is decision lies in the design of that coil that should be used for lung imaging. Surface coils are commonly used in MRI to obtain excellent high-resolution images. The surface coil design, which is based on the reception field, is only coupled to the region of interest and served to decrease the coupling to the environment which reduces the noise level. Surface coil is ideal for body surface imaging, such as the spine, but the lungs are set deep in the chest cavity and therefore difficult to image with the surface coil. To solve this problem, a pair of surface coils known as Helmholtz coils are used to create a homogeneous magnetic field in the medial plane of the body where the lungs are located (Figure 4.1). Transmit -~- ---- Solenoid Coil Receive Figure 4.1: Helmholtz Coil 40 The first step in building a Helmholtz coil is to construct two surface coils which are matched and tuned to resonate at the desired Larmor frequency. When a coil is matched, it means the input impedance is 50Q, so no power is reflected back to the coil. After completing the task of tuning and matching the two individual surface coils, the final Helmholtz coil can be build upon the two surface coils. The following sections will discuss the details of coil building and make comparisons between different coils. 4.1 Basic Coil Circuit Components Calculation To build an effective coil, it is important to choose the correct value for the capacitors in the coil circuit. In this section, circuit analyses will be used to calculate the values for the capacitors [16]. L02 C Figure 4.2:Circuit Diagram of the coil network 41 Figure 4.2 represents the general structure of surface coil circuit. L is the coil inductance, C, is the lumped capacitance to establish the input port, C2 is the total distributed capacitance, R is total coil loss, and Zi, is the input impedance of the circuit. From circuit analysis, Zi,, can be represented by: 1 R+ Zi = II_(1-- 2 LC 2 ) (Eqn 4.2). COC 2 1+ - -oLC, + jwRC, C2 At the desired resonance frequency, wo, 00 = 1 __ 4.3), IC(Eqn LCC2 C1 + C2 where the reactance of L and the sum of C1 + C2 are equal. Therefore the input impedance can be reduced to: Zn(cw = - 1 (wOC 1) 2 R + 1 jo 0 C, (Eqn 4.4). A simple circuit of a series of capacitors and resistors (Figure 4.3) can model this equation. 42 zin Figure 4.3: Simple Circuit model of coil After finding the capacitance, the tuned coil needs to be matched to 50Q, RO. When a coil is tuned to the Larmor frequency, the reactance of the circuit tends to be more capacitive, and the resistance of the coil depends on the values of the capacitors. The next step is to replace Zi, with the desired 50E2 impedance, and to solve equation 4.4 for the value of C, and to solve equation 4.2 for the C2 value. R CI = OJ~ L (Eqn 4.5), and 1-R (Eqn 4.6). C2 = 0z L 43 4.2 Method of Coil building Before the values of capacitors can be calculated, there is a systematic way of finding the resistance of the coil and then using the theoretical capacitor values to tune and match the coil. However, in practice, the theoretical value is only close to the actual value, and therefore the procedure of coil building is written below: Step One: Measure the inductance of the coil by connecting it across the reflection port of the network analyzer. Set the analyzer to the sl measurement and read the inductance value from the Smith chart at the resonate frequency of interest. Step Two: After obtaining the inductance value, use the relationship c>=2rf=1/V(L*C)to find the resonant capacitance value wheref is the Larmor frequency, L is the inductance, and C is the capacitance. Place the capacitor in parallel with the coil to obtain the desired resonant frequency. Use a s21 measurement on the network analyzer to tune the coil to the Larmor frequency. If the resonant frequency does not match withf, then adjust the capacitor value untilf is obtained. Step Three: Obtain quality (Q) factor experimentally, and using the equation Q=(w)*L)/R to calculate R, the resistance of coil. In order to measure the Q factor atf, first use the s21 measurement to capture the curve which represents the coil behavior. On the analyzer display, Q represents the 3dB bandwidth of the coil in the s21 measurement. Setf as the reference point by fixing the marker to be zero at that point. Then, use the bandwidth function to find the 3dB points on the curve. The the right corner of the screen. Then use the 44 Q value is Q value to calculate R. displayed on Step Four: Use the equations, Cparaiei= V(R/50)/(dw*L) and Cseries = (I-V(R/50))/(d*L), which were derived in section 4.1, to calculate theoretical values for capacitors in parallel and series to the coil. Place those capacitors in the circuit and obtain the resonant frequency from the sIl measurement. Step Five: If the measured frequency does not equal tof, then adjust the capacitor value by adding or removing capacitors, and by adding adjustable capacitors into the circuit until f is observed. Use the s 11 measurement and the smith chart to perform the final tuning and matching of the circuit. Step Six: After the RF coil is tuned and matched on the work bench, the final adjustments can be made in the simulator (the big shield) if the coil is built for an animal experiment. Otherwise, for the human experiment, the final adjustments have to be made inside of the MRI scanner. The simulator emulates the interior environment of MRI scanner, which couple to the coil and shield all other RF sources in that room that would have affected the tuning and matching processes. Step Seven: For a human sized coil, an extra set of capacitors is added between the coil and the ground to stop the oscillation in the coil. This procedure should be followed to build the first two individual surface coils, which will be used to build the Helmholtz coil. Wind both coils in the same direction, and make two leads at the end. Place the coils on top of each other, then solder the bottom lead of the top coil with the top lead of the bottom coil. After forming this connection, place a stick between the two coils to simulate the gap that will be created by the human chest area. During the experiment, one of the coils will be placed on top of 45 the chest, and the other on the back of the chest. The lungs are positioned in the middle of the Helmholtz coil to experience the homogeneous field that is created by the coil. After setting up the Helmholtz coil on the workbench, follow the procedure again to tune and match for the Helmholtz coil. However, when a human is placed between the coils, the frequency and the impedance can change dramatically. It is necessary to actually place the person between the coils and put him/her inside of the magnet to readjust the capacitors' value until the Larmor frequency is achieved. 46 Chapter 5 Results and Discussion 5.1 Coil Analysis in the Very Low-field In the study of Very Low-field HP-MRI spectroscopy, coil design plays an important role in obtaining a high signal to noise ratio during the imaging process. Analyses of different types of surface coils and different number of turns that optimize the coil efficiency were conducted to help determine which coil design should be use for the human lung imaging experiment. 5.1.1 Surface Coil Optimization with Number of Turns In order to understand how the number of loops of wire is affecting the properties of the coil, 1-turn, 5-turn, 10-turn, 20 turn and 30-turn human size (10.5inches in diameter) surface coils were built. To keep the consistency of the experiment and the convenience of testing the SNR in the scanner, these coils are tuned and matched to the 47 proton frequency, 634.87 KHz, instead of to the frequency of 3He or I 29Xe. The proton sample does not need to be hyperpolarized, therefore the magnetization is constant for every experiment as long as the same sample is used. If 3He or 129Xe is used, the polarization depends on the quality of the optical pumping, which means that the magnetization varies from experiment to experiment. Therefore an additional unknown, other than the number of loops used for the surface coil, affects the SNR. In the table 5.1, inductance, quality factor, calculated capacitance, measured capacitance, and resistance of the surface coil network at proton frequency are given. Coil Number Number of 1 turn 2 5 turns 3 10 turns 4 20 turns 5 30 turns 1.35pH 19.43 gH 80 pH 266 pH 570 pH 1 Turns Inductance (L) _ Quality 6 10 31 205 308 40.3 nF 1.25 nF 429 pF 160 pF 114 pF 6.2 nF 2.0 nF 356 pF 76 pF 52 pF 50 nF 2.953 nF 710 pF 200 pF 57 pF 11.5 nF 0.76 nF 820 pF 100 pF 39 pF N/A N/A 802 pF 180 pF N/A 0.90 Q 7.75 Q 10.29 Q 5.176 Q 7.38 £2 Factor (Q) Calculated Cparallel Calculated Cseries Actual Cparallel Actual Cseries1 Actual Cseries2 Resistance of Coil Network Table 5.1: Comparison of surface coils with different turns Figure 5.1 shows the values of inductance verses the number of turns for all surface coils. The relationship between inductance and the number of turns follows the physics equation: 48 i (Eqn. 5.1), where L is the inductance, N is the number of turns, 0P is the flux, and i is the current. Inductance is positively proportional to the number of turns in a coil. An increase L will lead to an increase in inductance. Figure 5.1 Human Size Proton Coil Inductance vs. Number of Turns 600500- £ 0 400- C.) E a)C.) Cu 300200- C.) V C 1000- - 0 10 20 30 40 Number of Turns in Coil It is necessary to determine which number of turns creates the most affective coil from the parameters that we measured and calculated. An efficient coil should have low resistance (R) and a large inductance value to produce a large Q value. In figure 5.2, the Q and R values are normalized to 1 individually, with 1 being the most desired value. To normalize Q, the larger the quality factor, the closer it is to 1. In normalizing R, the smaller the resistance, the closer it is to 1. In the graph, two normalized values are stacked on top of each other for evaluation. Coil #5 (30 turns) has the highest normalized 49 value, whereas coil #4 (20-turns) is more evenly distributed between the normalized Q and R. It is apparent that coil #5 and coil #4 are better than coils #1, #2, and #3. Figure 5.2: Normalized R and Q vs. Surface Coils 1.4 1.2 - 0 cc G) N E 0 z 1 - 0.8 -- - MNormalized R Normalized Q 0.60.40.202 4 3 5 Coil Number To test the efficiency of the surface coils, a water phantom was used to obtain FID, from which SNR can be calculated. In figure 5.3 we can see that the 1-turn surface coil gives a relatively small signal, which has been distorted by background. The FIDs of 5-turn and 10-turn coils look much clearer than the FID obtained by thel-turn coil, but still an enormous amount of background noise is riding on the signals. The FIDs obtained with the 20-turn and 30-turn coils are sufficient for MR imaging. However, the 20-turn coil has the highest SNR and lowest noise riding on the FID. 50 Figure 5.3 FID of Water Phantom using Human Size proton Surface Coils with different turns IJ - N E-- -- 5-turns, SNR: 22, File: 991030-18 1-turn, SNR: 7.50, File: 991030-24 -M - -E -l~ - 20-turns, SNR: 129, File: 000320-01 10-turns, SNR: 21.7, File: 991030-13 I - -ii -s 30-turns, SNR: 41.9, File: 000320-04 51 H 5.1.2 Optimizing Coil Design for Lung Imaging After determining the optimal number of turns to be used for a coil, the next step was to obtain the best shape for lung imaging. In the previous section, it was mentioned that coil design is unique to different part of the human body. In the case of lung imaging, the possible coil designs are surface, planar, and Helmholtz coils (Figure 5.4). The surface coil's wires are wound on top of each other and it looks like a short solenoid with a large diameter. The planar coil has its wires wound spirally on a flat surface. The Helmholtz coil is two surface coils on top of each other. Surface coil Planar coil Helmholtz coil Figure 5.4: Possible lung coil designs in Very Low-field The qualities of these three coils were tested with a water phantom. FIDs were obtained to compare the SNR difference between the coils (Figure 5.5). All FIDs were the results of 90-degree flip angle pulses with the shim turned on to fine tune the homogeneity of the magnetic field. The 90-degree flip angle refers to the RF power that is required to preccess the charged particles that spin along the Z-axis to X-axis plane. 52 By comparing the measured FIDs and calculated SNRs, it was found that the Helmholtz coil obtained a sufficiently high signal that noise could not be detected beyond of the FID. Besides the high signal to noise ratio, the FID obtained by Helmholtz coil decayed much more slowly and uniformly than the ones obtained by the planar and surface coils. The slow and uniform decay of FID indicated that it had a long relaxation time and hence a long lasting signal which can give a clear image. 5.1.3 Conclusion on Optimizing Lung Coils The experimental and theoretical results show that a 20-turn Helmholtz design is the most desired coil to be used for lung imaging in the Very Low-field. The theoretical calculation has shown that it is better to have coils which can create homogeneous magnetic fields in the region of imaging. Helmholtz coil satisfies this criterion conceptually and experimentally. Through the FID tests that I conducted, the numbers imply that the Helmholtz coil is the choice for lung imaging in the Very Low-field. Apart from the shape of the coil, the number of turns of the coil was proven to play an important role in SNR. The inductance (L) of the coil increases with the number of turns. The total loss of the coil network (r) is inversely proportional to the quality factor (Q). A coil with large Q, large L, and small r will have a high SNR. In our case, the bench test showed that a 30-turn coil has a large Q, but its total network loss is greater than that of a 20-turn coil. However, the FIDs that were obtained using these two coils showed that a 20-turn coil has a SNR of 129, and a 30-turn coil a SNR of 41.9. By comparing FID results the 20-turn coil was chosen for the Helmholtz design. 53 surface coils Figure 5.5 FIDs of Helmholtz, planar, and Helmholtz coil, SNR: 363, File 000421-02 i HMM EMEoE Eim Planar coil, SNR: 87, File 000312-01 Surface coil, SNR: 134, File 000227-88 54 5.2 Phantom Images in Very Low-field Proton FID was used as a reference for Very Low-field Lung imaging. Before the experiments with the hyperpolarized gas MR lung imaging was carried out, a water phantom was imaged with the Helmholtz coil to prove the proper functioning of the coil. (Figure 5.6) After successfully imaged the water phantom, a helium phantom was imaged with the same coil (Figure5.7). These tests concluded that both the coil and polarization of helium gas are ready for human lung imaging. Figure 5.6: Spin echo water phantom image with Helmholtz Coil, Dimension of the coil: 10.5"in diameter, 10" in height Dimension of sample: 6" in diameter Figure 5.7: Gradient echo helium phantom image. The bright spot in the center is from the cell's Rubidium pull off finger. Dimension of Coil: 10.5" in diameter Dimension of Sample: 30 mm in diameter 55 5.3 Very Low-field Lung Images of Rat in vivo While the human sized lung coil was being designed and constructed, the HP-MR lung imaging was carried out on a rat in vivo with 12 9Xe and 3He gas. After obtaining the very first image used 129Xe gas, He gas was also used for imaging. (Figure 5.8, 5.9) Figure 5.8: 12 9Xe lung image of rat in vivo Figure 5.9: 3He lung image of rat in vivo 56 By comparing the two images in figure 5.7 and 5.8, it can be seen that the image obtained by using 3He gas has a higher signal level than the one obtained by 12 9Xe gas. This phenomenon can be due to three main factors. First, the hyperpolarization process of 3He is much better understood then 129 Xe in our lab, even though both processes have not achieved the maximum polarization. Another factor is that the gyromagnetic ratio for 129 Xe is 2.8 times smaller than 3He, therefore even with maximum polarization, 129Xe polarization would still be lower than 3He by a factor of 2-4. Finally, the abundance of 129Xe isotope is only 26% of xenon at natural abundance, where 3He is at 100% isotopic enrichment. By observing the comparison between 129Xe and 3 He produced images, 3He was chosen to be used for human lung HP-MRI for the initial run. 5.4 Very Low-field Human Lung HP-MRI 5.4.1 Background Noise In conventional MRI, the two significant sources of noise are the biological sample and the coil network. Coil noise is proportional to o>P up to 1MHz, since we are operating in the Very Low-field regime, this effect can be ignored. The sample noise is proportional to o, therefore it would have more of an effect in high-field than the lowfield situations [2]. Therefore, sample noise dominates in the high-field, and coil network noise dominates in the low-field. The major difficulty that we encountered during HP human lung imaging that did not occur during animal experiments was the background noise picked up by the RF coil. The background noise increased with the size of the coil. In comparison, the size of the 57 coils used for humans is much larger than the one used for animals. Coil shielding, which was used in the animal experiments to reduce noise, was not used in the human experiment due to the coupling effect. The strategy we used to reduce background noise was to eliminate paths which might introduce noise into the HP-MRI hardware system. Figure 5.10 (a) shows a basic noise level test with large noise peaks. Often, the signal level of noise induced from the TX line is much larger than the NMR signal obtained from the lungs. The hardware connection for the test was the RF source connected to the RF port, 50Q connected to the coil port, and receive port connected to the pre-amplifier (Figure 5.10 (b)). In order to eliminate the noise peaks, a pair of cross diodes was interfaced between the transmission (TX) line and the TR switch (Figure 5.11 (b)). This component blocks all noise from the transmission line during the receiving mode. Figure 5.11(a) shows the cross diodes have reduced the noise peak from the background noise. Figure 5.10 (a): Basic noise data with noise peaks 58 Pre-Amp Receiver I RF Source T/R Switch Coil port - 500 Resistor Figure 5.10 (b): Hardware setup for the basic noise level test V Figure 5.11 (a): Basic noise data after cross diodes reduced noise peaks Pre-Amp Receiver R Coil port RF Source T/R Switch I 0 Resistor Figure 5.11 (b): The hardware setup with a pair of cross diodes 59 Other than adding the cross diodes, all active components such as gradient amplifier, MITEQ amplifier and the Resonance Pre-amplifier were grounded together to the superconducting magnet. Connecting all grounds together generated a common reference point for the system which reduced noise entering the system. Comparing part (a) and (b) of Figure5.12, the noise peaks in the background noise were reduced by connecting all electronics to the same ground. 1.1' 1 .. .. 1 111.1111 1 III.I j ,11. IJ I.1.1 fill 11.11 1 11 'fl (a) (b) Figure 5.12: (a) Basic noise data without connecting all active electronics to the same ground point. (b) Basic noise data after connecting all active electroinics to the same ground point. 60 Finally, with Dileep's help, a passive low-pass noise filter was built for the gradient amplifiers to reduce noise that was generated by the power source. (Figure 5.13) The noise source produced by the gradient amplifier did not affect the SNR for the animal coil, but distorted signal level for the human coil. (Figure 5.14 (a)) Background noise was measured again after altering the low-pass filter. Figure 5.14 (b) shows the elimination of the noise peaks. Gradient Amplifier L CGradient Coil Figure 5.13: Circuit diagram of low-pass filter. It is used to eliminate noise that was generated by the gradient amplifier. (b) (a) Figure 5.14: (a) Basic noise data before using gradient amplifier noise filter. (b) Basic noise data after using gradient amplifier noise filter. 61 5.4.2 Human lung image In HP-MRI, since the magnetization of the atom is set by its polarization, which is 2 independent of the external field, the signal scales with BO instead of B0 at low field and scales with BO at high field. The ratio between SNR(HP)/SNR(water) scales with 1/B, therefore HP-MRI has significant advantage over conventional MRI at Very Low-field. Since the noise factor of HP-MRI only depends on the coil network, this problem can be reduced by using low loss superconducting RF coils with unloaded Q factor greater than 50,000 [3]. However, this technique does not help conventional MRI, because the sample noise dominates [18]. Figure 5.15: HP-MRI human lung image From the theoretical point of view, HP-MRI is a certain possibility in the Very Low-Field, but no one has proved it experimentally. After testing the coil and the hyperpolarizing system, and reducing background noise, the very first HP-MRI human 62 lung image was taken with Mitch Albert as the subject. (Figure 5.15) In the image, the lobes of the lung can be seen clearly, however, the 10" in diameter coil is not big enough to cover the apex and the base of the lung. The successful imaging in the Very Low-field shows its possible clinical use for people with lung disease in the future. 5.5 Summary A 20-turn Helmholtz coil was built for obtaining human lung images. To design the coil, the number of turns was optimized to 20-turns. Among surface, planar, and Helmholtz coil shapes, Helmholtz was chosen to be used for its high SNR. Besides having an efficient human lung coil, noise sources were eliminated from the HP-MRI hardware system. A set of cross diodes were placed between the T/R switch and the transmitting line to reduce the noise entering the system during the receiving mode. All active electronics were connected to the same ground reference point to reduce the oscillation between each connection. Final alteration of the system was done by adding a low-pass noise filter between the gradient amplifier and the gradient coil. The filter reduced the noise in the same range as the Helium Larmor frequency. 5.6 Recommendations to Future Work Further work needs to be carried out to improve the Very Low-field HP-MRI system to obtain higher quality images. The 10" diameter Helmholtz coil is not large enough to image the full length of the lungs. Figure 5.14 depicts a human lung image, 63 where it is can be seen that the base of the lungs is missing. The use of a larger Helmholtz coil may allow more complete lung imaging. A circular shaped coil may not be the best design, therefore it is necessary to investigate the utility of other shapes, such as that of a rectangular coil. The experiments described used a 30-turn coil, which may not be optimal. Further studies should be conducted to determine the ideal number of turns. In order to understand how inductance, quality factors, and loss resistance in the coil network affect the SNR, 40-turn, 60-turn and 100-turn coils should be studied. When a new Very Low-field HP-MRI system is assembled, all active electronics should be placed close to the scanning magnet. A shielding system worked effectively in the animal experiments and should be considered for the human studies. Coil shielding was attempted for human experiments, but it coupled to the coil greatly so that the SNR was reduced. However, increasing the distance between the coil and the shielding might cause an augmentation in the SNR. 64 (i- ( -l AAA CL X, 2~- *4 -1- 10 ---- I -3 ) 1 F- 1 1 i 4 I i VEER a e imnt E'd 7 -1J-1 II - -A DiC 5 CC-L;TM Dl 16 "D2 O-=R -g,) -., Appendix B: Optical Pumping The figure shows the optical pumping setup and the procedure is described below. Heater Laser Temperature Control Box Gas Cell Ga Cell Glass Oven l.Setup the 3He or 129Xe cell and place it in the glass oven. 2.Aligned the laser beam and the glass oven with the external magnetic field. 3.Turn on the Omega Temperature Control Box and make sure the heater sensor is taped to the cell. Set the temperature to 150C for 3He, and 1 10C for 12 9Xe. 4.Turn on the air compressor then the Variac heater. (WEAR EYE PROTECTION) 5.When the cell temperature reaches 80C, turn on four channels of laser beams on. 3He CELL: WAIT FOR 2 HOURS 129Xe CELL: WAIT FOR 1.5 HOURS 6.After the waiting period, turn off the Variac heater. When the temperature stops dropping, then turn off two channels of laser beams. 7.When the temperature reaches 80C, turn off laser, and put the hyperpolarized cell inside of cold water. 8.The hyperpolarized cell is ready to be used. 66 References 1. Albert, M. et al.. Biological magnetic resonance imaging using laser-polarized I2 9 Xe. Nature Vol.370, 199-201 (1994). 2. Hoult, D.I. & Richards, R.E.. The signal-to-noise of the nuclear magnetic resonance experiment. J. Magn. Reson. Vol.24, 71-85 (1976). 3. Black, R.D. et al.. A high-temperature superconducting receiver for nuclear magnetic resonance microscopy. Science Vol.259, 793-795 (1993). 4. De Lange, E. E. et al.. Lung Air Spaces: MR Imaging Evaluation with Hyperpolarized 3He Gas. Radiology Vol.210, 851-857 (1999). 5. Venkatesh, A.K. et al.. MRI of the Lung Gas-Space at Very Low-Field Using Hyperpolarized Noble Gases. (2000). 6. Hashemi R.H. and Bradley W. G.. MRI The Basics. Williams & Wilkins, Philadelphia (1997) 7. Chen, C.N. and Hoult D. I.. Biomedical Magnetic Resonance Technology. Adam Hilger, Bristol and New York, 1989. 8. Rinck, P.A., Muller, R. N., and Petersen, S. B.. An Introduction to magnetic Resonance in Medicine, Thieme Medical Publishers, Inc. New York, 1990. 9. Ramirez, M.P. et al.. Physiological response of rats to delivery of helium and xenon: implications for hyperpolarized noble gas imaging. NMR in Biomedicine, Vol 13 253-264, 2000. 10. Albert, M. S. , Balamore, D., and Kornhauser, S. H.. Magnetic Resonance Imaging Using Hyperpolarized Xe. American Journal of Electromedicine, 72-80, 1994. 11. Albert, M. S. and Balamore, D. (1998). Development of Hyperpolarized Noble Gas MRI. Nuclear Instruments & methods in Physics Research, 441-453. 12. Chupp, T. and Swanson, S.. Medical Imaging with Laser Polarized Noble Gases. Unknow publisher, 2000. 13. Kacher, Dan. Thesis: Nuclear Magnetic Resonance measurements of the 12 9 Xe spinlattice relaxation time constant in human blood. Boston University, 1998. 67 14. Mansour, J.K.. Thesis: The Production of Hyperpolarization Gas for use in MRI Imaging. Boston Univerysity, 2000. 15. Yang, Y.J. et. al.. Low field RF Coil for Hyper-Polarized Noble Gas MRI. Conference Slices, Korea University, 1999. 16. Hashoian, R. S. Coil Interface Theory and Implementation. Conference Slices, 1-33. 17. Ocali, 0. and Atalar, E.. Ultimate Intrinsic Signal-to-Noise Ratio in MRI. MRM Vol.39, 462-473, 1998. 18. Edelstein, W.A. et. al.. The Intrinsic Signal-to-Noise Ratio in NMR Imaging. Magnetic Resonance in Medicine Vol 3, 604-618, 1986. 68